A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems

Open access
Navarro, Xavier, Krueger, Thilo B., Lago, Natalia, Micera, Silvestro, Stieglitz, Thomas, Dario, Paolo·2005·Journal of the Peripheral Nervous System

Abstract

Considerable scientific and technological efforts have been devoted to develop neuroprostheses and hybrid bionic systems that link the human nervous system with electronic or robotic prostheses, with the main aim of restoring motor and sensory functions in disabled patients.

AI snapshot

Written from the paper’s full text.

TL;DR

This review surveys peripheral nervous system interfaces for neuroprostheses and hybrid bionic systems, comparing non-invasive, muscle, extraneural, intraneural, and regenerative electrodes. It concludes that no single interface is universally optimal; selection depends on required selectivity, invasiveness, clinical maturity, and application.

Key findings

  • Electrode invasiveness generally tracks selectivity: surface and muscle electrodes are safer and easier but act only on underlying or implanted muscles; cuff and epineurial electrodes interface many axons; intrafascicular, penetrating, and sieve/regenerative electrodes access smaller axon groups but carry greater risk of nerve damage and are mostly research-stage.
  • Clinical PNS neuroprostheses are dominated by stimulation: sacral anterior root stimulators for bladder control reached about 2,000 patients worldwide; phrenic pacing, vagus nerve stimulation (postoperative infection under 3%, device removal 1%), cochlear implants, and pain/continence devices are clinical, while recording interfaces for prosthesis control remain largely research.
  • Penetrating arrays: Utah MEAs with 25 or 100 needles, including slanted lengths from 0.5 to 1.5 mm, recorded single-unit mechanoreceptor responses in only 10–20% of electrodes and evoked graded muscle recruitment with 1–20 µA currents; responses were stable over 36 hours, and in cat sciatic nerve up to 7 months stimulation stabilized in 80% of electrodes but sensory recordings did not.
  • LIFEs implanted in eight amputees for two days allowed voluntary motor signals linked to intended missing-limb movements and elicited graded touch, joint-movement, and position sensations; regenerative sieve electrodes require hole diameters around 40–65 µm, and no human implants have been reported. EMG-based control works for prostheses and exoskeletons but is limited in degrees of freedom, while ENG can provide event and kinematic feedback.

Methods

  • Design: critical narrative review, not a primary experimental study; it covers PNS organization, electrode types, materials and fabrication, and control strategies.
  • Data or sample: synthesizes prior animal and human studies and clinical experience; no systematic search, inclusion criteria, or primary dataset is described.
  • Main analysis: qualitative comparison using tables of advantages/disadvantages and application/status, with discussion of engineering, biological, and clinical tradeoffs.

Limitations

  • Stated: no universal best electrode exists; nerve electrodes may damage nerves; cuff, intraneural, and regenerative designs need more study; regenerative electrodes have no human implants; bioelectronic interfaces are early experimental.
  • Stated: long-term stability, selectivity, and biocompatibility remain unresolved, and many recording approaches are not clinically mature.
  • Evident: as a narrative review from 2005, it lacks a systematic search or meta-analysis and predates later technological and clinical developments.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1111/j.1085-9489.2005.10303.x

Cites

268 of 268 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

19 of 259 cited sources checked · 8 supported · 0 flagged · accuracy not yet measured

  • Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Helical electrodes are widely used for functional electrical stimulation of the vagus nerve for intractable epilepsy, sleep apnea, and depressive syndromes.

    “Helical electrodes are widely used for FES of the vagus nerve, which has found therapeutical applications for the control of intractable epilepsy, sleep apnea, and treatment of depressive syndromes (McLachlan, 1997; Fisher and Handforth, 1999) .”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites McLachlan, 1997 source
    • cites Fisher and Handforth, 1999 source

      The Abstract (page unavailable) describes an electrode implanted around the left vagus nerve and concludes that vagus nerve stimulation is effective for medically intractable partial seizures. It does not identify the electrode as helical, establish that helical electrodes are widely used, or discuss treatment of sleep apnea or depressive syndromes. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
  • Premisecitation
    § Book electrodes · p. 9

    An epidural implantation procedure was developed to increase the group of patients who may benefit from this neural prosthesis.

    “Besides the intradural surgical implantation, an epidural implantation procedure was developed (Sauerwein et al., 1990) to enhance the group of patients who may benefit from this kind of neural prosthesis.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Sauerwein et al., 1990 source
  • Premisecitation
    § Book electrodes · p. 9

    The sacral anterior root stimulator can also be used to restore male sexual and reproductive function.

    “The SARS can be also used to restore male sexual and reproductive function (Stief et al., 1992; Brindley, 1995) .”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites Stief et al., 1992 source

      The Abstract (p. 5) says the implanted stimulator could induce penile erection, and the Discussion (p. 6) reports full erections in all three men at certain stimulation frequencies. This supports restoration of an aspect of sexual function, but the paper does not establish restoration of reproductive function, such as ejaculation or fertility. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
    • cites Brindley, 1995

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Premisecitation
    § Penetrating microelectrodes · p. 14

    Electrode arrays provided neural contacts with low-force recruitment properties similar to those of single wires.

    “Although involving a more invasive insertion procedure, electrode arrays provided neural contacts with low-force recruitment properties similar to those of single wires.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Smit et al., 1999 source
  • Premisecitation
    § Penetrating microelectrodes · p. 14

    No strict inverse recruitment was observed, especially for intrafascicular electrodes.

    “Especially for intrafascicular electrodes, no strict inverse recruitment was observed (Veltink et al., 1989) .”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Veltink et al., 1989 source

      The supplied passages contain bibliographic metadata, not findings about recruitment. Searches for inverse recruitment with intrafascicular electrodes and for recruitment-order reversal returned no relevant passage, so our copy does not establish whether the paper observed strict inverse recruitment. (LLM, conf=0.72)

      Experimental — accuracy not yet measured
  • May combine two claims. 94
  • Classified as probably not a substantial claim. 46
  • The quote only partly states this claim. 1
  • May combine two claims.

    Premisecitation
    § Electrical · p. 2

    The most frequent use of PNS interfaces is in FES, with clinical applications for controlling micturition and defecation by stimulating the sacral roots.

    “Nowadays, the most frequent use of PNS interfaces resides in FES, which has clinical applications in various systems designed to control micturition and defecation by stimulating the sacral roots”
    Quote in the paperNot a substantial claim1 of 3 sources checked
    • cites Brindley et al., 1986 source

      The Discussion reports that most patients used sacral anterior root stimulators mainly for micturition. Methods and Results also reports that a few used S3 or S4 stimulation to aid defaecation. The paper does not establish that FES is the most frequent use of PNS interfaces overall. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
    • cites Creasey, 1993 source
    • cites Van Kerrebroeck et al., 1993 source
  • May combine two claims.

    Premisecitation
    § Non-electrical interfaces · p. 7

    Acoustic signals increase in parallel with surface EMG signals during voluntary and evoked contraction, but acoustic amplitude decays with fatigue whereas EMG amplitude does not.

    “Acoustic signals increase in parallel with surface EMG signals during voluntary and evoked contraction, but acoustic amplitude decays with fatigue whereas EMG amplitude does not”
    Quote in the paperNot a substantial claim2 of 2 sources checked
    • cites Barry et al., 1985 source

      The Abstract reports that acoustic amplitude decayed with fatigue while surface EMG amplitude did not, and the Figure 5 discussion describes both signals increasing together when voluntary effort increased. The copy does not establish that they increase in parallel during evoked contractions; the Discussion describes external nerve stimulation as a proposed future test. It also notes that surface EMG began to decay below 45% of maximal voluntary contraction, so the claim that EMG amplitude does not decay needs qualification. Page numbers are unavailable in our copy. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
    • cites Barry et al., 1992 source

      The Results (page unavailable) report that vibration amplitude declines with fatigue, but also state that the evoked compound muscle action potential (CMAP) declines, though less rapidly. The discussion (page unavailable) says surface EMG amplitude typically does not decline during moderate voluntary contractions. The paper does not support the broader claim that acoustic and EMG signals increase in parallel during both types of contraction, or that electrical-signal amplitude does not decline with evoked fatigue. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Book electrodes · p. 9

    After more than 10 years of development, Brindley's group reported the first 50 cases of a radio frequency-controlled sacral anterior root stimulator implant in humans.

