Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial
Open accessAbstract
Background: Standard insulin infusion sets (IISs) are to be replaced every 2 to 3 days to avoid complications and diabetic ketosis due to set failure. This pivotal trial evaluated the safety and performance of a new extended-wear infusion set (EIS) when used for 7 days by adults with type 1 diabetes (T1D). Methods: This single-arm, nonrandomized trial enrolled adults (18–80 years of age) with T1D, who used their own MiniMed™ 670G system with insulin lispro or insulin aspart and the EIS for up to 7 days, across 12 consecutive wears. Safety endpoints included incidence of serious adverse events (SAEs), serious adverse device effects (SADEs), unanticipated adverse device effects (UADEs), severe hypoglycemia (SevHypo), severe hyperglycemia (SevHyper), diabetic ketoacidosis (DKA), and skin infection. The EIS failure rate due to unexplained hyperglycemia (i.e., suspected occlusion), the overall EIS survival rate, glycemic control outcomes (i.e., A1C, mean sensor glucose and time spent in established glucose ranges), total daily insulin delivered, and satisfaction with the EIS were determined. Results: The intention to treat population ( n = 259, 48% men, 45.0 ± 14.1 years) wore a total of 3041 EIS devices. No SADE, UADE, or DKA events was reported. Overall rates of SAEs, SevHypo, SevHyper, and skin infection were 3.8, 2.5, 104.1, and 20.1 events per 100 participant-years. The rate of EIS failure due to unexplained hyperglycemia at the end of day 7 was 0.1% (95% confidence interval [CI]: 0.03–0.51) and 0.4% (95% CI: 0.16–1.00) for insulin lispro and aspart use, respectively. Overall EIS survival rate at the end of day 7 was 77.8% (95% CI: 76.2–79.3), glycemic control did not change, and participants reported greater satisfaction with the EIS compared with standard IISs worn before the study ( P < 0.001). Conclusions: This investigation demonstrates that the EIS, when worn for up to 7 days, was safe and rated with high satisfaction, without adversely affecting glycemic control in adults with T1D. Clinical Trial Registration number: NCT04113694 ( https://clinicaltrials.gov/ct2/show/NCT04113694 ).
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
In a single-arm pivotal trial, a new Medtronic extended-wear insulin infusion set (EIS) was worn up to 7 days by adults with type 1 diabetes using MiniMed 670G systems, with no serious adverse device effects, unanticipated adverse device effects, or DKA. The EIS had a 77.8% overall survival rate at day 7, low failure due to unexplained hyperglycemia, stable glycemic control, and higher user satisfaction than standard infusion sets.
Key findings
- Among 259 intention-to-treat adults (45.0 ± 14.1 years; 48% men) wearing 3041 EIS devices, there were no SADE, UADE, or DKA events; overall rates of SAEs, severe hypoglycemia, severe hyperglycemia, and skin infection were 3.8, 2.5, 104.1, and 20.1 events per 100 participant-years, respectively.
- EIS failure due to unexplained hyperglycemia at end of day 7 was 0.1% (95% CI 0.03–0.51) with insulin lispro and 0.4% (95% CI 0.16–1.00) with insulin aspart; overall EIS survival at end of day 7 was 77.8% (95% CI 76.2–79.3).
- Glycemic control did not change: A1C went from 7.2% ± 0.6% to 7.1% ± 0.6% (Δ −0.1 ± 0.4), and mean sensor glucose, coefficient of variation, time in glucose ranges, and total daily insulin delivered remained stable across 7 days of wear.
- Participants reported significantly greater satisfaction with the EIS than with standard IISs worn before the study across ease of insertion, comfort, duration, time to change, and convenience (all P < 0.001).
Methods
- Prospective, single-arm, nonrandomized trial at 15 U.S. investigational centers enrolling adults 18–80 years with type 1 diabetes who used their own MiniMed 670G system with insulin lispro or insulin aspart and wore the EIS for up to 7 days across 12 consecutive wears.
- The intention-to-treat population was n = 259; 3041 EIS devices were worn (1561 with insulin lispro, 1480 with insulin aspart), and 248 participants completed the study.
- Endpoints included safety events, EIS failure due to unexplained hyperglycemia, overall EIS survival excluding not-for-cause removals, glycemic outcomes (A1C, sensor glucose, time in ranges, total daily insulin), and satisfaction; satisfaction scores were compared with a Wilcoxon signed-rank test.
Limitations
- The paper states the study was open-label and nonrandomized without a comparator, which limits causal interpretation.
- The paper states participants were required to use Auto Mode on the MiniMed 670G, which may limit generalizability of glycemic findings to open-loop pump users.
- The paper states the population was majority white and had relatively well-controlled glycemia at baseline; the text also makes evident that the single-arm design and lack of a concurrent control group constrain comparisons with standard infusion sets.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
24 of 53 cited sources checked · 15 supported · 1 flagged · accuracy not yet measured
- Findingown data
The overall EIS survival rate by the end of day 7 was greater than previously reported survival rates for standard IIS devices worn for similar durations (69%, 63%, 43%, 33%, and 32%).
“The overall EIS survival rate of 77.8% (95% CI: 76.2–79.3) by the end of day 7 was greater than previously reported survival rates of 69% (n = 40 total participants and 80 total steel cannula sets),9 63% (n = 40 total participants, 80 total soft cannula sets),9 43% (n = 20 total participants, 40 total sets in non-lipohypertrophied tissue),18 33% (n = 20 total participants and 40 total sets in lipo”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
- cites [18] Karlin AW, Ly TT, Pyle L, et al.: Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther 2016; 18: 429– 435. 27227290 PMC4931738 source
- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
The citation is attached to the claimed 69% steel and 63% soft-cannula survival rates. The paper’s Results instead report that, at the end of day 7, 32% of steel sets and 33% of Teflon sets were functioning (page unavailable). (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
- Premisecitation
Prior rates of IIS failure due to unexplained hyperglycemia ranged from 19% to 30%.
“rates of IIS failure due to “unexplained hyperglycemia” (i.e., hyperglycemia >250 mg/dL and a failed correction bolus due to suspected IIS occlusion) ranged from a relatively high 19%9 to 30%.8”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
- Premisecitation
Early feasibility and pilot IIS studies showed significant device-related adverse events that included increased hyperglycemia and functional or mechanistic issues with IIS use during longer durations of wear.
“While early feasibility and pilot IIS studies showed significant device-related adverse events that included increased hyperglycemia and functional or mechanistic issues with IIS use during longer durations of wear,1–3”
Quote in the paperSubstantial claim3 of 3 sources checked- cites [1] Pfutzner A, Sachsenheimer D, Grenningloh M, et al.: Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: results from a prospective randomized crossover study. J Diabetes Sci Technol 2015; 9: 1292– 1298. 26341262 PMC4667306 source
- cites [2] Schmid V, Hohberg C, Borchert M, et al.: Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3. J Diabetes Sci Technol 2010; 4: 976– 982. 20663464 PMC2909532 source
- cites [3] Thethi TK, Rao A, Kawji H, et al.: Consequences of delayed pump infusion line change in patients with type 1 diabetes mellitus treated with continuous subcutaneous insulin infusion. J Diabetes Complications 2010; 24: 73– 78. 19395280 PMC2854665 source
The Results report significantly increased glucose levels and time spent above 180 mg/dL as pump infusion lines were worn from day 2 to day 5. The paper does not report significant mechanical problems with the infusion sets during its own trial; its Conclusions mention only a nonsignificant trend toward fewer catheter occlusions in another study. Thus it supports the prolonged-wear hyperglycemia finding, but not the full description of significant device-related events. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
Few randomized studies have reported failure and survival rates when IIS devices are worn out to 7 days.
