Multicentric study of monitoring alarms in the adult intensive care unit (ICU): a descriptive analysis
Abstract
Objectives To assess the relevance of current monitoring alarms as a warning system in the adult ICU.Design Prospective, observational study.Settings Two university hospital, and three general hospital, ICUs.Patients Hundred thirty-one patients, ventilated at admission, from different shifts (morning, evening, night) combined with different stages of stay, early (0-3 days), intermediate (4-6 days) and late (> 6 days).Interventions Experienced nurses were asked to record the patient's characteristics and, for each alarm event, the reason, type and consequence.Measurements and main results The mean age of the patients included was 59.8 +/- 16.4 and SAPS1 was 15.9 +/- 7.4. We recorded 1971 h of care. The shift distribution was 78 mornings, 85 evenings and 83 nights; the stage distribution was 88 early, 78 intermediate and 80 late. There were 3188 alarms, an average of one alarm every 37 min: 23.7% were due to staff manipulation, 17.5% to technical problems and 58.8% to the patients. Alarms originated from ventilators (37.8%), cardiovascular monitors (32.7%), pulse oximeters (14.9%) and capnography (13.5%). Of the alarms, 25.8% had a consequence such as sensor repositioning, suction, modification of the therapy (drug or ventilation). Only 5.9% of the alarms led to a physician's being called. The positive predictive value of an alarm was 27% and its negative predictive value was 99%. The sensitivity was 97% and the specificity 58%.Conclusions The study confirms that the level of monitoring in ICUs generates a great number of false-positive alarms.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Lawless (1994) · doi:10.1097/00003246-199406000-00017 · Cited by 406 · see citing works
Adverse Environmental Conditions in the Respiratory and Medical ICU Settings
Meyer, T. J., Eveloff, S. E., Bauer, M. S., Schwartz, W. A., Hill, N. S., & Millman, R. P. (1994) · CHEST Journal · doi:10.1378/chest.105.4.1211 · Cited by 259 · see citing works
Environmental Noise as a Cause of Sleep Disruption in an Intermediate Respiratory Care Unit
Aaron, J. N., Carlisle, C. C., Carskadon, M. A., Meyer, T. J., Hill, N. S., & Millman, R. P. (1996) · SLEEP · doi:10.1093/sleep/19.9.707 · Cited by 232 · see citing works
Are there too many alarms in the intensive care unit? An overview of the problems
Meredith, C., & Edworthy, J. R. (1995) · Journal of Advanced Nursing · doi:10.1046/j.1365-2648.1995.21010015.x · Cited by 137 · see citing works
Noise pollution in the anaesthetic and intensive care environment
Kam, P. C. A., KAM, A. C., & Thompson, J. F. (1994) · Anaesthesia · doi:10.1111/j.1365-2044.1994.tb04319.x · Cited by 131 · see citing works
Cropp, A. J., Woods, L. A., Raney, D. J., & Bredle, D. L. (1994) · CHEST Journal · doi:10.1378/chest.105.4.1217 · Cited by 113 · see citing works
Balogh, D., Kittinger, E., Benzer, A., & Hackl, J. M. (1993) · Intensive Care Medicine · doi:10.1007/bf01694709 · Cited by 112 · see citing works
Intelligent patient monitoring and management systems: a review
Mora, F., PASSARIELLO, G., Carrault, G., & Pichon, J. (1993) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.248164 · Cited by 93 · see citing works
Noise levels in a cardiac surgical intensive care unit: a preliminary study conducted in secret
McLaughlin, A., McLaughlin, B., Elliott, J., & Campalani, G. (1996) · Intensive and Critical Care Nursing · doi:10.1016/s0964-3397(96)80106-6 · Cited by 37 · see citing works
Clinicians' opinions on alarm limits and urgency of therapeutic responses
Koski, E. M. J., Mäkivirta, A., Sukuvaara, T., & Kari, A. (1995) · International journal of clinical monitoring and computing · doi:10.1007/bf01142488 · Cited by 28 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Monitor Alarm Fatigue : An Integrative Review
Cvach, M. M. (2012) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-46.4.268 · Cited by 641 · see citing works
Machine Learning and Decision Support in Critical Care
Johnson, A. E. W., Ghassemi, M. M., Nemati, S., Niehaus, K. E., Clifton, D. A., & Clifford, G. D. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2015.2501978 · Cited by 438 · see citing works
