Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers

Open access
Leng, G., Sabatier, N.·2016·Journal of Neuroendocrinology

Abstract

In this review, we consider the ways in which vasopressin and oxytocin have been measured since their first discovery. Two different ways of measuring oxytocin in widespread use currently give values in human plasma that differ by two orders of magnitude, and the values measured by these two methods in the same samples show no correlation. The notion that we should accept this seems absurd. Either one (or both) methods is not measuring oxytocin, or , by ‘oxytocin’, the scientists that use these different methods mean something very different. If these communities are to talk to each other, it is important to validate one method and invalidate the other, or else to establish exactly what each community understands by ‘oxytocin’. A similar issue concerns vasopressin: again, different ways of measuring vasopressin give values in human plasma that differ by two orders of magnitude, and it appears that the same explanation for discrepant oxytocin measurements applies to discrepant vasopressin measurements. The first assays for oxytocin and vasopressin measured biological activity directly. When immunoassays were introduced, they encountered problems: high molecular weight factors in raw plasma interfered with the binding of antibodies to the hormones, leading to high and erroneous readings. When these interfering factors were removed by extraction of plasma samples, immunoassays gave measurements consistent with bioassays, with measures of turnover and with the sensitivity of target tissues to exogenous hormone. However, many recent papers use an enzyme‐linked immunoassay to measure plasma levels without extracting the samples. Like the first radioimmunassays of unextracted plasma, this generates impossibly high and wholly erroneous measurements.

AI snapshot

Written from the paper’s full text.

TL;DR

This review argues that widely used unextracted-plasma immunoassays for oxytocin and vasopressin generate impossibly high, erroneous values, while extracted bioassays and radioimmunoassays converge on much lower physiological concentrations. It calls for assay validation and warns that much recent oxytocin/vasopressin literature may be based on measurements no better than random numbers.

Key findings

  • Two current methods for measuring oxytocin in human plasma differ by two orders of magnitude and show no correlation in the same samples; the same discrepancy applies to vasopressin.
  • Many recent studies using unextracted plasma ELISA report oxytocin and vasopressin levels typically of 100–500 pg/ml, whereas bioassays and extracted radioimmunoassays indicate normal circulating levels are <10 pg/ml; normal plasma vasopressin in man is below 2 μU/ml (5 pg/ml), rising several-fold after 24 h dehydration.
  • Comparisons of the same ELISA with and without extraction found no correlation: in seals, extracted samples gave approximately 8 pg/ml versus 300 pg/ml in unextracted heparin samples or 540 pg/ml in EDTA samples; in humans, extracted RIA gave 1.1 pg/ml while unextracted ELISA gave 97 pg/ml, and spiking unextracted samples exaggerated measurements 3.5-fold.
  • The authors attribute the high values to plasma protein interference with antibody binding, producing a high erroneous baseline; they conclude such assays in unextracted plasma are “no more than a random number generator.”

Methods

  • Narrative review of historical and current methods for measuring oxytocin and vasopressin, including bioassays, radioimmunoassays, ELISAs, LC-MS/MS, and copeptin as a vasopressin surrogate.
  • Synthesizes published studies and historical standards, including comparisons of extracted versus unextracted plasma, assay validation studies, clearance and secretion estimates, and target-tissue sensitivity data.
  • Main analysis is qualitative and quantitative plausibility checking rather than a systematic review or meta-analysis; it uses cited values to show that unextracted ELISA results are inconsistent with known production, storage, clearance, and bioactivity.

Limitations

  • The paper states no formal limitations section.
  • Evident from the text: it is a narrative review without a systematic search strategy or meta-analysis, and it presents no new primary experimental data.
  • Evident from the text: its conclusions depend on the cited literature and on calculations from previously published values; the proposed validation recommendations are not prospectively tested here.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1111/jne.12413

Cites

134 of 134 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 122 cited sources checked · 0 supported · 0 flagged

  • Findingnone given

    None of the four studies cited in the Introduction as examples of the use of ELISAs to measure oxytocin and vasopressin in unextracted samples made any correction for multiple comparisons in their analyses.

    “Indeed, none of the four studies cited in the Introduction 3, 4, 5, 6 as examples of the use of ELISAs to measure oxytocin and vasopressin in unextracted samples made any correction for multiple comparisons in their analyses.”
    Quote in the paperSubstantial claim0 of 4 sources checked
    • cites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source

      Not checked in the free run.

    • cites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source

      Not checked in the free run.

    • cites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source

      Not checked in the free run.

    • cites [6] Rubin LH, Carter CS, Bishop JR, Pournajafi‐Nazarloo H, Drogos LL, Hill SK, Ruocco AC, Keedy SK, Reilly JL, Keshavan MS, Pearlson GD, Tamminga CA, Gershon ES, Sweeney JA. Reduced levels of vasopressin and reduced behavioral modulation of oxytocin in psychotic disorders. Schizophr Bull 2014; 40: 1374– 1384. 24619535 PMC4193712 source

      Not checked in the free run.

  • Premisecitation

    More than 100 papers have recently been published using assays claimed to measure oxytocin and vasopressin in human plasma simply and reliably.

    “more than 100 papers have recently been published using assays that have been claimed to measure oxytocin and vasopressin in human plasma simply and reliably”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source

      Not checked in the free run.

    • cites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source

      Not checked in the free run.

    • cites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source

      Not checked in the free run.

  • Premisecitation

    Measurements by radioimmunoassay reported that, during suckling in women, intermittent pulses were observed with a mean amplitude of just approximately 9 pg/ml from a basal level of approximately 2 pg/ml, as expected for bolus release of approximately 17 ng.

    “Measurements by radioimmunoassay reported that, during suckling in women, intermittent pulses were observed with a mean amplitude of just approximately 9 pg/ml from a basal level of approximately 2 pg/ml, as expected for bolus release of approximately 17 ng 64.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [64] Ueda T, Yokoyama Y, Irahara M, Aono T. Influence of psychological stress on suckling‐induced pulsatile oxytocin release. Obstet Gynecol 1994; 84: 259– 262. 8041543

      Not checked in the free run.

  • Premisecitation

    High molecular weight factors in raw plasma interfere with the binding of antibodies to oxytocin and vasopressin, leading to high and erroneous readings.

    “high molecular weight factors in raw plasma interfered with the binding of antibodies to the hormones, leading to high and erroneous readings”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • Premisecitation

    Stimulation of the isolated posterior pituitary at 50 Hz releases approximately 7 pg of oxytocin per stimulus pulse, close to the release estimated from the intramammary pressure response to similar electrical stimulation in vivo.

    “Stimulation at 50 Hz releases approximately 7 pg of oxytocin per stimulus pulse 66, which is close to the release estimated from the intramammary pressure response to a similar electrical stimulation in vivo.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [66] Leng G, Bicknell R. Hormone release from the neural lobe In: Gross PM, ed. Circumventricular Organs and Body Fluids, vol. III Boca Raton, FL: CRC Press, 1987: 47– 62.

      Not checked in the free run.

  • May combine two claims. 69
  • Classified as probably not a substantial claim. 15
  • The quote only partly states this claim. 2
  • May combine two claims.

    Premisecitation

    The most sensitive oxytocin bioassays used mouse mammary strips and could detect 4 μU/ml (7 pg/ml) in 0.5 ml of sample, but even this was not sufficiently sensitive to measure levels in raw plasma.

