Noninvasive investigation of blood pressure changes using the pulse wave transit time: a novel approach in the monitoring of hemodialysis patients
Abstract
Severe blood pressure changes are well known in hemodialysis. Detection and prediction of these are important for the well-being of the patient and for optimizing treatment. New noninvasive methods for this purpose are required. The pulse wave transit time technique is an indirect estimation of blood pressure, and our intention is to investigate whether this technique is applicable for hemodialysis treatment. A measurement setup utilizing lower body negative pressure and isometric contraction was used to simulate dialysis-related blood pressure changes in normal test subjects. Systolic blood pressure levels were compared to different pulse wave transit times, including and excluding the cardiac preejection period. Based on the results of these investigations, a pulse wave transit time technique adapted for dialysis treatment was developed and tried out on patients. To determine systolic blood pressure in the normal group, the total pulse wave transit time was found most suitable (including the cardiac preejection period). Correlation coefficients were r = 0.80 +/- 0.06 (mean +/- SD) overall and r = 0.81 +/- 0.16 and r = 0.09 +/- 0.62 for the hypotension and hypertension phases, respectively. When applying the adapted technique in dialysis patients, large blood pressure variations could easily be detected when present. Pulse wave transit time is correlated to systolic blood pressure within the acceptable range for a trend-indicating system. The method's applicability for dialysis treatment requires further studies. The results indicate that large sudden pressure drops, like those seen in sudden hypovolemia, can be detected.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The Moens-Korteweg equation provides a linear relationship between PAT and BP (38 , 39).”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [38]
Citations
Cites
Chen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Lantelme, P., Mestre, C., Lièvre, M. M., Gressard, A., & Milon, H. (2002) · Hypertension · doi:10.1161/01.hyp.0000019132.41066.95 · Cited by 353
The difference in pulse transit time to the toe and finger measured by photoplethysmography
Nitzan, M., Khanokh, B., & Slovik, Y. (2001) · Physiological Measurement · doi:10.1088/0967-3334/23/1/308 · Cited by 218
Physiological effects of locally applied reduced pressure in man.
Wolthuis, R. A., Bergman, S. A., & NICOGOSSIAN, A. E. (1974) · Physiological Reviews · doi:10.1152/physrev.1974.54.3.566 · Cited by 214
Dialysis disequilibrium syndrome: Current Concepts on pathogenesis and prevention
Arieff, A. I. (1994) · Kidney International · doi:10.1038/ki.1994.84 · Cited by 199
Relation between heart rate and pulse transit time during paced respiration
Drinnan, M. J., Allen, J., & Murray, A. (2001) · Physiological Measurement · doi:10.1088/0967-3334/22/3/301 · Cited by 169
陳, 文. (2001) · Medical Entomology and Zoology · Cited by 115
Reducing symptoms during hemodialysis by continuously monitoring the hematocrit
Steuer, R. R., Leypoldt, J. K., Cheung, A. K., Senekjian, H. O., & Conis, J. M. (1996) · American Journal of Kidney Diseases · doi:10.1016/s0272-6386(96)90163-8 · Cited by 113
Pulse Transit Time and Blood Pressure: An Intensive Analysis
Lane, J. D., Greenstadt, L., Shapiro, D., & Rubinstein, E. H. (1983) · Psychophysiology · doi:10.1111/j.1469-8986.1983.tb00899.x · Cited by 107
Pollak, M. H., & Obrist, P. A. (1983) · Psychophysiology · doi:10.1111/j.1469-8986.1983.tb00895.x · Cited by 94
Flagged first, then by citations
Research works printed but not linked: 2
- D Franchi, 1996, Comput Cardiol, 23, 108
- WG Chochran, 1989, Statistical methods
Cited by
Sorted by citations
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“Similar foot location definitions have been applied to PPG waveforms, including the 1) diastolic minimum time [142]–[144], 2) intersecting tangent method [98], 3) times of maximum derivatives with respect to time [145]– [148], and 4) time to reach a fraction of the pulse height [143], [147].”
MethodBinodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
“The Moens-Korteweg equation provides a linear relationship between PAT and BP (38, 39).”
MethodHeart rate measurement based on a time-lapse image.
