Evolution in impedance at the electrode‐skin interface of two types of surface EMG electrodes during long‐term recordings
The DOI matches its Crossref record; 19 of the 22 references it cites were found in a registry or an index.
Abstract
The evolution in impedance at the electrode-skin interface of Beckman and Red Dot electrodes was assessed during long-term recordings. Impedance was measured between each pair of electrodes, arranged in a bipolar configuration on tibialis anterior (n=13). A waveform constructed of sinusoids of known frequencies, evenly distributed on a log scale from 1-16,384 Hz, was applied through the electrodes, and the result recorded by a DAQ system. SEMG signals were recorded at 1000 Hz during isometric dorsiflexion contractions of 30 s, performed every 15 min for 2 h. Impedance data were acquired at 65,536 Hz immediately before and after SEMG recordings. Large individual differences in impedance levels were observed at low frequencies. At high frequencies, impedance values depended only on the electrode type. Impedance decreased steadily with time for Beckman electrodes (p < 0.05), but did not decrease significantly for Red Dot electrodes. The magnitude of the reduction over time varied widely between individuals, and was related to the initial impedance values. The impedance-bandwidth product remained constant for each electrode type (95% confidence intervals 146.2-148.2 and 126.1-127.8 for Beckman and Red Dot electrodes respectively). When skin impedance is electrically modelled with a simple network containing a resistor and a capacitor, the capacitance varies with the properties of the electrode used, whereas resistance is dependent on the subject. Furthermore, the EMG spectrum is unaffected by impedance provided skin preparation is sufficient to reduce the impedance below 55 komega.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The impedance at the electrode-skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Hewson et al., 2003
Citations
Cites
Tabachnick, B. G., & Fidell, L. S. (1983) · Cited by 80,054
Muscles Alive—their functions revealed by electromyography
(1963) · Postgraduate Medical Journal · doi:10.1136/pgmj.39.449.162-b · Cited by 1,772
Myoelectrical manifestations of localized muscular fatigue in humans.
Luca, C. J. D. (1984) · PubMed · Cited by 1,121
Sampling, noise-reduction and amplitude estimation issues in surface electromyography
Clancy, E. A., Morin, E. L., & Merletti, R. (2002) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(01)00033-5 · Cited by 634
Kaplan, E. (1962) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-196244060-00031 · Cited by 589
Conduction velocity and EMG power spectrum changes in fatigue of sustained maximal efforts
Bigland-Ritchie, B., Donovan, E. F., & Roussos, C. S. (1981) · Journal of Applied Physiology · doi:10.1152/jappl.1981.51.5.1300 · Cited by 447
Muscular endurance and surface electromyogram in isometric and dynamic exercise
Hagberg, M. (1981) · Journal of Applied Physiology · doi:10.1152/jappl.1981.51.1.1 · Cited by 316
Duchêne, J., & Goubel, F. (1993) · PubMed · Cited by 178
Self-citationElectromyography and the study of sports movements: A review
Clarys, J. P., & Cabri, J. (1993) · Journal of Sports Sciences · doi:10.1080/02640419308730010 · Cited by 158
Noise characteristics of stainless-steel surface electrodes
Godin, D. T., Parker, P. A., & Scott, R. N. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446089 · Cited by 69
Research works printed but not linked: 2
- De Luca, 1985, Myoelectric manifestations of localized muscular fatigue, Crit. Rev. Biomed. Eng, 11, 251
- Thomasset, 1962, Propriétés électriques des tissus biologiques: mesure de l’impédance en clinique, Lyon Médical, 21, 107
Cited by
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
“The impedance at the electrode–skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type (Hewson et al., 2003).”
BackgroundMerletti, R., Botter, A., Troiano, A., Merlo, E., & Minetto, M. A. (2009) · Clinical Biomechanics · doi:10.1016/j.clinbiomech.2008.08.006 · Cited by 345
“During long term recordings, drying of the gel, changes of the skin properties and of the electrolyte-skin interface may cause significant variations of the electrode–skin impedance (Hewson et al., 2003; Dozio et al., 2007; Hoffmann et al., 2006).”
