A joint ESTRO and AAPM guideline for development, clinical validation and reporting of artificial intelligence models in radiation therapy
Open accessAbstract
BACKGROUND AND PURPOSE: Artificial Intelligence (AI) models in radiation therapy are being developed with increasing pace. Despite this, the radiation therapy community has not widely adopted these models in clinical practice. A cohesive guideline on how to develop, report and clinically validate AI algorithms might help bridge this gap. METHODS AND MATERIALS: A Delphi process with all co-authors was followed to determine which topics should be addressed in this comprehensive guideline. Separate sections of the guideline, including Statements, were written by subgroups of the authors and discussed with the whole group at several meetings. Statements were formulated and scored as highly recommended or recommended. RESULTS: The following topics were found most relevant: Decision making, image analysis, volume segmentation, treatment planning, patient specific quality assurance of treatment delivery, adaptive treatment, outcome prediction, training, validation and testing of AI model parameters, model availability for others to verify, model quality assurance/updates and upgrades, ethics. Key references were given together with an outlook on current hurdles and possibilities to overcome these. 19 Statements were formulated. CONCLUSION: A cohesive guideline has been written which addresses main topics regarding AI in radiation therapy. It will help to guide development, as well as transparent and consistent reporting and validation of new AI tools and facilitate adoption.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Acceptance testing of AI auto contouring algorithms typically involves both area‐based metrics (e.g., Dice Similarity Coefficient, Intersection over Union (IoU)) and distance‐based metrics (e.g., Hausdorff Distance) evaluated against reference standard contours. 11 , 12 , 13 , 14 These metrics are well established for assessing contour agreement, but they say nothing about where the model is confi”
From Uncertainty quantification of U‐Net based segmentation tool using conformal prediction (2026) · cites this paper as [11]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Nature of Statistical Learning Theory
Vapnik, V. (2000) · Statistics for Engineering and Information Science · doi:10.1007/978-1-4757-3264-1 · Cited by 46,387
PROBAST: A Tool to Assess the Risk of Bias and Applicability of Prediction Model Studies
Wolff, R. F., Moons, K. G. M., Riley, R. D., Whiting, P. F., Westwood, M., Collins, G. S., Reitsma, J. B., Kleijnen, J., Mallett, S., & for the PROBAST Group† (2019) · Annals of Internal Medicine · doi:10.7326/m18-1376 · Cited by 2,984
Checklist for Artificial Intelligence in Medical Imaging (CLAIM): a guide for authors and reviewers
Mongan, J., Moy, L., & Kahn, C. E. (2020) · Radiology: Artificial Intelligence · doi:10.1148/ryai.2020200029 · Cited by 1,443
Vasey, B., Nagendran, M., Campbell, B., Clifton, D. A., Collins, G. S., Denaxas, S., Denniston, A. K., Faes, L., Geerts, B., Ibrahim, M., et al. (2022) · Nature Medicine · doi:10.1038/s41591-022-01772-9 · Cited by 1,039
Self-citationCruz Rivera, S., Liu, X., Chan, A.-W., Denniston, A. K., Calvert, M. J., The SPIRIT-AI and CONSORT-AI Working Group, SPIRIT-AI and CONSORT-AI Steering Group, Darzi, A., Holmes, C., Yau, C., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-1037-7 · Cited by 933
Collins, G. S., Dhiman, P., Andaur Navarro, C. L., Ma, J., Hooft, L., Reitsma, J. B., Logullo, P., Beam, A. L., Peng, L., Van Calster, B., et al. (2021) · BMJ Open · doi:10.1136/bmjopen-2020-048008 · Cited by 852
Reporting of artificial intelligence prediction models
Collins, G. S., & Moons, K. G. M. (2019) · The Lancet · doi:10.1016/s0140-6736(19)30037-6 · Cited by 701
Minimum information about clinical artificial intelligence modeling: the MI-CLAIM checklist.