Takano, C., & Ohta, Y. (2007) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2006.09.006 · Cited by 345
Novel Wireless Devices for Cardiac Monitoring
Walsh, J. A., Topol, E. J., & Steinhubl, S. R. (2014) · Circulation · doi:10.1161/circulationaha.114.009024 · Cited by 175
Kim, C., Carek, A. M., Mukkamala, R., Inan, O. T., & Hahn, J. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2440291 · Cited by 155
Non-invasive continuous blood pressure monitoring: a review of current applications
Chung, E., Chen, G., Alexander, B. S., & Cannesson, M. P. (2013) · Frontiers of Medicine · doi:10.1007/s11684-013-0239-5 · Cited by 150
A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Buxi, D., Redouté, J., & Yuce, M. R. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/r1 · Cited by 140
Thomas, S. S., Nathan, V., Zong, C., Soundarapandian, K., Shi, X., & Jafari, R. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2015.2458779 · Cited by 140
Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
Zhang, G., Gao, M., Xu, D. Z., Olivier, N. B., & Mukkamala, R. (2011) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00980.2011 · Cited by 133
Ding, X., & Zhang, Y. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00096-x · Cited by 142
Noninvasive cuffless blood pressure estimation using pulse transit time and Hilbert–Huang transform
Choi, Y., Zhang, Q., & Ko, S. (2012) · Computers & Electrical Engineering · doi:10.1016/j.compeleceng.2012.09.005 · Cited by 123
Blood Pressure Estimation Using Pulse Transit Time From Bioimpedance and Continuous Wave Radar
Buxi, D., Jean-Michel, R., & Yuce, M. R. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2582472 · Cited by 123
Wang, L., Zhou, W., Xing, Y., & Zhou, X. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/7804243 · Cited by 109
Kim, C., Carek, A. M., Inan, O. T., Mukkamala, R., & Hahn, J. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2797239 · Cited by 110
Self-powered ultrasensitive pulse sensors for noninvasive multi-indicators cardiovascular monitoring
Xu, L., Zhang, Z., Gao, F., Zhao, X., Xun, X., Kang, Z., Liao, Q., & Zhang, Y. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.105614 · Cited by 108
Zhuo, B., Chen, S., Zhao, M., & Guo, X. (2017) · IEEE Journal of the Electron Devices Society · doi:10.1109/jeds.2017.2683558 · Cited by 97
Physiological Acoustic Sensing Based on Accelerometers: A Survey for Mobile Healthcare
Hu, Y., Kim, E. G., Cao, G., Liu, S., & Xu, Y. P. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1111-8 · Cited by 86
Pulse transit time as an indirect marker for variations in cardiovascular related reactivity
Foo, J. Y. A., & Lim, C. S. (2006) · Technology and Health Care · doi:10.3233/thc-2006-14205 · Cited by 83
Pulse Transit Time and Blood Pressure During Cardiopulmonary Exercise Tests
Wibmer, T., DOERING, K., Kropf-Sanchen, C., RÜDIGER, S., Blanta, I., Stoiber, K., Rottbauer, W., & SCHUMANN, C. (2014) · Physiological Research · doi:10.33549/physiolres.932581 · Cited by 81
Zhou, Z.-B., Cui, T., Li, D., Jian, J., Li, Z., Ji, S., Li, X., Xu, J., Liu, H., Yang, Y., et al. (2023) · Materials · doi:10.3390/ma16062133 · Cited by 85
McCarthy, B. M., Vaughan, C. J., O’Flynn, B., Mathewson, A., & Mathúna, C. Ó. (2013) · Journal of Human Hypertension · doi:10.1038/jhh.2013.41 · Cited by 72
Continuous Blood Pressure Monitoring using ECG and Finger Photoplethysmogram
Chua, C. P., & Heneghan, C. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.259612 · Cited by 65
Evaluation of blood pressure changes using vascular transit time
Foo, J. Y. A., Lim, C. S., & Wang, P. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/8/003 · Cited by 62
Chan, G. S. H., Middleton, P. M., Celler, B. G., Wang, L., & Lovell, N. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9086-8 · Cited by 62
Continuous blood pressure measurement using pulse transit time
Hennig, A., & Patzak, A. R. (2013) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-013-0617-x · Cited by 58
