Materials, Devices, and Systems of On‐Skin Electrodes for Electrophysiological Monitoring and Human–Machine Interfaces
Open accessThe DOI matches its Crossref record; 231 of the 233 references it cites were found in a registry or an index.
Abstract
On‐skin electrodes function as an ideal platform for collecting high‐quality electrophysiological (EP) signals due to their unique characteristics, such as stretchability, conformal interfaces with skin, biocompatibility, and wearable comfort. The past decade has witnessed great advancements in performance optimization and function extension of on‐skin electrodes. With continuous development and great promise for practical applications, on‐skin electrodes are playing an increasingly important role in EP monitoring and human–machine interfaces (HMI). In this review, the latest progress in the development of on‐skin electrodes and their integrated system is summarized. Desirable features of on‐skin electrodes are briefly discussed from the perspective of performances. Then, recent advances in the development of electrode materials, followed by the analysis of strategies and methods to enhance adhesion and breathability of on‐skin electrodes are examined. In addition, representative integrated electrode systems and practical applications of on‐skin electrodes in healthcare monitoring and HMI are introduced in detail. It is concluded with the discussion of key challenges and opportunities for on‐skin electrodes and their integrated systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Metal thin film electrodes. a) On‐skin electrodes with serpentine design. Reproduced with permission.[ 55 ] Copyright 2016, Wiley‐VCH. b) Stretchable electrodes with UL‐Sil and fabrics. Reproduced with permission.[ 56 ] Copyright 2014, Springer Nature. c‐i) Epidermal electrode systems incorporating EMG/ECG sensors and ii) image of the on‐skin electronics integrated onto a tattoo. Reproduced with permission.[ 57 ] Copyright 2011, AAAS. d) A schematic of TCR electrodes and the magnified views of Peano curves (inset). Reproduced with permission.[ 58 ] Copyright 2015, National Academy of Sciences USA. e) Photographs of gold film electrodes with wrinkled structures on the plasticized silk substrate. Reproduced with permission.[ 60 ] Copyright 2018, Wiley‐VCH. f‐i) An illustration of optimized on‐skin electrodes and ii) an SEM image shows the superior conformality of a 5 µm thick membrane. Reproduced with permission.[ 61 ] Copyright 2013, Wiley‐VCH. g‐i) Multifunctional device consisting of EP sensors and ii) an image shows conformal skin‐electrode interfaces during compression. Reproduced with permission.[ 66 ] Copyright 2015, Wiley‐VCH. h) Images of the SCG and ECG sensors with excellent conformal contact with skin. Reproduced with permission.[ 67 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/9100e896ece649828eaacff72284fbe2/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-2.jpg)
![Figure 3 Carbon material electrodes. a) Images of nine CNT/PDMS electrodes. Reproduced with permission.[ 72 ] Copyright 2012, IEEE. b) A schematic illustration of a CNTs/aPDMS ECG electrode. Reproduced with permission.[ 75 ] Copyright 2014, Springer Nature. c) The CNT/graphene electrode patch with the content of CNT and graphene at 9:1. Reproduced with permission.[ 76 ] Copyright 2016, American Chemical Society. d) A photograph of the graphene electronic tattoo containing EP sensors. Reproduced with permission.[ 82 ] Copyright 2017, American Chemical Society. e) rGO‐aPDMS electrodes for monitoring EP signals. Reproduced with permission.[ 85 ] Copyright 2020, Elsevier. f) EP electrodes with laser‐induced porous graphene. Reproduced with permission.[ 86 ] Copyright 2018, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/05f37dbfe3fd43509659163035bd1359/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-3.jpg)
![Figure 4 Electrodes based on Metallic nanomaterials. a‐i) On‐skin electrodes with Cu networked mesh structure (inset) and ii) schematic of Cu nanomesh electrodes. Reproduced with permission.[ 98 ] Copyright 2016, Wiley‐VCH. b‐i) Au nanomesh conductors and ii) the process of the polyvinyl alcohol nanofibers (PVA NFs) dissolution. Reproduced with permission.[ 44 ] Copyright 2017, Springer Nature . c‐i) A cross‐sectional SEM image of convoluted nanonetworks and ii) an image of the NEE intimately coupling with human skin. Reproduced with permission.[ 100 ] Copyright 2019, Wiley‐VCH. d‐i) SEM image of Ag‐PU network conductors, ii) an illustration of the hydrogen bonding, and iii) porous Ag‐PU electrodes on skin. Reproduced with permission.[ 101 ] Copyright 2019, Wiley‐VCH. e‐i) SEM image of the Ag flakes and Ag nanoparticle and ii) the ECCs‐based electrodes conformally attached to skin during compression. Reproduced with permission.[ 102 ] Copyright 2019, American Chemical Society. f‐i) SEM image of AgNW‐embedded aPDMS matrix and ii) the AgNWs‐aPDMS ECG electrodes on skin. Reproduced with permission.[ 103 ] Copyright 2018, American Chemical Society. g‐i) SEM image and backscattered electron (BSE) image (inset) of Ag–Au nanowires, ii) schematic illustrations of Ag–Au nanocomposites, and iii) photograph of Ag–Au nanocomposite EP electrodes. Reproduced with permission.[ 104 ] Copyright 2018, Springer Nature.](https://figures.cochat.ai/paper-figures/b5928ac01c404030b86c66085eea9895/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-4.jpg)