“few randomized studies have reported failure and survival rates when IIS devices are worn out to 7 days.8,9”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
- Premisecitation
Schmid et al.'s pilot study reported average BG levels that increased from 135.0 ± 68.4 mg/dL to 208.8 ± 39.6 mg/dL from day 1 to 7 of IIS wear.
“The Schmid et al.'s pilot study of 12 adults (40.3 ± 12.6 years of age) using the MiniMed Comfort infusion set or MiniMed Silhouette infusion set (ConvaTec Unomedical A/S, Lejre, Denmark) reported average BG levels that increased from 135.0 ± 68.4 mg/dL to 208.8 ± 39.6 mg/dL from day 1 to 7 of IIS wear.2”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [2] Schmid V, Hohberg C, Borchert M, et al.: Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3. J Diabetes Sci Technol 2010; 4: 976– 982. 20663464 PMC2909532 source
- Findingown data
The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was substantially lower than the 19%, 23%, 30%, and 35% reported to be attributable or partly attributable to unexplained hyperglycemia in studies investigating extended standard IIS wear.
“The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was quite low (≤2%) and substantially lower than the 19%,9 23%,18 30%,8 and 35%18 reported to be attributable or partly attributable to unexplained hyperglycemia in the studies investigating extended standard IIS wear.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
- cites [18] Karlin AW, Ly TT, Pyle L, et al.: Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther 2016; 18: 429– 435. 27227290 PMC4931738 source
- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
- Findingown data
Overall EIS survival rate at the end of day 7 was 77.8% (95% CI: 76.2-79.3).
“Overall EIS survival rate at the end of day 7 was 77.8% (95% CI: 76.2–79.3)”
Quote in the paperSubstantial claim - Findingown data
There were three serious adverse events not related to device or study procedure.
“There were three SAEs not related to device or study procedure.”
Quote in the paperNot a substantial claim - Findingown data
At the end of day 6, the EIS failure rate due to unexplained hyperglycemia was 0.3% for insulin aspart.
“Insulin aspart | 127 | 1480 | 4 | 0.3”
Quote in the paperNot a substantial claim - Findingown data
Participants reported greater satisfaction with the EIS compared with standard IISs worn before the study (P < 0.001).
“participants reported greater satisfaction with the EIS compared with standard IISs worn before the study (P < 0.001).”
Quote in the paperNot a substantial claim - Findingown data
An exploratory analysis of glycemic control and TDD across days of EIS wear showed that mean SG, CV, and the percentage of time spent at different glucose ranges aligned with appropriate glycemic control over time.
“An exploratory analysis of glycemic control and TDD across days of EIS wear showed that mean SG, CV, and the percentage of time spent at different glucose ranges aligned with appropriate glycemic control, over time (Table 4).”
Quote in the paperSubstantial claim - Findingown data
The rate of SevHyper (BG >250 mg/dL with blood ketones ≥0.6 mM) was 104.1 events per 100 participant-years.
“The rate of SevHyper (BG >250 mg/dL with blood ketones ≥0.6 mM) was 104.1 events per 100 participant-years.”
Quote in the paperNot a substantial claim - Findingnone given
There were two SevHypo events that were not related to the EIS.
“There were two SevHypo events that were not related to the EIS.”
Quote in the paperNot a substantial claim - Findingown data
Participant-reported satisfaction was significantly greater with EIS than with standard IIS used before study start (P < 0.001).
“The satisfaction rates based on the responses captured for each query are shown in Figure 4 and indicate significantly greater satisfaction with EIS versus standard IIS use (P < 0.001).”
- Also stated in: § Discussion · plain
“The satisfaction scores regarding device convenience and use and the proportion of participants responding with “very satisfied” and/or “extremely satisfied” were higher for the EIS, when compared with standard IIS devices used before study start.”
Quote in the paperSubstantial claim - Also stated in: § Discussion · plain
- Findingnone given
Glycemic control did not change during the study.
“glycemic control did not change”
Quote in the paperSubstantial claim - Findingown data
At the end of day 6, the EIS failure rate due to unexplained hyperglycemia was 0.06% for insulin lispro.
“Insulin lispro | 132 | 1561 | 1 | 0.06”
Quote in the paperNot a substantial claim - Findingown data
Overall EIS survival rate at the end of day 6 was 83.8% (95% CI: 82.4-85.1).
“On day 6, the overall rate was 83.8% (95% confidence interval [CI]: 82.4–85.1); on day 7, it was 77.8% (95% CI: 76.2–79.3).”
Quote in the paperSubstantial claim - Findingown data
The amount of TDD delivered remained stable across the 7 days of EIS wear.
“In addition, the amount of TDD delivered remained stable across the 7 days of EIS wear.”
- Also stated in: § Discussion · hedged
“TDD was relatively unchanged (54.3 ± 28.6 U, 53.5 ± 28.5 U, and 53.6 ± 28.0 U on days 1, 4, and 7, respectively).”
Quote in the paperSubstantial claim - Also stated in: § Discussion · hedged
- Findingown data
When the EIS was used with either insulin lispro or insulin aspart, A1C changed from 7.2% ± 0.6% to 7.1% ± 0.6% (Δ of −0.1 ± 0.4).
“When the EIS was used with either insulin lispro or insulin aspart, A1C changed from 7.2% ± 0.6% to 7.1% ± 0.6% (Δ of −0.1 ± 0.4).”
Quote in the paperSubstantial claim - Findingown data
The rate of EIS failure due to unexplained hyperglycemia at the end of day 7 was 0.1% (95% CI: 0.03-0.51) for insulin lispro use.
“The rate of EIS failure due to unexplained hyperglycemia at the end of day 7 was 0.1% (95% confidence interval [CI]: 0.03–0.51) and 0.4% (95% CI: 0.16–1.00) for insulin lispro and aspart use, respectively.”
Quote in the paperSubstantial claim - Findingown data
Using the definition of SevHyper in the MiniMed 670G trial and based on an analysis of similarly aged adults (≥18 years), this study had 41.4 events per 100 participant-years, which was similar to the 42.7 events per 100 participant-years in the MiniMed 670G system trial.
“Using the definition of SevHyper in that study and based on an analysis of similarly aged adults (≥18 years), this study had 41.4 events per 100 participant-years, which was similar to the 42.7 events per 100 participant-years in the MiniMed 670G system trial.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15] Bergenstal RM, Garg S, Weinzimer SA, et al.: Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA 2016; 316: 1407– 1408. 27629148 source
- Findingown data
The infusion site skin infection rate in this study was lower than the self-reported rates of 17.4% and 28% by individuals who completed survey reporting on a prior year of standard IIS use.