BackgroundTutorial. Surface EMG detection, conditioning and pre-processing: Best practices
Merletti, R., & Cerone, G. L. (2020) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2020.102440 · Cited by 233
Samuel, O. W., Zhou, H., Li, X., Wang, H., Zhang, H., Sangaiah, A. K., & Li, G. (2018) · Computers & Electrical Engineering · doi:10.1016/j.compeleceng.2017.04.003 · Cited by 193
MXene-infused bioelectronic interfaces for multiscale electrophysiology and stimulation
Driscoll, N., Erickson, B., Murphy, B. B., Richardson, A. G., Robbins, G., Apollo, N. V., Mentzelopoulos, G., Mathis, T., Hantanasirisakul, K., Bagga, P., et al. (2021) · Science Translational Medicine · doi:10.1126/scitranslmed.abf8629 · Cited by 180
Signal Acquisition Using Surface EMG and Circuit Design Considerations for Robotic Prosthesis
Zahak, M. (2012) · InTech eBooks · doi:10.5772/52556 · Cited by 139
“EMG Spectrum and noise influence on this spectrum [16]”
Nordez, A., Gennisson, J., Casari, P., Catheline, S., & Cornu, C. (2008) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2008.03.033 · Cited by 124
“The interelectrode resistance was reduced to less than 55 kO (Hewson et al., 2003) after standard skin preparation procedures: shaving, slight abrasion”
BackgroundMethodInterfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113
Chowdhury, A., Raza, H., Meena, Y. K., Dutta, A., & Prasad, G. (2018) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2018.11.010 · Cited by 111
Yang, L., Gan, L., Zhang, Z., Zhang, Z., Yang, H., Zhang, Y., & Wu, J. (2022) · ACS Omega · doi:10.1021/acsomega.2c00282 · Cited by 88
Sex differences in motor unit discharge rates at maximal and submaximal levels of force output
Inglis, J., & Gabriel, D. A. (2020) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2019-0958 · Cited by 72
“No practical significance can be placed on small changes in impedance below 10 kΩ, which is well within the accepted range for sEMG (Hewson et al. 2003).”
BackgroundPiervirgili, G., Petracca, F., & Merletti, R. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/10/2101 · Cited by 67
“It is known that the impedance of a pair of gelled electrodes decreases with time (Searle et al 2000, Hewson et al 2003, Grimnes et al 2008), but the behavior of single impedance magnitude and phase of the impedance unbalance ΔZ were not investigated for small electrodes.”
BackgroundMethodMurphy, B. B., Mulcahey, P. J., Driscoll, N., Richardson, A. G., Robbins, G. T., Apollo, N. V., Maleski, K., Lucas, T. H., Gogotsi, Y. G., Dillingham, T. R., et al. (2020) · Advanced Materials Technologies · doi:10.1002/admt.202000325 · Cited by 62
Time Evolution of the Skin–Electrode Interface Impedance under Different Skin Treatments
Murphy, B. B., Scheid, B. H., Hendricks, Q., Apollo, N. V., Litt, B., & Vitale, F. (2021) · Sensors · doi:10.3390/s21155210 · Cited by 53
“However, abrasion can irritate the skin and cause discomfort to the subject, and the initial effects of abrasive skin treatments do not persist over extended periods of time, since the impedance will naturally decrease once the skin rehydrates with sweat [7,12,22,25,27,28].”
