Signal‐to‐noise ratio and parallel imaging performance of a 16‐channel receive‐only brain coil array at 3.0 Tesla

Open access
de Zwart, Jacco A., Ledden, Patrick J., van Gelderen, Peter, Bodurka, Jerzy, Chu, Renxin, Duyn, Jeff H.·2004·Magnetic Resonance in Medicine

Abstract

The performance of a 16‐channel receive‐only RF coil for brain imaging at 3.0 Tesla was investigated using a custom‐built 16‐channel receiver. Both the image signal‐to‐noise ratio (SNR) and the noise amplification (g‐factor) in sensitivity‐encoding (SENSE) parallel imaging applications were quantitatively evaluated. Furthermore, the performance was compared with that of hypothetical coils with one, two, four, and eight elements (n) by combining channels in software during image reconstruction. As expected, both the g‐factor and SNR improved substantially with n. Compared to an equivalent (simulated) single‐element coil, the 16‐channel coil showed a 1.87‐fold average increase in brain SNR. This was mainly due to an increase in SNR in the peripheral brain (an up to threefold SNR increase), whereas the SNR increase in the center of the brain was 4%. The incremental SNR gains became relatively small at large n, with a 9% gain observed when n was increased from 8 to 16. Compared to the (larger) product birdcage head coil, SNR increased by close to a factor of 2 in the center, and by up to a factor of 6 in the periphery of the brain. For low SENSE acceleration (rate‐2), g‐factors leveled off for n > 4, and improved only slightly (1.4% averaged over brain) going from n = 8 to n = 16. Improvements in g for n > 8 were larger for higher acceleration rates, with the improvement for rate‐3 averaging 12.0%. Magn Reson Med 51:22–26, 2004. Published 2003 Wiley‐Liss, Inc.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/mrm.10678

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 176

Metadata

Authors
de Zwart, J. A., Ledden, P. J., van Gelderen, P., Bodurka, J., Chu, R., & Duyn, J. H.
Venue
Magnetic Resonance in Medicine
Publisher
Wiley
Published
2003-12-29
Volume
51
Issue
1
Pages
22-26
ISSN
  • 0740-3194 (ISSN-L)
  • 1522-2594
OpenAlex
W2125086636
Cited by
176
References
1
FWCI
9.83
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Signal‐to‐noise ratio and parallel imaging performance of a 16‐channel receive‐only brain coil array at 3.0 Tesla. https://app.cochat.ai/papers/10.1002/mrm.10678
The paper
de Zwart, J. A., Ledden, P. J., van Gelderen, P., Bodurka, J., Chu, R., & Duyn, J. H. (2004). Signal‐to‐noise ratio and parallel imaging performance of a 16‐channel receive‐only brain coil array at 3.0 Tesla. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.10678

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.