Norgeot, B., Quer, G., Beaulieu-Jones, B. K., Torkamani, A., Dias, R., Gianfrancesco, M., Arnaout, R., Kohane, I. S., Saria, S., Topol, E., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-1041-y · Cited by 665
Vollmer, S. J., Mateen, B. A., Bohner, G., Király, F. J., Ghani, R., Jonsson, P. F., Cumbers, S., Jonas, A., McAllister, K., Myles, P. R., et al. (2020) · BMJ · doi:10.1136/bmj.l6927 · Cited by 574
Langendijk, J. A., Lambin, P., Ruysscher, D. D., Widder, J., Bos, M., & Verheij, M. (2013) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2013.05.007 · Cited by 559
Research works printed but not linked: 6
- Fda, 2019, Proposed regulatory framework for modifications to artificial intelligence/machine learning (AI/ML)- based software as a medical device (SaMD). Food and Drug, Administration
- Holmes, 2023, Evaluating Large Language Models on a Highly-specialized Topic, Radiation Oncology Physics. Front, Oncol, 17, 1219326
- Huang, 2021, Virtual Patient-Specific Quality Assurance of IMRT Using UNet++: Classification, Gamma Passing Rates Prediction, and Dose Difference Prediction. Front, Oncol, %20;11
- Liu, Z, Wang, P, Li, Y, Holmes, J, Shu, P, Zhang, L, Liu, C, Liu, N, Zhu, D, Li, X, Li, Q, Patel, S. H., Sio, T. T, Liu, T, Liu, W. RadOnc-GPT: A Large Language Model for Radiation Oncology. arXiv:2309.10160 webpage visited 6-11-2023. · arXiv
- Prunaretty, 2023, A multi-centric evaluation of self-learning GAN based pseudo-CT generation software for low field pelvic magnetic resonance imaging. Front, Oncol, 13, 1245054
- van de, 2021, Artificial intelligence based treatment planning of radiotherapy for locally advanced breast cancer. Phys Imaging, Radiat Oncol, 20, 111
Cited by
Sorted by citations
Fountzilas, E., Pearce, T. E., Baysal, M. A., Chakraborty, A., & Tsimberidou, A. M. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01471-y · Cited by 233
Mastella, E., Calderoni, F., Manco, L., Ferioli, M., Medoro, S., Turra, A., Giganti, M., & Stefanelli, A. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100731 · Cited by 31
Rusanov, B., Ebert, M. A., Sabet, M., Rowshanfarzad, P., Barry, N., Kendrick, J., Alkhatib, Z., Gill, S., Dass, J., Bucknell, N. W., et al. (2025) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-024-01513-x · Cited by 21
Giordano, A., Romero-Mascarell, C., González-Suárez, B., & Guarner‐Argente, C. (2025) · Digestive Diseases and Sciences · doi:10.1007/s10620-025-09099-4 · Cited by 13
Carriero, S., Cannella, R., Cicchetti, F., Angileri, S. A., Bruno, A., Biondetti, P., Colciago, R. R., D’Antonio, A., Pepa, G. D., Grassi, F., et al. (2025) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/1754-9485.13880 · Cited by 16
Artificial intelligence in radiation therapy: from imaging to delivery—a comprehensive review
Duke, K., & Papanikolaou, N. (2025) · BJR|Artificial Intelligence · doi:10.1093/bjrai/ubaf012 · Cited by 13
Wong, Y. M., Yeap, P. L., Ong, A. L. K., Tuan, J. K. L., Lew, W. S., Lee, J. C. L., & Tan, H. Q. (2024) · Intelligence-Based Medicine · doi:10.1016/j.ibmed.2024.100163 · Cited by 12