![Figure 5 Other Materials for on‐skin Electrodes. a‐i) PEDOT:PSS electrodes with different diameters for EEG recording and ii) schematic illustrations show the cross section and top view of the electrode. Reproduced with permission.[ 111 ] Copyright 2014, Wiley‐VCH. b) Printed PEDOT:PSS electrodes on commercial paper for ECG measurements. Reproduced with permission.[ 112 ] Copyright 2017, Wiley‐VCH. c‐i) Schematics of the hybrid Au/hydrogel composite electrode on a forefinger and ii) SEM image of the Au arrays of 0.2 mm × 0.2 mm square tiles. Reproduced with permission.[ 114 ] Copyright 2016, Elsevier. d‐i) Schematics of interlocked electrodes, ii) SEM of the interlocking layer, and iii) an image of interlocking electrodes on the upper arm. Reproduced with permission.[ 115 ] Copyright 2020, Wiley‐VCH. e) A photo showing seven different types of electrodes including hydrogel‐PDMS electrode (Ag/AgCl and steel electrode as control groups). Reproduced with permission.[ 116 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/45d17db89cce4aaba3d6dcea202c8ab8/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-5.jpg)
![Figure 6 Electrodes with tuned adhesion. a‐i) Schematic illustrations of the liquid spray‐on‐bandage and ii) the multifunctional epidermal electronic system (EES) attached on forearm, the conditions of compression and stretch on skin and after wearing for a week. Reproduced with permission.[ 118 ] Copyright 2013, Wiley‐VCH. b‐i) Schematic illustrations of Ca‐modified hydrogel patch, which was ii) firmly attached to the forearm when subjected to tensile, compression, and twisting forces. Reproduced with permission.[ 121 ] Copyright 2018, Wiley‐VCH. c‐i) Illustrations of graphene‐based electrodes, ii) the SEM inset shows microstructures and comparison of adhesive force between 3M tape and bioinspired microstructures in repeated cycles. Reproduced with permission.[ 76 ] Copyright 2017, American Chemical Society. d‐i) Adhesion model for OIA attachment, ii) schematic illustrations, and iii) SEM images of OIA structure, and iv) ECG signals recording on wet skin in dynamic movements. Reproduced with permission.[ 135,136] Copyright 2017, Springer Nature; Copyright 2019, Wiley‐VCH. e‐i) Photograph of flexible ECG sensor on skin. Comparisons of peel forces of various silicone‐based adhesives with different ratios of H‐PDMS to V‐PDMS and MQ resin content. Reproduced with permission.[ 139 ] Copyright 2017, Wiley‐VCH. f‐i) Photograph of the on‐skin device and ii) correlation between PEIE concentration and adhesion. Reproduced with permission.[ 142 ] Copyright 2017, Wiley‐VCH. g‐i) Photograph of the GEPC‐based TEE and ii) schematic illustrations of the water‐triggered decomposition process, iii) comparisons of adhesion force before/during water treatment, and iv) the SNR of EMG signals obtained from the TEE and commercial patch during wet state. Reproduced with permission.[ 150 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/6794823c6541483e90546de5a6f2efc4/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-6.jpg)
![Figure 7 a‐i) SEM image of a microperforated silicone layer with patterned Au mesh electrode, ii) SEM tilted view of a bilayer microperforated silicone support laminated on skin replica, and iii) comparisons of WVTR values of different silicone films and the value without a film. Reproduced with permission.[ 152 ] Copyright 2019, Springer Nature. b‐i) Photograph, ii) SEM image, and iii) WVTR of an EP sensor based on porous graphene and porous elastomer sponges. Reproduced with permission.[ 86 ] Copyright 2018, Wiley‐VCH. c‐i) Schematic illustrations and ii) SEM image of dual‐layer PU/c‐PVA electrospun membrane, and iii) directional water transport from the hydrophobic (PU) side to the hydrophilic (PVA) side. Reproduced with permission.[ 158 ] Copyright 2012, The Royal Society of Chemistry.](https://figures.cochat.ai/paper-figures/0e2f6aa9bff54e04b1e55247bd2816f6/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-7.jpg)
![Figure 8 Large area, multichannel electrodes. a‐i) 5 × 5 interconnected electrode array ii) used for turning the fan or lamp on or off by squeezing the flexor or extensor muscles, and iii) two channel EMG signals measured from the flexor and extensor muscles (black) and corresponding control input signals (red). Reproduced with permission.[ 59 ] Copyright 2016, American Chemical Society. b‐i) 8‐channel sensor array and ii) an array laminated on the amputated upper limb of a subject for controlling prosthetic movements, iii) four sensor arrays covering the scalp of the subject, and iv) the scalp ERP map. Reproduced with permission.[ 152 ] Copyright 2019, Springer Nature. c‐i) 3 × 3 stretchable electrode array on the forearm and its 6 channels layout, ii) four gestures and the corresponding motion of the wheeled mobile robot. Reproduced with permission.[ 102 ] Copyright 2019, American Chemical Society.](https://figures.cochat.ai/paper-figures/c1255f36249a4f24a3d60d71dc7832ab/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-8.jpg)