“This rate was lower than the self-reported rates of 17.4%16 and 28%17 by individuals who completed survey reporting on a prior year of standard IIS use.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [16] Pickup JC, Yemane N, Brackenridge A, Pender S: Nonmetabolic complications of continuous subcutaneous insulin infusion: a patient survey. Diabetes Technol Ther 2014; 16: 145– 149. 24180294 PMC3934434 source
- cites [17] Taleb N, Messier V, Ott-Braschi S, et al.: Perceptions and experiences of adult patients with type 1 diabetes using continuous subcutaneous insulin infusion therapy: results of an online survey. Diabetes Res Clin Pract 2018; 144: 42– 50. 30077691 source
- Findingown data
The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was ≤2%.
“The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was quite low (≤2%) and substantially lower than the 19%,9 23%,18 30%,8 and 35%18 reported to be attributable or partly attributable to unexplained hyperglycemia in the studies investigating extended standard IIS wear.”
Quote in the paperSubstantial claim - Findingown data
The overall EIS survival rate by the end of day 7 was 77.8% (95% CI: 76.2-79.3).
“The overall EIS survival rate of 77.8% (95% CI: 76.2–79.3) by the end of day 7 was greater than previously reported survival rates of 69% (n = 40 total participants and 80 total steel cannula sets),9 63% (n = 40 total participants, 80 total soft cannula sets),9 43% (n = 20 total participants, 40 total sets in non-lipohypertrophied tissue),18 33% (n = 20 total participants and 40 total sets in lipo”
Quote in the paperSubstantial claim - Findingown data
Glucose metrics including mean SG and the percentage of time spent at SG ranges indicated glycemic stability for the 7-day wear duration.
“Glucose metrics that included mean SG and the percentage of time spent at SG ranges indicated glycemic stability for the 7-day wear duration.”
Quote in the paperSubstantial claim - Findingown data
There were 16 infusion site skin infections reported by 15 participants (5.8%), corresponding to 20.1 events per 100 participant-years.
“In this study, there were 16 infusion site skin infections reported by 15 participants (5.8%) that corresponded with 20.1 events per 100 participant-years.”
Quote in the paperNot a substantial claim - Premisecitation
Insulin degradation and preservative loss increased the inflammatory response.
“based on observations that insulin degradation and preservative loss increased the inflammatory response10”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [10] Chattaraj S, Zhang G, Anselmo E, Fusselman JC: Study of insulin stability impact on pump therapy: test model development. Diabetes 2020; 69: 1012- P. source
- Premisecitation
The new EIS includes a new extended-wear adhesive patch that improves skin adherence.
“in addition to a new extended-wear adhesive patch that improves skin adherence.13”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13] Zhang G, Chattaraj S, Anselmo E, et al.: Assessment of adhesive patches for an extended-wear infusion set. Diabetes 2020; 69: 986- P. source
- Premisecitation
Insulin-associated inflammation contributed to standard IIS failure.
“and insulin-associated inflammation contributed to standard IIS failure.11”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [11] Chattaraj S, Zhang G, Anselmo E, et al.: The Medtronic extended wear infusion set: determining mechanisms of action. Diabetes Technol Ther 2021; 23: A149.
- Interpretationreasoning
The requirement that participants use the Auto Mode function of their MiniMed 670G system throughout the study may limit the generalizability of glycemic control findings to populations of individuals with T1D using open-loop insulin pump systems.
“The trial also required that study participants use the Auto Mode function of their MiniMed 670G system throughout the course of the study, which may limit the generalizability of glycemic control findings to populations of individuals with T1D using open-loop insulin pump systems.”
Quote in the paperNot a substantial claim - Study design and conductnone given
A significant limitation of this study is the open-label and nonrandomized design without a comparator.
“A significant limitation of this study is the open-label and nonrandomized design without a comparator.”
Quote in the paperSubstantial claim
- Probably not a substantial claim. 19
- It may state two separate claims as one. 29
- The quote only partly states this claim. 2
Probably not a substantial claim.
PremisecitationStandard insulin infusion sets should be changed every 2 to 3 days to prevent or reduce adverse events.
“Standard IISs should be changed every 2 to 3 days,1–3 regardless of rapid-acting insulin used,4 to prevent or reduce adverse events (e.g., site failure resulting in hyperglycemia, skin inflammation, infection or pain at the IIS site, adhesive patch failure, etc.).”
Quote in the paperSubstantial claim4 of 4 sources checked- cites [1] Pfutzner A, Sachsenheimer D, Grenningloh M, et al.: Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: results from a prospective randomized crossover study. J Diabetes Sci Technol 2015; 9: 1292– 1298. 26341262 PMC4667306 source
- cites [2] Schmid V, Hohberg C, Borchert M, et al.: Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3. J Diabetes Sci Technol 2010; 4: 976– 982. 20663464 PMC2909532 source
- cites [3] Thethi TK, Rao A, Kawji H, et al.: Consequences of delayed pump infusion line change in patients with type 1 diabetes mellitus treated with continuous subcutaneous insulin infusion. J Diabetes Complications 2010; 24: 73– 78. 19395280 PMC2854665 source
- cites [4] Kerr D, Morton J, Whately-Smith C, et al.: Laboratory-based non-clinical comparison of occlusion rates using three rapid-acting insulin analogs in continuous subcutaneous insulin infusion catheters using low flow rates. J Diabetes Sci Technol 2008; 2: 450– 455. 19885209 PMC2769745 source
Probably not a substantial claim.
PremisecitationPrior overall findings determined that 7-day wear, irrespective of steel or Teflon cannula, did not adversely influence glycemic control.
“Overall findings from the latter determined that 7-day wear (irrespective of steel or Teflon cannula) did not adversely influence glycemic control”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
The Abstract reports no increase in hyperglycemia or daily insulin requirements when a set was successfully used for seven days. The Discussion makes clear that this analysis excluded sets that failed; the unqualified claim overlooks those failures, including failures associated with hyperglycemia. The finding therefore does not apply to seven-day wear overall. (LLM, conf=0.90)
Experimental — accuracy not yet measured - cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
It may state two separate claims as one.
PremisecitationWhile early standard IIS device studies confirmed manufacturer indications for 2- to 3-day use due to increased risk of IIS site reactions or compromised glycemic control, more recent studies with greater numbers of study participants have reported minimal or no issues with glycemia over time or fewer adverse events associated with extended IIS wear.