Poel, R., Rüfenacht, E., Scheib, S. G., Hemmatazad, H., Krcek, R., Tran, S., Romano, E., Rogers, S. J., Stieb, S., Poolakundan, M. R., et al. (2025) · Communications Medicine · doi:10.1038/s43856-025-01048-6 · Cited by 11
Konrad, M., Brink, C., Bertelsen, A. S., Lorenzen, E. L., Çelik, B., Nyborg, C. J., Dysager, L., Schytte, T., & Bernchou, U. (2024) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2024.100694 · Cited by 11
Finnegan, R., Quinn, A., Horsley, P. J., Chan, J. C. Y., Stewart, M., Bromley, R. M., & Booth, J. (2025) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70067 · Cited by 11
Descriptive overview of AI applications in x-ray imaging and radiotherapy
Damilakis, J. E., & Stratakis, J. (2024) · Journal of Radiological Protection · doi:10.1088/1361-6498/ad9f71 · Cited by 12
Chang, L., Li, H., Wu, W., Liu, X., Yan, J., Chen, Z., Wu, H., & Song, S. (2025) · Journal of Translational Medicine · doi:10.1186/s12967-025-07591-z · Cited by 10
Hoebers, F. J. P., Wee, L., Likitlersuang, J., Mak, R. H., Bitterman, D. S., Huang, Y., Dekker, A. L. A. J., Aerts, H. J. W. L., & Kann, B. H. (2023) · BJR|Open · doi:10.1093/bjro/tzae039 · Cited by 9
Uncertainties in outcome modelling in radiation oncology
Dünger, L., Mäusel, E., Zwanenburg, A., & Löck, S. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100774 · Cited by 7
AI in Radiation Oncology: A Comprehensive Review of Current Applications and Future Directions
Zafar, F., Vilsan, J., Mani, S., Yousif, A. R. A., Cano-Reyes, S. E., Abraham, G., Neto, J. F. D. B., Imam, B., Aluko, M., & Sikdar, S. (2025) · Cureus · doi:10.7759/cureus.92964 · Cited by 6
Vorbach, S. M., Putz, F., Ganswindt, U., Janssen, S., Grohmann, M., Knippen, S., Heinemann, F. E., Shafie, R. A. E., & Peeken, J. C. (2025) · Strahlentherapie und Onkologie · doi:10.1007/s00066-025-02403-1 · Cited by 5
Zamboglou, C., Doncker, W. D., Christoforou, A. T., Arcangeli, S., Berlín, A., Blanchard, P., Bauman, G. S., Campi, R., Castro, E., Choudhury, A., et al. (2025) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2025.111039 · Cited by 5
Self-citationYuan, L., Chen, Q., Al‐Hallaq, H., Yang, J., Yang, X., Geng, H., Latifi, K., Cai, B., Wu, Q. J., Xiao, Y., et al. (2025) · Practical Radiation Oncology · doi:10.1016/j.prro.2025.06.012 · Cited by 5
Goudschaal, K., Azzarouali, S., Visser, J., Admiraal, M. A., Wiersma, J., Wieringen, N. V., de la Fuente, A., Piet, M., Daniels, L. A., den Boer, D., et al. (2024) · Clinical and Translational Radiation Oncology · doi:10.1016/j.ctro.2024.100884 · Cited by 5
Unprepared and overwhelmed: A case for clinician-focused AI education
Siddiqui, N., Bouchi, Y., Kim, E., Hron, J. D., Park, J. S., & Kang, J. J. (2025) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2025.103252 · Cited by 4
ROADMAP: An Ontology of Medical AI Models and Datasets
Suri, A., Takahashi, M. S., Retson, T., Gonzales, R. A., Park, S. H., & Kahn, C. E. (2026) · Radiology Artificial Intelligence · doi:10.1148/ryai.260069 · Cited by 4
Li, J., Liu, W. G., Mu, Y., Wang, X., Zhang, H., Tang, K., & Zhang, D. (2026) · npj Precision Oncology · doi:10.1038/s41698-026-01345-w · Cited by 4
Perspectives for using artificial intelligence techniques in radiation therapy
