Microindentation for in vivo measurement of bone tissue mechanical properties in humans
Open accessAbstract
Bone tissue mechanical properties are deemed a key component of bone strength, but their assessment requires invasive procedures. Here we validate a new instrument, a reference point indentation (RPI) instrument, for measuring these tissue properties in vivo. The RPI instrument performs bone microindentation testing (BMT) by inserting a probe assembly through the skin covering the tibia and, after displacing periosteum, applying 20 indentation cycles at 2 Hz each with a maximum force of 11 N. We assessed 27 women with osteoporosis-related fractures and 8 controls of comparable ages. Measured total indentation distance (46.0 ± 14 versus 31.7 ± 3.3 µm, p = .008) and indentation distance increase (18.1 ± 5.6 versus 12.3 ± 2.9 µm, p = .008) were significantly greater in fracture patients than in controls. Areas under the receiver operating characteristic (ROC) curve for the two measurements were 93.1% (95% confidence interval [CI] 83.1–100) and 90.3% (95% CI 73.2–100), respectively. Interobserver coefficient of variation ranged from 8.7% to 15.5%, and the procedure was well tolerated. In a separate study of cadaveric human bone samples (n = 5), crack growth toughness and indentation distance increase correlated (r = –0.9036, p = .018), and scanning electron microscope images of cracks induced by indentation and by experimental fractures were similar. We conclude that BMT, by inducing microscopic fractures, directly measures bone mechanical properties at the tissue level. The technique is feasible for use in clinics with good reproducibility. It discriminates precisely between patients with and without fragility fracture and may provide clinicians and researchers with a direct in vivo measurement of bone tissue resistance to fracture. © 2010 American Society for Bone and Mineral Research
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
An introduction to ROC analysis
Fawcett, T. (2006) · Pattern Recognition Letters · doi:10.1016/j.patrec.2005.10.010 · Cited by 21,879
Osteoporosis Prevention, Diagnosis, and Therapy
NIH Consensus Development Panel on Osteoporosis Prevention, D. (2001) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.285.6.785 · Cited by 5,529
Bone Quality — The Material and Structural Basis of Bone Strength and Fragility
Seeman, E., & Delmas, P. D. (2006) · New England Journal of Medicine · doi:10.1056/nejmra053077 · Cited by 2,080
Structure and mechanical quality of the collagen–mineral nano-composite in bone
Fratzl, P., Gupta, H. S., Paschalis, E. P., & Roschger, P. (2004) · J. Mater. Chem. · doi:10.1039/b402005g · Cited by 1,222
Fantner, G. E., Hassenkam, T., Kindt, J. H., Weaver, J. C., Birkedal, H., Pechenik, L., Cutroni, J. A., Cidade, G. A. G., Stucky, G. D., Morse, D. E., et al. (2005) · Nature Materials · doi:10.1038/nmat1428 · Cited by 933
Self-citationSaito, M., & Marumo, K. (2010) · Osteoporosis International · doi:10.1007/s00198-009-1066-z · Cited by 906
Cummings, S. R., Karpf, D. B., Harris, F., Genant, H. K., Ensrud, K., LaCroix, A. Z., & Black, D. M. (2002) · The American Journal of Medicine · doi:10.1016/s0002-9343(01)01124-x · Cited by 752
Mechanical Testing of Bone and the Bone-Implant Interface
(1999) · CRC Press eBooks · doi:10.1201/9781420073560 · Cited by 709
Garnero, P., Sornay-Rendu, E., Claustrat, B., & Delmas, P. D. (2000) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.2000.15.8.1526 · Cited by 709
The true toughness of human cortical bone measured with realistically short cracks
Koester, K. J., Ager, J. W., & Ritchie, R. O. (2008) · Nature Materials · doi:10.1038/nmat2221 · Cited by 558
Self-citation
Cited by
In Vivo Assessment of Bone Quality in Postmenopausal Women With Type 2 Diabetes
Farr, J. N., Drake, M. T., Amin, S., Melton, L. J., McCready, L. K., & Khosla, S. (2014) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.2106 · Cited by 499
Harvey, N. C., Glüer, C. C., Binkley, N., McCloskey, E. V., Brandi, M.-L., Cooper, C., Kendler, D., Lamy, O., Laslop, A., Camargos, B. M., et al. (2015) · Bone · doi:10.1016/j.bone.2015.05.016 · Cited by 486
Epidemiology of fracture risk with advancing age
Ensrud, K. E. (2013) · The Journals of Gerontology Series A: Biological Sciences and Medical Sciences · doi:10.1093/gerona/glt092 · Cited by 456
Osteoporosis: the evolution of a diagnosis
Lorentzon, M., & Cummings, S. R. (2015) · Journal of Internal Medicine · doi:10.1111/joim.12369 · Cited by 379
Systemic Complications of Acromegaly and the Impact of the Current Treatment Landscape: An Update
Gadelha, M. R., Kasuki, L., Lim, D. S. T., & Fleseriu, M. (2019) · Endocrine Reviews · doi:10.1210/er.2018-00115 · Cited by 363
Metadata
- Authors
- Diez-Perez, A., Güerri, R., Nogues, X., Cáceres, E., Peña, M. J., Mellibovsky, L., Randall, C., Bridges, D., Weaver, J. C., Proctor, A., Brimer, D., Koester, K. J., Ritchie, R. O., & Hansma, P. K.
- Venue
- Journal of Bone and Mineral Research
- Publisher
- Oxford University Press (OUP)
- Published
- 2010-07-30
- Volume
- 25
- Issue
- 8
- Pages
- 1877-1885
- ISSN
- 0884-0431 (ISSN-L)
- 1523-4681
- OpenAlex
- W2035965297
- Cited by
- 269
- References
- 46
- FWCI
- 12.66
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Microindentation for in vivo measurement of bone tissue mechanical properties in humans. https://app.cochat.ai/papers/10.1002/jbmr.73
- The paper
- Diez-Perez, A., Güerri, R., Nogues, X., Cáceres, E., Peña, M. J., Mellibovsky, L., Randall, C., Bridges, D., Weaver, J. C., Proctor, A., Brimer, D., Koester, K. J., Ritchie, R. O., & Hansma, P. K. (2010). Microindentation for in vivo measurement of bone tissue mechanical properties in humans. Journal of Bone and Mineral Research. https://doi.org/10.1002/jbmr.73
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.