Sendelbach, S. E., & Funk, M. (2013) · AACN Advanced Critical Care · doi:10.1097/nci.0b013e3182a903f9 · Cited by 497 · see citing works
Steinberg, J. S., Varma, N., Cygankiewicz, I., Aziz, P., Balsam, P., Baranchuk, A., Cantillon, D. J., Dilaveris, P., Dubner, S. J., El-Sherif, N., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.03.038 · Cited by 303 · see citing works
ECG Signal Quality During Arrhythmia and Its Application to False Alarm Reduction
Behar, J. A., Oster, J., Li, Q., & Clifford, G. D. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2240452 · Cited by 294 · see citing works
Sendelbach, S., & Funk, M. (2013) · AACN Advanced Critical Care · doi:10.4037/nci.0b013e3182a903f9 · Cited by 497 · see citing works
Alarm Algorithms in Critical Care Monitoring
Imhoff, M., & Kuhls, S. (2006) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000204385.01983.61 · Cited by 266 · see citing works
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 212 · see citing works
Continuous Monitoring in an Inpatient Medical-Surgical Unit: A Controlled Clinical Trial
Brown, H. V., Terrence, J., Vásquez, P., Bates, D. W., & Zimlichman, E. (2013) · The American Journal of Medicine · doi:10.1016/j.amjmed.2013.12.004 · Cited by 212 · see citing works
Intensive care unit alarms—How many do we need?*
Siebig, S., Kuhls, S., Imhoff, M., Ursula Gather, C., Schölmerich, J., & Wrede, C. E. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181cb0888 · Cited by 206 · see citing works
Escobar, G. J., LaGuardia, J. C., Turk, B. J., Ragins, A. I., Kipnis, P., & Draper, D. (2012) · Journal of Hospital Medicine · doi:10.1002/jhm.1929 · Cited by 185 · see citing works
Bonafide, C. P., Lin, R., Zander, M. I., Graham, C. S., Paine, C. W., Rock, W. L., Rich, A., Roberts, K. E., Fortino, M., Nadkarni, V., et al. (2015) · Journal of Hospital Medicine · doi:10.1002/jhm.2331 · Cited by 182 · see citing works
Neonatal and Pediatric Pulse Oximetry
Salyer, J. W. (2003) · Respiratory Care · doi:10.4187/respcare.03480386 · Cited by 183 · see citing works
Paine, C. W., Goel, V. V., Ely, E., Stave, C. D., Stemler, S., Zander, M., & Bonafide, C. P. (2016) · Journal of Hospital Medicine · doi:10.1002/jhm.2520 · Cited by 183 · see citing works
Alarms in the intensive care unit: how can the number of false alarms be reduced?
Chambrin, M.-C. (2001) · Critical Care · doi:10.1186/cc1021 · Cited by 176 · see citing works
Self-citationImproving detection of patient deterioration in the general hospital ward environment
Vincent, J.-L., Einav, S., Pearse, R., Jaber, S., Kranke, P., Overdyk, F. J., Whitaker, D. K., Gordo, F., Dahan, A., & Hoeft, A. (2018) · European Journal of Anaesthesiology · doi:10.1097/eja.0000000000000798 · Cited by 174 · see citing works
Validation of an electronic surveillance system for acute lung injury
Herasevich, V., Yılmaz, M., Khan, H., Hubmayr, R. D., & Gajic, O. (2009) · Intensive Care Medicine · doi:10.1007/s00134-009-1460-1 · Cited by 164 · see citing works
Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia
Ginosar, Y., Weiniger, C. F., Meroz, Y., Kurz, V., Bdolah‐Abram, T., Babchenko, A., Nitzan, M., & Davidson, E. M. (2009) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2009.01968.x · Cited by 157 · see citing works
Petter, A. H., Chioléro, R. L., Cassina, T., Chassot, P., Müller, X. M., & Revelly, J. (2003) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000086728.36285.be · Cited by 118 · see citing works
Steinberg, J. S., Varma, N., Cygankiewicz, I., Aziz, P. F., Balsam, P., Baranchuk, A. M., Cantillon, D. J., Dilaveris, P., Dubner, S., El‐Sherif, N., et al. (2017) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12447 · Cited by 118 · see citing works
Borowski, M., Görges, M., Fried, R., Such, O., Wrede, C. E., & Imhoff, M. (2011) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmt.2011.005 · Cited by 114 · see citing works
The Wolf Is Crying in the Operating Room
Schmid, F., Goepfert, M. S., Kuhnt, D., Eichhorn, V., Diedrichs, S., Reichenspurner, H. C., Goetz, A. E., & Reuter, D. A. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181fcc504 · Cited by 114 · see citing works
Zhao, Z., & Zhang, Y. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00727 · Cited by 113 · see citing works
A national online survey on the effectiveness of clinical alarms.