    “The most sensitive bioassays used mouse mammary strips and could detect 4 μU/ml (7 pg/ml) in 0.5 ml of sample 50, although even this was still not sufficiently sensitive to measure levels in raw plasma.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [50] Robinson IC, Walker JM. Bioassay of oxytocin on the superfused mammary gland of the mouse, using an automatic apparatus. Br J Pharmacol 1974; 50: 409– 417. 4368453 PMC1776677 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    The human posterior pituitary contains approximately 21 μg of oxytocin, whereas that of the rat contains 0.5-1 μg.

    “the human posterior pituitary contains approximately 21 μg of oxytocin 72, 73, 74, whereas that of the rat contains 0.5–1 μg.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites [72] Chard T, Boyd NR, Forsling ML, McNeilly AS, Landon J. The development of a radioimmunoassay for oxytocin: the extraction of oxytocin from plasma, and its measurement during parturition in human and goat blood. J Endocrinol 1970; 48: 223– 234. 5471863 source

      Not checked in the free run.

    • cites [73] Chard T, Forsling ML, James MA, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: sensitivity of the assay in aqueous buffer solution, specificity and the dissociation of immunological and biological activity. J Endocrinol 1970; 46: 533– 542. 5443195 source

      Not checked in the free run.

    • cites [74] Chard T, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: radioiodination, antibody production and separation techniques. J Endocrinol 1970; 46: 269– 278. 5492003 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    The normal plasma concentration of vasopressin in man is below 2 μU/ml (5 pg/ml), increasing several fold after 24 h of dehydration.

    “the normal plasma concentration of vasopressin in man is below 2 μU/ml (5 pg/ml) 39, with this increasing several fold after 24 h of dehydration.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [39] Share L. Vasopressin, its bioassay and the physiological control of its release. Am J Med 1967; 42: 701– 712. 5337374 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Across mammalian species, blood volume is linearly proportional to body mass, although the mass of the pituitary is proportional only to the log of body mass.

    “Across mammalian species, blood volume is linearly proportional to body mass, although the mass of the pituitary is proportional only to the log of body mass 71.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [71] Stahl WR. Organ weights in primates and other mammals. Science 1965; 150: 1039– 1042. 5843618 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In lactating rats, each suckling-induced burst of oxytocin neurones is followed by an increase in intramammary pressure that can be mimicked by i.v. injection of approximately 0.5 mU (0.8 ng) of oxytocin.

    “Each burst is followed by an increase in intramammary pressure that can be mimicked by i.v. injection of approximately 0.5 mU (0.8 ng) of oxytocin 16, 65.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [16] Leng G, Pineda R, Sabatier N, Ludwig M. 60 YEARS OF NEUROENDOCRINOLOGY: the posterior pituitary, from Geoffrey Harris to our present understanding. J Endocrinol 2015; 226: T173– T185. 25901040 source

      Not checked in the free run.

    • cites [65] Wakerley JB, Lincoln DW. Intermittent release of oxytocin during suckling in the rat. Nat New Biol 1971; 233: 180– 181. 5287087 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    The kidney responds to changes in the rate of vasopressin secretion of less than 1 pg/min/kg, and the rate required to produce isosmolar urine during overhydration is approximately 3 pg/min/kg.

    “the kidney responds to changes in the rate of vasopressin secretion of < 1 pg/min/kg, and also that the rate required to produce isosmolar urine during overhydration is approximately 3 pg/min/kg”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites [44] Andersen LJ, Andersen JL, Schutten HJ, Warberg J, Bie P. Antidiuretic effect of subnormal levels of arginine vasopressin in normal humans. Am J Physiol 1990; 259: R53– R60. 2375429

      Not checked in the free run.

    • cites [45] Hollander W Jr, Williams TF, Fordham CC 3rd, Welt LG. A study of the quantitative relationship between antidiuretic hormone (vasopressin) and the renal tubular reabsorption of water. J Clin Invest 1957; 36: 1059– 1071. 13449158 PMC1072692 source

      Not checked in the free run.

    • cites [46] Raisz LG, Mc NW, Saxon L. Studies on the renal concentrating mechanism. I. Role of vasopressin. J Lab Clin Med 1958; 52: 437– 445. 13575934

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Saameli calculated that the half-life of oxytocin in pregnant women was approximately 3-4 min, and that uterine contractions were initiated by infusions that raised plasma concentrations by 3 μU/ml (5 pg/ml).

    “he calculated that the half‐life of oxytocin in pregnant women was approximately 3–4 min, and also that uterine contractions were initiated by infusions that raised plasma concentrations by 3 μU/ml (5 pg/ml).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [52] Saameli K. An indirect method for the estimation of oxytocin blood concentration and half‐life in pregnant women near term. Am J Obstet Gynecol 1963; 85: 186– 192. 13975829 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    The plasma volume of a rat is approximately 4 ml/100 g, and a burst that releases 0.5 mU (0.8 ng) oxytocin should raise circulating concentrations in a 300 g rat by approximately 70 pg/ml.

    “The plasma volume of a rat is approximately 4 ml/100 g 67, and so a burst that releases 0.5 mU (0.8 ng) oxytocin should raise circulating concentrations in a 300 g rat by approximately 70 pg/ml.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [67] Lee HB, Blaufox MD. Blood volume in the rat. J Nucl Med 1985; 26: 72– 76. 3965655

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Higuchi et al. developed a radioimmunoassay that allowed oxytocin to be measured in unextracted rat plasma, and during suckling-induced milk let-down plasma concentrations rose from 16.5 pg/ml to approximately 65 pg/ml, declining with a half-life of approximately 1.5 min.

    “Higuchi et al. 68 developed a radioimmunoassay that allowed oxytocin to be measured in unextracted rat plasma, and took samples at the time of suckling‐induced milk let‐down, as recognised by stretching movements of the suckling pups. Plasma concentrations rose from 16.5 pg/ml to approximately 65 pg/ml, and declined with a half‐life of approximately 1.5 min.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [68] Higuchi T, Tadokoro Y, Honda K, Negoro H. Detailed analysis of blood oxytocin levels during suckling and parturition in the rat. J Endocrinol 1986; 110: 251– 256. 3755747 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In pigs, suckling evokes increases in intramammary pressure equivalent to those induced by i.v. injection of 50 mU (83 ng) of oxytocin, which for a plasma volume of 5 l would raise plasma concentrations by approximately 17 pg/ml, close to the peak concentration (approximately 18 pg/ml) measured by radioimmunoassay during suckling-induced milk ejection.

    “In pigs (body weight 150–200 kg), suckling evokes increases in intramammary pressure equivalent to those induced by i.v. injection of 50 mU (83 ng) of oxytocin 69. For a plasma volume of 5 l, these would raise plasma concentrations by approximately 17 pg/ml, which is close to the peak concentration (approximately 18 pg/ml) measured by radioimmunoassay during suckling‐induced milk ejection.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [69] Ellendorff F, Forsling ML, Poulain DA. The milk ejection reflex in the pig. J Physiol 1982; 333: 577– 594. 6892113 PMC1197265 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Such interference has been recognised in immunoassays for cholecystokinin, parathyroid hormone, thyrotropic hormone, glucagon, secretin, angiotensin, growth hormone, vasoactive intestinal peptide, gonadotrophins and gonadotrophin-releasing hormone, all of which were resolved by sample extraction.

