Calibration and standardization of bone mineral densitometers
The DOI matches its Crossref record.
Abstract
Measurements of bone mineral density (BMD) by four 153Gd and two x-ray bone densitometers were compared utilizing spine phantoms that simulated the human lumbar spine. The six instruments provided BMD values that differed by as much as 16%, due to differences as large as 8% in bone mineral content and as large as 7% in bone area. Instrument calibration curves, determined by measuring thin, medium, and thick hydroxyapatite blocks, were linear (r = 0.99) but had different (p < 0.0001) slopes and intercepts. Serial measurements of spine and total body phantoms were employed to evaluate the long-term stability of 153Gd bone densitometry. These measurements of spine phantom BMD increased 1.0–2.6% (p < 0.0001) following a software change, and measurements of total body bone density increased 4–6% after 153Gd source replacement. The changes occurred at a time when serial measurements of cylindrical calibration standards were stable, indicating that such simple standards are unable to detect and correct for changes in instrumental response. We conclude that investigators, manufacturers, and government regulatory agencies must develop and implement the following: (1) effective calibration procedures that would assure comparability among instruments, and (2) appropriate quality control phantoms that would allow the confident interpretation of serial patient measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
TSH Is a Negative Regulator of Skeletal Remodeling
Abe, E., Marians, R. C., Yu, W., Wu, X.-B., Ando, T., Li, Y., Iqbal, J., Eldeiry, L., Rajendren, G., Blair, H. C., et al. (2003) · Cell · doi:10.1016/s0092-8674(03)00771-2 · Cited by 691
Kalender, W. A., Felsenberg, D., Genant, H. K., Fischer, M., Dequeker, J., & Reeve, J. (1995) · European Journal of Radiology · doi:10.1016/0720-048x(95)00631-y · Cited by 263
Horber, F. F., Thomi, F., Casez, J. P., Fonteille, J., & Jaeger, P. (1992) · The British Journal of Radiology · doi:10.1259/0007-1285-65-778-895 · Cited by 166
Heymsfield, S. B., Ebbeling, C. B., Zheng, J., Pietrobelli, A., Strauss, B. J., Silva, A. M., & Ludwig, D. S. (2015) · Obesity Reviews · doi:10.1111/obr.12261 · Cited by 111
Teigen, L. M., Kuchnia, A. J., Mourtzakis, M., & Earthman, C. P. (2017) · Nutrition in Clinical Practice · doi:10.1177/0884533616676264 · Cited by 75
Metadata
- Authors
- Kelly, T. L., Slovik, D. M., & Neer, R. M.
- Venue
- Journal of Bone and Mineral Research
- Publisher
- Oxford University Press (OUP)
- Published
- 1989-10-01
- Volume
- 4
- Issue
- 5
- Pages
- 663-669
- ISSN
- 0884-0431 (ISSN-L)
- 1523-4681
- OpenAlex
- W2045601732
- Cited by
- 61
- References
- 0
- FWCI
- 4.13
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Calibration and standardization of bone mineral densitometers. https://app.cochat.ai/papers/10.1002/jbmr.5650040504
- The paper
- Kelly, T. L., Slovik, D. M., & Neer, R. M. (1989). Calibration and standardization of bone mineral densitometers. Journal of Bone and Mineral Research. https://doi.org/10.1002/jbmr.5650040504
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.