The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis
Open accessThe DOI matches its Crossref record; 67 of the 69 references it cites were found in a registry or an index; 4 of the 16 sources its claims cite were checked for support.
Abstract
Not much is known about how accurate and reproducible different thermometers are at diagnosing patients with suspected fever. The study aims at evaluating which peripheral thermometers are more accurate and reproducible. We searched Medline, Embase, Scopus, WOS, CENTRAL, and Cinahl to perform: (1) diagnostic accuracy meta-analysis (MA) using rectal mercury-in-glass or digital thermometry as reference, and bivariate models for pooling; (2) network MA to estimate differences in mean temperature between devices; (3) Bland-Altman method to estimate 95% coefficient of reproducibility. PROSPERO registration: CRD42020174996. We included 46 studies enrolling more than 12,000 patients. Using 38 °C (100.4 ℉) as cut-off temperature, temporal infrared thermometry had a sensitivity of 0.76 (95% confidence interval, 0.65, 0.84; low certainty) and specificity of 0.96 (0.92, 0.98; moderate certainty); tympanic infrared thermometry had a sensitivity of 0.77 (0.60, 0.88; low certainty) and specificity of 0.98 (0.95, 0.99; moderate certainty). For all the other index devices, it was not possible to pool the estimates. Compared to the rectal mercury-in-glass thermometer, mean temperature differences were not statistically different from zero for temporal or tympanic infrared thermometry; the median coefficient of reproducibility ranged between 0.53 °C [0.95 ℉] for infrared temporal and 1.2 °C [2.16 ℉] for axillary digital thermometry. Several peripheral thermometers proved specific, but not sensitive for diagnosing fever with rectal thermometry as a reference standard, meaning that finding a temperature below 38 °C does not rule out fever. Fixed differences between temperatures together with random error means facing differences between measurements in the order of 2 °C [4.5 ℉]. This study informs practitioners of the limitations associated with different thermometers; peripheral ones are specific but not sensitive.
Overview
CoChat Summary
## TL;DR
This systematic review and network meta-analysis found that peripheral thermometers are generally specific but not sensitive for detecting fever when compared with rectal thermometry at a 38 °C threshold. Temporal artery infrared and tympanic infrared thermometers performed best overall, but substantial random error remained across devices.
## Key findings
- The review included 46 studies with 12,602 patients; 43 studies contributed to quantitative analyses.
- For fever detection at 38 °C, temporal artery infrared thermometry had pooled sensitivity 0.76 (95% CI 0.65–0.84) and specificity 0.96 (0.92–0.98) in 9 studies (2,533 participants; 885 with fever).
- For fever detection at 38 °C, tympanic infrared thermometry had pooled sensitivity 0.77 (0.60–0.88) and specificity 0.98 (0.95–0.99) in 9 studies (3,862 participants; 1,279 with fever).
- Compared with rectal mercury-in-glass thermometry, axillary digital temperature was significantly lower by −0.67 °C (−0.98, −0.37), and axillary mercury-in-glass by −0.55 °C (−0.87, −0.23); the median 95% coefficient of reproducibility ranged from 0.53 °C for infrared temporal to 1.2 °C for axillary digital thermometry.
- The authors report that the mean 95% coefficient of reproducibility across 73 direct comparisons was 1.06 °C, and that fixed plus random error could lead to differences of about 2 °C in the worst case.
## Methods
- Systematic review and meta-analysis conducted according to Cochrane DTA guidance and reported per PRISMA-DTA; registered on PROSPERO CRD42020174996.
- Searched Medline, Embase, Web of Science, Scopus, CENTRAL, and CINAHL up to March 2020 for studies of adults or children in emergency, hospital inpatient, or neonatal settings comparing thermometer types.
- Used bivariate meta-analysis for diagnostic accuracy, and network meta-analysis for mean differences between thermometer types; analyses were centered on a 38 °C fever threshold and rectal mercury-in-glass or digital thermometry as reference standard.
## Limitations
- The paper states that there were too few studies to estimate accuracy for some devices/sites and settings, including a specific setting such as the ED.
- The authors state that overall certainty was moderate because of limitations in study design and considerable heterogeneity across studies, including different body sites, mixed child/adult populations, threshold effects, and operator variability.
