Harmonizing FDG PET quantification while maintaining optimal lesion detection: prospective multicentre validation in 517 oncology patients
Open accessAbstract
PURPOSE: Point-spread function (PSF) or PSF + time-of-flight (TOF) reconstruction may improve lesion detection in oncologic PET, but can alter quantitation resulting in variable standardized uptake values (SUVs) between different PET systems. This study aims to validate a proprietary software tool (EQ.PET) to harmonize SUVs across different PET systems independent of the reconstruction algorithm used. METHODS: NEMA NU2 phantom data were used to calculate the appropriate filter for each PSF or PSF+TOF reconstruction from three different PET systems, in order to obtain EANM compliant recovery coefficients. PET data from 517 oncology patients were reconstructed with a PSF or PSF+TOF reconstruction for optimal tumour detection and an ordered subset expectation maximization (OSEM3D) reconstruction known to fulfil EANM guidelines. Post-reconstruction, the proprietary filter was applied to the PSF or PSF+TOF data (PSFEQ or PSF+TOFEQ). SUVs for PSF or PSF+TOF and PSFEQ or PSF+TOFEQ were compared to SUVs for the OSEM3D reconstruction. The impact of potential confounders on the EQ.PET methodology including lesion and patient characteristics was studied, as was the adherence to imaging guidelines. RESULTS: For the 1380 tumour lesions studied, Bland-Altman analysis showed a mean ratio between PSF or PSF+TOF and OSEM3D of 1.46 (95%CI: 0.86-2.06) and 1.23 (95%CI: 0.95-1.51) for SUVmax and SUVpeak, respectively. Application of the proprietary filter improved these ratios to 1.02 (95%CI: 0.88-1.16) and 1.04 (95%CI: 0.92-1.17) for SUVmax and SUVpeak, respectively. The influence of the different confounding factors studied (lesion size, location, radial offset and patient's BMI) was less than 5%. Adherence to the European Association of Nuclear Medicine (EANM) guidelines for tumour imaging was good. CONCLUSION: These data indicate that it is not necessary to sacrifice the superior lesion detection and image quality achieved by newer reconstruction techniques in the quest for harmonizing quantitative comparability between PET systems.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Compared with non-PSF reconstructed images, PSF can increase lesion SUVmax, SUVmean and SUVpeak by up to 30% on average.
“Compared with non-PSF reconstructed images, this can increase lesion SUVmax, SUVmean and SUVpeak by up to 30% on average [100,101,102].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [100]
The Results (Patient studies; page unavailable) show that PSF reconstruction increased lesion SUVmax and SUVpeak relative to conventional OSEM, but the reported mean increases varied by centre: SUVmax reached 57%, while SUVpeak reached 28%. The cited passage does not establish an increase of up to 30% on average for lesion SUVmean, so the three-metric claim is not fully supported. (LLM, conf=0.89)
Experimental — accuracy not yet measuredWith appropriate filter width based on sphere CR from standardized phantom measurements, PSF-induced SUV increases can be negated, resulting in comparable lesion SUVs to non-PSF data.
“With appropriate filter width (FWHM) based on sphere CR from standardized phantom measurements, PSF-induced SUV increases can be negated, resulting in comparable lesion SUVs to those in non-PSF data.”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [100]
PSF reconstruction improves reconstructed spatial resolution and increases lesion SUV but can lead to overestimation of true activity due to Gibbs' artifacts.
“This improves reconstructed spatial resolution [97,98] and increases lesion SUV [99,100] but can lead to overestimation of the true activity due to so-called Gibbs’ artifacts [92,93].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [100]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid Tumors
Wahl, R. L., Jacene, H., Kasamon, Y., & Lodge, M. A. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.057307 · Cited by 3,777 · see citing works
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
Boellaard, R., Delgado-Bolton, R., Oyen, W. J. G., Giammarile, F., Tatsch, K., Eschner, W., Verzijlbergen, F. J., Barrington, S. F., Pike, L. C., Weber, W. A., et al. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2961-x · Cited by 3,303 · see citing works
Young, H. E., Baum, R. P., Cremerius, U., Herholz, K., Hoekstra, O. S., Lammertsma, A. A., Pruim, J., & Price, P. (1999) · European Journal of Cancer · doi:10.1016/s0959-8049(99)00229-4 · Cited by 1,716 · see citing works
Barrington, S. F., Mikhaeel, N. G., Kostakoğlu, L., Meignan, M. A., Hutchings, M., Müeller, S., Schwartz, L. H., Zucca, E., Fisher, R. I., Trotman, J., et al. (2014) · Journal of Clinical Oncology · doi:10.1200/jco.2013.53.5229 · Cited by 1,663 · see citing works
Self-citationFDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0
Boellaard, R., O’Doherty, M. J., Weber, W., Mottaghy, F. M., Lonsdale, M. N., Stroobants, S. G., Oyen, W. J. G., Kotzerke, J., Hoekstra, O. S., Pruim, J., et al. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1297-4 · Cited by 1,329 · see citing works
Procedure guideline for tumor imaging with F-18 FDG PET/CT
Delbeke, D., Coleman, R. E., Guiberteau, M. J., Brown, M. L., Royal, H. D., Siegel, B. A., Townsend, D. W., Berland, L. L., Parker, J. A., Hübner, K. F., et al. (2006) · Cited by 620 · see citing works
Physical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 487 · see citing works
Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls
Rahmim, A., Qi, J., & Sossi, V. (2013) · Medical Physics · doi:10.1118/1.4800806 · Cited by 354 · see citing works
Procedure guideline for tumor imaging with 18F-FDG PET/CT 1.0.