    “Such interference has also been recognised in immunoassays for cholecystokinin 80, parathyroid hormone 81, thyrotropic hormone 82, glucagon 83, secretin 84, 85, angiotensin 86, growth hormone 87, vasoactive intestinal peptide 88, gonadotrophins 89 and gonadotrophin‐releasing hormone 90, all of which were resolved by sample extraction.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites [80] Harvey RF, Dowsett L, Hartog M, Read AE. Radioimmunoassay of cholecystokinin‐pancreozymin. Gut 1974; 15: 690– 699. 4435583 PMC1413038 source

      Not checked in the free run.

    • cites [81] Christensen MS. A sensitive radioimmunoassay of parathyroid hormone in human serum using a specific extraction procedure. Scand J Clin Lab Invest 1976; 36: 313– 322. 959749 source

      Not checked in the free run.

    • cites [82] Spencer CA, Challand GS. Interference in a radioimmunoassay for human thyrotropin. Clin Chem 1977; 23: 584– 588. 556979 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    The first peak accounted for a 67% loss of binding of antibody to vasopressin, and the second peak accounted for a further 30% loss; neither apparent immunoreactivity was displaced by excess cold vasopressin, indicating occlusion of antibody recognition rather than competitive binding.

    “The first peak accounted for a 67% loss of binding of antibody to vasopressin, and the second peak accounted for a further 30% loss. Neither of these apparent ‘immunoreactivities’ were displaced by an excess of cold vasopressin, indicating that they did not involve competitive binding of vasopressin, but occlusion of antibody recognition of vasopressin.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given

    It is unlikely that an antibody will recognise a different ligand with higher affinity than it has for the ligand to which it was raised; it may recognise other ligands with lower affinity, although this presents a major problem only if that different ligand is present at higher concentrations.

    “However, it is unlikely that an antibody will recognise a different ligand with higher affinity than it has for the ligand to which it was raised; it may recognise other ligands with lower affinity, although this presents a major problem only if that different ligand is present at higher concentrations.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisereasoning

    Close attention was given to extraction and concentration procedures in developing immunoassays for vasopressin and oxytocin.

    “Thus, close attention was given to extraction and concentration procedures 76.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In unextracted plasma from healthy subjects, Fyhrquist et al. found a mean vasopressin concentration of 2 pg/ml supine and 6.2 pg/ml upright, in agreement with bioassay levels measured by Segar and Moore.

    “In unextracted plasma from healthy subjects, they found a mean vasopressin concentration of 2 pg/ml in the supine position and 6.2 pg/ml in the upright position, which is in agreement with the levels measured in bioassays by Segar and Moore 33.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Fyhrquist et al. 92

      Not checked in the free run.

    • cites Segar and Moore 33

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Robertson et al. resolved three peaks of immunoreactivity by filtering plasma on G-25 Sephadex; only the third and smallest was consistent with true vasopressin, being recovered in the same location as pure vasopressin.

    “By filtering plasma samples on G‐25 Sephadex, they resolved three peaks of ‘immunoreactivity’, only the third and smallest of which was consistent with true vasopressin, being recovered in the same location as pure vasopressin 78.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Enzymatic degradation of oxytocin is a problem only when measuring in plasma from pregnant women, and extraction procedures can remove such enzymes efficiently.

    “Fortunately, enzymatic degradation of oxytocin is a problem only when measuring in plasma from pregnant women 54, and extraction procedures can remove such enzymes efficiently.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [54] Leake RD, Weitzman RE, Glatz TH, Fisher DA. Plasma oxytocin concentrations in men, nonpregnant women, and pregnant women before and during spontaneous labor. J Clin Endocrinol Metab 1981; 53: 730– 733. 7287862 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In 1973, Chard noted that the best immunoassays for vasopressin were no more sensitive than the best bioassays; the same is still true today for most radioimmunoassays and most enzyme-linked immunoassays (ELISAs).

    “In 1973, Chard 76 noted that best immunoassays for vasopressin were no more sensitive than the best bioassays; the same is still true today for most radioimmunoassays and most enzyme‐linked immunoassays (ELISAs).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

  • May combine two claims. The quote only partly states this claim.

    Premisecitation

    The third peak reproduced the expected relationship with plasma osmotic pressure in normal adults, the expected high levels in nephrogenic diabetes insipidus, and the expected low levels in pituitary diabetes insipidus.

    “The immunoreactivity in the third peak reproduced the expected relationship with plasma osmotic pressure in normal adults, the expected high levels in nephrogenic diabetes insipidus”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    After accounting for plasma matrix interference, Robertson et al. calculated that normal plasma concentrations of vasopressin in man were approximately 1 pg/ml, increasing to approximately 10 pg/ml with dehydration.

    “After accounting for the plasma matrix interference, Robertson et al. 78 calculated that normal plasma concentrations of vasopressin in man were approximately 1 pg/ml, increasing to approximately 10 pg/ml with dehydration.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    There were considerable difficulties in developing immunoassays for vasopressin and oxytocin because they have a low molecular weight, are poor immunogens, making high-affinity antisera difficult, and circulating levels were clearly very low.

    “There were considerable difficulties in developing immunoassays for vasopressin and oxytocin. Because these have a low molecular weight, they are poor immunogens, making the preparation of high‐affinity antisera difficult, and circulating levels were clearly very low.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    By the end of the 20th century, bioassays were still of critical importance in the discovery of novel antagonists and agonists of vasopressin and oxytocin because biological activity cannot be inferred from structure alone.

    “By the end of the 20th century, bioassays were still ‘of critical importance’ in the discovery of novel antagonists and antagonists of vasopressin and oxytocin because biological activity cannot be inferred from structure alone 75.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [75] Chan WY, Wo NC, Stoev ST, Cheng LL, Manning M. Discovery and design of novel and selective vasopressin and oxytocin agonists and antagonists: the role of bioassays. Exp Physiol 2000; 85S: 7S– 18S. 10795902 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Overestimation can arise when enzymes are present that degrade oxytocin; the plasma of late pregnant women contains abundant oxytocinase, which inactivates oxytocin by splitting the peptide linkage between the cysteine and tyrosine residues.

    “Overestimation can also arise when enzymes are present that degrade oxytocin. The plasma of late pregnant women contains abundant oxytocinase, which inactivates oxytocin by splitting the peptide linkage between the cysteine and tyrosine residues 102.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [102] Nomura S, Ito T, Yamamoto E, Sumigama S, Iwase A, Okada M, Shibata K, Ando H, Ino K, Kikkawa F, Mizutani S. Gene regulation and physiological function of placental leucine aminopeptidase/oxytocinase during pregnancy. Biochim Biophys Acta 2005; 1751: 19– 25. 15894523 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    By the 1980s, many laboratories had generated radioimmunoassays for oxytocin using independently produced antibodies, and most but not all required extraction.

    “By the 1980s, many laboratories had generated radioimmunoassays for oxytocin, using independently produced antibodies 58, 72, 73, 74, 94, 95, 96, 97, 98] and most, but not all 99, 100, required extraction.”
    Quote in the paperNot a substantial claim0 of 10 sources checked
    • cites [58] Dawood MY. Evolving concepts of oxytocin for induction of labor. Am J Perinatol 1989; 6: 167– 172. 2653335 source

      Not checked in the free run.