Landry, G., Kurz, C., & Thummerer, A. (2024) · The European Physical Journal Plus · doi:10.1140/epjp/s13360-024-05660-8 · Cited by 4
Self-citationBungyi, G., Pócza, T., Zongor, Á., Takácsi‐Nagy, Z., & Major, T. (2024) · Orvosi Hetilap · doi:10.1556/650.2024.33180 · Cited by 4
Vellini, L., Quaranta, F., Menna, S., Pilloni, E., Catucci, F., Lenkowicz, J., Votta, C., Aquilano, M., D’Aviero, A., Iezzi, M., et al. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100708 · Cited by 4
Royal College of Radiologists Guidance Statements on the Use of Auto-contouring in Radiotherapy
Mackay, K., Banfill, K., Bernstein, D., Daniel, J., Díez, P., Gwynne, S., Hoole, A. C. F., Jena, R., Marchant, T., Nix, M. G., et al. (2025) · Clinical Oncology · doi:10.1016/j.clon.2025.104004 · Cited by 3
Montero, A. M. B., Huet‐Dastarac, M., Herranz-Hernández, S. M., Tengler, B., Buhl, E. S., Galapon, A., Cárdenas, C. E. S., Fusella, M., Herbin, G., de Hond, Y. J. M., et al. (2026) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2026.100940 · Cited by 4
Self-citationRen, J., Hochreuter, K., Kallehauge, J. F., & Korreman, S. (2025) · Lecture notes in computer science · doi:10.1007/978-3-031-83274-1_9 · Cited by 3
Self-citationGalapon, A., Wagenaar, D., Langendijk, J. A., & Both, S. (2025) · Medical Physics · doi:10.1002/mp.17937 · Cited by 3
Mazonakis, M., Tzanis, E., Kachris, S., Lyraraki, E., & Damilakis, J. E. (2025) · Physica Medica · doi:10.1016/j.ejmp.2025.104896 · Cited by 3
Mehmood, R., Parvez, M. A., Qureshi, M., & Sagheer, Z. (2025) · Review of Applied Management and Social Sciences · doi:10.47067/ramss.v8i1.442 · Cited by 3
Ren, J., Hochreuter, K., Rasmussen, M. E., Kallehauge, J. F., & Korreman, S. S. (2025) · Lecture notes in computer science · doi:10.1007/978-3-031-83274-1_2 · Cited by 3
Self-citationFontaine, J., Suszko, M., di Franco, F., Leroux, A., Bonnet, É., Bosset, M., Langrand-Escure, J., Clippe, S., Fleury, B., & Guy, J. (2025) · Physica Medica · doi:10.1016/j.ejmp.2025.105065 · Cited by 2
Buzzi, S., Mancosu, P., Bresolin, A., Gallo, P., Fauci, F. M. L., Lobefalo, F., Paganini, L., Pelizzoli, M., Reggiori, G., Franzese, C., et al. (2025) · Bioengineering · doi:10.3390/bioengineering12080897 · Cited by 2
Zhang, X., Xiao, N., Liang, H., Li, P., Zhang, Y., Zhang, S., Zhou, B., Yao, S., Yang, Z., & Chen, J. (2025) · Urologic Oncology Seminars and Original Investigations · doi:10.1016/j.urolonc.2025.11.013 · Cited by 2
Jinia, A. J., Chapman, K., Liu, S., Biancia, C. D., Hipp, E., Lin, E. C., Moulder, R., Parikh, D., Cordero, J., Pistone, C., et al. (2026) · npj Digital Medicine · doi:10.1038/s41746-026-02390-2 · Cited by 3
Self-learning GAN based synthetic CT generation: unlocking CBCT-based adaptive radiotherapy
Prunaretty, J., Colombo, L., Romdhani, S., Teboul, O., Azria, D., Paragyios, N., & Fenoglietto, P. (2026) · Frontiers in Oncology · doi:10.3389/fonc.2026.1756153 · Cited by 2
Auditing the clinical usage of deep-learning based organ-at-risk auto-segmentation in radiotherapy
Mason, J., Doherty, J., Robinson, S., Bastide, M. D. L., Miskell, J., & McLauchlan, R. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100716 · Cited by 2