“Thus, while early standard IIS device studies confirmed manufacturer indications for 2- to 3-day use due to increased risk of IIS site reactions or compromised glycemic control, more recent studies with greater numbers of study participants have reported minimal or no issues with glycemia over time or fewer adverse events associated with extended IIS wear.8,9,18,19”
Quote in the paperNot a substantial claim1 of 4 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
- cites [18] Karlin AW, Ly TT, Pyle L, et al.: Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther 2016; 18: 429– 435. 27227290 PMC4931738 source
- cites [19] Freckmann G, Arndt S, Fiesselmann A, et al.: Randomized cross-over study comparing two infusion sets for CSII in daily life. J Diabetes Sci Technol 2017; 11: 253– 259. 27605591 PMC5478026 source
Probably not a substantial claim. The paper’s conclusions may not depend on it.
InterpretationcitationThe variability in reporting may, in part, be due to the variability in clinical infusion set failures, where some are partial and/or transient and can be seen frequently when monitoring infusion pressures that do not result in alarm (i.e., silent occlusions).
“The variability in reporting may, in part, be due to the variability in clinical infusion set failures, where some are partial and/or transient and can be seen frequently when monitoring infusion pressures that do not result in alarm (i.e., silent occlusions).20”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [20] Gibney M, Xue Z, Swinney M, et al.: Reduced silent occlusions with a novel catheter infusion set (BD FlowSmart): results from two open-label comparative studies. Diabetes Technol Ther 2016; 18: 136– 143. 26701357 PMC4790215 source
Split into 2 atoms — 1 confirmed, 1 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Probably not a substantial claim. The paper’s conclusions may not depend on it.
PremisecitationThere were multiple definitions of a failed correction dose, with 1 h or 60-90 min as the elapsed time by when a single correction dose was required to decrease BG by 50 mg/dL.
“There were also multiple definitions of a failed correction dose where 1 h8,18 or 60–90 min9 was the elapsed time by when a single correction dose was required to decrease BG by 50 mg/dL.”
Quote in the paperNot a substantial claim1 of 3 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
- cites [18] Karlin AW, Ly TT, Pyle L, et al.: Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther 2016; 18: 429– 435. 27227290 PMC4931738 source
- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
It may state two separate claims as one.
PremisecitationThe follow-up 3-month randomized crossover study reported a trending increase in IIS-related issues including hyperglycemia, increased frequency of IIS changes (from 73 to 143 changes), increased number of infusion set-related adverse events (from 305 to 517 events), and site reactions (from 1.7 ± 2.2 to 7.5 ± 14.0 events/participant), all of which were significant.
“The follow-up 3-month randomized crossover study1 with 22 adults (39.0 ± 11.0 years of age) using the MiniMed Mio infusion set or Inset II™ infusion set (ConvaTec Unomedical A/S) also reported a trending increase in IIS-related issues that included hyperglycemia, increased frequency of IIS changes (from 73 to 143 changes), increased number of infusion set-related adverse events (from 305 to 517”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [1] Pfutzner A, Sachsenheimer D, Grenningloh M, et al.: Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: results from a prospective randomized crossover study. J Diabetes Sci Technol 2015; 9: 1292– 1298. 26341262 PMC4667306 source
It may state two separate claims as one.
PremisecitationIn Schmid et al.'s study, events resulting in IIS change occurred more frequently after day 3 and impaired insulin absorption resulting in hyperglycemia was significant by day 7.
“While only a few study participants reached 7-day IIS wear, events resulting in IIS change occurred more frequently after day 3 and impaired insulin absorption resulting in hyperglycemia was significant by day 7.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [2] Schmid V, Hohberg C, Borchert M, et al.: Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3. J Diabetes Sci Technol 2010; 4: 976– 982. 20663464 PMC2909532 source
The Patient Characteristics section reports that events requiring infusion-set changes began on day 3 and became more frequent during and after that day. The Figure 4 discussion reports that the rise in mean blood glucose reached statistical significance on day 7, but the Discussion says changes in insulin absorption only “appeared” to occur; it does not establish impaired absorption as the cause or report a statistically significant absorption measurement. (LLM, conf=0.87)
Experimental — accuracy not yet measured
It may state two separate claims as one.
PremisecitationA recent pilot study of 22 adults using varied insulin delivery systems demonstrated a survival rate of 82% at 6.8 days of wear during use of a novel IIS with multiple perforations within the cannula shaft, without significant change in glucose or daily insulin delivered at this time point.
“In addition to these findings, a recent pilot study of 22 adults (40.1 ± 14.3 years of age) using varied insulin delivery systems demonstrated a survival rate of 82% at 6.8 days of wear, during use of a novel IIS with multiple perforations within the cannula shaft (i.e., Lantern technology [ConvaTec, Inc., Deeside, United Kingdom]) without significant change in glucose or daily insulin delivered”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [22] Lal RA, Hsu L, Zhang J, et al.: Longevity of the novel ConvaTec infusion set with Lantern technology. Diabetes Obes Metab 2021; 23: 1973– 1977. 33822472 PMC8720264 source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Probably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
PremisecitationPatel et al. indicated that kinking at initial insertion of Teflon-cannula IISs contributed to 15% of the total IIS failure rate, although investigational staff inserted all IIS devices during the study.
“For instance, Patel et al., indicated that kinking at initial insertion of Teflon-cannula IISs (MiniMed Quick-Set infusion set [Medtronic]) contributed to 15% of the total IIS failure rate, although investigational staff inserted all IIS devices during the study.8”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
The Results report that 15% of Teflon catheters failed on initial insertion because of kinking, and the Methods state that study staff inserted all infusion sets. The 15% figure is a failure rate for Teflon sets, not the proportion of the total IIS failure rate attributable to kinking as the citing sentence suggests. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Probably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
PremisecitationSome of the aforementioned studies reported that standard IIS survival appeared to depend heavily on the individual user, as ∼42% of Patel et al.'s study participants were able to wear an IIS for 6-7 days and 75% of those in Waldenmaier et al.'s study wore at least one standard IIS for 7 days.
“Interestingly, some of the aforementioned studies reported that standard IIS survival appeared to depend heavily on the individual user, as ∼42% of the Patel et al.'s study participants were able to wear an IIS for 6–7 days and 75% of those in Waldenmaier et al.'s study wore at least one standard IIS for 7 days.”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [8] Patel PJ, Benasi K, Ferrari G, et al.: Randomized trial of infusion set function: steel versus teflon. Diabetes Technol Ther 2014; 16: 15– 19. 24090124 PMC3887416 source
- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
It may state two separate claims as one.
PremisecitationEarly studies of extended standard IIS wear demonstrated deterioration in glycemic control and significant IIS- or IIS site-related adverse events.