- Evident from the text: many studies had unclear enrollment, limited blinding, and some thermometers/sites had only one to three studies, which restricts precision and generalizability.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures




Verification
Claims the paper makes
4 Valid · 0 Not valid · 7 Not checked · 41 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“axillary digital thermometry was significantly lower by − 0.67 °C (− 0.98, − 0.37)”
§ The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis > Results > Mean differences between thermometers using network meta-analyses
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mean 95% CR value of 73 direct comparisons between devices was 1.06 °C”
§ The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis > Results > Description of random error for each direct comparison
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Compared to the rectal mercury-in-glass thermometer, mean temperature differences were not statistically different from zero for temporal or tympanic infrared thermometry”
§ The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“For all the other index devices, it was not possible to pool the estimates.”
- Also stated in: § Diagnostic accuracy estimates · plain
“There were three or fewer studies on other devices/sites and cut-offs (Appendix 2), thus meta-analyses were not possible.”
§ The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Diagnostic accuracy estimates · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“the median coefficient of reproducibility ranged between 0.53 °C [0.95 ℉] for infrared temporal and 1.2 °C [2.16 ℉] for axillary digital thermometry.”
- Also stated in: § Description of random error for each direct comparison · plain
“The median 95%CR vs. rectal mercury-in-glass thermometry was 1.16 °C for axillary digital thermometry (three studies), 0.79 °C for digital oral thermometry (one study), 0.70 °C for digital rectal thermometry (one study), 0.73 °C for tympanic infrared thermometry (six studies), 1.08 °C for infrared forehead thermometry (three studies) and 0.53 °C for infrared temporal thermometry (one study).”
§ The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Description of random error for each direct comparison · plain
13 more claims
Create a free account to see all 52 claims and to verify the 3 references still open.
- Probably not a substantial claim. 17
- It may state two separate claims as one. 23
Citations
Cites
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · cited 2×
“96 times the standard deviation (SD) of these differences (random error), which are the two components of the 95% coefficient of reproducibility [14].”
Reitsma, J. B., Glas, A. S., Rutjes, A. W. S., Scholten, R. J. P. M., Bossuyt, P. M., & Zwinderman, A. H. (2005) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2005.02.022 · Cited by 3,369 · cited 1×
“We used mixed models to fit bivariate meta-analyses, which model sensitivity and specificity while accounting for their correlation [16].”
McInnes, M. D. F., Moher, D., Thombs, B. D., McGrath, T. A., Bossuyt, P. M., and the PRISMA-DTA Group, Clifford, T., Cohen, J. F., Deeks, J. J., Gatsonis, C., et al. (2018) · JAMA · doi:10.1001/jama.2017.19163 · Cited by 3,210 · cited 1×
“The reporting was in accordance with the Preferred Reporting Items for Systematic reviews and Meta-Analyses of Diagnostic Test Accuracy Studies (PRISMA-DTA) criteria [10].”
Graphical tools for network meta-analysis in STATA
Chaimani, A., Higgins, J. P. T., Mavridis, D., Spyridonos, P., & Salanti, G. (2013) · PLoS ONE · doi:10.1371/journal.pone.0076654 · Cited by 2,473 · cited 11×
“(Mercury-in-glass or digital rectal at 38 °C as cut-off as reference device) NMA Device No.”
InfluentialMeta-analyses involving cross-over trials: methodological issues
Elbourne, D. R., Altman, D. G., Higgins, J. P., Curtin, F., Worthington, H. V., & Vail, A. (2002) · International Journal of Epidemiology · doi:10.1093/ije/31.1.140 · Cited by 2,305 · cited 3×
“Niven et al. calculated 95% coefficients of reproducibility as twice the SDs but did not explain how SDs were pooled across studies [2].”