Delbeke, D., Coleman, R. E., Guiberteau, M. J., Brown, M. L., Royal, H. D., Siegel, B. A., Townsend, D. W., Berland, L. L., Parker, J. A., Hubner, K., et al. (2006) · PubMed · Cited by 297 · see citing works
Westerterp, M., Pruim, J., Oyen, W. J. G., Hoekstra, O. S., Paans, A. M. J., Visser, E. P., van Lanschot, J. J. B., Sloof, G. W., & Boellaard, R. (2006) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-006-0224-1 · Cited by 294 · see citing works
Noise Considerations for PET Quantification Using Maximum and Peak Standardized Uptake Value
Lodge, M. A., Chaudhry, M. Q., & Wahl, R. L. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.101733 · Cited by 226 · see citing works
Application and Evaluation of a Measured Spatially Variant System Model for PET Image Reconstruction
Alessio, A. M., Stearns, C. W., Tong, S., Ross, S. G., Kohlmyer, S. G., Ganin, A., & Kinahan, P. E. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2010.2040188 · Cited by 214 · see citing works
Need for Standardization of 18F-FDG PET/CT for Treatment Response Assessments
Boellaard, R. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.085662 · Cited by 159 · see citing works
Hasenclever, D., Kurch, L., Mauz-Körholz, C., Elsner, A., Georgi, T., Wallace, H., Landman-Parker, J., Moryl-Bujakowska, A., Cepelová, M., Karlén, J., et al. (2014) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2715-9 · Cited by 157 · see citing works
Sunderland, J. J., & Christian, P. E. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.148056 · Cited by 144 · see citing works
Lasnon, C., Desmonts, C., Quak, E., Gervais, R., Dô, P., Dubos‐Arvis, C., & Aide, N. (2013) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-013-2391-1 · Cited by 126 · see citing works
Self-citationSummary of the UPICT Protocol for 18F-FDG PET/CT Imaging in Oncology Clinical Trials
Graham, M. M., Wahl, R. L., Hoffman, J. M., Yap, J. T., Sunderland, J. J., Boellaard, R., Perlman, E. S., Kinahan, P. E., Christian, P. E., Hoekstra, O. S., et al. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.158402 · Cited by 103 · see citing works
Akamatsu, G., Mitsumoto, K., Taniguchi, T., Tsutsui, Y., Baba, S., & Sasaki, M. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2013.09.030 · Cited by 96 · see citing works
Makris, N. E., Huisman, M. C., Kinahan, P. E., Lammertsma, A. A., & Boellaard, R. (2013) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-013-2465-0 · Cited by 90 · see citing works
Lasnon, C., Hicks, R. J., Beauregard, J., Milner, A. D., Paciencia, M., Guizard, A., Bardet, S., Gervais, R., Lemoel, G., Zalcman, G., et al. (2012) · Clinical Nuclear Medicine · doi:10.1097/rlu.0b013e318251e3d1 · Cited by 65 · see citing works
Self-citationBrendle, C., Kupferschläger, J., Nikolaou, K., la Fougère, C. J., Gatidis, S., & Pfannenberg, C. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.10.018 · Cited by 63 · see citing works
SUVref: reducing reconstruction-dependent variation in PET SUV
Kelly, M. D., & Declerck, J. M. (2011) · EJNMMI Research · doi:10.1186/2191-219x-1-16 · Cited by 58 · see citing works
Bellèvre, D., Blanc‐Fournier, C., Switsers, O., Dugué, A. E., Lévy, C., Allouache, D., Desmonts, C., Crouet, H., Guilloit, J., Grellard, J., et al. (2014) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2689-7 · Cited by 58 · see citing works
Self-citationEuropean Association of Nuclear Medicine
&NA; (1986) · Nuclear Medicine Communications · doi:10.1097/00006231-198609000-00001 · Cited by 43 · see citing works
Lee, Y. S., Kim, J. S., Kim, K. M., Kang, J. H., Lim, S. M., & Kim, H.-J. (2014) · Annals of Nuclear Medicine · doi:10.1007/s12149-014-0815-z · Cited by 39 · see citing works
Gibbs Artifact Reduction by Nonnegativity Constraint
Zeng, G. L. (2011) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.110.086439 · Cited by 34 · see citing works
Bergmann, H., Geist, B. K., Schaffarich, M. P., Hirtl, A., Hacker, M., Beyer, T., & Rausch, I. (2014) · Nuklearmedizin - NuclearMedicine · doi:10.3413/nukmed-0665-14-05 · Cited by 33 · see citing works
Rausch, I., Bergmann, H., Geist, B. K., Schaffarich, M. P., Hirtl, A., Hacker, M., & Beyer, T. (2014) · Nuklearmedizin - NuclearMedicine · Cited by 27 · see citing works
Quak, E., Hovhannisyan, N., Lasnon, C., Fruchart, C., Vilque, J., Musafiri, D., & Aide, N. (2014) · Haematologica · doi:10.3324/haematol.2014.104125 · Cited by 25 · see citing works
Self-citation(2018) · Springer Publishing Company eBooks · doi:10.1891/9780826168733 · Cited by 23 · see citing works
