Evaluation of a HbA1c point-of-care analyzer
Abstract
Objectives A better glycemic monitoring of diabetic patients and avoiding complications of poorly controlled diabetes could be possible with point-of-care testing technology (POCT) for HbA1c determination. B-Analyst® was studied to check whether it complied with the quality requirements for this purpose. Design and methods We evaluated the B-Analyst® (Menarini Diagnostics), which is based in the principle of latex agglutination immunoturbidimetry, to assess the validity of the technique of HbA1c. We carried out the method comparison with the HA-8180® (Menarini Diagnostics) as a reference method [High Performance Liquid Chromatography (HPLC)]. We assessed the analytical quality of the B-Analyst® studying the accuracy: inter-assay variability and intra-assay study. Furthermore, possible interferences by hemoglobinopathies were studied. Results Regression analysis of the data for the method comparison between HA-8180® and B-Analyst® showed a slope of 1.0085 and an intercept of 0.1208. The Pearson's correlation coefficient was 0.9958 (p<0.0001). Bias study showed a mean difference from B-Analyst® with respect to HA-8180® of 0.1872 with a 95% confidence interval. The standard error of the estimate (Syx) was 0.2091. The concordance correlation coefficient to assess accuracy was 0.9922 (0.9891-0.9945). The CV for the inter-assay study was 1.4%. For the intra-assay study we analyzed 3 samples with different HbA1c % whose CV were 1.03% [4.7% HbA1c (28 mmol/mol)], 0.46% [6.4% HbA1c (46 mmol/mol)] and 0.78% [8.1% HbA1c (65 mmol/mol)]. Conclusion The B-Analyst® evaluated not only showed good correlation with HA-8180®, but also it presented a great accuracy both in the inter-assay and in the intra-assay studies. The B-Analyst® complies with quality specifications required for monitoring of diabetic patients.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The systematic review and meta-analysis included 11 studies involving 1,593 participants and 3,285 paired HbA1c measurements comparing POC devices with reference HPLC-based laboratory methods.
“A total of 11 studies involving 1,593 participants and 3,285 paired measurements were included.”
From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [26]
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
Standards of Medical Care in Diabetes
Power, D. V. (2006) · Diabetes Care · doi:10.2337/diacare.29.02.06.dc05-1593 · Cited by 8,672 · see citing works
Turner, R. C., Fox, C., Matthews, McElroy, H. H., Cull, C. A., Holman, R. R., Neil, P., Hadden, D. R., Wright, D., Manley, E., et al. (1998) · The Lancet · Cited by 8,419 · see citing works
Standards of medical care in diabetes—2014
American Diabetes Association (2014) · Diabetes Care · doi:10.2337/dc14-s014 · Cited by 4,572 · see citing works
The UK Prospective Diabetes Study (UKPDS): clinical and therapeutic implications for type 2 diabetes
King, P., Peacock, I., & Donnelly, R. (1999) · British Journal of Clinical Pharmacology · doi:10.1046/j.1365-2125.1999.00092.x · Cited by 703 · see citing works
Approved IFCC reference method for the measurement of HbA1c in human blood
Jeppsson, J.-O., Kobold, U., Barr, J., Finke, A., Hoelzel, W., Hoshino, T., Miedema, K., Mosca, A., Mauri, P., Paroni, R., et al. (2002) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2002.016 · Cited by 655 · see citing works
HÆMOGLOBIN A1: AN INDICATOR OF THE METABOLIC CONTROL OF DIABETIC PATIENTS
Gonen, B., Rochman, H., Rubenstein, A. H., Tanega, S., & Horwitz, D. (1977) · The Lancet · doi:10.1016/s0140-6736(77)90237-9 · Cited by 343 · see citing works
Lenters-Westra, E., & Slingerland, R. J. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2009.130641 · Cited by 228 · see citing works
Nichols, J. H., Christenson, R. H., Clarke, W., Gronowski, A., Hammett-Stabler, C. A., Jacobs, E., Kazmierczak, S., Lewandrowski, K., Price, C., Sacks, D. B., et al. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2006.12.025 · Cited by 169 · see citing works
Lenters‐Westra, E., & Slingerland, R. J. (2014) · Clinical Chemistry · doi:10.1373/clinchem.2014.224311 · Cited by 130 · see citing works
The Diabetes Research in Children Network (DirecNet) Study Group (2005) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2005.00088.x · Cited by 121 · see citing works