    “After more than 10 years of development, Brindley's group reported the first 50 cases of a radio frequency-controlled implant from series production with medical approval called sacral anterior root stimulator (SARS) in humans (Brindley et al., 1986) .”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Brindley et al., 1986 source

      The Abstract reports the first 50 human recipients of sacral anterior root stimulators; Methods and Results describes external radio transmitters, and the References include development work dating to 1972. But the printed sentence goes further: Methods and Results says the first 11 implants were made in the Neurological Prostheses Unit, not series-produced, and our copy does not establish medical approval. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Book electrodes · p. 9

    The Finetech-Brindley stimulator was commercialized, and less than a decade later 500 patients had benefited from the system.

    “The implant was commercialized as FinetechBrindley stimulator, and less than a decade later 500 patients benefited from the system (Brindley, 1994) .”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Brindley, 1994 source

      The Introduction says the 500th London-design stimulator was implanted in February 1992, and Sources of information identifies Finetech Medical Ltd as the manufacturer. The paper does not date commercialization, so it cannot establish the claimed interval of less than a decade. Its benefits discussion describes potential benefits but does not establish that all 500 patients benefited. (LLM, conf=0.86)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Penetrating microelectrodes · p. 14

    In acute experiments, multiple-wire-microelectrode arrays were inserted into rat nerves to investigate selective stimulation of motor units.

    “In acute experiments, multiple-wire-microelectrode arrays were inserted into rat nerves to investigate selective stimulation of motor units (Smit et al., 1999) .”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Smit et al., 1999 source
  • May combine two claims.

    Premisecitation
    § Intrafascicular electrodes · p. 13

    Histology of animal nerves with metal or polymer longitudinal intrafascicular electrodes implanted for six months showed the implants were biocompatible and caused no damage, although a slight increase in connective tissue around the implant site was evident long-term.

    “The histology of animal nerves that had metal or polymer LIFEs implanted for a period of 6 months revealed that the implants are biocompatible and that no damage was caused by the presence of the electrode, although a slight increase in connective tissue surrounding the implant site was evident in the longterm (Lefurge et al., 1991; Lawrence et al., 2002) .”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites Lefurge et al., 1991 source
    • cites Lawrence et al., 2002 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Activity of embryonic mouse spinal cord cells has been recorded simultaneously over several months on a planar glass substrate with 64 iridium tin oxide electrodes.

    “The activity of embryonic spinal cord cells of mice has been recorded simultaneously over a period of several months on a planar glass substrate with 64 iridium tin oxide electrodes (Droge et al., 1986)”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Droge et al., 1986 source

      The Cell culture section describes recordings from embryonic mouse spinal-cord cells, and the paper describes simultaneous multisite recording. But its recording matrix had 36 gold-conductor electrodes, not 64 iridium tin oxide electrodes; the Results describe cultures examined 3–5 weeks after seeding, not recordings over several months. No page numbers are available in our copy. (LLM, conf=0.90)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Premisecitation
    § EMG-based control of exoskeletons · p. 21

    Different complex algorithms have been developed to extract force/torque information from sEMG signals to address these limitations, although in some cases good results have been achieved using a simple proportional control law.

    “Different complex algorithms have been developed to extract force/torque information from sEMG signals that address these limitations, although in some cases good results have been achieved using a simple proportional control law (Gordon and Ferris, 2004)”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Gordon and Ferris, 2004 source

      The Virtual object interface section describes a simple proportional law: processed surface EMG amplitudes generated virtual forces, and the Discussion reports that subjects improved their control with practice. The Discussion also mentions signal-processing algorithms for myoelectric prostheses, but our copy does not say those algorithms extract force or torque from sEMG; the study's proportional forces were virtual, not measured force or torque estimates. Page numbers are unavailable in our copy. (LLM, conf=0.85)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingcitation
    § Non-electrical interfaces · p. 7

    Patients learned to open and close the hand reliably after a few minutes of practice.

    “The patients learned to open and close the hand reliably after a few minutes of practice”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Barry et al., 1986

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationnone given
    § Interface Electrodes · p. 4

    Attempts to develop material surfaces that may be recognized as biological and form truly biocompatible interfaces with tissue are under investigation, although they have yet to be integrated in functional neural interfaces.

    “Attempts to develop material surfaces that may be recognized as biological and form truly biocompatible interfaces with the tissue are under investigation, although they have yet to be integrated in functional neural interfaces.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § Acritical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems · p. 1

    Extraneural electrodes such as cuff and epineurial electrodes provide simultaneous interface with many axons in the nerve.

    “Extraneural electrodes, such as cuff and epineurial electrodes, provide simultaneous interface with many axons in the nerve”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisenone given
    § Electrical · p. 2

    Nerve electrodes have the potential to damage the nerves on which they are implanted, whereas muscle electrodes are generally considered safer.

    “However, nerve electrodes also have the potential to damage the nerves on which they are implanted, whereas muscle electrodes are generally considered as safer”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given
    § Organization of the PNS · p. 3

    Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.

    “Signals are transmitted by the corresponding axons in series of impulses or action potentials, with intensity of the signal mainly coded in impulse frequency along the peripheral axon.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.
  • May combine two claims.

    Interpretationnone given
    § Electrical · p. 2

    Patients are able to use implantable neuroprostheses for activities of daily living and to enhance their quality of life and independence.

    “Patients are able to use the system for activities of daily living and enhancing their quality of life and independency.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Organization of the PNS · p. 3

    Nerve fibers are grouped in fascicles surrounded by connective tissue in the peripheral nerve.

    “Nerve fibers, both afferent and efferent, are grouped in fascicles surrounded by connective tissue in the peripheral nerve”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Peters et al., 1991

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Electrical · p. 2

    Most patients with nervous system injury suffer from an incomplete lesion, so modular, controllable systems for supporting functions are being developed.

    “New technological developments are taking into account the fact that the most patients with nervous system injury suffer from an incomplete lesion, and thus modular, controllable systems for supporting these functions are being developed”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rupp and Gerner, 2004 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given
    § Organization of the PNS · p. 3

    Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.

    “Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § Organization of the PNS · p. 3

    The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.

    “The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal with respect to proximal segments.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.
  • May combine two claims.

    Premisecitation
    § Electrical · p. 2

    Nerve electrodes implanted adjacent to, around, or within a peripheral nerve trunk or spinal root reduce the stimulus intensity required for activation compared with muscular or surface electrodes, thereby decreasing hazardous electrochemical processes and power consumption.

    “Due to the proximity to the nerve, the stimulus intensity required for activation is reduced in comparison with muscular or surface electrodes and, consequently, hazardous electrochemical processes and power consumption of the stimulator system can be decreased”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Loeb and Peck, 1996 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given
    § Organization of the PNS · p. 3

    The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.

    “The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given
    § Acritical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems · p. 1

    Intrafascicular, penetrating, and regenerative electrodes may contact small groups of axons within a nerve fascicle.

    “intrafascicular, penetrating, and regenerative electrodes may contact small groups of axons within a nerve fascicle”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Interface Electrodes · p. 4

    Compatibility between a technical and a biological system can be divided into structural biocompatibility and surface biocompatibility.

    “In general, the compatibility between a technical and a biological system (Bronzino, 1995) can be divided into the structural biocompatibility and the surface biocompatibility”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Bronzino, 1995

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisereasoning
    § Epimysial and intramuscular electrodes · p. 7

    Epimysial electrodes offer a good signal-to-noise ratio because of their position and large surface and provide exact selectivity, namely the implanted muscle.

