CoChat

Electric field estimation of deep transcranial magnetic stimulation clinically used for the treatment of neuropsychiatric disorders in anatomical head models

Parazzini, Marta, Fiocchi, Serena, Chiaramello, Emma, Roth, Yiftach, Zangen, Abraham, Ravazzani, Paolo · 2017 · Medical Engineering & Physics
Cited by 39

The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.

Abstract

Literature studies showed the ability to treat neuropsychiatric disorders using H1 coil, developed for the deep Transcranial Magnetic Stimulation (dTMS). Despite the positive results of the clinical studies, the electric field (E) distributions inside the brain induced by this coil when it is positioned on the scalp according to the clinical studies themselves are not yet precisely estimated. This study aims to characterize the E distributions due to the H1 coil in the brain of two realistic human models by computational electromagnetic techniques and to compare them with the ones due to the figure-of-8 coil, traditionally used in TMS and positioned as such to simulate the clinical experiments. Despite inter-individual differences, our results show that the dorsolateral prefrontal cortex is the region preferentially stimulated by both H1 and figure-of-8 coil when they are placed in the position on the scalp according to the clinical studies, with a more broad and non-focal distribution in the case of H1 coil. Moreover, the H1 coil spreads more than the figure-of-8 coil both in the prefrontal cortex and medial prefrontal cortex and towards some deeper brain structures and it is characterized by a higher penetration depth in the frontal lobe. This work highlights the importance of the knowledge of the electric field distribution in the brain tissues to interpret the outcomes of the experimental studies and to optimize the treatments.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.medengphy.2017.02.003

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 8 references

Cited by

See all 39

Metadata

Authors
Parazzini, M., Fiocchi, S., Chiaramello, E., Roth, Y., Zangen, A., & Ravazzani, P.
Venue
Medical Engineering & Physics
Published
2017-02-21
Volume
43
Issue
1
Pages
30-38
ISSN
  • 1350-4533 (ISSN-L)
  • 1873-4030
OpenAlex
W2590808930
Cited by
39
References
8
FWCI
1.45
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Electric field estimation of deep transcranial magnetic stimulation clinically used for the treatment of neuropsychiatric disorders in anatomical head models. https://app.cochat.ai/papers/10.1016/j.medengphy.2017.02.003
The paper
Parazzini, M., Fiocchi, S., Chiaramello, E., Roth, Y., Zangen, A., & Ravazzani, P. (2017). Electric field estimation of deep transcranial magnetic stimulation clinically used for the treatment of neuropsychiatric disorders in anatomical head models. Medical Engineering & Physics. https://doi.org/10.1016/j.medengphy.2017.02.003

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.