Student or educator? Sign up with your school email and get 6 months of Pro free.
Cited by 248

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Quantification of hepatic steatosis with MRI: The effects of accurate fat spectral modeling

Open access
Reeder, Scott B., Robson, Philip M., Yu, Huanzhou, Shimakawa, Ann, Hines, Catherine DG, McKenzie, Charles A., Brittain, Jean H.·2009·Journal of Magnetic Resonance Imaging

Abstract

Purpose: To develop a chemical‐shift–based imaging method for fat quantification that accounts for the complex spectrum of fat, and to compare this method with MR spectroscopy (MRS). Quantitative noninvasive biomarkers of hepatic steatosis are urgently needed for the diagnosis and management of nonalcoholic fatty liver disease (NAFLD). Materials and Methods: Hepatic steatosis was measured with “fat‐fraction” images in 31 patients using a multiecho chemical‐shift–based water‐fat separation method at 1.5T. Fat‐fraction images were reconstructed using a conventional signal model that considers fat as a single peak at –210 Hz relative to water (“single peak” reconstruction). Fat‐fraction images were also reconstructed from the same source images using two methods that account for the complex spectrum of fat; precalibrated and self‐calibrated “multipeak” reconstruction. Single‐voxel MRS that was coregistered with imaging was performed for comparison. Results: Imaging and MRS demonstrated excellent correlation with single peak reconstruction (r 2= 0.91), precalibrated multipeak reconstruction (r 2= 0.94), and self‐calibrated multipeak reconstruction (r 2= 0.91). However, precalibrated multipeak reconstruction demonstrated the best agreement with MRS, with a slope statistically equivalent to 1 (0.96 ± 0.04;P= 0.4), compared to self‐calibrated multipeak reconstruction (0.83 ± 0.05,P= 0.001) and single‐peak reconstruction (0.67 ± 0.04,P< 0.001). Conclusion: Accurate spectral modeling is necessary for accurate quantification of hepatic steatosis with MRI. J. Magn. Reson. Imaging 2009;29:1332–1339. © 2009 Wiley‐Liss, Inc.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/jmri.21751

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 248

Metadata

Authors
Reeder, S. B., Robson, P. M., Yu, H., Shimakawa, A., Hines, C. D., McKenzie, C. A., & Brittain, J. H.
Venue
Journal of Magnetic Resonance Imaging
Publisher
Wiley
Published
2009-05-26
Volume
29
Issue
6
Pages
1332-1339
ISSN
  • 1053-1807 (ISSN-L)
  • 1522-2586
OpenAlex
W2162992000
Cited by
248
References
2
FWCI
13.62
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Quantification of hepatic steatosis with MRI: The effects of accurate fat spectral modeling. https://app.cochat.ai/papers/10.1002/jmri.21751
The paper
Reeder, S. B., Robson, P. M., Yu, H., Shimakawa, A., Hines, C. D., McKenzie, C. A., & Brittain, J. H. (2009). Quantification of hepatic steatosis with MRI: The effects of accurate fat spectral modeling. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.21751

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.