    • cites [72] Chard T, Boyd NR, Forsling ML, McNeilly AS, Landon J. The development of a radioimmunoassay for oxytocin: the extraction of oxytocin from plasma, and its measurement during parturition in human and goat blood. J Endocrinol 1970; 48: 223– 234. 5471863 source

      Not checked in the free run.

    • cites [73] Chard T, Forsling ML, James MA, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: sensitivity of the assay in aqueous buffer solution, specificity and the dissociation of immunological and biological activity. J Endocrinol 1970; 46: 533– 542. 5443195 source

      Not checked in the free run.

    • cites [74] Chard T, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: radioiodination, antibody production and separation techniques. J Endocrinol 1970; 46: 269– 278. 5492003 source

      Not checked in the free run.

    • cites [94] Dawood MY, Raghavan KS, Pociask C. Radioimmunoassay of oxytocin. J Endocrinol 1978; 76: 261– 270. 627820 source

      Not checked in the free run.

    • cites [95] Schams D. Oxytocin determination by radioimmunoassay. III. Improvement to subpicogram sensitivity and application to blood levels in cyclic cattle. Acta Endocrinol (Copenh) 1983; 103: 180– 183. 6683055

      Not checked in the free run.

    • cites [96] Chiodera P, Louis F, Legros JJ. Simultaneous radioimmunoassay for plasma arginine‐vasopressin and oxytocin using DEAE Sephadex A 25 extraction. J Endocrinol Invest 1984; 7: 287– 293. 6501802 source

      Not checked in the free run.

    • cites [97] Sheldrick EL, Flint AP. Circulating concentrations of oxytocin during the estrous cycle and early pregnancy in sheep. Prostaglandins 1981; 22: 631– 636. 7323312 source

      Not checked in the free run.

    • cites [98] Landgraf R. Plasma oxytocin concentrations in man after different routes of administration of synthetic oxytocin. Exp Clin Endocrinol 1985; 85: 245– 248. 4018162 source

      Not checked in the free run.

    • cites [99] Burd JM, Weightman DR, Baylis PH. Solid phase radioimmunoassay for direct measurement of human plasma oxytocin. J Immunoassay 1985; 6: 227– 243. 3900138 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationcitation

    When oxytocin is incubated in late-pregnancy plasma, there is a rapid loss of biological activity but after 2 h oxytocin immunoreactivity exceeds bioactivity by three-fold; the explanation is that oxytocinase degrades labelled oxytocin, reducing label binding to antibody and leading to overestimation.

    “When oxytocin is incubated in this plasma, there is a rapid loss of biological activity but, after 2 h, oxytocin ‘immunoreactivity’ exceeds bioactivity by three‐fold. The obvious explanation is that oxytocinase in the sample also degrades the labelled oxytocin. The consequence will be that less label will bind to antibody, leading to overestimation of the amount of oxytocin 76.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

    • Rests on an assumption: When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.
  • May combine two claims.

    Premisecitation

    In 1970, Robertson et al. described one of the first radioimmunoassays to measure plasma vasopressin in man; it was sensitive to 1 pg/ml (0.3 mU/ml) but when whole plasma was assayed, apparent concentrations were several hundredfold higher than anticipated, failed to fluctuate appropriately in response to physiological manipulation or disease states, and were not reduced substantially by dialysis.

    “This assay was sensitive to 1 pg/ml (0.3 mU/ml) but, when whole plasma was assayed, apparent concentrations were ‘several hundredfold higher than anticipated, failed to fluctuate appropriately in response to physiological manipulation or in disease states, and were not reduced substantially by dialysis of the plasma under conditions that completely removed vasopressin’.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    An antibody might recognise elements unrelated to the antigen against which it was raised, and unrelated molecules might cross-react.

    “In some cases, an antibody might recognise elements unrelated to the antigen against which it was raised. Antibodies are sensitive to differences in amino acid sequence, although they recognise the structure of a molecule, not its amino acid sequence per se, and so unrelated molecules might cross‐react 76, 79.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

    • cites [79] Martin MJ, Chard T, Landon J. The development of a radioimmunoassay for bovine neurophysin. J Endocrinol 1972; 52: 481– 495. 4336118 source

      Not checked in the free run.

  • May combine two claims.

    Premisenone given

    At that time, seven other vasopressin radioimmunoassays, using different antibodies and extraction procedures, all reported plasma concentrations of 1-10 pg/ml, and all showed that concentrations changed with physiological conditions in the manner expected from bioassays.

    “At that time, seven other vasopressin radioimmunoassays had been described, using different antibodies and different extraction procedures. All had reported plasma concentrations of 1–10 pg/ml, and all showed that concentrations changed with physiological conditions in the manner expected from bioassays.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation

    Baylis and Heath emphasised that even small increases in protein concentration lower binding and careful control of extract protein concentration is needed to avoid misinterpretation that a fall in binding is due to immunoreactive vasopressin.

    “When describing their vasopressin assay, Baylis and Heath 91 emphasised that ‘even small increases in protein concentration lower binding and therefore careful control of extract protein concentration is needed to avoid the misinterpretation that a fall in binding is due to immunoreactive vasopressin’.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [91] Baylis PH, Heath DA. The development of a radioimmunoassay for the measurement of human plasma arginine vasopressin. Clin Endocrinol (Oxf) 1977; 7: 91– 102. 891002 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Antibodies do not recognise the whole molecule, so an immunoassay may overestimate hormone concentrations by measuring fragments of no physiological significance.

    “Antibodies do not recognise the whole molecule, and so an immunoassay may overestimate hormone concentrations by measuring fragments of no physiological significance 76.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    With few exceptions, assays in extracted plasma reported basal oxytocin levels below 5 pg/ml in man; of 17 studies reviewed by Szeto et al., only two reported higher levels, the highest being 71 pg/ml in samples from the corpus cavernosum of the penis.

    “With few exceptions, assays in extracted plasma reported basal levels of < 5 pg/ml in man: of 17 studies reviewed by Szeto et al. 101, only two reported higher levels; the highest (71 pg/ml) were in samples taken from the corpus cavernosum of the penis, a circumstance that might be expected to affect hormone secretion.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Szeto et al. 101

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Henneberry et al. assayed unextracted samples using standard curves generated in charcoal-filtered plasma and reported basal vasopressin levels of 1.8 pg/ml, rising to 6.3 pg/ml after a saline infusion that raised plasma osmotic pressure from 280 to 293 mOsm/l.

    “Henneberry et al. 93 took a different approach: they assayed unextracted samples using standard curves generated in charcoal‐filtered plasma, and reported basal levels of 1.8 pg/ml, rising to 6.3 pg/ml after a saline infusion that raised plasma osmotic pressure from 280 to 293 mOsm/l.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Henneberry et al. 93

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given

    High salt concentrations have a nonlinear effect, and serial dilutions of samples with a high salt content produce measurements that do not parallel the standard curve.

    “High salt concentrations have a nonlinear effect, and serial dilutions of samples with a high salt content produce measurements that do not parallel the standard curve.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning

    To sustain a plasma concentration of 350 pg/ml in man, oxytocin must be produced at 40,000 times the rate measured in the rat; given a difference in pituitary mass and oxytocin content of just 40-fold, this seems beyond plausibility.