Clinical implementation of patient-specific quality assurance for synthetic computed tomography
Nella, F., Tanadini‐Lang, S., & Bello, R. D. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100764 · Cited by 2
Emin, S., Rossi, E., Hedman, M., Giovenco, M., Villegas, F., & Onjukka, E. (2025) · Physica Medica · doi:10.1016/j.ejmp.2025.105089 · Cited by 1
Cao, Z., Zhang, J., Chen, L., Li, T., Wu, H., & Zhang, Y. (2025) · Current Medical Imaging Formerly Current Medical Imaging Reviews · doi:10.2174/0115734056401610250827114351 · Cited by 1
Podobnik, G., Borg, C., Debono, C. J., Mercieca, S., & Vrtovec, T. (2025) · Scientific Reports · doi:10.1038/s41598-025-18598-3 · Cited by 1
Shimizu, Y., Kishi, N., Mizowaki, T., & Nakamura, M. (2025) · Medical Physics · doi:10.1002/mp.70190 · Cited by 1
Franzese, C., Tanadini‐Lang, S., Verellen, D., Wiersema, L., Hörner‐Rieber, J., Romero, A. M., Pasquier, D., Bruynzeel, A. M. E., Boda‐Heggemann, J., Depuydt, T., et al. (2025) · Clinical and Translational Radiation Oncology · doi:10.1016/j.ctro.2025.101096 · Cited by 1
Sohoni, N., Sohoni, N. S., Sutherland, R., Sundaresan, V. M., Smani, S., Ananth, P. J., Onofrey, J. A., Aneja, S., Miszczyk, M., Lee, H.-J., et al. (2026) · Urology · doi:10.1016/j.urology.2026.01.008 · Cited by 1
Capone, L., Raza, G. H., D’Ambrosio, C., Tortorelli, F., Aquilanti, F. M., & Gentile, P. C. (2026) · AI · doi:10.3390/ai7030097 · Cited by 1
Dionne, P., Martin, A., Ouellet, É., Roy, J., Sidibé, I., Froment, M., Vigneault, É., Foster, W., Bachand, F., Archambault, L., et al. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70563 · Cited by 1
He, Y., Lu, Y., & Hu, L. (2026) · Frontiers in Oncology · doi:10.3389/fonc.2026.1781422 · Cited by 1
Clinical Implementation of Artificial Intelligence Systems in Medical Imaging and Radiotherapy
Vienna International Centre, 100, P. B., Vienna, A.-1., Austria, 2600-0, T. +. (., & 2600-7, F. +. (. (2026) · IAEA human health series · doi:10.61092/iaea.9olq-aqa9 · Cited by 2
Simões, R., van der Velden, S., Gooding, M. J., Djamal Boukerroui, A., Remeijer, P., & Janssen, T. M. (2026) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2026.111684 · Cited by 1
Azevedo, T. A., Abrantes, A. M., & Carvalho, J. (2025) · Technology in Cancer Research & Treatment · doi:10.1177/15330338251350909 · Cited by 2
Delopoulos, N., Marschner, S. N., Lombardo, E., Ribeiro, M. F., Rogowski, P., Losert, C., Winderl, T., Albarqouni, S., Belka, C., Corradini, S., et al. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100819 · Cited by 1
Self-citationCurrent advances in automation in radiotherapy
Nijkamp, J. A., Knäusl, B., Aznar, M. C., Georg, D., Thorwarth, D., Thwaites, D., Muren, L. P., & van der Heide, U. A. (2025) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2025.110779 · Cited by 1
Self-citationUncertainty quantification of U‐Net based segmentation tool using conformal prediction
Borden, B. J., Graham‐Knight, J. B., Lasserre, P., Lucas, S., & Rajapakshe, R. D. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70663
ESTRO congress and AI: (No) more questions!
