Reducing the gradient artefact in simultaneous EEG-fMRI by adjusting the subject's axial position
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Large artefacts which compromise EEG data quality are generated when electroencephalography (EEG) and functional magnetic resonance imaging (fMRI) are carried out concurrently. The gradient artefact produced by the time-varying magnetic field gradients is the largest of these artefacts. Although average artefact correction (AAS) and related techniques can remove the majority of this artefact, the need to avoid amplifier saturation necessitates the use of a large dynamic range and strong low-pass filtering in EEG recording. Any intrinsic reduction in the gradient artefact amplitude would allow data with a higher bandwidth to be acquired without amplifier saturation, thus increasing the frequency range of neuronal activity that can be investigated using combined EEG-fMRI. Furthermore, gradient artefact correction methods assume a constant artefact morphology over time, so their performance is compromised by subject movement. Since the resulting, residual gradient artefacts can easily swamp signals from brain activity, any reduction in their amplitude would be highly advantageous for simultaneous EEG-fMRI studies. The aim of this work was to investigate whether adjustment of the subject's axial position in the MRI scanner can reduce the amplitude of the induced gradient artefact, before and after artefact correction using AAS. The variation in gradient artefact amplitude as a function of the subject's axial position was first investigated in six subjects by applying gradient pulses along the three Cartesian axes. The results of this study showed that a significant reduction in the gradient artefact magnitude can be achieved by shifting the subject axially by 4 cm towards the feet relative to the standard subject position (nasion at iso-centre). In a further study, the 4 cm shift was shown to produce a 40% reduction in the RMS amplitude (and a 31% reduction in the range) of the gradient artefact generated during the execution of a standard multi-slice, EPI sequence. By picking out signals occurring at harmonics of the slice acquisition frequency, it was also shown that the 4 cm shift led to a 36% reduction in the residual gradient artefact after AAS. Functional and anatomical MR data quality is not affected by the 4 cm shift, as the head remains in the homogeneous region of the static magnet field and gradients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Method for Removing Imaging Artifact from Continuous EEG Recorded during Functional MRI
Allen, P. J., Josephs, O., & Turner, R. (2000) · NeuroImage · doi:10.1006/nimg.2000.0599 · Cited by 1,301 · cited 13×
“The optimised recording position also produced a 36% reduction in the residual gradient artefact after AAS has been applied to EEG data recorded from human subjects executing cued foot movements that produced changes in head position representative of a typical fMRI experiment.”
InfluentialLaufs, H., Krakow, K., Sterzer, P., Eger, E., Beyerle, A., Salek-Haddadi, A., & Kleinschmidt, A. (2003) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1831638100 · Cited by 1,028 · cited 1×
“Applications of this technique are now far reaching, ranging from the study of brain networks associated with the resting state (Laufs et al. 2003) to the investigation of epileptic foci (Iannetti et al.”
Single-trial EEG–fMRI reveals the dynamics of cognitive function
Debener, S., Ullsperger, M., Siegel, M., & Engel, A. K. (2006) · Trends in Cognitive Sciences · doi:10.1016/j.tics.2006.09.010 · Cited by 428
Eichele, T., Specht, K., Moosmann, M., Jongsma, M. L. A., Quiroga, R. Q., Nordby, H., & Hugdahl, K. (2005) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0505508102 · Cited by 308
Cited by
Methods for Simultaneous EEG-fMRI: An Introductory Review
Huster, R. J., Debener, S., Eichele, T., & Herrmann, C. S. (2012) · The Journal of Neuroscience · doi:10.1523/jneurosci.0447-12.2012 · Cited by 308
Pisauro, M. A., Fouragnan, E., Retzler, C., & Philiastides, M. G. (2017) · Nature Communications · doi:10.1038/ncomms15808 · Cited by 208
EEG-Informed fMRI: A Review of Data Analysis Methods
Abreu, R., Leal, A., & Figueiredo, P. (2018) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2018.00029 · Cited by 171
“This model allowed the derivation of the optimal head orientation and position inside the MRI scanner in subsequent experiments, which was shown to significantly minimize the impact of MR gradients on the EEG recordings (Mullinger et al., 2011).”
MethodFouragnan, E., Queirazza, F., Retzler, C., Mullinger, K. J., & Philiastides, M. G. (2017) · Scientific Reports · doi:10.1038/s41598-017-04507-w · Cited by 61
“To minimize the influence of the GA at source, electrodes Fp1 and Fp2 were positioned at the scanner’s z = 0 position (i.e. the scanner’s isocenter), corresponding to a 4 cm shift of the head in the foot-head direction 75 .”
Self-citationVallat, R., Nicolas, A., & Ruby, P. (2020) · Sleep · doi:10.1093/sleep/zsaa116 · Cited by 34
Metadata
- Authors
- Mullinger, K. J., Yan, W. X., & Bowtell, R.
- Venue
- NeuroImage
- Publisher
- Elsevier BV
- Published
- 2011-02
- Volume
- 54
- Issue
- 3
- Pages
- 1942-1950
- ISSN
- 1053-8119 (ISSN-L)
- OpenAlex
- W2079034885
- Cited by
- 70
- References
- 4
- FWCI
- 1.81
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.