Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Open accessAbstract
Ubiquitous blood pressure (BP) monitoring is needed to improve hypertension detection and control and is becoming feasible due to recent technological advances such as in wearable sensing. Pulse transit time (PTT) represents a well-known potential approach for ubiquitous BP monitoring. The goal of this review is to facilitate the achievement of reliable ubiquitous BP monitoring via PTT. We explain the conventional BP measurement methods and their limitations; present models to summarize the theory of the PTT-BP relationship; outline the approach while pinpointing the key challenges; overview the previous work toward putting the theory to practice; make suggestions for best practice and future research; and discuss realistic expectations for the approach.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Models that require a single calibration cuff measurement are convenient but potentially less accurate than those that use multiple measurements to form a patient-specific calibration curve.
“Models that require a single calibration cuff measurement are convenient but potentially less accurate than those that use multiple measurements to form a patient-specific calibration curve”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [20]
Pulse wave propagation delay between the genders indicates baseline differences in arterial stiffness with women displaying lower degrees of pulse arrival latency delay indicating higher vascular stiffness.
“Pulse wave propagation delay between the genders indicates baseline differences in arterial stiffness with women displaying lower degrees of pulse arrival latency delay indicating higher vascular stiffness [106,107].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [106]
New devices that promise to change the current paradigm of hypertension monitoring use either inflating wristband cuffs or optical pulse wave-based metrics of BP.
“New devices that promise to change the current paradigm of hypertension monitoring use either inflating wristband cuffs (e.g., Omron HeartGuide, currently undergoing clinical testing) or optical pulse wave-based metrics of BP [75,76].”
From Wearables and the Medical Revolution (2018) · cites this paper as [76]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · see citing works
Expert consensus document on arterial stiffness: methodological issues and clinical applications
Laurent, S., Cockcroft, J., Van Bortel, L., Boutouyrie, P., Giannattasio, C., Hayoz, D., Pannier, B., Vlachopoulos, C., Wilkinson, I., Struijker-Boudier, H., et al. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl254 · Cited by 6,077 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Burt, V. L., Whelton, P., Roccella, E. J., Brown, C., Cutler, J. A., Higgins, M., Horan, M. J., & Labarthe, D. (1995) · Hypertension · doi:10.1161/01.hyp.25.3.305 · Cited by 2,998 · see citing works
Anderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456 · see citing works
Lakatta, E. G., & Levy, D. (2003) · Circulation · doi:10.1161/01.cir.0000048892.83521.58 · Cited by 2,244 · see citing works
The Reference Values for Arterial Stiffness' Collaboration (2010) · European Heart Journal · doi:10.1093/eurheartj/ehq165 · Cited by 2,085 · see citing works
Schwartz, G., Tee, B. C.-K., Mei, J., Appleton, A. L., Kim, D. H., Wang, H., & Bao, Z. (2013) · Nature Communications · doi:10.1038/ncomms2832 · Cited by 2,042 · cited 1× · see citing works
“Flexible strain or pressure sensors placed over a superficial artery can measure waveforms indicative of BP via the tonometric principle [109]–[112].”
BackgroundHuman blood pressure determination by sphygmomanometry.
Perloff, D., Grim, C. M., Flack, J. M., Frohlich, E. D., Hill, M. D., McDonald, M., & Morgenstern, B. Z. (1993) · Circulation · doi:10.1161/01.cir.88.5.2460 · Cited by 1,592 · see citing works
Advancements in Noncontact, Multiparameter Physiological Measurements Using a Webcam
Poh, M., McDuff, D., & Picard, R. W. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2086456 · Cited by 1,488 · see citing works
Asmar, R., Benetos, A., Topouchian, J., Laurent, P., Pannier, B., Brisac, A.-M., Target, R., & Levy, B. I. (1995) · Hypertension · doi:10.1161/01.hyp.26.3.485 · Cited by 1,376 · see citing works
Applying the right statistics: analyses of measurement studies
Bland, J. M., & Altman, D. G. (2003) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.122 · Cited by 1,305 · see citing works
Mechanical Factors in Arterial Aging
O’Rourke, M. F., & Hashimoto, J. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.12.050 · Cited by 1,245 · see citing works
Methodological Guidelines for Impedance Cardiography
Sherwood, A., Allen, M. T., Fahrenberg, J., Kelsey, R. M., Lovallo, W. R., & van Doornen, L. J. P. (1990) · Psychophysiology · doi:10.1111/j.1469-8986.1990.tb02171.x · Cited by 1,169 · see citing works
Remoissenet, M. (1994) · Springer eBooks · doi:10.1007/978-3-662-03057-8 · Cited by 924 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · cited 4× · see citing works
“The advantage of reflectance-mode PPG is that it is less restrictive in terms of measurement locations [66], [67], [69].”
BackgroundMethodInfluentialThe dynamic elastic properties of the arterial wall
Bergel, D. H. (1961) · The Journal of Physiology · doi:10.1113/jphysiol.1961.sp006687 · Cited by 907 · see citing works
Parati, G., Casadei, R., Groppelli, A., Di Rienzo, M., & Mancia, G. (1989) · Hypertension · doi:10.1161/01.hyp.13.6.647 · Cited by 899 · see citing works
Applications of adaptive filtering to ECG analysis: noise cancellation and arrhythmia detection
Thakor, N. V., & Zhu, Y.-S. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83591 · Cited by 886 · see citing works
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology
Imholz, B. P. M. (1998) · Cardiovascular Research · doi:10.1016/s0008-6363(98)00067-4 · Cited by 837 · see citing works
McDONALD'S BLOOD FLOW IN ARTERIES: THEORETICAL, EXPERIMENTAL AND CLINICAL PRINCIPLES
Czuwala, P. J. (1991) · Echocardiography · doi:10.1111/j.1540-8175.1991.tb01207.x · Cited by 829 · see citing works
Relation of Structure to Function of the Tissues of the Wall of Blood Vessels
Burton, A. C. (1954) · Physiological Reviews · doi:10.1152/physrev.1954.34.4.619 · Cited by 800 · see citing works
Ambulatory Blood-Pressure Monitoring
Pickering, T. G., Shimbo, D., & Haas, D. C. (2006) · New England Journal of Medicine · doi:10.1056/nejmra060433 · Cited by 786 · see citing works
Ballistocardiography and Seismocardiography: A Review of Recent Advances
Inan, O. T., Migeotte, P.-F., Park, K.-S., Etemadi, M., Tavakolian, K., Casanella, R., Zanetti, J., Tank, J., Funtova, I., Prisk, G. K., et al. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2361732 · Cited by 775 · see citing works
Unobtrusive Sensing and Wearable Devices for Health Informatics
Zheng, Y.-L., Ding, X., Poon, C. C. Y., Lo, B. P. L., Zhang, H., Zhou, X.-L., Yang, G., Zhao, N., & Zhang, Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2309951 · Cited by 725 · see citing works
Predominance of Isolated Systolic Hypertension Among Middle-Aged and Elderly US Hypertensives
Franklin, S. S., Jacobs, M. J., Wong, N. D., L'Italien, G. J., & Lapuerta, P. (2001) · Hypertension · doi:10.1161/01.hyp.37.3.869 · Cited by 705 · see citing works
Bland, M., & Altman, D. G. (1990) · Computers in Biology and Medicine · doi:10.1016/0010-4825(90)90013-f · Cited by 689 · see citing works
Alterations with Age in the Viscoelastic Properties of Human Arterial Walls
Learoyd, B. M., & Taylor, M. G. (1966) · Circulation Research · doi:10.1161/01.res.18.3.278 · Cited by 684 · see citing works
Mechanical Properties of Arteries in Vivo
PETERSON, L. H., JENSEN, R. E., & PARNELL, J. (1960) · Circulation Research · doi:10.1161/01.res.8.3.622 · Cited by 679 · cited 1× · see citing works
“The composition of the arterial wall and the function of its components are well understood [19]–[23].”
