Materials, Devices, and Systems of On‐Skin Electrodes for Electrophysiological Monitoring and Human–Machine Interfaces
Open accessThe DOI matches its Crossref record; 231 of the 233 references it cites were found in a registry or an index.
Abstract
On‐skin electrodes function as an ideal platform for collecting high‐quality electrophysiological (EP) signals due to their unique characteristics, such as stretchability, conformal interfaces with skin, biocompatibility, and wearable comfort. The past decade has witnessed great advancements in performance optimization and function extension of on‐skin electrodes. With continuous development and great promise for practical applications, on‐skin electrodes are playing an increasingly important role in EP monitoring and human–machine interfaces (HMI). In this review, the latest progress in the development of on‐skin electrodes and their integrated system is summarized. Desirable features of on‐skin electrodes are briefly discussed from the perspective of performances. Then, recent advances in the development of electrode materials, followed by the analysis of strategies and methods to enhance adhesion and breathability of on‐skin electrodes are examined. In addition, representative integrated electrode systems and practical applications of on‐skin electrodes in healthcare monitoring and HMI are introduced in detail. It is concluded with the discussion of key challenges and opportunities for on‐skin electrodes and their integrated systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Metal thin film electrodes. a) On‐skin electrodes with serpentine design. Reproduced with permission.[ 55 ] Copyright 2016, Wiley‐VCH. b) Stretchable electrodes with UL‐Sil and fabrics. Reproduced with permission.[ 56 ] Copyright 2014, Springer Nature. c‐i) Epidermal electrode systems incorporating EMG/ECG sensors and ii) image of the on‐skin electronics integrated onto a tattoo. Reproduced with permission.[ 57 ] Copyright 2011, AAAS. d) A schematic of TCR electrodes and the magnified views of Peano curves (inset). Reproduced with permission.[ 58 ] Copyright 2015, National Academy of Sciences USA. e) Photographs of gold film electrodes with wrinkled structures on the plasticized silk substrate. Reproduced with permission.[ 60 ] Copyright 2018, Wiley‐VCH. f‐i) An illustration of optimized on‐skin electrodes and ii) an SEM image shows the superior conformality of a 5 µm thick membrane. Reproduced with permission.[ 61 ] Copyright 2013, Wiley‐VCH. g‐i) Multifunctional device consisting of EP sensors and ii) an image shows conformal skin‐electrode interfaces during compression. Reproduced with permission.[ 66 ] Copyright 2015, Wiley‐VCH. h) Images of the SCG and ECG sensors with excellent conformal contact with skin. Reproduced with permission.[ 67 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/9100e896ece649828eaacff72284fbe2/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-2.jpg)
![Figure 3 Carbon material electrodes. a) Images of nine CNT/PDMS electrodes. Reproduced with permission.[ 72 ] Copyright 2012, IEEE. b) A schematic illustration of a CNTs/aPDMS ECG electrode. Reproduced with permission.[ 75 ] Copyright 2014, Springer Nature. c) The CNT/graphene electrode patch with the content of CNT and graphene at 9:1. Reproduced with permission.[ 76 ] Copyright 2016, American Chemical Society. d) A photograph of the graphene electronic tattoo containing EP sensors. Reproduced with permission.[ 82 ] Copyright 2017, American Chemical Society. e) rGO‐aPDMS electrodes for monitoring EP signals. Reproduced with permission.[ 85 ] Copyright 2020, Elsevier. f) EP electrodes with laser‐induced porous graphene. Reproduced with permission.[ 86 ] Copyright 2018, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/05f37dbfe3fd43509659163035bd1359/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-3.jpg)
![Figure 4 Electrodes based on Metallic nanomaterials. a‐i) On‐skin electrodes with Cu networked mesh structure (inset) and ii) schematic of Cu nanomesh electrodes. Reproduced with permission.[ 98 ] Copyright 2016, Wiley‐VCH. b‐i) Au nanomesh conductors and ii) the process of the polyvinyl alcohol nanofibers (PVA NFs) dissolution. Reproduced with permission.[ 44 ] Copyright 2017, Springer Nature . c‐i) A cross‐sectional SEM image of convoluted nanonetworks and ii) an image of the NEE intimately coupling with human skin. Reproduced with permission.[ 100 ] Copyright 2019, Wiley‐VCH. d‐i) SEM image of Ag‐PU network conductors, ii) an illustration of the hydrogen bonding, and iii) porous Ag‐PU electrodes on skin. Reproduced with permission.[ 101 ] Copyright 2019, Wiley‐VCH. e‐i) SEM image of the Ag flakes and Ag nanoparticle and ii) the ECCs‐based electrodes conformally attached to skin during compression. Reproduced with permission.[ 102 ] Copyright 2019, American Chemical Society. f‐i) SEM image of AgNW‐embedded aPDMS matrix and ii) the AgNWs‐aPDMS ECG electrodes on skin. Reproduced with permission.[ 103 ] Copyright 2018, American Chemical Society. g‐i) SEM image and backscattered electron (BSE) image (inset) of Ag–Au nanowires, ii) schematic illustrations of Ag–Au nanocomposites, and iii) photograph of Ag–Au nanocomposite EP electrodes. Reproduced with permission.[ 104 ] Copyright 2018, Springer Nature.](https://figures.cochat.ai/paper-figures/b5928ac01c404030b86c66085eea9895/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-4.jpg)