    “In general, they offer a good signal-to-noise ratio because of their position and large surface and provide exact selectivity, namely the implanted muscle.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § Surface electrodes · p. 5

    For recording purposes, electrodes made of silver/silver chloride in the form of bars (10-1 mm) have been found to give adequate signal-to-noise ratio.

    “For recording purposes, electrodes made of silver/silver chloride in the form of bars (10 -1 mm) have been found to give adequate signal-to-noise ratio.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Surface electrodes · p. 5

    Multichannel surface EMG electrodes have been developed that enable measurements of motor unit potentials size and number and provide topographical information beyond the capability of intramuscular needle electrodes.

    “Multichannel (linear and high-density arrays) surface EMG electrodes have been developed that enable measurements of motor unit potentials size and number and provide topographical information that is beyond the capability of intramuscular needle electrodes”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Zwarts and Stegeman, 2003 source
  • May combine two claims.

    Premisecitation
    § Surface electrodes · p. 5

    Surface electrodes can also supply a perceivable sensation to the skin or excite nerve tracts or muscles underneath the skin.

    “they can also be employed to supply a perceivable sensation to the skin or to excite nerve tracts or muscles underneath the skin”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Patterson and Lockwood, 1993 source
    • cites Taylor et al., 1999 source
  • May combine two claims.

    Interpretationnone given
    § Non-electrical interfaces · p. 7

    Disadvantages of acoustic myography include susceptibility to interference by environmental noise and vibrations and limited experience with its use.

    “Disadvantagesinclude the susceptibility to interference by environmental noise and vibrations and limited experience with its use.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 8

    A final issue with percutaneous electrodes is that muscle contractions may result in eventual breakage, although these problems have become less frequent with the improvement of electrodes and lead wires.

    “A final issue with percutaneous electrodes is that muscle contractions may result in eventual breakage, although these problems have become less frequent with the improvement of electrodes and lead wires”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Onishi et al., 2000 source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Surface electrodes · p. 5

    The active recording electrode should be located near but not on the motor point because at this point the stability of signals suffers from minor physical displacement of the electrode.

    “The active recording electrode should be located near but not on the motor point because at this point the stability of signals suffers from minor physical displacement of the electrode”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites De Luca, 2002

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationreasoning
    § Non-electrical interfaces · p. 7

    Further technical developments to increase sensitivity of measurements are needed before magnetic interfaces can be used for single-unit activity recordings.

    “Further technical developments to increase sensitivity of measurements are needed before magnetic interfaces can be used for single-unit activity recordings.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisenone given
    § Epimysial and intramuscular electrodes · p. 7

    Although epimysial electrodes can be used for EMG recording, their main application area is for muscle stimulation.

    “Although they can be used for EMG recording, the main application area is for muscle stimulation”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given
    § Interface Electrodes · p. 5

    The number of electrodes to be used depends on the working application: a low number for robust use and limited functionality, or high numbers for good spatial resolution and selectivity.

    “The number of electrodes to be used depends on the working application: low number of electrodes for robust use and limited functionality or high numbers for good spatial resolution and selectivity”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisenone given
    § Non-electrical interfaces · p. 7

    Muscle sounds have been used to measure twitch force, monitor fatigue, and in the diagnosis of muscle diseases.

    “Muscle sounds have been used to measure twitch force, monitor fatigue and in the diagnosis of muscle diseases.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given
    § Non-electrical interfaces · p. 7

    As a control signal for externally powered prostheses, acoustic myography may present advantages over surface EMG because it is unaffected by changes in skin impedance, less sensitive to placement on the muscle, and requires less amplification and electrical shielding.

    “As a control signal for externally powered prostheses, acoustic myography may present some advantages over surface EMG because it is unaffected by changes in skin impedance, less sensitive to placement on the muscle, and requires less amplification and electrical shielding.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Surface electrodes · p. 5

    The impedance at the electrode-skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type.

    “The impedance at the electrode-skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Hewson et al., 2003 source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisereasoning
    § Interface Electrodes · p. 5

    Cuff and intrafascicular electrodes can be applied to intact nerves in acute or chronic studies, whereas regenerative sieve electrodes are implanted in transected nerves that need to regenerate across the electrode sieve over several months.

    “cuff and intrafascicular electrodes can be applied to intact nerves in acute or chronic studies, whereas, by definition, regenerative sieve electrodes are implanted in transected nerves that need to regenerate across the electrode sieve over several months.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 7

    Intramuscular electrodes may damage many muscle fibers, but in most instances this loss will not represent a noticeable reduction in muscle force.

    “These electrodes may damage many muscle fibers, but in most instances this loss will not represent a noticeable reduction in muscle force”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Mortimer et al., 1980 source
  • May combine two claims.

    Interpretationreasoning
    § Non-electrical interfaces · p. 7

    Measurements for magnetoneurography are usually performed inside a magnetically shielded room to minimize external contamination, limiting its application for control of neuroprostheses.

    “To minimize external contamination, measurements are usually performed inside a magnetically shielded room, thus limiting its application for control of neuroprostheses.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 7

    Compared with nerve cuff electrodes, intramuscular electrodes have been shown to produce more reliable, more graded, and less fatigable recruitment of motor units.

    “Compared with nerve cuff electrodes, intramuscular electrodes have been shown to produce more reliable, more graded, and less fatigable recruitment of motor units”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Singh et al., 2000
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § Epimysial and intramuscular electrodes · p. 7

    High energy consumption is required to activate the muscle with epimysial electrodes.

    “In addition, high energy consumption is required to activate the muscle.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning
    § Epimysial and intramuscular electrodes · p. 8

    To avoid problems derived from excess charge delivery, stimulation is given in cycles, and the system is explanted once regenerating axons reinnervate the muscles of interest.

    “To avoid problems derived from excess charge delivery, stimulation is given in cycles, and the system is explanted once regenerating axons reinnervate the muscles of interest.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 8

    The BION implantable microstimulator consists of a small cylindrical capsule (16-2 mm) whose internal components are connected to electrodes sealed hermetically into its ends.

    “This device consists of a small cylindrical capsule (16 -2 mm) whose internal components are connected to electrodes sealed hermetically into its ends.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Loeb et al., 2001 source
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 7

    Electrode and lead failures or infections have been found to be rare and the electrode response was stable over time in patients with intramuscular or epimysial electrodes during more than 3 years.

    “Electrode and lead failures or infections have been found to be rare and the electrode response was stable over time in patients with intramuscular or epimysial electrodes during more than 3 years”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Kilgore et al., 2003 source
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 8

    The currents required to produce contractions are high and may induce discomfort and tissue damage.

    “In addition, the currents required to produce contractions are high and may induce discomfort and tissue damage”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Chae and Hart, 1998 source
  • May combine two claims.

    Interpretationnone given
    § Epimysial and intramuscular electrodes · p. 8

    Adequate EMG signals can be obtained with epimysial and intramuscular electrodes, but the fatigue of the muscle and the derived change of the recorded signal have to be taken into account for long-term use.

    “Adequate EMG signals can be obtained with epimysial and intramuscular electrodes, but the fatigue of the muscle and the derived change of the recorded signal have to be taken into account for long-term use.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 9

    The BION electrode was found to be stable, biocompatible, and effective in experimental animals when implanted passively or used for muscle stimulation over long-term implantation.

    “Tests performed indicated that the BION is a stable, biocompatible, and effective electrode when implanted passively or used to stimulate muscles in experimental animals for long-term implantation (Cameron et al., 1998) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Cameron et al., 1998 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given
    § Epimysial and intramuscular electrodes · p. 7

    Management may be difficult when many epimysial electrodes on different muscles are used because of the interconnection between them, providing additional risks and possible source of failure.

    “Management may be difficult when many epimysial electrodes on different muscles are used because of the interconnection between them, providing additional risks and possible source of failure.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given
    § Surface electrodes · p. 5

    The main advantage of surface electrodes being non-invasive and easily adaptable is counteracted by disadvantages such as the need for daily placement and frequent calibration, and low reproducibility and quality of acquired signals.