Peters, S., & Hertel, T. (2024) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2024.110428
Radiotherapy quality assurance for the PEACE 1 trial: An individual case review analysis
Alyamani, N., Clementel, E., Sargos, P., Blanchard, P., Supiot, S., Ronchin, P., Pommier, P., Dubergé, T., Silva, M., Hammoud, Y., et al. (2025) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2025.110780
Barateau, A. (2025) · Annales des Mines - Responsabilité et environnement · doi:10.3917/re1.117.0104
Radiation oncology at crossroads: Rise of AI and managing the unexpected
Bakhtiari, M. (2025) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70043
Recent innovations in offline and online Magnetic Resonance Imaging guided radiation oncology
Boldrini, L., Thorwarth, D., & Placidi, L. (2025) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2025.100753
Wang, T. (2025) · International Conference on Health Big Data · doi:10.1145/3733006.3733040
Future Directions in Radiobiology and Clinical Radiation Oncology
Pal, D., Bhui, U., Bandyapadhyay, S., Ashique, S., & Debnath, B. (2025) · Springer eBooks · doi:10.1007/978-3-031-95724-6_20
Siddiqui, N. S., Bouchi, Y., Shah, S. J. H., AlQarni, S., Sood, S., Lee, Y., Park, J. S., & Kang, J. (2025) · JCO Clinical Cancer Informatics · doi:10.1200/cci-25-00067
Alyamani, N., Abrunhosa-Branquinho, A. N., Corning, C., Sharabiani, M., Castadot, P., Giralt, J., Kaźmierska, J., Grant, W., Christiaens, M., Tomsej, M., et al. (2025) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2025.111141
Self-citationModality-AGnostic image Cascade (MAGIC) for multi-modality cardiac substructure segmentation
Summerfield, N., He, Q., Kuo, A., Ruff, C., Pan, J., Ghanem, A. I., Zhu, S., Kumar, A., Nagpal, P., Zhao, J., et al. (2025) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2025.111296 · Cited by 2
Guo, Y., Aboagye, E. O., & Gujral, D. M. (2025) · Current Oncology Reports · doi:10.1007/s11912-025-01730-x
Roy, S., Singh, R., Paikarathodi, A., Maurya, S. M., Luharia, A. A., & Dube, P. (2025) · Journal of Cancer Treatment and Research · doi:10.11648/j.jctr.20251304.17
Radiation oncology as a complex adaptive system
Bakhtiari, M. (2026) · Frontiers in Oncology · doi:10.3389/fonc.2025.1686835 · Cited by 1
Real-Time AI-Based Radiotherapy Planning for Nasopharyngeal Carcinoma: Development and Validation
Wang, G., Yang, X., Ni, Q., Tang, J., Huang, K., Guo, H., Jiang, X., Diao, W., Li, H., Yang, Y., et al. (2026) · Cyborg and Bionic Systems · doi:10.34133/cbsystems.0544
Principles and techniques of stereotactic radiosurgery
(731854), H. Z., Liu, M., & Wen, N. W. (2026) · Chinese Science Bulletin (Chinese Version) · doi:10.1360/csb-2025-5596
Beddok, A., Cohen, E., Raymond, G., Rabenjason, S., Donnarieix, D., & Delporte, L. (2026) · Journal of Cancer Education · doi:10.1007/s13187-026-02836-x
Comparison of Four Auto-Contouring Software for Prostate Cancer Structures
Ath, V., Choi, S. H., Kim, K. B., Lee, J. C., & Nguyen, D. C. (2026) · Progress in Medical Physics · doi:10.14316/pmp.2026.37.1.8
Marini, L., Mozzi, C., Avanzo, M., Lizzi, F., Meattini, I., Pirrone, G., Retico, A., Uwitonze, E., Kraan, A. C., & Talamonti, C. (2026) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2026.100966