![Figure 5 Other Materials for on‐skin Electrodes. a‐i) PEDOT:PSS electrodes with different diameters for EEG recording and ii) schematic illustrations show the cross section and top view of the electrode. Reproduced with permission.[ 111 ] Copyright 2014, Wiley‐VCH. b) Printed PEDOT:PSS electrodes on commercial paper for ECG measurements. Reproduced with permission.[ 112 ] Copyright 2017, Wiley‐VCH. c‐i) Schematics of the hybrid Au/hydrogel composite electrode on a forefinger and ii) SEM image of the Au arrays of 0.2 mm × 0.2 mm square tiles. Reproduced with permission.[ 114 ] Copyright 2016, Elsevier. d‐i) Schematics of interlocked electrodes, ii) SEM of the interlocking layer, and iii) an image of interlocking electrodes on the upper arm. Reproduced with permission.[ 115 ] Copyright 2020, Wiley‐VCH. e) A photo showing seven different types of electrodes including hydrogel‐PDMS electrode (Ag/AgCl and steel electrode as control groups). Reproduced with permission.[ 116 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/45d17db89cce4aaba3d6dcea202c8ab8/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-5.jpg)
![Figure 6 Electrodes with tuned adhesion. a‐i) Schematic illustrations of the liquid spray‐on‐bandage and ii) the multifunctional epidermal electronic system (EES) attached on forearm, the conditions of compression and stretch on skin and after wearing for a week. Reproduced with permission.[ 118 ] Copyright 2013, Wiley‐VCH. b‐i) Schematic illustrations of Ca‐modified hydrogel patch, which was ii) firmly attached to the forearm when subjected to tensile, compression, and twisting forces. Reproduced with permission.[ 121 ] Copyright 2018, Wiley‐VCH. c‐i) Illustrations of graphene‐based electrodes, ii) the SEM inset shows microstructures and comparison of adhesive force between 3M tape and bioinspired microstructures in repeated cycles. Reproduced with permission.[ 76 ] Copyright 2017, American Chemical Society. d‐i) Adhesion model for OIA attachment, ii) schematic illustrations, and iii) SEM images of OIA structure, and iv) ECG signals recording on wet skin in dynamic movements. Reproduced with permission.[ 135,136] Copyright 2017, Springer Nature; Copyright 2019, Wiley‐VCH. e‐i) Photograph of flexible ECG sensor on skin. Comparisons of peel forces of various silicone‐based adhesives with different ratios of H‐PDMS to V‐PDMS and MQ resin content. Reproduced with permission.[ 139 ] Copyright 2017, Wiley‐VCH. f‐i) Photograph of the on‐skin device and ii) correlation between PEIE concentration and adhesion. Reproduced with permission.[ 142 ] Copyright 2017, Wiley‐VCH. g‐i) Photograph of the GEPC‐based TEE and ii) schematic illustrations of the water‐triggered decomposition process, iii) comparisons of adhesion force before/during water treatment, and iv) the SNR of EMG signals obtained from the TEE and commercial patch during wet state. Reproduced with permission.[ 150 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/6794823c6541483e90546de5a6f2efc4/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-6.jpg)
![Figure 7 a‐i) SEM image of a microperforated silicone layer with patterned Au mesh electrode, ii) SEM tilted view of a bilayer microperforated silicone support laminated on skin replica, and iii) comparisons of WVTR values of different silicone films and the value without a film. Reproduced with permission.[ 152 ] Copyright 2019, Springer Nature. b‐i) Photograph, ii) SEM image, and iii) WVTR of an EP sensor based on porous graphene and porous elastomer sponges. Reproduced with permission.[ 86 ] Copyright 2018, Wiley‐VCH. c‐i) Schematic illustrations and ii) SEM image of dual‐layer PU/c‐PVA electrospun membrane, and iii) directional water transport from the hydrophobic (PU) side to the hydrophilic (PVA) side. Reproduced with permission.[ 158 ] Copyright 2012, The Royal Society of Chemistry.](https://figures.cochat.ai/paper-figures/0e2f6aa9bff54e04b1e55247bd2816f6/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-7.jpg)
![Figure 8 Large area, multichannel electrodes. a‐i) 5 × 5 interconnected electrode array ii) used for turning the fan or lamp on or off by squeezing the flexor or extensor muscles, and iii) two channel EMG signals measured from the flexor and extensor muscles (black) and corresponding control input signals (red). Reproduced with permission.[ 59 ] Copyright 2016, American Chemical Society. b‐i) 8‐channel sensor array and ii) an array laminated on the amputated upper limb of a subject for controlling prosthetic movements, iii) four sensor arrays covering the scalp of the subject, and iv) the scalp ERP map. Reproduced with permission.[ 152 ] Copyright 2019, Springer Nature. c‐i) 3 × 3 stretchable electrode array on the forearm and its 6 channels layout, ii) four gestures and the corresponding motion of the wheeled mobile robot. Reproduced with permission.[ 102 ] Copyright 2019, American Chemical Society.](https://figures.cochat.ai/paper-figures/c1255f36249a4f24a3d60d71dc7832ab/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-8.jpg)
