Noninvasive Glucose Monitoring: Comprehensive Clinical Results
The DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
CONTEXT: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management. OBJECTIVE: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements. DESIGN: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study. SETTING: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States. PARTICIPANTS: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin. MAIN OUTCOME MEASURES: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid. RESULTS: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis. CONCLUSION: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“One of the first commercial wearable sensing devices, a wrist-worn glucose monitor, used iontophoresis to extract ISF for analysis77.”
From End-to-end design of wearable sensors (2022) · cites this paper as [77]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Turner, R. C., Fox, C., Matthews, McElroy, H. H., Cull, C. A., Holman, R. R., Neil, P., Hadden, D. R., Wright, D., Manley, E., et al. (1998) · The Lancet · Cited by 8,419
Ohkubo, Y., Kishikawa, H., Araki, E., Miyata, T., Isami, S., Motoyoshi, S., Kojima, Y., Furuyoshi, N., & Shichiri, M. (1995) · Diabetes Research and Clinical Practice · doi:10.1016/0168-8227(95)01064-k · Cited by 3,079
DRAIZE, J. H., WOODARD, G., & CALVERY, H. O. (1944) · The Journal of Pharmacology and Experimental Therapeutics · doi:10.1016/s0022-3565(25)08751-8 · Cited by 2,602
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Measurement of glucose in diabetic subjects using noninvasive transdermal extraction
Tamada, J. A., Bohannon, N. J. V., & Potts, R. O. (1995) · Nature Medicine · doi:10.1038/nm1195-1198 · Cited by 182
Self-citationReverse Iontophoresis: Noninvasive Glucose Monitoring in Vivo in Humans
Rao, G. S., Guy, R. H., Glikfeld, P., LaCourse, W. R., Leung, L., Tamada, J. A., Potts, R. O., & Azimi, N. (1995) · Pharmaceutical Research · doi:10.1023/a:1016271301814 · Cited by 138
Self-citationMeyerhoff, C., Bischof, F., Sternberg, F., Zier, H., & Pfeiffer, E. F. (1992) · Diabetologia · doi:10.1007/bf02221686 · Cited by 126
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119
Cited by
Prausnitz, M. R., & Langer, R. (2008) · Nature Biotechnology · doi:10.1038/nbt.1504 · Cited by 3,517
Current status and future potential of transdermal drug delivery
Prausnitz, M. R., Mitragotri, S., & Langer, R. (2004) · Nature Reviews Drug Discovery · doi:10.1038/nrd1304 · Cited by 1,462
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306
Transdermal drug delivery: overcoming the skin’s barrier function
Naik, A., Kalia, Y. N., & Guy, R. H. (2000) · Pharmaceutical Science & Technology Today · doi:10.1016/s1461-5347(00)00295-9 · Cited by 720
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660
Fluorescence-based glucose sensors
Pickup, J. C., Hussain, F., Evans, N. D., Rolinski, O. J., & Birch, D. J. S. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.10.002 · Cited by 616
Correlation Between Sweat Glucose and Blood Glucose in Subjects with Diabetes
Moyer, J., Wilson, D., Finkelshtein, I., Wong, B., & Potts, R. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0262 · Cited by 570
“To show correlation between SG and BG it is necessary to have a reasonable change in BG.”
Self-citationBoyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390
Self-citationBiosensors: the new wave in cancer diagnosis
Mousa, S. (2010) · Nanotechnology, Science and Applications · doi:10.2147/nsa.s13465 · Cited by 381
Sampling interstitial fluid from human skin using a microneedle patch
Samant, P. P., Niedzwiecki, M. M., Raviele, N., Tran, V., Mena-Lapaix, J., Walker, D. I., Felner, E. I., Jones, D. P., Miller, G. W., & Prausnitz, M. R. (2020) · Science Translational Medicine · doi:10.1126/scitranslmed.aaw0285 · Cited by 379
Yonzon, C. R., Haynes, C. L., Zhang, X., Walsh, & Van Duyne, R. P. (2004) · Analytical Chemistry · doi:10.1021/ac035134k · Cited by 376
Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 374
Toward Flexible and Wearable Human‐Interactive Health‐Monitoring Devices
Takei, K., Honda, W., Harada, S., Arie, T., & Akita, S. (2015) · Advanced Healthcare Materials · doi:10.1002/adhm.201400546 · Cited by 371
Lipani, L., Dupont, B. G. R., Doungmene, F., Marken, F., Tyrrell, R. M., Guy, R. H., & Ilie, A. (2018) · Nature Nanotechnology · doi:10.1038/s41565-018-0112-4 · Cited by 340
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288
Tura, A., Maran, A., & Pacini, G. (2007) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2006.10.027 · Cited by 283
“Participants wore up to two GlucoWatch during the 15-h study session and performed two fingersticks per hour for comparative blood glucose measurements [53].”
MethodPhysiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 251
“IGS(T Ti Tp 60) IGS(T Ti Tp) IGS(T Ti Tp 60) IGS(T Ti Tp) (4)”
Self-citationFirst human experiments with a novel non-invasive, non-optical continuous glucose monitoring system
Caduff, A., Hirt, E., Feldman, Y., Ali, Z., & Heinemann, L. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00196-9 · Cited by 247
“There is a need for a glucose monitoring system that can provide detailed information on glucose patterns throughout the day (Tamada et al., 1999).”
BackgroundGlucose monitoring by reverse iontophoresis
Potts, R. O., A. Tamada, J., & J. Tierney, M. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.210 · Cited by 243
Self-citationComing of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223
“Minimally invasive, realtime continuous glucose measurement [5, 6] is progressing to accurate, insulin-dosing approved, factory-calibrated systems [7].”
BackgroundCan Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 213
“To-day, almost all minimal and noninvasive glucose detection methods, such as iontophoresis 7,8 , hypodermic needles 9,10 , electrochemical/ enzymatic sensors, 11–15 ultrasound, 16 and optical approaches, 17,18 have focused on estimating glucose in the interstitial fluid (ISF) compartment of the…”
In vivo glucose monitoring: the clinical reality and the promise
Pickup, J. C., Hussain, F., Evans, N. D., & Sachedina, N. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.08.016 · Cited by 203
Wireless on-demand drug delivery
Mirvakili, S. M., & Langer, R. (2021) · Nature Electronics · doi:10.1038/s41928-021-00614-9 · Cited by 202
Trends in Nanomaterial-Based Non-Invasive Diabetes Sensing Technologies
Makaram, P., Owens, D., & Aceros, J. (2014) · Diagnostics · doi:10.3390/diagnostics4020027 · Cited by 200
Delays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300528 · Cited by 198
Continuous Subcutaneous Glucose Monitoring in Children With Type 1 Diabetes
Chase, H. P., Kim, L. M., Owen, S. L., MacKenzie, T. A., Klingensmith, G. J., Murtfeldt, R., & Garg, S. K. (2001) · Pediatrics · doi:10.1542/peds.107.2.222 · Cited by 194
Self-citationKoschwanez, H. E., & Reichert, W. M. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.034 · Cited by 181
Emerging biomedical sensing technologies and their applications
Cote, G. L., Lec, R. M., & Pishko, M. V. (2003) · IEEE Sensors Journal · doi:10.1109/jsen.2003.814656 · Cited by 180
“The type of the deformation accompanying the wave is important because it determines resultantacousticsensingprocesses.Compressionaldeformation is associated with the structural relaxational processes in the medium, while shear is coupled with medium viscoelastic properties and therefore they are…”
BackgroundDevices for overcoming biological barriers: The use of physical forces to disrupt the barriers
Mitragotri, S. (2013) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2012.07.016 · Cited by 174
“nomenon has been used to monitor blood glucose levels in diabetic patients [43].”
