Reflectance pulse oximetry in fetal lambs: Subcutaneous vessels and vasoconstriction affect its reliability
Open accessAbstract
Objective Reflectance pulse oximetry (RPOX) has been introduced for intrapartum fetal surveillance. The purpose of this study was to describe two possible effects on the reliability of RPOX, namely the effect of the presence of a subcutaneous vein and the effect of vasoconstriction by adrenaline, both at fetal SaO2 levels.Methods In four anesthetized fetal lambs, a prototype 660/890 nm reflectance sensor (Nellcor Inc.) was placed on the fetal head, with the photodiode of the sensor precisely over a superficial subcutaneous vein. Measurements were made before and after coagulation of the vein. In five anesthetized fetal lambs, one or two reflectance sensors were placed on the fetal head and/or neck and adrenaline was administered in doses of 0.02 to 0.04 mg via a brachial artery. Pulse oximeter saturation readings (SpO2) were compared with continuous arterial oxygen saturation (SaO2) values obtained using a fiberoptic catheter (Opticath, Abbott) in the carotid artery.Results When the sensor was placed over the vein, the pulse oximeter read 18% to 24% too low at a SaO2 level of 20% to 50%. After coagulation of the vein, SpO2 readings were in agreement with fiberoptic SaO2 values. Administration of adrenaline resulted in a large overestimation of the SaO2 in 6 of the 7 measurements.Conclusions Subcutaneous veins and vasoconstriction can affect the reliability of reflectance pulse oximetry. As comparable situations may occur during labor, SpO2 readings should be interpreted with caution when this kind or comparable types of RPOX sensors are used at low SaO2 levels.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Nijland et al. confirmed in fetal lambs that reflectance pulse oximetry SpO2 readings were significantly lower than fiberoptic SaO2 values, and when the vessel was coagulated, this difference became negligible.
“Nijland et al. also confirmed these results in fetal lambs, where reflectance pulse oximetry SpO2 readings were significantly lower than fiberoptic SaO2 values. However, when the vessel was coagulated, this difference became negligible [128].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [128]
Cites
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 438 · see citing works
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 337 · see citing works
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617328 · Cited by 305 · see citing works
Kelleher, J. F. (1989) · The Journal of Clinical Monitoring · doi:10.1007/bf01618369 · Cited by 205 · see citing works
Dildy, G. A., van den Berg, P. P., Katz, M., Clark, S. L., Jongsma, H. W., Nijhuis, J. G., & Loucks, C. A. (1994) · American Journal of Obstetrics and Gynecology · doi:10.1016/0002-9378(94)90081-7 · Cited by 133 · see citing works
Self-citationSimple photon diffusion analysis of the effects of multiple scattering on pulse oximetry
Schmitt, J. M. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.137285 · Cited by 118 · see citing works
Fetal monitoring with pulse oximetry
Johnson, N. A., JOHNSON, V. A., Fisher, J. D., et al. (1991) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1991.tb10308.x · Cited by 104 · see citing works
Kim, J. M., Arakawa, K., Benson, K. T., & Fox, D. K. (1986) · Anesthesia & Analgesia · doi:10.1213/00000539-198612000-00015 · Cited by 86 · see citing works
Postoperative monitoring of conjunctival oxygen tension and temperature
Fink, S., Abraham, E., & Ehrlich, H. (1988) · International journal of clinical monitoring and computing · doi:10.1007/bf01739231 · Cited by 77 · see citing works
Adaptation of pulse oximetry for fetal monitoring during labour
Gardosi, J. O., Schram, C. M. H., & Symonds, M. E. (1991) · The Lancet · doi:10.1016/0140-6736(91)92928-u · Cited by 74 · see citing works
Artifacts in fetal pulse oximetry: Incomplete sensor-to-skin contact
Gardosi, J. O., Damianou, D., & Schram, C. M. H. (1994) · American Journal of Obstetrics and Gynecology · doi:10.1016/s0002-9378(94)70116-4 · Cited by 49 · see citing works
Reflectance Pulse Oximetry in Fetal Lambs
Dassel, A. C. M., Graaff, R., Aarnoudse, J. G., Elstrodt, J. M., Heida, P., Koelink, M. H., de Mul, F. F. M., & Greve, J. (1992) · Pediatric Research · doi:10.1203/00006450-199203000-00014 · Cited by 47 · see citing works
The effect of caput succedaneum on oxygen saturation measurements
Johnson, N., JOHNSON, V. A., Bannister, J., et al. (1990) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1990.tb02518.x · Cited by 47 · see citing works
Validation of Arterial Oxygen Saturation Measurements in Utero Using Pulse Oximetry
Harris, A. P., SENDAK, M. J., Chung, D., et al. (2008) · American Journal of Perinatology · doi:10.1055/s-2007-994731 · Cited by 30 · see citing works
Artifacts in fetal pulse oximetry: Nonarterial pulsatile signals
Schram, C. M. H., & Gardosi, J. O. (1994) · American Journal of Obstetrics and Gynecology · doi:10.1016/s0002-9378(94)70117-2 · Cited by 28 · see citing works
The effect of pulsating arteries on reflectance pulse oximetry: Measurements in adults and neonates
Nijland, R., Jongsma, W., van den Berg, P. P., et al. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617733 · Cited by 24 · see citing works
Self-citationMendelson, Y., & Yocum, B. L. (1992) · PubMed · Cited by 16 · see citing works
Tissue optics applied to reflectance pulse oximetry.
Graaff, R. (1993) · University of Groningen research database (University of Groningen / Centre for Information Technology) · Cited by 16 · see citing works
Measurement of oxygen saturation by multiwavelength analyzer influenced by interspecies differences
Nijland, R., Ringnalda, B. E. M., Jongsma, H. W., et al. (1994) · Clinical Chemistry · doi:10.1093/clinchem/40.10.1971 · Cited by 15 · see citing works
Self-citationNijland, R., Jongsma, H. W., NlJHUIS, J. G., & Oeseburg, B. (1995) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1995.tb04336.x · Cited by 5 · see citing works
Self-citationNijland, R., Jongsma, H. W., Menssen, J., Nijhuis, J. G., van den Berg, P. P., & Oeseburg, B. (1996) · Advances in experimental medicine and biology · doi:10.1007/978-1-4613-0333-6_8 · Cited by 5 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
- Authors
- Nijland, R., Jongsma, H. W., Verbruggen, I. M., & Nijhuis, J. G.
- Affiliations
- Radboud University Nijmegen (NL)
- ORCID
- Nijhuis, Jan G.
- Venue
- Journal of Clinical Monitoring
- Published
- 1996-05
- Volume
- 12
- Issue
- 3
- Pages
- 225-230
- ISSN
- 0748-1977 (ISSN-L)
- OpenAlex
- W1999181681
- Cited by
- 8
- References
- 21
- Citations by year
- 2021: 1
- FWCI
- 1.23
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- other-oa
- www.springer.com/tdm (tdm)
- Locations
- The Journal of Clinical Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Radboud Repository (Radboud University) (submittedVersion)
- Radboud Repository (Radboud University) (submittedVersion, other-oa)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- reflectance pulse oximetry, fetal oxygen saturation, subcutaneous veins, vasoconstriction, epinephrine, intrapartum fetal monitoring, fetal sheep, arterial oxygen saturation, sensor reliability
- MeSH terms
- Animals, Epinephrine, Female, Fetal Blood, Fetal Monitoring, Oximetry, Pregnancy, Sheep, Vasoconstriction, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review