Consequences of delayed pump infusion line change in patients with type 1 diabetes mellitus treated with continuous subcutaneous insulin infusion

Open access
Thethi, Tina K., Rao, Ajay, Kawji, Haytham, Mallik, Tilak, Yau, C. Lillian, Christians, Uwe, Fonseca, Vivian·2010·Journal of Diabetes and its Complications

Abstract

Objective To systematically investigate the effect of lack of adherence to the recommended change in insulin pump infusion line use beyond 48 h and determine whether the type of insulin made a difference. Research design and methods This was a double-blind, randomized, crossover trial with 20 patients with diabetes mellitus I using insulins aspart and lispro without a line change for up to 100 h. Using retrospective continuous glucose monitoring, we analyzed the average glucose over the day. Changes in serum 1,5-anhydroglucitol, carboxymethyllysine, and free 15-F(2t) isoprostane were also studied. Results From Day 2 to Day 5 of the pump line use, the daily average glucose level increased from 122.7 to 163.9 mg/dl (P<.05), fasting glucose from 120.3 to 154.5 mg/dl (P<.05), postprandial glucose from 114.6 to 172.1 mg/dl (P<.05), and the daily maximum glucose from 207.7 to 242.8 dl (P<.05 for the trend). Time period that the glucose was >180 mg/dl increased from 14.5% to 38.3% (P<.05). Loss of control occurred despite increase in total daily insulin dose from 48.5+/-11.8 to 55.3+/-17.9 U (P=.05). There was no difference in loss of control between insulin types, and biomarkers measured did not change significantly. Conclusions The insulin pump infusion should be changed every 48 h in patients using continuous subcutaneous insulin infusion (CSII), to avoid loss of glycemic control. In the short-term, this loss of glycemic control has no impact on oxidative stress and glycation.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.jdiacomp.2009.03.002

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Longer IIS wear duration >3 days has been associated with higher number of catheter-related adverse events, including hyperglycemia, site reactions, and silent occlusions.

    “In addition, longer IIS wear duration >3 days has been associated with higher number of catheter-related adverse events, including hyperglycemia (blood glucose >250 mg/dL), site reactions, and ''silent occlusions.''7,12,17”

    From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [12]

    The Results and Conclusions (pages unavailable) report worsening glycemic control as infusion-line use continued beyond 48 hours, including more time with glucose above 180 mg/dL. They do not establish an increase in events defined by blood glucose above 250 mg/dL, site reactions, or silent occlusions with wear beyond three days. The Introduction mentions an infusion-site infection in an earlier case, but not an observed increase in site reactions in this study. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • Silent occlusions contribute to the occurrence of unexplained hyperglycemia and glycemic variability in patients using CSII.

    “Based on our findings as well as previous reports,15,16,18 we believe that silent occlusions contribute to the occurrence of unexplained hyperglycemia and glycemic variability in patients using CSII.”

    From Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies (2016) · cites this paper as [18]

    The Conclusions report that glycemic control deteriorated after 48 hours of pump infusion line use, with increases in several glucose measures, and discuss a possible effect on glucose variability. They do not identify silent occlusions as the cause or characterize the hyperglycemia as unexplained; the mention of catheter occlusions concerns a separate study. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Early feasibility and pilot IIS studies showed significant device-related adverse events that included increased hyperglycemia and functional or mechanistic issues with IIS use during longer durations of wear.

    “While early feasibility and pilot IIS studies showed significant device-related adverse events that included increased hyperglycemia and functional or mechanistic issues with IIS use during longer durations of wear,1–3”

    From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [3]

    The Results report significantly increased glucose levels and time spent above 180 mg/dL as pump infusion lines were worn from day 2 to day 5. The paper does not report significant mechanical problems with the infusion sets during its own trial; its Conclusions mention only a nonsignificant trend toward fewer catheter occlusions in another study. Thus it supports the prolonged-wear hyperglycemia finding, but not the full description of significant device-related events. (LLM, conf=0.85)

    Experimental — accuracy not yet measured
  • Early studies of extended standard IIS wear demonstrated deterioration in glycemic control and significant IIS- or IIS site-related adverse events.

    “Early studies of extended standard IIS wear demonstrated deterioration in glycemic control and significant IIS- or IIS site-related adverse events.1–3”

    From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [3]

    The Conclusions (page unavailable) report that glycemic control deteriorated after 48 hours of pump infusion line use. The Introduction (page unavailable) describes a serious infusion-site infection after prolonged wear, but attributes it to an earlier case report; this study does not demonstrate significant IIS- or site-related adverse events of its own. (LLM, conf=0.85)

    Experimental — accuracy not yet measured
  • Standard insulin infusion sets should be changed every 2 to 3 days to prevent or reduce adverse events.