![Figure 9 Standalone on‐skin systems through integration. a‐i) A flexible sensing array with integrated FPCB, ii) schematic of the sensor array for multiplexed perspiration analysis, and iii) the smart wristband worn on forearm. Reproduced with permission.[ 174 ] Copyright 2016, Springer Nature. b‐i) The component side and ii) sensor side of an integrated wearable patch. Reproduced with permission.[ 175 ] Copyright 2016, Wiley‐VCH. c) The mechanoacoustic–electrophysiological sensing i) platform and ii) its stretchability. Reproduced with permission.[ 160 ] Copyright 2016, AAAS. d) The NFC‐enabled optoelectronic system. Reproduced with permission.[ 181 ] Copyright 2016, AAAS.](https://figures.cochat.ai/paper-figures/927c81c174624988b4880ac5555b617d/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-9.jpg)
![Figure 10 On‐skin electrodes for healthcare. a‐i) The mSC electrode system and ii) captured ECG and pulse signals, iii) the alarm function of the system. Reproduced with permission.[ 183 ] Copyright 2016, Wiley‐VCH. b‐i) On‐skin sensor patch with remote control unit and ii) comparisons with commercial sensors, iii) pilot study on a patient diagnosed with Parkinson's disease and dysphagia. Reproduced with permission.[ 189 ] Copyright 2019, AAAS. c‐i) On‐skin patches for recognizing mental fatigue, ii) prediction accuracy of different models, and iii) output fatigue level labels by the DT‐based model during daily mental works. Reproduced with permission.[ 164 ] Copyright 2020, American Chemical Society. d‐i) The chest‐mounted E‐tattoo and ii) captured ECG, SCG and STI signals, iii) the correlation between SBP/DBP and RAC measured by an e‐tattoo, and iv) continuous estimation of SBP and DBP by the SOMNO and EMAC tattoo. Reproduced with permission.[ 67 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/a95d758d31644e878ae923479c1d7a12/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-10.jpg)
![Figure 11 Applications of on‐skin electrodes in HMI. a‐i) On‐skin devices on the bicep and tricep (inset) during control of the virtual arm and ii) accuracy associated with a virtual arm targeting task. Reproduced with permission.[ 55 ] Copyright 2016, Wiley‐VCH. b‐i) Quadrotor control by sEMG signals from forearms based on the control gestures and ii) sEMG signals acquired from forearms corresponding to four gestures. Reproduced with permission.[ 61 ] Copyright 2013, Wiley‐VCH. c‐i) LTE electrodes for EEG recordings and ii) demonstration of the spelling task based on the BCI. Reproduced with permission.[ 58 ] Copyright 2015, National Academy of Sciences USA. d) The AgPMS electrode BCI system for controlling a driverless car. Reproduced with permission.[ 163 ] Copyright 2019, American Chemical Society. e‐i) Flexible wireless electrode systems, ii) CNN and SVM test accuracy on using time‐domain data and frequency‐domain data, and iii) in vivo demonstration of the on‐skin BMI with human subjects. Reproduced with permission.[ 191 ] Copyright 2019, Springer Nature.](https://figures.cochat.ai/paper-figures/80d5ef792637494ebb7637c9e5f1d8bb/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-11.jpg)

Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Capture bioelectrical signals (e.g., EMG) and | bioelectrical impedance (BI) variations via | electrodes 6,207 .”
From Recent advances in intelligent wearable systems: from multiscale biomechanical features towards human motion intent prediction (2026) · cites this paper as [6]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Surface energy and the contact of elastic solids
Johnson, K. L., Kendall, K., & Roberts, A. D. (1971) · Proceedings of the Royal Society of London. A. Mathematical and Physical Sciences · doi:10.1098/rspa.1971.0141 · Cited by 8,061
Wireless Power Transfer via Strongly Coupled Magnetic Resonances
Kurs, A., Karalis, A., Moffatt, R., Joannopoulos, J. D., Fisher, P., & Soljačić, M. (2007) · Science · doi:10.1126/science.1143254 · Cited by 5,576
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Materials and Mechanics for Stretchable Electronics
Rogers, J. A., Someya, T., & Huang, Y. (2010) · Science · doi:10.1126/science.1182383 · Cited by 5,042
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
The Use of Surface Electromyography in Biomechanics
De Luca, C. J. (1997) · Journal of Applied Biomechanics · doi:10.1123/jab.13.2.135 · Cited by 3,268
Capillarity and Wetting Phenomena
de Gennes, P.-G., Brochard-Wyart, F., & Quéré, D. (2004) · Springer eBooks · doi:10.1007/978-0-387-21656-0 · Cited by 3,121
Thermal conductivity of carbon nanotubes and their polymer nanocomposites: A review
Han, Z., & Fina, A. (2011) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2010.11.004 · Cited by 2,603
Skin electronics from scalable fabrication of an intrinsically stretchable transistor array
Wang, S., Xu, J., Wang, W., Wang, G.-J. N., Rastak, R., Molina-Lopez, F., Chung, J. W., Niu, S., Feig, V. R., Lopez, J., et al. (2018) · Nature · doi:10.1038/nature25494 · Cited by 2,198
Yuk, H., Lu, B., & Zhao, X. (2019) · Chemical Society Reviews · doi:10.1039/c8cs00595h · Cited by 2,108
Research works printed but not linked: 4
- Gallagher A. J., 2012, IRCOBI Conf., 40, 494
- Torres F., 2009, Clin. Cosmet. Invest. Dermatol., 2, 39
- Jagota A., 2011, Mater. Sci. Eng., R, 72, 253
- Zeng Z., 2020, ACS Sens.