MethodIn Vivo Blood Glucose Quantification Using Raman Spectroscopy
Shao, J., Lin, M., Li, Y., Li, X., Liu, J., Liang, J., & Yao, H. (2012) · PLoS ONE · doi:10.1371/journal.pone.0048127 · Cited by 171
The GlucoWatch®biographer: a frequent, automatic and noninvasive glucose monitor
Tierney, M. J., Tamada, J. A., Potts, R. O., Eastman, R. C., Pitzer, K., Ackerman, N. R., & Fermi, S. J. (2000) · Annals of Medicine · doi:10.3109/07853890009002034 · Cited by 154
Self-citationSkin Barrier and Transdermal Drug Delivery
Prausnitz, M. R., Elias, P. M., Franz, T., Schmuth, M., Tsai, J., Menon, G. K., Holleran, W. M., & Feingold, K. R. (2012) · Time to knit · doi:10.1016/b978-0-7020-6275-9.00124-0 · Cited by 143
Carbon-dot-based ratiometric fluorescence glucose biosensor
Cho, M.-J., & Park, S. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.11.055 · Cited by 143
Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring
Goodarzi, M., Sharma, S., Ramón, H., & Saeys, W. (2015) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2014.12.005 · Cited by 142
Reverse iontophoresis for non-invasive transdermal monitoring
Leboulanger, B., Guy, R. H., & Delgado-Charro, M. B. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/3/r01 · Cited by 134
Electrical biochip technology?a tool for microarrays and continuous monitoring
Albers, J., Grunwald, T., Nebling, E., Piechotta, G., & Hintsche, R. (2003) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-003-2192-7 · Cited by 125
Detection of Hypoglycemia With the GlucoWatch Biographer
Pitzer, K. R., Desai, S., Dunn, T., Edelman, S., Jayalakshmi, Y., Kennedy, J., Tamada, J. A., & Potts, R. O. (2001) · Diabetes Care · doi:10.2337/diacare.24.5.881 · Cited by 122
Self-citation*The Diabetes Research in Children Network (DirecNet) Study Group (2004) · Diabetes Care · doi:10.2337/diacare.27.3.722 · Cited by 115
Noninvasive and Minimally Invasive Methods for Transdermal Glucose Monitoring
Sieg, A., Guy, R. H., & Delgado-Charro, M. B. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.174 · Cited by 108
Biosensors for in vivo glucose measurement: can we cross the experimental stage
Abel, P., & von Woedtke, T. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00099-4 · Cited by 104
Advancements in Skin‐Mediated Drug Delivery: Mechanisms, Techniques, and Applications
Lee, D. H., Lim, S., Kwak, S. S., & Kim, J. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202302375 · Cited by 104
Iscoe, K. E., Campbell, J. E., Jamnik, V., Perkins, B. A., & Riddell, M. C. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.627 · Cited by 101
Nonenzymatic continuous glucose monitoring in human whole blood using electrified nanoporous Pt
Park, S., Park, S., Jeong, R.-A., Boo, H., Park, J., Kim, H. C., & Chung, T. D. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2011.10.033 · Cited by 101
Burmeister, J. J., Arnold, M. A., & Small, G. W. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/152091500316683 · Cited by 99
Gušev, M., Poposka, L., Spasevski, G., Kostoska, M., Koteska, B., Simjanoska, M., Ackovska, N., Stojmenski, A., Tasič, J. F., & Trontelj, J. K. (2020) · Journal of Sensors · doi:10.1155/2020/9628281 · Cited by 98
Fluidics and sample handling in clinical chemical analysis
Cunningham, D. D. (2001) · Analytica Chimica Acta · doi:10.1016/s0003-2670(00)01256-3 · Cited by 98
Performance Evaluation of Blood Glucose Monitoring Devices
Chen, E. T., Nichols, J. H., Duh, S.-H., & Hortin, G. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526969 · Cited by 96
In Vivo Near-Infrared Spectroscopy of Rat Skin Tissue with Varying Blood Glucose Levels
Olesberg, J. T., Liu, L.-Z., Zee, V. V., & Arnold, M. A. (2005) · Analytical Chemistry · doi:10.1021/ac051036i · Cited by 94
Tunable Laser Diode System for Noninvasive Blood Glucose Measurements
Olesberg, J. T., Arnold, M. A., Mermelstein, C., Schmitz, J., & Wagner, J. (2005) · Applied Spectroscopy · doi:10.1366/000370205775142485 · Cited by 92
Use of the GlucoWatch Biographer in Children With Type 1 Diabetes
Chase, H. P., Roberts, M. D., Wightman, C., Klingensmith, G., Garg, S. K., Van Wyhe, M., Desai, S., Harper, W., Lopatin, M., Bartkowiak, M., et al. (2003) · Pediatrics · doi:10.1542/peds.111.4.790 · Cited by 81