BackgroundSurface electrode placement affects the EMG recordings of the quadriceps muscles
Wong, Y. M., & Ng, G. Y. F. (2006) · Physical Therapy in Sport · doi:10.1016/j.ptsp.2006.03.006 · Cited by 51
Assessment of muscle hardness changes induced by a submaximal fatiguing isometric contraction
Nordez, A., Guével, A., Casari, P., Catheline, S., & Cornu, C. (2008) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2007.11.005 · Cited by 50
“The inter-electrode resistance was reduced to less than 55 kX (Hewson et al., 2003) after standard skin preparation procedures: shaving, slight abrasion and cleaning with an alcohol-acetone solution (Maı̈setti et al., 2002a). sEMG signals were pre-amplified (gain = 600).”
MethodYoun, W., & Kim, J. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0641-y · Cited by 50
“Second, while proper skin preparations, like shaving the electrode sites or cleaning with alcohol, are required to obtain the acceptable level of skin impedance for EMG [14], no skin preparation is needed for MMG, and MMG is not influenced by changes in skin impendence due to sweating [36].”
BackgroundRemaud, A., Cornu, C., & Guével, A. (2008) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2007.11.001 · Cited by 51
“Electrode-skin impedance was reduced below 55 kX (Hewson et al., 2003) using standard skin preparation procedures (Maisetti et al.”
MethodMadeleine, P., Vedsted, P., Blangsted, A. K., Sjøgaard, G., & Søgaard, K. (2006) · Ergonomics · doi:10.1080/00140130600665828 · Cited by 48
“The use of conventional or dry surface EMG electrodes can also cause problems especially during long-term recording as the skin-electrode impedance can change, altering the quality of the surface EMG signal (Hewson et al. 2003).”
BackgroundJacob, D., Kristensen, I. U., Aubonnet, R., Recenti, M., Donisi, L., Ricciardi, C., Svansson, H. Á. R., Agnarsdóttir, S., Colacino, A., Jónsdóttir, M. K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-12822-0 · Cited by 57
Wang, S., Tang, H. J., Wang, B., & Mo, J. (2021) · IEEE Transactions on Neural Networks and Learning Systems · doi:10.1109/tnnls.2021.3124330 · Cited by 52
Martinez, S. R., Floch, P. L., Liu, J., & Howe, R. D. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202202661 · Cited by 50
Neuromuscular adaptations to 8-week strength training: isotonic versus isokinetic mode
Remaud, A., Cornu, C., & Guével, A. (2009) · European Journal of Applied Physiology · doi:10.1007/s00421-009-1164-9 · Cited by 54
“Electrode-skin impedance was reduced below 55 k (Hewson et al. 2003) using standard skin preparation procedures (i.”
MethodEEG Signal Quality of a Subcutaneous Recording System Compared to Standard Surface Electrodes
Duun‐Henriksen, J., Kjær, T. W., Looney, D., Atkins, M. D., Sørensen, J. A., Rose, M. H., Mandic, D. P., Madsen, R. E., & Juhl, C. B. (2015) · Journal of Sensors · doi:10.1155/2015/341208 · Cited by 40
Sex differences in the modulation of the motor unit discharge rate leads to reduced force steadiness
Inglis, J., & Gabriel, D. A. (2021) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2020-0953 · Cited by 43
Analysis of surface EMG spike shape across different levels of isometric force
Gabriel, D. A., Lester, S., Lenhardt, S. A., & Cambridge, E. D. J. (2006) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2006.07.004 · Cited by 40
Pozzo, M., Bottin, A., Ferrabone, R., & Merletti, R. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02350986 · Cited by 38
“This performance is similar to other types of commercially available electrode for bipolar EMG using a similar 'recessed electrode' technology (HEWSON et al., 2003).”
