The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Imbalanced biphasic electrical stimulation: Muscle tissue damage
Abstract
The effects of imbalanced biphasic stimulation were studied on cat skeletal muscle to determine if greater charge densities can be safely used than with balanced or monophasic stimulation. The results of the study indicate that imbalanced biphasic stimulation can be tolerated safely by tissue at or below a net dc current density of 35 microA/mm2 and not safely tolerated at or above a net dc current of 50 microA/mm2. Monophasic stimulation has been shown to be safe at or below net dc current levels of 10 microA/mm2 and in these studies we found it was not safe at or above net dc current levels of 20 microA/mm2. Stimuli were applied to muscles via coiled wire intramuscular electrodes using a regulated current source. Since the safe average current density was higher for imbalanced biphasic stimulation than for monophasic stimulation, this suggests that: (a) pH change is not the primary reaction causing tissue damage and (b) the damaging electrochemical process that takes place during a cathodic stimulation pulse can be reversed by an anodic pulse having substantially less charge than its companion cathodic pulse. We conclude that greater cathodic charge densities can be safely employed with imbalanced biphasic stimulation than with either monophasic stimulation or balanced charge biphasic stimulation.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Charge-balanced, or at least biphasic, waveforms are now commonly used (see Table 4) in order to limit any electro-chemical change [77, 86].”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [86]
Citations
Cites
Statistics for experimenters. An introduction to design, data analysis, and model building.
Box, G. E. P., Hunter, W. G., & Hunter, J. S. (1978) · Cited by 4,021
Statistics for Experimenters: An Introduction to Design, Data Analysis and Model Building.
Pearce, S. C., Box, G. E. P., Hunter, W. G., & Hunter, J. S. (1979) · Biometrics · doi:10.2307/2530361 · Cited by 2,003
Statistical Analysis in Psychology and Education.
Jackson, P. H., & Ferguson, G. A. (1972) · Journal of the Royal Statistical Society. Series A (General) · doi:10.2307/2345050 · Cited by 644
Brummer, S. B., Robblee, L. S., & Hambrecht, F. T. (1983) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1983.tb31628.x · Cited by 205
Intramuscular electrical stimulation: Tissue damage
Mortimer, J. T., Kaufman, D., & Roessmann, U. (1980) · Annals of Biomedical Engineering · doi:10.1007/bf02364479 · Cited by 123
Self-citationDecreased defibrillator-induced dysfunction with biphasic rectangular waveforms
Jones, J. L., & Jones, R. E. (1984) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1984.247.5.h792 · Cited by 104
Ultrastructural injury to chick myocardial cells in vitro following "electric countershock".
Jones, J. L., Proskauer, C. C., Paull, W. K., et al. (1980) · Circulation Research · doi:10.1161/01.res.46.3.387 · Cited by 93
A Percutaneous Wire Electrode for Chronic Research Use
Caldwell, C. W., & RESWICK, J. B. (1975) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1975.324516 · Cited by 79
An approach to corrosion control during electrical stimulation
McHardy, J., Geller, D., & Brummer, S. B. (1977) · Annals of Biomedical Engineering · doi:10.1007/bf02364014 · Cited by 45
Calculations of temperature rise produced in body tissue by a spherical electrode
Ruggeri, R. T., & Beck, T. R. (1985) · Annals of Biomedical Engineering · doi:10.1007/bf02584237 · Cited by 8
Cited by
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078
“9(c), may be used to reduce the most positive potentials during the anodic phase with respect to the charge-balanced waveform, and prevent electrode corrosion (Scheiner and Mortimer, 1990).”
BackgroundInfluentialAccelerated Skin Wound Healing by Electrical Stimulation
Luo, R., Dai, J., Zhang, J., & Li, Z. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100557 · Cited by 510
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198
Cogan, S. F., Troyk, P. R., Ehrlich, J., Plante, T. D., & Detlefsen, D. E. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.862572 · Cited by 138
Metadata
- Authors
- Scheiner, A., Mortimer, J. T., & Roessmann, U.
- Affiliations
- Case Western Reserve University (US)
- ORCID
- Mortimer, J. Thomas
- Venue
- Annals of Biomedical Engineering
- Published
- 1990-07
- Volume
- 18
- Issue
- 4
- Pages
- 407-425
- ISSN
- 0090-6964 (ISSN-L)
- 1573-9686
- OpenAlex
- W3157488099
- Cited by
- 117
- References
- 10
- Citations by year
- 2026: 5 · 2025: 2 · 2024: 2 · 2023: 3 · 2022: 3 · 2021: 7 · 2020: 2 · 2019: 3 · 2018: 5 · 2017: 7 · 2016: 6 · 2015: 3 · 2014: 4 · 2013: 6 · 2012: 7
- FWCI
- 0.14
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (N0I-NS-4-2362)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Keywords
- muscle damage, skeletal muscle, current density, intramuscular electrodes, pH change, stimulation safety
- MeSH terms
- Animals, Cats, Electric Injuries, Electric Stimulation, Electrodes, Muscles, Necrosis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Imbalanced biphasic electrical stimulation: Muscle tissue damage. https://app.cochat.ai/papers/10.1007/bf02364157
- The paper
- Scheiner, A., Mortimer, J. T., & Roessmann, U. (1990). Imbalanced biphasic electrical stimulation: Muscle tissue damage. Annals of Biomedical Engineering. https://doi.org/10.1007/bf02364157
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.