Flagged first, then by citations
1 of the 33 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Aide, N., Lasnon, C., Veit-Haibach, P., Sera, T., Sattler, B., & Boellaard, R. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3740-2 · Cited by 344 · see citing works
Self-citationMeignan, M. A., Cottereau, A., Versari, A., Chartier, L., Dupuis, J., Boussetta, S., Grassi, I., Casasnovas, O., Haïoun, C., Tilly, H., et al. (2016) · Journal of Clinical Oncology · doi:10.1200/jco.2016.66.9440 · Cited by 325 · see citing works
Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET
van der Vos, C. S., Koopman, D., Rijnsdorp, S., Arends, A. J., Boellaard, R., van Dalen, J. A., Lubberink, M., Willemsen, A. T. M., & Visser, E. P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3727-z · Cited by 206 · see citing works
How We Read Oncologic FDG PET/CT
Hofman, M. S., & Hicks, R. J. (2016) · Cancer Imaging · doi:10.1186/s40644-016-0091-3 · Cited by 155 · see citing works
Self-citationRadiomics in Nuclear Medicine Applied to Radiation Therapy: Methods, Pitfalls, and Challenges
Reuzé, S., Schernberg, A., Orlhac, F., Sun, R., Chargari, C., Dercle, L., Deutsch, É., Buvat, I., & Robert, C. (2018) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2018.05.022 · Cited by 126 · see citing works
Clinical evaluation of the radiolanthanide terbium-152: first-in-human PET/CT with 152 Tb-DOTATOC
Baum, R. P., Singh, A., Benešová, M., Vermeulen, C. E., Gnesin, S., Köster, U., Johnston, K., Müller, D., Senftleben, S., Kulkarni, H., et al. (2017) · Dalton Transactions · doi:10.1039/c7dt01936j · Cited by 94 · see citing works
Point-spread function reconstructed PET images of sub-centimeter lesions are not quantitative
Munk, O. L., Tolbod, L. P., Hansen, S. B., & Bogsrud, T. V. (2017) · EJNMMI Physics · doi:10.1186/s40658-016-0169-9 · Cited by 84 · see citing works
New PET technologies – embracing progress and pushing the limits
Aide, N., Lasnon, C., Kesner, A., Levin, C. S., Buvat, I., Iagaru, A., Hermann, K., Badawi, R. D., Cherry, S. R., Bradley, K. M., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05390-4 · Cited by 73 · see citing works
Self-citationWong, A. N. M., Callahan, J., Keyaerts, M., Neyns, B., Mangana, J., Aberle, S., Herschtal, A., Fullerton, S., Milne, D., Iravani, A., et al. (2020) · Cancer Imaging · doi:10.1186/s40644-020-00313-2 · Cited by 71 · see citing works
Self-citationMetabolic Tumor Volume in Lymphoma: Hype or Hope?
Schöder, H. M., & Moskowitz, C. H. (2016) · Journal of Clinical Oncology · doi:10.1200/jco.2016.69.3747 · Cited by 68 · see citing works
Cottereau, A.-S., Hapdey, S., Chartier, L., Modzelewski, R., Casasnovas, O., Itti, E., Tilly, H., Vera, P., Meignan, M. A., & Becker, S. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.180406 · Cited by 64 · see citing works
Repetitive 18F-FDG-PET/CT in patients with large-vessel giant-cell arteritis and controlled disease
de Boysson, H., Aide, N., Liozon, É., Lambert, M., Parienti, J., Monteil, J., Huglo, D. G., Bienvenu, B., Manrique, A., & Aouba, A. (2017) · European Journal of Internal Medicine · doi:10.1016/j.ejim.2017.08.013 · Cited by 63 · see citing works
Self-citationWeyts, K., Lasnon, C., Ciappuccini, R., Lequesne, J., Corroyer-Dulmont, A., Quak, E., Clarisse, B., Roussel, L., Bardet, S., & Jaudet, C. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05800-1 · Cited by 59 · see citing works
Self-citationNCTN Assessment on Current Applications of Radiomics in Oncology
Nie, K., Al‐Hallaq, H., Li, X. A., Benedict, S., Sohn, J. W., Moran, J. M., Yong Fan, A., Huang, M., Knopp, M. V., Michalski, J. M., et al. (2019) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2019.01.087 · Cited by 63 · see citing works
Fournier, L., de Geus-Oei, L. D., Regge, D., Oprea-Lager, D., D'Anastasi, M., Bidaut, L., Bäuerle, T., Lopci, E., Cappello, G., Lecouvet, F., et al. (2022) · Frontiers in Oncology · doi:10.3389/fonc.2021.800547 · Cited by 49 · see citing works
Lasnon, C., Majdoub, M., Lavigne, B., Dô, P., Madelaine, J., Visvikis, D., Hatt, M., & Aide, N. (2016) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-016-3441-2 · Cited by 53 · see citing works
Self-citationInfluences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 50 · see citing works
“To correct for these increases, an additional Gaussian filter can be applied during image reconstruction or to the final images [100,102].”