MethodHubble, R. P., Naughton, G. A., Silburn, P. A., & Cole, M. H. (2014) · BMJ Open · doi:10.1136/bmjopen-2014-006095 · Cited by 43
“surface electrodes over the muscles of interest, the skin will be prepared with an abrasive gel (Nuprep; Weaver Company, Aurora, Colorado, USA), and then cleaned thoroughly with an isopropyl alcohol wipe to minimise impedance at the electrode-skin interface and improve clarity of the myoelectric signal.(54) For indivi…”
MethodRemaud, A., Cornu, C., & Guével, A. (2005) · Europe PMC (PubMed Central) · Cited by 33
On-line signal quality estimation of multichannel surface electromyograms
Grönlund, C., Roeleveld, K., Holtermann, A., & Karlsson, J. S. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345813 · Cited by 30
Johnson, R. E., Kording, K. P., Hargrove, L. J., & Sensinger, J. W. (2016) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2016.2598095 · Cited by 32
“In contrast, factors such as subtle posture shifts, fatigue [29], or skin impedance changes [30] may cause systematic changes in the control mapping, which users may still perceive as random errors.”
BackgroundCharacterization of dry and wet Electrode-Skin interfaces on different skin treatments for HDsEMG
Cattarello, P., & Merletti, R. (2016) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2016.7533808 · Cited by 33
“The effect of the contact time, that is known to have some influence on the interface [4, 5, 9, 12] and the variability of the measurements were also investigated.”
BackgroundBureau, C., Dres, M., Morawiec, É., Mayaux, J., Delemazure, J., Similowski, T., & Demoule, A. (2022) · Annals of Intensive Care · doi:10.1186/s13613-022-01025-5 · Cited by 26
“Due to the variability of skin impedance [16, 17] and the patient’s morphology, EMGmax and EMGauc values were normalized to EMG activity at initiation of the SBT.”
MethodMadeleine, P., Xie, Y., Szeto, G. P. Y., & Samani, A. (2016) · Clinical Neurophysiology · doi:10.1016/j.clinph.2016.06.015 · Cited by 30
“Checker, USA) 10 kΩ (Konrad 2006) and 50 kΩ (Hewson et al., 2003) for respectively the neck-shoulder-forearm muscles and APB muscle were considered acceptable.”
A Review of EMG recording technique
Ahmad, I., Ansari, F., & Dhanbad, S. (2012) · Cited by 23
Farinha, A., Kellogg, S., Dickinson, K., & Davison, T. (2006) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205(2006)40[72:sirfem]2.0.co;2 · Cited by 25
Real-time actuation of a dielectric elastomer actuator neuroprosthesis for facial paralysis
Konstantinidi, S., Imholz, C., Martinez, T., Benouhiba, A., Walter, A., Civet, Y., Lindenblatt, N., & Perriard, Y. (2023) · Smart Materials in Medicine · doi:10.1016/j.smaim.2023.06.003 · Cited by 26
“Moreover, for the clinical application of the DEA, the location of the neural interface under the skin is more favorable than surface EMG electrodes which must be repositioned daily and can be seen from the outside [27].”
InfluentialCorticomuscular Co-Activation Based Hybrid Brain-Computer Interface for Motor Recovery Monitoring
Chowdhury, A., Dutta, A., & Prasad, G. (2020) · IEEE Access · doi:10.1109/access.2020.3026037 · Cited by 23
“However, CMC at this frequency range (8-12 Hz) is not visible because EMG activations are prominent at a higher frequency range as its power spectrum peaks around 50–60 Hz [53].”
BackgroundLong-Term Electrode–Skin Impedance Variation for Electromyographic Measurements
Sousa, A. S. P., Noites, A., Vilarinho, R., & Santos, R. S. (2023) · Sensors · doi:10.3390/s23208582 · Cited by 21
“This may result from the decreased initial impedance values obtained in the present study, as this factor has been demonstrated to affect the magnitude of the reduction over time [12].”
BackgroundInfluentialGavriel, C., & Faisal, A. A. (2014) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2014.23 · Cited by 18
“Its long-term applicability has been extensively investigated in several studies [23] [7] [8], showing a systematic decrease in the precision of the signal throughout the day, inline with our”
BackgroundA segmentation approach to long duration surface EMG recordings
Falou, W. E., Duchêne, J., Hewson, D. J., Khalil, M. A., Michel Grabisch, B., & Lino, F. (2004) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2004.07.004 · Cited by 18
“Although the level of noise introduced by the electrode-skin contact was not measured, a recent study reported that impedance levels up to 55 kX had no effect on median frequency estimates [12].”