    “Their main advantage of being non-invasive and easily adaptable is counteracted by several disadvantages, such as the need for daily placement and frequent calibration, and by the low reproducibility and quality of acquired signals.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisenone given
    § Surface electrodes · p. 5

    Recorded EMG signals are used to determine the activation time of a muscle, estimate the force produced during muscle contraction, and estimate the rate of muscle fatigue.

    “Currently, recorded EMG signals are used to determine the activation time of a muscle, to estimate the force produced during muscle contraction and to estimate the rate of muscle fatigue.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Interface Electrodes · p. 5

    Capacitive electrodes such as titanium nitride might reduce the risk of corrosion under stimulation conditions, but their performance highly varies with the fabrication technology.

    “Capacitive electrodes such as titanium nitride might reduce the risk of corrosion under stimulation conditions, but their performance highly varies with the fabrication technology”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Janders et al., 1996

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Weiland et al., 2002 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § Epimysial and intramuscular electrodes · p. 8

    For large muscles, the EMG activity mainly reflects the action potentials of motor units located more superficial or closer to the electrode.

    “For large muscles, the EMG activity mainly reflects the action potentials of motor units located more superficial or closer to the electrode.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationcitation
    § Epimysial and intramuscular electrodes · p. 8

    This requires the implantation and maintenance of many electrodes subjected to considerable mechanical stress caused by strong muscle contractions.

    “This requires the implantation and maintenance of many electrodes subjected to considerable mechanical stress caused by strong muscle contractions”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Devasahayam, 1992 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given
    § Interface Electrodes · p. 5

    The choice of an adequate electrode size and material, in combination with a structurally compatible design, is always a compromise between electrode impedance, signal-to-noise ratio, and selectivity.

    “The choice of an adequate electrode size and material, in combination with a 'structural compatible' design, is always a compromise between electrode impedance, signal-to-noise ratio, and selectivity.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Epimysial and intramuscular electrodes · p. 8

    Direct muscle stimulation may be used to recover chronically denervated muscles, although activation requires long duration pulses and extremely high current that are usually not generated by common FES devices.

    “Direct muscle stimulation may be also used to recover chronically denervated muscles, although activation requires long duration pulses and extremely high current that are usually not generated by common FES devices”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Kern et al., 2002 source
  • May combine two claims.

    Premisecitation
    § Interfascicular electrodes · p. 12

    There was no gross evidence of axonal damage one day after implantation of the slowly penetrating interfascicular nerve electrode, but chronic implants have not been reported.

    “There was no gross evidence of axonal damage after 1 day, but chronic implants have not been reported.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Tyler and Durand, 1997 source
  • May combine two claims.

    Premisecitation
    § Cuff electrodes · p. 10

    Quasi-trapezoidal stimulation pulses induce more gradual contraction than rectangular pulses, an effect attributed to activation of fatigue-resistant motor units before large fast-fatigue units with cuff electrodes.

    “For example, quasi-trapezoidal stimulation pulses have been shown to induce more gradual contraction than rectangular pulses, an effect that was attributed to activation of fatigue-resistant motor units before activation of large fast-fatigue units with cuff electrodes (Fang and Mortimer, 1991; Rozman et al., 1993) .”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Fang and Mortimer, 1991 source
    • cites Rozman et al., 1993 source
  • May combine two claims.

    Premisecitation
    § Cuff electrodes · p. 10

    Using short pulsewidths and a subthreshold transverse current from a steering anode restricts the region of excitation of the nerve trunk and significantly improves stimulation selectivity.

    “The use of short pulsewidths and the introduction of a subthreshold transverse current from a steering anode allows restriction of the region of excitation of the nerve trunk, significantly improving the selectivity of stimulation (Gorman and Mortimer, 1986; Sweeney et al., 1990; Grill and Mortimer, 1996b; Navarro et al., 2001) .”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Gorman and Mortimer, 1986

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Sweeney et al., 1990 source
    • cites Grill and Mortimer, 1996b

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § Cuff electrodes · p. 10

    Chronic implantation of snug cuffs modifies nerve shape and produces loss of large nerve fibers, which are the most sensitive to compression.

    “Although snug-fitting nerve cuffs have been advocated to reduce the stimulus charge injection or to obtain a high signal-to-noise ratio for neural recordings (Naples et al., 1990) , different studies have shown that chronic implantation of snug cuffs modifies the nerve shape and produces a loss of large nerve fibers, which are the most sensitive to compression (Krarup et al., 1989; Grill and Mort”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Krarup et al., 1989 source
    • cites Grill and Mortimer, 1998 source
    • cites Larsen et al., 1998 source
  • May combine two claims.

    Premisecitation
    § Cuff electrodes · p. 11

    A tripolar electrode configuration is preferred to reduce noise pickup and provides a biophysically ideal way to use a differential amplifier to reject large bioelectric signals arising outside the cuff.

    “Tripolar (central active electrode vs. two reference electrodes near the ends of the cuff) electrode configuration is preferred to reduce noise pickup (Stein et al., 1977) ; it provides a biophysically ideal way to use a differential amplifier to reject almost completely the large bioelectric signals arising from outside the cuff (e.g., EMG).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Stein et al., 1977 source
  • May combine two claims.

    Premisecitation
    § Interfascicular electrodes · p. 12

    Histological cross-sections showed that the slowly penetrating interfascicular nerve electrode rearranges the epineurium and penetrates deep within a multifascicular nerve but does not disrupt the perineurium.

    “Histological cross-sections showed that the SPINE rearranges the epineurium and penetrates deep within a multifascicular nerve but did not disrupt the perineurium (Tyler and Durand, 1997) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Tyler and Durand, 1997 source
  • May combine two claims.

    Premisecitation
    § Flat-interface nerve electrodes · p. 11

    Flattening a nerve into a more elliptical shape makes fascicles more accessible and moves central fibers closer to the stimulating electrode compared with cylindrical cuffs.

    “By flattening the nerve into a more elliptical shape, fascicles become more accessible and central fibers are moved closer to the stimulating electrode in comparison with cylindrical cuffs.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Tyler and Durand, 2002 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Book electrodes · p. 9

    The first micturition caused by electrical nerve root stimulation was reported in 1973.

    “The first micturition because of electrical nerve root stimulation was reported in 1973 (Brindley, 1973) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Brindley, 1973

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Postoperative infections after helicoidal electrode implants in vagus nerve patients occur in less than 3% of cases, and removal of the device is required in 1%.

    “Postoperative infections after helicoidal electrode implants in vagus nerve patients occur in less than 3% of cases, requiring removal of the device in 1%.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites McLachlan, 1997 source
  • May combine two claims.

    Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Helicoidal electrodes conform to the shape of the nerve and minimize mechanical trauma.

    “Helicoidal electrodes, placed circumjacent to the nerve, are made of flexible platinum ribbon with an open helical design (Fig. 5), allowing the electrode to conform to the shape of the nerve and minimizing mechanical trauma (Naples et al., 1990) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Naples et al., 1990

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § Flat-interface nerve electrodes · p. 12

    Flat-interface nerve electrodes that moderately flattened the nerve demonstrated the best selectivity for limb motion measurements, and this selectivity was maintained throughout the implant time.

    “Electrodes that moderately flattened the nerve demonstrated the best selectivity for limb motion measurements, which were maintained throughout the implant time (Tyler and Durand, 2003; Leventhal and Durand, 2004) .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Tyler and Durand, 2003 source
    • cites Leventhal and Durand, 2004 source
  • May combine two claims.

    Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Using epineurial electrodes for neuropathic pain has limited benefit because it should be applied only when pain involves one nerve territory in the head or limbs, and it has frequent morbidity from electrode approximation, electrode displacement, and lead failure.