    “Thus, to sustain a plasma concentration of 350 pg/ml in man, oxytocin must be produced at 40 000 times the rate measured in the rat. Given a difference in pituitary mass and oxytocin content of just 40‐fold, this seems beyond plausibility.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    In rats, Ginsburg and Heller estimated that the initial half-life of exogenously administered vasopressin was less than 1 min; from experiments in acutely nephrectomised rats and by ligating arteries to reduce liver blood flow, they concluded that most vasopressin was cleared by the kidneys and liver.

    “In rats, Ginsburg and Heller 104 estimated that the initial half‐life of exogenously administered vasopressin was less than 1 min. From experiments in acutely nephrectomised rats, and by ligating arteries to reduce blood flow through the liver, it was concluded that most of the vasopressin was cleared by the kidneys and liver.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Ginsburg and Heller 104

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Claims that the Enzo ELISA is a valid way of measuring oxytocin in unextracted plasma do not rest on the manufacturer's advice, which strongly recommends extraction, but instead converge on two papers, neither of which actually investigated human plasma.

    “Claims that the Enzo ELISA is a valid way of measuring oxytocin in unextracted plasma do not rest on manufacturer's advice, which, as we have noted, strongly recommends extraction, but, instead, converge on two papers, neither of which actually investigated human plasma at all.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Ginsburg and Smith found that oxytocin added to rat plasma in vitro at 37 °C was relatively stable, losing just 6-13% of biological activity after 40 min, whereas when oxytocin was injected into rats its biological activity disappeared with a half-life of 1.65 min.

    “Ginsburg and Smith 49, using a rat uterus bioassay, found that, when oxytocin was added to rat plasma in vitro at 37 °C, it was relatively stable, losing just 6–13% of biological activity after 40 min of incubation. By contrast, when oxytocin was injected into rats, the biological activity disappeared with a half‐life of 1.65 min.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Ginsburg and Smith 49

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In dogs, Ang and Jenkins injected radiolabelled vasopressin and used thin-layer chromatography and autoradiography to quantify labelled vasopressin in plasma; it disappeared with an initial half-life of 3 min and a later half-life of 13 min.

    “In dogs, rather than measure vasopressin by either radioimmunoassay or bioassay, Ang and Jenkins 106 injected radiolabelled vasopressin and used thin‐layer chromatography and autoradiography to quantify labelled vasopressin in plasma samples: it disappeared with an initial half‐life of 3 min and a later half‐life of 13 min.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Ang and Jenkins 106

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In patients with chronic renal failure, vasopressin clearance is much slower, with metabolic clearance rate reduced by approximately half, consistent with the conclusion that vasopressin is mainly cleared from the circulation through the kidneys.

    “In patients with chronic renal failure, the clearance is much slower: MCR is reduced by approximately half 108, which is consistent with the conclusion that vasopressin is mainly cleared from the circulation through the kidneys.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [108] Argent NB, Wilkinson R, Baylis PH. Metabolic clearance rate of arginine vasopressin in severe chronic renal failure. Clin Sci (Lond) 1992; 83: 583– 587. 1335395 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Ultrafiltration of plasma indicated that approximately 30% of similarly infused vasopressin was bound to large molecules, whereas there was no apparent binding of oxytocin; in vitro oxytocin bioactivity was stable, with only approximately 20% lost after 24 h at 37 °C.

    “Ultrafiltration of the plasma indicated that, whereas approximately 30% of similarly infused vasopressin was bound to large molecules, there was no apparent binding of oxytocin. In vitro, oxytocin bioactivity was stable, as found in rats 49; only approximately 20% was lost after 24 h at 37 °C.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [49] Ginsburg M, Smith MW. The fate of oxytocin in male and female rats. Br J Pharmacol Chemother 1959; 14: 327– 333. 13850241 PMC1481739 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    The reported mean oxytocin and vasopressin concentrations of typically 100-500 pg/ml in unextracted human plasma are inconsistent by two orders of magnitude with bioassays and with radioimmunoassay measurements on extracted samples, and are also inconsistent with the known sensitivity of peripheral targets and with what is known of the production, storage and clearance of these hormones.

    “These values are inconsistent, by two orders of magnitude, with both bioassays of oxytocin and vasopressin and with radioimmunoassay measurements on extracted samples. They are also inconsistent with the known sensitivity of peripheral targets to oxytocin and vasopressin, and with what is known of the production, storage and clearance of these hormones.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    In nephrectomised rats, oxytocin bioactivity declined with a half-life of approximately 3 min in the first 7 min and much more slowly thereafter; when the splanchnic vessels were also ligated, the initial decline was not further changed but after 7 min oxytocin concentrations stuck at a high, steady level.

    “In nephrectomised rats, bioactivity declined with a half‐life of approximately 3 min in the first 7 min, and much more slowly thereafter. When the splanchnic vessels were also ligated, the initial decline was not further changed but, after 7 min, oxytocin concentrations ‘stuck’ at a high, steady level.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Ginsburg and Smith 49

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Ginsburg and Smith deduced that when oxytocin enters the circulation it is rapidly removed by the kidneys and liver, and that without these organs injected oxytocin diffuses within approximately 7 min throughout the extracellular fluid, where it is degraded only very slowly.

    “Ginsburg and Smith deduced that, when oxytocin enters the circulation, it is rapidly removed by the kidneys and liver. Without these organs, injected oxytocin diffuses within approximately 7 min throughout the extracellular fluid, where it is degraded only very slowly.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Zhang et al. recognised that one common challenge in biomarker analysis is to select an appropriate surrogate matrix to prepare the standard curve, free of target analytes and identical or similar to unknown sample matrix with regard to ionisation effects and extraction recovery.

    “As Zhang et al. 125 recognised, ‘One common challenge in biomarker analysis is to select an appropriate surrogate matrix to prepare the standard curve, which is free of target analytes and is identical or similar to unknown sample matrix with regard to ionisation effects and extraction recovery’.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Zhang et al. 125

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In man, Fabian et al. measured oxytocin with a mammary gland bioassay and found that after the end of an i.v. infusion bioactivity disappeared with a half-life of 4.8 min and a distribution volume of 10.5 ml/100 g, close to the extracellular fluid volume, whereas an acute large injection disappeared with a half-life of 3.2 min and a distribution volume of 7.4 ml/100 g.

    “In man, Fabian et al. 43 measured oxytocin with a mammary gland bioassay and studied its clearance after the end of an i.v. infusion. Bioactivity disappeared with a half‐life of 4.8 min with a distribution volume of 10.5 ml/100 g, which is close to the extracellular fluid volume, whereas an acute large injection disappeared with a half‐life of 3.2 min and a distribution volume of 7.4 ml/100 g.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Fabian et al. 43

      Not checked in the free run.

  • May combine two claims. The quote only partly states this claim.

