Tutorial. Surface EMG detection, conditioning and pre-processing: Best practices
Open accessAbstract
This tutorial is aimed primarily to non-engineers, using or planning to use surface electromyography (sEMG) as an assessment tool for muscle evaluation in the prevention, monitoring, assessment and rehabilitation fields. The main purpose is to explain basic concepts related to: (a) signal detection (electrodes, electrode-skin interface, noise, ECG and power line interference), (b) basic signal properties, such as amplitude and bandwidth, (c) parameters of the front-end amplifier (input impedance, noise, CMRR, bandwidth, etc.), (d) techniques for interference and artifact reduction, (e) signal filtering, (f) sampling and (g) A/D conversion, These concepts are addressed and discussed, with examples. The second purpose is to outline best practices and provide general guidelines for proper signal detection, conditioning and A/D conversion, aimed to clinical operators and biomedical engineers. Issues related to the sEMG origin and to electrode size, interelectrode distance and location, have been discussed in a previous tutorial. Issues related to signal processing for information extraction will be discussed in a subsequent tutorial.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The frequency range of the input sine wave signals used in the performance tests (10-210 Hz) spans much of the relevant range of EEG, ECG, and EMG signals.
“Hence, the frequency range of the input sine wave signals used in our performance tests spans much of the relevant range of these three common bioelectric signals.”
From Comparison between Two Time Synchronization and Data Alignment Methods for Multi-Channel Wearable Biosensor Systems Using BLE Protocol (2023) · cites this paper as [37]
The paper describes an sEMG bandwidth of 20–500 Hz (Analog low-pass, high-pass and notch filters; page unavailable), which overlaps the citing paper’s 10–210 Hz test range. Our copy does not establish the relevant frequency ranges for EEG or ECG, and 10–210 Hz does not cover the full stated sEMG bandwidth. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Hewson, D. J., Hogrel, J.-Y., Langeron, Y., & Duchêne, J. (2003) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(02)00097-4 · Cited by 99 · see citing works
Flagged first, then by citations
Research works printed but not linked: 101
- Barbero, M., Merletti, R. and Rainoldi, A. (2012) Atlas of Muscle Innervation Zones: Understanding surface EMG and its applications. 1st edn. Milano, Italy: Springer-Verlag Italia. doi: 10.1007/978-88-470-2463-2. · DOI
- Bareket, L. et al. (2016) 'Temporary-tattoo for long-term high fidelity biopotential recordings', Scientific Reports, 10(6), pp. 1-8. doi: 10.1038/srep25727. · DOI
- Basmajian, J. V and De Luca, C. J. (1985) Muscles Alive: their functions revealed by electromyography. doi: 10.1227/01.NEU.0000028086.48597.4F. · DOI
- Bates, J. B. and Chu, Y. T. (1992) 'Electrode-electrolyte interface impedance: Experiments and model', Annals of Biomedical Engineering, 20, pp. 349-362. doi: 10.1007/BF02368536. · DOI
- Besomi, M. et al. (2019) 'Consensus for experimental design in electromyography (CEDE) project: Electrode selection matrix', Journal of Electromyography and Kinesiology, 48, pp. 128-144. doi: 10.1016/j.jelekin.2019.07.008. · DOI
- Barbero, M., Merletti, R. and Rainoldi, A. (2012) Understanding surface EMG and its applications doi: 10.1007/978-88-470-2463-2. · DOI
- Bareket, L. et al. Barbero, M., Merletti, R. and Rainoldi, A. (2012) Understanding surface EMG and its applications doi: 10.1007/978-88-470-2463-2. · DOI
- Blinowska, K. and Zygierewicz, J. (2011) Practical Biomedical Signal Analysis Using MATLAB. 1st edn, Practical Biomedical Signal Analysis Using MATLAB. 1st edn. CRC Press. doi: 10.1201/b11148. · DOI
- Botter, A. et al. (2019) 'Development and testing of acoustically-matched hydrogel-based electrodes for simultaneous EMG-ultrasound detection', Medical Engineering and Physics, 64(1), pp. 74-79. doi: 10.1016/j.medengphy.2018.12.002. · DOI
- Botter, A. and Vieira, T. M. (2015) 'Filtered virtual reference: A new method for the reduction of power line interference with minimal distortion of monopolar surface EMG', IEEE Transactions on Biomedical Engineering, 62(11), pp. 2638-2647. doi: 10.1109/TBME.2015.2438335. · DOI
- Burbank, D. P. and Webster, J. G. (1978) 'Reducing skin potential motion artefact by skin abrasion', Medical and Biological Engineering and Computing, 16(1), pp. 31-38. doi: 10.1007/BF02442929. · DOI
- Campanini, I. et al. (2020) 'Surface Electromyography: barriers limiting widespread use of sEMG in clinical assessment and neurorehabilitation.', In press in Frontiers in Neurology/Neurorehabilitation.