BackgroundLangewouters, G. J., Wesseling, K. H., & Goedhard, W. J. A. (1984) · Journal of Biomechanics · doi:10.1016/0021-9290(84)90034-4 · Cited by 544 · see citing works
Brien, E. O., Petrie, J. C., Little, W. C., de Swiet, M., Padfield, P. L., Altma, D. G., Bland, M., Coats, A. J. S., & Atkins, N. (1993) · Journal of Hypertension · doi:10.1097/00004872-199306000-00013 · Cited by 518 · see citing works
Hines, E. A., & Brown, G. E. (1936) · American Heart Journal · doi:10.1016/s0002-8703(36)90370-8 · Cited by 506 · see citing works
Electrical Impedance Plethysmography
Nyboer, J., Mm, K., & Hannapel, L. (1950) · Circulation · doi:10.1161/01.cir.2.6.811 · Cited by 493 · see citing works
Gesche, H., Grosskurth, D., Küchler, G., & Patzak, A. R. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1983-3 · Cited by 472 · see citing works
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466 · see citing works
Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart Rate
Shelley, K. H. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269512.82836.c9 · Cited by 450 · see citing works
Textbook of Medical Physiology
Mehta, V. V. M. A. A. (2011) · Cited by 442 · see citing works
Agarwal, R., Bills, J. E., Hecht, T. J. W., & Light, R. P. (2011) · Hypertension · doi:10.1161/hypertensionaha.110.160911 · Cited by 430 · see citing works
Chen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425 · see citing works
Cardiovascular and neuroendocrine adjustment to public speaking and mental arithmetic stressors
Al’Absi, M., Bongard, S., Buchanan, T. W., Pincomb, G. A., Licinio, J., & Lovallo, W. R. (1997) · Psychophysiology · doi:10.1111/j.1469-8986.1997.tb02397.x · Cited by 423 · see citing works
Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure
Payne, R. A., Symeonides, C. N., Webb, D. J., & Maxwell, S. R. J. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00657.2005 · Cited by 418 · see citing works
Contact-Free Measurement of Cardiac Pulse Based on the Analysis of Thermal Imagery
Garbey, M., Sun, N., Merla, A., & Pavlidis, I. T. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.891930 · Cited by 402 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Starr, I., Rawson, A. J., Schroeder, H. A., & Joseph, N. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.1.1 · Cited by 398 · see citing works
Kubicek, W. G., PATTERSON, R. P., & Witsoe, D. A. (1970) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1970.tb17735.x · Cited by 396 · see citing works
Joseph, C. N., Porta, C., Casucci, G., Casiraghi, N., Maffeis, M., Rossi, M., & Bernardi, L. (2005) · Hypertension · doi:10.1161/01.hyp.0000179581.68566.7d · Cited by 394 · see citing works
Invited Review: Aging and the cardiovascular system
Ferrari, A. U., Radaelli, A., & Centola, M. (2003) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00601.2003 · Cited by 381 · see citing works
Cuff-less and Noninvasive Measurements of Arterial Blood Pressure by Pulse Transit Time
Poon, C. C. Y., & Zhang, Y. (2005) · doi:10.1109/iembs.2005.1615827 · Cited by 376 · see citing works
Utility of the Photoplethysmogram in Circulatory Monitoring
Reisner, A. T., Shaltis, P., McCombie, D. B., Asada, H. H., Warner, D. S., & Warner, M. A. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31816c89e1 · Cited by 365 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
An open-source algorithm to detect onset of arterial blood pressure pulses
Zong, W., Heldt, T., Moody, G. B., & Mark, R. G. (2003) · Computers in cardiology · doi:10.1109/cic.2003.1291140 · Cited by 326 · see citing works
Influence of vascular smooth muscle on contractile mechanics and elasticity of arteries
Dobrin, P., & Rovick, A. (1969) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1969.217.6.1644 · Cited by 322 · see citing works
Determination of pulse wave velocities with computerized algorithms
Chiu, Y. C., Arand, P. W., Shroff, S. G., Feldman, T. E., & Carroll, J. D. (1991) · American Heart Journal · doi:10.1016/0002-8703(91)90153-9 · Cited by 313 · see citing works
Keren, H., Burkhoff, D., & Squara, P. (2007) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00195.2007 · Cited by 312 · see citing works
STUDIES ON THE ELASTIC PROPERTIES OF HUMAN ISOLATED AORTA
Hallock, P., & Benson, I. C. (1937) · Journal of Clinical Investigation · doi:10.1172/jci100886 · Cited by 299 · see citing works
Arterial Wall Mechanics in Conscious Dogs
Armentano, R. L., Barra, J. G., Levenson, J. A., Simon, A. C., & Pichel, R. H. (1995) · Circulation Research · doi:10.1161/01.res.76.3.468 · Cited by 297 · see citing works
Pulse Transit Time as an Indicator of Arterial Blood Pressure
Geddes, L. A., Voelz, M. H., Babbs, C. F., Bourland, J. D., & Tacker, W. A. (1981) · Psychophysiology · doi:10.1111/j.1469-8986.1981.tb01545.x · Cited by 293 · see citing works
Silke, B., & McAuley, D. F. (1998) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1000600 · Cited by 289 · see citing works
Harris, W. S., Schoenfeld, C. D., & Weissler, A. M. (1967) · Journal of Clinical Investigation · doi:10.1172/jci105661 · Cited by 268 · see citing works
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263 · see citing works
Theory of the oscillometric maximum and the systolic and diastolic detection ratios
Drzewiecki, G. M., Hood, R. R., & Apple, H. (1994) · Annals of Biomedical Engineering · doi:10.1007/bf02368225 · Cited by 252 · see citing works
Bank, A. J., Wang, H., Holte, J. E., Mullen, K. A., Shammas, R. A., & Kubo, S. H. (1996) · Circulation · doi:10.1161/01.cir.94.12.3263 · Cited by 249 · see citing works
Bevegård, B. S., & Shepherd, J. T. (1966) · Journal of Clinical Investigation · doi:10.1172/jci105317 · Cited by 237 · see citing works
Robust ballistocardiogram acquisition for home monitoring
Inan, O. T., Etemadi, M., Wiard, R. M., Giovangrandi, L., & Kovacs, G. T. A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/2/005 · Cited by 234 · see citing works
Physiocal, calibrating finger vascular physiology for finapres
Wesseling, K. H., Wit, B. D., van der Hoeven, G. M. A., Goudoever, J. V., & Settels, J. J. (1995) · TU/e Research Portal · Cited by 229 · see citing works
Pulse Wave Velocity as a Measure of Blood Pressure Change
Gribbin, B., Steptoe, A., & Sleight, P. (1976) · Psychophysiology · doi:10.1111/j.1469-8986.1976.tb03344.x · Cited by 227 · see citing works
Autonomic Mechanisms in Hemodynamic Responses to Isometric Exercise
Martin, C. E., Shaver, J. A., Leon, D. F., Thompson, M. E., REDDY, P. S., & Leonard, J. J. (1974) · Journal of Clinical Investigation · doi:10.1172/jci107731 · Cited by 227 · see citing works
The collagen and elastin content of the arterial wall in the dog
Harkness, M. L. R., Harkness, R. D., & McDonald, D. A. (1957) · Proceedings of the Royal Society B Biological Sciences · doi:10.1098/rspb.1957.0029 · Cited by 224 · see citing works
Ambulatory monitoring of the impedance cardiogram
Willemsen, G. A. H. M., de Geus, E. J. C., Klaver, C. H. A. M., VanDoornen, L. J. P., & Carrofl, D. (1996) · Psychophysiology · doi:10.1111/j.1469-8986.1996.tb02122.x · Cited by 222 · see citing works
Adaptive Beat-to-Beat Heart Rate Estimation in Ballistocardiograms
Brüser, C., Stadlthanner, K., de Waele, S., & Leonhardt, S. K. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2128337 · Cited by 219 · see citing works
The difference in pulse transit time to the toe and finger measured by photoplethysmography
Nitzan, M., Khanokh, B., & Slovik, Y. (2001) · Physiological Measurement · doi:10.1088/0967-3334/23/1/308 · Cited by 218 · see citing works
Oscillometric blood pressure: a review for clinicians
Alpert, B. S., Quinn, D. E., & Gallick, D. (2014) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2014.08.014 · Cited by 213 · see citing works
Adaptive threshold method for the peak detection of photoplethysmographic waveform
Shin, H. S., Lee, C., & Lee, M. (2009) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2009.10.006 · Cited by 211 · see citing works