A Cuffless Blood Pressure Measurement Based on the Impedance Plethysmography Technique
Liu, S.-H., Cheng, D., & Su, C.-H. (2017) · Sensors · doi:10.3390/s17051176 · Cited by 55
Thomas, S. S., Nathan, V., Zong, C., Akinbola, E., Aroul, A. L. P., Philipose, L., Soundarapandian, K., Shi, X., & Jafari, R. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6944076 · Cited by 52
Effect of confounding factors on blood pressure estimation using pulse arrival time
Kim, J. S., Kim, K. K., Baek, H. J., & Park, K. S. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/5/007 · Cited by 44
Pant, S., Umesh, S., & Asokan, S. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2973342 · Cited by 42
Noninvasive Techniques for Prevention of Intradialytic Hypotension
Sörnmo, L., Sandberg, F., Gil, E., & Solem, K. (2012) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2012.2210036 · Cited by 41
Pulse wave transit time for monitoring respiration rate
Johansson, A., Ahlstrom, C. H. G., Länne, T., & Ask, P. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0064-y · Cited by 38
Self-citationA method for accurate localization of the first heart sound and possible applications
Ahlstrom, C. H. G., Länne, T., Ask, P., & Johansson, A. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/3/011 · Cited by 28
Self-citationNonlinear phonocardiographic Signal Processing
Ahlstrom, C. H. G. (2008) · KTH Publication Database DiVA (KTH Royal Institute of Technology) · Cited by 28
Self-citationPrediction of Intradialytic Hypotension Using Photoplethysmography
Solem, K., Olde, B., & Sörnmo, L. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2042170 · Cited by 26
Marcinkevičs, Z., Grēve, M., Aivars, J. I., Erts, R., & Zehtabi, A. H. (2009) · Cited by 24
Blood Pressure Measurement Using Finger ECG and Photoplethysmogram for IoT
Dinh, A., Luu, L., & Cao, T. (2017) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-4361-1_14 · Cited by 24
Cuffless and Continuous Blood Pressure Monitoring Using a Single Chest-Worn Device
Park, J., Yang, S.-M., Sohn, J., Lee, J., Lee, S., Ku, Y., & Kim, H. C. (2019) · IEEE Access · doi:10.1109/access.2019.2942184 · Cited by 25
Haegele-Link, S., Claus, D., Ducker, S., Vogt, T., & Birklein, F. (2008) · The Open Neurology Journal · doi:10.2174/1874205x00802010012 · Cited by 20
Pulse transit time-based blood pressure estimation using hilbert-huang transform
Zhang, Q., Shi, Y., Teng, D., Dinh, A., Ko, S., Chen, L., Basran, J., Bello-Haas, V. P. M. D., & Choi, Y. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5334008 · Cited by 18
Pulse transit times to the capillary bed evaluated by laser Doppler flowmetry
Bernjak, A., & Stefanovska, A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/3/002 · Cited by 17
Ghosh, S. K., Chattopadhyay, B. P., Roy, R. M., Mukherjee, J., & Mahadevappa, M. (2019) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2019.02.002 · Cited by 17
Clinical applications of pulse transit time in paediatric critical care
Foo, J. Y. A., & Wilson, S. J. (2009) · Journal of Medical Engineering & Technology · doi:10.1080/03091900701860210 · Cited by 17
Zhang, G., McCombie, S., Greenstein, R., & McCombie, D. B. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6943656 · Cited by 16
ReNN: Rule-embedded Neural Networks
Wang, H. (2018) · Proceedings - International Conference on Pattern Recognition/Proceedings/International Conference on Pattern Recognition · doi:10.1109/icpr.2018.8545379 · Cited by 17
High Fidelity MZI-BCG Sensor With Homodyne Demodulation for Unobtrusive HR and BP Monitoring
Yang, F., Xu, W. L., Lyu, W., Tan, F., Yu, C., & Dong, B. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3158070 · Cited by 16
Liu, S.-H., Liu, L.-J., Pan, K., Chen, W., & Tan, T.-H. (2019) · Applied Sciences · doi:10.3390/app9142922 · Cited by 15
Parameters for mean blood pressure estimation based on electrocardiography and photoplethysmography
