Safety of long-term electrical peripheral nerve stimulation: review of the state of the art

Open access
Günter, Clara, Delbeke, Jean, Ortiz-Catalan, Max · 2019 · Journal of NeuroEngineering and Rehabilitation

Abstract

BACKGROUND: Electrical stimulation of peripheral nerves is used in a variety of applications such as restoring motor function in paralyzed limbs, and more recently, as means to provide intuitive sensory feedback in limb prostheses. However, literature on the safety requirements for stimulation is scarce, particularly for chronic applications. Some aspects of nerve interfacing such as the effect of stimulation parameters on electrochemical processes and charge limitations have been reviewed, but often only for applications in the central nervous system. This review focuses on the safety of electrical stimulation of peripheral nerve in humans. METHODS: We analyzed early animal studies evaluating damage thresholds, as well as more recent investigations in humans. Safety requirements were divided into two main categories: passive and active safety. We made the distinction between short-term (< 30 days) and chronic (> 30 days) applications, as well as between electrode preservation (biostability) and body tissue healthy survival (harmlessness). In addition, transferability of experimental results between different tissues and species was considered. RESULTS: At present, extraneural electrodes have shown superior long-term stability in comparison to intraneural electrodes. Safety limitations on pulse amplitude (and consequently, charge injection) are dependent on geometrical factors such as electrode placement, size, and proximity to the stimulated fiber. In contrast, other parameters such as stimulation frequency and percentage of effective stimulation time are more generally applicable. Currently, chronic stimulation at frequencies below 30 Hz and percentages of effective stimulation time below 50% is considered safe, but more precise data drawn from large databases are necessary. Unfortunately, stimulation protocols are not systematically documented in the literature, which limits the feasibility of meta-analysis and impedes the generalization of conclusions. We therefore propose a standardized list of parameters necessary to define electrical stimulation and allow future studies to contribute to meta-analyses. CONCLUSION: The safety of chronic continuous peripheral nerve stimulation at frequencies higher than 30 Hz has yet to be documented. Precise parameter values leading to stimulation-induced depression of neuronal excitability (SIDNE) and neuronal damage, as well as the transition between the two, are still lacking. At present, neural damage mechanisms through electrical stimulation remain obscure.

Overview

CoChat Summary

TL;DR

This review synthesizes what is known about the safety of long-term electrical stimulation of human peripheral nerves, separating passive electrode safety from active stimulation-parameter safety. It concludes that extraneural electrodes are currently the safer chronic option, while chronic stimulation above 30 Hz and the effects of uninterrupted long-term stimulation remain insufficiently documented.

Key findings

  • Extraneural electrodes showed superior long-term stability versus intraneural electrodes; the paper states that extraneural electrodes had proven biostable in humans for at least 12 years (peroneal nerve), up to 18 years (phrenic nerve), up to 8 years (median, radial, sciatic, and ulnar nerves), and up to 11 years in feedback-controlled applications.
  • For chronic stimulation, the review says frequencies below 30 Hz and percentages of effective stimulation time below 50% are currently considered safe; however, the safety of chronic continuous stimulation above 30 Hz has “yet to be documented.”
  • Reported safe amplitudes depend strongly on geometry and electrode type: up to 3100 μA in cat sciatic nerves, up to 20 mA in humans with cuff/needle electrodes, up to 3 mA in human vagus nerve cuff stimulation, and typically 10–300 μA for intrafascicular electrodes.
  • The paper highlights that stimulation protocols are poorly documented, which blocks meta-analysis and generalization, and proposes a standardized parameter list for future studies.

Methods

  • Narrative review of the literature focused on safety of peripheral nerve stimulation in humans, with supporting evidence drawn from early animal studies of damage thresholds and more recent human investigations.
  • The authors organized safety into passive safety and active safety, and distinguished short-term (< 30 days) from chronic (> 30 days) applications, and biostability from harmlessness.
  • The review compares transferability across species and tissue types and summarizes effects of parameters including charge, pulse duration, amplitude, waveform, inter-phase delay, frequency, percentage of effective stimulation time, and treatment duration.