![Figure 9 Standalone on‐skin systems through integration. a‐i) A flexible sensing array with integrated FPCB, ii) schematic of the sensor array for multiplexed perspiration analysis, and iii) the smart wristband worn on forearm. Reproduced with permission.[ 174 ] Copyright 2016, Springer Nature. b‐i) The component side and ii) sensor side of an integrated wearable patch. Reproduced with permission.[ 175 ] Copyright 2016, Wiley‐VCH. c) The mechanoacoustic–electrophysiological sensing i) platform and ii) its stretchability. Reproduced with permission.[ 160 ] Copyright 2016, AAAS. d) The NFC‐enabled optoelectronic system. Reproduced with permission.[ 181 ] Copyright 2016, AAAS.](https://figures.cochat.ai/paper-figures/927c81c174624988b4880ac5555b617d/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-9.jpg)
![Figure 10 On‐skin electrodes for healthcare. a‐i) The mSC electrode system and ii) captured ECG and pulse signals, iii) the alarm function of the system. Reproduced with permission.[ 183 ] Copyright 2016, Wiley‐VCH. b‐i) On‐skin sensor patch with remote control unit and ii) comparisons with commercial sensors, iii) pilot study on a patient diagnosed with Parkinson's disease and dysphagia. Reproduced with permission.[ 189 ] Copyright 2019, AAAS. c‐i) On‐skin patches for recognizing mental fatigue, ii) prediction accuracy of different models, and iii) output fatigue level labels by the DT‐based model during daily mental works. Reproduced with permission.[ 164 ] Copyright 2020, American Chemical Society. d‐i) The chest‐mounted E‐tattoo and ii) captured ECG, SCG and STI signals, iii) the correlation between SBP/DBP and RAC measured by an e‐tattoo, and iv) continuous estimation of SBP and DBP by the SOMNO and EMAC tattoo. Reproduced with permission.[ 67 ] Copyright 2019, Wiley‐VCH.](https://figures.cochat.ai/paper-figures/a95d758d31644e878ae923479c1d7a12/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-10.jpg)
![Figure 11 Applications of on‐skin electrodes in HMI. a‐i) On‐skin devices on the bicep and tricep (inset) during control of the virtual arm and ii) accuracy associated with a virtual arm targeting task. Reproduced with permission.[ 55 ] Copyright 2016, Wiley‐VCH. b‐i) Quadrotor control by sEMG signals from forearms based on the control gestures and ii) sEMG signals acquired from forearms corresponding to four gestures. Reproduced with permission.[ 61 ] Copyright 2013, Wiley‐VCH. c‐i) LTE electrodes for EEG recordings and ii) demonstration of the spelling task based on the BCI. Reproduced with permission.[ 58 ] Copyright 2015, National Academy of Sciences USA. d) The AgPMS electrode BCI system for controlling a driverless car. Reproduced with permission.[ 163 ] Copyright 2019, American Chemical Society. e‐i) Flexible wireless electrode systems, ii) CNN and SVM test accuracy on using time‐domain data and frequency‐domain data, and iii) in vivo demonstration of the on‐skin BMI with human subjects. Reproduced with permission.[ 191 ] Copyright 2019, Springer Nature.](https://figures.cochat.ai/paper-figures/80d5ef792637494ebb7637c9e5f1d8bb/materials-devices-and-systems-of-onskin-electrodes-for-electrophysiological-figure-11.jpg)

Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Capture bioelectrical signals (e.g., EMG) and | bioelectrical impedance (BI) variations via | electrodes 6,207 .”
From Recent advances in intelligent wearable systems: from multiscale biomechanical features towards human motion intent prediction (2026) · cites this paper as [6]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Surface energy and the contact of elastic solids
Johnson, K. L., Kendall, K., & Roberts, A. D. (1971) · Proceedings of the Royal Society of London. A. Mathematical and Physical Sciences · doi:10.1098/rspa.1971.0141 · Cited by 8,061
Wireless Power Transfer via Strongly Coupled Magnetic Resonances
Kurs, A., Karalis, A., Moffatt, R., Joannopoulos, J. D., Fisher, P., & Soljačić, M. (2007) · Science · doi:10.1126/science.1143254 · Cited by 5,576
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Materials and Mechanics for Stretchable Electronics
Rogers, J. A., Someya, T., & Huang, Y. (2010) · Science · doi:10.1126/science.1182383 · Cited by 5,042
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
The Use of Surface Electromyography in Biomechanics
De Luca, C. J. (1997) · Journal of Applied Biomechanics · doi:10.1123/jab.13.2.135 · Cited by 3,268
Capillarity and Wetting Phenomena
de Gennes, P.-G., Brochard-Wyart, F., & Quéré, D. (2004) · Springer eBooks · doi:10.1007/978-0-387-21656-0 · Cited by 3,121
Thermal conductivity of carbon nanotubes and their polymer nanocomposites: A review
Han, Z., & Fina, A. (2011) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2010.11.004 · Cited by 2,603
Skin electronics from scalable fabrication of an intrinsically stretchable transistor array
Wang, S., Xu, J., Wang, W., Wang, G.-J. N., Rastak, R., Molina-Lopez, F., Chung, J. W., Niu, S., Feig, V. R., Lopez, J., et al. (2018) · Nature · doi:10.1038/nature25494 · Cited by 2,198
Yuk, H., Lu, B., & Zhao, X. (2019) · Chemical Society Reviews · doi:10.1039/c8cs00595h · Cited by 2,108
Research works printed but not linked: 4
- Gallagher A. J., 2012, IRCOBI Conf., 40, 494
- Torres F., 2009, Clin. Cosmet. Invest. Dermatol., 2, 39
- Jagota A., 2011, Mater. Sci. Eng., R, 72, 253
- Zeng Z., 2020, ACS Sens.
Cited by
Items 301–400 of 446
Forood, A. M. K., Pandian, S. S. R., McNaboe, R. Q., Lachos, T. D. C., Lantigua, D. O., & Posada-Quintero, H. F. (2025) · Micromachines · doi:10.3390/mi16111255 · Cited by 6
“Furthermore, reliance on adhesive gel electrodes limits long-term wear due to discomfort, skin irritation, and signal degradation, reducing feasibility for daily monitoring in real-world contexts [15,16].”