“Early studies of extended standard IIS wear demonstrated deterioration in glycemic control and significant IIS- or IIS site-related adverse events.1–3”
Quote in the paperSubstantial claim3 of 3 sources checked- cites [1] Pfutzner A, Sachsenheimer D, Grenningloh M, et al.: Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: results from a prospective randomized crossover study. J Diabetes Sci Technol 2015; 9: 1292– 1298. 26341262 PMC4667306 source
- cites [2] Schmid V, Hohberg C, Borchert M, et al.: Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3. J Diabetes Sci Technol 2010; 4: 976– 982. 20663464 PMC2909532 source
- cites [3] Thethi TK, Rao A, Kawji H, et al.: Consequences of delayed pump infusion line change in patients with type 1 diabetes mellitus treated with continuous subcutaneous insulin infusion. J Diabetes Complications 2010; 24: 73– 78. 19395280 PMC2854665 source
The Conclusions (page unavailable) report that glycemic control deteriorated after 48 hours of pump infusion line use. The Introduction (page unavailable) describes a serious infusion-site infection after prolonged wear, but attributes it to an earlier case report; this study does not demonstrate significant IIS- or site-related adverse events of its own. (LLM, conf=0.85)
Experimental — accuracy not yet measured
Probably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
Findingown dataThe intention to treat population comprised 259 adults with type 1 diabetes, mean age 45.0 ± 14.1 years, who wore a total of 3041 EIS devices, including 1561 with insulin lispro and 1480 with insulin aspart.
“The demographics of the intention to treat population (n = 259, 45.0 ± 14.1 years of age, min–max 18.7–77.4 years) who wore a total of 3041 EIS devices (1561 used with insulin lispro and 1480 used with insulin aspart) are listed in Table 1.”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Findingown dataThe rate of EIS failure due to unexplained hyperglycemia at the end of day 7 was 0.4% (95% CI: 0.16-1.00) for insulin aspart use.
“Insulin aspart | 127 | 1480 | 6 | 0.4”
Quote in the paperSubstantial claimIt may state two separate claims as one.
Findingown dataThere were no serious adverse device effects, unanticipated adverse device effects, or diabetic ketoacidosis events.
“There were no SADE, UADE, or DKA events.”
Quote in the paperSubstantial claimIt may state two separate claims as one.
Findingown dataThe overall rates of serious adverse events, severe hypoglycemia, severe hyperglycemia, and skin infection were 3.8, 2.5, 104.1, and 20.1 events per 100 participant-years, respectively.
“The overall rates of SAEs, SevHypo, SevHyper, and skin infection were 3.8, 2.5, 104.1, and 20.1 events per 100 participant-years.”
Quote in the paperSubstantial claimIt may state two separate claims as one.
Findingown dataThere was no serious adverse device event, UADE, or episode of DKA.
“There was no serious adverse device event, UADE, or episode of DKA.”
Quote in the paperSubstantial claimIt may state two separate claims as one.
Findingown dataGlucose variability was 35.6% ± 5.7% on day 1 and 30.5% ± 4.7% on day 7.
“Glucose variability was 35.6% ± 5.7% on day 1 and 30.5% ± 4.7% on day 7 and TDD was relatively unchanged (54.3 ± 28.6 U, 53.5 ± 28.5 U, and 53.6 ± 28.0 U on days 1, 4, and 7, respectively).”
Quote in the paperSubstantial claimIt may state two separate claims as one.
Findingown dataIn this study, excluding the initial training visit, participants inserted the EIS on their own, the rate of kinking at insertion was 0.5%, and 99.6% of participants were able to wear at least one EIS for 7 days.
“In this study, and excluding the initial training visit, participants inserted the EIS on their own and the rate of kinking at insertion was a low 0.5% with 99.6% participants being able to wear at least one EIS for 7 days.”
Quote in the paperSubstantial claimIt may state two separate claims as one. The quote only partly states this claim.
Findingown dataThe EIS failure rate due to unexplained hyperglycemia and both unexplained hyperglycemia and device-related SevHyper was 1.1% for insulin lispro and 1.2% for insulin aspart.
“| Insulin lispro | 132 | 1561 | 17 | 1.1 |”
- Also stated in: § Effectiveness endpoint evaluation · plain
“| Insulin aspart | 127 | 1480 | 18 | 1.2 |”
Quote in the paperSubstantial claim- Also stated in: § Effectiveness endpoint evaluation · plain
It may state two separate claims as one.
PremisecitationThe MiniMed 670G system pivotal trial used a definition of SevHyper of BG >300 mg/dL and blood ketones >0.6 mM and involved use of a 3-day IIS.
“The MiniMed 670G system pivotal trial15 used a definition of SevHyper of BG >300 mg/dL and blood ketones >0.6 mM and involved use of a 3-day IIS.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15] Bergenstal RM, Garg S, Weinzimer SA, et al.: Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA 2016; 316: 1407– 1408. 27629148 source
It may state two separate claims as one.
PremisecitationFor users of both CSII and CGM technologies, the combined application and intervals of device change can be burdensome because of numerous adhesive patch removals and increased risk of skin irritation, skin reactions, or adhesive tape allergies, with minimal days for recovery.
“Thus, for users of both CSII and CGM technologies, the combined application and intervals of device change can be burdensome because of numerous adhesive patch removals and increased risk of skin irritation,5,6 skin reactions,5,6 or adhesive tape allergies,7 with minimal days for recovery.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [5] Berg AK, Norgaard K, Thyssen JP, et al.: Skin problems associated with insulin pumps and sensors in adults with type 1 diabetes: a cross-sectional study. Diabetes Technol Ther 2018; 20: 475– 482. 29893593 source
- cites [6] Christensen MO, Berg AK, Rytter K, et al.: Skin problems due to treatment with technology are associated with increased disease burden among adults with type 1 diabetes. Diabetes Technol Ther 2019; 21: 215– 221. 30943072 source
- cites [7] Heinemann L, Kamann S: Adhesives used for diabetes medical devices: a neglected risk with serious consequences? J Diabetes Sci Technol 2016; 10: 1211– 1215. 27566734 PMC5094339 source
It may state two separate claims as one.
Study design and conductown dataThe survival rate analysis was based on 2788 of 3041 infusion sets after excluding not-for-cause removals unrelated to the investigational device, including 234 early removals, 14 removals due to insulin depletion, and 5 due to pump replacement.
“The survival rate analysis was based on 2788 of 3041 infusion sets, excluding not-for-cause removals unrelated to the investigational device, where 234 sets were inadvertently removed by the participants early (<174 h) due to time calculation error and 19 sets were removed due to insulin depletion (n = 14) and pump replacement (n = 5).”
Quote in the paperSubstantial claimIt may state two separate claims as one.
PremisecitationThe new EIS includes a new H-cap connector and tubing with new fluid path design that improve insulin preservative retention and stability.
“Iterative advancements relative to earlier IIS devices designed for 2 to 3 days of wear included a new H-cap connector and tubing with new fluid path design that improve insulin preservative retention and stability,12”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Ilany J, Cohen O, Konvalina N, et al.: Clinical study of a new extended wear infusion set design. Diabetes Technol Ther 2020; 22: A- 27. source
Probably not a substantial claim. The paper’s conclusions may not depend on it.
Study design and conductcitationSample size estimation was based on previously published single-center data.
“Sample size estimation was based on previously published single-center data.14”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Buckingham BA, Marcal T, Hoffman L, et al.: Testing a novel infusion set for extended wear duration. Diabetes 2020; 69: 997- P. source
Probably not a substantial claim. It may state two separate claims as one.
Study design and conductreasoningApproximately 300 enrolled participants were estimated to ensure that 240 met eligibility criteria, with approximately half using insulin lispro and half using insulin aspart.