Wahid, K. A., Nailon, W. H., & Solberg, T. D. (2026) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2026.100968
Sarchosoglou, A., Genitsarios, I., Silvis-Cividjian, N., Papavasileiou, P., Bakas, A., Papanikolaou, N., & Pappas, E. (2026) · Technical Innovations & Patient Support in Radiation Oncology · doi:10.1016/j.tipsro.2026.100404
Sun, H., Li, J., Liu, Y., Meng, F., An, Z., Cui, Y., Lv, B., Chai, G., Jia, L., Shi, Z., et al. (2026) · BMC Medical Imaging · doi:10.1186/s12880-026-02400-5
Chang, J., & Vaccarelli, M. J. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70592
Felice, F. D., Lechner, W., & Blanchard, P. (2026) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2026.111585
Cardenas, C. E., Harms, J. M., & Wang, D. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70618
Image-Guided Radiooncology: The Potential of Artificial Intelligence in Clinical Application
Vorbach, S. M., Combs, S. E., Wiestler, B., & Peeken, J. C. (2026) · Recent results in cancer research/Recent Results in Cancer Research · doi:10.1007/978-3-032-21415-7_9
AI-Driven Ultrasound: Automation, Image Interpretation and Decision Support
Sarkar, N., Stancliff, H., & Srinivasareddy, S. (2026) · IntechOpen eBooks · doi:10.5772/intechopen.1015391
Malone, C., Callens, D., Forde, E., Leech, M., Cardenas, C. E., Gooding, M. J., Ryan, S. M., Thirion, P. G., Fitzpatrick, C., O’Donovan, T., et al. (2026) · Clinical and Translational Radiation Oncology · doi:10.1016/j.ctro.2026.101210
Marsilla, J., Welch, M. L., Siraj, J., McElcheran, C., Kim, J. W., Tkachuk, D., Kim, S., Cho, J., Hahn, E., Hosni, A., et al. (2026) · Physics and Imaging in Radiation Oncology · doi:10.1016/j.phro.2026.101015
Amer, A., Ghatwary, N., Fayed, S. F., Magdy, S., Hussein, A., Kadry, R., & Abdelmaksoud, A. I. (2026) · Machine Learning and Knowledge Extraction · doi:10.3390/make8060158
Survey evidence demands professional guidance on large language model use in medical physics
Norris, S. A., Kron, T., Douglass, M. J. J., & Badawy, M. K. (2026) · Physica Medica · doi:10.1016/j.ejmp.2026.105835
An, D., Dong, L., Sun, Z., Wu, J., Wang, T., Kang, J., He, Z., & Hu, Z. (2026) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ae81bd
Yang, D. B., Lei, M., Yang, Q., Sun, Z., Hou, X., Hu, W., & Wang, J. (2026) · JCO Clinical Cancer Informatics · doi:10.1200/cci-25-00380
Beyond auto-segmentation: the case for planning and dosimetry AI in head and neck radiation oncology
Elkhadrawy, A., Choi, S. J., Dee, E. C., Patel, M. S., Wu, Y., Riaz, N., Veeraraghavan, H., Lee, D., Elguindi, S., & Lee, N. Y. (2026) · BMJ Oncology · doi:10.1136/bmjonc-2026-001115
Hindmarsh, J., Jessop, N., & Baldock, C. (2026) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-026-01738-y
Zhu, M., Gou, L., Zhang, C., & Zheng, D. (2026) · Frontiers in Oncology · doi:10.3389/fonc.2026.1912362
Tomori, S., Narazaki, K., & Tasaka, S. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70715
Ahmed, E. N., Ahmed, M. S., & Mushtaq, A. H. (2026) · Precision Oncology · doi:10.3390/precisoncol1030011
Chow, J. C. L., & Giap, H. (2026) · International Journal of Particle Therapy · doi:10.1016/j.ijpt.2026.101954
Xuan, Z., Shan, G., Yu, S., Lai, Z., Hao, X., Guo, J., Cao, Y., Yang, S., & Wang, D. (2026) · Radiation Medicine and Protection · doi:10.1016/j.radmp.2026.08.002