BackgroundSelf-citationBeaudette, S. M., Unni, R. R., & Brown, S. H. M. (2014) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2014.03.006 · Cited by 17
“, 2009), changes in tissue impedance (Hewson et al., 2003), subject testing day (Frost et al.”
BackgroundBioimpedance Measurement of Knee Injuries Using Bipolar Electrode Configuration
Ye, X., Wu, L., Mao, K., Feng, Y., Li, J., Ning, L., & Chen, J. (2022) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2022.3200355 · Cited by 18
Lee, D., You, D., Cho, G., Lee, H., Shin, E., Choi, T., Kim, S., Lee, S., & Nam, W. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105600 · Cited by 23
Characterization of a New Type of Dry Electrodes for Long-Term Recordings of Surface-Electromyogram
García, G. A., Zaccone, F., Ruff, R., Micera, S., Hoffmann, K., & Dario, P. (2007) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2007.4428523 · Cited by 16
“Some authors have simplified the model to a circuit composed of a resistor in series with a capacitor in parallel with another resistor [12].”
BackgroundSensors Based on Conducting Polymers for Measurement of Physiological Parameters
Terán-Jiménez, O., Rodríguez-Roldán, G., Hernández-Rivera, D., & Suaste-Gómez, E. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2671448 · Cited by 15
Cornu, C., Nordez, A., & Bideau, B. (2009) · International Journal of Sports Medicine · doi:10.1055/s-0029-1237390 · Cited by 15
Activity Recognition Based on Pattern Recognition of Myoelectric Signals for Rehabilitation
Samuel, O. W., Fang, P., Chen, S., Geng, Y., & Li, G. (2017) · Scalable computing and communications · doi:10.1007/978-3-319-58280-1_16 · Cited by 11
Investigation of lower limb’s muscles activity during performance of salat between two age groups
Malik, N. A., Wahid, Z., Zulkipili, A. F., Ibrahim, S. N., Gunawan, T. S., & Khan, S. (2019) · Indonesian Journal of Electrical Engineering and Computer Science · doi:10.11591/ijeecs.v14.i2.pp608-617 · Cited by 11
“Variation in resistance of individual subjects that depends on the skin and underlying tissue [18] may be contributing to this difference.”
BackgroundMascia, A., Collu, R., Makni, N., Concas, M., Barbaro, M., & Cosseddu, P. (2025) · Sensors · doi:10.3390/s25154568 · Cited by 11
Micropatterned Biphasic Printed Electrodes for High‐Fidelity on‐Skin Bioelectronics
Carneiro, M. R., Lopes, T., Silva, A. F., Majidi, C., & Tavakoli, M. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202505419 · Cited by 12
Bingham, A., Jelfs, B., Arjunan, S. P., & Kumar, D. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512846 · Cited by 9
“For sEMG recording systems the quality of signal is usually determined by measuring the skin impedance before recording [5] or by visually inspecting the recorded signal for abnormalities [6].”
Neuromuscular fatigue in autoimmune myasthenia gravis: A cross-sectional study
Birnbaum, S., Sharshar, T., Ropers, J., Portero, P., & Hogrel, J. (2023) · Neurophysiologie Clinique · doi:10.1016/j.neucli.2023.102844 · Cited by 10
Self-citationMuscular activity level during pedalling is not affected by crank inertial load
Duc, S., Villerius, V., Bertucci, W., Pernin, J. N., & Grappe, F. (2005) · European Journal of Applied Physiology · doi:10.1007/s00421-005-1401-9 · Cited by 9
“Sites were shaved and cleaned with alcohol swab in order to reduce skin impedance to less than 15 kX, which is in line with the recommendation of Hewson et al. (2003).”