InfluentialGrading quality of evidence and strength of recommendations for diagnostic tests and strategies
Schünemann, H. J., Oxman, A. D., Brozek, J., Glasziou, P., Jaeschke, R., Vist, G. E., Williams, J. W., Kunz, R., Craig, J., Montori, V. M., et al. (2008) · BMJ · doi:10.1136/bmj.39500.677199.ae · Cited by 1,782
White, I. R. (2015) · The Stata Journal: Promoting communications on statistics and Stata · doi:10.1177/1536867x1501500403 · Cited by 766
metandi: Meta-analysis of diagnostic accuracy using hierarchical logistic regression
Harbord, R. M., & Whiting, P. (2009) · The Stata Journal: Promoting communications on statistics and Stata · doi:10.1177/1536867x0900900203 · Cited by 453 · cited 1×
“We used the STATA metandi package [20] to fit bivariate models, the STATA network routine to perform NMA [21] and the STATA metan function to obtain pairwise metaanalyses [22].”
Nikolakopoulou, A., White, I. R., & Salanti, G. (2020) · Handbook of Meta-Analysis · doi:10.1201/9781315119403-10 · Cited by 303
Cited by
Exploring Thermal Dynamics in Wound Healing: The Impact of Temperature and Microenvironment
Huang, J., Fan, C., Ma, Y., & Huang, G. (2024) · Clinical Cosmetic and Investigational Dermatology · doi:10.2147/ccid.s468396 · Cited by 53
“The archetypical modalities for quantifying thermal variations predominantly rely on single-point detection mechanisms, incorporating instruments such as thermocouple thermometers, 21 infrared thermometers, 22 and infrared thermographic cameras (Figure 1).”
BackgroundAjčević, M., Stella, A. B., Furlanis, G., Caruso, P., Naccarato, M., Accardo, A., & Manganotti, P. (2022) · Sensors · doi:10.3390/s22134760 · Cited by 41
Chen, A., Zhu, J., Lin, Q., & Liu, W. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph192315883 · Cited by 35
“[3] provided a summary of the research on detecting body temperature in hospital systems using electronic thermometers, infrared thermometers, and mercury thermometers.”
BackgroundMicrogels sense wounds' temperature, pH and glucose
Xiao, Y., Xu, K., Zhao, P., Ji, L., Hua, C., Jia, X., Wu, X., Diao, L., Zhong, W., Lyu, G., et al. (2024) · Biomaterials · doi:10.1016/j.biomaterials.2024.122813 · Cited by 35
Comparative Analysis of Human Body Temperatures Measured with Noncontact and Contact Thermometers
Dolibog, P., Pietrzyk, B., Kierszniok, K., & Pawlicki, K. (2022) · Healthcare · doi:10.3390/healthcare10020331 · Cited by 40
“Their disadvantage is the fact that, like traditional thermometers, they are contact devices, which require the implementation of disinfecting agents [7].”
BackgroundTemperature sensor based on an erbium-doped fiber Sagnac interferometer
Pizzaia, J. P. L., Silva, R. L., Junior, A. G. L., & Castellani, C. E. S. (2022) · Applied Optics · doi:10.1364/ao.448377 · Cited by 27
Reliability of Non-Contact Infrared Thermometers for Fever Screening Under COVID-19
Lai, F., Li, X., Wang, Q., Luo, Y., Wang, X., Huang, X., Zhang, J., Peng, J., Wang, Q., Fan, L., et al. (2022) · Risk Management and Healthcare Policy · doi:10.2147/rmhp.s357567 · Cited by 27
Post‐craniotomy fever and its associated factors in patients with traumatic brain injury
Huang, J., Wang, T.-J. T., Wu, S., et al. (2021) · Nursing in Critical Care · doi:10.1111/nicc.12640 · Cited by 13
Ehlers, U. E., Ulmer, J., Keller, M., Klein, C., & Pietsch, U. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01199-2 · Cited by 10
“However, the fever detection rate of the tympanic temperature probe was also lower than reported in other studies at 86% [33].”