    “Standard IISs should be changed every 2 to 3 days,1–3 regardless of rapid-acting insulin used,4 to prevent or reduce adverse events (e.g., site failure resulting in hyperglycemia, skin inflammation, infection or pain at the IIS site, adhesive patch failure, etc.).”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 31 · Cited by 53

Research works printed but not linked: 27

  1. [1]The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus The Diabetes Control and Complications Trial Research Group 1 N Engl J Med 1993 Sep 30 329 14 977 86 8366922
  2. [2]Hanaire-Broutin H Melki V Bessieres-Lacombe S Tauber JP Comparison of continuous subcutaneous insulin infusion and multiple daily injection regimens using insulin lispro in type 1 diabetic patients on intensified treatment: a randomized study. The Study Group for the Development of Pump Therapy in Diabetes 1 Diabetes Care 2000 Sep 23 9 1232 5 10977011
  3. [3]Hirsch IB Bode BW Garg S Lane WS Sussman A Hu P Continuous subcutaneous insulin infusion (CSII) of insulin aspart versus multiple daily injection of insulin aspart/insulin glargine in type 1 diabetic patients previously treated with CSII 1 Diabetes Care 2005 Mar 28 3 533 8 15735183
  4. [4]Toxic-Shock Syndrome in a Patient Using a Continuous Subcutaneous Insulin Infusion Pump -- Idaho MMWR 1983 August 1232 31 404 406 6410172
  5. [5]Mecklenburg RS Benson EA Benson JW Jr Fredlund PN Guinn T Metz RJ Acute complications associated with insulin infusion pump therapy. Report of experience with 161 patients 1 JAMA 1984 Dec 21 252 23 3265 9 6439896
  6. [6]Pietri A Raskin P Cutaneous complications of chronic continuous subcutaneous insulin infusion therapy 1 Diabetes Care 1981 Nov 4 6 624 6 6751739
  7. [8]Senstius J Poulsen C Hvass A Comparison of in vitro stability for insulin aspart and insulin glulisine during simulated use in insulin pumps 1 Diabetes Technol Ther 2007 Dec 9 6 517 21 18034606
  8. [9]Haschke M Zhang YL Kahle C Klawitter J Korecka M Shaw LM HPLC-atmospheric pressure chemical ionization MS/MS for quantification of 15-F2t-isoprostane in human urine and plasma 1 Clin Chem 2007 Mar 53 3 489 97 17259231
  9. [10]DeVries JH Snoek FJ Kostense PJ Masurel N Heine RJ A randomized trial of continuous subcutaneous insulin infusion and intensive injection therapy in type 1 diabetes for patients with long-standing poor glycemic control 1 Diabetes Care 2002 Nov 25 11 2074 80 12401759
  10. [11]Lenhard MJ Reeves GD Continuous subcutaneous insulin infusion: a comprehensive review of insulin pump therapy 1 Arch Intern Med 2001 Oct 22 161 19 2293 300 11606144
  11. [12]Pickup J Keen H Continuous subcutaneous insulin infusion at 25 years: evidence base for the expanding use of insulin pump therapy in type 1 diabetes Diabetes Care 2002 Mar 25 3 593 8 11874953
  12. [13]Tamborlane WV Bonfig W Boland E Recent advances in treatment of youth with Type 1 diabetes: better care through technology 1 Diabet Med 2001 Nov 18 11 864 70 11703429
  13. [14]Rosilio M Cotton JB Wieliczko MC Gendrault B Carel JC Couvaras O Factors associated with glycemic control. A cross-sectional nationwide study in 2,579 French children with type 1 diabetes. The French Pediatric Diabetes Group 1 Diabetes Care 1998 Jul 21 7 1146 53 9653610
  14. [15]Hoogma RP Schumicki D Safety of insulin glulisine when given by continuous subcutaneous infusion using an external pump in patients with type 1 diabetes 1 Horm Metab Res 2006 Jun 38 6 429 33 16823727
  15. [16]Burdick J Chase HP Slover RH Knievel K Scrimgeour L Maniatis AK Missed insulin meal boluses and elevated hemoglobin A1c levels in children receiving insulin pump therapy 1 Pediatrics 2004 Mar 113 3 Pt 1 e221 e224 14993580
  16. [17]Monnier L Mas E Ginet C Michel F Villon L Cristol JP Activation of oxidative stress by acute glucose fluctuations compared with sustained chronic hyperglycemia in patients with type 2 diabetes 1 JAMA 2006 Apr 12 295 14 1681 7 16609090
  17. [18]Nicolls MR Haskins K Flores SC Oxidant stress, immune dysregulation, and vascular function in type I diabetes 2 Antioxid Redox Signal 2007 Jul 9 7 879 89 17508913
  18. [19]Reynolds LR Reemergence of insulin pump therapy in the 1990s 1 South Med J 2000 Dec 93 12 1157 61 11142448
  19. [20]Bischof F Meyerhoff C Pfeiffer EF Quality control of intensified insulin therapy: HbA1 versus blood glucose 1 Horm Metab Res 1994 Dec 26 12 574 8 7705761
  20. [21]Lauritzen T Pramming S Deckert T Binder C Pharmacokinetics of continuous subcutaneous insulin infusion 1 Diabetologia 1983 May 24 5 326 9 6347780
  21. [22]Gerich JE Clinical significance, pathogenesis, and management of postprandial hyperglycemia 1 Arch Intern Med 2003 Jun 9 163 11 1306 16 12796066
  22. [23]Pickup JC Kidd J Burmiston S Yemane N Determinants of glycaemic control in type 1 diabetes during intensified therapy with multiple daily insulin injections or continuous subcutaneous insulin infusion: importance of blood glucose variability 1 Diabetes Metab Res Rev 2006 May 22 3 232 7 16389648
  23. [25]van Deutekom AW Niessen HW Schalkwijk CG Heine RJ Simsek S Increased Nepsilon-(carboxymethyl)-lysine levels in cerebral blood vessels of diabetic patients and in a (streptozotocin-treated) rat model of diabetes mellitus 1 Eur J Endocrinol 2008 May 158 5 655 60 18426823
  24. [26]Miura J Yamagishi S Uchigata Y Takeuchi M Yamamoto H Makita Z Serum levels of non-carboxymethyllysine advanced glycation endproducts are correlated to severity of microvascular complications in patients with Type 1 diabetes 1 J Diabetes Complications 2003 Jan 17 1 16 21 12505751
  25. [27]Roberts LJ Morrow JD Measurement of F(2)-isoprostanes as an index of oxidative stress in vivo 1 Free Radic Biol Med 2000 Feb 15 28 4 505 13 10719231
  26. [28]Flores L Rodela S Abian J Claria J Esmatjes E F2 isoprostane is already increased at the onset of type 1 diabetes mellitus: effect of glycemic control 1 Metabolism 2004 Sep 53 9 1118 20 15334370
  27. [29]Hammer S Jonker JT Lamb HJ van der Meer RW Zondag W Sepers JM Short-term Hyperglycemic Dysregulation in Patients with Type 1 Diabetes Does Not Change Myocardial Triglyceride Content or Myocardial Function 1 Diabetes Care 2008 May 5 PMC2494624 18458137