BackgroundA review of harmonization strategies for quantitative PET
Akamatsu, G., Tsutsui, Y., Daisaki, H., Mitsumoto, K., Baba, S., & Sasaki, M. (2023) · Annals of Nuclear Medicine · doi:10.1007/s12149-022-01820-x · Cited by 49 · see citing works
Rufini, V., Garganese, G., Ieria, F. P., Pasciuto, T., Fragomeni, S. M., Gui, B., Florit, A., Inzani, F. S., Zannoni, G. F., Scambia, G., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05257-8 · Cited by 49 · see citing works
Bone Metastases Are Measurable: The Role of Whole-Body MRI and Positron Emission Tomography
Oprea‐Lager, D. E., Cysouw, M. C. F., Boellaard, R., Deroose, C. M., de Geus‐Oei, L., Lopci, E., Bidaut, L. M., Herrmann, K., Fournier, L. S., Bäuerle, T., et al. (2021) · Frontiers in Oncology · doi:10.3389/fonc.2021.772530 · Cited by 49 · see citing works
Experimental validation of absolute SPECT/CT quantification for response monitoring in breast cancer
Collarino, A., Arias‐Bouda, L. M. P., Olmos, R. A. V., van der Tol, P., Dibbets‐Schneider, P., de Geus‐Oei, L., & van Velden, F. H. P. (2018) · Medical Physics · doi:10.1002/mp.12880 · Cited by 42 · see citing works
Does PET SUV Harmonization Affect PERCIST Response Classification?
Quak, E., Roux, P. L., Lasnon, C., Robin, P., Hofman, M. S., Bourhis, D., Callahan, J., Binns, D., Desmonts, C., Salaün, P., et al. (2016) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.171983 · Cited by 41 · see citing works
Self-citationLasnon, C., Salomon, T., Desmonts, C., Dô, P., Oulkhouir, Y., Madelaine, J., & Aide, N. (2016) · Annals of Nuclear Medicine · doi:10.1007/s12149-016-1135-2 · Cited by 41 · see citing works
Self-citationDevriese, J., Beels, L., Maes, A., de Wiele, C. V., & Pottel, H. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0235-6 · Cited by 37 · see citing works
Lasnon, C., Enilorac, B., Popotte, H., & Aide, N. (2017) · EJNMMI Research · doi:10.1186/s13550-017-0279-y · Cited by 35 · see citing works
Self-citationThe value of the Standardized Uptake Value (SUV) and Metabolic Tumor Volume (MTV) in lung cancer.
Hicks, R. (2022) · Seminars in nuclear medicine · doi:10.1053/j.semnuclmed.2022.04.007 · Cited by 33 · see citing works
Self-citationRufini, V., Lorusso, M., Inzani, F. S., Pasciuto, T., Triumbari, E. K. A., Grillo, L. R., Locco, F., Margaritora, S., Pescarmona, E. O., & Rindi, G. (2022) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-022-05848-z · Cited by 31 · see citing works
Reducing Radiation Exposure to Paediatric Patients Undergoing [18F]FDG-PET/CT Imaging
Kertész, H., Beyer, T., London, K., Saleh, H., Chung, D. K. V., Rausch, I., Cal-González, J., Kitsos, T., & Kench, P. L. (2021) · Molecular Imaging and Biology · doi:10.1007/s11307-021-01601-4 · Cited by 31 · see citing works
J-Hart study group, Tsutsui, Y., Daisaki, H., Akamatsu, G., Umeda, T., Ogawa, M., Kajiwara, H., Kawase, S., Sakurai, M., Nishida, H., et al. (2018) · EJNMMI Research · doi:10.1186/s13550-018-0438-9 · Cited by 29 · see citing works
Houdu, B., Lasnon, C., Licaj, I., Thomas, G., Do, P., Guizard, A.-V., Desmonts, C., & Aide, N. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4151-8 · Cited by 37 · see citing works
Self-citationBoellaard, R. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-015-3182-7 · Cited by 28 · see citing works
Beichel, R., Smith, B. J., Bauer, C., Ulrich, E. J., Ahmadvand, P., Budzevich, M. M., Gillies, R. J., Goldgof, D. B., Grkovski, M., Hamarneh, G., et al. (2016) · Medical Physics · doi:10.1002/mp.12041 · Cited by 28 · see citing works
Taralli, S., Scolozzi, V., Foti, M., Ricciardi, S., Forcione, A. R., Cardillo, G., & Calcagni, M. L. (2018) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4226-6 · Cited by 28 · see citing works
Lasnon, C., Quak, E., Roux, P. L., Robin, P., Hofman, M. S., Bourhis, D., Callahan, J., Binns, D., Desmonts, C., Salaün, P., et al. (2017) · EJNMMI Physics · doi:10.1186/s40658-017-0185-4 · Cited by 27 · see citing works
Self-citationDoes PET Reconstruction Method Affect Deauville Score in Lymphoma Patients?