- Cattarello, P. and Merletti, R. (2016) 'Characterization of dry and wet Electrode-Skin interfaces on different skin treatments for HDsEMG', in 2016 IEEE International Symposium on Medical Measurements and Applications, MeMeA 2016 -Proceedings, pp. 1-6. doi: 10.1109/MeMeA.2016.7533808. · DOI
- Cerone, G. L., Botter, A. and Gazzoni, M. (2019) 'A Modular, Smart, and Wearable System for High Density sEMG Detection', IEEE Transactions on Biomedical Engineering, 66(12), pp. 3371- 3380. doi: 10.1109/TBME.2019.2904398. · DOI · Paper page
- Cerone, G. L. and Gazzoni, M. (2018) 'A wireless, minaturized multi-channel sEMG acquisition system for use in dynamic tasks', in 2017 IEEE Biomedical Circuits and Systems Conference, BioCAS 2017 -Proceedings, pp. 1-4. doi: 10.1109/BIOCAS.2017.8325129. · DOI
- DeLuca, C. J. et al. (2010) 'Filtering the surface EMG signal: Movement artifact and baseline noise contamination', Journal of Biomechanics, 43, pp. 1573-1579. doi: 10.1016/j.jbiomech.2010.01.027. · DOI
- Dobrev, D. and Daskalov, I. (2002) 'Two-electrode biopotential amplifier with current-driven inputs', Medical and Biological Engineering and Computing, 40(1), pp. 122 -127. doi: 10.1007/BF02347705. · DOI
- Dobrev, D., Neycheva, T. and Mudrov, N. (2005) 'Simple two-electrode biosignal amplifier', Medical and Biological Engineering and Computing, 43(1), pp. 725 -730. doi: 10.1007/BF02430949. European Network of Transmission System Operators for Electricity, E.-E. A. (2014) ENTSO2014. Available at: www. entsoe.eu. · DOI
- Farina, D. (2006) 'Surface Electromyography (EMG) Signal Processing', in Wiley Encyclopedia of Biomedical Engineering. doi: 10.1002/9780471740360.ebs1378. · DOI
- Farina, D., Jensen, W. and Akay, M. (2013) Introduction to Neural Engineering for Motor Rehabilitation, Introduction to Neural Engineering for Motor Rehabilitation. IEEE Press and J. Wiley. doi: 10.1002/9781118628522. · DOI
- Fernández, M. and Pallás-Areny, R. (2000) 'Ag-AgCl electrode noise in high-resolution ECG measurements', Biomedical Instrumentation and Technology, 34(2), pp. 125-130.