Špulák, D., Čmejla, R., & Fabián, V. (2011) · International Conference on Applied Electronics · Cited by 14
Non-invasive cuffless blood pressure and heart rate monitoring using impedance cardiography
Ghosh, S. K., Chattopadhyay, B. P., Roy, R. M., Mukherjee, J., & Mahadevappa, M. (2021) · Intelligent Medicine · doi:10.1016/j.imed.2021.11.001 · Cited by 14
Min, S. D., & Shin, H. (2015) · Journal of Electrical Engineering and Technology · doi:10.5370/jeet.2015.10.5.2126 · Cited by 14
Park, Y., Kim, S., Lee, Y. S., Choi, S., Ku, S., & Hwang, G. (2020) · Journal of Medical Internet Research · doi:10.2196/13156 · Cited by 13
Liu, S.-H., Zhang, B.-H., Chen, W., Su, C.-H., & Chin, C.-L. (2022) · Nutrients · doi:10.3390/nu14122552 · Cited by 13
Liu, S., Lai, S.-H., Wang, J.-J., Tan, T.-H., & Huang, Y. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8856749 · Cited by 12
Cuff-free blood pressure estimation using signal processing techniques
Zhang, Q. (2010) · University Library - University of Saskatchewan (University of Saskatchewan) · Cited by 12
Photoplethysmogram signal analysis for detecting vital physiological parameters: An evaluating study
Reguig, F. B. (2016) · International Symposium on Signal, Image, Video and Communications (ISIVC) · doi:10.1109/isivc.2016.7893981 · Cited by 13
Xu, Y., Ping, P., Wang, D., & Zhang, W. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/5396030 · Cited by 11
Peng, R., Li, Y., & Yan, W. (2021) · Scientific Reports · doi:10.1038/s41598-021-90056-2 · Cited by 12
Journal of Artificial Organs 2005: the year in review
Sawa, Y., Horiuchi, T., Kishida, A., Masuzawa, T., Mizuguchi, K., Nishimura, M., Okoshi, T., Shinzato, T., Tatsumi, E., Tomizawa, Y., et al. (2006) · Journal of Artificial Organs · doi:10.1007/s10047-006-0333-4 · Cited by 11
Pulse transit time differential measurement by fiber Bragg grating pulse recorder
Umesh, S., Padma, S., Ambastha, S., Kalegowda, A., & Asokan, S. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.5.057005 · Cited by 10
Calculation and Validation of Continuous Pulse Transit Time Based on Normalized Pulse Wave Velocity
Shin, H., & Choi, A. (2020) · IEEE Access · doi:10.1109/access.2020.3041498 · Cited by 8
Njoum, H., & Kyriacou, P. A. (2016) · Physiological Measurement · doi:10.1088/1361-6579/38/2/87 · Cited by 8
Mamun, M. M. R. K., & Alouani, A. T. (2019) · Conference proceedings - Canadian Conference on Electrical and Computer Engineering · doi:10.1109/ccece.2019.8861521 · Cited by 8
Demonstration abstract: BioWatch — A wrist watch based physiological signal acquisition system
Thomas, S. S., Nathan, V., Zong, C., Aroul, A. L. P., Philipose, L., Soundarapandian, K., Shi, X., & Jafari, R. (2014) · ACM/IEEE International Conference on Information Processing in Sensor Networks (IPSN) · doi:10.1109/ipsn.2014.6846797 · Cited by 8
Factors that Affect Pulse Wave Time Transmission in the Monitoring of Cardiovascular System
Foo, J. Y. A., Wilson, S. J., & Wang, P. (2008) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-008-9115-2 · Cited by 7
Watanabe, K., Kurihara, Y., Watanabe, K., Azami, T., Nukaya, S., & Tanaka, H. (2014) · IEEE Transactions on Human-Machine Systems · doi:10.1109/thms.2014.2320945 · Cited by 7
Douniama, C., Sauter, C. U., & Couronné, R. (2009) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-03882-2_571 · Cited by 5
Photoplethysmography: Towards a non-invasive pressure measurement technique
Njoum, H., & Kyriacou, P. A. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7590776 · Cited by 5
Djupedal, H., Nøstdahl, T., Hisdal, J., Landsverk, S. A., & Høiseth, L. O. (2022) · Physiological Reports · doi:10.14814/phy2.15355 · Cited by 4
Nordine, M., Pille, M., Kraemer, J. F., Berger, C., Brandhorst, P., Kaeferstein, P., Kopetsch, R., Wessel, N., Trauzeddel, R. F., & Treskatsch, S. (2023) · Sensors · doi:10.3390/s23063304 · Cited by 6
How much influence does inspiration have on pulse transit time in sleep apnea?