Enilorac, B., Lasnon, C., Nganoa, C., Fruchart, C., Gac, A., Damaj, G. L., & Aide, N. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.202721 · Cited by 27 · see citing works
Self-citationFour-dimensional quantitative analysis using FDG-PET in clinical oncology
Tamaki, N., Hirata, K., Kotani, T., Nakai, Y., Matsushima, S., & Yamada, K. (2023) · Japanese Journal of Radiology · doi:10.1007/s11604-023-01411-4 · Cited by 22 · see citing works
Anatomical Grading for Metabolic Activity of Brown Adipose Tissue
Becker, A. S., Nagel, H. W., Wolfrum, C., & Burger, I. A. (2016) · PLoS ONE · doi:10.1371/journal.pone.0149458 · Cited by 24 · see citing works
Cottereau, A., Buvat, I., Kanoun, S., Versari, A., Casasnovas, O., Chauvie, S., Clerc, J., Gallamini, A., & Meignan, M. A. (2018) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4005-4 · Cited by 24 · see citing works
Optimization of SPECT/CT imaging protocols for quantitative and qualitative 99mTc SPECT
Kupitz, D., Wissel, H., Wuestemann, J., Bluemel, S., Pech, M., Amthauer, H., Kreissl, M. C., & Großer, O. S. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00405-3 · Cited by 26 · see citing works
Prenosil, G. A., Hentschel, M., Weitzel, T., Sari, H., Shi, K., Afshar‐Oromieh, A., & Rominger, A. (2022) · EJNMMI Physics · doi:10.1186/s40658-022-00455-1 · Cited by 23 · see citing works
Pontoizeau, C., Girard, A., Mesbah, H., Haumont, L.-A., Devillers, A., Tempescul, A., Salaün, P., Lamy, T., Jeune, F. L., & Palard-Novello, X. (2019) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000002879 · Cited by 22 · see citing works
Daisaki, H., Kitajima, K., Nakajo, M., Watabe, T., Ito, K., Sakamoto, F., Nakahara, T., Ishibashi, M., & Toriihara, A. (2021) · Scientific Reports · doi:10.1038/s41598-021-87942-0 · Cited by 21 · see citing works
Keller, H., Shek, T. L. T., Driscoll, B., Xu, Y., Nghiem, B., Nehmeh, S. A., Grkovski, M., Schmidtlein, C. R., Budzevich, M. M., Balagurunathan, Y., et al. (2022) · Tomography · doi:10.3390/tomography8020091 · Cited by 21 · see citing works
Miwa, K., Wagatsuma, K., Iimori, T., Sawada, K., Kamiya, T., Sakurai, M., Miyaji, N., Murata, T., & Sato, E. (2018) · Physica Medica · doi:10.1016/j.ejmp.2018.07.001 · Cited by 20 · see citing works
[18F]FDG PET/MRI enables early chemotherapy response prediction in pancreatic ductal adenocarcinoma
Harder, F. N., Jungmann, F., Kaissis, G. A., Lohöfer, F. K., Ziegelmayer, S., Havel, D., Quante, M., Reichert, M., Schmid, R. M., Demir, İ. E., et al. (2021) · EJNMMI Research · doi:10.1186/s13550-021-00808-4 · Cited by 20 · see citing works
Ptáček, J., Karhan, P., & Fiala, P. (2017) · Physica Medica · doi:10.1016/j.ejmp.2017.06.002 · Cited by 18 · see citing works
Validation of phantom-based harmonization for patient harmonization
Panetta, J. V., Daube-Witherspoon, M. E., & Karp, J. S. (2017) · Medical Physics · doi:10.1002/mp.12311 · Cited by 17 · see citing works
Lasnon, C., Houdu, B., Kammerer, E., Salomon, T., Devreese, J., Lebasnier, A., & Aide, N. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-015-3214-3 · Cited by 17 · see citing works
Self-citationTsutsui, Y., Awamoto, S., Himuro, K., Umezu, Y., Baba, S., & Sasaki, M. (2018) · Europe PMC (PubMed Central) · doi:10.22038/aojnmb.2017.26684.1186 · Cited by 15 · see citing works
Monsef, A., Ay, M., Sheikhzadeh, P., Geramifar, P., Rahmim, A., & Ghafarian, P. (2022) · Nuclear medicine communications · doi:10.1097/mnm.0000000000001598 · Cited by 11 · see citing works
Advancing Precision Nuclear Medicine and Molecular Imaging for Lymphoma
Wright, C., Maly, J. J., Zhang, J., & Knopp, M. V. (2016) · PET Clinics · doi:10.1016/j.cpet.2016.08.005 · Cited by 14 · see citing works
Song, Y., Meng, X., Cao, Z., Zhao, W., Zhang, Y., Guo, R., Zhou, X. H., Yang, Z. H., & Li, N. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00540-z · Cited by 13 · see citing works
