Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Open accessAbstract
BACKGROUND: Electrical stimulation of peripheral nerves is used in a variety of applications such as restoring motor function in paralyzed limbs, and more recently, as means to provide intuitive sensory feedback in limb prostheses. However, literature on the safety requirements for stimulation is scarce, particularly for chronic applications. Some aspects of nerve interfacing such as the effect of stimulation parameters on electrochemical processes and charge limitations have been reviewed, but often only for applications in the central nervous system. This review focuses on the safety of electrical stimulation of peripheral nerve in humans. METHODS: We analyzed early animal studies evaluating damage thresholds, as well as more recent investigations in humans. Safety requirements were divided into two main categories: passive and active safety. We made the distinction between short-term (< 30 days) and chronic (> 30 days) applications, as well as between electrode preservation (biostability) and body tissue healthy survival (harmlessness). In addition, transferability of experimental results between different tissues and species was considered. RESULTS: At present, extraneural electrodes have shown superior long-term stability in comparison to intraneural electrodes. Safety limitations on pulse amplitude (and consequently, charge injection) are dependent on geometrical factors such as electrode placement, size, and proximity to the stimulated fiber. In contrast, other parameters such as stimulation frequency and percentage of effective stimulation time are more generally applicable. Currently, chronic stimulation at frequencies below 30 Hz and percentages of effective stimulation time below 50% is considered safe, but more precise data drawn from large databases are necessary. Unfortunately, stimulation protocols are not systematically documented in the literature, which limits the feasibility of meta-analysis and impedes the generalization of conclusions. We therefore propose a standardized list of parameters necessary to define electrical stimulation and allow future studies to contribute to meta-analyses. CONCLUSION: The safety of chronic continuous peripheral nerve stimulation at frequencies higher than 30 Hz has yet to be documented. Precise parameter values leading to stimulation-induced depression of neuronal excitability (SIDNE) and neuronal damage, as well as the transition between the two, are still lacking. At present, neural damage mechanisms through electrical stimulation remain obscure.
Overview
CoChat Summary
TL;DR
This review synthesizes what is known about the safety of long-term electrical stimulation of human peripheral nerves, separating passive electrode safety from active stimulation-parameter safety. It concludes that extraneural electrodes are currently the safer chronic option, while chronic stimulation above 30 Hz and the effects of uninterrupted long-term stimulation remain insufficiently documented.
Key findings
- Extraneural electrodes showed superior long-term stability versus intraneural electrodes; the paper states that extraneural electrodes had proven biostable in humans for at least 12 years (peroneal nerve), up to 18 years (phrenic nerve), up to 8 years (median, radial, sciatic, and ulnar nerves), and up to 11 years in feedback-controlled applications.
- For chronic stimulation, the review says frequencies below 30 Hz and percentages of effective stimulation time below 50% are currently considered safe; however, the safety of chronic continuous stimulation above 30 Hz has “yet to be documented.”
- Reported safe amplitudes depend strongly on geometry and electrode type: up to 3100 μA in cat sciatic nerves, up to 20 mA in humans with cuff/needle electrodes, up to 3 mA in human vagus nerve cuff stimulation, and typically 10–300 μA for intrafascicular electrodes.
- The paper highlights that stimulation protocols are poorly documented, which blocks meta-analysis and generalization, and proposes a standardized parameter list for future studies.
Methods
- Narrative review of the literature focused on safety of peripheral nerve stimulation in humans, with supporting evidence drawn from early animal studies of damage thresholds and more recent human investigations.
- The authors organized safety into passive safety and active safety, and distinguished short-term (< 30 days) from chronic (> 30 days) applications, and biostability from harmlessness.
- The review compares transferability across species and tissue types and summarizes effects of parameters including charge, pulse duration, amplitude, waveform, inter-phase delay, frequency, percentage of effective stimulation time, and treatment duration.
Limitations
- The paper explicitly states that stimulation protocols are not systematically documented in the literature, limiting the feasibility of meta-analysis and the generalization of conclusions.
- The paper explicitly states that precise parameter values for stimulation-induced depression of neuronal excitability (SIDNE), neuronal damage, and the transition between them are still lacking.
- Evident from the text: much of the safety evidence comes from animal studies and short-term human experiments, so chronic human safety conclusions are based on extrapolation rather than direct long-term human data.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
16 Valid · 10 Not valid · 73 Not checked · 38 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“and up to 11 years stimulating motor and sensory nerves in a feedback controlled application [26]”
§ Safety of long-term electrical peripheral nerve stimulation: review of the state of the art > Passive safety > Biostability and harmlessness
What the sources say
- Partly supportedcites [26] Christie BP, Freeberg M, Memberg WD, Pinault GJC, Hoyen HA, Tyler DJ, et al. Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves. J Neuroeng Rehabil. 2017;14:70. sourceExperimental — accuracy not yet measured
The Conclusion reports that non-penetrating spiral nerve cuffs remained operational for up to 11 years in motor and sensory neuroprostheses. The Methods describe sensory feedback for one participant and control signals for motor-system users, but do not establish that the up-to-11-year result was obtained in a feedback-controlled application combining motor and sensory stimulation. (LLM, conf=0.84)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Data collected by Navarro et al. point to the fact that currently in clinical practice, implanted electrodes are generally extraneural [36].”