All-day skin wiring for high-fidelity epidermal sensing with liquid metal tattoos
Yi, J., Yan, W., Luo, J., Wang, Y., Wang, Z., Sun, Q., Wang, M., Li, L., Ye, X., Yang, J., et al. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.172444 · Cited by 6
Wei, Q., Wang, X., Liu, H., Li, M., An, Y., & Ma, Z. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.172761 · Cited by 6
Li, K., Li, Y., Wei, H., Tian, J., Zhang, J., Wang, H., Dai, Z., Men, Y., Cao, W., & Huang, Z. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.175525 · Cited by 6
Multi-Modal Portable Respiratory Rate Monitoring Device for Childhood Pneumonia Detection
Khan, S. R., Wang, X., Jiang, T., Ju, W., Radacsi, N., Kadir, M. A., Rabbani, K. S. E., Cunningham, S., & Mitra, S. (2023) · Micromachines · doi:10.3390/mi14040708 · Cited by 6
“However, it can cause skin irritation including redness and itching for long-term use [17].”
BackgroundHuang, H., Cai, X., Wan, S., & Sun, L. (2026) · Materials Today Bio · doi:10.1016/j.mtbio.2026.103241 · Cited by 5
“As the fundamental element for capturing EP signals [12], electrodes convert ionic currents into measurable voltages [13].”
InfluentialUsing BCI and EEG to process and analyze driver’s brain activity signals during VR simulation
Nader, M., Jacyna-Gołda, I., Nader, S., & Nehring, K. (2021) · Archives of Transport · doi:10.5604/01.3001.0015.6305 · Cited by 5
Van, N. D., Mills, D. E., Tran, C., Nguyen, T., Nguyen, H., Tran, T. L., Song, P., Phan, H., Nguyen, N., Dao, D. V., et al. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.3c13541 · Cited by 5
Zhao, Z., Yu, H. Y., Wisniewski, D. J., Cea, C., Ma, L., Trautmann, E. M., Churchland, M. M., Gelinas, J. N., & Khodagholy, D. (2024) · Advanced Science · doi:10.1002/advs.202308014 · Cited by 5
Cai, Y., Wang, F., Yang, L., Chen, L., Ying, J., Liu, Z., Wang, G., Ma, C., & Liu, J. (2024) · Journal of Materials Chemistry C · doi:10.1039/d4tc02801e · Cited by 5
Stretchable and Self‐Adhesive Conductors for Smart Epidermal Electronics
Wang, L., & Kong, D. (2024) · Macromolecular Rapid Communications · doi:10.1002/marc.202400774 · Cited by 5
Water-Removable Paste Electrode for Electrophysiological Monitoring on Hairy Skin
Zhang, R., Wen, S., Zhao, Y., & Ji, S. (2025) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.4c02190 · Cited by 5
Li, X., Zhang, J., Shi, B., Li, Y., Wang, Y., Shuai, K., Li, Y., Ming, G., Song, T., Pei, W., et al. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c00565 · Cited by 5
Flexible Networks of Patterned Conducting Polymer Nanowires for Fully Polymeric Bioelectronics
Chapman, C. A. R., Cuttaz, E. A., Tahirbegi, I. B., Novikov, A. A., Petkos, K., Koutsoftidis, S., Drakakis, E. M., Goding, J. A., & Green, R. A. (2021) · Advanced NanoBiomed Research · doi:10.1002/anbr.202100102 · Cited by 5
He, X. D., Yang, R., Xu, C., Zhao, Z., Cheng, Y. F., Egberts, P., Zeng, H., & Lu, Q. (2023) · Nanoscale · doi:10.1039/d3nr02821f · Cited by 5
Lin, Z., Luo, M., Liang, J., Li, Z., Lin, Y.-T., Chen, X., Chen, B., Peng, L., Ouyang, Y., & Mou, L. (2025) · Journal of Materials Chemistry B · doi:10.1039/d4tb01711k · Cited by 5
Self-citationFlexible and Scalable Dry Conductive Elastomeric Nanocomposites for Surface Stimulation Applications
Seshadri, D. R., Radwan, A. N., Bianco, N. D., Zorman, C. A., & Bogie, K. M. (2023) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2023.3289059 · Cited by 5
“This can be attributed to the adhesive and/or heat generation during electrode cycling during electroceutical therapy [27].”
BackgroundMukherjee, A., Pancham, P. P., & Yang, H. (2021) · IEEE International Flexible Electronics Technology Conference (IFETC) · doi:10.1109/ifetc49530.2021.9580512 · Cited by 5
Nica, S. L., Hulubei, C., Popovici, D., & Dobromir, M. (2022) · Polymer Engineering and Science · doi:10.1002/pen.25872 · Cited by 5
Varnaitė‐Žuravliova, S., Rukuižienė, Ž., Skurkytė-Papievienė, V., Bekampienė, P., Trakšelytė, V., & Baltušnikaitė‐Guzaitienė, J. (2026) · Journal of Functional Biomaterials · doi:10.3390/jfb17020100 · Cited by 4
Stimulation Modalities in Wearable Haptic Systems: Single-Mode Feedback to Multiphysics Actuation
Guo, X. M., Avila, R. O., Wu, Z., Wang, X., Xu, P., Ju, P., Yang, Q., Liu, P., Choi, S., Nikitsin, A. V., et al. (2026) · Research · doi:10.34133/research.1192 · Cited by 4
“Long-term skin-conformal wear also faces challenges from sweat, friction, temperature, and humidity fluctuations, as well as repeated donning and doffing [ 45 , 122 , 123 ].”
Zohra, F. T., Abdullah, A. M., Yuan, Y., Wang, Y., Lin, J., Zhang, W., Shahed, K. S., Zhang, X., Biswas, M. A. S., Li, B., et al. (2026) · Science Advances · doi:10.1126/sciadv.aee5890 · Cited by 4
“…detection of biosignals from muscles, eyes, brain, and heart with epidermal bioelectrodes can provide enhanced diagnostic accuracy for mental health monitoring in personalized and user-friendly settings ( 3 , 4 ) as these physiological activities are interrelated with each other ( 5 , 6 ).”