An Automatic Beat Detection Algorithm for Pressure Signals
Aboy, M., McNames, J., Thong, T., Tsunami, D., Ellenby, M. S., & Goldstein, B. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.855725 · Cited by 208 · see citing works
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203 · see citing works
Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., & Yamakoshi, T. (2013) · 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609852 · Cited by 202 · see citing works
Non-constrained Blood Pressure Monitoring Using ECG and PPG for Personal Healthcare
Yoon, Y.-Z., Cho, J. H., & Yoon, G. (2008) · Journal of Medical Systems · doi:10.1007/s10916-008-9186-0 · Cited by 199 · see citing works
A Smart Health Monitoring Chair for Nonintrusive Measurement of Biological Signals
Baek, H. J., Chung, G. S., Kim, K. K., & Park, K. S. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2175742 · Cited by 199 · see citing works
Wong, M. Y. M., Poon, C. C. Y., & Zhang, Y. (2009) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-009-9070-7 · Cited by 191 · see citing works
Measurements of Young's Modulus of Elasticity of the Canine Aorta with Ultrasound
Hughes, D. J., Babbs, C. F., Geddes, L. A., & BOURLAND, J. D. (1979) · Ultrasonic Imaging · doi:10.1177/016173467900100406 · Cited by 189 · see citing works
Monitoring Arterial Blood Pressure: What You May Not Know
McGhee, B. H., & Bridges, E. J. (2002) · Critical Care Nurse · doi:10.4037/ccn2002.22.2.60 · Cited by 189 · see citing works
Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes
Allen, J., & Murray, A. (2002) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001478 · Cited by 179 · see citing works
An Armband Wearable Device for Overnight and Cuff-Less Blood Pressure Measurement
Zheng, Y.-L., Yan, B. P., Zhang, Y., & Poon, C. C. Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2318779 · Cited by 176 · see citing works
Pulse Wave Velocity and Blood Pressure Change: Calibration and Applications
Steptoe, A., Smulyan, H., & Gribbin, B. (1976) · Psychophysiology · doi:10.1111/j.1469-8986.1976.tb00866.x · Cited by 172 · see citing works
Cattivelli, F. S., & Garudadri, H. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.35 · Cited by 172 · see citing works
Large Artery Remodeling During Aging
Gaballa, M. A., Jacob, C. T., Raya, T. E., Liu, J., SIMON, B. R., & Goldman, S. (1998) · Hypertension · doi:10.1161/01.hyp.32.3.437 · Cited by 170 · see citing works
Reddy, K. A., George, B., & Kumar, V. J. (2008) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2008.2009136 · Cited by 165 · see citing works
Westerhof, N., Stergiopulos, N., & Noble, M. I. M. (2010) · Springer eBooks · doi:10.1007/978-1-4419-6363-5 · Cited by 165 · see citing works
Basal perfusion of the cutaneous microcirculation: Measurements as a function of anatomic position
Tur, E., Tur, M., Maibach, H. I., & Guy, R. H. (1983) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12522619 · Cited by 163 · see citing works
Noncontact imaging photoplethysmography to effectively access pulse rate variability
Sun, Y., Hu, S., Azorin-Peris, V., Kalawsky, R. S., & Greenwald, S. E. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.6.061205 · Cited by 163 · see citing works
Kim, E. J., Park, C. G., Park, J., Suh, S. Y., Choi, C. U., Kim, J. W., Kim, S. H., Lim, H. E., Rha, S., Seo, H. S., et al. (2006) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1002120 · Cited by 158 · see citing works
Castiglioni, P., Faini, A., Parati, G., & Rienzo, M. D. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353199 · Cited by 156 · see citing works
Pulse Transit Time: Relationship to Blood Pressure and Myocardial Performance
Obrist, P. A., Light, K. C., McCubbin, J. A., Hutcheson, J. S., & Hoffer, J. L. (1979) · Psychophysiology · doi:10.1111/j.1469-8986.1979.tb02993.x · Cited by 153 · see citing works
Fundamentals of impedance cardiography
PATTERSON, R. P. (1989) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.32403 · Cited by 152 · see citing works
Radar sensing of heartbeat and respiration at a distance with applications of the technology
Greneker, E. F. (1997) · doi:10.1049/cp:19971650 · Cited by 149 · see citing works
In vivo reflectance of blood and tissue as a function of light wavelength
Cui, W., Ostrander, L. E., & Lee, B. Y. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55667 · Cited by 149 · see citing works
Muehlsteff, J., Aubert, X. L., & Schuett, M. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.260275 · Cited by 137 · see citing works
Peng, F., Zhang, Z., Gou, X., Liu, H., & Wang, W. (2014) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-13-50 · Cited by 136 · see citing works
Beta-Receptor Activity in Aorta
Fleisch, J. H., MOLING, H. M., & Brodie, B. B. (1970) · Circulation Research · doi:10.1161/01.res.26.2.151 · Cited by 135 · see citing works
Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
Zhang, G., Gao, M., Xu, D. Z., Olivier, N. B., & Mukkamala, R. (2011) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00980.2011 · Cited by 133 · see citing works
Aging, isometric strength and endurance, and cardiovascular responses to static effort
Petrofsky, J. S., & LIND, A. R. (1975) · Journal of Applied Physiology · doi:10.1152/jappl.1975.38.1.91 · Cited by 133 · see citing works
A novel method to detect Heart Beat Rate using a mobile phone
Pelegris, P., Banitsas, K. A., Orbach, T., & Kostas Marias, A. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626580 · Cited by 126 · see citing works
Noninvasive cuffless blood pressure estimation using pulse transit time and Hilbert–Huang transform
Choi, Y., Zhang, Q., & Ko, S. (2012) · Computers & Electrical Engineering · doi:10.1016/j.compeleceng.2012.09.005 · Cited by 123 · see citing works
Cardiac and respiratory related electrical impedance changes in the human thorax
Brown, B. H., Barber, D. C., Morice, A. H., & Leathard, A. D. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.310088 · Cited by 123 · see citing works
A new method for pulse oximetry possessing inherent insensitivity to artifact
Hayes, M. J., & Smith, P. R. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.915711 · Cited by 122 · see citing works
Circulatory reflexes from carotid and extracarotid baroreceptor areas in man.
Mancia, G., Ferrari, A. U., Gregorini, L., Valentini, R., Ludbrook, J., & Zanchetti, A. (1977) · Circulation Research · doi:10.1161/01.res.41.3.309 · Cited by 120 · see citing works
Noninvasive and Nonocclusive Blood Pressure Estimation Via a Chest Sensor
Solà, J., Proença, M., Ferrario, D., Porchet, J.-A., Falhi, A., Grossenbacher, O., Allemann, Y., Rimoldi, S. F., & Sartori, C. S. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2272699 · Cited by 117 · see citing works
Measurement of Local Pulse Wave Velocity: Effects of Signal Processing on Precision
Hermeling, E., Reesink, K. D., Reneman, R. S., & Hoeks, A. P. G. (2007) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2006.11.018 · Cited by 117 · see citing works
Chen, Y., Wen, C., Tao, G., Bi, M., & Li, G. (2009) · Annals of Biomedical Engineering · doi:10.1007/s10439-009-9759-1 · Cited by 115 · see citing works
陳, 文. (2001) · Medical Entomology and Zoology · Cited by 115 · see citing works
The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals
Lee, H.-W., Lee, J., Jung, W.-G., & Lee, G.-K. (2007) · International Journal of Control Automation and Systems · Cited by 114 · see citing works
Liu, Q., Yan, B. P., Yu, C., Zhang, Y., & Poon, C. C. Y. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2286998 · Cited by 109 · see citing works
Comparison of physiological responses of women and men to isometric exercise
Petrofsky, J. S., Burse, R. L., & Lind, A. R. (1975) · Journal of Applied Physiology · doi:10.1152/jappl.1975.38.5.863 · Cited by 107 · cited 1× · see citing works
“8(b) shows the extent to which these and other well-known interventions perturb BP as well as their limitations, with references providing further information on the interventions [189]–[200].”