- Dobrev, D., Neycheva, T. and Mudrov, N. (2005) 'Simple two-electrode biosignal amplifier', Medical and Biological Engineering and Computing 10.1007/BF02430949. European Network of Transmission System Operators for Electricity, E.-E. A. (2014) 10.1007/BF02347705. · DOI
- Dobrev, D. and Daskalov, I. (2002) 'Two-electrode biopotential amplifier with current-driven Medical and Biological Engineering and Computing 10.1007/BF02347705. 10.1016/j.jbiomech.2010.01.027. · DOI
- Dobrev, D. and Daskalov, I. (2002) 'Two-electrode biopotential amplifier with current-driven Medical and Biological Engineering and Computing (2010) 'Filtering the surface EMG signal: Movement artifact and baseline noise Journal of Biomechanics 10.1016/j.jbiomech.2010.01.027. · DOI
- Dobrev, D. and Daskalov, I. (2002) 'Two-electrode biopotential amplifier with current-driven DeLuca, C. J. (2006) 'Electromygraphy', in (2010) 'Filtering the surface EMG signal: Movement artifact and baseline noise Journal of Biomechanics system for use in dynamic tasks', in , pp. 1-4. doi: 10.1109/BIOCAS.2017.8325129. · DOI
- DeLuca, C. J. (2006) 'Electromygraphy', in IEEE Transactions on Biomedical Engineering 3380. doi: 10.1109/TBME.2019.2904398. · DOI · Paper page
- Cerone, G. L. and Gazzoni, M. (2018) 'A wireless, minaturized multi-channel sEMG acquisition 2017 IEEE Biomedical Circuits and Systems Conference, , pp. 1-4. doi: 10.1109/BIOCAS.2017.8325129. · DOI
- Cerone, G. L., Botter, A. and Gazzoni, M. (2019) 'A Modular, Smart, and Wearable System for IEEE Transactions on Biomedical Engineering
- Cerone, G. L., Botter, A. and Gazzoni, M. (2019) 'A Modular, Smart, and Wearable System for IEEE Transactions on Biomedical Engineering , pp. 1-6. doi:
- Cattarello, P. and Merletti, R. (2016) 'Characterization of dry and wet Electrode-Skin interfaces on 2016 IEEE International Symposium on Medical , pp. 1-6. doi:
- Cattarello, P. and Merletti, R. (2016) 'Characterization of dry and wet Electrode-Skin interfaces on 2016 IEEE International Symposium on Medical (2020) 'Surface Electromyography: barriers limiting widespread use of sEMG , 16(1), pp. 31-38. doi: (2020) 'Surface Electromyography: barriers limiting widespread use of sEMG , 62(11), pp. 2638-2647. doi: 10.1109/TBME.2015.2438335. · DOI
- Burbank, D. P. and Webster, J. G. (1978) 'Reducing skin potential motion artefact by skin , 16(1), pp. 31-38. doi:
- Floyd, W. F. and Silver, P. H. S. (1950) 'Electromyographic study of patterns of activity of the anterior abdominal wall muscles in man.', Journal of anatomy, 84, pp. 132-145.
- Gazzoni, M. (2015) 'Low cost inkjet printing for the fast prototyping of surface EMG detection systems', in 2015 IEEE International Symposium on Medical Measurements and Applications, MeMeA 2015 -Proceedings, pp. 79-83. doi: 10.1109/MeMeA.2015.7145176. · DOI
- Glover, J. R. (1977) 'Adaptive Noise Canceling Applied to Sinusoidal Interferences', in IEEE Transactions on Acoustics, Speech, and Signal Processing, pp. 484-491. doi: 10.1109/TASSP.1977.1162997. · DOI
- Godin, D. T., Parker, P. A. and Scott, R. N. (1991) 'Noise characteristics of stainless-steel surface electrodes', Medical & Biological Engineering & Computing, 29, pp. 585-590. doi: 10.1007/BF02446089. · DOI
- Griss, P. et al. (2002) 'Characterization of micromachined spiked biopotential electrodes', IEEE Transactions on Biomedical Engineering, 49(6), pp. 597-604. doi: 10.1109/TBME.2002.1001974. · DOI