    Premisecitation

    Three studies compared measurements of oxytocin in the same samples by the same ELISA with and without extraction; Robinson et al. found approximately 8 pg/ml in extracted seal plasma versus 300 pg/ml in unextracted heparin samples or 540 pg/ml in unextracted EDTA samples with no correlation, Christensen et al. found radioimmunoassay extracted mean levels of 1.1 pg/ml and spiked levels higher by t

    “Three studies have compared measurements of oxytocin in the same samples by the same ELISA, with and without extraction 101, 116, 117. Robinson et al. 116 measured oxytocin levels in seals: in extracted samples, they found plasma levels of approximately 8 pg/ml, compared to 300 pg/ml in unextracted samples collected in heparin vacutainers or 540 pg/ml in unextracted samples collected in ethylenedi”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites [101] Szeto A, McCabe PM, Nation DA, Tabak BA, Rossetti MA, McCullough ME, Schneiderman N, Mendez AJ. Evaluation of enzyme immunoassay and radioimmunoassay methods for the measurement of plasma oxytocin. Psychosom Med 2011; 73: 393– 400102. 21636661 PMC3118424 source

      Not checked in the free run.

    • cites [116] Robinson KJ, Hazon N, Lonergan M, Pomeroy PP. Validation of an enzyme‐linked immunoassay (ELISA) for plasma oxytocin in a novel mammal species reveals potential errors induced by sampling procedure. J Neurosci Meth 2014; 226: 73– 79. 24485867 source

      Not checked in the free run.

    • cites [117] Christensen JC, Shiyanov PA, Estepp JR, Schlager JJ. Lack of association between human plasma oxytocin and interpersonal trust in a Prisoner's Dilemma paradigm. PLoS ONE 2014; 9: e116172. 25549255 PMC4280178 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationnone given

    As a proportion of body weight, the rat pituitary contains much more oxytocin and vasopressin than the human pituitary, and in rats and other small mammals circulating levels are higher; in man and other large animals lower circulating concentrations are compensated for by higher sensitivity of peripheral targets.

    “As a proportion of body weight, the rat pituitary contains much more oxytocin and vasopressin than the human pituitary and, in rats, as in other small mammals, circulating levels are higher. In man and other large animals, lower circulating concentrations are compensated for by higher sensitivity of the peripheral targets.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    In rabbits, vasopressin was cleared with a half-life of approximately 4.5 min, mainly by the kidneys and liver.

    “In similar studies on rabbits, vasopressin was cleared with a half‐life of approximately 4.5 min, again mainly by the kidneys and liver 105.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [105] Hasan SH. The clearance of vasopressin and deaminovasopressin from the circulation of normal and hypercalcaemic rabbits. J Endocrinol 1970; 47: 135– 136. 5428912 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Plasma oxytocin was measured using human plasma diluted with water (1:6) as a surrogate matrix, and basal levels of 2.4 pg/ml for human but 298 pg/ml for rat were reported; these high levels in the rat may indicate that human plasma is not a suitable surrogate matrix for measurements in rat.

    “Plasma oxytocin was measured using human plasma diluted with water (1 : 6) as a surrogate matrix, and basal levels of 2.4 pg/ml for human but 298 pg/ml for rat were reported; these high levels in the rat may indicate that human plasma is not a suitable surrogate matrix for measurements in rat.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [125] Zhang G, Zhang Y, Fast DM, Lin Z, Steenwyk R. Ultra sensitive quantitation of endogenous oxytocin in rat and human plasma using a two‐dimensional liquid chromatography‐tandem mass spectrometry assay. Anal Biochem 2011; 416: 45– 52. 21609710 source

      Not checked in the free run.

  • Classified as probably not a substantial claim. May combine two claims.

    Premisecitation

    The Enzo ELISAs draw attention to plasma matrix interference, illustrate the benefits of plasma extraction, and strongly recommend that users extract plasma samples; they state that dilution might minimise interference but warn that users must verify that dilutions are appropriate for their samples.

    “The ELISAs in most widespread use are marketed by Enzo, who draw attention to problems of plasma matrix interference, illustrate the benefits of plasma extraction, and ‘strongly recommend’ that users extract plasma samples. They state that it might be possible to dilute plasma samples enough to minimise interference (suggesting at least an eight‐fold dilution) but warn that the user ‘must verify’”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [113] Designs A. Oxytocin Enzyme Immunoassay (EIA) kit: Manufacturer's instructions. Available at: http://www.enzolifesciences.com/fileadmin/redacteur/pdf/adi/ADI-900-153.pdf.

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Carter et al. reported that in voles the ELISA measured spiked levels of oxytocin with a gain of 0.79, consistent with a reduction in antibody affinity although not sufficient to account for the discrepancy; in HPLC fractions apparent immunoreactivity appeared only in fractions that eluted with oxytocin, suggesting no plasma interference, but the same method had previously been used by Wismer

    “Carter et al. 119 further reported that, in voles, the ELISA measured spiked levels of oxytocin with a gain of 0.79. This is consistent with a reduction in antibody affinity, although not sufficient to account for the discrepancy between expected and measured oxytocin in Kramer's measurements. In fractions separated by high‐performance liquid chromatography (HPLC), apparent immunoreactivity”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Carter et al. 119

      Not checked in the free run.

    • cites Wismer Fries et al. 120

      Not checked in the free run.

    • cites Anderson 121

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Lauson and Bacanegra concluded that at physiological levels below 25 pg/ml most plasma vasopressin is bound to protein and cleared by tubular cells of the kidney by enzymatic inactivation and secretion, while at higher concentrations more vasopressin is unbound and some is filtered and excreted into urine.

    “Lauson and Bacanegra 103 concluded that, at physiological levels (< 25 pg/ml), most of the vasopressin in the plasma is bound to protein and is cleared by the tubular cells of the kidney by enzymatic inactivation and secretion. At higher concentrations, more of the vasopressin is unbound, and some of this is filtered at the glomerulus and excreted into urine.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Lauson and Bacanegra 103

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Binding of plasma proteins to an immunoglobulin can hinder its ability to bind an antigen, reducing its affinity, and/or may completely prevent some of the antibody from binding to antigen.

    “The binding of plasma proteins to an immunoglobulin can hinder its ability to bind an antigen, reducing its affinity, and/or may completely prevent some of the antibody from binding to antigen.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Bjerner et al. 111

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Interpretationreasoning

    A plasma concentration of 350 pg/ml oxytocin in man implies a total blood content of approximately 1 μg given a plasma volume of 3 l, and for a half-life of 10 min implies a secretion and production rate of 500 ng/min, or 720 IU/day.

    “A plasma concentration of 350 pg/ml oxytocin in man implies a total blood content of approximately 1 μg (given a plasma volume of 3 l) and, for a half‐life of 10 min, this implies a secretion (and production) rate of 500 ng/min, or 720 IU/day.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    In the baboon, oxytocin clearance was consistent with a two-compartmental system corresponding to the plasma volume and the extravascular fluid volume; the initial half-life in the first phase of distribution was approximately 1 min, followed by a terminal elimination half-life of approximately 10 min.

    “In the baboon, oxytocin clearance was also consistent with a two‐compartmental system, corresponding to the plasma volume and the extravascular fluid volume. The initial half‐life in the first phase of distribution was approximately 1 min, followed by a terminal elimination half‐life of approximately 10 min 109.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [109] Kowalski WB, Diveky L, Mehendale R, Parsons M, Wilson L Jr. Effect of pregnancy on the metabolic clearance rate and the volume of distribution of oxytocin in the baboon. Am J Physiol 1998; 274: E791– E795. 9612235 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    The manufacturer's advice to extract plasma has been largely ignored; many studies have used the Enzo assay on unextracted human plasma, reporting mean concentrations of oxytocin and vasopressin typically 100-500 pg/ml, and those that diluted samples did so by less than the manufacturer's recommendations without confirming that their dilution was adequate.