“Approximately 300 enrolled participants were estimated to ensure that 240 met eligibility criteria, with approximately half using insulin lispro and half using insulin aspart.”
Quote in the paperSubstantial claimProbably not a substantial claim. The paper’s conclusions may not depend on it.
Study design and conductown dataA total of 248 participants completed the study.
“A total of 248 participants completed the study.”
Quote in the paperNot a substantial claimProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
Study design and conductown dataThere were 291 adults enrolled in the trial, 29 screen failures, 4 early withdrawals, and 10 withdrawals after at least one EIS insertion.
“There were 291 adults who enrolled in the trial, 29 screen failures, 4 early withdrawals and 10 withdrawals after, at least, one insertion of an EIS (Fig. 2).”
Quote in the paperNot a substantial claimIt may state two separate claims as one.
Interpretationown dataThe EIS, when worn for up to 7 days, was safe and rated with high satisfaction, without adversely affecting glycemic control in adults with type 1 diabetes.
“This investigation demonstrates that the EIS, when worn for up to 7 days, was safe and rated with high satisfaction, without adversely affecting glycemic control in adults with T1D.”
Quote in the paperSubstantial claimProbably not a substantial claim. The paper may give no reason why it holds. It may state two separate claims as one.
InterpretationreasoningExtending the wear duration of IIS devices to closely align with CGM sensor change intervals would reduce diabetes management burden, IIS waste, and insulin loss associated with more frequent IIS changes.
“Extending the wear duration of IIS devices to closely align with that of CGM sensor change intervals would not only reduce diabetes management burden but also reduce IIS waste and insulin loss associated with more frequent IIS changes.”
Quote in the paperNot a substantial claim- Rests on an assumption: Continuous glucose monitoring systems often used in conjunction with CSII pump therapy can function for at least 7 days before a sensor change.
Probably not a substantial claim.
Study design and conductcitationA sample size of at least 100 for each insulin type would yield power >80% to demonstrate a failure rate due to unexplained hyperglycemia, on wear-day 6 or 7, of ≤20% with a two-sided 0.05 significance level.
“A sample size of at least 100 for each insulin type would yield power >80% to demonstrate a failure rate due to unexplained hyperglycemia, on wear-day 6 or 7, of ≤20% (with two-sided 0.05 significance level).”
Quote in the paperSubstantial claimProbably not a substantial claim. The paper’s conclusions may not depend on it.
PremisecitationIn Freckmann et al.'s study, 18.7% had kinked cannulas.
“In Freckmann et al.'s study, where the study participants (N = 80 total) inserted their IIS devices (Accu-Chek™ FlexLink infusion set and Accu-Chek FlexLink Plus infusion set [Roche Diabetes Care, Mannheim, Germany]), 18.7% had kinked cannulas.19”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [19] Freckmann G, Arndt S, Fiesselmann A, et al.: Randomized cross-over study comparing two infusion sets for CSII in daily life. J Diabetes Sci Technol 2017; 11: 253– 259. 27605591 PMC5478026 source
Probably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it. It may state two separate claims as one.
InterpretationreasoningThese different results likely reflect iterative modifications in IIS device performance, IIS insertion devices, and user experience, or a combination of these factors.
“These different results likely reflect iterative modifications in IIS device performance, IIS insertion devices, and user experience, or a combination of these factors.”
Quote in the paperNot a substantial claimIt may state two separate claims as one.
PremisecitationA recent medical economics Markov Chain Monte Carlo model analysis of standard IIS clinical data versus this study EIS trial data showed a greater average lifetime of 151.6 versus 66.0 h, resulting in 75 fewer set changes per year, for the EIS.
“A recent medical economics Markov Chain Monte Carlo model analysis of standard IIS clinical data8,18 versus this study EIS trial data showed a greater average lifetime of 151.6 versus 66.0 h, resulting in 75 fewer set changes per year, for the EIS.23”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [23] Kwa T, Zhang G, Shepard K, et al.: The improved survival rate and cost-effectiveness of a 7-day continuous subcutaneous insulin infusion set. J Med Econ 2021; 24: 837– 845. 34154504 source
Probably not a substantial claim. The paper’s conclusions may not depend on it.
Interpretationnone givenThe differences in failure rates observed across these studies were likely due to multiple factors, including study participant selection and definitions of infusion set failure.
“The differences in failure rates observed across these studies were likely due to multiple factors, including study participant selection and definitions of infusion set failure.”
Quote in the paperNot a substantial claimProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
PremisecitationIn Waldenmaier et al.'s study, participants could use an accompanying insertion device, and no issues with kinking at IIS insertion were reported.
“While a different Teflon-cannula IIS device (YpsoPump™ Orbit™ soft infusion set [Ypsomed AG, Burgdorf, Switzerland]) was used in Waldenmaier et al.'s study, participants could use an accompanying insertion device, and no issues with kinking at IIS insertion were reported.9”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Waldenmaier D, Zschornack E, Buhr A, et al.: A prospective study of insulin infusion set use for up to 7 days: early replacement reasons and impact on glycemic control. Diabetes Technol Ther 2020; 22: 734– 741. 32167382 source
Probably not a substantial claim. The paper’s conclusions may not depend on it.
Study design and conductreasoningThe Auto Mode function, which included multiple periods of time when insulin delivery was suspended to prevent hypoglycemia, served as a more rigorous test of infusion set function than an open-loop system where there is continuous insulin delivery.
“However, the Auto Mode function, which included multiple periods of time (some lasting hours) when insulin delivery was suspended to prevent hypoglycemia, served as a more rigorous test of infusion set function than an open-loop system where there is continuous insulin delivery.”
Quote in the paperNot a substantial claimIt may state two separate claims as one.
Interpretationown dataGlycemic control and participant-reported satisfaction during EIS use were favorable.
“Glycemic control and participant-reported satisfaction during EIS use were favorable.”
Quote in the paperSubstantial claimProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
Study design and conductown dataThe study population included a majority of white participants with relatively well-controlled glycemia at study start.
“In addition, the study population also included participants, a majority of whom were white, with relatively well-controlled glycemia at study start.”
Quote in the paperNot a substantial claimIt may state two separate claims as one.
Interpretationown dataThe Medtronic extended infusion set described in this study was safely worn for up to 7 days by adults without adversely affecting glycemic control and with favorable user satisfaction.
“The Medtronic extended infusion set described in this study was safely worn for up to 7 days by adults without adversely affecting glycemic control and with favorable user satisfaction.”
Quote in the paperSubstantial claimIt may state two separate claims as one.
PremisecitationOne randomized crossover study investigating intended real-world 3-day use of standard IIS devices reported comparable failure rates (or unplanned IIS changes) of 16%-20%, and ∼40% were due to unexpected high glucose, occlusion alarms, or backflow of insulin.