Cited by
Items 201–300 of 446
Polyphenol‐Mediated Multifunctional Human–Machine Interface Hydrogel Electrodes in Bioelectronics
Jiang, L., Gan, D., Xu, C., Zhang, T., Gao, M., Xie, C., Zhang, D., & Lu, X. (2024) · Small Science · doi:10.1002/smsc.202400362 · Cited by 20
Highly Conformal Polymers for Ambulatory Electrophysiological Sensing
Hu, M., Zhang, J. P., Liu, Y., Zheng, X., Li, X., Li, X., & Yang, H. (2022) · Macromolecular Rapid Communications · doi:10.1002/marc.202200047 · Cited by 20
Electronic tattoos based on large-area Mo2C grown by chemical vapor deposition for electrophysiology
Wang, S., Wang, X., Zhang, W., Shi, X., Song, D., Zhang, Y., Zhao, Y., Zhao, Z., & Liu, N. (2023) · Nano Research · doi:10.1007/s12274-023-5423-y · Cited by 19
Soft and Conductive Polyethylene Glycol Hydrogel Electrodes for Electrocardiogram Monitoring
Lee, D., Song, J., Kim, J., Lee, J., Son, D., & Shin, M. (2023) · Gels · doi:10.3390/gels9120957 · Cited by 19
“…and capability to mimic the extracellular matrix (ECM) present in living tissues, PEGDA hydrogels have diverse applications in various biomedical 2 of 12 onto their skin, the silver-based electrodes caused skin irritations, such as rashes and itching, during electrophysiological monitoring [8–13].”
BackgroundFang, L. B., Dong, X., Weng, Z., & Chen, T. (2022) · Journal of theoretical and applied electronic commerce research · doi:10.3390/jtaer18010003 · Cited by 19
“Based on the metaverse, the human–machine interface should provide several types of value for humans, such as ease of use [34], leading the flow experience [33] and building an immersive environment [35].”
BackgroundPress-N-Go On-Skin Sensor with High Interfacial Toughness for Continuous Healthcare Monitoring
Hou, C., Cao, C., Ma, R., Ai, L., Hu, Z., Huang, Y., & Yao, X. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c22936 · Cited by 19
IoMT-Enabled Stress Monitoring in a Virtual Reality Environment and at Home
Vela, L. M., Crandall, H., Lim, T., Zhang, F., Gibbs, A., Mitchell, A. R. J., Condon, A., Diamond, L. M., Zhang, H., & Sanchez, B. (2023) · IEEE Internet of Things Journal · doi:10.1109/jiot.2023.3240099 · Cited by 18
“To address this unmet need, the research community is continuously looking for better electrode materials and for ways to better collect relevant physiological signals, including ECG [48].”
BackgroundWang, W., Lu, L., Ma, H., Li, Z., Lu, X., & Xie, Y. (2024) · Microsystems & Nanoengineering · doi:10.1038/s41378-024-00838-7 · Cited by 18
Self-citationUltrapermeable and Wet-Adhesive Monolayer Porous Film for Stretchable Epidermal Electrode
Tian, Q., Zhao, H., Zhou, R., Li, T., Huang, J., Tong, W., Xie, R., Li, Q., Li, G., & Liu, Z. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c16489 · Cited by 18
Chen, B., Qian, Z., Song, G., Zhuang, Z., Sun, X., Ma, S., Liang, Y., Ren, L., & Ren, L. (2023) · Chemical Engineering Journal · doi:10.1016/j.cej.2023.142350 · Cited by 18
A reversible gel-free electrode for continuous noninvasive electrophysiological signal monitoring
Liu, Q., Zhou, J., Yang, L., Xie, J., Guo, C., Li, Z., Qi, J., Shi, S., Zhang, Z., Yang, H., et al. (2023) · Journal of Materials Chemistry C · doi:10.1039/d3tc00948c · Cited by 18
Self-citationAnisotropic Electronic Skin for Neurofeedback
Hu, W., Song, D., Shi, X., Zhou, L., Zhao, Z., Xue, T., Lin, X., & Liu, N. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202309359 · Cited by 18
Wang, C., Hong, W., Deng, Y., Lan, L., Zhang, S., Ping, J. F., Ying, Y. B., Yu, C., Luo, J., Chen, W., et al. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202422980 · Cited by 17
Liu, L., Feng, X., Du, J., Wang, H., Xue, E., Chen, S., Yokota, T., Someya, T., & Wang, B. (2025) · Matter · doi:10.1016/j.matt.2025.102428 · Cited by 17
Chen, B., Qian, Z., Song, G., Niu, X., Yu, Y., Wang, S., Wu, J., Ma, S., Liang, Y., Ren, L., et al. (2025) · Nano Letters · doi:10.1021/acs.nanolett.4c05791 · Cited by 17
Huang, H., Wu, N., Liu, H., Dong, Y., Han, L., Wan, S., Dou, G., & Sun, L. (2021) · Advanced Materials Interfaces · doi:10.1002/admi.202101602 · Cited by 17
Zhang, J., Hu, M., Zheng, X., Wang, H., Li, X., Zhang, X., & Yang, H. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202300243 · Cited by 17
Singh, K., Lin, C.-C., Huang, W., Lei, W., Chiueh, H., Wang, Y., Chang, P.-H., Lin, R.-Z., & Huang, W.-C. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c23013 · Cited by 16
Hou, S., Fan, X., Jin, T., Li, H., & Han, Y. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202519643 · Cited by 16
Yang, G., Xie, B., Tian, J., Zhang, Y., Liu, Y., Gong, H., Pang, B., Qiu, Y., Gong, C., Bu, T., et al. (2025) · Advanced Materials · doi:10.1002/adma.202513302 · Cited by 16
Self-citationDesign, Fabrication, and Application of Stretchable Electronic Conductors
Cheng, B., Zhuo, J., Zhou, Y., Chen, J., Cao, L., He, J., Chen, Z., Ma, X., Wang, J., Li, H., et al. (2026) · Nano-Micro Letters · doi:10.1007/s40820-025-02009-3 · Cited by 16
Self-citationLiu, Y., Xu, B., Xie, Z., Yang, J., Liu, Y., Yang, Y., & Xu, H. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.3c13942 · Cited by 16
Evolution of Musculoskeletal Electronics
Li, J., Zhang, F., Lyu, H., Yin, P., Shi, L., Li, Z., Zhang, L., Di, C., & Tang, P. (2024) · Advanced Materials · doi:10.1002/adma.202303311 · Cited by 16
Self-citationHorii, T., Yamashita, K., Ito, M., Okada, K., & Fujie, T. (2024) · NPG Asia Materials · doi:10.1038/s41427-024-00553-9 · Cited by 17
“Polymer ultrathin fi lms ( “ nanosheets ” ) have thicknesses of tens to hundreds of nanometers and areas of several square centimeters (i.e., aspect ratios greater than or equal to 10 6 ) 33,34 .”