Korniewicz, D., Clark, T., & David, Y. (2008) · American Journal of Critical Care · doi:10.4037/ajcc2008.17.1.36 · Cited by 117 · see citing works
Medical audible alarms: a review
Edworthy, J. R. (2012) · Journal of the American Medical Informatics Association · doi:10.1136/amiajnl-2012-001061 · Cited by 105 · see citing works
Managing alarm systems for quality and safety in the hospital setting
Bach, T. A., Berglund, L.-M., & Turk, E. (2018) · BMJ Open Quality · doi:10.1136/bmjoq-2017-000202 · Cited by 101 · see citing works
Multimodal heart beat detection using signal quality indices
Johnson, A. E. W., Behar, J. A., Andreotti, F., Clifford, G. D., & Oster, J. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/8/1665 · Cited by 97 · see citing works
Patient monitoring alarms in the ICU and in the operating room
Schmid, F., Goepfert, M. S., & Reuter, D. A. (2013) · Critical Care · doi:10.1186/cc12525 · Cited by 96 · see citing works
Siebig, S., Kuhls, S., Imhoff, M., Langgartner, J., Reng, M., Schölmerich, J., Ursula Gather, C., & Wrede, C. E. (2009) · Journal of Critical Care · doi:10.1016/j.jcrc.2008.09.001 · Cited by 96 · see citing works
Attitudes and Practices Related to Clinical Alarms
Funk, M., Clark, J. T., Bauld, T. J., Ott, J. C., & Coss, P. (2014) · American Journal of Critical Care · doi:10.4037/ajcc2014315 · Cited by 91 · see citing works
Smart alarms from medical devices in the OR and ICU
Imhoff, M., Kuhls, S., Ursula Gather, C., & Fried, R. (2009) · Best practice & research. Clinical anaesthesiology · doi:10.1016/j.bpa.2008.07.008 · Cited by 87 · see citing works
Reducing Hospital Noise: A Review of Medical Device Alarm Management
Konkani, A., Oakley, B. A., & Bauld, T. J. (2012) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-46.6.478 · Cited by 85 · see citing works
Alarm fatigue and its influence on staff performance
Deb, S., & Claudio, D. (2015) · IIE Transactions on Healthcare Systems Engineering · doi:10.1080/19488300.2015.1062065 · Cited by 74 · see citing works
Rivera-Fernández, R., Nap, R. E., Vázquez-Mata, G., & Miranda, D. R. (2007) · Journal of Critical Care · doi:10.1016/j.jcrc.2006.09.005 · Cited by 66 · see citing works
Hall, T. C., Bilku, D. K., Al-Leswas, D., Neal, C. P., Horst, C., Cooke, J., Metcalfe, M. S., & Dennison, A. R. (2014) · Journal of Parenteral and Enteral Nutrition · doi:10.1177/0148607113518945 · Cited by 67 · see citing works
Borowski, M., Siebig, S., Wrede, C. E., & Imhoff, M. (2011) · Computational and Mathematical Methods in Medicine · doi:10.1155/2011/143480 · Cited by 63 · see citing works
Use of Pagers With an Alarm Escalation System to Reduce Cardiac Monitor Alarm Signals
Cvach, M. M., Frank, R. J., Doyle, P. A., & Khouri-Stevens, Z. (2013) · Journal of Nursing Care Quality · doi:10.1097/ncq.0b013e3182a61887 · Cited by 63 · see citing works
Luo, X., Li, H., Huang, T., & Rose, T. M. (2016) · Safety Science · doi:10.1016/j.ssci.2015.12.026 · Cited by 60 · see citing works
Nurses' Perspectives on Clinical Alarms
Honan, L., Funk, M., Maynard, M. A., Fahs, D. B., Clark, J. T., & David, Y. B. (2015) · American Journal of Critical Care · doi:10.4037/ajcc2015552 · Cited by 59 · see citing works
GAZARIAN, P. K. (2013) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2013.05.014 · Cited by 59 · see citing works
Patient-specific learning in real time for adaptive monitoring in critical care