- Hermens, H. J. et al. (2000) 'Development of recommendations for SEMG sensors and sensor placement procedures', Journal of Electromyography and Kinesiology, 10(2), pp. 361-374. doi: 10.1016/S1050-6411(00)00027-4. · DOI · Paper page
- Huigen, E., Peper, A. and Grimbergen, C. A. (2002) 'Investigation into the origin of the noise of surface electrodes', Medical and Biological Engineering and Computing, 40(1), pp. 332 -338. doi: 10.1007/BF02344216. · DOI
- Jahanmiri-Nezhad, F. et al. (2015) 'A practice of caution: spontaneous action potentials or artifactual spikes?', Journal of NeuroEngineering and Rehabilitation, 12(1), pp. 1-5. doi: 10.1186/1743-0003-12-5. · DOI
- Keshtkaran, M. R. and Yang, Z. (2014) 'A fast, robust algorithm for power line interference cancellation in neural recording', Journal of Neural Engineering, 11(2), pp. 1-14. doi: 10.1088/1741-2560/11/2/026017. · DOI
- Kim, M. et al. (2018) 'Wireless sEMG system with a microneedle-based high-density electrode array on a flexible substrate', Sensors (Switzerland), 30(18), pp. 1-12. doi: 10.3390/s18010092. · DOI
- Koch, K. P., Schuettler, M. and Stieglitz, T. (2002) 'Considerations on noise of electrodes in combination with amplifiers for bioelectrical signal recording.', Biomedizinische Technik. Biomedical engineering, 47, pp. 514-516. doi: 10.1515/bmte.2002.47.s1b.514. · DOI
- Lu, F. et al. (2018) 'Review of stratum corneum impedance measurement in non-invasive penetration application', Biosensors, 8(2), pp. 1-20. doi: 10.3390/bios8020031. · DOI
- McCool, P. et al. (2014) 'Identification of contaminant type in surface electromyography (EMG) signals', IEEE Transactions on Neural Systems and Rehabilitation Engineering, 22, pp. 774-783. doi: 10.1109/TNSRE.2014.2299573. · DOI
- Merletti, R. et al. (2009) 'Technology and instrumentation for detection and conditioning of the surface electromyographic signal: State of the art', Clinical Biomechanics, 24(2), pp. 122-134. doi: 10.1016/j.clinbiomech.2008.08.006. · DOI
- Koch, K. P., Schuettler, M. and Stieglitz, T. (2002) 'Considerations on noise of electrodes in combination with amplifiers for bioelectrical signal recording.', Biomedical engineering et al. array on a flexible substrate', Koch, K. P., Schuettler, M. and Stieglitz, T. (2002) 'Considerations on noise of electrodes in combination with amplifiers for bioelectrical signal recording.', 10.1088/1741-2560/11/2/026017. (2018) 'Wireless sEMG system with a microneedle-based high-density electrode array on a flexible substrate', Keshtkaran, M. R. and Yang, Z. (2014) 'A fast, robust algorithm for power line interference cancellation in neural recording', 10.1088/1741-2560/11/2/026017. (2018) 'Wireless sEMG system with a microneedle-based high-density electrode · DOI
- Keshtkaran, M. R. and Yang, Z. (2014) 'A fast, robust algorithm for power line interference cancellation in neural recording', 10.1088/1741-2560/11/2/026017. · DOI
- Keshtkaran, M. R. and Yang, Z. (2014) 'A fast, robust algorithm for power line interference Journal of NeuroEngineering and Rehabilitation EMG signal processing Medical and Biological Engineering and Computing (2015) 'A practice of caution: spontaneous action potentials or Journal of NeuroEngineering and Rehabilitation