Yin, M., Nakayama, M., Miyazaki, S., & Ishikawa, K. (2008) · Otolaryngology · doi:10.1016/j.otohns.2007.12.031 · Cited by 5
Moghimi, S., Chau, T. T., & Guerguerian, A. (2013) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2013.6696175 · Cited by 5
Pulse Signal Acquisition Based on LabView Software
Chen, X. (2008) · International Conference on Future BioMedical Information Engineering · doi:10.1109/fbie.2008.104 · Cited by 5
Ding, X. (2019) · Springer eBooks · doi:10.1007/978-3-030-24701-0_9 · Cited by 5
Wunderle, M., Bachler, M., Stadler, D., Hametner, B., Orter, S., Küchle, C., Krenn, S., Heemann, U., Wassertheurer, S., Schmaderer, C., et al. (2025) · Clinical Kidney Journal · doi:10.1093/ckj/sfaf255 · Cited by 4
A Computational Approach to Predict Pulse Transit Time Variations During Postural Change
Foo, J. Y. A. (2007) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-007-9034-8 · Cited by 4
Combination of Denoising Algorithms for Video-Based Non-contact Heart Rate Measurement
Xiao, H., Xu, J., Hu, D., & Wang, J. (2022) · Information Communication Technologies Conference (ICTC) · doi:10.1109/ictc55111.2022.9778685 · Cited by 4
Demonstration abstract: BioWatch: a wrist watch based physiological signal acquisition system
Thomas, S. S., Nathan, V., Zong, C., Aroul, A. L. P., Philipose, L., Soundarapandian, K., Shi, X., & Jafari, R. (2014) · doi:10.5555/2602339.2602409 · Cited by 4
Using impedance-plethysmography technique for cuffless blood pressure measurement
Liu, S.-H., Cai, G.-H., Cheng, D., & Huang, Y. (2017) · Conference proceedings/Conference proceedings - IEEE International Conference on Systems, Man, and Cybernetics · doi:10.1109/smc.2017.8122636 · Cited by 3
Using feature optimization and fuzzy logic to detect hypertensive heart diseases
Mamun, M. M. R. K., & Alouani, A. T. (2020) · Proceedings of the World Congress on Electrical Engineering and Computer Systems and Science · doi:10.11159/icbes20.128 · Cited by 3
Smart Watches for Physiological Monitoring: A Case Study on Blood Pressure Measurement
Nathan, V., Thomas, S. S., & Jafari, R. (2016) · Springer eBooks · doi:10.1007/978-3-319-45142-8_14 · Cited by 2
Variations of Heart Rate, Pulse Arrival Time and Blood Pressure in a Versatile Laboratory Protocol
Ahmaniemi, T., Rajala, S., Lindholm, H. J., Taipalus, T., & Müller, K. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512876 · Cited by 2
A NEW LOW-COMPLEXITY ALGORITHM FOR THE PULSE TRANSIT TIME EVALUATION
Lilia, R., & Omari, T. (2023) · Bioscience & Engineering An International Journal · doi:10.5121/bioej.2023.10202 · Cited by 1
Nonlinear relationship between systolic blood pressure and pulse transit time in anesthetized dogs
Tang, H. T., Sun, J. H., & Park, Y. (2014) · International Conference on Biomedical Engineering and Informatics (BMEI) · doi:10.1109/bmei.2014.7002800 · Cited by 1
A new approach for detecting sudden hypotension in hemodialysis by using dual-channel optical system
Heravi, M. A. Y., Golmakani, E., & Joharinia, S. (2016) · Journal of Innovative Optical Health Sciences · doi:10.1142/s1793545816500334 · Cited by 1
Grigonytė, E., Gil, E., Laguna, P., & Sörnmo, L. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.09.020 · Cited by 1
Measurement of Blood Pressure in Conscious Cats and Dogs
Jepson, R. E. (2020) · Springer eBooks · doi:10.1007/978-3-030-33020-0_2 · Cited by 1
Comparison of Heart Rate Variability with Pulse Transit Time during General Anesthesia