    “The latter application is of limited benefit because it should be applied only when pain involves one nerve territory in the head or the limbs and has frequent morbidity associated with electrode approximation, electrode displacement, and lead failure (Hassenbusch et al., 1996) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Hassenbusch et al., 1996 source
  • May combine two claims.

    Premisecitation
    § Flat-interface nerve electrodes · p. 12

    Acute experiments showed that individual fascicles of the cat sciatic nerve, and groups of fibers within fascicles, can be selectively activated with the flat-interface nerve electrode, and selectivity strongly depends on the relative locations of the fascicle and electrode contacts.

    “Results on acute experiments showed that it is possible to selectively activate individual fascicles of the cat sciatic nerve with the FINE, as well as groups of fibers within the fascicles, and revealed the strong dependency of selectivity on the relative locations of the fascicle and the electrode contacts (Tyler and Durand, 2002; Leventhal and Durand, 2003) .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Tyler and Durand, 2002 source
    • cites Leventhal and Durand, 2003 source
  • May combine two claims.

    Premisereasoning
    § Book electrodes · p. 9

    Ventral root stimulation causes simultaneous contraction of the sphincter and detrusor muscles, so voiding occurs in brief spurts at the end of stimulation when the striated sphincter relaxes faster than the smooth detrusor.

    “Because ventral root stimulation produces simultaneous contraction of both sphincter and detrusor muscles, voiding is achieved in brief spurts at the end of the stimulation, when the striated sphincter muscle relaxes faster than the smooth detrusor muscle.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Occasional fracture of the electrode wire has been reported for helicoidal electrodes.

    “Occasional fracture of the electrode wire has been reported (McLachlan, 1997) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites McLachlan, 1997 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Book electrodes · p. 9

    There are now 2,000 patients worldwide with a urinary bladder stimulator.

    “Nowadays, there are 2,000 patients worldwide that carry a urinary bladder stimulator (Rijkhoff, 2004) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rijkhoff, 2004 source
  • May combine two claims.

    Premisecitation
    § Epineurial and helicoidal electrodes · p. 9

    Epineurial electrodes have been used in biomedical functional electrical stimulation for breathing control by phrenic nerve stimulation, foot-drop improvement, and relief of neuropathic pain.

    “Epineurial electrodes are used in biomedical FES applications for breathing control by phrenic nerve stimulation (Creasey et al., 1996) , foot-drop improvement (Liberson et al., 1961) , and relief of neuropathic pain (Strege et al., 1994) .”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Creasey et al., 1996

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Liberson et al., 1961

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Strege et al., 1994 source
  • May combine two claims.

    Premisecitation
    § Flat-interface nerve electrodes · p. 12

    Chronic studies in laboratory animals with flat-interface nerve electrodes implanted for 1 to 3 months showed that high reshaping force induced nerve damage, whereas moderate and small forces caused no detectable change in nerve physiology and histology.

    “Chronic studies in laboratory animals with FINE implanted over 1-3 months showed that electrodes applying high reshaping force induced nerve damage, whereas those with moderate and small forces did not cause any detectable change in nerve physiology and histology.”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Tyler and Durand, 2003 source
    • cites Leventhal and Durand, 2004 source
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    The applicability of regenerative electrodes depends on successful axonal regeneration through the perforations or holes, the possibility of nerve damage from mechanical load or constrictive forces within the holes, and the biocompatibility of the components.

    “Applicability of regenerative electrodes is dependent on the success of axonal regeneration through the perforations or holes, the possibility of nerve damage from the mechanical load imposed by the electrode or from constrictive forces within the holes, and the biocompatibility of the components (Rosen et al., 1990; Navarro et al., 1996) .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Rosen et al., 1990 source
    • cites Navarro et al., 1996 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    Recruitment curves for the Utah electrode array were broader, with twitch thresholds starting at much lower currents than usual for cuff electrodes.

    “Recruitment curves for the electrode array were broader with twitch thresholds starting at much lower currents than usual for cuff electrodes.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Branner et al., 2001 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Penetrating microelectrodes · p. 13

    A study used a microelectrode to stimulate the sensory nerve of an operator to indicate contact with an object by a remote robotic hand.

    “An interesting study used a microelectrode to stimulate the sensory nerve of an operator to indicate contact with an object by a remote robotic hand (Shimojo et al., 2003) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Shimojo et al., 2003

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Penetrating microelectrodes · p. 15

    A negative aspect in the volunteer case was gradual degradation of the electrode wire bundle.

    “One negative aspect was the gradual degradation of the electrode wire bundle.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Warwick et al., 2003 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    A slanted array with electrodes of varying length from 0.5 to 1.5 mm and 0.1 mm difference between rows of neighboring electrodes was developed to decrease redundant electrodes and provide access to more fascicles within the nerve.

    “To decrease the number of redundant electrodes and provide access to more fascicles within the nerve, a slanted array with electrodes of varying length, ranging from 0.5 to 1.5 mm with 0.1 mm difference in length between rows of neighboring electrodes, has been developed (Branner et al., 2001) .”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Branner et al., 2001 source
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    Using multiple-hole silicon arrays, researchers demonstrated axonal regeneration and neural activity recording in peripheral nerves of rat, frog, and fish.

    “Using multiple-hole silicon arrays, the researchers demonstrated axonal regeneration and even neural activity recording was demonstrated in peripheral nerves of rat, frog, and fish (Kovacs et al., 1994; Navarro et al., 1996; Bradley et al., 1997; Della Santina et al., 1997; Mensinger et al., 2000) .”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Kovacs et al., 1994 source
    • cites Navarro et al., 1996 source
    • cites Bradley et al., 1997 source
  • May combine two claims.

    Premisecitation
    § Intrafascicular electrodes · p. 13

    Activity from cutaneous receptors in response to stimuli has been recorded in several studies, but the activity was small and difficult to distinguish from background noise.

    “Activity arising from cutaneous receptors in response to stimuli has been recorded in several studies, although activity was found to be small and difficult to distinguish from the background noise in the recording (Malmstrom et al., 1998; Yoshida et al., 2000) .”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Malmstrom et al., 1998 source
    • cites Yoshida et al., 2000 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Penetrating microelectrodes · p. 14

    Rutten and colleagues concluded that an electrode separation of 120 μm was optimal for selective interfacing of peripheral nerves such as the rat peroneal nerve containing around 350 alphamotor fibers.

    “In a series of experimental and modeling studies, Rutten and colleagues concluded that an electrode separation of 120 m m was optimal for selective interfacing peripheral nerves, such as the rat peroneal”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rutten, 2002 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 14

    A trade-off must be made between selectivity of stimulation and the total number of nerve fibers that can be activated, and optimal selectivity requires redundant electrodes in a 3D configuration.

    “Thus, a trade-off must be made between selectivity of stimulation and the total number of nerve fibers that can be activated; optimal selectivity requires the use of redundant electrodes in the 3D configuration (Rutten et al., 1999) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rutten et al., 1999 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 13

    Microstimulation through a microneurography electrode can activate single axons, although in most cases it affects small groups of axons in the fascicle.

    “Microstimulation through the electrode can be used to activate single axons, although in most cases stimulation affects small groups of axons in the fascicle (Ochoa and Torebjo ¨ rk, 1983; Torebjo ¨ rk et al., 1987) .”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Ochoa and Torebjo ¨ rk, 1983 source
    • cites Torebjo ¨ rk et al., 1987

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    Regenerating axons grow through the holes of regenerative electrodes, making it possible to record action potentials from and stimulate individual axons or small fascicles.

    “Regenerating axons eventually grow through the holes (Fig. 12), making it possible to record action potentials from and to stimulate individual axons or small fascicles.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Llina ´ s et al., 1973

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Edell, 1986 source
    • cites Kovacs et al., 1992 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    The volunteer received feedback from force and slip sensors in a prosthetic hand and used the array to control the hand for grasping an unseen object.

    “The subject received feedback information from force and slip sensors in a prosthetic hand and subsequently used the array to control the hand for grasping an unseen object.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Warwick et al., 2003 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    Stimulation properties stabilized in 80% of the Utah array electrodes over the course of the experiment, but the recorded sensory signals were not stable over time.