§ Safety of long-term electrical peripheral nerve stimulation: review of the state of the art > Passive safety > Structural classification
What the sources say
- Supportedcites [36] Navarro X, Krueger TB, Lago N, Micera S, Stieglitz T, Dario P. A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems. J Peripher Nerv Syst. 2005;10:229–58. source
The electrode-status table (Surface electrodes, p. 6) lists implanted epineurial and helical electrodes as in clinical practice, while listing intrafascicular electrodes for artificial-limb control as research. The Epineurial and helicoidal electrodes section (p. 9) describes epineurial electrodes as placed on the nerve, supporting the review’s general distinction between clinical extraneural interfaces and intraneural research devices. (LLM, conf=0.87)
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“At present, extraneural electrodes have shown superior long-term stability in comparison to intraneural electrodes.”
- Also stated in: § Structural classification · plain
“Presently, with regards to long-term human implantations, extraneural electrodes still emerge as the most favorable choice for clinical applications [26, 37].”
§ Safety of long-term electrical peripheral nerve stimulation: review of the state of the art > Abstract > Results
What the sources say
- Partly supportedcites [26] Christie BP, Freeberg M, Memberg WD, Pinault GJC, Hoyen HA, Tyler DJ, et al. Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves. J Neuroeng Rehabil. 2017;14:70. sourceExperimental — accuracy not yet measured
The Conclusion reports that non-penetrating spiral nerve cuff electrodes remained reliable for up to 11 years. But the Discussion says the study focused solely on those electrodes; it does not establish that extraneural electrodes are more stable than intraneural electrodes. (LLM, conf=0.85)
- Supportedcites [37] Ortiz-Catalan M, Brånemark R, Håkansson B, Delbeke J. On the viability of implantable electrodes for the natural control of artificial limbs: review and discussion. Biomed Eng Online. 2012;11:33. source
The Discussion section, “Suitable IEs for long-term and natural prosthetic control” (page unavailable), says intraneural electrodes can lose long-term recording quality through encapsulation, whereas extra-type epimysial and cuff electrodes have remained functional for years. The Review section, “Nerve-based Electrodes” (page unavailable), also reports stable cuff-electrode performance over time. Together, these passages support the comparison as a summary of the review. (LLM, conf=0.91)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Structural classification · plain
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Presently, with regards to long-term human implantations, extraneural electrodes still emerge as the most favorable choice for clinical applications [26, 37].”
§ Safety of long-term electrical peripheral nerve stimulation: review of the state of the art > Passive safety > Structural classification
What the sources say
- Partly supportedcites [26] Christie BP, Freeberg M, Memberg WD, Pinault GJC, Hoyen HA, Tyler DJ, et al. Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves. J Neuroeng Rehabil. 2017;14:70. sourceExperimental — accuracy not yet measured
The Conclusion (page not provided) finds non-penetrating spiral nerve cuff electrodes reliable for long-term clinical use in human neuroprostheses, with performance reported up to 11 years after implantation. The Discussion (page not provided) says the study examined only spiral cuffs and notes tradeoffs among interfaces; it does not establish that extraneural electrodes generally are the most favorable choice. (LLM, conf=0.88)
- Supportedcites [37] Ortiz-Catalan M, Brånemark R, Håkansson B, Delbeke J. On the viability of implantable electrodes for the natural control of artificial limbs: review and discussion. Biomed Eng Online. 2012;11:33. source
The Discussion section, “Suitable IEs for long-term and natural prosthetic control” (page unavailable), says intraneural electrodes carry greater risk, while extra-type electrodes, including cuffs, have demonstrated safety and functionality over several years in clinical applications. The Review’s “Nerve-based Electrodes” section also reports long-term functionality of cuff electrodes implanted in humans. Together, these passages support the cited preference for extraneural electrodes in long-term clinical implantation. (LLM, conf=0.88)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“In the case of eliciting somatosensory perception via peripheral nerve stimulation, Tan et al. found that the quality of a sensation can be altered by pulse width modulation [17]”
§ Safety of long-term electrical peripheral nerve stimulation: review of the state of the art > Active safety – Stimulation parameters > Duration
What the sources say
- Supportedcites [17] Tan DW, Schiefer MA, Keith MW, Anderson JR, Tyler J, Tyler DJ. A neural interface provides long-term stable natural touch perception. Sci Transl Med. 2014;6:257ra138. source
In the results (pp. 3–4), Tan et al. report that modulating stimulation pulse width changed perceived sensations, including a shift from tingling to natural pressure. They also report that reducing the modulation changed pulsing pressure to continuous pressure or, at some sites, vibration, itch, or tingling. (LLM, conf=0.94)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
74 more claims · 5 with their cited sources checked
Create a free account to see all 137 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 20
- It may state two separate claims as one. 43
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Complications due to changes in the endoneurial environment include increased endoneurial pressure, nerve fiber compression, variable shifts in thresholds and loss of axons [29, 30].”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [30]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“These events have particularly been observed in larger nerve fibers [4, 31, 32] and in chronically implanted intrafascicular devices [30, 33, 34].”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [30]
The cited source says what the claim attributes to it.