“It is also important to note that one randomized crossover study investigating intended real-world 3-day use of standard IIS devices reported comparable failure rates (or unplanned IIS changes) of 16%–20% and that ∼40% were due to unexpected high glucose, occlusion alarms, or backflow of insulin.19”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [19] Freckmann G, Arndt S, Fiesselmann A, et al.: Randomized cross-over study comparing two infusion sets for CSII in daily life. J Diabetes Sci Technol 2017; 11: 253– 259. 27605591 PMC5478026 source
The paper relies on this without support.
Continuous glucose monitoring systems often used in conjunction with CSII pump therapy can function for at least 7 days before a sensor change.
“However, continuous glucose monitoring (CGM) systems often used in conjunction with CSII pump therapy can function for ≥7 days before a sensor change.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Severe hypoglycemia while not wearing an EIS counted as occurring during wear
MinorThe paper says one of the two severe hypoglycemia events occurred when the participant was not wearing an EIS, but Table 2 counts both events under safety outcomes during EIS wear.
“A second event involved a woman, not wearing an EIS, who lost consciousness at work due to SevHypo”
“Table: Table 2. Safety Outcomes During Extended Infusion Set Wear”
Citations
Cites
Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
Bergenstal, R. M., Garg, S., Weinzimer, S. A., Buckingham, B. A., Bode, B. W., Tamborlane, W. V., & Kaufman, F. R. (2016) · JAMA · doi:10.1001/jama.2016.11708 · Cited by 644
Self-citationInsulin Infusion Set: The Achilles Heel of Continuous Subcutaneous Insulin Infusion
Heinemann, L., & Krinelke, L. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600429 · Cited by 148
Nonmetabolic Complications of Continuous Subcutaneous Insulin Infusion: A Patient Survey
Pickup, J. C., Yemane, N., Brackenridge, A., & Pender, S. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0192 · Cited by 123
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97
Randomized Trial of Infusion Set Function: Steel Versus Teflon
Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0119 · Cited by 90
Self-citationBerg, A. K., Nørgaard, K., Thyssen, J. P., Zachariae, C., Hommel, E., Rytter, K., & Svensson, J. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0088 · Cited by 87
Adhesives Used for Diabetes Medical Devices: A Neglected Risk With Serious Consequences?
Heinemann, L., & Kamann, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816662949 · Cited by 74
Kerr, D., Morton, J., Whately-Smith, C., Everett, J., & Begley, J. P. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200314 · Cited by 60
Pfützner, A., Sachsenheimer, D., Grenningloh, M., Heschel, M., Walther-Johannesen, L., Gharabli, R., & Klonoff, D. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815604438 · Cited by 53
Thethi, T. K., Rao, A., Kawji, H., Mallik, T., Yau, C. L., Christians, U., & Fonseca, V. (2010) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2009.03.002 · Cited by 53
Gibney, M., Xue, Z., Swinney, M., Bialonczyk, D., & Hirsch, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0342 · Cited by 32
Karlin, A. W., Ly, T. T., Pyle, L., Forlenza, G. P., Messer, L., Wadwa, R. P., DeSalvo, D. J., Payne, S. L., Hanes, S., Clinton, P., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0432 · Cited by 32
Self-citationChristensen, M. O., Berg, A. K., Rytter, K., Hommel, E., Thyssen, J. P., Svensson, J., & Nørgaard, K. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0007 · Cited by 28
Taleb, N., Messier, V., Ott-Braschi, S., Ardilouze, J.-L., & Rabasa-Lhoret, R. (2018) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2018.07.035 · Cited by 25
Waldenmaier, D., Zschornack, E., Buhr, A., Pleus, S., Haug, C., & Freckmann, G. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0445 · Cited by 23
Longevity of the novel ConvaTec infusion set with Lantern technology
Lal, R. A., Hsu, L., Zhang, J., Schøndorff, P. K., Heschel, M., & Buckingham, B. (2021) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14395 · Cited by 22
Self-citationRandomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life
Freckmann, G., Arndt, S., Fießelmann, A., Klausmann, G., Pralle, K., Künsting, T., & Petersen, B. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816667510 · Cited by 15
Kwa, T., Zhang, G., Shepard, K., Wherry, K., & Chattaraj, S. (2021) · Journal of Medical Economics · doi:10.1080/13696998.2021.1945784 · Cited by 11
Self-citation1012-P: Study of Insulin Stability Impact on Pump Therapy: Test Model Development
CHATTARAJ, S., ZHANG, G., ANSELMO, E., & FUSSELMAN, J. (. C. (2020) · Diabetes · doi:10.2337/db20-1012-p · Cited by 5
Self-citation986-P: Assessment of Adhesive Patches for an Extended-Wear Infusion Set
ZHANG, G., CHATTARAJ, S., ANSELMO, E., HOFFMAN, L., TRAN, M., & BONDY, S. (2020) · Diabetes · doi:10.2337/db20-986-p · Cited by 4
Self-citation994-P: Clinical Study of a New Extended Wear Infusion Set
ILANY, J., COHEN, O., KONVALINA, N., ZHANG, G., & CHATTARAJ, S. (2020) · Diabetes · doi:10.2337/db20-994-p · Cited by 4
Self-citation997-P: Testing a Novel Infusion Set for Extended Wear Duration
BUCKINGHAM, B. A., MARCAL, T., HOFFMAN, L., MUSOLINO, G., EKHLASPOUR, L., ZHANG, G., & CHATTARAJ, S. (2020) · Diabetes · doi:10.2337/db20-997-p · Cited by 2
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- Chattaraj S, , Zhang G, , Anselmo E, et al.: The Medtronic extended wear infusion set: determining mechanisms of action. Diabetes Technol Ther, 2021; 23:A149.
