NMR measurement of LDL particle number using the Vantera® Clinical Analyzer

Open access
Matyus, Steven P., Braun, Paul J., Wolak-Dinsmore, Justyna, Jeyarajah, Elias J., Shalaurova, Irina, Xu, Yuan, Warner, Suzette M., Clement, Thomas S., Connelly, Margery A., Fischer, Timothy J. · 2014 · Clinical Biochemistry

Abstract

BACKGROUND: The Vantera Clinical Analyzer was developed to enable fully-automated, high-throughput nuclear magnetic resonance (NMR) spectroscopy measurements in a clinical laboratory setting. NMR-measured low-density lipoprotein particle number (LDL-P) has been shown to be more strongly associated with cardiovascular disease outcomes than LDL cholesterol (LDL-C) in individuals for whom these alternate measures of LDL are discordant. OBJECTIVE: The aim of this study was to assess the analytical performance of the LDL-P assay on the Vantera Clinical Analyzer as per Clinical Laboratory Standards Institute (CLSI) guidelines. RESULTS: Sensitivity and linearity were established within the range of 300-3500 nmol/L. For serum pools containing low, medium and high levels of LDL-P, the inter-assay, intra-assay precision and repeatability gave coefficients of variation (CVs) between 2.6 and 5.8%. The reference interval was determined to be 457-2282 nmol/L and the assay was compatible with multiple specimen collection tubes. Of 30 substances tested, only 2 exhibited the potential for assay interference. Moreover, the LDL-P results from samples run on two NMR platforms, Vantera Clinical Analyzer and NMR Profiler, showed excellent correlation (R(2)=0.96). CONCLUSIONS: The performance characteristics suggest that the LDL-P assay is suitable for routine testing in the clinical laboratory on the Vantera Clinical Analyzer, the first automated NMR platform that supports NMR-based clinical assays.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.clinbiochem.2014.07.015

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 168

Metadata

Authors
Matyus, S. P., Braun, P. J., Wolak-Dinsmore, J., Jeyarajah, E. J., Shalaurova, I., Xu, Y., Warner, S. M., Clement, T. S., Connelly, M. A., & Fischer, T. J.
Venue
Clinical Biochemistry
Publisher
Elsevier BV
Published
2014-11
Volume
47
Issue
16-17
Pages
203-210
ISSN
0009-9120 (ISSN-L)
OpenAlex
W2023797086
Cited by
168
References
7
FWCI
5.08
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review