MethodZhang, J., Wang, M. D. S., Alam, M., Zheng, Y.-P., Ye, F., & Hu, X. (2024) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2024.1372158 · Cited by 10
Prabhavathy, T., Kumar, E. V., & Balaji, E. (2024) · Applied Soft Computing · doi:10.1016/j.asoc.2024.112382 · Cited by 9
Smartware electrodes for ECG measurements : Design, evaluation and signal processing
Rattfält, L. (2013) · Linköping University Electronic Press eBooks · doi:10.3384/diss.diva-100134 · Cited by 17
“Methods for evaluating clinical electrodes have been developed throughout the years [3, 29, 42].”
BackgroundWang, L., Zeng, Z., Sun, W., Lin, X., Cao, J., Sun, K., & Ouyang, J. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.170106 · Cited by 12
Muscle Fatigue Detection and Analysis Using EMG Sensor
Ahmed, S. F., Joyo, M. K., Mahdi, H. F., & Kiwarkis, I. J. (2020) · IEEE International Conference on Engineering Technologies and Applied Sciences (ICETAS) · doi:10.1109/icetas51660.2020.9484230 · Cited by 7
Viruega, H., Gaillard, I., Briatte, L., & Gaviria, M. (2020) · Brain Sciences · doi:10.3390/brainsci10050281 · Cited by 8
“When muscle activation is at such a low level, an acceptable SNR can only be obtained by markedly reducing the noise level in various ways [28].”
BackgroundInfluentialNogales, R., Guilcapi, J., Benalcazar, F., & Vargas, J. (2022) · Lecture notes in networks and systems · doi:10.1007/978-3-030-97719-1_16 · Cited by 6
Abbiss, C. R., Peiffer, J. J., Netto, K., & Laursen, P. B. (2006) · Journal of Musculoskeletal Research · doi:10.1142/s021895770600187x · Cited by 7
Wiedemann, L. G., & McDaid, A. (2017) · IEEE Life Sciences Conference (LSC) · doi:10.1109/lsc.2017.8268162 · Cited by 6
“higher frequencies of alternating current applied through a pair of electrodes with skin impedances ranging between 1kΩ and 1411kΩ at a frequency of 1Hz [16].”
BackgroundObjective and parametric methods used in functional classification disabled swimmers
Uścinowicz, N., Seidel, W., Zostawa, P., & Klich, S. (2013) · Physiotherapy · doi:10.2478/physio-2013-0022 · Cited by 6
Microneedle electrodes for electromyography
Krieger, K. J., Lijnse, T. M., Lowery, M. M., & O’Cearbhaill, E. D. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.117945 · Cited by 7
Rong, S. O. N. G., Liu, W., RuiLiang, L. I. U., Qiu, Q., Jiang, L., & Zhou, W. (2015) · Chinese Science Bulletin (Chinese Version) · doi:10.1360/n972014-01093 · Cited by 5
Characterization of alternating current impedance properties of biomedical electrodes
Zhou, W., Cheng, D.-C., Song, R., Zhang, C.-J., Xu, W.-P., & Pan, X. (2013) · Journal of Central South University · doi:10.1007/s11771-013-1609-5 · Cited by 5
“HEWSON et al [15] conducted the preliminary studies on two types of Ag/AgCl electrodes for EMG signal detection, and obtained the impedance properties of electrodes under different time conditions and current frequencies.”
BackgroundSpanu, A., Taki, M., Baldazzi, G., Mascia, A., Pietrabissa, R., Pani, D., Cosseddu, P., & Bonfiglio, A. (2024) · Advanced Engineering Materials · doi:10.1002/adem.202302195 · Cited by 4
Hu, H., Lo, W. L. A., Wang, X., Li, L., & Zhou, P. (2021) · Journal of Healthcare Engineering · doi:10.1155/2021/7296322 · Cited by 4
Manzo, R., Apicella, A., Arpaïa, P., Caputo, F., & Moccaldi, N. (2025) · IEEE Access · doi:10.1109/access.2025.3605928 · Cited by 4
“Movement, sweating of the patient, as well as prolonged use of the device, can cause slippage and partial/total detachment of the electrodes used for measurement, affecting the electrode-skin interface impedance and consequently the quality of signals [7], [8].”