ResultKasbekar, R. S., Naz, A., Marcos, L. S., Liu, Y., Hendrickson, K., Gorsich, J. C., & Baun, M. (2021) · Nursing Open · doi:10.1002/nop2.831 · Cited by 10
Physiological strain in outdoor workers: The hidden danger of high humidity
Isaac, T. S., Ranjith, S., Latha, P. K., Shanmugam, R., & Venugopal, V. (2025) · Environmental Research · doi:10.1016/j.envres.2025.121495 · Cited by 12
Optimal diagnostic fever thresholds using non-contact infrared thermometers under COVID-19
Lai, F., Li, X., Liu, T., Wang, X., Wang, Q., Chen, S., Wei, S., Xiong, Y., Hou, Q., Zeng, X., et al. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.985553 · Cited by 8
Malmberg, S., Khan, T., Gunnarsson, R. K., Jacobsson, G., & Sundvall, P. (2022) · Scandinavian Journal of Infectious Diseases · doi:10.1080/23744235.2022.2080249 · Cited by 9
“For many years, contactless body temperature estimation has been tried with various accuracy [21].”
BackgroundFever Management in Children and Adolescents
Niehues, T., Krafft, H. S., Schemmer, J., Schwarz, S., Möhler, R., & Martin, D. (2026) · Deutsches Ärzteblatt international · doi:10.3238/arztebl.m2026.0002 · Cited by 4
Optimizing pharmacological management of the febrile child
Bianchi, M., Costa, M. G., Cardinale, F., Nardo, G. D., Mennini, M., Orsini, A. A., Foiadelli, T., Striano, P., Parisi, P., & Ferretti, A. (2025) · Expert Opinion on Pharmacotherapy · doi:10.1080/14656566.2025.2592793 · Cited by 5
“However, this method is invasive and may cause discomfort in children [28].”
InfluentialALPER, F. M. P. (2024) · IEEE Instrumentation & Measurement Magazine · doi:10.1109/mim.2024.10622085 · Cited by 3
“Among these optical systems, infrared tympanic ther-mometers have emerged as practical tools for accurate and non-invasive temperature measurements within the tympa-num’s range (36–41 °C) [3],[4].”
BackgroundFever assessment in children under five: Are we following the guidelines?
Isa, H. M. A., Isa, A. J., Alnasheet, M. A., & Mansoor, M. H. (2024) · World Journal of Clinical Pediatrics · doi:10.5409/wjcp.v13.i1.88864 · Cited by 4
Evaluating the Accuracy of Low-Cost Wearable Sensors for Healthcare Monitoring
Filgueiras, T. P., Bertemes-Filho, P., & Noveletto, F. (2025) · Micromachines · doi:10.3390/mi16070791 · Cited by 10
USING INFRARED THERMOGRAPHIC TECHNIQUE AS AN ALTERNATIVE TO CONVENTIONAL RECTAL THERMOMETER IN SHEEP
Hussein, N. J., Habib, R. S., Khalaf, F. M., Fazal, R. M., & Owais, M. (2025) · Science Journal of University of Zakho · doi:10.25271/sjuoz.2025.13.1.1446 · Cited by 2
“This close agreement agrees with the high accuracy predicted by digital thermometers for measuring the core body temperature (Pecoraro et al ., 2021), and the ability of the IR camera to capture precise temperature readings as a non-invasive instrument, providing that it is correctly calibrated and…”
Li, R., Shen, M., Liu, H., et al. (2023) · JMIR Formative Research · doi:10.2196/42548 · Cited by 2
“The use of infrared thermometers for fever detection is a mature technology, with at least 80% sensitivity and 95% specificity [4,5].”
BackgroundAliabadi, M., Motlagh, M. S., Golmohammadi, R., Heidarimoghadam, R., & Farhadian, M. (2022) · PLoS ONE · doi:10.1371/journal.pone.0279170 · Cited by 3
“The main reason for selecting ET to represent body temperature was the simple measurement of ET using an accurate and non-invasive thermometer [45].”
