The Double Sensor—A non-invasive device to continuously monitor core temperature in humans on earth and in space
Open accessAbstract
The objective of our study was to establish whether rectal temperature recordings in humans could be replaced by a non-invasive skin temperature sensor combined with a heat flux sensor (Double Sensor) located at the forehead to monitor core body temperature changes due to circadian rhythms. Rectal and Double Sensor data were collected continuously for 24h in seven men undertaking strict head-down tilt bed-rest. Individual differences between the two techniques varied between -0.72 and +0.55 degrees C. Nonetheless, when temperature data were approximated by cosinor analysis in order to compare circadian rhythm profiles between methods, it was observed that there were no significant differences between mesor, amplitude, and acrophase (P>0.310). It was therefore concluded that the Double Sensor technology is presently not accurate enough for performing single individual core body temperature measurements under resting conditions at normal ambient room temperature. Yet, it seems to be a valid, non-invasive alternative for monitoring circadian rhythm profiles.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“a validated heat-flux sensor (Tcore, Dräger, Germany), connected to a miniaturized data logger (HFUM, KORA Industrie Elektronik, Germany) [44–48]”
In the citing paper
“The double heat-flux sensor technology has been successfully tested during hypothermia, bedrest, and exercise on Earth and in space [44–48].”
In the citing paper
“This approach has been tested in various space analogs and clinical settings and has been found to provide a reliable and valid estimate of CBT20–23.”
From Increased core body temperature in astronauts during long-duration space missions (2017) · cites this paper as [21]
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
A Note on the Concordance Correlation Coefficient
Steichen, T. J., & Cox, N. J. (2002) · The Stata Journal Promoting communications on statistics and Stata · doi:10.1177/1536867x0200200206 · Cited by 430
Barratt, M. (2016) · CRC Press eBooks · doi:10.1201/b13197-27 · Cited by 349
A note on concordance correlation coefficient.
Liao, J. J. Z., & Lewis, J. W. (2000) · PubMed · Cited by 319
Mirkin, D. B. (1994) · JAMA · doi:10.1001/jama.1994.03520050087038 · Cited by 281
A new technique for monitoring the deep body temperature in man from the intact skin surface.
Fox, R. H., & Solman, A. (1971) · PubMed · Cited by 179
BODY TEMPERATURE AND ANAESTHESIA
HALL, G. M. (1978) · British Journal of Anaesthesia · doi:10.1093/bja/50.1.39 · Cited by 130
A non-invasive device to continuously determine heat strain in humans
Gunga, H.-C., Sandsund, M., Reinertsen, R. E., Sattler, F., & Koch, J. (2008) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2008.03.004 · Cited by 129
Self-citationEsophageal and tympanic temperature responses to core blood temperature changes during hyperthermia
Shiraki, K., Konda, N., & Sagawa, S. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.61.1.98 · Cited by 123
Temperature in the external auditory meatus as an index of central temperature changes
Cooper, K., Cranston, W. I., & Snell, E. S. (1964) · Journal of Applied Physiology · doi:10.1152/jappl.1964.19.5.1032 · Cited by 96
Holmér, I., Kuklane, K., & Gao, C. (2005) · Lund University Publications (Lund University) · Cited by 87
BODY TEMPERATURE AND ANAESTHESIA
IMRIE, M. M., & HALL, G. M. (1990) · British Journal of Anaesthesia · doi:10.1093/bja/64.3.346 · Cited by 81