Glycated hemoglobin as a marker of cardiovascular risk
Khaw, K., & Wareham, N. J. (2006) · Current Opinion in Lipidology · doi:10.1097/mol.0b013e3280106b95 · Cited by 110 · see citing works
Global Harmonization of Hemoglobin A1c
Sacks, D. B. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2004.047431 · Cited by 94 · see citing works
Nurse-based evaluation of point-of-care assays for glycated haemoglobin
John, A. S., Davis, T. M. E., Goodall, I., Townsend, M. A., & Price, C. P. (2005) · Clinica Chimica Acta · doi:10.1016/j.cca.2005.09.003 · Cited by 47 · see citing works
Eldor, R., & Raz, I. (2009) · The Review of Diabetic Studies · doi:10.1900/rds.2009.6.6 · Cited by 29 · see citing works
Heylen, O., Neyghem, S. V., Exterbille, S., Wehlou, C., Gorus, F., & Weets, I. (2014) · American Journal of Clinical Pathology · doi:10.1309/ajcpru5qc2jbansv · Cited by 22 · see citing works
Keramati, T., Razi, F., Tootee, A., & Larijani, B. A. (2014) · Journal of Diabetes & Metabolic Disorders · doi:10.1186/s40200-014-0094-1 · Cited by 17 · see citing works
Haliassos, A., Drakopoulos, I., Katritsis, D. G., Chiotinis, N., Korovesis, S., & Makris, K. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.041 · Cited by 17 · see citing works
Armonización de resultados de HbA1c en España
Goberna, R., Aguilar‐Diosdado, M., Santos-Rey, K., & Mateo, J. (2009) · Revista del Laboratorio Clínico · doi:10.1016/j.labcli.2008.11.002 · Cited by 5 · see citing works
Amez, J. M. F., Molina, E. A., Jiménez, N. O., Sancho, L. G., Gómez, M. A. G., & Valle, M. B. (2009) · Atención Primaria · doi:10.1016/j.aprim.2009.03.013 · Cited by 4 · see citing works
Flagged first, then by citations
1 of the 23 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Impact of HbA1c Testing at Point of Care on Diabetes Management
Schnell, O., Crocker, J. B., & Weng, J. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816678263 · Cited by 139 · see citing works
Hirst, J. A., McLellan, J. H., Price, C. P., English, E., Feakins, B. G., Stevens, R. J., & Farmer, A. J. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0303 · Cited by 102 · see citing works
Trends in Quantification of HbA1c Using Electrochemical and Point-of-Care Analyzers
Mandali, P. K., Prabakaran, A., Annadurai, K., & Krishnan, U. M. (2023) · Sensors · doi:10.3390/s23041901 · Cited by 16 · see citing works
Elliott, T. G., Dooley, K., Zhang, M., Campbell, H., & Thompson, D. J. S. (2017) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2017.10.053 · Cited by 10 · see citing works
Point of care testing in family practice: common myths debunked
Howick, J., Bossuyt, P. M., & Cals, J. (2016) · Family Practice · doi:10.1093/fampra/cmw082 · Cited by 9 · see citing works
Wei, Y., Gu, M., Ren, Y., & Hu, Z. (2026) · Biomedical Reports · doi:10.3892/br.2026.2120 · Cited by 3 · see citing works
Szybowska, P., Wilusz, M., Rogatko, I., & Sztefko, K. (2016) · Diagnostyka Laboratoryjna · doi:10.5604/01.3001.0009.3634
Prospecção Tecnológica e Científica de Testes Rápidos para Hemoglobina Glicada (HbA1c)
Farrapo, J. S. T., Gris, E. F., Nascimento, P. G. B. D., & Ferreira, E. A. (2022) · Cadernos de Prospecção · doi:10.9771/cp.v15i4.49263
- Authors
- Menéndez-Valladares, P., Fernández-Riejos, P., Sánchez-Mora, C., Pérez-Pérez, A., Sánchez-Margalet, V., & González-Rodríguez, C.
- Affiliations
- Hospital Universitario Virgen Macarena (ES)
- Corresponding author
- Menéndez-Valladares, Paloma
- Venue
- Clinical Biochemistry
- Publisher
- Elsevier BV
- Published
- 2015-07
- Volume
- 48
- Issue
- 10-11
- Pages
- 686-689
- ISSN
- 0009-9120 (ISSN-L)
- OpenAlex
- W1985278698
- Cited by
- 8
- References
- 23
- Citations by year
- 2023: 2 · 2022: 1 · 2017: 1 · 2016: 4
- FWCI
- 1.23
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Biochemistry (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Advanced Glycation End Products research
- Keywords
- HbA1c, point-of-care testing, method comparison, analytical accuracy, inter-assay variability, hemoglobinopathies, assay interference
- MeSH terms
- Chromatography, High Pressure Liquid, Diabetes Mellitus, Glycated Hemoglobin, Humans, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review