Kit, N. H., Dugué, A. E., Sevin, E., Allouache, N., Lesaunier, F., Joly, F., & Aide, N. (2015) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000000457 · Cited by 13 · see citing works
Self-citationChen, W., Li, Y., Li, Z., Jiang, Y., Cui, Y., Zeng, J., Mo, Y., Tang, S., Li, S., Liu, L., et al. (2024) · Journal of Nuclear Medicine · doi:10.2967/jnumed.123.266948 · Cited by 11 · see citing works
Andela, S. B., Amthauer, H., Furth, C., Rogasch, J. M. M., Beck, M., Mehrhof, F. B., Ghadjar, P., van den Hoff, J., Klatte, T., Tahbaz, R., et al. (2024) · Radiation Oncology · doi:10.1186/s13014-024-02483-w · Cited by 12 · see citing works
Cysouw, M. C. F., Kramer, G. M., Heijtel, D., Schuit, R. C., Morris, M. J., van den Eertwegh, A. J. M., Voortman, J., Hoekstra, O. S., Oprea‐Lager, D. E., & Boellaard, R. (2019) · EJNMMI Research · doi:10.1186/s13550-019-0531-8 · Cited by 11 · see citing works
Kawashima, K., Kato, K., Tomabechi, M., Matsuo, M., Otsuka, K., Ishida, K., Nakamura, R., & Ehara, S. (2016) · British Journal of Radiology · doi:10.1259/bjr.20150938 · Cited by 11 · see citing works
A Bio-Imaging Signature as a Predictor of Clinical Outcomes in Locally Advanced Pancreatic Cancer
Fiore, M., Taralli, S., Trecca, P., Scolozzi, V., Marinelli, L., Triumbari, E. K. A., Caputo, D. D., Angeletti, S., Ciccozzi, M., Coppola, A., et al. (2020) · Cancers · doi:10.3390/cancers12082016 · Cited by 10 · see citing works
Jiménez-Ortega, E., Ureba, A., Baeza, J. A., Barbeiro, A. R., Balcerzyk, M., Parrado‐Gallego, Á., Wals-Zurita, A., Gómez, F. J. G., & Leal, A. (2019) · PLoS ONE · doi:10.1371/journal.pone.0210549 · Cited by 9 · see citing works
Hicks, R. J., & den Abbeele, A. D. V. (2022) · Cancer Imaging · doi:10.1186/s40644-022-00486-y · Cited by 9 · see citing works
Self-citationMattoli, M. V., Calcagni, M. L., Taralli, S., Indovina, L., Spottiswoode, B. S., & Giordano, A. (2019) · Molecular Imaging and Biology · doi:10.1007/s11307-019-01342-5 · Cited by 9 · see citing works
Taralli, S., Scolozzi, V., Boldrini, L., Lenkowicz, J., Pelliccioni, A., Lorusso, M., Attieh, O. A., Ricciardi, S., Carleo, F., Cardillo, G., et al. (2021) · Frontiers in Medicine · doi:10.3389/fmed.2021.664529 · Cited by 9 · see citing works
On the Cutoff of Baseline Total Metabolic Tumor Volume in High–Tumor-Burden Follicular Lymphoma
Laffon, É., & Marthan, R. (2017) · Journal of Clinical Oncology · doi:10.1200/jco.2016.70.6150 · Cited by 9 · see citing works
Wu, L., Zhang, Z.-S., Jiang, C., Li, L., Sun, X.-J., Bai, M., Liu, M., Xiong, K., Shang, J., Yu, J.-M., et al. (2025) · Advancement of science · doi:10.1002/advs.202413125 · Cited by 13 · see citing works
Machado, M. A. D., Menezes, V. O. M. D. O., Namías, M., Vieira, N. S., Queiroz, C. C., Matheoud, R., Alessio, A. M., & Oliveira, M. L. (2018) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.118.213405 · Cited by 8 · see citing works
Allioux, F., Damaj, G. L., Vilque, J., Nganoa, C., Gac, A., Aide, N., & Lasnon, C. (2021) · Leukemia & lymphoma/Leukemia and lymphoma · doi:10.1080/10428194.2021.1948028 · Cited by 7 · see citing works
Self-citationLasnon, C., Enilorac, B., & Aide, N. (2018) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-3938-y · Cited by 7 · see citing works
Self-citationRadioguided surgery with β decay: A feasibility study in cervical cancer
Collarino, A., Florit, A., Bizzarri, N., Lanni, V., Morganti, S., Summa, M. D., Vizzielli, G., Fanfani, F., Mirabelli, R., Ferrandina, G., et al. (2023) · Physica Medica · doi:10.1016/j.ejmp.2023.102658 · Cited by 6 · see citing works
Using PET for therapy monitoring in oncological clinical trials: challenges ahead