BackgroundPulse Transit Time and Blood Pressure: An Intensive Analysis
Lane, J. D., Greenstadt, L., Shapiro, D., & Rubinstein, E. H. (1983) · Psychophysiology · doi:10.1111/j.1469-8986.1983.tb00899.x · Cited by 107 · see citing works
Zong, W., Moody, G. B., & Mark, R. G. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02347553 · Cited by 106 · see citing works
Piezoelectric sensor determination of arterial pulse wave velocity
McLaughlin, J. A., McNeill, M. D. J., Braun, B., & McCormack, P. D. (2003) · Physiological Measurement · doi:10.1088/0967-3334/24/3/306 · Cited by 105 · see citing works
Ahlstrom, C., Johansson, A., Uhlin, F., Länne, T., & Ask, P. (2005) · Journal of Artificial Organs · doi:10.1007/s10047-005-0301-4 · Cited by 104 · see citing works
A Technical Assessment of Pulse Wave Velocity Algorithms Applied to Non-invasive Arterial Waveforms
Gaddum, N. R., Alastruey, J., Beerbaum, P., Chowienczyk, P., & Schaeffter, T. (2013) · Annals of Biomedical Engineering · doi:10.1007/s10439-013-0854-y · Cited by 104 · see citing works
Spīgulis, J., Gailite, L., Lihachev, A., & Erts, R. (2007) · Applied Optics · doi:10.1364/ao.46.001754 · Cited by 98 · see citing works
Rapid Assessment of Cardiac Contractility on a Home Bathroom Scale
Etemadi, M., Inan, O. T., Giovangrandi, L., & Kovacs, G. T. A. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2161998 · Cited by 97 · see citing works
Effect of nitroglycerin on aortic impedance, diameter, and pulse-wave velocity.
Latson, T. W., Hunter, W. C., Katoh, N., & SAGAWA, K. (1988) · Circulation Research · doi:10.1161/01.res.62.5.884 · Cited by 94 · see citing works
Theoretical study on the effect of sensor contact force on pulse transit time
Teng, X.-F., & Zhang, Y.-T. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.900815 · Cited by 94 · see citing works
Ubiquitous Wireless ECG Recording: A Powerful Tool Physicians Should Embrace
Saxon, L. A. (2013) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.12097 · Cited by 92 · see citing works
Li, Q., Mark, R. G., & Clifford, G. D. (2009) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-8-13 · Cited by 92 · see citing works
Shelley, K. H., Tamai, D., Jablonka, D., Gesquiere, M., Stout, R. G., & Silverman, D. G. (2005) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000145063.01043.4b · Cited by 91 · see citing works
Lindberg, L., & Öberg, P. Å. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446295 · Cited by 89 · see citing works
Relationships ol Pulse Transmission Times to Pre‐ejection Period and Blood Pressure
Newlin, D. B. (1981) · Psychophysiology · doi:10.1111/j.1469-8986.1981.tb03042.x · Cited by 89 · see citing works
Regional variation of series elasticity in canine arterial smooth muscles
Cox, R. H. (1978) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1978.234.5.h542 · Cited by 89 · see citing works
Pulse Transit Time and Blood Pressure During Cardiopulmonary Exercise Tests
Wibmer, T., DOERING, K., Kropf-Sanchen, C., RÜDIGER, S., Blanta, I., Stoiber, K., Rottbauer, W., & SCHUMANN, C. (2014) · Physiological Research · doi:10.33549/physiolres.932581 · Cited by 81 · see citing works
Coefficient-Free Blood Pressure Estimation Based on Pulse Transit Time–Cuff Pressure Dependence
Forouzanfar, M., Ahmad, S., Batkin, I., Dajani, H. R., Groza, V. Z., & Bolić, M. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2243148 · Cited by 80 · see citing works
Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure
Masè, M., Mattei, W., Cucino, R., Faes, L., & Nollo, G. (2011) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2010.11.019 · Cited by 78 · see citing works
Hermeling, E., Hoeks, A. P. G., Winkens, M. H. M., Waltenberger, J. L., Reneman, R. S., Kroon, A. A., & Reesink, K. D. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.143867 · Cited by 77 · see citing works
Variability of photoplethysmographyperipheral pulse measurementsat the ears, thumbs and toes
Allen, J., & Murray, A. (2000) · IEE Proceedings - Science Measurement and Technology · doi:10.1049/ip-smt:20000846 · Cited by 76 · see citing works
Jago, J. R., & Murray, A. (1988) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/9/4/003 · Cited by 74 · see citing works
Li, K., & Warren, S. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2167717 · Cited by 73 · see citing works
McCarthy, B. M., Vaughan, C. J., O’Flynn, B., Mathewson, A., & Mathúna, C. Ó. (2013) · Journal of Human Hypertension · doi:10.1038/jhh.2013.41 · Cited by 72 · see citing works
Cuffless Blood Pressure Monitoring Using Hydrostatic Pressure Changes
Shaltis, P., Reisner, A. T., & Asada, H. H. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.919142 · Cited by 71 · see citing works
Imaging the Cardiovascular Pulse
Sun, N., Garbey, M., Merla, A., & Pavlidis, I. T. (2005) · IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings - IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings · doi:10.1109/cvpr.2005.184 · Cited by 67 · see citing works
Towards continuous monitoring of pulse rate in neonatal intensive care unit with a webcam
Mestha, L. K., Kyal, S., Xu, B., Lewis, L., & Kumar, V. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6944455 · Cited by 66 · see citing works
Pulse arrival time as a method of obtaining systolic and diastolic blood pressure indirectly
Geddes, L. A., Voelz, M., James, S., & Reiner, D. (1981) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442787 · Cited by 65 · see citing works
Carotid–femoral pulse wave velocity: Impact of different arterial path length measurements
Sugawara, J., Hayashi, K., Yokoi, T., & Tanaka, H. (2009) · Artery Research · doi:10.1016/j.artres.2009.11.001 · Cited by 65 · see citing works
Smart smooth muscle spring-dampers
Armentano, R. L., Barra, J. G., Pessana, F., Craiem, D. O., Graf, S., Santana, D. B., & Sanchez, R. A. (2007) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2007.289123 · Cited by 65 · see citing works
A new approach for non-intrusive monitoring of blood pressure on a toilet seat
Kim, J. S., Chee, Y. J., Park, J. W., Choi, J. W., & Park, K. S. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/2/010 · Cited by 64 · see citing works
Marie, G. V., Lo, C. R., Jones, J. V., & Johnston, D. W. (1984) · Psychophysiology · doi:10.1111/j.1469-8986.1984.tb00235.x · Cited by 63 · see citing works
Constrained independent component analysis approach to nonobtrusive pulse rate measurements
Tsouri, G. R., Kyal, S., Dianat, S. A., & Mestha, L. K. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.7.077011 · Cited by 62 · see citing works
(1997) · CRC Press eBooks · doi:10.1201/9780367802592 · Cited by 62 · see citing works
Vascular Contraction: Effect of Age and Extracellular Calcium
Cohen, M. L., & Berkowitz, B. A. (2008) · Blood Vessels · doi:10.1159/000158085 · Cited by 61 · see citing works
Impedance cardiographyThe impact of new technology
Critchley, L. A. H. (1998) · Anaesthesia · doi:10.1046/j.1365-2044.1998.437-az0550.x · Cited by 60 · see citing works