A Highly Sensitive Surface Electrode for Electrophysiological Monitoring
Jiang, L., Li, Q., Huang, Q., Li, D., Jiao, Y., Wang, Y., Li, Y., Zou, K., Chen, Z., Gu, J., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202421132 · Cited by 16
Niu, H., Lin, Z., Yu, P., Luo, H., Peng, C., Ning, W., Li, B., Wei, Y., Guo, Q., Yang, J., et al. (2025) · Journal of Colloid and Interface Science · doi:10.1016/j.jcis.2025.137984 · Cited by 16
Nayani, S., Rao, P. S., & Lakshmi, D. R. (2023) · Concurrency and Computation Practice and Experience · doi:10.1002/cpe.7601 · Cited by 16
“Itoptimizes 43,44 theweightoftheRoughSetTheoryinthefeatureminingphasewiththeconsiderationofminimizing the correlation as well as the kernel, hidden neuron count, number of estimators, epoch size, and hidden neuron count of SVM, DNN, AdaBoost, autoencoder, and LSTM to maximize the accuracy and…”
BackgroundXing, L., Liu, L., Jin, R., Zhang, H., Shen, Y., Zhang, S., He, Z., Li, D., Ren, H., Huang, Q., et al. (2024) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c12768 · Cited by 15
Zhao, Z., Huang, C., & Zeng, H. (2024) · Langmuir · doi:10.1021/acs.langmuir.4c03975 · Cited by 15
Li, N., Xiao, Y., Yang, D., & He, J. (2025) · International Journal of Biological Macromolecules · doi:10.1016/j.ijbiomac.2025.140966 · Cited by 15
Chen, X., Fu, Y., Yan, X., Liu, Z., Gan, S., Yang, W., Chen, S., & Liu, L. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202501130 · Cited by 15
Raouf, A., Yu, Y., Hao, S., Zhang, T., Ahmad, M., Rashid, M., Ullah, I., Yu, B., & Cong, H. (2025) · Journal of Materials Chemistry A · doi:10.1039/d5ta05697g · Cited by 15
Shi, Y., Ding, T., Yuan, Z., Li, R., Wang, B., & Wu, Z. (2022) · Nanoenergy Advances · doi:10.3390/nanoenergyadv2010003 · Cited by 15
Adhesive cutaneous conducting polymer electrodes
Dimov, I. B., Sautter, A., Lövenich, W., Neumann, C., & Malliaras, G. G. (2022) · Applied Physics Reviews · doi:10.1063/5.0079616 · Cited by 15
Perspective Chapter: Tissue-Electronics Interfaces
Seyedkhani, S. A., & Mohammadpour, R. (2022) · IntechOpen eBooks · doi:10.5772/intechopen.108129 · Cited by 15
High performance zwitterionic hydrogels for ECG/EMG signals monitoring
Shi, Y., Ding, Y., Wang, W., & Yu, D. (2023) · Colloids and Surfaces A Physicochemical and Engineering Aspects · doi:10.1016/j.colsurfa.2023.132081 · Cited by 14
Abd-Elbaki, M. K. M., Ragab, T. M., Ismael, N. E. R., & Khalil, A. S. G. (2023) · RSC Advances · doi:10.1039/d3ra05730e · Cited by 15
“21 Recently, several approaches have been introduced for the development of high-performance wearable epidermal electrodes such as metal thin lm electrodes, 22,23 carbon material electrodes, 24 – 26 conductive elastomeric laments (CEFs) electrodes, 27 metallic nanomaterials-based electrodes,…”
BackgroundMethodAdvancing Flexible Sensors through On-Demand Regulation of Supramolecular Nanostructures
Wu, B., Wu, T., Huang, Z., & Ji, S. (2024) · ACS Nano · doi:10.1021/acsnano.4c08310 · Cited by 14
Toward Hand Gesture Recognition Using a Channel-Wise Cumulative Spike Train Image-Driven Model
Yu, Y., Zhou, Z., Xu, Y., Chen, C., Guo, W., & Sheng, X. (2025) · Cyborg and Bionic Systems · doi:10.34133/cbsystems.0219 · Cited by 14
“Fortunately, the emerging new flexible sensing technologies might be able to provide promising and effective solutions [ 39 ].”
Self-citationAutonomic dysregulation and atrial arrhythmias
Mikhailov, A., Arora, R., & Pfenniger, A. (2025) · American Journal of Physiology-Cell Physiology · doi:10.1152/ajpcell.00537.2025 · Cited by 14
Non-Standard Electrode Placement Strategies for ECG Signal Acquisition
Metshein, M., Krivošei, A., Abdullayev, A., Annus, P., & Märtens, O. (2022) · Sensors · doi:10.3390/s22239351 · Cited by 14
“The most used system is the 12-lead system, followed by the 5-lead and the standard 3-lead system [3].”