MethodLiu, T., Peng, C., Hsu, F., Chang, L., Wang, H., & Chang, W. (2022) · DEN Open · doi:10.1002/deo2.159 · Cited by 2
Continuous wireless temperature monitoring using TempTraq® in children
Liu, S., Nataraja, R. M., Medhurst, A., & Pacilli, M. (2025) · Pediatric Research · doi:10.1038/s41390-025-04553-5 · Cited by 1
Fieber bei Kindern und Jugendlichen in der HNO-Heilkunde
Goncalves, M., Hackenberg, S., & Martin, D. (2026) · HNO · doi:10.1007/s00106-026-01775-9
Pinheiro, A. C., Abbud, R. S., Perestrelo, K. G., Amorim, J. E., Freitas, J. M. N., & Sucupira, M. C. A. (2026) · Tropical Animal Health and Production · doi:10.1007/s11250-026-05059-w · Cited by 2
Evolution of medical thermometers: from antique prototypes to the modern devices
Zakharova, I. N., Berezhnaya, I. V., Kolushkin, D. S., et al. (2026) · Pediatrics Consilium Medicum · doi:10.26442/26586630.2026.1.203474
Erste AWMF-S3-Leitlinie zum ambulanten Fiebermanagement bei Kindern und Jugendlichen erschienen
Martin, D., Schemmer, J., Krafft, H. S., Möhler, R., Schwarz, S., & Niehues, T. (2026) · Zeitschrift für Komplementärmedizin · doi:10.1055/a-2765-8288
Fu, H., Guo, Y., Shi, Q., et al. (2026) · Journal of Optics · doi:10.1007/s12596-025-02988-x
Implementation of Bioheat Equation in Lumped-Parameter Blood Circulation Model
Akay, S., Yıldırım, C., Ertürk, H., et al. (2026) · Lecture notes in mechanical engineering · doi:10.1007/978-3-032-16138-3_4
Bridging the Gap in Pain Measurement with a Brain-Based Index
Segning, C. M., Bouhali, A., de Oliveira, L. V. F., Oliveira, C. S., da Silva, R. A., Fernandes, K. B. P., & Ngomo, S. (2025) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph23010033
“For body temperature, differences of up to ± 0.5 ◦ C are commonly accepted between contact digital thermometers and infrared forehead thermometers, corresponding to a mean bias of around ± 2–3% and SD typically under 0.3 ◦ C [38].”
InfluentialMcGee, S. R. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-25013-2.00018-4
Inclusion of S-Boxes in Speck Cipher: Impact on Differential Cryptanalysis and Security
Rajput, M., & Choudhari, P. C. (2025) · International Conference on Electrical, Computer and Energy Technologies (ICECET) · doi:10.1109/icecet63943.2025.11471931
“The temperature measurement should be accurately obtained and transmitted with precision up to the first decimal place, as even minor deviations in decimal values can significantly impact the diagnosis and subsequent treatment plan [17].”
Bouclet, F., Pénichoux, J., Mihailescu, S., Seris, S., Burel, L., Richard, D., Collard, S., Jennet, E., Bertrand, E., Alani, M., et al. (2025) · Acta Haematologica · doi:10.1159/000544693
“Among them, a large meta-analysis of more than 8,000 patients [1] showed a limited accuracy of these methods especially among patients with fever or hypothermia, and a limited sensitivity for the detection of fever that was corroborated by Pecoraro et al. [14] and Bijur et al. [13].”
Not All Thermometers Are Created Equal
Phuong, L. K., & Foley, D. A. (2025) · Springer eBooks · doi:10.1007/978-981-96-7234-9_17
Samae, M., Kaewkhanitarak, H., Leeudomwong, T., et al. (2024) · Journal of Health Science and Medical Research · doi:10.31584/jhsmr.20241136
“However, there is a limited understanding of their accuracy and reproducibility in diagnosing patients with suspected fever 11 .”
Nepesov, M. İ., Kılınç, D., & Sezer, R. G. (2024) · Iranian Journal of Pediatrics · doi:10.5812/ijp-144744 · Cited by 1
Ehlers, U. E., Ulmer, J., Mirja, K., Carsten, K., & Pietsch, U. (2024) · Research Square · doi:10.21203/rs.3.rs-4554391/v1
Design Implementation of a Logger-Based Digital Thermometer
Gesa, N. F., Ahemen, I., Amah, A. N., et al. (2024) · European Modern Studies Journal · doi:10.59573/emsj.8(2).2024.31
Contact-Less Body Temperature Monitoring by Infrared Camera: Accuracy Preliminary Assessment
Leogrande, E., Botrugno, C., Natale, T., & Dell’Olio, F. (2023) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/sensors56945.2023.10325125
“Mercury, digital, and infrared thermometers are just a few of the tools that can be used to measure a person’s body temperature, and many certified devices for measuring this vital sign are available on the market [4].”