Wearable patterned nanocomposite circuits for temperature compensated strain sensing
Lin, Y., & Loh, K. J. (2025) · Smart Materials and Structures · doi:10.1088/1361-665x/ade5dd · Cited by 4
“The skin, being the largest organ in the human body, is not only a convenient surface for sensor attachment but is also the biochemical and mechanical interface between the person and environment where relevant sensing streams can be acquired [1, 2].”
Ni, K., Shao, Q., Wang, K., & Wang, Z. (2021) · Mechanics of Advanced Materials and Structures · doi:10.1080/15376494.2021.2006375 · Cited by 4
“Inspired by this geometry-dominated adhesion principle, many researchers have developed microfibrillar structures and investigated their potential applications, such as wall-climbing robots [3, 4], transfer printing [5–7], and on-skin electrodes [8, 9].”
BackgroundZhao, Z., Yang, C., & Li, D. (2024) · Micromachines · doi:10.3390/mi15050598 · Cited by 4
Low-temperature flexible composite preform based on silver nanowires aerogel and SnBi58 solder
Fu, B., Li, X., Peng, Z., He, J., Fang, Y. W., Dong, B., Hu, Q., Zhu, W., & Li, M. (2025) · Journal of Materials Research and Technology · doi:10.1016/j.jmrt.2025.02.218 · Cited by 4
Laser Beam Machining of Janus Porous Electrodes for Long‐term Electrophysiological Sensing
Zhou, Y., Huang, Z., Chen, Y., Dong, X., Fan, F., Chen, L., & Feng, B. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202401812 · Cited by 4
Continuous Electrochemical Glucose Detection Using Ferrocene Carboxylic Acid Chemistry
Nayak, K. K., & Bhat, N. (2024) · IEEE Applied Sensing Conference (APSCON) · doi:10.1109/apscon60364.2024.10466091 · Cited by 4
Kim, H., Jo, Y., Lee, Y., Cho, E., Kim, J., & Shin, D. (2025) · Journal of Applied Polymer Science · doi:10.1002/app.58011 · Cited by 3
Gao, Y. K., Yang, X.-C., Shang, B. Z., & Qiao, S. (2025) · ACS Applied Polymer Materials · doi:10.1021/acsapm.5c03258 · Cited by 3
Xu, A., Xiao, M., Sui, Z., Wang, Y., Zheng, D., Liu, Y., Wang, L., Liu, H., & Lee, C. (2026) · Advanced Materials Technologies · doi:10.1002/admt.202502093 · Cited by 3
Advances in MXene‐Based Electrochemical Sensors for Environmental Monitoring of Heavy Metal Ions
Wannassi, J., Kahri, H., & Barhoumi, H. (2026) · Electroanalysis · doi:10.1002/elan.70136 · Cited by 3
Pang, B., Yang, G., Wen, J., Gong, H., Gong, C., Qiu, Y., Wang, Z., Zheng, Q., Zhou, S., Bu, T., et al. (2026) · Advanced Science · doi:10.1002/advs.76396 · Cited by 3
“These signals are captured by recording electrodes at the target site [ 10– 12 ].”
Self-citationZhang, W., Xin, X., Wang, Y., Zhang, X., Zhang, S., Yuan, Y., Xia, S., Abdullah, A. M., Zohra, F. T., Li, B., et al. (2026) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2615835123 · Cited by 3
Biomimetic Bilayer‐Structured CNF/GO‐Reinforced CS–PAA Hydrogel for Flexible Strain Sensors
Liu, T., Wang, G., Wang, M., Liu, C., Lv, M., Liu, H., Li, C., & Xu, C. (2026) · Advanced Functional Materials · doi:10.1002/adfm.77646 · Cited by 3
“Human skin exhibits functionally graded mechanical properties: the epidermis, responsible for tactile sensing, has a Young’s modulus of 0.1–0.5 MPa and an elongation at break of 100%– 150%, while the dermis, providing mechanical support, shows 0.5–1 MPa and 150%–200%, respectively [ 41–43 ].”
InfluentialSmart Prosthetics Controller Types: Review
Ahmed, A., Mashhadany, Y. I. M. A., Ahmad, R. B., & khaleefah, F. (2024) · Anbar journal for engineering sciences/Mağallaẗ al-anbār li-l-ʻulūm al-handasiyyaẗ · doi:10.37649/aengs.2024.152744.1097 · Cited by 3
Patterning Adhesive Layers for Array Electrodes via Electrochemically Grafted Polymers
Wen, S., Zhang, R., Zhao, Y., Xu, X., & Ji, S. (2025) · ACS Omega · doi:10.1021/acsomega.4c10830 · Cited by 3
Assessing contact impedance of embroidered textrodes for biopotential measurements
Komal, K., Cleary, F., Wells, J. S. G., Narayanan, R. P., & Bennett, L. (2025) · The Design Journal · doi:10.1080/14606925.2025.2565156 · Cited by 3
“As the human body’s curvature can prevent the traditional 2D embroidery design of textrodes from maintaining a uniform pressure, this may degrade signal quality (Wu et al. 2021).”
Alcan, V., & Özkendir, O. M. (2021) · Materialia · doi:10.1016/j.mtla.2021.101170 · Cited by 3
Liu, G., Xu, C., Ren, J., Liu, Y., Yao, L., Xie, Y., Zhang, T., & Liu, Y. (2023) · Journal of Materials Science Materials in Electronics · doi:10.1007/s10854-023-10599-0 · Cited by 3
Chansaengsri, K., Tunhoo, B., Onlaor, K., & Thiwawong, T. (2024) · IEEE Open Journal of Engineering in Medicine and Biology · doi:10.1109/ojemb.2024.3431030 · Cited by 3
“Wearable health care sensing devices as dry flexible electrodes have been classified into several types depending on the substrate such as rigid, bulky devices [10], tattoo-like skin devices [11], flexible devices with compression [12], implantation [13], and application for textile device [14].”