Authors
Thethi, T. K., Rao, A., Kawji, H., Mallik, T., Yau, C. L., Christians, U., & Fonseca, V.
Affiliations
  • Tulane University (US)
  • University of Colorado Denver (US)
Corresponding author
Thethi, Tina K.
Venue
Journal of Diabetes and its Complications
Publisher
Elsevier BV
Published
2010-03
Volume
24
Issue
2
Pages
73-78
ISSN
  • 0891-6632 (ISSN-L)
  • 1056-8727
OpenAlex
W2038371745
Cited by
53
References
31
Citations by year
2025: 2 · 2024: 2 · 2023: 10 · 2022: 5 · 2021: 3 · 2020: 3 · 2019: 2 · 2018: 4 · 2017: 1 · 2016: 4 · 2015: 7 · 2014: 2 · 2013: 4 · 2012: 3
FWCI
0.97
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes and its Complications (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion)
Funders
  • National Heart, Lung, and Blood Institute (T32HL007609)
  • National Institute of Diabetes and Digestive and Kidney Diseases (P30DK048520)
  • National Center for Research Resources (M01RR005096, M01RR000827)
  • Eunice Kennedy Shriver National Institute of Child Health and Human Development (K12HD043451)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
Keywords
type 1 diabetes, insulin pump therapy, insulin aspart, insulin lispro, glycemic control, hyperglycemia, oxidative stress, glycation, 1,5-anhydroglucitol, carboxymethyllysine, isoprostanes, continuous glucose monitoring
MeSH terms
Adolescent, Adult, Aged, Blood Glucose, Deoxyglucose, Diabetes Mellitus, Type 1, Double-Blind Method, Fasting, Female, Humans, Hyperglycemia, Insulin, Insulin Infusion Systems, Lysine, Male, Middle Aged, Cross-Over Studies, F2-Isoprostanes, Infusions, Subcutaneous, Young Adult, Insulin Lispro

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review