- Huigen, E., Peper, A. and Grimbergen, C. A. (2002) 'Investigation into the origin of the noise of Medical and Biological Engineering and Computing , 13(3), pp. 273-279. doi: https://doi.org/10.1016/S1050-6411(02)00097-4. · DOI · Paper page
- Huigen, E., Peper, A. and Grimbergen, C. A. (2002) 'Investigation into the origin of the noise of Medical and Biological Engineering and Computing Journal of Electromyography and , 13(3), pp. 273-279. doi: https://doi.org/10.1016/S1050-6411(02)00097-4. (2003) 'Evolution in impedance at the electrode-skin interface of two types of Journal of Electromyography and , 13(3), pp. 273-279. doi: https://doi.org/10.1016/S1050-6411(02)00097-4. (2000) 'Development of recommendations for SEMG sensors and sensor Journal of Electromyography and Kinesiology , 49(6), pp. 597-604. doi: 10.1109/TBME.2002.1001974. (2000) 'Development of recommendations for SEMG sensors and sensor , 10(2), pp. 361-374. doi: (2002) 'Characterization of micromachined spiked biopotential electrodes', , 49(6), pp. 597-604. doi: 10.1109/TBME.2002.1001974. · DOI · Paper page
- Godin, D. T., Parker, P. A. and Scott, R. N. (1991) 'Noise characteristics of stainless-steel surface , 29, pp. 585-590. doi: (2002) 'Characterization of micromachined spiked biopotential electrodes', Godin, D. T., Parker, P. A. and Scott, R. N. (1991) 'Noise characteristics of stainless-steel surface Applications', Critical Reviews in Biomedical Engineering, 38(4), pp. 347-379. doi: 10.1615/CritRevBiomedEng.v38.i4. · DOI
- Merletti, R. (2010) 'The electrode-skin interface and optimal detection of bioelectric signals', Physiological Measurement, 31(10), pp. 1-4. doi: 10.1088/0967-3334/31/10/e01. · DOI
- Merletti, R. and Bonato, P. (2008) 'Electromyography (EMG), Surface', in Wiley Encyclopedia of Biomedical Engineering. John Wiley & Sons, pp. 1-21. doi: 10.1002/9780471740360.ebs1156. · DOI
- Merletti, R. and Farina, D. (2016) Surface Electromyography: Physiology, Engineering and Applications. Edited by R. Merletti and D. Farina. IEEE Press and J. Wiley. doi: 10.1002/9781119082934. · DOI · Paper page
- Merletti, R., Farina, D. and Holobar, A. (2018) 'Surface Electromyography (sEMG)', in Wiley Encyclopedia of Electrical and Electronics Engineering. doi: 10.1002/047134608x.w8371. · DOI
- Merletti, R. and Muceli, S. (2019) 'Tutorial. Surface EMG detection in space and time: Best practices', Journal of Electromyography and Kinesiology, 49, pp. 1-16. doi: 10.1016/j.jelekin.2019.102363. · DOI
- Merlo, A. and Campanini, I. (2016) 'Applications in movement and gait analysis', in Merletti, R. and Farina, D. (eds) Surface Electromyography: Physiology, Engineering and Applications. IEEE Press and Wiley Interscience, p. 447. doi: 10.1002/9781119082934.ch16. · DOI
- Mewett, D. T., Reynolds, K. J. and Nazeran, H. (2004) 'Reducing power line interference in digitised electromyogram recordings by spectrum interpolation', Medical and Biological Engineering and Computing, 42(4), pp. 524-531. doi: 10.1007/BF02350994. · DOI
- Pani, D. et al. (2019) 'Validation of Polymer-Based Screen-Printed Textile Electrodes for Surface EMG Detection', IEEE Transactions on Neural Systems and Rehabilitation Engineering, 27(7), pp. 1370-1377. doi: 10.1109/TNSRE.2019.2916397. · DOI