Wang, Z., Yi, H., & Li, Z. (2024) · National Science Open · doi:10.1360/nso/20240025 · Cited by 3
Self-citationSpray-Coated Ultrathin and Porous Films for Physiological Sensing and Force Detection
Li, T., Ding, Y., Teng, C., Zheng, Y., Wang, X., & Zhou, D. (2024) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c11179 · Cited by 3
Functional Hydrogels for Wearable Electronics
Zameer, A., Qin, Y., Xu, H., Liang, Q., & Zhang, Q. M. (2025) · Macromolecular Chemistry and Physics · doi:10.1002/macp.202400491 · Cited by 3
Preparation and application of self-adhesive ultra-thin epidermal electrophysiological electrodes
Hu, W., Song, D., Shi, X., & Liu, N. (2022) · Scientia Sinica Chimica · doi:10.1360/ssc-2022-0019 · Cited by 3
A Skin Sensor for Epileptic Seizure Detection and Notification Applications
Leon, L., Contreras, N., Polanco, F. C., Fentry, E., Sajal, S., & Parvez, I. (2023) · IEEE SoutheastCon · doi:10.1109/southeastcon51012.2023.10114958 · Cited by 3
“By measuring the low-level electrical potential in the skin and comparing its values over time we can differentiate the normal status of the user against an epileptic event [16].”
BackgroundXue, K., Hao, T., Zhang, C., Ding, J., Chen, X., Wang, J., He, L., Wang, M., & Xu, J. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3634355 · Cited by 2
“However, conventional ECG monitoring typically relies on rigid Ag/AgCl wet electrodes, which are unsuitable for long-term and dynamic use due to issues such as poor comfort, gel dehydration [5], and susceptibility to skin irritation associated with wet electrodes.”
Clinical use of electrodermal activity and current developments
Kara, A. Y., Dolu, N., & Esen, F. (2026) · European Biophysics Journal · doi:10.1007/s00249-026-01820-z · Cited by 2
“Novel dry electrodes incorporating conductive polymers have yielded enhanced conductivity, self-adhe-siveness, and mechanical flexibility that improve comfort and signal quality during prolonged wear (Zhang et al. 2020; Wu et al. 2021).”
Cheng, S., Chen, L., Gu, T., Qiu, B., Pan, Z., Li, W., Zhu, J., Xie, R., Wu, J., & Huo, F. (2026) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.6c07695 · Cited by 2
Integrated Sensing Devices for Brain-Computer Interfaces
Do, T., Duong, N. M. H., & Lin, C. (2023) · Springer eBooks · doi:10.1007/978-3-031-21610-7_7 · Cited by 2
Miyamoto, K., Miller, R. M., Voors‐Pette, C., Oosterhaven, J. A. F., van den Dobbelsteen, M., Mihara, K., Geldof, M., Sato, Y., Matsuda, N., Kirita, S., et al. (2024) · Clinical and Translational Science · doi:10.1111/cts.70060 · Cited by 2
Dry Film Photoresist Enabled Sustainable Large-Area Manufacturing for Flexible Electronics
Wang, Q., Chen, L., Bi, X., Dong, X., Xie, J., Feng, B., & Duan, H. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c09006 · Cited by 2
A flexible 3D force sensor for handwriting recognition
Zhao, M., Geng, J., Yan, J., Chen, X., & Nie, B. (2021) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp-bmei53629.2021.9624406 · Cited by 2
A User-customized and Untethered Electro-haptic Device for Immersive Human-Machine InteractionPreprint
Cui, Z., Wang, S., Zhao, Y., Wang, Y., Wen, X., Chen, S., & Xiong, Z. (2025) · arXiv (Cornell University) · doi:10.48550/arxiv.2511.20578 · Cited by 2
“In similar works done before, metal electrodes were widely used for sensing and stimulation on the skin surface [79].”
Soft, wearable devices to monitor electrophysiological signals and gaseous biomarkers
Herbert, R., Popoola, M. O., Zhang, W., & Cheng, H. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-99147-6.00006-5 · Cited by 1
Reliable transfer enabled by UV-curable stamp with tunable rigidity
Chen, L., Niu, Y., Liu, C., Fan, R., Liu, P., Ma, D., Zhang, X., Liu, C., & Duan, H. (2024) · Surfaces and Interfaces · doi:10.1016/j.surfin.2024.105348 · Cited by 1
Du, X., Wu, L., Yu, R., Fan, Z., Chen, Y., Zhou, T., Yang, Z., Liu, Y., Zhu, M., & Pan, S. (2025) · ACS Sensors · doi:10.1021/acssensors.5c02247 · Cited by 1
Raju, R., Rajanna, K., Yashaswini, V. L., Prasad, A. S., Mohan, K. M., Swamy, N. K., Norek, M., Madhukar, B. S., & Shivarudraiah, V. H. (2025) · Polymers for Advanced Technologies · doi:10.1002/pat.70426 · Cited by 1
Savelyev, M. S., Kuksin, A. V., Murashko, D. T., Otsupko, E. P., Suchkova, V. V., Popovich, K. D., Vasilevsky, P. N., Kurilova, U. E., Selishchev, S. V., & Gerasimenko, A. Y. (2025) · Gels · doi:10.3390/gels11121004 · Cited by 1
“…tissue (metals, as a rule, cannot deform by more than 5% without failure, which affects their functionality and long-term use), and ensure a tight fit considering the mobility of nervous tissues (the material must be able to stretch while maintaining its shape without breaking adhesion) [5,18,23].”