MethodNucleobase-Modified Adhesive and Conductive Hydrogel Interface for Bioelectronics
Sun, S., Xu, M., Zhao, Y., Cheng, T., Xiang, Y., Liu, X., Wang, J., & Pei, R. (2023) · ACS Applied Nano Materials · doi:10.1021/acsanm.3c04282 · Cited by 14
Gan, D., Wang, Q., Zhang, X., Qu, X., Sun, H., Cui, Y., Wang, W., Qu, L., & Dong, X. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202411588 · Cited by 14
Yang, M.-C., Li, C., Ni, H., Song, Y., & Sun, J. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c08008 · Cited by 13
Seeing inside the Body Using Wearable Sensing and Imaging Technologies
Kim, S., Han, S., Park, S., & Koo, J. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202502480 · Cited by 13
Shen, Z., Li, J., Hu, H., Du, C., Ding, X., Pan, T., & Yu, X. (2025) · AI Sensors · doi:10.3390/aisens1020008 · Cited by 13
“For instance, sweat secretion on the skin surface alters the electrical properties of the sensor-skin interface [152], while hair and the stratum corneum can create contact gaps, thereby increasing impedance [153].”
Engineering Liquid Metal Nanoparticles for Wearable Devices
Chen, Y., Zhang, Z., He, S., & Liu, G. (2026) · ACS Nano · doi:10.1021/acsnano.5c18099 · Cited by 13
Loh, J. Y. Y., Zeineddine, A., Shayegannia, M., McNeil, R., McRae, L., & Kherani, N. P. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00420-z · Cited by 13
“Equation (2) provides the basis for surface wrinkling metrology.”
BackgroundGao, X., Su, T., Bao, Y., Lu, J., Zhang, L., He, C., & Ouyang, J. (2023) · Journal of Materials Chemistry C · doi:10.1039/d3tc01988h · Cited by 13
Neural repair and regeneration interfaces: a comprehensive review
Sha, B., & Du, Z. J. (2024) · Biomedical Materials · doi:10.1088/1748-605x/ad1f78 · Cited by 13
A Stretchable and Sweat-Adhesive 3D Graphene Eutectogel Electrode for EMG Monitoring
Pan, Y., Cui, X., Song, D., Hu, W., Lin, X., & Liu, N. (2024) · ACS Applied Nano Materials · doi:10.1021/acsanm.4c01676 · Cited by 13
Niu, X., Dou, R., Sun, H., Luo, D., & Liu, H. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.154214 · Cited by 13
Ren, G., Yang, J., Wang, X., Hu, G., Chen, H., Fang, Q., & Yang, A. (2025) · ACS Applied Polymer Materials · doi:10.1021/acsapm.4c04052 · Cited by 13
Review of Medical Adhesive Technology in the Context of Medical Adhesive-Related Skin Injury
Holm, D., Schommer, K., & Kottner, J. (2024) · Journal of Wound Ostomy and Continence Nursing · doi:10.1097/won.0000000000001115 · Cited by 11
Chen, M., Qin, W., Wang, Y., Song, Y., Tian, F., Zhai, H., Gu, C., Ge, L., Xue, S., Li, X., et al. (2025) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2025.138385 · Cited by 11
Temporary Tattoo-Inspired, Skin-Adaptable Epidermal Electrode from an Ultrathin PU–PVA Film
Ye, X., Li, L., Wang, Z., Wang, Y., Yang, J., Zheng, M., Wang, M., Ji, Z., Lin, S., Zhang, Y., et al. (2025) · ACS Sensors · doi:10.1021/acssensors.5c02018 · Cited by 11
Kim, K., Jeong, J. Y., Shin, J. H., & Kim, T. (2025) · Small Methods · doi:10.1002/smtd.202501708 · Cited by 11
Morphology-adaptive Au-Ag nanowire elastronics for integrated FlexoSERS and bioelectrical sensing
Zhang, H., Chen, Y., Chen, G., Zhang, W., Yang, C. G., Zhou, F., Zhao, Y., Deng, H., Huang, X., An, Y., et al. (2026) · Science Advances · doi:10.1126/sciadv.aec2162 · Cited by 11
Lee, M., Kim, J.-S., Khine, M. T., Kim, S., & Gandla, S. (2023) · Nanomaterials · doi:10.3390/nano13172467 · Cited by 11
“The electrodes used in clinical and research settings for EP signal recording are predominantly gel and dry electrodes [11,12].”
Design strategies for skin-interfaced sensors
Guo, S., & Huang, H. (2024) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2024.115671 · Cited by 12
A Scalable Laser‐Centric Fabrication of an Epidermal Cardiopulmonary Patch
Rachim, V. P., Lee, J., Kim, Y., Oh, J., Jeong, U., & Park, S. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200242 · Cited by 11
Liquid Metal-Enabled Epidermal Interfaces
Fang, T., Sun, Y., & Kong, D. (2024) · ACS Applied Bio Materials · doi:10.1021/acsabm.3c01298 · Cited by 10
Carbon-based iontronics – current state and future perspectives
Li, P., Galek, P., Grothe, J., & Kaskel, S. (2025) · Chemical Science · doi:10.1039/d4sc06817c · Cited by 10
Indoor Abnormal Behavior Detection for the Elderly: A Review
Gu, T., & Tang, M. (2025) · Sensors · doi:10.3390/s25113313 · Cited by 10
“…application due to skin–sensor modulus mismatch, leading to poor biocompatibility [9]; unstable voltage output from self-powered sensors [10,11]; motion artifacts contributing to interface noise [12]; and signal distortion caused by material plasticity and cyclic degradation of nanomaterials [13].”