Butterworth, V., Young, T., King, A. P., & Urbano, T. G. (2026) · Frontiers in Oncology · doi:10.3389/fonc.2026.1893002
Wang, N., Xu, T., Kang, Y., Liang, Y., Wang, W., Nie, W., Dai, J., Tang, Y., & Men, K. (2026) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.70768
Maddaloni, F. S., Broggi, S., Fodor, A., Palazzo, G., Ubeira-Gabellini, M. G., Riani, C., Pasetti, M., Tummineri, R., Saufi, S., Mangili, P., et al. (2026) · Physica Medica · doi:10.1016/j.ejmp.2026.107200
Drivers of Change in Radiotherapy Practice: Lessons From a UK Tertiary Cancer Centre (2013–2023)
Kapadi, A., Price, G. J., Finn, C. F., Holley, R., Banfill, K., Walls, G., Marchant, T., Radhakrishna, G., Price, J., Choudhary, A., et al. (2026) · Clinical Oncology · doi:10.1016/j.clon.2026.104342
David, S. R., Kifli, A. A. A., & Rajabalaya, R. (2026) · Journal of Drug Delivery Science and Technology · doi:10.1016/j.jddst.2026.108966
Optimizing the delivery of radiotherapy with artificial intelligence
Katsoulakis, E., & Naqa, I. E. (2026) · Nature Reviews Clinical Oncology · doi:10.1038/s41571-026-01204-4 · Cited by 1
Mondal, K., Vijay, A., Mandal, A., & Patel, G. (2026) · Journal of Medical Physics · doi:10.4103/jmp.jmp_105_26
Metadata
- Authors
- Hurkmans, C., Bibault, J.-E., Brock, K. K., van Elmpt, W., Feng, M., David Fuller, C., Jereczek-Fossa, B. A., Korreman, S., Landry, G., Madesta, F., Mayo, C., McWilliam, A., Moura, F., Muren, L. P., El Naqa, I., Seuntjens, J., Valentini, V., & Velec, M.
- Affiliations
- Catharina Ziekenhuis (NL)
- Hôpital Européen Georges-Pompidou (FR)
- The University of Texas MD Anderson Cancer Center (US)
- Maastricht University Medical Centre (NL)
- University of California, San Francisco (US)
- Aarhus University (DK)
- LMU Klinikum (DE)
- University Medical Center Hamburg-Eppendorf (DE)
- University of Michigan (US)
- University of Manchester (GB)
- University Health Network (CA)
- Università Cattolica del Sacro Cuore (IT)
- University of Toronto (CA)
- Corresponding author
- Hurkmans, Coen
- ORCID
- Venue
- Radiotherapy and Oncology
- Publisher
- Elsevier BV
- Published
- 2024-08
- Volume
- 197
- Pages
- 110345
- ISSN
- 0167-8140 (ISSN-L)
- OpenAlex
- W4399327635
- Cited by
- 110
- References
- 85
- Citations by year
- 2026: 56 · 2025: 46 · 2024: 7
- FWCI
- 10.19
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Radiotherapy and Oncology (publishedVersion, cc-by)
- PubMed (publishedVersion)
- TU/e Research Portal (publishedVersion, cc-by)
- Digital Commons-TMC (Texas Medical Center) (submittedVersion)
- Research Publications (Maastricht University) (submittedVersion, cc-by)
- Funders
- European SocieTy for Radiotherapy and Oncology
- Topics
- Health Sciences › Medicine › Health Informatics › Artificial Intelligence in Healthcare and Education
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiomics and Machine Learning in Medical Imaging
- Keywords
- artificial intelligence, radiotherapy, treatment planning, adaptive radiotherapy, image segmentation, patient-specific quality assurance, outcome prediction, AI ethics, clinical adoption of AI, Delphi consensus
- MeSH terms
- Algorithms, Artificial Intelligence, Delphi Technique, Humans, Radiotherapy, Radiotherapy Planning, Computer-Assisted, Radiation Oncology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review