CoChat

Evaluation of three indirect calorimetry devices in mechanically ventilated patients: Which device compares best with the Deltatrac II®? A prospective observational study

Open access
Graf, Séverine, Karsegard, Véronique Laurie, Viatte, Valérie, Heidegger, Claudia Paula, Fleury, Yvan, Pichard, Claude, Genton, Laurence · 2015 · Clinical Nutrition
Cited by 88

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

BACKGROUND & AIMS Indirect calorimetry (IC) is the gold standard to measure energy expenditure (EE) in hospitalized patients. The popular 30 year-old Deltatrac II(®) (Datex) IC is no more commercialized, but other manufacturers have developed new devices. This study aims at comparing for the first time simultaneously, two new IC, the CCM express(®) (Medgraphics) and the Quark RMR(®) (Cosmed) with the Deltatrac II(®) to assess their potential use in intensive care unit (ICU) patients. METHODS ICU patients on mechanical ventilation, with positive end-expiratory pressure <9 cm H2O and fraction of inspired oxygen <60%, underwent measurements by the three IC simultaneously connected during 20 min to the ventilator (Evita XL(®), Dräger). Patients' characteristics, VO2 consumption, VCO2 production, respiratory quotient and EE were recorded. Data were presented as mean (SD) and compared by linear regression, repeated measure one-way ANOVA and Bland & Altman diagrams. RESULTS Forty patients (23 males, 60(17) yrs, BMI 25.4(7.0) kg/m(2)) were included. For the Deltatrac II(®), VO2 was 227(61) ml/min, VCO2 189(52) ml/min and EE 1562(412) kcal/d. VO2, VCO2, and EE differed significantly between Deltatrac II(®) and CCM express(®) (p < 0.001), but not between Deltatrac II(®) and Quark RMR(®). For EE, diagrams showed a mean difference (2SD) of 25.2(441) kcal between Deltatrac II(®) vs. the Quark RMR(®), and -273 (532) kcal between Deltatrac II(®) vs CCM express(®). CONCLUSION Quark RMR(®) compares better with Deltatrac II(®) than CCM express(®), but it suffers an EE variance of 441 kcal, which is not acceptable for clinical practice. New indirect IC should be further improved before recommending their clinical use in ICU.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.clnu.2014.01.008

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 7 references

Cited by

Metadata

Authors
Graf, S., Karsegard, V. L., Viatte, V., Heidegger, C. P., Fleury, Y., Pichard, C., & Genton, L.
Venue
Clinical Nutrition
Publisher
Elsevier BV
Published
2015-02
Volume
34
Issue
1
Pages
60-65
ISSN
0261-5614 (ISSN-L)
OpenAlex
W2025059392
Cited by
88
References
7
FWCI
8.65
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.