Limitations

  • The paper explicitly states that stimulation protocols are not systematically documented in the literature, limiting the feasibility of meta-analysis and the generalization of conclusions.
  • The paper explicitly states that precise parameter values for stimulation-induced depression of neuronal excitability (SIDNE), neuronal damage, and the transition between them are still lacking.
  • Evident from the text: much of the safety evidence comes from animal studies and short-term human experiments, so chronic human safety conclusions are based on extrapolation rather than direct long-term human data.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1186/s12984-018-0474-8

Cites

119 of 119 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 Classification of electrodes used in neurostimulation. Flat Interface Nerve Electrode (FINE), Transversal Intrafascicular Multichannel Electrode (TIME), Utah Slanted Electrode Array (USEA), Coiled Wire Intra-neural Electrode (CWIE), Longitudinal IntraFascicular Electrode (LIFE), Micro-/MultiChannel Roll Electrode (MCRE), Regenerative MultiElectrode Array (REMI)
Fig. 1 Classification of electrodes used in neurostimulation. Flat Interface Nerve Electrode (FINE), Transversal Intrafascicular Multichannel Electrode (TIME), Utah Slanted Electrode Array (USEA), Coiled Wire Intra-neural Electrode (CWIE), Longitudinal IntraFascicular Electrode (LIFE), Micro-/MultiChannel Roll Electrode (MCRE), Regenerative MultiElectrode Array (REMI)CC BY · PubMed Central
Fig. 2 Biphasic, asymmetric, square pulse with inter-phase delay. Current of stimulating phase of stimulation (I), duration of stimulating phase of stimulation (D), inter-phase delay (w), stimulation pulse frequency (F)
Fig. 2 Biphasic, asymmetric, square pulse with inter-phase delay. Current of stimulating phase of stimulation (I), duration of stimulating phase of stimulation (D), inter-phase delay (w), stimulation pulse frequency (F)CC BY · PubMed Central
Fig. 3 Waveform parameters
Fig. 3 Waveform parametersCC BY · PubMed Central

Verification

Claims the paper makes

16 Valid · 10 Not valid · 73 Not checked · 38 Own results and interpretations

74 more claims · 5 with their cited sources checked

Create a free account to see all 137 claims and to verify the 1 reference still open.

See all claims — free account
  • Probably not a substantial claim. 20
  • It may state two separate claims as one. 43
  • The quote only partly states this claim. 1

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 120 references

Research works printed but not linked: 2

  1. Ortiz-Catalan M, Mastinu E, Brånemark R, Håkansson B. Direct Neural Sensory Feedback and Control via Osseointegration. Cape Town: XVI World Congr Int Soc Prosthetics Orthot (ISPO); 2017.
  2. Lilly JC. Injury and excitation by electric currents. In: Sheer DE, editor. Electr Stimul brain. Austin: University of Texas Press; 1961. p. 60–4.

Cited by

See all 198

Metadata

Authors
Günter, C., Delbeke, J., & Ortiz-Catalan, M.
Affiliations
  • Chalmers University of Technology (SE)
  • Ghent University (BE)
  • Integrum (Sweden) (SE)
Corresponding author
Ortiz-Catalan, Max
Venue
Journal of NeuroEngineering and Rehabilitation
Published
2019-01-18
Volume
16
Issue
1
Pages
13-13
ISSN
1743-0003 (ISSN-L)
OpenAlex
W2917203101
Cited by
198
References
120
Citations by year
2026: 12 · 2025: 32 · 2024: 30 · 2023: 36 · 2022: 27 · 2021: 30 · 2020: 24 · 2019: 7
FWCI
2.93
Type
journal article
Language
en
Open access
Licence
Locations
  • Journal of NeuroEngineering and Rehabilitation (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, other-oa)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Ghent University Academic Bibliography (Ghent University) (submittedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Chalmers Research (Chalmers University of Technology) (submittedVersion)
Funders
  • Stiftelsen Promobilia
  • VINNOVA
  • Horizon 2020 Framework Programme
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
  • Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
Keywords
peripheral nerve stimulation, charge injection limit, stimulation parameters, stimulation frequency, motor recovery
MeSH terms
Animals, Electric Stimulation, Humans, Peripheral Nerves

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Safety of long-term electrical peripheral nerve stimulation: review of the state of the art. https://app.cochat.ai/papers/10.1186/s12984-018-0474-8
The paper
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019). Safety of long-term electrical peripheral nerve stimulation: review of the state of the art. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/s12984-018-0474-8

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.