    “The stimulation properties stabilized in 80% of the electrodes over the course of the experiment, but the recorded sensory signals were not stable over time.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Branner et al., 2004 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    The Utah multielectrode array can record more effectively from more dorsal root ganglion neurons than conventional recording techniques, providing a more stable option for chronic implantation.

    “The Utah MEA is also capable of recording more effectively from more dorsal root ganglion neurons than has been achieved by conventional recording techniques, providing a more stable option for chronic implantation (Aoyagi et al., 2003) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Aoyagi et al., 2003 source
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    A one-to-one design allowing access to each individual regenerated axon through one hole of 2-10 μm diameter has proved impossible, because nerve regeneration fails with holes of such small size.

    “Ideally a one-to-one design would allow access to each individual regenerated axon grown through one hole (2-10 m m in diameter). However, this has been proved impossible; nerve regeneration fails with holes of such small size.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Edell, 1986 source
    • cites Rosen et al., 1990 source
    • cites Navarro et al., 1996 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    Electrodes in the Utah array selectively recorded single-unit responses from mechanoreceptors, although only in 10-20% of the electrodes, and evoked graded recruitment of force in muscle groups in a highly selective fashion with current injections in the 1-20 μA range.

    “Electrodes in the array were capable of selectively recording single-unit responses from mechanoreceptors, although only in 10-20% of the electrodes, and evoked graded recruitment of force in muscle groups in a highly selective fashion with current injections in the 1-20 m A range (Branner and Norman, 2000; Branner et al., 2001) .”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Branner and Norman, 2000 source
    • cites Branner et al., 2001 source
  • May combine two claims.

    Premisecitation
    § Intrafascicular electrodes · p. 13

    Longitudinal intrafascicular electrodes offer good selectivity for stimulation and for multiunit extracellular recording, and their geometry makes them suitable for long-term implantation.

    “The LIFEs offer good selectivity for stimulation (Nannini and Horch, 1991; Yoshida and Horch, 1993) and for multiunit extracellular recording (Goodall et al., 1991; McNaughton and Horch, 1994; Yoshida and Stein, 1999) , and their geometry, make them suitable for longterm implantations (Lefurge et al., 1991) .”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Nannini and Horch, 1991 source
    • cites Yoshida and Horch, 1993 source
    • cites Lefurge et al., 1991 source
  • May combine two claims.

    Premisecitation
    § Intrafascicular electrodes · p. 13

    The use of multistrand Kevlar longitudinal intrafascicular electrodes increases tensile strength and signal-to-noise ratio.

    “The use of multistrand Kevlar LIFEs (m-polyLIFE) increases the tensile strength and signal-to-noise ratio (Lawrence et al., 2004) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Lawrence et al., 2004 source
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    Silicon regenerative interfaces cause frequent signs of axonopathy and constitute a physical barrier that limits elongation of regenerating axons depending on the size of the holes.

    “However, such silicon interfaces cause frequent signs of axonopathy and constitute a physical barrier that limits the elongation of regenerating axons depending on the size of the holes (Edell, 1986; Rosen et al., 1990; Navarro et al., 1996; Zhao et al., 1997) .”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites Edell, 1986 source
    • cites Rosen et al., 1990 source
    • cites Navarro et al., 1996 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 15

    Both recording and stimulation with the Utah array were stable over the 36-hour-long period of the experiments.

    “Both the recording and stimulation were stable over the 36-h-long period of the experiments.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Branner et al., 2001 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 14

    Loss of neurons in the spiral ganglion and cochlear nucleus was found after cochlear nerve microelectrode implants, thought to be due to difficulty with electrode insertion and chronic motion of the implants.

    “Nevertheless, the loss of neurons in spiral ganglion and cochlear nucleus was found, thought to be due to difficulty with electrode insertion and chronic motion of the implants.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Arts et al., 2003 source
  • May combine two claims.

    Premisecitation
    § Penetrating microelectrodes · p. 14

    A silicon-based ribbon electrode with high flexibility functioned successfully after chronic implantation of up to one year, but its interconnect cable was not designed for the highly mobile somatic peripheral nerve and was too stiff and brittle for intrafascicular application.

    “A silicon-based ribbon electrode with high flexibility demonstrated successful function after chronic implantation of up to 1 year (Hetke et al., 1994) , but this interconnect cable was not designed for implants in the highly mobile somatic peripheral nerve and was too stiff and brittle for intrafascicular application (Najafi and Hetke, 1990) .”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Hetke et al., 1994 source
    • cites Najafi and Hetke, 1990 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given
    § Bioelectronic interfaces · p. 19

    In the PNS, bioelectronic interfaces are in an early and experimental stage.

    “In the PNS, bioelectronic interfaces are in an early and experimental stage.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Pine et al. placed single rat hippocampus cells into a 3D silicon structure and demonstrated that the transplanted cells retained capacity for axonal outgrowth and that nerve action potentials could be recorded extracellularly from the soma.

    “Pine et al . (1994) managed to place single hippocampus cells of the rat into a 3D silicon structure, which was termed 'cultured neuron probe.' They could demonstrate that the transplanted cells inside the microcompartments retained their capacity for axonal outgrowth and that nerve action potentials could be directly recorded extracellularly from the soma of the cells.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Pine et al., 1994

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Fromherz et al. obtained a current-free measurement of the membrane potential of a single leech neuron by direct coupling to the gate of a field effect transistor.

    “they obtained a current-free measurement of the membrane potential of a single neuron of a leech by direct coupling to the gate of a field effect transistor (FET) (Fromherz et al., 1991a)”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Fromherz et al., 1991a

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    Chronically implanted regenerative electrodes have allowed stimulation of different nerve bundles and recording of nerve action potentials in response to functional stimulation.

    “been possible to stimulate different nerve bundles and to record nerve action potentials in response to functional stimulation (Navarro et al., 1998; Ceballos et al., 2002)”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Navarro et al., 1998

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Ceballos et al., 2002 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § EMG-based control of hand prostheses · p. 20

    The formal scheme for EMG acquisition and analysis for prosthetic control includes signal conditioning, pre-processing, onset detection, feature extraction, dimensionality reduction, pattern recognition, and off-line and on-line learning.

    “the formal scheme for the acquisition and analysis of EMG signals for the control of prosthetic devices is composed of several modules (Zecca et al., 2002) : (1) signal conditioning, pre-processing, and detection of onset movement (Micera et al., 1999) ; (2) feature extraction; (3) dimensionality reduction; (4) pattern recognition; and (5) off-line and on-line learning.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Zecca et al., 2002 source
    • cites Micera et al., 1999 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Materials for substrates, insulation, and encapsulation · p. 18

    Fabrication of silicon-based interconnection cables is technologically feasible.

    “The fabrication of silicon-based interconnection cables was shown to be technologically feasible (Hetke et al., 1990; 1994)”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Hetke et al., 1990; 1994

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • May combine two claims.

    Interpretationreasoning
    § Regenerative electrodes · p. 16

    Interfacing regenerated axons with regenerative electrodes requires time, so acute experiments are not possible.

    “some time is needed for interfacing the regenerated axons, thus precluding acute experiments”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    The presence of a polyimide-based regenerative electrode showed no chronic foreign body response, and the immunohistochemically measured pattern of characteristic proteins was comparable with a healing reaction without any implant.

    “The presence of a polyimide-based regenerative electrode showed no chronic foreign body response, and the immunohistochemically measured pattern of characteristic proteins was comparable with a healing reaction without any implant (Klinge et al., 2001a) .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Klinge et al., 2001a source
  • May combine two claims.

    Interpretationnone given
    § EMG-based control of hand prostheses · p. 20

    In EMG-based control of hand prostheses, homologous muscles usually cannot be used to control prosthetic movements, so complex pattern recognition algorithms are required.