Fan, F., Chen, L., Ma, Y., Wang, P., Dong, X., Huang, F., Niu, Y., Zhou, Y., Shu, Z., Xie, J., et al. (2025) · Small Methods · doi:10.1002/smtd.202500219 · Cited by 10
Khodami, S., Gharakhloo, M., Dagdelen, S., Fita, P., Karbarz, M., & Stojek, Z. J. (2025) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2025.116909 · Cited by 10
Han, L., Zhang, W., Fan, C., Qi, Z., Liu, Y., Wang, Y., Zhao, Y., Zhang, B., Chen, X., Dong, Y., et al. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202512101 · Cited by 10
Self‐Powered Flexible Electronics: Biomass Hydrogels for Integrated Sensing and Energy Harvesting
Song, B., Wang, C., Hao, S., Shao, C., Wen, J., Yang, J. H., & Cong, H. (2026) · Small · doi:10.1002/smll.202511724 · Cited by 10
Fan, L., Gao, F., Xu, L., Xun, X., Zhao, S., Yang, B., Bi, H., Zhao, X., Liao, Q., & Zhang, Y. (2026) · Nano-Micro Letters · doi:10.1007/s40820-026-02132-9 · Cited by 10
“The impedance of the conformal electrodes can be mathematically represented as: where w is the frequency [10, 35].”
Guo, J., Wang, X., Bai, R., Zhang, Z., Chen, H., Xue, K., Ma, C., Zang, D., Yin, E., Gao, K., et al. (2024) · Biosensors · doi:10.3390/bios14090432 · Cited by 10
Bioinspired Intelligent Electronic Skin for Medicine and Healthcare
Liu, Z., Hu, J., & Shen, G. (2025) · Small Methods · doi:10.1002/smtd.202402164 · Cited by 10
Opportunities and challenges for flexible and printable electrodes in electroencephalography
Xing, L., Batchelor, J. C., & Casson, A. J. (2021) · IEEE International Conference on Flexible and Printable Sensors and Systems (FLEPS) · doi:10.1109/fleps51544.2021.9469782 · Cited by 10
Biocomposites - Recent Advances
Elnashar, M. M. M., & Karakuş, S. (2022) · IntechOpen eBooks · doi:10.5772/intechopen.104003 · Cited by 10
Chen, S., Peng, C., Yang, B., Lin, J., Zhou, L., Jiang, Z., Liu, Z., Liu, Y., & Tang, L. (2026) · npj Artificial Intelligence · doi:10.1038/s44387-026-00083-5 · Cited by 9
Self-citationDesign and applications of conductive polymers in wearable electrophysiological sensing
Zhou, N., Zhang, L., Cui, T., Wu, P., & Lei, Z. (2025) · Science China Technological Sciences · doi:10.1007/s11431-025-3114-6 · Cited by 9
“Furthermore, consistent adhesion is crucial for maintaining a stable interfacial impedance, which prevents signal loss or distortion that can occur if the sensor partially detaches [72].”
Sarma, S., Jain, N., Bansal, L., Vishwakarma, R., Prasun, A., Sahu, T. K., Kumar, R., & Sarma, T. K. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c08151 · Cited by 9
Tang, J., Luk, P. C. K., & Zhou, Y. (2023) · Bioengineering · doi:10.3390/bioengineering10050514 · Cited by 9
A High-Entropy Electrode Material for Multimodal Biosignal Detection
Yang, L., Hu, Z., Xiang, Z., Zhou, J., Wang, X., Liu, Q., Gan, L., Shi, S., Yang, W., Zhang, Y., et al. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.156209 · Cited by 9
Zhang, H., Shen, Y., Xing, L., Wang, W., Liu, L., Jin, R., He, Z., Wang, Y., Yao, S., Cao, X., et al. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.160973 · Cited by 9
Tan, K. Y., Poh, W. S., Endot, N. A., Khiew, P. S., Foo, C. Y., & Lim, . N. (2025) · Additive manufacturing · doi:10.1016/j.addma.2025.104706 · Cited by 8
Direct-ink-writing biosensors: From material innovation to personalized healthcare applications
Li, D., Liang, H., & Zhang, Y. (2025) · Review of Materials Research · doi:10.1016/j.revmat.2025.100071 · Cited by 8
Deformation Driven Suction Cups: A Mechanics‐Based Approach to Wearable Electronics
Lee, S., Akerson, A., Pardakhtim, R., Hajiesmaili, E., Rhodes, K., Li, Z., Stanley, A. A., Amini, A., Piazza, D., Daraio, C., et al. (2026) · Advanced Science · doi:10.1002/advs.202520417 · Cited by 8
Self-citationThayil, R., Thayil, A., & Parne, S. R. (2026) · Advanced Engineering Materials · doi:10.1002/adem.202502674 · Cited by 8
A high-performance composite fiber with an organohydrogel sheath for electrocardiogram monitoring
Chang, B., Luo, J., Liu, J., Zhang, B., Zhu, M. L., Li, K., Li, Y., Zhang, Q., Shi, G., & Hou, C. (2024) · Journal of Materials Chemistry C · doi:10.1039/d4tc01165a · Cited by 8
Learning from Octopuses: Cutting-Edge Developments and Future Directions
Duan, J., Lei, Y., Fang, J., Qi, Q., Zhan, Z., & Wu, Y. (2025) · Biomimetics · doi:10.3390/biomimetics10040224 · Cited by 8
“Regarding multifunctional integrated bionic systems, robots that can monitor environmental changes in real time and autonomously adjust their behavior will be developed by combining the octopus’s perceptual abilities with the responsiveness of smart materials [32,101–106].”