Zhang, Y., & Peter Szolovits, B. (2008) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2008.03.011 · Cited by 58 · see citing works
Hu, S. B., Wong, D. J. L., Correa, A., Li, N., & Deng, J. C. (2016) · PLoS ONE · doi:10.1371/journal.pone.0161401 · Cited by 58 · see citing works
Zimlichman, E., Szyper‐Kravitz, M., Shinar, Z., Klap, T., Levkovich, S., Unterman, A., Rozenblum, R., Rothschild, J. M., Amital, H., & Shoenfeld, Y. (2012) · Journal of Hospital Medicine · doi:10.1002/jhm.1963 · Cited by 56 · see citing works
Sendelbach, S. E. (2012) · Nursing Clinics of North America · doi:10.1016/j.cnur.2012.05.009 · Cited by 52 · see citing works
Measurement of Physiological Monitor Alarm Accuracy and Clinical Relevance in Intensive Care Units
Ruppel, H., Funk, M., & Whittemore, R. (2018) · American Journal of Critical Care · doi:10.4037/ajcc2018385 · Cited by 49 · see citing works
False alarm reduction in critical care
Clifford, G. D., Silva, I., Moody, B., Li, Q., Kella, D., Chahin, A. B., Kooistra, T., Perry, D., & Mark, R. G. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/e5 · Cited by 48 · see citing works
Kobayashi, L., Parchuri, R., Gardiner, F. G., Paolucci, G. A., Tomaselli, N. M., Al-Rasheed, R. S., Bertsch, K., Devine, J., Boss, R. M., Gibbs, F. J., et al. (2012) · BMJ Quality & Safety · doi:10.1136/bmjqs-2012-001134 · Cited by 47 · see citing works
Phillips, J. K., & Barnsteiner, J. (2005) · Critical Care Nursing Quarterly · doi:10.1097/00002727-200510000-00003 · Cited by 45 · see citing works
R-Peak Estimation using Multimodal Lead Switching
Johnson, A. E. W., Behar, J. A., Andreotti, F., Clifford, G. D., & Oster, J. (2014) · HAL · Cited by 45 · see citing works
Mechanical Ventilation Alarms and Alarm Fatigue
Scott, J. B., Vaux, L. D., Dills, C. C., & Strickland, S. L. (2019) · Respiratory Care · doi:10.4187/respcare.06878 · Cited by 45 · see citing works
Reducing False Intracranial Pressure Alarms Using Morphological Waveform Features
Scalzo, F., Liebeskind, D. S., & Hu, X. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2210042 · Cited by 43 · see citing works
Predictive combinations of monitor alarms preceding in-hospital code blue events
Hu, X., Sapo, M., Nenov, V., Barry, T., Kim, S., Do, D. H., Boyle, N., & Martin, N. A. (2012) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2012.03.001 · Cited by 43 · see citing works
Olive, M. K., & Owens, G. E. (2018) · Translational Pediatrics · doi:10.21037/tp.2018.04.03 · Cited by 43 · see citing works
Rossetti, A. (2020) · Music and Medicine · doi:10.47513/mmd.v12i2.742 · Cited by 41 · see citing works
Detection of PVC by using a wavelet-based statistical ECG monitoring procedure
Jung, Y., & Kim, H. (2017) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2017.03.023 · Cited by 42 · see citing works
Addressing the flaws of current critical alarms: a fuzzy constraint satisfaction approach
Otero, A., Félix, P., Barro, S., & Palacios, F. (2009) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2009.08.002 · Cited by 42 · see citing works
Pimentel, M. A. F., Santos, M., Springer, D., & Clifford, G. D. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/8/1717 · Cited by 41 · see citing works
Effect of Altering Alarm Settings: A Randomized Controlled Study
Cvach, M. M., Rothwell, K. J., Cullen, A. M., Nayden, M. G., Cvach, N., & Pham, J. C. (2015) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-49.3.214 · Cited by 40 · see citing works
Intra‐operative monitoring – many alarms with minor impact