    “In this case, it is usually not possible to use the 'homologous' muscles to control the movements of the prosthetic device, and the development of a complex algorithm exploiting the potentialities of using advanced pattern recognition techniques is required.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § EMG-based control of hand prostheses · p. 20

    The EMG signal was modeled as amplitude-modulated Gaussian noise whose variance is related to the force developed by the muscle.

    “the EMG signal was modeled as the amplitude-modulated Gaussian noise whose variance is related to the force developed by the muscle.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Graupe and Cline, 1975; 1982

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Graupe et al., 1982 source
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given
    § EMG-based control of neuroprostheses and hybrid bionic systems · p. 20

    sEMG signals are easy to record and provide important access to the neuromuscular system of the user.

    “sEMG signals are easy to record and provide an important access to the neuromuscular system of the user.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Bioelectronic interfaces · p. 19

    The structure's topology, such as the width of grooves, has an important impact on the growth of cultured cells.

    “The structure's topology (e.g., the width of grooves) has an important impact on the growth of the cultured cells (Wilkinson et al., 1987; Clark et al., 1990; Connolly et al., 1992)”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Wilkinson et al., 1987 source
    • cites Clark et al., 1990 source
    • cites Connolly et al., 1992 source
  • May combine two claims.

    Premisecitation
    § Materials for substrates, insulation, and encapsulation · p. 18

    Silicone elastomer remains the material of choice for encapsulating cables and for additional coating of hermetically sealed electronic circuits.

    “Silicone elastomer is still the material of choice for encapsulation of cables and also as additional coating for hermetically sealed electronic circuits (Stieglitz et al., 2004)”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Stieglitz et al., 2004 source
  • May combine two claims.

    Premisecitation
    § Regenerative electrodes · p. 16

    Polyimide-based electrodes have been shown to be biocompatible and stable over months of in vivo implantation and allow much better regeneration than silicon dice.

    “Polyimide-based electrodes have been shown to be biocompatible and stable over months of in vivo implantation and allow for much better regeneration than silicon dice (Navarro et al., 1998; Ceballos et al., 2002; Lago et al., 2005) .”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Navarro et al., 1998

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Ceballos et al., 2002 source
    • cites Lago et al., 2005 source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Fromherz et al. demonstrated guided growth of leech nerve cells on a laminin-coated substrate.

    “Fromherz et al . (1991b) could demonstrate the guided growth of nerve cells from leeches on a laminin-coated substrate.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Fromherz et al., 1991b

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationreasoning
    § Regenerative electrodes · p. 16

    Regenerative electrodes can be implanted in severed nerves of an amputee's limb to create a bidirectional interface for a feedback-controlled neuroprosthesis.

    “One of the most logical and challenging applications of regenerative electrodes consists of their implantation in severed nerves of an amputee's limb for bidirectional interface in a feedback-controlled neuroprosthesis.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Bioelectronic interfaces · p. 19

    In the cone electrode, a piece of rat sciatic nerve was placed into a cone, and after implantation in rat brain, neurite ingrowth was observed and stable recordings were obtained over 6 months.

    “Here, a piece of a rat sciatic nerve was placed into a cone of 1.5 mm length and 100-200 m mdiameter. After implantation in the brain of a rat, neurite ingrowth was observed and stable recordings were obtained over 6 months.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Kennedy, 1989 source
  • May combine two claims.

    Premisereasoning
    § Bioelectronic interfaces · p. 19

    Neither the cultured neuron probe nor the cone electrode approach has been investigated in the PNS.

    “Neither approach has been investigated in the PNS.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Materials for the electrodes · p. 18

    Platinum, iridium, tungsten, and stainless steel are typically used as conductors, and silicone elastomer, polytetrafluoroethylene, and polyimide as insulating carriers in neural electrodes.

    “Materials typically used are platinum, iridium, tungsten, and stainless steel as conductors and silicone elastomer, polytetrafluoroethylene, and polyimide as insulating carriers (Naples et al., 1990; Heiduschka and Thanos, 1998)”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Naples et al., 1990

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites Heiduschka and Thanos, 1998 source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § Bioelectronic interfaces · p. 19

    With surface topologies on the substrate, axonal outgrowth of embryonic rat nerve cells was observed and electrical activity was recorded in a microstructure with 16 electrodes connected with grooves.

    “Having surface topologies on the substrate, axonal outgrowth of embryonic nerve cells from rats was observed and electrical activity was recorded in a microstructure with 16 electrodes that were connected with grooves (Jimbo and Kawana, 1992; Jimbo et al., 1993)”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Jimbo and Kawana, 1992 source
    • cites Jimbo et al., 1993 source
  • May combine two claims.

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Reinnervation of the gastrocnemius and tibialis muscles via the distal stump of the sciatic nerve was observed after a few weeks, although there was no communication with the CNS.

    “Reinnervation of the gastrocnemius and tibialis muscles via the distal stump of the sciatic nerve was observed after a few weeks, although there was no communication with the CNS.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Katsuki et al., 1997 source
  • Classified as probably not a substantial claim.

    Premisecitation
    § Regenerative electrodes · p. 16

    Stimulation of afferent nerve fibers in the residual limb can provide sensory feedback from tactile and force sensors to an amputee.

    “sensory feedback from tactile and force sensors might be provided to the user through stimulation of afferent nerve fibers within the residual limb (Riso, 1999)”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Riso, 1999 source
  • May combine two claims.

    Interpretationreasoning
    § Materials for substrates, insulation, and encapsulation · p. 18

    Cost, technology, and limited cable length restrict widespread use of silicon-based interconnection cables in neural prostheses.

    “but cost, technology, and limited cable length limit widespread use in neural prostheses applications”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given
    § Choice of fabrication technology · p. 18

    Devices made by precision mechanics are robust, stable, and reliable but have low spatial resolution, low electrode selectivity, and quite large size.

    “The devices are robust, stable, and reliable but suffer from a low spatial resolution and low selectivity of the electrodes as well as from a quite large size.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § Bioelectronic interfaces · p. 19

    Functional regeneration was proven, but the number of surviving cells decreased from 10,000 injected to some tens that finally innervated the muscle.

    “Functional regeneration was proven but the number of surviving cells tremendously decreased from the 10,000 that had been injected to some tens that finally innervated the muscle (T. Stieglitz, 2004. The neuron micro probe project, personal communication)”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites T. Stieglitz, 2004. The neuron micro probe project, personal communication source
  • May combine two claims.

    Interpretationreasoning
    § Materials for the electrodes · p. 18

    Deterioration of an implanted electrode may cause implant failure and release toxic products.

    “deterioration of the device may result in implant failure and the release of toxic products”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation
    § EMG-based control of exoskeletons · p. 21

    Using this neuro-fuzzy approach, it was possible to control different degrees of freedom of the shoulder while reducing the muscular effort of the subjects.

    “Using this approach, it was possible to control the different degrees of freedom of the shoulder reducing the muscular effort of the subjects.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Kiguchi et al., 2001; 2003

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisereasoning
    § ENG-based control of neuroprostheses and hybrid bionic systems · p. 22

    Neuro-fuzzy models seem most adequate for processing ENG because of their model-free structure.

    “Neuro-fuzzy models seem the most adequate because of their model-free structure.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation
    § EMG-based control of exoskeletons · p. 21

    Experimental tests have shown that use of sEMG significantly improved the mechanical gain of the system while maintaining natural control.

    “Experimental tests have shown that the use of sEMG significantly improved the mechanical gain of the system (i.e., the possibility of augmenting capabilities and/or reducing human muscular efforts to carry out the different tasks) while maintaining natural control of the system.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rosen et al., 2001 source
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given
    § Conclusion · p. 23

    At present, there is no universal best choice for all possible uses of the described PNS interfaces.

    “At the moment, there is no universal best choice for all possible uses of the described interfaces”
    Quote in the paperNot a substantial claim
  • The quote only partly states this claim.

