B1 mapping by Bloch‐Siegert shift
Open accessAbstract
A novel method for amplitude of radiofrequency field (B) mapping based on the Bloch‐Siegert shift is presented. Unlike conventionally applied double‐angle or other signal magnitude–based methods, it encodes the B1 information into signal phase, resulting in important advantages in terms of acquisition speed, accuracy, and robustness. The Bloch‐Siegert frequency shift is caused by irradiating with an off‐resonance radiofrequency pulse following conventional spin excitation. When applying the off‐resonance radiofrequency in the kilohertz range, spin nutation can be neglected and the primarily observed effect is a spin precession frequency shift. This shift is proportional to the square of the magnitude of B. Adding gradient image encoding following the off‐resonance pulse allows one to acquire spatially resolved B1 maps. The frequency shift from the Bloch‐Siegert effect gives a phase shift in the image that is proportional to B. The phase difference of two acquisitions, with the radiofrequency pulse applied at two frequencies symmetrically around the water resonance, is used to eliminate undesired off‐resonance effects due to amplitude of static field inhomogeneity and chemical shift. In vivo Bloch‐Siegert B1 mapping with 25 sec/slice is demonstrated to be quantitatively comparable to a 21‐min double‐angle map. As such, this method enables robust, high‐resolution B mapping in a clinically acceptable time frame. Magn Reson Med 63:1315–1322, 2010. © 2010 Wiley‐Liss, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Magnetic Resonance for Nonrotating Fields
Bloch, F., & Siegert, A. (1940) · Physical Review · doi:10.1103/physrev.57.522 · Cited by 1,179
Katscher, U., Börnert, P., Leussler, C., & van den Brink, J. S. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10353 · Cited by 719
Rapid magnetic resonance quantification on the brain: Optimization for clinical usage
Warntjes, J. B. M., Leinhard, O. D., West, J., & Lundberg, P. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21635 · Cited by 500
Saturated double‐angle method for rapid B1+ mapping
Cunningham, C. H., Pauly, J. M., & Nayak, K. S. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20896 · Cited by 353
Hua, J., Jones, C. K., Blakeley, J., Smith, S. A., van Zijl, P. C. M., & Zhou, J. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21387 · Cited by 222
Radiofrequency map of an NMR coil by imaging
Akoka, S., Franconi, F., Seguin, F., & Le Pape, A. (1993) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(93)90078-r · Cited by 164
A phase‐sensitive method of flip angle mapping
Morrell, G. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21729 · Cited by 152
Fast, accurate, and precise mapping of the RF field in vivo using the 180° signal null
Dowell, N. G., & Tofts, P. S. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21368 · Cited by 118
In vivo method for correcting transmit/receive nonuniformities with phased array coils
Wang, J., Qiu, M., & Constable, R. T. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20377 · Cited by 110
Hornak, J. P., Szumowski, J., & Bryant, R. G. (1988) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910060204 · Cited by 104
Fast 3D radiofrequency field mapping using echo‐planar imaging
Jírů, F., & Klose, U. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21083 · Cited by 91
Simultaneous Soft Pulses Applied at Nearby Frequencies
Steffen, M., Vandersypen, L. M. K., & Chuang, I. L. (2000) · Journal of Magnetic Resonance · doi:10.1006/jmre.2000.2178 · Cited by 39
Cited by
Pros and cons of ultra-high-field MRI/MRS for human application
Ladd, M. E., Bachert, P., Meyerspeer, M., Moser, E., Nagel, A. M., Norris, D. G., Schmitter, S., Speck, O., Straub, S., & Zaiss, M. (2018) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/j.pnmrs.2018.06.001 · Cited by 544
Alexander, A. L., Hurley, S. A., Samsonov, A. A., Adluru, N., Hosseinbor, A. P., Mossahebi, P., Tromp, D. P. M., Zakszewski, E., & Field, A. S. (2011) · Brain Connectivity · doi:10.1089/brain.2011.0071 · Cited by 449
Shukla‐Dave, A., Obuchowski, N. A., Chenevert, T. L., Jambawalikar, S., Schwartz, L. H., Malyarenko, D., Huang, W., Noworolski, S. M., Young, R. J., Shiroishi, M. S., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26518 · Cited by 389
Using high-resolution quantitative mapping of R1 as an index of cortical myelination
Lutti, A., Dick, F., Sereno, M. I., & Weiskopf, N. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.06.005 · Cited by 346
What are normal relaxation times of tissues at 3 T?
Bojorquez, J. Z., Bricq, S., Acquitter, C., Brunotte, F., Walker, P. M., & Lalande, A. (2017) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2016.08.021 · Cited by 286
Metadata
- Authors
- Sacolick, L. I., Wiesinger, F., Hancu, I., & Vogel, M. W.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2010-04-23
- Volume
- 63
- Issue
- 5
- Pages
- 1315-1322
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2145535148
- Cited by
- 529
- References
- 12
- FWCI
- 21.94
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.