Development and validation of a highly sensitive LC–MS/MS assay for the quantification of arginine vasopressin in human plasma and urine: Application in preterm neonates and child
Abstract
Arginine vasopressin is an endogenous neuropeptide secreted in response to situations such as hyperosmolality, hypotension and hypovolemia. The purpose of this study was to develop a reliable assay using small volumes of plasma and urine samples to quantify vasopressin levels in preterm infants. Weak cation solid-phase extraction was used to extract vasopressin from 200μl human plasma and urine samples. Separation was achieved on a Waters Acquity UPLC BEH C18 column by gradient elution at 0.55ml/min, with a mobile phase composed of methanol and 0.02% aqueous acetic acid solution. Analysis was performed under a hybrid triple quadrupole linear ion trap mass spectrometer, operated in multiple reaction monitoring mode using positive ionization. The linear response range was 1.0-40pg/ml for vasopressin, with the lower limit of quantification (LLOQ) of 1.0pg/ml in human plasma and urine. Recoveries at concentrations of 3, 10 and 32pg/ml were all greater than 70%, and matrix effects were within 15%. The method was validated with intra-day and inter-day precision of less than 8% for human plasma and urine. The intra-day and inter-day accuracy for human plasma were 91.9-100.6% and 92.3-104.8%, respectively. The intra-day and inter-day accuracy for human urine were 89.2-95.9% and 89.3-91.3%, respectively. The validated method was successfully applied to analyze two preterm neonate plasma samples and one child urine sample. In conclusion, the developed and validated method was sensitive and reliable, and was successfully used to quantify endogenous vasopressin levels in neonate plasma and child urine.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A similar LC-MS/MS assay was used to measure vasopressin in human plasma, again using human plasma diluted with water (1:6) as a surrogate matrix, and reported levels below 2 pg/ml in children.
“A similar LC‐MS/MS assay was used to measure vasopressin in human plasma, again using human plasma diluted with water (1 : 6) as a surrogate matrix, and this reported levels < 2 pg/ml in children 126.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [126]
Cites
Mallet, C. R., Lu, Z., & Mazzeo, J. R. (2003) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.1276 · Cited by 442 · see citing works
Bankir, L. (2001) · Cardiovascular Research · doi:10.1016/s0008-6363(01)00328-5 · Cited by 300 · see citing works
Zhang, G., Zhang, Y., Fast, D. M., Lin, Z., & Steenwyk, R. (2011) · Analytical Biochemistry · doi:10.1016/j.ab.2011.04.041 · Cited by 79 · see citing works
Vasopressin: physiology and clinical use in patients with vasodilatory shock: a review.
den Ouden, D. T., & Meinders, A. E. (2005) · PubMed · Cited by 71 · see citing works
Low‐dose vasopressin infusion therapy for refractory hypotension in ELBW infants
Ikegami, H., Funato, M., Tamai, H., et al. (2009) · Pediatrics International · doi:10.1111/j.1442-200x.2009.02967.x · Cited by 53 · see citing works
Vasopressin: A Review of Therapeutic Applications
Holt, N. F., & Haspel, K. L. (2009) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2009.09.006 · Cited by 41 · see citing works
Mabrouk, O. S., & Kennedy, R. T. (2012) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2012.06.006 · Cited by 38 · see citing works
Price, J. F., Towbin, J. A., Denfield, S. W., et al. (2004) · Circulation · doi:10.1161/01.cir.0000129764.84596.eb · Cited by 26 · see citing works
Serum Vasopressin Response in Patients With Intradialytic Hypotension
Rho, M., Perazella, M. A., Parikh, C. R., et al. (2008) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.05341107 · Cited by 24 · see citing works
Therapeutic applications of vasopressin in pediatric patients
Agrawal, A., Singh, V., Varma, A., et al. (2012) · Indian Pediatrics · doi:10.1007/s13312-012-0046-0 · Cited by 17 · see citing works
Cool, D. R., & DeBrosse, D. (2003) · Journal of Chromatography B · doi:10.1016/s1570-0232(03)00284-8 · Cited by 16 · see citing works
Gordon, A. C. (2011) · Journal of the Intensive Care Society · doi:10.1177/175114371101200105 · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers
Leng, G., & Sabatier, N. (2016) · Journal of Neuroendocrinology · doi:10.1111/jne.12413 · Cited by 237 · see citing works
- Authors
- Zhang, D., Rios, D. R., Tam, V. H., & Chow, D. S.-L.
- Affiliations
- University of Houston (US)
- Baylor College of Medicine (US)
- Corresponding author
- Chow, Diana S.-L.
- Venue
- Journal of Pharmaceutical and Biomedical Analysis
- Publisher
- Elsevier BV
- Published
- 2014-10
- Volume
- 99
- Pages
- 67-73
- ISSN
- 0731-7085 (ISSN-L)
- OpenAlex
- W2076034666
- Cited by
- 29
- References
- 12
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 3 · 2023: 4 · 2021: 1 · 2020: 1 · 2019: 1 · 2018: 1 · 2016: 4 · 2015: 6
- FWCI
- 2.98
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Pharmaceutical and Biomedical Analysis (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Thrasher Research Fund
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Methods in Pharmaceuticals
- Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
- Social Sciences › Psychology › Social Psychology › Neuroendocrine regulation and behavior
- Keywords
- vasopressin, LC-MS/MS method development, preterm infants, neonatal plasma, urine, multiple reaction monitoring, method validation, small-volume sample analysis
- MeSH terms
- Arginine Vasopressin, Child, Preschool, Chromatography, High Pressure Liquid, Female, Humans, Infant, Infant, Newborn, Infant, Premature, Male, Reference Standards, Reproducibility of Results, Tandem Mass Spectrometry, Limit of Detection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review