- Piervirgili, G., Petracca, F. and Merletti, R. (2014) 'A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag-AgCl electrodes', Physiological Measurement, 35(10), pp. 2101-2118. doi: 10.1088/0967-3334/35/10/2101. · DOI
- Puurtinen, M. M. et al. (2006) 'Measurement of noise and impedance of dry and wet textile electrodes, and textile electrodes with hydrogel', in Annual International Conference of the IEEE Engineering in Medicine and Biology -Proceedings, pp. 6012-6015. doi: 10.1109/IEMBS.2006.260155. · DOI
- Radüntz, T. (2018) 'Signal quality evaluation of emerging EEG devices', Frontiers in Physiology, 9(98), pp. 1-12. doi: 10.3389/fphys.2018.00098. · DOI
- Shustak, S. et al. (2019) 'Home monitoring of sleep with a temporary-tattoo EEG, EOG and EMG electrode array: A feasibility study', Journal of Neural Engineering, 16(2). doi: 10.1088/1741- 2552/aafa05. · DOI
- Puurtinen, M. M. electrodes, and textile electrodes with hydrogel', in Engineering in Medicine and Biology -Proceedings 10.1109/IEMBS.2006.260155. , 35(10), pp. 2101-2118. doi: 10.1088/0967-3334/35/10/2101. · DOI
- Puurtinen, M. M. et al. electrodes, and textile electrodes with hydrogel', in Engineering in Medicine and Biology -Proceedings
- Piervirgili, G., Petracca, F. and Merletti, R. (2014) 'A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag-AgCl electrodes', , 35(10), pp. 2101-2118. doi: 10.1088/0967-3334/35/10/2101. 1370-1377. doi: 10.1109/TNSRE.2019.2916397. · DOI
- Piervirgili, G., Petracca, F. and Merletti, R. (2014) 'A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag-AgCl electrodes', (2019) 'Validation of Polymer-Based Screen-Printed Textile Electrodes for Surface IEEE Transactions on Neural Systems and Rehabilitation Engineering 1370-1377. doi: 10.1109/TNSRE.2019.2916397. · DOI
- Piervirgili, G., Petracca, F. and Merletti, R. (2014) 'A new method to assess skin treatments for . 1st edn. Edited by D. Mitchell. Nova Science Publisher. (2019) 'Validation of Polymer-Based Screen-Printed Textile Electrodes for Surface IEEE Transactions on Neural Systems and Rehabilitation Engineering , 42(4), pp. 524-531. doi: 10.1007/BF02350994. Surface Electromyography: Fundamentals, Computational Techniques and . 1st edn. Edited by D. Mitchell. Nova Science Publisher. · DOI
- Mewett, D. T., Reynolds, K. J. and Nazeran, H. (2004) 'Reducing power line interference in digitised electromyogram recordings by spectrum interpolation', , 42(4), pp. 524-531. doi: 10.1007/BF02350994. Surface Electromyography: Fundamentals, Computational Techniques and Press and Wiley Interscience, p. 447. doi: 10.1002/9781119082934.ch16. · DOI
- Merlo, A. and Campanini, I. (2016) 'Applications in movement and gait analysis', in Merletti, R. Surface Electromyography: Physiology, Engineering and Applications Press and Wiley Interscience, p. 447. doi: 10.1002/9781119082934.ch16. · DOI
- Merlo, A. and Campanini, I. (2016) 'Applications in movement and gait analysis', in Merletti, R. Merletti, R. and Muceli, S. (2019) 'Tutorial. Surface EMG detection in space and time: Best , 49, pp. 1-16. doi:
- Merlo, A. and Campanini, I. (2016) 'Applications in movement and gait analysis', in Merletti, R. doi: 10.1002/047134608x.w8371. · DOI
- Merletti, R. and Muceli, S. (2019) 'Tutorial. Surface EMG detection in space and time: Best , 49, pp. 1-16. doi:
- Merletti, R., Farina, D. and Holobar, A. (2018) 'Surface Electromyography (sEMG)', in doi: 10.1002/047134608x.w8371. . Edited by R. Merletti and D. Farina. IEEE Press and J. Wiley. doi: · DOI
- Merletti, R., Farina, D. and Holobar, A. (2018) 'Surface Electromyography (sEMG)', in . John Wiley & Sons, pp. 1-21. doi: 10.1002/9780471740360.ebs1156. Surface Electromyography: Physiology, Engineering and Wiley Encyclopedia of . John Wiley & Sons, pp. 1-21. doi: 10.1002/9780471740360.ebs1156. Surface Electromyography: Physiology, Engineering and · DOI
- Sinderby, C., Lindström, L. and Grassino, A. E. (1995) 'Automatic assessment of electromyogram qualityitle', Journal of Applied Physiology, 79(5), pp. 1803-1815. doi: doi: 10.1152/jappl.1995.79.5.1803. · DOI
- Spinelli, E. et al. (2016) 'A simple and reproducible capacitive electrode', Medical Engineering and Physics, 38(3), pp. 286-289. doi: 10.1016/j.medengphy.2015.12.006. · DOI
- Spinelli, E. M. and Mayosky, M. A. (2005) 'Two-electrode biopotential measurements: Power line interference analysis', IEEE Transactions on Biomedical Engineering, 52(8), pp. 1436-1442. doi: 10.1109/TBME.2005.851488. · DOI
- De Talhouet, H. and Webster, J. G. (1996) 'The origin of skin-stretch-caused motion artifacts under electrodes', Physiological Measurement, 17(2), pp. 1-8. doi: 10.1088/0967-3334/17/2/003. · DOI
- Tam, H. W. and Webster, J. G. (1977) 'Minimizing Electrode Motion Artifact by Skin Abrasion', IEEE Transactions on Biomedical Engineering, 24(2), pp. 134-139. doi: 10.1109/TBME.1977.326117. · DOI
- Turner, I. N. et al. (1995) 'Noise coherence in closely-spaced electrodes: The implications for spatial averaged EGG recordings', Journal of Medical Engineering and Technology, 19(5), pp. 158-161. doi: 10.3109/03091909509009987. · DOI
- Webster, J. G. (1984) 'Reducing Motion Artifacts and Interference in Biopotential Recording', IEEE Transactions on Biomedical Engineering, 31(12), pp. 823-826. doi: 10.1109/TBME.1984.325244. · DOI
- Webster, J. G. (2013) Medical Instrumentation-Application and Design, John Wiley & Sons. doi: 10.1097/00004669-197807000-00017. · DOI
- Winder, S. (2002) Analog and digital filter design. Elsevier Inc.
- Winter, B. B. and Webster, J. G. (1983) 'Reduction of Interference Due to Common Mode Voltage in Biopotential Amplifiers', IEEE Transactions on Biomedical Engineering, 30(1), pp. 58-62. doi: 10.1109/TBME.1983.325167. · DOI
- Winter, B. B. and Webster, J. G. (1983) 'Driven-Right-Leg Circuit Design', IEEE Transactions on Biomedical Engineering, 30(1), pp. 62-66. doi: 10.1109/TBME.1983.325168. · DOI
- Yousef, H., Alhajj, M. and Sharma, S. (2019) Anatomy, skin (integument), epidermis. StatPearls Publishing. Available at: https://www.ncbi.nlm.nih.gov/books/NBK470464/.
- Zhao, Z. and Zhang, Y. (2018) 'SQI quality evaluation mechanism of single-lead ECG signal based on simple heuristic fusion and fuzzy comprehensive evaluation', Frontiers in Physiology, 9(727), pp. 1-13. doi: 10.3389/fphys.2018.00727. · DOI
- Zucca, A. et al. (2015) 'Tattoo Conductive Polymer Nanosheets for Skin-Contact Applications', Advanced Healthcare Materials, 4(7), pp. 983-990. doi: 10.1002/adhm.201400761. · DOI
- Yousef, H., Alhajj, M. and Sharma, S. (2019) Publishing. Available at: https://www.ncbi.nlm.nih.gov/books/NBK470464/.
- Yousef, H., Alhajj, M. and Sharma, S. (2019) Publishing. Available at: https://www.ncbi.nlm.nih.gov/books/NBK470464/.