de Man, F. R., Greuters, S., Boer, C., Veerman, D. P., & Loer, S. A. (2013) · Anaesthesia · doi:10.1111/anae.12289 · Cited by 38 · see citing works
A Framework for Reducing Alarm Fatigue on Pediatric Inpatient Units
Karnik, A., & Bonafide, C. P. (2015) · Hospital Pediatrics · doi:10.1542/hpeds.2014-0123 · Cited by 38 · see citing works
Connecting the dots: rule-based decision support systems in the modern EMR era
Herasevich, V., Kor, D. J., Subramanian, A., & Pickering, B. W. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9445-6 · Cited by 37 · see citing works
Machine learning-data mining integrated approach for premature ventricular contraction prediction
Mastoi, Q.-U.-A., Memon, M. S., Lakhan, A. R., Mohammed, M. A., Qabulio, M., Al-Turjman, F. M., & Abdulkareem, K. H. (2021) · Neural Computing and Applications · doi:10.1007/s00521-021-05820-2 · Cited by 36 · see citing works
Cantillon, D. J., Loy, M., Burkle, A., Pengel, S., Brosovich, D., Hamilton, A., Khot, U. N., & Lindsay, B. D. (2016) · JAMA · doi:10.1001/jama.2016.10258 · Cited by 35 · see citing works
Bridi, A. C., da Silva, R. C. L., de Farias, C. C. P., Franco, A. S., & dos Santos, V. D. L. Q. (2014) · Revista Brasileira de Terapia Intensiva · doi:10.5935/0103-507x.20140005 · Cited by 35 · see citing works
Advanced Integrated Real-Time Clinical Displays
Kruger, G. H., & Tremper, K. K. (2011) · Anesthesiology Clinics · doi:10.1016/j.anclin.2011.05.004 · Cited by 35 · see citing works
Over-monitoring and alarm fatigue: For whom do the bells toll?
Feder, S. L., & Funk, M. (2013) · Heart & Lung · doi:10.1016/j.hrtlng.2013.09.001 · Cited by 32 · see citing works
Ginosar, Y., Weiniger, C. F., Kurz, V., Babchenko, A., Nitzan, M., & Davidson, E. M. (2009) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/s12630-008-9036-z · Cited by 32 · see citing works
Schlafstörungen bei kritisch kranken Patienten
Walder, B., Haase, U., & Rundshagen, I. (2006) · Der Anaesthesist · doi:10.1007/s00101-006-1086-4 · Cited by 32 · see citing works
Görges, M., Westenskow, D. R., & Markewitz, B. A. (2012) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-012-9370-0 · Cited by 31 · see citing works
Wolken, R., Woodruff, R. J., Smith, J., Albert, R. K., & Douglas, I. S. (2011) · Respiratory Care · doi:10.4187/respcare.01453 · Cited by 31 · see citing works
Balaguera, H. U., Wise, D., Ng, C. Y., Tso, H.-W., Chiang, W., Hutchinson, A. M., Galvin, T., Hilborne, L. H., Hoffman, C., Huang, C.-C., et al. (2017) · Journal of Medical Internet Research · doi:10.2196/jmir.7131 · Cited by 30 · see citing works
Differences in Alarm Events Between Disposable and Reusable Electrocardiography Lead Wires
Albert, N. M., Murray, T., Bena, J., Slifcak, E., Roach, J. D., Spence, J., & Burkle, A. (2014) · American Journal of Critical Care · doi:10.4037/ajcc2015663 · Cited by 30 · see citing works
Managing Alarms in Acute Care Across the Life Span: Electrocardiography and Pulse Oximetry
(2018) · Critical Care Nurse · doi:10.4037/ccn2018468 · Cited by 31 · see citing works
Inokuchi, R., Satō, H., Nanjo, Y., Echigo, M., Tanaka, A., Ishii, T., Matsubara, T., Doi, K., Gunshin, M., Hiruma, T., et al. (2013) · BMJ Open · doi:10.1136/bmjopen-2013-003354 · Cited by 29 · see citing works
Blum, J. M., Kruger, G. H., Sanders, K. L., Gutierrez, J., & Rosenberg, A. L. (2009) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-008-9159-3 · Cited by 29 · see citing works