Ochiai, R., Takeda, J., Hosaka, H., Sugo, Y., Tanaka, R., & Soma, T. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009950731297 · Cited by 58 · see citing works
Currens, J. H. (1948) · American Heart Journal · doi:10.1016/0002-8703(48)90650-4 · Cited by 58 · see citing works
Estimation Of Systolic And Diastolic Pressure Using The Pulse Transit Time
Ye, S.-Y., Kim, G.-R., Jung, D. K., Baik, S.-W., & Jeon, G.-R. (2010) · doi:10.5281/zenodo.1077836 · Cited by 57 · see citing works
Young, C. C., Mark, J. B., White, W. D., DeBree, A., Vender, J. S., & Fleming, A. J. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617519 · Cited by 56 · see citing works
Hansen, S. K., & Staber, M. (2006) · European Journal of Anaesthesiology · doi:10.1017/s0265021506000688 · Cited by 54 · see citing works
Filtering noncorrelated noise in impedance cardiography
Barros, A. K., Yoshizawa, M., & Yasuda, Y. (1995) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.364522 · Cited by 54 · see citing works
The effect of applied sensor contact force on pulse transit time
Teng, X. F., & Zhang, Y. T. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/8/002 · Cited by 53 · see citing works
Measurement of pulse-wave velocity using a beat-sampling technique
Pruett, J. D., BOURLAND, J. D., & Geddes, L. A. (1988) · Annals of Biomedical Engineering · doi:10.1007/bf02364622 · Cited by 53 · see citing works
Thomas, S. S., Nathan, V., Zong, C., Akinbola, E., Aroul, A. L. P., Philipose, L., Soundarapandian, K., Shi, X., & Jafari, R. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6944076 · Cited by 52 · see citing works
Hermeling, E., Vermeersch, S. J., Rietzschel, E. R., Buyzere, M. D., Gillebert, T. C., van de Laar, R. J., Ferreira, I., Hoeks, A. P. G., Bortel, L. M. V., Reneman, R. S., et al. (2011) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834e4b75 · Cited by 51 · see citing works
An Investigation of Pulse Transit Time as a Non-Invasive Blood Pressure Measurement Method
McCarthy, B. M., O’Flynn, B., & Mathewson, A. (2011) · Journal of Physics Conference Series · doi:10.1088/1742-6596/307/1/012060 · Cited by 50 · see citing works
Use of EMFi as a Blood Pressure Pulse Transducer
Sorvoja, H. S. S., Kokko, V.-M., Myllylä, R. A., & Miettinen, J. (2005) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2005.853345 · Cited by 50 · see citing works
Gaddum, N. R., Keehn, L., Guilcher, A., Gómez, A., Brett, S., Beerbaum, P., Schaeffter, T., & Chowienczyk, P. (2014) · Hypertension · doi:10.1161/hypertensionaha.114.04370 · Cited by 49 · see citing works
A Blood Pressure Monitoring Method for Stroke Management
Ma, H. T. (2014) · BioMed Research International · doi:10.1155/2014/571623 · Cited by 46 · see citing works
Poon, C. C. Y., Zhang, Y., & Liu, Y. (2006) · IEEE/EMBS International Summer School on Medical Devices and Biosensors (ISSS-MDBS) · doi:10.1109/issmdbs.2006.360098 · Cited by 41 · see citing works
Reflectance pulse oximetry at the forehead improves by pressure on the probe
Dassel, A. C. M., Graaff, R., Sikkema, M., Meijer, A., Zijlstra, W. G., & Aarnoudse, J. G. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617518 · Cited by 40 · see citing works
Hill, D. W., & Lowe, H. J. B. (1973) · Medical & Biological Engineering & Computing · doi:10.1007/bf02477397 · Cited by 39 · see citing works
Design challenges for camera oximetry on a mobile phone
Karlen, W., Lim, J., Ansermino, J. M., Dumont, G. A., & Scheffer, C. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346459 · Cited by 39 · see citing works
Effect of adaptive motion-artifact reduction on QRS detection.
Hamilton, P. S., Curley, M., & Aimi, R. (2000) · PubMed · Cited by 39 · see citing works
Adaptive algorithm for continuous monitoring of blood pressure using a pulse transit time
Jadooei, A., Zaderykhin, O., & Shulgin, V. (2013) · IEEE International Conference on Electronics and Nanotechnology (ELNANO) · doi:10.1109/elnano.2013.6552042 · Cited by 35 · see citing works
Chen, Z., Yang, X., Teo, J. T., & Ng, S. H. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610029 · Cited by 35 · see citing works
A pulse transit time measurement method based on electrocardiography and bioimpedance
Bang, S., Lee, C., Park, J., Cho, M., Yoon, Y., & Cho, S. (2009) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2009.5372060 · Cited by 34 · see citing works
Douniama, C., Sauter, C. U., & Couronné, R. (2009) · Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) · Cited by 33 · see citing works
Continuous Blood Pressure Monitoring in Daily Life
Lopez, G. F., Shuzo, M., Ushida, H., Hidaka, K., Yanagimoto, S., Imai, Y., KOSAKA, A., Delaunay, J., & Yamada, I. (2010) · Journal of Advanced Mechanical Design Systems and Manufacturing · doi:10.1299/jamdsm.4.179 · Cited by 32 · see citing works
Static‐charge‐sensitive bed ballistocardiography in cardiovascular monitoring
Lindqvist, A. E., Pihlajamäki, K. K., Jalonen, J. R., LAAKSONEN, V. O., & Alihanka, J. (1996) · Clinical Physiology · doi:10.1111/j.1475-097x.1996.tb00553.x · Cited by 32 · see citing works
Studies on the effect of controlled volume change on the thoracic electrical impedance
PATTERSON, R. P., Kubicek, W. G., Witsoe, D. A., & From, A. H. L. (1978) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457804 · Cited by 31 · see citing works
Estimation of pulse transit time using two diametric blood pressure waveform measurements
Hahn, J., Reisner, A. T., & Asada, H. H. (2010) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2010.04.017 · Cited by 30 · see citing works
Impedance plethysmography for the evaluation of pulse-wave velocity in limbs
Risacher, F., JOSSINET, J., McAdams, E. T., McLaughlin, J. A., Mann, Y., Schmitt, M., Matias, A., & Jarry, R. (1993) · Medical & Biological Engineering & Computing · doi:10.1007/bf02458053 · Cited by 30 · see citing works
McCombie, D. B., Reisner, A. T., & Asada, H. H. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649321 · Cited by 30 · see citing works
Cybulski, G., Książkiewicz, A., Lukasik, W., Niewiadomski, W., & Pałko, T. (2002) · Computers in cardiology · doi:10.1109/cic.1995.482712 · Cited by 30 · see citing works
Ono, T., Miyamura, M., Yasuda, Y., Ito, T., Saito, T., Ishiguro, T., Yoshizawa, M., & Yambe, T. (2004) · The Tohoku Journal of Experimental Medicine · doi:10.1620/tjem.203.17 · Cited by 29 · see citing works
Xu, D. Z., Ryan, K. L., Rickards, C. A., Zhang, G., Convertino, V. A., & Mukkamala, R. (2011) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00443.2011 · Cited by 29 · see citing works
Smart healthcare textile sensor system for unhindered-pervasive health monitoring
Rai, P., Kumar, P. S., Oh, S., Kwon, H., Mathur, G. N., Varadan, V. K., & Agarwal, M. P. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.921253 · Cited by 29 · see citing works
A Novel System Identification Technique for Improved Wearable Hemodynamics Assessment
Wiens, A. D., & Inan, O. T. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2387354 · Cited by 29 · see citing works
[Mobile Health: IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices].