Deroose, C. M., Stroobants, S. G., Liu, Y., Shankar, L. K., & Bourguet, P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3689-1 · Cited by 6 · see citing works
Using EQ·PET to reduce reconstruction-dependent variations in [ 18 F]FDG-PET brain imaging
Vanhoutte, M., Semah, F., Lopes, R., Jaillard, A., Petyt, G., Aziz, A.-L., Lahousse, H., Declerck, J. M., Pasquier, F., Spottiswoode, B. S., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab35b4 · Cited by 5 · see citing works
Jin, S. O., Mérida, I., Stavropoulos, I., Elwes, R. D. C., Lam, T., Guedj, É., Girard, N. J., Costes, N., & Hammers, A. (2023) · EJNMMI Research · doi:10.1186/s13550-023-01023-z · Cited by 5 · see citing works
Iizuka, H., Daisaki, H., Ogawa, M., Yoshida, K., & Kaneta, T. (2019) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001037 · Cited by 5 · see citing works
Interim PET in lymphoma: from Deauville to Peking criteria. On the road, again…
Meignan, M. A., & Cottereau, A. (2017) · Leukemia & lymphoma/Leukemia and lymphoma · doi:10.1080/10428194.2017.1368078 · Cited by 5 · see citing works
Taralli, S., Orlandi, A., Pafundi, P. C., Tempesta, V., Leone, A. D., Pontolillo, L., Scardina, L., Lorusso, M., Paris, I., & Calcagni, M. L. (2025) · La radiologia medica · doi:10.1007/s11547-025-01961-9 · Cited by 4 · see citing works
You, H., Sanli, Y., & Subramaniam, R. M. (2019) · American Journal of Roentgenology · doi:10.2214/ajr.18.21067 · Cited by 4 · see citing works
Impact of scanner harmonization on PET-based treatment response assessment in metastatic melanoma
Weisman, A. J., Bradshaw, T., Namías, M., & Jeraj, R. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abb6bb · Cited by 4 · see citing works
Terbium radionuclides for theranostics
Müller, C., Schibli, R., Bernhardt, P., Köster, U., & van der Meulen, N. P. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822960-6.00076-4 · Cited by 4 · see citing works
Weyts, K., Lequesne, J., Johnson, A., Curcio, H., Parzy, A., Coquan, E., & Lasnon, C. (2024) · EJNMMI Research · doi:10.1186/s13550-024-01128-z · Cited by 4 · see citing works
Taralli, S., Scolozzi, V., Triumbari, E. K. A., Carleo, F., Martino, M., Massimi, A. R. D., Ricciardi, S., Cardillo, G., & Calcagni, M. L. (2020) · The Quarterly Journal of Nuclear Medicine and Molecular Imaging · doi:10.23736/s1824-4785.19.03140-6 · Cited by 4 · see citing works
Samim, M., Prevoo, W., der Veen, B. J. D. W., Kuhlmann, K. F. D., Ruers, T. J. M., van Hillegersberg, R., van den Bosch, M. A. A. J., Verkooijen, H. M., Lam, M. G. E. H., & Stokkel, M. P. M. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3637-0 · Cited by 4 · see citing works
Sipilä, O., Liukkonen, J., Halme, H.-L., Tolvanen, T., Sohlberg, A. O., Hakulinen, M. A., Manninen, A.-L., Tahvanainen, K. U. O., Tunninen, V., Ollikainen, T., et al. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00551-w · Cited by 3 · see citing works
Tanaka, A., Sekine, T., ter Voert, E. T., Zeimpekis, K., Delso, G., Barbosa, F. D. G., Warnock, G., Kumita, S., Veit-Haibach, P., & Huellner, M. (2022) · Frontiers in Medicine · doi:10.3389/fmed.2022.796085 · Cited by 5 · see citing works
Balcerzyk, M., Fernández‐López, R., Parrado‐Gallego, Á., Pachón-Garrudo, V. M., -Royan, J. C., Hevilla, J., Jiménez-Ortega, E., & Leal, A. (2017) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2017.04.035 · Cited by 3 · see citing works
Lowe, G. J., Spottiswoode, B. S., Declerck, J. M., Sullivan, K. R., Sharif, M. S., Wong, W., & Sanghera, B. (2020) · BJR|Open · doi:10.1259/bjro.20190035 · Cited by 3 · see citing works
Gutman, F., Poncin, D., Guedec-Ghelfi, R., Génin, R., David, P.-M., Cuif, A., Popinat, G., Nogueira, V., & Aide, N. (2025) · EJNMMI Reports · doi:10.1186/s41824-025-00253-9 · Cited by 2 · see citing works
Self-citationDo phantom harmonization efforts translate into harmonized patient images?