Self-citationHu, Y., Pan, Z., Bock, M. D., Tan, T. X., Wang, Y., Shi, Y., Yan, N., & Yetisen, A. K. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116542 · Cited by 79
Schiaffini, R., Ciampalini, P., Fierabracci, A., Spera, S., Borrelli, P., Bottazzo, G. F., & Crinò, A. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.309 · Cited by 77
Clinical Opportunities for Continuous Biosensing and Closed-Loop Therapies
Li, J., Liang, J., Laken, S. J., Langer, R., & Traverso, G. (2020) · Trends in Chemistry · doi:10.1016/j.trechm.2020.02.009 · Cited by 77
Active enhancement methods for intra- and transdermal drug delivery
Zorec, B. S., Préat, V., Miklavčič, D., & Pavšelj, N. (2013) · Slovenian Medical Journal · Cited by 76
Closed-Loop Insulin Delivery in Type 1 Diabetes
Thabit, H., & Hovorka, R. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2011.12.003 · Cited by 74
Corrie, S. R., Fernando, G. J. P., Crichton, M. L., Brunck, M. E. G., Anderson, C. D., & Kendall, M. A. F. (2010) · Lab on a Chip · doi:10.1039/c0lc00068j · Cited by 74
Non-invasive assessment of the effects of iontophoresis on human skin in-vivo
Curdy, C., Kalia, Y. N., & Guy, R. H. (2001) · Journal of Pharmacy and Pharmacology · doi:10.1211/0022357011776117 · Cited by 74
Electronic home healthcare: concepts and challenges
Demiris, G. (2004) · International Journal of Electronic Healthcare · doi:10.1504/ijeh.2004.004655 · Cited by 73
Continuous Glucose Monitoring: An Overview of Its Development
Skyler, J. S. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0045 · Cited by 72
Shang, T., Zhang, J. Y., Thomas, A., Arnold, M. A., Vetter, B. N., Heinemann, L., & Klonoff, D. C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007212 · Cited by 71
Modeling of spherical fluorescent glucose microsensor systems: Design of enzymatic smart tattoos
Brown, J. Q., & McShane, M. J. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2005.08.013 · Cited by 71
Management of diabetes using adaptive control
Hovorka, R. (2004) · International Journal of Adaptive Control and Signal Processing · doi:10.1002/acs.851 · Cited by 70
Emerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 67
Self-citationPoint-of-Care Testing in Diabetes Mellitus
Price, C. P. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.185 · Cited by 67
Sensitive sensing of biomarkers in interstitial fluid
Kim, Y., & Prausnitz, M. R. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00679-5 · Cited by 64
Davey, R. J., Low, C., Jones, T. W., & Fournier, P. A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400614 · Cited by 62
Chuang, H., Taylor, E., & Davison, T. W. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504322783378 · Cited by 60
(2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526996 · Cited by 60
Geng, Z., Tang, F., Ding, Y., Li, S., & Wang, X. (2017) · Scientific Reports · doi:10.1038/s41598-017-13018-7 · Cited by 59
Hypoglycemia in diabetes: common, often unrecognized.
Gabriely, I., & Shamoon, H. (2004) · Cleveland Clinic Journal of Medicine · doi:10.3949/ccjm.71.4.335 · Cited by 58
Continuous Glucose Sensing: Future Technology Developments
Brauker, J. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0137 · Cited by 58
Dunn, T. C., Eastman, R. C., & Tamada, J. A. (2004) · Diabetes Care · doi:10.2337/diacare.27.9.2161 · Cited by 55
Self-citationAnalysis: Overcoming Skin's Barrier: The Search for Effective and User-Friendly Drug Delivery
Prausnitz, M. R. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/152091501300209606 · Cited by 55
Transdermal Reverse Iontophoresis of Valproate: A Noninvasive Method for Therapeutic Drug Monitoring
Delgado-Charro, M. B., & Guy, R. H. (2003) · Pharmaceutical Research · doi:10.1023/a:1025730815971 · Cited by 55
Toward continuous glucose monitoring with planar modified biosensors and microdialysis
Ricci, F., Caprio, F., Poscia, A., Valgimigli, F., Messeri, D., Lepori, E., Dall’Oglio, G., Palleschi, G., & Moscone, D. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.08.041 · Cited by 54
Novel micromachined silicon sensor for continuous glucose monitoring
Piechotta, G., Albers, J., & Hintsche, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2005.02.008 · Cited by 53
Reverse Iontophoresis for Noninvasive Glucose Monitoring: The Internal Standard Concept