Zhou, X., Wu, W., & Bao, S. (2015) · PubMed · Cited by 29 · see citing works
A piezo-film-based measurement system for global haemodynamic assessment
Clemente, F., Arpaïa, P., & Cimmino, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/5/007 · Cited by 27 · see citing works
Franzblau, S. A. (1968) · JAMA · doi:10.1001/jama.1968.03140020085031 · Cited by 26 · see citing works
A Robust Approach Toward Recognizing Valid Arterial-Blood-Pressure Pulses
Asgari, S., Bergsneider, M., & Hu, X. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2034845 · Cited by 26 · see citing works
Affordable technology: blood pressure measuring devices for low resource settings
Asmar, R. G., O’Brien, E., Mieke, S., et al. (2005) · Newcastle University ePrints (Newcastle Univesity) · doi:10.25405/ncl.34271367.v1 · Cited by 26 · see citing works
Affordable technology: blood pressure measuring devices for low resource settings
Asmar, R. G., O'Brien, E., Mieke, S., van, M. G., Murray, A., Parati, G., & Steenvoorden, G. K. (2005) · Newcastle University ePrints (Newcastle Univesity) · doi:10.25405/ncl.34271367 · Cited by 26 · see citing works
Improved Pulse Wave Velocity Estimation Using an Arterial Tube-Load Model
Gao, M., Zhang, G., Olivier, N. B., & Mukkamala, R. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2291385 · Cited by 21 · see citing works
Heravi, M. A. Y., Khalilzadeh, M. A., & Joharinia, S. (2014) · PubMed · Cited by 21 · see citing works
Robust estimation of pulse wave transit time using group delay
Meloni, A., Zymeski, H., Pepe, A., Lombardi, M., & Wood, J. C. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24207 · Cited by 20 · see citing works
Comparative Study on Tube-Load Modeling of Arterial Hemodynamics in Humans
Rashedi, M., Fazeli, N., Chappell, A., Wang, S., MacArthur, R. G. G., McMurtry, M. S., Finegan, B. A., & Hahn, J. (2013) · Journal of Biomechanical Engineering · doi:10.1115/1.4023373 · Cited by 20 · see citing works
A Novel Calibration Method for Noninvasive Blood Pressure Measurement Using Pulse Transit Time
Yan, Y. S., & Zhang, Y. (2007) · IEEE/EMBS International Summer School on Medical Devices and Biosensors (ISSS-MDBS) · doi:10.1109/issmdbs.2007.4338283 · Cited by 17 · see citing works
Smooth muscle contraction and viscoelasticity of arterial wall
Wurzel, M., Cowper, G. R., & McCooK, J. M. (1970) · Canadian Journal of Physiology and Pharmacology · doi:10.1139/y70-079 · Cited by 17 · see citing works
Liao, C.-F., Cheng, H.-M., Sung, S.-H., Yu, W.-C., & Chen, C.-H. (2010) · Journal of Human Hypertension · doi:10.1038/jhh.2010.106 · Cited by 17 · see citing works
Using Individualized Pulse Transit Time Calibration to Monitor Blood Pressure During Exercise
cheol, J. I., Jeffrey, W., & Finkelstein, J. (2013) · Studies in health technology and informatics · doi:10.3233/978-1-61499-276-9-39 · Cited by 17 · see citing works
Electromechanical delay in the intact dog heart
Kettunen, R., TIMISJÄRVI, J., Kouvalainen, E., Anttila, P., & Hirvonen, L. (1985) · International Journal of Cardiology · doi:10.1016/0167-5273(85)90195-0 · Cited by 16 · see citing works
Automatic detection of characteristic points in impedance cardiogram
Naidu, S. M. M., Pandey, P. C., & Pandey, V. K. (2011) · Computing in cardiology · Cited by 15 · see citing works
Parameters for mean blood pressure estimation based on electrocardiography and photoplethysmography
Špulák, D., Čmejla, R., & Fabián, V. (2011) · International Conference on Applied Electronics · Cited by 14 · see citing works
Data-Driven Lossy Tube-Load Modeling of Arterial Tree: In-Human Study
Abdollahzade, M., Kim, C., Fazeli, N., Finegan, B. A., McMurtry, M. S., & Hahn, J. (2014) · Journal of Biomechanical Engineering · doi:10.1115/1.4028089 · Cited by 14 · see citing works
Flagged first, then by citations
3 of the 212 references couldn’t be matched to a known paper.
Research works printed but not linked: 5
- [16]Eckerle JS Webster JG Tonometry arterial Encyclopedia of Medical Devices and Instrumentation 1988 New York John Wiley & Sons
- [85]Patterson R Development of an electrical impedance plethysmography system to monitor cardiac output Proc of the First Ann Rocky Mountain Bioengineering Symposium 1964 56 71
- [97]Katz AM Physiology of the Heart 2010 5th ed Philadelphia, PA Lippincott Williams & Wilkins
- [161]Costanzo LS Physiology 2003 Saunders
- [199]Froelicher VF Myers J Exercise and the Heart 2006 5 ed Saunders
Cited by
Sorted by citations
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895 · see citing works
Wearables and the Medical Revolution
Dunn, J., Runge, R., & Snyder, M. (2018) · Personalized Medicine · doi:10.2217/pme-2018-0044 · Cited by 654 · see citing works
Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621 · see citing works
Blood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network
Slapničar, G., Mlakar, N., & Luštrek, M. (2019) · Sensors · doi:10.3390/s19153420 · Cited by 387 · see citing works
Self-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348 · see citing works
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 300 · see citing works
Novel Methods for Pulse Wave Velocity Measurement
Pereira, T., Correia, C. M. B. A., & Cardoso, J. M. R. (2015) · Journal of Medical and Biological Engineering · doi:10.1007/s40846-015-0086-8 · Cited by 284 · see citing works
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Management of Hypertension in the Digital Era
Kario, K. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14742 · Cited by 240 · see citing works
Stergiou, G. S., Mukkamala, R., Avolio, A. P., Kyriakoulis, K. G., Mieke, S., Murray, A., Parati, G., Schutte, A. E., Sharman, J. E., Asmar, R. G., et al. (2022) · Journal of Hypertension · doi:10.1097/hjh.0000000000003224 · Cited by 231 · see citing works
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226 · see citing works
Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals
Mukkamala, R., Yavarimanesh, M., Natarajan, K., Hahn, J.-O., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2021) · Hypertension · doi:10.1161/hypertensionaha.121.17747 · Cited by 211 · see citing works
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215 · see citing works
Dieffenderfer, J., Goodell, H., Mills, S., McKnight, M., Yao, S., Lin, F., Beppler, E., Bent, B., Lee, B., Misra, V., et al. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2573286 · Cited by 212 · see citing works
Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method
Chandrasekhar, A., Kim, C., Naji, M. B., Natarajan, K., Hahn, J., & Mukkamala, R. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aap8674 · Cited by 208 · see citing works
Ding, X., Yan, B. P., Zhang, Y., Liu, J., Zhao, N., & Tsang, H. K. (2017) · Scientific Reports · doi:10.1038/s41598-017-11507-3 · Cited by 202 · see citing works
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 210 · see citing works
Kim, K., Choi, J., Jeong, Y., Cho, I., Kim, M., Kim, S., Oh, Y. S., & Park, I. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201900978 · Cited by 186 · see citing works
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180 · see citing works
Huynh, T. H., Jafari, R., & Chung, W. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2865751 · Cited by 181 · see citing works
Photoplethysmography and Deep Learning: Enhancing Hypertension Risk Stratification
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Biosensors · doi:10.3390/bios8040101 · Cited by 170 · see citing works
Introducing Contactless Blood Pressure Assessment Using a High Speed Video Camera
Jeong, I. C., & Finkelstein, J. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0439-z · Cited by 161 · see citing works
Local Pulse Wave Velocity: Theory, Methods, Advancements, and Clinical Applications
Nabeel, P. M., Kiran, V. R., Joseph, J., Abhidev, V. V., & Sivaprakasam, M. (2019) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2019.2931587 · Cited by 158 · see citing works