Baik, S.-W., Kim, T. K., Kim, J. H., Jeon, G. R., & Ye, S. Y. (2008) · Journal of the Korean Institute of Electrical and Electronic Material Engineers · doi:10.4313/jkem.2008.21.8.770 · Cited by 1
Castro, A., Carvalho, P., Muehlsteff, J., Mattos, S. D. S., & Coimbra, M. T. (2016) · Advances in bioinformatics and biomedical engineering book series · doi:10.4018/978-1-5225-0660-7.ch002 · Cited by 1
Ghosh, S. K., Chattopadhyay, B. P., Roy, R. M., Mukhopadhyay, J., & Mahadevappa, M. (2020) · IEEE International Conference on Bioinformatics and Bioengineering (BIBE) · doi:10.1109/bibe50027.2020.00113 · Cited by 1
Kern, F., & Bernhard, S. (2019) · medRxiv · doi:10.1101/2019.12.19.19015321 · Cited by 1
Commercialization of contact-free blood pressure monitoring technology
Radhakrishna, S. (2010) · OhioLink ETD Center (Ohio Library and Information Network)
Relationship between HsCRP and Pulse Transit Time
Kim, Y.-J., Min, H.-G., Kim, Y., Jeon, A.-Y., Jeon, G.-R., & Ye, S.-Y. (2007) · Journal of Life Science · doi:10.5352/jls.2007.17.2.218
Aguilar–Pelaez, E., & Rodriguez–Villegas, E. (2008) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2008.4541615
A Novel Portable Pulse Signal Recorder
Chen, X. (2008) · International Conference on Future BioMedical Information Engineering · doi:10.1109/fbie.2008.105
Detecting Emotional Response to Music using Near-infrared Spectroscopy of the Prefrontal Cortex
Moghimi, S. (2014) · TSpace (University of Toronto)
PPG Signal-Based Blood Pressure Estimation Using Modified Hilbert-Huang Transform
赵, 晓. (2018) · Journal of Sensor Technology and Application · doi:10.12677/jsta.2018.62008
Comparison of the Correlation of Different Pulse Transit Time Parameters to Blood Pressure
Lin, W.-H., Samuel, O. W., Liu, Q., Zhang, Y., & Li, G. (2018) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-4505-9_36
A Study on the Wearable and Cuffless Continuous Blood Pressure Monitoring System
박종현 (2019) · Seoul National University Open Repository (Seoul National University)
Detection and localization of Coronary Arterial Lesion with the Aid of Impedance CardiographyPreprint
Ghosh, S. K., Chattopadhyay, B. P., Roy, R. M., Mukherjee, J., & Mahadevappa, M. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2003.12067
Castro, A., Carvalho, P., Muehlsteff, J., Mattos, S. D. S., & Coimbra, M. T. (2017) · IGI Global eBooks · doi:10.4018/978-1-5225-3158-6.ch045
Metadata
- Authors
- Ahlstrom, C., Johansson, A., Uhlin, F., Länne, T., & Ask, P.
- Affiliations
- Linköping University (SE)
- Linköping University Hospital (SE)
- Corresponding author
- Ahlstrom, Christer
- Venue
- Journal of Artificial Organs
- Published
- 2005-10-20
- Volume
- 8
- Issue
- 3
- Pages
- 192-197
- ISSN
- 1434-7229 (ISSN-L)
- 1619-0904
- OpenAlex
- W2047693489
- Cited by
- 104
- References
- 20
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 4 · 2022: 4 · 2021: 2 · 2020: 8 · 2019: 10 · 2018: 7 · 2017: 5 · 2016: 8 · 2015: 6 · 2014: 10 · 2013: 4 · 2012: 2
- FWCI
- 4.68
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Journal of Artificial Organs (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- pulse transit time, blood pressure monitoring, hemodialysis, non-invasive blood pressure estimation, systolic blood pressure, hypotension, hypertension, hypovolemia, lower body negative pressure, isometric contraction
- MeSH terms
- Adult, Aged, Blood Pressure, Blood Pressure Determination, Female, Renal Dialysis, Humans, Male, Monitoring, Physiologic, Pulse
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review