CRITERIA FOR SELECTING ELECTRODES FOR ELECTRICAL STIMULATION: THEORETICAL AND PRACTICAL CONSIDERATIONS
Abstract
Smaller, more charge-intensive electrodes are needed for "safe" stimulation of the nervous system. In this paper we review critical concepts and the state of the art in electrodes. Control of charge density and charge balance are essential to avoid tissue electrolysis. Chemical criteria for "safe" stimulation are reviewed ("safe" is equated with "chemically reversible"). An example of a safe, but generally impractical, charge-injection process is double-layer charging. The limit here is the onset of irreversible faradaic processes. More charge can be safely injected with so-called "capacitor" electrodes, such as porous intermixtures of Ta/Ta2O5. BaTiO3 has excellent dielectric properties and may provide a new generation of capacitor electrodes. Faradaic charge injection is usually partially irreversible since some of the products escape into the solution. With Pt, up to 400 muc/cm2 real area can be absorbed by faradaic reactions of surface-adsorbed species, but a small part is lost due to metal dissolution. The surface of "activated" Ir is covered with a multilayer hydrated oxide. Charge injection occurs via rapid valence change within this oxide. Little or no metal dissolution is observed, and gassing limits are not exceeded even under stringent conditions.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage.
“Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Brummer et al., 1983
Cited by
Sorted by citations
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · see citing works
In vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes
Weiland, J. D., Anderson, D. J., & Humayun, M. S. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.805487 · Cited by 382 · see citing works
Weiland, J. D., Liu, W., & Humayun, M. S. (2005) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.7.060804.100435 · Cited by 371 · see citing works
The Argus® II Retinal Prosthesis System
Luo, Y. H.-L., & da Cruz, L. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2015.09.003 · Cited by 351 · see citing works
Lewis, P. M., Ackland, H. M., Lowery, A. J., & Rosenfeld, J. V. (2015) · Brain Research · doi:10.1016/j.brainres.2014.11.020 · Cited by 228 · see citing works
Cuff electrodes for chronic stimulation and recording of peripheral nerve activity
Loeb, G. E., & Peck, R. A. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00123-9 · Cited by 215 · see citing works
Factors affecting perceptual thresholds in epiretinal prostheses
de Balthasar, C., Patel, S., Roy, A., Freda, R., Greenwald, S., Horsager, A., Mahadevappa, M., Yanai, D., McMahon, M. J., Humayun, M. S., et al. (2008) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.07-0696 · Cited by 204 · see citing works
A model for intracortical visual prosthesis research
Troyk, P., Bak, M., Berg, J., Bradley, D., Cogan, S., Erickson, R., Kufta, C., McCreery, D., Schmidt, E., & Towle, V. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07308.x · Cited by 166 · see citing works
Sputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation
Slavcheva, E., Vitushinsky, R., Mokwa, W., & Schnakenberg, U. (2004) · Journal of The Electrochemical Society · doi:10.1149/1.1747881 · Cited by 154 · see citing works
Biostability of micro‐photodiode arrays for subretinal implantation
Hämmerle, H., Kobuch, K., Kohler, K., Nisch, W., Sachs, H., & Stelzle, M. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00185-5 · Cited by 146 · see citing works
Imbalanced biphasic electrical stimulation: Muscle tissue damage
Scheiner, A., Mortimer, J. T., & Roessmann, U. (1990) · Annals of Biomedical Engineering · doi:10.1007/bf02364157 · Cited by 117 · see citing works
Predicting visual sensitivity in retinal prosthesis patients
Horsager, A., Greenwald, S. H., Weiland, J. D., Humayun, M. S., Greenberg, R. J., McMahon, M. J., Boynton, G. M., & Fine, I. (2009) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.08-2595 · Cited by 107 · see citing works
Lewis, P. M., & Rosenfeld, J. V. (2016) · Brain Research · doi:10.1016/j.brainres.2015.08.038 · Cited by 107 · see citing works
Recent advances in neural electrode–tissue interfaces
Woeppel, K., Yang, Q., & Cui, X. T. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.09.003 · Cited by 99 · see citing works
Current Density Profiles of Surface Mounted and Recessed Electrodes for Neural Prostheses
Rubinstein, J. T., Spelman, F. A., Soma, M., & Suesserman, M. F. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326007 · Cited by 89 · see citing works
The effect of chronic intracortical microstimulation on the electrode–tissue interface
Chen, K. H., Dammann, J. F., Boback, J. L., Tenore, F. V., Otto, K. J., Gaunt, R. A., & Bensmaia, S. J. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/2/026004 · Cited by 59 · see citing works
Polymeric C-shaped cuff electrode for recording of peripheral nerve signal
Xue, N., Sun, T., Tsang, W. M., Delgado-Martinez, I., Lee, S.-H., Sheshadri, S., Xiang, Z., Merugu, S., Gu, Y., Yen, S.-C., et al. (2015) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2015.01.006 · Cited by 48 · see citing works
Brunton, E. K., Winther-Jensen, B., Wang, C., Yan, E. B., Hagh Gooie, S., Lowery, A. J., & Rajan, R. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2015.00265 · Cited by 46 · see citing works
- Authors
- Brummer, S. B., Robblee, L. S., & Hambrecht, F. T.
- Affiliations
- EIC Laboratories (US)
- National Institutes of Health (US)
- Venue
- Annals of the New York Academy of Sciences
- Publisher
- Wiley
- Published
- 1983-06
- Volume
- 405
- Issue
- 1
- Pages
- 159-171
- ISSN
- 0077-8923 (ISSN-L)
- 1749-6632
- OpenAlex
- W2014316686
- Cited by
- 205
- References
- 0
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 4 · 2023: 4 · 2022: 3 · 2021: 4 · 2020: 3 · 2019: 8 · 2018: 8 · 2017: 6 · 2016: 5 · 2015: 9 · 2014: 15 · 2013: 3 · 2012: 2
- FWCI
- 0.71
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Annals of the New York Academy of Sciences (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advancements in Battery Materials
- Keywords
- electrode selection, charge density, charge balance, double-layer charging, capacitor electrodes, BaTiO3 dielectric, platinum electrode, iridium oxide, metal dissolution
- MeSH terms
- Electric Injuries, Electric Stimulation, Electrochemistry, Electrodes, Humans, Iridium, Neurons, Platinum
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review