Esmaili, A., Kachuee, M., & Shabany, M. (2017) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2017.2745081 · Cited by 155 · see citing works
Wearable Devices for Precision Medicine and Health State Monitoring
Jeong, I. C., Bychkov, D., & Searson, P. C. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2871638 · Cited by 155 · see citing works
Kim, C., Carek, A. M., Mukkamala, R., Inan, O. T., & Hahn, J. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2440291 · Cited by 155 · see citing works
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102301 · Cited by 153 · see citing works
Miao, F., Wen, B., Hu, Z., Fortino, G., Wang, X., Liu, Z.-D., Tang, M., & Li, Y. (2020) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2020.101919 · Cited by 145 · see citing works
Sharman, J. E., O’Brien, E., Alpert, B., Schutte, A. E., Delles, C., Hecht Olsen, M., Asmar, R., Atkins, N., Barbosa, E., Calhoun, D., et al. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002246 · Cited by 139 · see citing works
Liu, J., Yan, B. P., Zhang, Y., Ding, X., Su, P., & Zhao, N. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2874957 · Cited by 139 · see citing works
Li, H., Ma, Y., Liang, Z., Wang, Z., Cao, Y. K., Xu, Y., Zhou, H., Lu, B., Chen, Y., Han, Z., et al. (2020) · National Science Review · doi:10.1093/nsr/nwaa022 · Cited by 142 · see citing works
A review of wearable and unobtrusive sensing technologies for chronic disease management
Guo, Y., Liu, X., Peng, S., Jiang, X., Xu, K., Chen, C., Wang, Z., Dai, C., & Chen, W. (2020) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2020.104163 · Cited by 137 · see citing works
Multi-Sensor Fusion Approach for Cuff-Less Blood Pressure Measurement
Miao, F., Liu, Z.-D., Liu, J., Wen, B., He, Q., & Li, Y. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2901724 · Cited by 136 · see citing works
Continuous PPG-Based Blood Pressure Monitoring Using Multi-Linear Regression
Haddad, S., Boukhayma, A., & Caizzone, A. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3128229 · Cited by 139 · see citing works
Monitoring vital signs over multiplexed radio by near-field coherent sensing
Hui, X., & Kan, E. C.-C. (2017) · Nature Electronics · doi:10.1038/s41928-017-0001-0 · Cited by 135 · see citing works
A Wearable Tele-Health System towards Monitoring COVID-19 and Chronic Diseases
Jiang, W., Majumder, S., Kumar, S., Subramaniam, S., Li, X., Khedri, R., Mondal, T. K., Abolghasemian, M., Satia, I., & Deen, M. J. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3069815 · Cited by 137 · see citing works
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138 · see citing works
Martin, S. L.-O., Carek, A. M., Kim, C., Ashouri, H., Inan, O. T., Hahn, J., & Mukkamala, R. (2016) · Scientific Reports · doi:10.1038/srep39273 · Cited by 131 · see citing works
Hypertension Assessment via ECG and PPG Signals: An Evaluation Using MIMIC Database
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Diagnostics · doi:10.3390/diagnostics8030065 · Cited by 132 · see citing works
Cuffless blood pressure estimation from PPG signals and its derivatives using deep learning models
El-Hajj, C., & Kyriacou, P. A. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102984 · Cited by 136 · see citing works
Zhang, Q., Zhou, D., & Zeng, X. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0317-z · Cited by 132 · see citing works
Blood Pressure Estimation Using Pulse Transit Time From Bioimpedance and Continuous Wave Radar
Buxi, D., Jean-Michel, R., & Yuce, M. R. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2582472 · Cited by 121 · see citing works
Pramanik, P. K. D., Upadhyaya, B. K., Pal, S., & Pal, T. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-815368-0.00001-4 · Cited by 121 · see citing works
Emerging sensing and modeling technologies for wearable and cuffless blood pressure monitoring
Zhao, L., Liang, C., Huang, Y., Zhou, G., Xiao, Y., Ji, N., Zhang, Y., & Zhao, N. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00835-6 · Cited by 120 · see citing works
Franklin, D., Tzavelis, A., Lee, J. Y., Chung, H. U., Trueb, J. T., Arafa, H., Kwak, S. S., Huang, I. A., Liu, Y., Rathod, M., et al. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01098-y · Cited by 119 · see citing works
Sharifi, I., Goudarzi, S., & Khodabakhshi, M. B. (2018) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2018.12.005 · Cited by 115 · see citing works
End-to-End Blood Pressure Prediction via Fully Convolutional Networks
Baek, S., Jang, J.-Y., & Yoon, S. (2019) · IEEE Access · doi:10.1109/access.2019.2960844 · Cited by 114 · see citing works
Zhang, Q., Zeng, X., Hu, W. W., & Zhou, D. (2017) · IEEE Access · doi:10.1109/access.2017.2707472 · Cited by 113 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring.
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · KTUePubl (Repository of Kaunas University of Technology) · doi:10.17863/cam.81148 · Cited by 113 · see citing works
PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement
Chandrasekhar, A., Yavarimanesh, M., Natarajan, K., Hahn, J., & Mukkamala, R. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2976989 · Cited by 113 · see citing works
Definition of Fiducial Points in the Normal Seismocardiogram
Sørensen, K., Schmidt, S. E., Jensen, A. S., Søgaard, P., & Struijk, J. J. (2018) · Scientific Reports · doi:10.1038/s41598-018-33675-6 · Cited by 111 · see citing works
Konijnenburg, M., Stanzione, S., Yan, L., Jee, D., Pettine, J., van Wegberg, R., Kim, H. J., van Liempd, C., Fish, R., Schuessler, J., et al. (2016) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2016.2605660 · Cited by 109 · see citing works
Aguirre, N. A., Grall‐Maës, E., Cymberknop, L. J., & Armentano, R. L. (2021) · Sensors · doi:10.3390/s21062167 · Cited by 110 · see citing works
Ibrahim, B., & Jafari, R. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2946661 · Cited by 106 · see citing works
Holz, C., & Wang, E. J. (2017) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3132024 · Cited by 106 · see citing works
Kim, C., Carek, A. M., Inan, O. T., Mukkamala, R., & Hahn, J. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2797239 · Cited by 104 · see citing works
Conventional pulse transit times as markers of blood pressure changes in humans
Block, R., Yavarimanesh, M., Natarajan, K., Carek, A. M., Mousavi, A. S., Chandrasekhar, A., Kim, C., Zhu, J., Schifitto, G., Mestha, L. K., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-73143-8 · Cited by 107 · see citing works
A novel art of continuous noninvasive blood pressure measurement
Fortin, J., Rogge, D. E., Fellner, C., Flotzinger, D., Grond, J., Lerche, K., & Saugel, B. (2021) · Nature Communications · doi:10.1038/s41467-021-21271-8 · Cited by 104 · see citing works
Carek, A. M., Conant, J., Joshi, A. N., Kang, H., & Inan, O. T. (2017) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3130905 · Cited by 103 · see citing works
Cuffless blood pressure estimation using only a smartphone
Matsumura, K., Rolfe, P. J., Toda, S., & Yamakoshi, T. (2018) · Scientific Reports · doi:10.1038/s41598-018-25681-5 · Cited by 102 · see citing works
Wang, W., Mohseni, P., Kilgore, K. L., & Najafizadeh, L. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3128383 · Cited by 107 · see citing works
Wang, E. J., Zhu, J., Jain, M., Lee, T.-J., Saba, E., Nachman, L., & Patel, S. (2018) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3173574.3173999 · Cited by 103 · see citing works
Cuffless Blood Pressure Monitoring: Promises and Challenges
Pandit, J. A., Lores, E., & Batlle, D. (2020) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.03680320 · Cited by 101 · see citing works
Wearable Cuff-Less Blood Pressure Estimation at Home via Pulse Transit Time
Ganti, V. G., Carek, A. M., Nevius, B. N., Heller, J. A., Etemadi, M., & Inan, O. T. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3021532 · Cited by 101 · see citing works
Smets, E., Raedt, W. D., & Hoof, C. A. V. (2018) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2018.2883751 · Cited by 102 · see citing works
