Multi‐system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice‐water phantom
Open accessAbstract
Purpose: To determine quantitative quality control procedures to evaluate technical variability in multi‐center measurements of the diffusion coefficient of water as a prerequisite to use of the biomarker apparent diffusion coefficient (ADC) in multi‐center clinical trials. Materials and Methods: A uniform data acquisition protocol was developed and shared with 18 participating test sites along with a temperature‐controlled diffusion phantom delivered to each site. Usable diffusion weighted imaging data of ice water at five b‐values were collected on 35 clinical MRI systems from three vendors at two field strengths (1.5 and 3 Tesla [T]) and analyzed at a central processing site. Results: Standard deviation of bore‐center ADCs measured across 35 scanners was <2%; error range: −2% to +5% from literature value. Day‐to‐day repeatability of the measurements was within 4.5%. Intra‐exam repeatability at the phantom center was within 1%. Excluding one outlier, inter‐site reproducibility of ADC at magnet isocenter was within 3%, although variability increased for off‐center measurements. Significant (>10%) vendor‐specific and system‐specific spatial nonuniformity ADC bias was detected for the off‐center measurement that was consistent with gradient nonlinearity. Conclusion: Standardization of DWI protocol has improved reproducibility of ADC measurements and allowed identifying spatial ADC nonuniformity as a source of error in multi‐site clinical studies. J. Magn. Reson. Imaging 2013;37:1238–1246. © 2012 Wiley Periodicals, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
On behalf of the EUSOBI international Breast Diffusion-Weighted Imaging working group, Baltzer, P., Mann, R. M., Iima, M., Sigmund, E. E., Clauser, P., Gilbert, F. J., Martincich, L., Partridge, S. C., Patterson, A., et al. (2020) · European Radiology · doi:10.1007/s00330-019-06510-3 · Cited by 437
Shukla‐Dave, A., Obuchowski, N. A., Chenevert, T. L., Jambawalikar, S., Schwartz, L. H., Malyarenko, D., Huang, W., Noworolski, S. M., Young, R. J., Shiroishi, M. S., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26518 · Cited by 389
Self-citationKessler, L. G., Barnhart, H. X., Buckler, A. J., Choudhury, K. R., Kondratovich, M. V., Toledano, A., Guimaraes, A. R., Filice, R., Zhang, Z., Sullivan, D. C., et al. (2015) · Statistical Methods in Medical Research · doi:10.1177/0962280214537333 · Cited by 293
Multi-centre reproducibility of diffusion MRI parameters for clinical sequences in the brain
Grech‐Sollars, M., Hales, P. W., Miyazaki, K., Raschke, F., Rodriguez, D., Wilson, M., Gill, S. K., Banks, T., Saunders, D. E., Clayden, J. D., et al. (2015) · NMR in Biomedicine · doi:10.1002/nbm.3269 · Cited by 213
Diffusion-MRI in neurodegenerative disorders
Goveas, J., O’Dwyer, L., Mascalchi, M., Cosottini, M., Diciotti, S., De Santis, S., Passamonti, L., Tessa, C., Toschi, N., & Giannelli, M. (2015) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2015.04.006 · Cited by 104
Metadata
- Authors
- Malyarenko, D., Galbán, C. J., Londy, F. J., Meyer, C. R., Johnson, T. D., Rehemtulla, A., Ross, B. D., & Chenevert, T. L.
- Venue
- Journal of Magnetic Resonance Imaging
- Publisher
- Wiley
- Published
- 2012-09-28
- Volume
- 37
- Issue
- 5
- Pages
- 1238-1246
- ISSN
- 1053-1807 (ISSN-L)
- 1522-2586
- OpenAlex
- W2024314995
- Cited by
- 197
- References
- 0
- FWCI
- 10.24
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Multi‐system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice‐water phantom. https://app.cochat.ai/papers/10.1002/jmri.23825
- The paper
- Malyarenko, D., Galbán, C. J., Londy, F. J., Meyer, C. R., Johnson, T. D., Rehemtulla, A., Ross, B. D., & Chenevert, T. L. (2013). Multi‐system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice‐water phantom. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.23825
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.