In the citing paper
“Typical stimulation pulse widths range up to a few hundred μs , and current amplitudes vary from tens of μA to a few mA in peripheral nerve interfaces [9], [101].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [9]
Citations
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078
Coding and use of tactile signals from the fingertips in object manipulation tasks
Johansson, R. S., & Flanagan, J. R. (2009) · Nature Reviews Neuroscience · doi:10.1038/nrn2621 · Cited by 2,063
Which elements are excited in electrical stimulation of mammalian central nervous system: A review
Ranck, J. B. (1975) · Brain Research · doi:10.1016/0006-8993(75)90364-9 · Cited by 2,024 · cited 1×
“Stimulation safety limits cannot be applied as a global concept to neural tissue due to the diversity of cells forming it [103].”
BackgroundBiological Responses to Materials
Anderson, J. M. (2001) · Annual Review of Materials Research · doi:10.1146/annurev.matsci.31.1.81 · Cited by 1,465
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751
Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708
Ben‐Menachem, E., Mañon‐Espaillat, R., Ristanovic, R., Wilder, B. J., Stefan, H., Mirza, W., Tarver, W. B., Wernicke, J. F., & First International Vagus Nerve Stimulation Study Group (1994) · Epilepsia · doi:10.1111/j.1528-1157.1994.tb02482.x · Cited by 683
Research works printed but not linked: 2
- Ortiz-Catalan M, Mastinu E, Brånemark R, Håkansson B. Direct Neural Sensory Feedback and Control via Osseointegration. Cape Town: XVI World Congr Int Soc Prosthetics Orthot (ISPO); 2017.
- Lilly JC. Injury and excitation by electric currents. In: Sheer DE, editor. Electr Stimul brain. Austin: University of Texas Press; 1961. p. 60–4.
Cited by
Electroactive electrospun nanofibers for tissue engineering
Zhang, X., Li, L., Ouyang, J., Zhang, L., Xue, J., Zhang, H., & Tao, W. (2021) · Nano Today · doi:10.1016/j.nantod.2021.101196 · Cited by 197
Zuo, K. J., Gordon, T., Chan, K. M., & Borschel, G. H. (2020) · Experimental Neurology · doi:10.1016/j.expneurol.2020.113397 · Cited by 161
Ejneby, M. S., Jakešová, M., Ferrero, J. J., Migliaccio, L., Sahalianov, I. Y., Zhao, Z., Magnus Berggren, A., Khodagholy, D., Đerek, V., Gelinas, J. N., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00817-7 · Cited by 185
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133
In Vivo Organic Bioelectronics for Neuromodulation
Magnus Berggren, A., Głowacki, E. D., Simon, D. T., Stavrinidou, E., & Tybrandt, K. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00390 · Cited by 129
Metadata
- Authors
- Günter, C., Delbeke, J., & Ortiz-Catalan, M.
- Affiliations
- Chalmers University of Technology (SE)
- Ghent University (BE)
- Integrum (Sweden) (SE)
- Corresponding author
- Ortiz-Catalan, Max
- Venue
- Journal of NeuroEngineering and Rehabilitation
- Published
- 2019-01-18
- Volume
- 16
- Issue
- 1
- Pages
- 13-13
- ISSN
- 1743-0003 (ISSN-L)
- PMC
- PMC6339286
- OpenAlex
- W2917203101
- Cited by
- 198
- References
- 120
- Citations by year
- 2026: 12 · 2025: 32 · 2024: 30 · 2023: 36 · 2022: 27 · 2021: 30 · 2020: 24 · 2019: 7
- FWCI
- 2.93
- Type
- journal article
- Language
- en
- Update notices
- correction (2020-06-15)
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0/ (tdm)
- creativecommons.org/licenses/by/4.0/ (vor)
- Locations
- Journal of NeuroEngineering and Rehabilitation (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Ghent University Academic Bibliography (Ghent University) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Chalmers Research (Chalmers University of Technology) (submittedVersion)
- Funders
- Stiftelsen Promobilia
- VINNOVA
- Horizon 2020 Framework Programme
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- peripheral nerve stimulation, charge injection limit, stimulation parameters, stimulation frequency, motor recovery
- MeSH terms
- Animals, Electric Stimulation, Humans, Peripheral Nerves
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Safety of long-term electrical peripheral nerve stimulation: review of the state of the art. https://app.cochat.ai/papers/10.1186/s12984-018-0474-8
- The paper
- Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019). Safety of long-term electrical peripheral nerve stimulation: review of the state of the art. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/s12984-018-0474-8
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.