    “This advice has been largely ignored. Many studies have used this assay on unextracted human plasma, reporting mean concentrations of oxytocin and vasopressin of typically 100–500 pg/ml 3, 4, 5, 6, 114. Those that have diluted samples have done so by less than the manufacturer's recommendations, without confirming that their dilution is adequate.”
    Quote in the paperSubstantial claim0 of 5 sources checked
    • cites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source

      Not checked in the free run.

    • cites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source

      Not checked in the free run.

    • cites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source

      Not checked in the free run.

    • cites [6] Rubin LH, Carter CS, Bishop JR, Pournajafi‐Nazarloo H, Drogos LL, Hill SK, Ruocco AC, Keedy SK, Reilly JL, Keshavan MS, Pearlson GD, Tamminga CA, Gershon ES, Sweeney JA. Reduced levels of vasopressin and reduced behavioral modulation of oxytocin in psychotic disorders. Schizophr Bull 2014; 40: 1374– 1384. 24619535 PMC4193712 source

      Not checked in the free run.

    • cites [114] McCullough ME, Churchland PS, Mendez AJ. Problems with measuring peripheral oxytocin: can the data on oxytocin and human behavior be trusted? Neurosci Biobehav Rev 2013; 37: 1485– 1492. 23665533 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    With a reduction in maximum binding, an assay will report erroneously high values even in the absence of antigen and will exaggerate measurements of added antigen; with a reduction in affinity, the assay will under-report added antigen; in both cases there is a linear relationship between amount added and amount measured, with slope depending on the nature of the interference.

    “With a reduction in maximum binding, an assay will report erroneously high values even in the absence of any antigen, and will exaggerate measurements of added antigen. The effect of a reduction in affinity is different; in this case, the assay will under‐report added antigen. In both cases, there will be a linear relationship between the amount of oxytocin added and the amount measured, although”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given

    Bolus injections of oxytocin and vasopressin disappear rapidly, with an initial half-life that varies between species and with physiological state: in the rat both hormones disappear with an initial half-life of approximately 2 min, and in man oxytocin has a half-life of 4-10 min except in pregnancy when oxytocinase shortens it to 2-3 min.

    “Bolus injections disappear rapidly, with an initial half‐life that varies between species and with physiological state: in the rat, both hormones disappear with an initial half‐life of approximately 2 min; in man, oxytocin has a half‐life of 4–10 min, except in pregnancy, when oxytocinase shortens this to 2–3 min.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Later radioimmunoassay studies indicated that vasopressin disappeared from the circulation with a bi-exponential decay: the fast phase had a half-life of 5.4 min and a distribution volume close to the plasma volume, and the slower phase had a half-life of 78 min with a distribution volume close to the extracellular fluid volume.

    “Later studies using radioimmunoassay indicated that vasopressin disappeared from the circulation with a bi‐exponential decay 107. The fast phase had a half‐life of 5.4 min and a distribution volume close to the plasma volume; the slower phase had a half‐life of 78 min with a distribution volume close to the extracellular fluid volume.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [107] Benmansour M, Rainfray M, Paillard F, Ardaillou R. Metabolic clearance rate of immunoreactive vasopressin in man. Eur J Clin Invest 1982; 12: 475– 480. 6818035 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Carter et al. claimed a biological validation for the ELISA in reporting that breastfeeding in women lowered salivary concentrations of oxytocin; because suckling-induced oxytocin release is indispensable for milk let-down in mammals, it is difficult to understand how such an anomalous observation qualifies as a validation.

    “Finally, Carter et al. 119 claimed a biological validation for the ELISA in reporting that breastfeeding in women lowered salivary concentrations of oxytocin. Because suckling‐induced oxytocin release is indispensable for milk let‐down in mammals 123, it is difficult to understand how such an anomalous observation qualifies as a validation.”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Carter et al. 119

      Not checked in the free run.

    • cites [123] Russell JA, Leng G, Douglas AJ. The magnocellular oxytocin system, the fount of maternity: adaptations in pregnancy. Front Neuroendocrinol 2003; 24: 27– 61. 12609499 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Interpretationnone given

    Plasma matrix proteins interact differently with different antibodies: there is no single radioimmunoassay and no single ELISA, and their susceptibility to plasma interference will vary with the antibody.

    “Plasma matrix proteins interact differently with different antibodies: there is no single ‘radioimmunoassay’ and no single ‘ELISA’, and their susceptibility to plasma interference will vary with the antibody.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given

    In mice or rats, validation of vasopressin or oxytocin assays could be supported by measurements in gene knockout animals or hypophysectomised animals, where vasopressin and/or oxytocin are completely absent.

    “In mice or rats, validation could be supported by measurements in gene knockout animals or hypophysectomised animals, where vasopressin and/or oxytocin are completely absent.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning

    Measuring oxytocin or vasopressin in cerebrospinal fluid, urine or saliva should have fewer problems because of the relative absence of proteins, but in urine high salt concentrations may cause problems.

    “Measuring oxytocin or vasopressin in cerebrospinal fluid, urine or saliva should have fewer problems because of the relative absence of proteins but, in urine, high salt concentrations may cause problems.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Interpretationreasoning

    Plasma proteins differ from individual to individual and differ substantially from species to species, so any assay in unextracted plasma should be validated for the species in which it is measured and ideally in the individual in which it is measured.

    “Plasma proteins differ from individual to individual and differ substantially from species to species, and so any assay in unextracted plasma should be validated for the species in which it is measured and, ideally, in the individual in which it is measured.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Plasma proteins differ from individual to individual and differ substantially from species to species.
  • May combine two claims.

    Interpretationreasoning

    If the high levels arise from plasma protein interference, then these assays are, for unextracted plasma, no more than a random number generator, returning arbitrary values with a high variance and a skewed distribution.

    “If so, these assays are, for unextracted plasma, no more than a random number generator, returning arbitrary values with a high variance and a skewed distribution.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationcitation

    Levels from an individual could be verified by constructing a standard curve in hormone-stripped plasma from that individual, although it would be impractical to do this for every individual sampled.

    “Levels from an individual could be verified by constructing a standard curve in hormone‐stripped plasma from that individual 93, although it would be impractical to do this for every individual sampled.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [93] Henneberry HP, Slater JD, Eisen V, Fuhr S. Arginine vasopressin response to hypertonicity in hypertension studied by arginine vasopressin assay in unextracted plasma. J Hypertens 1992; 10: 221– 228. 1315818 source

      Not checked in the free run.

  • Classified as probably not a substantial claim.

    Premisenone given

    The current impetus for measuring vasopressin and oxytocin in plasma derives from the assumption that such measurements are informative of the release of these peptides in the brain.

    “The current impetus for measuring vasopressin and oxytocin in plasma derives from the assumption that such measurements are informative of the release of these peptides in the brain.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Interpretationreasoning

    There are no known conditions in which oxytocin is absent or grossly elevated in man; however, given the sensitivity of peripheral targets in man to very low concentrations, levels exceeding 10 pg/ml should be treated with scepticism.