Huang, T., Li, L., Liang, J., Zhu, Z., Yang, Z., & Xi, Y. (2026) · Analytical Methods · doi:10.1039/d5ay01980j · Cited by 1
Zhang, X., Deng, R., Du, J., Ma, Q., Li, D., Zhang, B., Jian, N., & Liu, G. (2026) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2026.117775 · Cited by 1
Capacitive Touch Sensing-Based Human–Machine Interface for Upper-Limb Prosthetics
Jan, B., Ehshan, K. M., Abbas, Z., Venkataraman, S., Panigrahi, B. K., & Malik, S. (2026) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2026.3697066 · Cited by 1
“Nevertheless, a critical bottleneck that persists in prosthetic HMIs is the dependence on direct skin-electrode adhesion [30], [31].”
Zhuang, T., Yao, Y., Hao, X., Zhu, Z., Chen, K., Liu, Y., Wu, Q., Yin, F., & Wang, H. (2026) · Small · doi:10.1002/smll.74045 · Cited by 2
Research progress on MXene-based electrophysiological electrodes
Xinyun, L. I. N., Wenya, H. U., Song, D., & Nan, L. I. U. (2026) · Gongneng cailiao yu qijian xuebao · doi:10.3724/jfmd.2511093 · Cited by 1
Flexible Sensing Electrodes: From Materials Design to Device Fabrication
Zhang, H., Wang, Y., Duan, X., Zhang, M., Han, L., Mahmud, S., & Zhang, R. (2026) · Materials · doi:10.3390/ma19143036 · Cited by 1
“However, achieving the simultaneous optimization of these often conflicting properties remains a major challenge [5,6].”
Xu, K., Ke, K., Hu, Y., Guo, J.-X., Niu, S.-J., Wang, Y., Yin, B., Jiang, Y., Yang, M., & Yang, W. M. (2026) · Progress in Materials Science · doi:10.1016/j.pmatsci.2026.101786 · Cited by 1
Auricular Electrophysiology Demonstrates Enhanced Stability Under Vehicle Vibration and Acceleration
Orsini, A., Liu, C., Shen, B., Cao, M., Xu, H., Ananthakumar, T., Cho, S., & Thakor, N. V. (2026) · Military Medicine · doi:10.1093/milmed/usag045 · Cited by 1
Ni, J., Lu, J., Wu, N., Tang, J., Chen, T., Zhang, X., Wang, W., & Zhao, Z. (2026) · Rare Metals · doi:10.1002/rar2.70499 · Cited by 1
Clear Patch: Transparent Wearable Sensors for Human and Plant Health Monitoring
Zare, M., Jamalzadegan, S., Poonam, A. D., Jaseem, S. A., Kim, J. Y., Xu, J., Choi, H., Atar, N., Chen, B., Dickey, M. D., et al. (2026) · ACS Nano · doi:10.1021/acsnano.6c06211 · Cited by 1
Xie, Y., Liu, G., Ren, J., Liu, Y., Yao, L., Xu, B., Chen, D., & Liu, Y. (2023) · Journal of Physics Conference Series · doi:10.1088/1742-6596/2463/1/012019 · Cited by 1
“For that, a one-mask micro-fabrication method was developed for a cost-down objective [6].”
MethodNcube, T., Olawoore, P., Maphosa, P., Mogashoa, S., Aljohani, F. S., & Swanepoel, M. (2024) · Minerals Engineering · doi:10.1016/j.mineng.2024.109128 · Cited by 1
Evaluation of Patchable Inkjet-Printed Wearable Sensors for Measuring Muscle Activities
Lim, J., Zhang, H., Sun, M., Lin, H., Jia, B., Liu, J. Z., & Tan, Y. (2024) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/sensors60989.2024.10784475 · Cited by 1
“There have been attempts to develop sensors in various form factors, such as electronic tattoo [3], electronic skin [4], and electronic textile [5], to enhance usability in everyday life [6].”
Systematic Investigation of Heart Sound Propagation Using Continuous Wave Radar
Oesten, M., Abel, L., Albrecht, N. C., Richer, R., Langer, D., Grießhammer, S. G., Ghanem, K., Steigleder, T., Ostgathe, C., Koelpin, A., et al. (2025) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2025.3530821 · Cited by 1
“Besides the limitations on patient autonomy and participation in daily activities, contact-based methods encounter several challenges, such as skin irritations from long-term measurements with electrodes [13] or sensor placement for patients with burns [14].”