Peng, C., Chen, M., Sim, H. K., Zhu, Y., & Jiang, X. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3021923 · Cited by 98 · see citing works
Boosting the Battery Life of Wearables for Health Monitoring Through the Compression of Biosignals
Hooshmand, M., Zordan, D., Testa, D. D., Grisan, E., & Rossi, M. (2017) · IEEE Internet of Things Journal · doi:10.1109/jiot.2017.2689164 · Cited by 97 · see citing works
Hypertension Assessment Using Photoplethysmography: A Risk Stratification Approach
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Journal of Clinical Medicine · doi:10.3390/jcm8010012 · Cited by 99 · see citing works
González, S., Hsieh, W., & Chen, T. P.-C. (2023) · Scientific Data · doi:10.1038/s41597-023-02020-6 · Cited by 99 · see citing works
Mieloszyk, R. J., Twede, H., Lester, J., Wander, J., Basu, S., Cohn, G., Smith, G., Morris, D., Gupta, S., Tan, D., et al. (2022) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2022.3153259 · Cited by 96 · see citing works
Xing, X., Ma, Z., Zhang, M., Zhou, Y., Dong, W., & Song, M. (2019) · Scientific Reports · doi:10.1038/s41598-019-45175-2 · Cited by 93 · see citing works
Pulse arrival time as a surrogate of blood pressure
Finnegan, E., Davidson, S. M., Harford, M., Jorge, J., Watkinson, P., Young, D., Tarassenko, L., & Villarroel, M. (2021) · Scientific Reports · doi:10.1038/s41598-021-01358-4 · Cited by 93 · see citing works
A Chair–Based Unobtrusive Cuffless Blood Pressure Monitoring System Based on Pulse Arrival Time
Tang, Z., TAMURA, T., Sekine, M., Huang, M., Chen, W., Yoshida, M., Sakatani, K., Kobayashi, H., & Kanaya, S. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2614962 · Cited by 91 · see citing works
Liang, Y., Abbott, D., Howard, N., Lim, K. I., Ward, R. K., & Elgendi, M. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8030337 · Cited by 94 · see citing works
Rong, M., & Li, K. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102328 · Cited by 89 · see citing works
Javaid, A. Q., Ashouri, H., Dorier, A., Etemadi, M., Heller, J. A., Roy, S., & Inan, O. T. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2600945 · Cited by 88 · see citing works
Gao, M., Olivier, N. B., & Mukkamala, R. (2016) · Physiological Reports · doi:10.14814/phy2.12768 · Cited by 86 · see citing works
Self-citationCuff-Less Blood Pressure Estimation Using Pulse Waveform Analysis and Pulse Arrival Time
Yoon, Y.-Z., Kang, J. M., Kwon, Y., Park, S., Noh, S., Kim, Y., Park, J., & Hwang, S. W. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2714674 · Cited by 88 · see citing works
Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device
Watanabe, N., Bando, Y., Kawachi, T., Yamakita, H., Futatsuyama, K., Honda, Y., Yasui, H., Nishimura, K., Kamihara, T., Okumura, T., et al. (2017) · JACC Basic to Translational Science · doi:10.1016/j.jacbts.2017.07.015 · Cited by 87 · see citing works
Mukkamala, R., & Hahn, J. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2756018 · Cited by 86 · see citing works
Bi-Modal Arterial Compliance Probe for Calibration-Free Cuffless Blood Pressure Estimation
Nabeel, P. M., Joseph, J., Srinivasa, K., Sivaprakasam, M., & Chenniappan, M. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2866332 · Cited by 83 · see citing works
Cuff-Less Methods for Blood Pressure Telemonitoring
Bard, D. M., Joseph, J. I., & van Helmond, N. (2019) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2019.00040 · Cited by 84 · see citing works
Gurel, N. Z., Huang, M., Wittbrodt, M. T., Jung, H., Ladd, S. L., Shandhi, M. M. H., Ko, Y., Shallenberger, L. H., Nye, J. A., Pearce, B. D., et al. (2019) · Brain stimulation · doi:10.1016/j.brs.2019.08.002 · Cited by 84 · see citing works
Bed-Based Ballistocardiography: Dataset and Ability to Track Cardiovascular Parameters
Carlson, C., Turpin, V. G., Suliman, A., Ade, C. J., Warren, S., & Thompson, D. E. (2020) · Sensors · doi:10.3390/s21010156 · Cited by 84 · see citing works
Barvik, D., Černý, M., Penhaker, M., & Noury, N. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3109643 · Cited by 85 · see citing works
The Microsoft Research Aurora Project: Important Findings on Cuffless Blood Pressure Measurement
Mukkamala, R., Shroff, S. G., Landry, C., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2022) · Hypertension · doi:10.1161/hypertensionaha.122.20410 · Cited by 82 · see citing works
Continuous Non-Invasive Blood Pressure Monitoring: A Methodological Review on Measurement Techniques
Le, T., Ellington, F., Lee, T.-Y., Vo, K., Khine, M., Krishnan, S. K., Dutt, N. D., & Cao, H. (2020) · IEEE Access · doi:10.1109/access.2020.3040257 · Cited by 84 · see citing works
Ibrahim, B., & Jafari, R. (2022) · Scientific Reports · doi:10.1038/s41598-021-03612-1 · Cited by 86 · see citing works
Estimating blood pressure trends and the nocturnal dip from photoplethysmography
Radha, M., de Groot, K., Rajani, N. B., Wong, C. C. P., Kobold, N., Vos, V., Fonseca, P. M., Mastellos, N., Wark, P. A., Velthoven, N., et al. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab030e · Cited by 85 · see citing works
Wearable Millimeter-Wave Device for Contactless Measurement of Arterial Pulses
Johnson, J. E., Shay, O., Kim, C., & Liao, C. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2948581 · Cited by 78 · see citing works
Rastegar, S., GholamHosseini, H., & Lowe, A. (2019) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-019-00813-x · Cited by 80 · see citing works
Zheng, Y., Poon, C. C. Y., Yan, B. P., & Lau, J. Y. W. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0558-6 · Cited by 82 · see citing works
Nabeel, P. M., Jayaraj, J., & Sivaprakasam, M. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa9550 · Cited by 80 · see citing works
Gurel, N. Z., Wittbrodt, M. T., Jung, H., Shandhi, M. M. H., Driggers, E. G., Ladd, S. L., Huang, M., Ko, Y., Shallenberger, L. H., Beckwith, J., et al. (2020) · Neurobiology of Stress · doi:10.1016/j.ynstr.2020.100264 · Cited by 79 · see citing works
Finnegan, E., Davidson, S. M., Harford, M., Watkinson, P., Tarassenko, L., & Villarroel, M. (2023) · Scientific Reports · doi:10.1038/s41598-022-27170-2 · Cited by 83 · see citing works
An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
Huynh, T. H., Jafari, R., & Chung, W. (2018) · Sensors · doi:10.3390/s18072095 · Cited by 80 · see citing works
Lin, W.-H., Wang, H., Samuel, O. W., Liu, G., Huang, Z., & Li, G. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aaa454 · Cited by 78 · see citing works
Maqsood, S., Xu, S., Tran, S. N., Garg, S., Springer, M., Karunanithi, M., & Mohawesh, R. (2022) · Expert Systems with Applications · doi:10.1016/j.eswa.2022.116788 · Cited by 78 · see citing works
- Authors
- Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S.
- Affiliations
- Michigan State University (US)
- University of Maryland, College Park (US)
- Georgia Institute of Technology (US)
- Palo Alto Research Center (US)
- Corresponding author
- Mukkamala, Ramakrishna
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2015-08
- Volume
- 62
- Issue
- 8
- Pages
- 1879-1901
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- PMC
- PMC4515215
- OpenAlex
- W2294418644
- Cited by
- 935
- References
- 212
- Citations by year
- 2026: 92 · 2025: 93 · 2024: 96 · 2023: 89 · 2022: 93 · 2021: 87 · 2020: 84 · 2019: 92 · 2018: 83 · 2017: 71 · 2016: 50 · 2015: 4
- FWCI
- 34.44
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Biomedical Imaging and Bioengineering
- National Institutes of Health (U01 EB018818)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Keywords
- blood pressure monitoring, pulse transit time, ubiquitous health monitoring, wearable sensors, hypertension detection, non-invasive blood pressure estimation
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Clothing, Equipment Design, Humans, Signal Processing, Computer-Assisted, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review