Boppudi, S. T., Belal, S. Y., & Guntur, S. R. (2020) · Biomedical Engineering Letters · doi:10.1007/s13534-020-00164-7 · Cited by 3
Activation pattern masticatory muscles in dysphonics woman.
Rodrigues-Bigaton, D., Schwarzenbeck, A., Berni, K. C. D. S., Guirro, R. R. D. J., & Silvério, K. C. A. (2010) · PubMed · Cited by 2
Kuruganti, U., Pradhan, A., & Chester, V. L. (2017) · JPO Journal of Prosthetics and Orthotics · doi:10.1097/jpo.0000000000000167 · Cited by 3
“These steps were necessary as the electrode-skin impedance must be below 55 kΩ to not disrupt the validity of the EMG signal.(23)”
BackgroundElectrode potential and selective ionic adsorption
Alexe-Ionescu, A. L., Barbero, G., & Merletti, R. (2009) · Physics Letters A · doi:10.1016/j.physleta.2009.03.031 · Cited by 2
Experimental Studies of Alternating and Direct Current Impedance Properties of Ag/AgCl Electrodes
Liu, W., Zhou, W., Li, M. G., Qiu, Q., Liu, R. L., Lin, W. S., & Song, R. (2015) · Applied Mechanics and Materials · doi:10.4028/www.scientific.net/amm.778.110 · Cited by 2
Comparison of two different orthoses used in the treatment of lateral epicondylitis
Sari, M. E., Öztürk, D., & Bek, N. (2024) · Prosthetics and Orthotics International · doi:10.1097/pxr.0000000000000345 · Cited by 2
“11 Measurements were made during maximum grip strength and during NO, WO, and CB conditions.”
An EMG Controlled Bionic Arm with Machine Learning
Uygun, M. I., Recber, B., Celikli, M. A., & Öniz, Y. (2023) · International Symposium on Multidisciplinary Studies and Innovative Technologies (ISMSIT) · doi:10.1109/ismsit58785.2023.10304962 · Cited by 1
“To achieve dependable signal acquisition, it is crucial to make sure that the electrodes are positioned correctly [9-10].”
BackgroundAbou-Hamde, A., Philippi, L., Jones, E., Martin, C., Wu, K., Kundell, M., Mathur, S., Sadeghian, A., Davoudpour, M., Batt, J. A. E., et al. (2024) · Sensors · doi:10.3390/s24113599 · Cited by 1
Zhu, H., Tan, Z., Li, Y., Sun, Q., & Qi, H. (2025) · Sensor Review · doi:10.1108/sr-06-2025-0418 · Cited by 2
Electrodes based on PPy polymer for electrocardiography and impedance plethysmography
Terán-Jiménez, O., Hernández-Rivera, D., & Suaste-Gómez, E. (2017) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-4086-3_58 · Cited by 1
Electrical Conductivity of Skin Compared to Skin Perfusion Recordings
Jarløv, A., & Jensen, T. T. (2017) · International Journal of Biomedical and Clinical Engineering · doi:10.4018/ijbce.2017070101 · Cited by 1
Can Textile Electrode for ECG Apply to EMG Measurement?