van Pul, C., Mortel, H. P. M. E. V. D., Bogaart, J. J. L. V. D., Mohns, T., & Andriessen, P. (2015) · Acta Paediatrica · doi:10.1111/apa.12907 · Cited by 29 · see citing works
Welinder, A., Sörnmo, L., Feild, D. Q., Feldman, C. L., Pettersson, J., Wagner, G. S., & Pahlm, O. (2004) · American Journal of Critical Care · doi:10.4037/ajcc2004.13.3.228 · Cited by 29 · see citing works
Blum, J. M., & Tremper, K. K. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181bfe97f · Cited by 29 · see citing works
A data mining approach to reduce the false alarm rate of patient monitors
Baumgartner, B., Rodel, K., & Knoll, A. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6347345 · Cited by 28 · see citing works
Integrating monitor alarms with laboratory test results to enhance patient deterioration prediction
Bai, Y., Do, D. H., Harris, P. R. E., Schindler, D. M., Boyle, N. G., Drew, B. J., & Hu, X. (2014) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2014.09.006 · Cited by 28 · see citing works
Blankush, J. M., Freeman, R., McIlvaine, J., Tran, T. L., Nassani, S. J., & Leitman, I. M. (2016) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-016-9943-4 · Cited by 28 · see citing works
A fuzzy logic-based system for anaesthesia monitoring
Mirza, M., GholamHosseini, H., & Harrison, M. J. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5627987 · Cited by 27 · see citing works
Semi-supervised detection of intracranial pressure alarms using waveform dynamics
Scalzo, F., & Hu, X. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/4/465 · Cited by 27 · see citing works
Charbonnier, S., & Gentil, S. (2010) · International Journal of Adaptive Control and Signal Processing · doi:10.1002/acs.1123 · Cited by 26 · see citing works
Safety analysis of proposed data‐driven physiologic alarm parameters for hospitalized children
Goel, V. V., Poole, S. F., Longhurst, C., Platchek, T. S., Pageler, N. M., Sharek, P. J., & Palma, J. P. (2016) · Journal of Hospital Medicine · doi:10.1002/jhm.2635 · Cited by 26 · see citing works
Borkowska, M., Labeau, S. O., Schepens, T., Vandijck, D. M., de Vyver, K. V., Christiaens, D., Lizy, C., Blackwood, B., & Blot, S. I. (2018) · American Journal of Critical Care · doi:10.4037/ajcc2018959 · Cited by 25 · see citing works
Forecasting intracranial hypertension using multi-scale waveform metrics
Hüser, M., Kündig, A., Karlen, W., Luca, V. D., & Jaggi, M. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab6360 · Cited by 25 · see citing works
Video Methods for Evaluating Physiologic Monitor Alarms And Alarm Responses
Bonafide, C. P., Zander, M. I., Graham, C. S., Paine, C. W., Rock, W. L., Rich, A., Roberts, K. E., Fortino, M., Nadkarni, V., Lin, R., et al. (2014) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-48.3.220 · Cited by 24 · see citing works
A remote monitoring system with early diagnosis of hypertension and Hypotension
Baig, M. M., & GholamHosseini, H. (2013) · doi:10.1109/pht.2013.6461278 · Cited by 24 · see citing works
Implementation and experience of an innovative smart patient care system: a cross-sectional study
Wen, M.-H., Bai, D. L., Lin, S., Chu, C., & Hsu, Y. (2022) · BMC Health Services Research · doi:10.1186/s12913-022-07511-7 · Cited by 22 · see citing works
Users' Opinions on Intensive Care Unit Alarms — a Survey of German Intensive Care Units