BackgroundTemperature Monitoring in Children: An Agreement Study
Patel, D. V., Barot, R., Cecil, R., Phatak, A. G., Shinde, M. K., Patel, A., & Nimbalkar, S. M. (2023) · Journal of Neonatology · doi:10.1177/09732179231164527 · Cited by 1
Alper, M. P. (2023) · Research Square · doi:10.21203/rs.3.rs-2761413/v1
IoT based sensor operated automatic potion (SOAP) dispenser
Garimella, R. C., Madeti, S. R. K., Neetoori, R., Gandreti, P., & Kumar, K. (2023) · Elsevier eBooks · doi:10.1016/b978-0-323-95876-9.00001-x
Hegge, H. H. M., & Franssen, R. (2023) · Bohn Stafleu van Loghum eBooks · doi:10.1007/978-90-368-2931-1_37
Risk Factors and Individual Protection Measures for COVID-19 in Federal Police Officers
Dias, J. A., Chagas, E. F. B., Detregiachi, C. R. P., Serva, F. M., Biteli, P., Mendes, C. G., Xavier, E. F. D. S., Jorge, C. F. B., Zutin, T. L. M., dos Santos, M. C., et al. (2022) · Hygiene · doi:10.3390/hygiene2040017 · Cited by 1
Li, R., Shen, M., Liu, H., et al. (2022) · doi:10.2196/preprints.42548
Fever in adults, differential diagnosis and administration of antipyretics
Krížek, B., & Matalová, P. (2022) · Medicína pro praxi · doi:10.36290/med.2022.002
Hegge, H. H. M., & Franssen, R. (2022) · Bohn Stafleu van Loghum eBooks · doi:10.1007/978-90-368-2787-4_37
Review of Medical Thermometry: From the Advent to Modern Applications
Borisov, I. V., Bondar, V. A., Kudinov, D. A., Kanarskii, M. M., Nekrasova, J. J., & Dmitriev, D. A. (2021) · Medicina · doi:10.29234/2308-9113-2021-9-3-75-90
Metadata
- Authors
- Pecoraro, V., Petri, D., Costantino, G., Squizzato, A., Moja, L., Virgili, G., & Lucenteforte, E.
- Affiliations
- University of Pisa (IT)
- University of Milan (IT)
- University of Insubria (IT)
- Azienda Ospedaliero-Universitaria Careggi (IT)
- Corresponding author
- Lucenteforte, Ersilia
- ORCID
- Venue
- Internal and Emergency Medicine
- Published
- 2020-11-25
- Volume
- 16
- Issue
- 4
- Pages
- 1071-1083
- ISSN
- 1828-0447 (ISSN-L)
- 1970-9366
- PMC
- PMC7686821
- OpenAlex
- W3107110570
- Semantic Scholar
- e6bfbad0712f698c2c504dda8d4670280ed1075b
- Cited by
- 49
- References
- 69
- Citations by year
- 2026: 9 · 2025: 10 · 2024: 9 · 2023: 6 · 2022: 12 · 2021: 3
- FWCI
- 0.77
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Internal and Emergency Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Florence Research (University of Florence) (submittedVersion, cc-by)
- Archivio Istituzionale della Ricerca (Universita Degli Studi Di Milano) (submittedVersion, other-oa)
- CINECA IRIS Institutial research information system (University of Pisa) (submittedVersion)
- IrInSubria (University of Insubria) (submittedVersion)
- Funders
- Fondazione Cassa si Risparmio di Firenze (24383)
- Università di Pisa
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
- Health Sciences › Medicine › Physiology (medical) › Thermoregulation and physiological responses
- Keywords
- body temperature measurement, fever diagnosis, infrared thermometry, tympanic thermometer, digital thermometer, mercury-in-glass thermometer, rectal thermometry, diagnostic accuracy, reproducibility, sensitivity and specificity
- MeSH terms
- Body Temperature, Diagnostic Tests, Routine, Equipment Design, Humans, Sensitivity and Specificity, Thermometers, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.