    “There are no known conditions in which oxytocin is absent or grossly elevated in man. However, given the sensitivity of peripheral targets in man to very low concentrations, levels exceeding 10 pg/ml should be treated with scepticism.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: There are no known conditions in which oxytocin is absent or grossly elevated in man.
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationcitation

    Analytical verification by HPLC with chromatographic separation can confirm that measurements reflect the actual peptide, although care is needed to ensure that all high molecular weight proteins are eluted from the column and are checked for assay interference; there should be quantitative agreement between the amount of hormone in the appropriate fraction and the level reported by the assay.

    “Analytical verification by HPLC with chromatographic separation 78 can confirm that measurements reflect the actual peptide, although care is needed to ensure that all high molecular weight proteins are eluted from the column and are checked for assay interference. There should be quantitative agreement between the amount of hormone in the appropriate fraction, and the level reported by the assay.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Interference will be apparent if, when samples are spiked with known amounts of hormone, the assay recovers amounts different to the expected amount; however, the absence of a difference is not conclusive because interference that reduces antibody affinity will lead to underestimation of added hormone, whereas interference that blocks antibody recognition will lead to overestimation, and both

    “Such interference will be apparent if, when samples are spiked with known amounts of hormone, the assay recovers amounts different to the expected amount. However, the absence of a difference is not conclusive because interference that reduces antibody affinity will lead to underestimation of added hormone, whereas interference that blocks antibody recognition will lead to overestimation, and both”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    Samples with high levels will not dilute in parallel to the standard curve if the levels arise from factors such as high salt levels or pH that interfere nonlinearly with antibody binding, although this test may not recognise protein interference that blocks antibody recognition of the antigen.

    “Samples with high levels will not dilute in parallel to the standard curve if the levels arise from factors such as high salt levels or pH that interfere nonlinearly with antibody binding, although this test may not recognise protein interference that blocks antibody recognition of the antigen.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisecitation

    The fallacies in the supposition that plasma measurements are informative of brain peptide release have been discussed elsewhere.

    “The fallacies in this supposition have been discussed elsewhere 2.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [2] Leng G, Ludwig M. Intranasal oxytocin: myths and delusions. Biol Psychiatry 2016; 79: 243– 250. 26049207 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationcitation

    In a steady state, unbound oxytocin and vasopressin should distribute in extravascular fluids, and measurements in saliva might be useful for estimating steady-state secretion.

    “In a steady‐state, unbound oxytocin and vasopressin should distribute in extravascular fluids, and measurements in saliva might be useful for estimating steady‐state secretion 119, 130, 131.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites [119] Carter CS, Pournajafi‐Nazarloo H, Kramer KM, Ziegler TE, White‐Traut R, Bello D, Schwertz D. Oxytocin: behavioral associations and potential as a salivary biomarker. Ann N Y Acad Sci 2007; 1098: 312– 322. 17435137 source

      Not checked in the free run.

    • cites [130] Jong TR, Menon R, Bludau A, Grund T, Biermeier V, Klampfl SM, Jurek B, Bosch OJ, Hellhammer J, Neumann ID. Salivary oxytocin concentrations in response to running, sexual self‐stimulation, breastfeeding and the TSST: the Regensburg Oxytocin Challenge (ROC) study. Psychoneuroendocrinology 2015; 62: 381– 388. 26385109 source

      Not checked in the free run.

    • cites [131] Frijling JL, van Zuiden M, Nawijn L, Koch SB, Neumann ID, Veltman DJ, Olff M. Salivary oxytocin and vasopressin levels in police officers with and without post‐traumatic stress disorder. J Neuroendocrinol 2015; 27: 743– 751. 26184739 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Tsukazaki et al. developed an LC-MS/MS assay for vasopressin and showed that, correcting for plasma matrix effects, this assay gave close quantitative agreement with radioimmunoassay results from extracted human plasma samples, and had higher sensitivity and needed smaller sample volumes.

    “Tsukazaki et al. 127 developed an LC‐MS/MS assay for vasopressin, and showed that, correcting for plasma matrix effects, this assay gave close quantitative agreement with radioimmunoassay results from extracted human plasma samples, and had higher sensitivity and needed smaller sample volumes.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Tsukazaki et al. 127

      Not checked in the free run.

74 of 74
  • The paper relies on this without support.

    Antibodies are generally raised to the tail end of oxytocin, which confers its biological activity.

    “Generally, antibodies are raised to the tail end of oxytocin, which confers its biological activity.”
  • The paper relies on this without support.

    When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.

    “When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.”
  • The paper relies on this without support.

    There are no known conditions in which oxytocin is absent or grossly elevated in man.

    “There are no known conditions in which oxytocin is absent or grossly elevated in man.”
  • The paper relies on this without support.

    Plasma proteins differ from individual to individual and differ substantially from species to species.

    “Plasma proteins differ from individual to individual and differ substantially from species to species”
  • The paper relies on this without support.

    Oxytocin injected subcutaneously is expected to distribute throughout the extracellular fluid volume.

    “Oxytocin injected s.c. is expected to distribute throughout the extracellular fluid volume”
5 of 5

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Daily oxytocin production is inconsistent with the stated infusion rate

    Major

    At the paper's stated potency of 600 IU/mg, 500 ng/min amounts to 0.72 mg/day, or 432 IU/day-not 720 IU/day. The erroneous daily value underlies the subsequent comparison with production in rats.

  • Vasopressin assay sensitivity uses the wrong unit scale

    Minor

    The paper elsewhere equates 4 μU with 10 pg of vasopressin, so 1 pg corresponds to about 0.4 μU-not 0.3 mU (300 μU). The stated sensitivity is therefore hundreds of times larger than its stated equivalent.

2 of 2

Cites

Flagged first, then by citations

5 of the 139 references couldn’t be matched to a known paper.

Research works printed but not linked: 1

  1. Heller J, 1959, [Significance of a new method of titration of antidiuretic hormone], Cesk Fysiol, 8, 194

Cited by

Sorted by citations

Authors
Leng, G., & Sabatier, N.
Affiliations
University of Edinburgh (GB)
Corresponding author
Leng, G.
Venue
Journal of Neuroendocrinology
Publisher
Wiley
Published
2016-09-23
Volume
28
Issue
10
ISSN
  • 0953-8194 (ISSN-L)
  • 1365-2826
OpenAlex
W2498209759
Cited by
237
References
139
Citations by year
2026: 14 · 2025: 22 · 2024: 29 · 2023: 21 · 2022: 16 · 2021: 36 · 2020: 22 · 2019: 24 · 2018: 29 · 2017: 21 · 2016: 3
FWCI
4.04
Type
journal article
Language
en
Open access
Locations
  • Journal of Neuroendocrinology (publishedVersion, cc-by-nc-nd)
  • PubMed Central (publishedVersion, cc-by-nc-nd)
  • PubMed (publishedVersion)
  • Edinburgh Research Explorer (University of Edinburgh) (submittedVersion, cc-by-nc-nd)
Funders
Seventh Framework Programme (607310)
Topics
  • Social Sciences › Psychology › Social Psychology › Neuroendocrine regulation and behavior
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuropeptides and Animal Physiology
Keywords
oxytocin, vasopressin, plasma hormone measurement, immunoassay interference, bioassay, enzyme-linked immunoassay, radioimmunoassay, plasma extraction, method validation
MeSH terms
Animals, Biological Assay, Humans, Immunoassay, Oxytocin, Vasopressins, Reproducibility of Results

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-08.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-08.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review