- Winter, B. B. and Webster, J. G. (1983) 'Driven-Right-Leg Circuit Design', Biomedical Engineering, 30(1), pp. 62-66. doi: 10.1109/TBME.1983.325168. in Biopotential Amplifiers', 10.1109/TBME.1983.325167. · DOI
- Winter, B. B. and Webster, J. G. (1983) 'Driven-Right-Leg Circuit Design', Winter, B. B. and Webster, J. G. (1983) 'Reduction of Interference Due to Common Mode Voltage in Biopotential Amplifiers', IEEE Transactions on Biomedical Engineering 10.1109/TBME.1983.325167. 10.1097/00004669-197807000-00017. · DOI
- Analog and digital filter design Winter, B. B. and Webster, J. G. (1983) 'Reduction of Interference Due to Common Mode Voltage Medical Instrumentation-Application and Design 10.1097/00004669-197807000-00017. · DOI
- Webster, J. G. (1984) 'Reducing Motion Artifacts and Interference in Biopotential Recording', IEEE Transactions on Biomedical Engineering, 31(12), pp. 823-826. doi: Medical Instrumentation-Application and Design
- Webster, J. G. (1984) 'Reducing Motion Artifacts and Interference in Biopotential Recording', , 31(12), pp. 823-826. doi: (1995) 'Noise coherence in closely-spaced electrodes: The implications for Journal of Medical Engineering and Technology
- Webster, J. G. (1984) 'Reducing Motion Artifacts and Interference in Biopotential Recording', (1995) 'Noise coherence in closely-spaced electrodes: The implications for Journal of Medical Engineering and Technology Understanding data converters, SLAA013, Mixed-Signal Products (1995) 'Noise coherence in closely-spaced electrodes: The implications for Journal of Medical Engineering and Technology , 24(2), pp. 134-139. doi: Mixed-Signal Products
- Tam, H. W. and Webster, J. G. (1977) 'Minimizing Electrode Motion Artifact by Skin Abrasion', , 24(2), pp. 134-139. doi:
- De Talhouet, H. and Webster, J. G. (1996) 'The origin of skin-stretch-caused motion artifacts under , 17(2), pp. 1-8. doi: 10.1088/0967-3334/17/2/003. · DOI
- De Talhouet, H. and Webster, J. G. (1996) 'The origin of skin-stretch-caused motion artifacts under , 17(2), pp. 1-8. doi: 10.1088/0967-3334/17/2/003. Electromyography: New developments, procedures and applications. 1st edn. · DOI
Cited by
Sorted by citations
Wang, H., Li, J., McDonald, B. E., Farrell, T. R., Huang, X., & Clancy, E. A. (2023) · Sensors · doi:10.3390/s23052465 · Cited by 18 · see citing works
- Authors
- Merletti, R., & Cerone, G. L.
- Affiliations
- Politecnico di Torino (IT)
- Corresponding author
- Merletti, R.
- Venue
- Journal of Electromyography and Kinesiology
- Publisher
- Elsevier BV
- Published
- 2020-10
- Volume
- 54
- Pages
- 102440
- ISSN
- 1050-6411 (ISSN-L)
- OpenAlex
- W3037535647
- Cited by
- 233
- References
- 1
- Citations by year
- 2026: 46 · 2025: 52 · 2024: 56 · 2023: 37 · 2022: 21 · 2021: 17 · 2020: 4
- FWCI
- 6.21
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- Locations
- Journal of Electromyography and Kinesiology (publishedVersion)
- PubMed (publishedVersion)
- PORTO@Iris (submittedVersion, cc-by-nc-nd)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Keywords
- surface electromyography, signal detection, electrodes, skin-electrode interface, ECG interference, power line interference, front-end amplifier, signal filtering, sampling, analog-to-digital conversion, artifact reduction, muscle assessment, best practices
- MeSH terms
- Amplifiers, Electronic, Electrodes, Electromyography, Humans, Artifacts, Practice Guidelines as Topic, Muscle, Skeletal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review