Bestimmung der Körperkerntemperatur während der postoperativen Wiedererwärmung
Bräuer, A., Weyland, W., Fritz, U., Schuhmann, M. U., Schmidt, J. H., & Braun, U. (1997) · Der Anaesthesist · doi:10.1007/s001010050454 · Cited by 16
Smith, P. M., DAVIES, G., & Christie, M. J. (1980) · Ergonomics · doi:10.1080/00140138008924721 · Cited by 10
Flagged first, then by citations
Research works printed but not linked: 1
- Braeuer, 1997, Bestimmung der Körpertemperatur, Anaesthesist, 46, 683
Cited by
Sorted by citations
Depner, C. M., Cheng, P., Devine, J. K., Khosla, S., de Zambotti, M., Robillard, R., Vakulin, A., & Drummond, S. P. A. (2019) · SLEEP · doi:10.1093/sleep/zsz254 · Cited by 245
Long-term dry immersion: review and prospects
Navasiolava, N. M., Custaud, M., Tomilovskaya, E., Larina, I. M., Mano, T., Gauquelin‐Koch, G., Gharib, C., & Kozlovskaya, I. B. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1750-x · Cited by 205
Dow, J., Giesbrecht, G., Danzl, D., Brugger, H., Sagalyn, E., Walpoth, B., Auerbach, P., Mcintosh, S., Némethy, M., McDevitt, M. C., et al. (2019) · Wilderness & environmental medicine (Print) · doi:10.1016/j.wem.2019.10.002 · Cited by 151
Estimation of human core temperature from sequential heart rate observations
Buller, M. J., Tharion, W. J., Cheuvront, S. N., Montain, S. J., Kenefick, R. W., Castellani, J. W., Latzka, W. A., Roberts, W. S., Richter, M., Jenkins, O. C., et al. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/7/781 · Cited by 172
Zafren, K., Giesbrecht, G., Danzl, D., Brugger, H., Sagalyn, E., Walpoth, B., Weiss, E., Auerbach, P., Mcintosh, S., Némethy, M., et al. (2014) · Wilderness & environmental medicine (Print) · doi:10.1016/j.wem.2014.09.002 · Cited by 84
Metadata
- Authors
- Gunga, H.-C., Werner, A., Stahn, A., Steinach, M., Schlabs, T., Koralewski, E., Kunz, D., Belavý, D. L., Felsenberg, D., Sattler, F., & Koch, J.
- Affiliations
- Charité – Universitätsmedizin Berlin (DE)
- Stiftung Charité (DE)
- Dräger (Germany) (DE)
- Corresponding author
- Gunga, H.-C.
- Venue
- Respiratory Physiology & Neurobiology
- Publisher
- Elsevier BV
- Published
- 2009-10
- Volume
- 169
- Pages
- S63-S68
- ISSN
- 1569-9048 (ISSN-L)
- OpenAlex
- W2025601974
- Cited by
- 122
- References
- 15
- Citations by year
- 2026: 3 · 2025: 8 · 2024: 8 · 2023: 8 · 2022: 9 · 2021: 11 · 2020: 6 · 2019: 11 · 2018: 10 · 2017: 8 · 2016: 5 · 2015: 7 · 2014: 7 · 2013: 7 · 2012: 5
- FWCI
- 3.09
- Type
- journal article
- Language
- en
- Open access
- green
- figshare.com
- Locations
- Respiratory Physiology & Neurobiology (publishedVersion)
- PubMed (publishedVersion)
- Figshare (submittedVersion, cc-by)
- Funders
- Deutsches Zentrum für Luft- und Raumfahrt
- Alexander von Humboldt-Stiftung (14431/02/NL/SH2, BBR2-2)
- European Space Agency (50WB0720)
- Topics
- Health Sciences › Medicine › Physiology (medical) › Thermoregulation and physiological responses
- Life Sciences › Neuroscience › Endocrine and Autonomic Systems › Circadian rhythm and melatonin
- Health Sciences › Medicine › Physiology (medical) › Spaceflight effects on biology
- Keywords
- core body temperature, non-invasive temperature monitoring, heat flux sensors, frons, rectal temperature, circadian rhythm, head-down bed rest, cosinor analysis
- MeSH terms
- Adult, Body Temperature, Extraterrestrial Environment, Humans, Male, Monitoring, Physiologic, Sensitivity and Specificity, Skin Temperature, Thermography, Reproducibility of Results, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review