Opu, M. M. R., & Ruqnuzzaman, M. (2026) · Research Square · doi:10.21203/rs.3.rs-8464801/v1 · Cited by 1
A Facile Bonding Material to Enable Interconnection among complex Surfaces through AgNWs Aerogel
Han, Z., Li, M., Zhu, W., Peng, F., & Fang, Y. (2021) · International Conference on Electronic Packaging Technology (ICEPT) · doi:10.1109/icept52650.2021.9567966
Nanoparticle Chain based Materials for Shielding and Flexible Devices
Fan, H., Boeuf, W. L., & Maheshwari, V. (2022) · IEEE International Conference on Flexible and Printable Sensors and Systems (FLEPS) · doi:10.1109/fleps53764.2022.9781562
Minimally-invasive and non-invasive flexible devices for robust characterizations of deep tissues
Yuan, Y., Tang, W., Wu, H., Chen, S., Ren, X., Hu, B., & Gu, N. (2022) · Scientia Sinica Chimica · doi:10.1360/ssc-2022-0100
Self-citationHighly adhesive stretchable polymer and highly dynamic stable human electrophysiological monitoring
Hu, M., & Yang, H. (2022) · Scientia Sinica Chimica · doi:10.1360/ssc-2022-0155
Rahmawati, F., Wijayadi, L. J., & Sari Mariyati Dewi, N. (2022) · Science Midwifery · doi:10.35335/midwifery.v10i4.746
“Data collection was carried out by filling out questionnaires of the research subjects after which the examination of skin hydration levels using the SK-IV corneometer/ It is able to provide efficient and sensitive results for skin hydration measurements in four skin areas, namely the back of the left right hand, and…”
MethodSung, M. (2023) · International Journal of Students Research in Technology & Management · doi:10.18510/ijsrtm.2023.1113
Devices based on the ECG signal
Bolić, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-820947-9.00001-5
Mishra, R. K., Joseph, K., Chianella, I., Nezhad, H. Y., & Goel, S. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-24861-0.00005-8
Analysis of the Possibilities of Using the EOG Signal in Controlling a Mechanical Device
Kot, A., Nawrocka, A., & Perciun, A. (2025) · International Carpathian Control Conference · doi:10.1109/iccc65605.2025.11022884
“Electrodes around the eye sockets detect voltage changes, which are processed and interpreted as commands for controlling an external device [3].”
Vellus Hair-Inspired Triboelectric Antenna for Approach and Pressure Sensing
Shen, L., Jiang, J., Lei, H., Lu, B., Zhang, P., Miao, D., Sun, X., & Wen, Z. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c12911
Soft Sensors for Disease Diagnosis
Hu, H., Su, Y., & Nan, K. (2025) · Wiley eBooks · doi:10.1002/9783527845507.ch3
Spray‐Coated Fully Water‐Soluble Composites for Waterproof Breathable E‐Tattoos
Zhu, Y., Dai, L., Hu, M., Wang, F., Guo, H., Zhao, X., & Yang, H. (2025) · Advanced Engineering Materials · doi:10.1002/adem.202501149
Mechanical characterization of the human auricle for earable device design
Li, Y., Bance, M., & Sutcliffe, M. P. F. (2025) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2025.107181
Zhang, Y., Li, Y., Zhang, J., Zheng, Y., He, Y., He, Y., Yang, J., Li, L., Xiong, P., Zhang, J., et al. (2025) · Journal of Science Advanced Materials and Devices · doi:10.1016/j.jsamd.2025.101016
Smart wearable TENS device for overactive bladder management
Ju, W. (2025) · ERA · doi:10.7488/era/6273
Estrada Petrocelli, L., Torres, A., Rodríguez, Y., Caballero, C., Molino, J., & Ibarra Ramírez, E. (2022) · Gente Clave
Giangrande, A. (2025) · PORTO@Iris · doi:10.13121/polito/porto/2997444
Hartel, M. C. (2023) · eScholarship (California Digital Library)
Inkjet-Printed ECG Electrodes and AI-based Data Reliability for Smart-Health CPS
Momota, M. M. R. (2023) · ThinkTech (Texas Tech University)
Yang, L., Bi, Y., Xia, S., & Gao, G. (2025) · ACS Applied Polymer Materials · doi:10.1021/acsapm.5c03606
Design and applications of conductive polymers in wearable electrophysiological sensing
Zhou, N., Zhang, L., Cui, T., Wu, P., & Lei, Z. (2026) · Scientia Sinica Technologica · doi:10.1360/sst-2025-0347
Fang, D., Yin, Z., Zhou, Q., Peng, C., Ji, B., Wu, J., & Zhou, B. (2026) · Materials Today Chemistry · doi:10.1016/j.mtchem.2026.103424
E-textile sensor for monitoring and evaluation of atopic dermatitis in skin
Todorov, A. R. (2026) · ePrints Soton (University of Southampton)
Naz, T., Hussain, S. M., & Rizwan, M. (2025) · Pakistan Journal of Engineering Technology & Science · doi:10.22555/pjets.v13i2.1451
Kim, H., Saha, S., Wachnin, A. J., Ban, S., Lee, Y., Kwon, Y., Lee, J., Chhabra, I., Ram, S., Nyan, C., et al. (2026) · JMIR Formative Research · doi:10.2196/80286
“Other wearable systems with smaller form factors, such as those worn around or in the ear, either have to be customized for an individual for optimal fit [39] or contain wires [21,39], which could contribute to relatively low signal quality due to mechanical noise [40].”
Metadata
- Authors
- Wu, H., Yang, G., Zhu, K., Liu, S., Guo, W., Jiang, Z., & Li, Z.
- Affiliations
- State Key Laboratory of Digital Manufacturing Equipment and Technology (CN)
- Fudan University (CN)
- Corresponding author
- Wu, Hao
- Li, Zhuo
- Venue
- Advanced Science
- Publisher
- Wiley
- Published
- 2020-12-04
- Volume
- 8
- Issue
- 2
- Pages
- 2001938-2001938
- ISSN
- 2198-3844 (ISSN-L)
- PMC
- PMC7816724
- OpenAlex
- W3110558114
- Cited by
- 395
- References
- 233
- Citations by year
- 2026: 56 · 2025: 95 · 2024: 77 · 2023: 69 · 2022: 59 · 2021: 38 · 2020: 1
- FWCI
- 3.49
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Advanced Science (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- National Natural Science Foundation of China (51820105008, 91648115)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Nanomaterials and Printing Technologies
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Keywords
- electrophysiological monitoring, human-machine interaction, wearable sensors, stretchable materials, biocompatibility, electrode adhesion, breathability, electrode materials, healthcare monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.