Siebig, S., Sieben, W., Kollmann, F., Imhoff, M., Bruennler, T., Rockmann, F., Ursula Gather, C., & Wrede, C. E. (2009) · Anaesthesia and Intensive Care · doi:10.1177/0310057x0903700116 · Cited by 22 · see citing works
Hernandez, L., Kim, R. B., Tokcan, N., Derksen, H., Biesterveld, B. E., Croteau, A. J., Williams, A. M., Mathis, M. R., Najarian, K., & Gryak, J. (2021) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2021.102032 · Cited by 22 · see citing works
Enhanced Notification of Critical Ventilator Events
Evans, R. S., Johnson, K. V., Flint, V. B., Kinder, T., Lyon, C. R., Hawley, W. L., Vawdrey, D. K., & Thomsen, G. E. (2005) · Journal of the American Medical Informatics Association · doi:10.1197/jamia.m1863 · Cited by 22 · see citing works
Understanding Clinical Alarm Safety
Lukasewicz, C. L., & Mattox, E. A. (2015) · Critical Care Nurse · doi:10.4037/ccn2015113 · Cited by 21 · see citing works
Kawagishi, T., Kanaya, N., Nakayama, M., Kurosawa, S., & Namiki, A. (2004) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000130343.66453.37 · Cited by 21 · see citing works
THE FUTURE OF BEDSIDE MONITORING
Kohli-Seth, R. D., & Oropello, J. M. (2000) · Critical Care Clinics · doi:10.1016/s0749-0704(05)70134-2 · Cited by 20 · see citing works
Critical Care Nurses’ Cognitive Ergonomics Related to Medical Device Alarms
Wung, S., & Schatz, M. R. (2018) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2018.02.002 · Cited by 20 · see citing works
Cvach, M. M., Stokes, J. E., Manzoor, S. H., Brooks, P., Burger, T., Gottschalk, A., & Pustavoitau, A. (2018) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000003801 · Cited by 20 · see citing works
Wen, M.-H., Chen, P.-Y., Lin, S., Lien, C.-W., Tu, S.-H., Chueh, C.-Y., Wu, Y. ‐., Kian, K. T. C., Hsu, Y.-L., & Bai, D. L. (2024) · Journal of Medical Internet Research · doi:10.2196/58380 · Cited by 19 · see citing works
Safe Use of Vasopressin and Angiotensin II for Patients with Circulatory Shock
Bauer, S. R., Sacha, G. L., & Lam, S. W. (2018) · Pharmacotherapy The Journal of Human Pharmacology and Drug Therapy · doi:10.1002/phar.2147 · Cited by 18 · see citing works
- Authors
- Chambrin, M.-C., Ravaux, P., Calvelo-Aros, D., Jaborska, A., Chopin, C., & Boniface, B.
- Affiliations
- Inserm (FR)
- Université de Lille (FR)
- ORCID
- Ravaux, P.
- Venue
- Intensive Care Medicine
- Published
- 1999-12
- Volume
- 25
- Issue
- 12
- Pages
- 1360-1366
- ISSN
- 0342-4642 (ISSN-L)
- 1432-1238
- OpenAlex
- W1997522188
- Cited by
- 227
- References
- 10
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 7 · 2023: 6 · 2022: 4 · 2021: 6 · 2020: 9 · 2019: 6 · 2018: 14 · 2017: 11 · 2016: 13 · 2015: 22 · 2014: 19 · 2013: 21 · 2012: 15
- FWCI
- 3.16
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- Intensive Care Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Respiratory Support and Mechanisms
- Keywords
- alarm fatigue, false positive alerts, intensive care unit, mechanical ventilation, patient monitoring, alarm response, positive predictive value, pulse oximetry, capnography
- MeSH terms
- Adult, Electrocardiography, Equipment Failure, False Positive Reactions, Female, France, Hospitals, General, Hospitals, University, Humans, Intensive Care Units, Length of Stay, Male, Middle Aged, Monitoring, Physiologic, Oximetry, Predictive Value of Tests, Prospective Studies, Respiration, Artificial, Sensitivity and Specificity, Severity of Illness Index, Safety Management, Capnography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review