Wearable, Wireless, Stretchable, High‐Density Surface Electromyography Interface
Zhao, Y., Zhang, N., Li, J., Ren, J., & Gu, G. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202500607 · Cited by 8
Electrically functionalized body surface for deep-tissue bioelectrical recording
Zhang, D., Zhao, G., Zhang, Y. E., Xu, D., Wang, S., Wang, K., Zhou, B., Ling, Y., Liu, Y., Cui, Q., et al. (2026) · Nature Biomedical Engineering · doi:10.1038/s41551-026-01663-1 · Cited by 7
Highly Stable Polymeric Electrooculography Electrodes for Contactless Human-Machine Interactions
Yu, X., Luo, Z., Ouyang, X., Wang, W., Rao, Y., Yuan, Y., Cai, Z. Y., Hu, Y., & Li, X. (2025) · ACS Sensors · doi:10.1021/acssensors.5c00031 · Cited by 7
Electronic Tattoos for Next Generation Wearable Electronics and Healthcare Applications
Singh, A., Ahmed, E., Sharma, S., Dubey, A., Sundramoorthy, A. K., & Arya, S. K. (2025) · ECS Sensors Plus · doi:10.1149/2754-2726/ae23e2 · Cited by 7
“137 In addition, the skin’s bioimpedance may be actively monitored using bioimpedance sensors, which can provide information about blood flow, sweating, and skin moisture.”
Lian, Z., Wang, L., Xue, Y., Jiang, Y., Yu, D., & Wang, W. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.175720 · Cited by 7
Ng, S. H., Seah, G. E. K. K., Safanama, D., & Goh, S. S. (2025) · Analytica Chimica Acta · doi:10.1016/j.aca.2025.343772 · Cited by 7
A Novel Wearable Device for Continuous Blood Pressure Monitoring Utilizing Strain Gauge Technology
McMurray, J. P., DeVries, A., Frazee, K. P., Sizemore, B., Branan, K. L., Jennings, R., & Coté, G. L. (2025) · Biosensors · doi:10.3390/bios15070413 · Cited by 7
“Long-term adhesion can be difficult and cause skin irritation, motion sensitivity, and slow delamination of the flexible substrate, which will cause a drift in the BP calibration, all of which are affected by environmental factors like temperature, humidity, and water exposure [34–36].”
Continuous biosignal acquisition beyond the limit of epidermal turnover
Bhatia, A., Kasper, K. A., & Gutruf, P. (2025) · Materials Horizons · doi:10.1039/d5mh00758e · Cited by 7
Lee, J., Rachim, V. P., Hwang, Y. M., & Park, S. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3359215 · Cited by 7
“As the importance of continuous monitoring of biosignals becomes increasingly apparent, research on biosignal monitoring for various purposes has accelerated with the rapid development of wearable sensor technologies along with the miniaturization of electronics [1], [2], [3].”
BackgroundOctopus-Inspired Soft Robot for Slow Drug Release
Tong, D., Zhao, Y., Wu, Z., Chen, Y., Xu, X., Chen, Q., Fan, X., & Yang, Z. (2024) · Biomimetics · doi:10.3390/biomimetics9060340 · Cited by 7
Seo, B., Choi, Y. H., Noh, D., Kim, J., Chen, X., & Choi, W. (2024) · Device · doi:10.1016/j.device.2024.100602 · Cited by 7
Zimo, 蔡. C., Cuifang, 匡. K., Huayong, 杨. Y., Minghui, 洪. H., & Kaichen, 徐. X. (2024) · Chinese Journal of Lasers · doi:10.3788/cjl231372 · Cited by 7
Bioelectronic Sensors for Neuromuscular Perception in Human‐Machine Interfaces
Li, J., Niu, B., Wu, K., Xiao, J., Liu, X., & Wang, Y. (2025) · Advanced Robotics Research · doi:10.1002/adrr.202500047 · Cited by 6
Wang, C., Chen, J., Xiao, T., Ling, T., An, X., Jiang, T., Shao, R., Yang, L., & Sun, S. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.168785 · Cited by 6
All‐Dry Self‐Adhesive Soft Electrodes with Subsurface Pores for Long‐Term Skin Recording
Jeon, M. S., Yi, J., Hong, S., Lee, J., Jang, H., Kim, B., Lee, H., Lee, C. H., & Kim, D. R. (2025) · Small · doi:10.1002/smll.202507416 · Cited by 6
Metadata
- Authors
- Wu, H., Yang, G., Zhu, K., Liu, S., Guo, W., Jiang, Z., & Li, Z.
- Affiliations
- State Key Laboratory of Digital Manufacturing Equipment and Technology (CN)
- Fudan University (CN)
- Corresponding author
- Wu, Hao
- Li, Zhuo
- Venue
- Advanced Science
- Publisher
- Wiley
- Published
- 2020-12-04
- Volume
- 8
- Issue
- 2
- Pages
- 2001938-2001938
- ISSN
- 2198-3844 (ISSN-L)
- PMC
- PMC7816724
- OpenAlex
- W3110558114
- Cited by
- 395
- References
- 233
- Citations by year
- 2026: 56 · 2025: 95 · 2024: 77 · 2023: 69 · 2022: 59 · 2021: 38 · 2020: 1
- FWCI
- 3.49
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Advanced Science (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- National Natural Science Foundation of China (51820105008, 91648115)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Nanomaterials and Printing Technologies
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Keywords
- electrophysiological monitoring, human-machine interaction, wearable sensors, stretchable materials, biocompatibility, electrode adhesion, breathability, electrode materials, healthcare monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.