Sieg, A., Guy, R. H., & Delgado-Charro, M. B. (2003) · Journal of Pharmaceutical Sciences · doi:10.1002/jps.10492 · Cited by 51
Identification of Hypoglycemia and Hyperglycemia in Type 1 Diabetic patients using ECG parameters
Nguyen, L. L., Su, S. W., & Nguyen, H. T. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346525 · Cited by 51
Glucose sensors: toward closed loop insulin delivery
Chia, C. W., & Saudek, C. D. (2004) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2003.12.001 · Cited by 50
Thomas, K. J., Sherman, D. B., Amiss, T. J., Andaluz, S. A., & Pitner, J. B. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.261 · Cited by 49
Carroll, M. F., Burge, M. R., & Schade, D. S. (2003) · Reviews in Endocrine and Metabolic Disorders · doi:10.1023/a:1022990003161 · Cited by 48
Continuous glucose monitoring in children with glycogen storage disease type I
Hershkovitz, E., Rachmel, A., Ben‐Zaken, H., & Phillip, M. (2001) · Journal of Inherited Metabolic Disease · doi:10.1023/a:1013996325720 · Cited by 47
Reverse iontophoresis: a non-invasive technique for measuring blood urea level
Değim, İ. T., Ilbasmiş-Tamer, S., Dündaröz, R., & Oğuz, Y. (2003) · Pediatric Nephrology · doi:10.1007/s00467-003-1217-y · Cited by 47
Psychological aspects of continuous glucose monitoring in pediatric type 1 diabetes
The Diabetes Research in Children Network (DirecNet) Study Group* (2006) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2006.00142.x · Cited by 46
Yamakoshi, Y., Ogawa, M., Yamakoshi, T., Satoh, M., Nogawa, M., Tanaka, S., TAMURA, T., Rolfe, P. J., & Yamakoshi, K. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4352951 · Cited by 45
Reverse iontophoretic monitoring in premature neonates: feasibility and potential
Sekkat, N., Naik, A., Kalia, Y. N., Glikfeld, P., & Guy, R. H. (2002) · Journal of Controlled Release · doi:10.1016/s0168-3659(02)00046-9 · Cited by 44
Self-monitoring of blood glucose in children and teens with diabetes
Bui, H., Perlman, K., & Daneman, D. (2005) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2005.00095.x · Cited by 42
Connolly (2008) · International Journal of Nanomedicine · doi:10.2147/ijn.s1728 · Cited by 42
THEDIABETESRESEARCHINCHILDREN (2004) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2004.03.042 · Cited by 41
Point of Care Testing: Potential for Tracking Disease Management Outcomes
Price, C. P., & John, A. S. (2002) · Elsevier eBooks · doi:10.1016/b978-1-4160-6164-9.00020-2 · Cited by 39
Rizvi, A. A. (2004) · Southern Medical Journal · doi:10.1097/01.smj.0000125108.06314.25 · Cited by 39
Blood Glucose: Measurement in the Point-of-Care Setting
Kiechle, F. L., & Main, R. I. (2000) · Laboratory Medicine · doi:10.1309/4bf1-et6t-wfe3-m7xa · Cited by 39
Use of the GlucoWatch® biographer in children and adolescents with diabetes
Eastman, R. C., Chase, H. P., Buckingham, B. A., Hathout, E. H., Fuller-Byk, L., Leptien, A. D., Wyhe, M. M. V., Davis, T. L., Fermi, S. J., Pechler, H., et al. (2002) · Pediatric Diabetes · doi:10.1034/j.1399-5448.2002.30302.x · Cited by 38
Self-citationMamkin, I., Ten, S., Bhandari, S., & Ramchandani, N. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200520 · Cited by 38
Investigations on Multisensor-Based Noninvasive Blood Glucose Measurement System
Yadav, J., Rani, A., Singh, V., & Mohan Murari, B. (2017) · Journal of Medical Devices · doi:10.1115/1.4036580 · Cited by 38
Continuous glucose monitoring and external insulin pump: towards a subcutaneous closed loop
Hanaire, H. P. (2006) · Diabetes & Metabolism · doi:10.1016/s1262-3636(06)72808-7 · Cited by 38
Garg, S. K., & Aktürk, H. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0142 · Cited by 37
Self-citationHypo- and Hyperglycemic Alarms: Devices and Algorithms
Howsmon, D. P., & Bequette, B. W. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815583507 · Cited by 37
Metadata
- Authors
- Tamada, J. A., Garg, S., Jovanovic, L., Pitzer, K. R., Fermi, S., Potts, R. O., & and the Cygnus Research Team
- Venue
- JAMA
- Publisher
- American Medical Association (AMA)
- Published
- 1999-11-17
- Volume
- 282
- Issue
- 19
- Pages
- 1839
- ISSN
- 0002-9955 (ISSN-L)
- 0098-7484
- OpenAlex
- W2088032108
- Cited by
- 282
- References
- 18
- FWCI
- 11.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.