Panetta, J. V., Scheuermann, J. S., Karp, J. S., & Daube-Witherspoon, M. E. (2015) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2015.7582196 · Cited by 1 · see citing works
Jiménez-Ortega, E., Agüera, R., Ureba, A., Balcerzyk, M., Wals-Zurita, A., Gómez, F. J. G., & Leal, A. (2022) · Tomography · doi:10.3390/tomography8020090 · Cited by 1 · see citing works
Novel biomarkers for melanoma immunotherapy
Wong, N. M. A. (2020) · Minerva Access (University of Melbourne)
Prenosil, G. A., Hentschel, M., Weitzel, T., Sari, H., Shi, K., & Rominger, A. (2021) · Research Square · doi:10.21203/rs.3.rs-864693/v1
Rufini, V., Lorusso, M., Inzani, F. S., Pasciuto, T., Triumbari, E. K. A., Grillo, L. R., Lococo, F., Margaritora, S., Pescarmona, E. O., & Rindi, G. (2022) · Research Square · doi:10.21203/rs.3.rs-1273213/v1
Okazaki, E., Koyama, K., Higashikawa, M., Nishida, N., & Fukuda, H. (2024) · European Archives of Oto-Rhino-Laryngology · doi:10.1007/s00405-024-08905-5
Babacan, G. B., Tamam, M. Ö., Demirel, N., Mansuroğlu, İ., & Bilgi, H. (2025) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000002047
Kupitz, D., Wissel, H., Wuestemann, J., Bluemel, S., Pech, M., Amthauer, H., Kreissl, M. C., & Großer, O. S. (2021) · Figshare · doi:10.6084/m9.figshare.15086222
Kupitz, D., Wissel, H., Wuestemann, J., Bluemel, S., Pech, M., Amthauer, H., Kreissl, M. C., & Großer, O. S. (2021) · Figshare · doi:10.6084/m9.figshare.15086228
Wong, A. N. M., Callahan, J., Keyaerts, M., Neyns, B., Mangana, J., Aberle, S., Herschtal, A., Fullerton, S., Milne, D., Amir, I., et al. (2020) · Figshare · doi:10.6084/m9.figshare.12307817
Self-citationBela Andela, S., Amthauer, H., Furth, C., Rogasch, J. M. M., Beck, M., Mehrhof, F. B., Ghadjar, P., van den Hoff, J., Klatte, T., Tahbaz, R., et al. (2024) · Figshare · doi:10.6084/m9.figshare.26755660
Kupitz, D., Wissel, H., Wuestemann, J., Bluemel, S., Pech, M., Amthauer, H., Kreissl, M. C., & Großer, O. S. (2021) · Figshare · doi:10.6084/m9.figshare.15086225
Jin, S. O., Mérida, I., Stavropoulos, I., Elwes, R. D. C., Lam, T., Guedj, E., Girard, N. J., Costes, N., & Hammers, A. (2024) · Figshare · doi:10.6084/m9.figshare.26640244
Harder, F. N., Jungmann, F., Kaissis, G. A., Lohöfer, F. K., Ziegelmayer, S., Havel, D., Quante, M., Reichert, M., Schmid, R. M., Demir, İ. E., et al. (2021) · Figshare · doi:10.6084/m9.figshare.15071563
- Authors
- Quak, E., Le Roux, P.-Y., Hofman, M. S., Robin, P., Bourhis, D., Callahan, J., Binns, D., Desmonts, C., Salaun, P.-Y., Hicks, R. J., & Aide, N.
- Affiliations
- Centre François Baclesse (FR)
- The University of Melbourne (AU)
- Centre Hospitalier Régional Universitaire de Brest (FR)
- Centre Hospitalier Universitaire de Caen Normandie (FR)
- Inserm (FR)
- Corresponding author
- Aide, Nicolas
- ORCID
- Venue
- European Journal of Nuclear Medicine and Molecular Imaging
- Published
- 2015-07-30
- Volume
- 42
- Issue
- 13
- Pages
- 2072-2082
- ISSN
- 1619-7070 (ISSN-L)
- 1619-7089
- PMC
- PMC4623085
- OpenAlex
- W941202616
- Cited by
- 107
- References
- 33
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 7 · 2023: 6 · 2022: 12 · 2021: 16 · 2020: 8 · 2019: 9 · 2018: 12 · 2017: 16 · 2016: 11 · 2015: 4
- FWCI
- 11.62
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- European Journal of Nuclear Medicine and Molecular Imaging (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- HAL (submittedVersion)
- Minerva Access (University of Melbourne) (submittedVersion, cc-by)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Lung Cancer Diagnosis and Treatment
- Keywords
- point spread function reconstruction, time-of-flight PET, SUVmax, SUVpeak, EANM guidelines, standardized uptake value, multicentre validation
- MeSH terms
- Aged, Carcinoma, Non-Small-Cell Lung, Female, Humans, Image Processing, Computer-Assisted, Liver Neoplasms, Lymphoma, Non-Hodgkin, Male, Melanoma, Middle Aged, Radiopharmaceuticals, Fluorodeoxyglucose F18, Positron-Emission Tomography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review