Goto, D., & Shiozawa, N. (2018) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-9038-7_81 · Cited by 1
Adaptations nerveuses du système neuromusculaire à l'exercice isotonique versus isocinétique
Remaud, A. (2007) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 1
Investigation of the Relationship of Muscle Mechanical Characteristics with Biosignal Energy
Mariūnas, M., & Kojelyte, K. (2006) · Diffusion and defect data, solid state data. Part B, Solid state phenomena/Solid state phenomena · doi:10.4028/www.scientific.net/ssp.113.151
Accelerometric Detection of Intentional Movements in the Presence of Dyskinesia
Villada, M. C. (2015) · TSpace (University of Toronto)
Proprioceptive Neuromuscular Facilitation of the Wrist Flexors
McGuire, J. (2014) · Brock University Digital Repository (Brock University)
Relationship between Surface and Indwelling EMG Spike Shape Measures
Parro, J. J. (2014) · Brock University Digital Repository (Brock University) · Cited by 1
Shoulder proprioception and motor control
Rodrigues, S. (2016)
Yamaguchi, T., Tsukahara, A., Tanaka, Y., & Ueno, A. (2025) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2025.3541655 · Cited by 1
“11 is widely used [36], [37], [38], [39], [40], [41].”
RIME-Optimized CEEMDAN-LSTM Framework for sEMG Prediction: Cross-Subjects, Tasks, and Ages Study
Xiu, Q., Wang, Y., Chen, X., Zhang, D., Ma, T., Yu, J., & Xie, P. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3655884
“This results in the nonlinearity, nonstationarity, and individual differences [9], [10], [11].”
Birnbaum, S. (2018) · theses.fr (ABES) · doi:10.70675/4895700azc979z4d13z8b43za419d641793c
EARICHAPPAN, K., Hamzaid, N. A., & Ibitoye, M. O. (2026) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519426300073
Capacitive Touch Sensing-Based Human–Machine Interface for Upper-Limb Prosthetics
Jan, B., Ehshan, K. M., Abbas, Z., Venkataraman, S., Panigrahi, B. K., & Malik, S. (2026) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2026.3697066 · Cited by 1
“Nevertheless, a critical bottleneck that persists in prosthetic HMIs is the dependence on direct skin-electrode adhesion [30], [31].”
Schroeer, A., Biehl, T., Latzel, M., Launer, S., Strauß, D. J., & Corona–Strauss, F. I. (2026) · Frontiers in Neuroscience · doi:10.3389/fnins.2026.1838748
“However, it should also be noted that the EMG is very robust for impedances below 55 k (Hewson et al., 2003).”
Karunanithi, R., Veluswamy, P., & Palanisamy, R. (2026) · Springer eBooks · doi:10.1007/978-3-032-19984-3_2
Metadata
- Authors
- Hewson, D. J., Hogrel, J.-Y., Langeron, Y., & Duchêne, J.
- Affiliations
- Université de Technologie de Troyes (FR)
- Institut de Myologie (FR)
- Corresponding author
- Hewson, D.J.
- Venue
- Journal of Electromyography and Kinesiology
- Publisher
- Elsevier BV
- Published
- 2003-06
- Volume
- 13
- Issue
- 3
- Pages
- 273-279
- ISSN
- 1050-6411 (ISSN-L)
- OpenAlex
- W2070908442
- Semantic Scholar
- 533ea51acbec0a3fb19706b65d62cb46f05389dd
- Cited by
- 99
- References
- 22
- Citations by year
- 2026: 5 · 2025: 7 · 2024: 5 · 2023: 6 · 2022: 5 · 2021: 5 · 2020: 8 · 2019: 2 · 2018: 4 · 2017: 7 · 2016: 4 · 2015: 4 · 2014: 6 · 2013: 3 · 2012: 2
- FWCI
- 3.67
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- Journal of Electromyography and Kinesiology (publishedVersion)
- PubMed (publishedVersion)
- HAL (submittedVersion)
- University of Bedfordshire Repository (University of Bedfordshire) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrical and Bioimpedance Tomography
- Keywords
- skin capacitance, skin resistance, tibialis anterior
- MeSH terms
- Adult, Electrodes, Electromyography, Female, Galvanic Skin Response, Humans, Leg, Male, Muscle Contraction, Electric Impedance, Muscle, Skeletal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.