Cited by
Sorted by citations
Digital Technology for Diabetes
Hughes, M. S., Addala, A., & Buckingham, B. (2023) · New England Journal of Medicine · doi:10.1056/nejmra2215899 · Cited by 59
Self-citationClosed-loop insulin delivery: update on the state of the field and emerging technologies
Ware, J., & Hovorka, R. (2022) · Expert Review of Medical Devices · doi:10.1080/17434440.2022.2142556 · Cited by 47
The changing landscape of automated insulin delivery in the management of type 1 diabetes
Lakshman, R., Boughton, C. K., & Hovorka, R. (2023) · Endocrine Connections · doi:10.1530/ec-23-0132 · Cited by 43
Closed-loop systems: recent advancements and lived experiences
Kadiyala, N., Hovorka, R., & Boughton, C. K. (2024) · Expert Review of Medical Devices · doi:10.1080/17434440.2024.2406901 · Cited by 29
Battelino, T., Kuusela, S., Shetty, A., Rabbone, I., Cherubini, V., Campbell, F., Ahomäki, R., Tuomaala, A., Peters, C. J., Iafusco, D., et al. (2025) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(25)00091-9 · Cited by 27
Self-citationLipohypertrophy and Insulin: An Old Dog That Needs New Tricks
Huang, J. K., Yeung, A. M., Kerr, D. E., Gentile, S., Heinemann, L., Al‐Sofiani, M. E., Joseph, J. I., Seley, J. J., & Klonoff, D. C. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.04.006 · Cited by 17
Development of the Extended Infusion Set and Its Mechanism of Action
ZHANG, G., Cohen, O., & CHATTARAJ, S. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221112120 · Cited by 17
Self-citationHughes, M. S., Pasquel, F. J., Davis, G. M., Lal, R., Buckingham, B. A., Boughton, C. K., Brown, S. A., & Bally, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0343 · Cited by 13
Self-citationClosed-loop therapy: recent advancements and potential predictors of glycemic outcomes
Royston, C., Hovorka, R., & Boughton, C. K. (2025) · Expert Opinion on Drug Delivery · doi:10.1080/17425247.2025.2492363 · Cited by 11
Advances in Insulin Infusion Set in the New Era of Automated Insulin Delivery: A Systematic Review
ZHANG, G., ANSELMO, E., Kwa, T., Cohen, O., Vigersky, R. A., & CHATTARAJ, S. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221145731 · Cited by 8
Self-citationInsights for personalized choice of a hybrid closed–loop system: an expert opinion
Molfetta, S. D., Boscari, F., Rossi, A., Girelli, A., Lepore, G., Bozzetto, L., Irace, C., Laviola, L., & Bruttomesso, D. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113120 · Cited by 4
Walsh, J. S. A., & Heinemann, L. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241233607 · Cited by 4
Garhyan, P., Pratt, E. J., Klein, O., Famulla, S., Zijlstra, E., Lalonde, A., Swinney, M. R., KAZDA, C. M., & Dassau, E. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221145200 · Cited by 2
Pachołek, A., Oniszczuk, E. R. S. , Mierzwa, J., Wróbel, A., Piotrowicz, K., Gąsowski, J., & Tomasik, T. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2023.145426 · Cited by 2
Huang, Q., Ni, Y., Lei, M., Ling, P., Yan, J., Guo, X., Yang, D., & Wang, C. (2025) · BMJ Open · doi:10.1136/bmjopen-2024-094333 · Cited by 2
Progressive Acceleration of Insulin Exposure Over 7 Days of Infusion Set Wear
Kastner, J. R., Bailey, T. S., Strange, P., Shi, L., Oberg, K. A., Strasma, P. J., Joseph, J. I., & Muchmore, D. B. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0323 · Cited by 2
Self-citationLal, R., Lum, J. W., Reed, Z. W., Kruger, D. F., Reed, J. H. C., Weinstock, R. S., Carlson, A. L., Kudva, Y. C., Liljenquist, D. R., Cobry, E. C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1177/15209156251364548 · Cited by 3
Self-citationDevice Insertion Versus Material: Drivers of Inflammation in Diabetes Device Interfaces
Cunegundes, P. S., Wood, K., de Souza, J. G., Bhangu, A., Mao, L., & Klueh, U. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251389945 · Cited by 1
Bjerrekær, J., Berg, A. K., Christensen, M. H., Suvitaival, T., Nørgaard, K., & Svensson, J. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261433362 · Cited by 1
Kastner, J. R., Poettler, T., Kopanz, J., Hochfellner, D. A., Romey, M., Baumann, P. M., Muchmore, D. B., Strasma, P. J., & Mader, J. K. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14935 · Cited by 2
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.2503 · Cited by 1
Zhou, L., Huber, D. E., Antwerp, B. V., & Pennathur, S. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221123083 · Cited by 1
Ucieklak, D., Mrozińska, S., Wojnarska, A., Malecki, M. T., Klupa, T., & Matejko, B. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2022.144079
Automatisierte Insulindosierung: Ideale Diabeteskontrolle mit vielen offenen Fragen
Freckmann, G., Deiss, D., Ziegler, R., & Schlüter, S. (2023) · Deutsches Ärzteblatt · doi:10.3238/persdia.2023.03.24.01
Pump, Infusion Set, and Insulin Options and Selection
(2024) · American Diabetes Association eBooks · doi:10.2337/9781580408363.ch02
Study of pump-delivered insulin propagation into the subcutaneous tissue
Jacquemier, P. (2023) · theses.fr (ABES) · doi:10.70675/aee3d56aze546z4909za3d2zc16bb77f0750
Nieto de la Marca, M. D. L. O., Fernández Velasco, P., Del Amo Simón, S., Sánchez Ibáñez, M., Collantes Matallana, A., Gómez Hoyos, E., Torres Torres, B., Ortolá Buigues, A., de Luis Román, D. A., & Díaz Soto, G. (2026) · Endocrine · doi:10.1007/s12020-026-04743-z
Groisman, L., Thorson, T. J., Saudi, B., Myrick, R. J., Gonthier, A. R., Nelson, G., Loeum, C., Torjman, M. C., Joseph, J. I., Mohraz, A., et al. (2026) · Biomaterials Advances · doi:10.1016/j.bioadv.2026.215155
Metadata
- Authors
- Brazg, R., Garg, S. K., Bhargava, A., Thrasher, J. R., Latif, K., Bode, B. W., Bailey, T. S., Horowitz, B. S., Cavale, A., Kudva, Y. C., Kaiserman, K. B., Grunberger, G., Reed, J. C., Chattaraj, S., Zhang, G., Shin, J., Chen, V., Lee, S. W., Cordero, T. L., Rhinehart, A. S., Vigersky, R. A., & Buckingham, B. A.
- Affiliations
- Rainier Clinical Research Center, Renton, Washington, USA.
- Barbara Davis Center for Diabetes, Aurora, Colorado, USA.
- Iowa Diabetes and Endocrinology Research Center (US)
- Medical Investigations, Inc., Little Rock, Arkansas, USA.
- AM Diabetes and Endocrinology Center, Bartlett, Tennessee, USA.
- Atlanta Diabetes Associates (US)
- AMCR Institute (US)
- Metabolic Research Institute, Inc., West Palm Beach, Florida, USA.
- Diabetes and Endocrinology Consultants of Pennsylvania, Feasterville Trevose, Pennsylvania, USA.
- Mayo Clinic (US)
- SoCal Diabetes, Torrance, California, USA.
- Grunberger Diabetes Institute (US)
- Endocrine Research Solutions, Roswell, Georgia, USA.
- Medtronic (United States) (US)
- Stanford Medicine (US)
- Corresponding author
- Buckingham, Bruce A.
- ORCID
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2022-07-28
- Volume
- 24
- Issue
- 8
- Pages
- 535-543
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC9353978
- OpenAlex
- W4221083646
- Semantic Scholar
- 6336acfa4f845ee5d308430be7b61c6759fd590b
- Cited by
- 28
- References
- 21
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 5 · 2023: 7 · 2022: 7
- FWCI
- 4.44
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Technology & Therapeutics (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, continuous glucose monitoring, insulin pump therapy, MiniMed 670G, insulin lispro, insulin aspart, diabetic ketoacidosis, severe hypoglycemia, severe hyperglycemia, glycemic control, HbA1c, time in range, device safety, patient satisfaction
- MeSH terms
- Adult, Blood Glucose, Diabetes Mellitus, Type 1, Female, Humans, Hyperglycemia, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Survival Rate, Diabetic Ketoacidosis, Insulin Lispro
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-06, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-06 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review