    Premisecitation
    § EMG-based control of exoskeletons · p. 21

    Using this myoprocessor scheme, a one-degree-of-freedom exoskeleton was controlled using dynamic, kinematic, and physiological feedback.

    “Using this schemea one-degree-of-freedom exoskeleton has been controlled using a multiple-feedback approach: (1) dynamic feedback - the moments generated at the interfaces between the human arm, the external load, and the exoskeleton structure; (2) kinematic feedback -the joint angular trajectories recorded using encoders, and the angular velocity and acceleration calculated by finite differences;”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rosen et al., 2001 source
  • May combine two claims.

    Interpretationnone given
    § EMG-based control of hand prostheses · p. 21

    The continuous classifier approach seems promising for EMG-controlled hand prostheses, but EMG signals seem limited to prostheses with few degrees of freedom, probably no more than four.

    “This approach seems promising for the development of EMG-based controlled hand prostheses although the use of EMG signals seems to be limited to prosthesis with a few degrees of freedom (probably no more than four).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationnone given
    § EMG-based control of hand prostheses · p. 21

    With the EMG approach, the user must initiate all contractions to control the prosthesis from rest, making it difficult to switch between classes and increasing response time in unexpected situations.

    “by using this approach, the user must initiate all the contractions to control the prosthesis from rest, making it difficult to switch from one class to another and increasing the response time of the user when dealing with unexpected situations.”
    Quote in the paperSubstantial claim
127 of 127
  • The paper relies on this without support.

    Nerve fibers are recruited by electrical stimulation according to their thickness, so large motor fibers are activated earlier than thinner fibers.

    “nerve fibers are recruited by electrical stimulation according to their thickness, therefore large motor fibers innervating the fast-fatigue and strong motor units are activated earlier than the physiologically first-recruited thinner fibers controlling slow fatigue-resistant motor units”
  • The paper relies on this without support.

    Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.

    “Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.”
  • The paper relies on this without support.

    Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.

    “Signals are transmitted by the corresponding axons in series of impulses or action potentials, with intensity of the signal mainly coded in impulse frequency along the peripheral axon.”
  • The paper relies on this without support.

    The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.

    “The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal with respect to proximal segments.”
  • The paper relies on this without support.

    The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.

    “The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.”
  • The paper relies on this without support.

    Using multiple wire needles to interface a high number of nerve fibers has practical problems for long-term use.

    “However, the use of multiple wire needles for interfacing a high number of nerve fibers has practical problems for long-term use.”
  • The paper relies on this without support.

    No human implants of regenerative electrodes have been reported.

    “No human implants of regenerative electrodes have been reported.”
  • The paper relies on this without support.

    Cuff electrodes are the type of peripheral nervous system interface electrode most investigated in basic and applied research.

    “Cuff electrodes are the type of PNS interface electrode most investigated in basic and applied research.”
  • The paper relies on this without support.

    The neuromuscular system has non-linear and time-variant characteristics.

    “the non-linear and time-variant characteristics of the neuromuscular system”
  • The paper relies on this without support.

    A traumatic lesion of a peripheral nerve leads to Wallerian degeneration of the distal nerve.

    “A traumatic lesion of a peripheral nerve leads to Wallerian degeneration of the distal nerve”
  • The paper relies on this without support.

    Healing of a peripheral nerve lesion occurs only under favorable conditions involving sprouting of axons from the proximal stump, outgrowth along the distal stump, and reinnervation of target tissues.

    “healing of such a lesion occurs only under favorable conditions involving sprouting of axons from the proximal nerve stump, outgrowth along the distal nerve stump, and reinnervation of target tissues”
  • The paper relies on this without support.

    Recovery from a peripheral nerve lesion is usually delayed and limited if the lesion site is very proximal.

    “Recovery is usually delayed and limited if the lesion site is very proximal.”
  • The paper relies on this without support.

    Muscle atrophy prevents formation of new neuromuscular junctions at a later time when regenerating nerve fibers may reach the muscle.

    “muscle atrophy prevents the formation of new neuromuscular junctions at a later time point when regenerating nerve fibers may reach the muscle”
  • The paper relies on this without support.

    Regenerated nerve fibers have smaller than normal diameter and internodal length.

    “the smaller than normal diameter and internodal length of regenerated fibers”
14 of 14

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Motor and sensory nerve signals gathered using cuff electrodes have been used to guide and fine-tune prostheses such as artificial limbs in human patients.

    “Although more challenging to carry out compared to nerve stimulation due to the distance between electrodes and the weak electrical signals transmitted by axons, with an ECM-rich epineurium separating the two, motor and sensory nerve signals gathered using cuff electrodes have been used to guide and fine-tune prostheses such as artificial limbs in human patients (Micera et al., 2001; Navarro et al”

    From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005

    The ENG-based control section (p. 22) reports that cuff electrodes recorded sensory signals used to regulate grasping in a hand neuroprosthesis, and that cuff-recorded signals predicted ankle movements in an animal model. It separately reports hand-prosthesis control in human amputees using LIFE electrodes, not cuffs. The paper therefore does not support the claim that cuff-recorded motor and sensory signals were used to guide artificial limbs in human patients. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • For nerve stimulation at the clinical level, epineurial cuffs remain the preferred choice.

    “For nerve stimulation at the clinical level, epineurial cuffs remain the preferred choice (Navarro et al., 2005).”

    From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005

    The Cuff electrodes section (p. 10) says cuffs have been implanted in patients for decades and can stimulate nerves. But it does not call “epineurial cuffs” the preferred clinical choice; the Epineurial and helicoidal electrodes section (p. 9) instead describes button-type epineurial electrodes replacing cuffs in some applications to reduce injury risk. (LLM, conf=0.85)

    Experimental — accuracy not yet measured
  • A thickening of the epineurium as part of FBR can dissipate the already weak electrical signals transmitted by axons, limiting the use of these devices for electrical recording applications in long-term implantations.

    “A thickening of the epineurium as part of FBR can dissipate the already weak electrical signals transmitted by axons, limiting the use of these devices for electrical recording applications in long-term implantations (Navarro et al., 2005).”

    From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005

    The Intrafascicular electrodes section (pp. 12–13) links fibrous encapsulation around an electrode to a gradual decrease in recorded axon-potential amplitude, supporting a recording-related effect of tissue growth. It does not identify thickening of the epineurium as the cause or say that this response limits long-term recording generally; the Cuff electrodes section (p. 11) reports successful long-term nerve-activity recording. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Based on the degree of invasiveness, nerve neuroprosthetic designs can be categorized into three groups: epineurial cuff electrodes, intraneural penetrating electrodes, and regenerative nerve electrodes.

    “Based on the degree of invasiveness nerve neuroprosthetic designs can be categorized into three different groups (summarized in Figures 4, 5).”

    From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005

    The Abstract (p. 1) distinguishes extraneural cuff and epineurial electrodes from intrafascicular, penetrating, and regenerative electrodes; the Cuff electrodes section (p. 10) also compares their invasiveness. The paper does not present the asserted three-group classification: it treats cuff and epineurial electrodes as distinct designs and discusses intrafascicular and interfascicular electrodes separately (p. 12). (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Most commonly the nerve lesion for regenerative implants is a full nerve transection where the two ends are cleanly separated and a gap forms.

    “Most commonly this is in the form of a full nerve transection, where the two ends of the nerve are cleanly separated and a gap forms between the two (Navarro et al., 2005; Thompson et al., 2015).”

    From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005

    The Regenerative electrodes section (p. 16) says these electrodes are implanted between severed nerve stumps and that multichannel regenerative electrodes can be applied only to transected nerves. It does not say that full transection is the most common lesion for regenerative implants, or specify that the ends are cleanly separated and a gap forms. (LLM, conf=0.86)

    Experimental — accuracy not yet measured

Cites

Flagged first, then by citations

Research works printed but not linked: 18

  1. Brindley GS, 1995, Neuroprostheses., 15
  2. Bronzino JD, 1995, Biomedical Engineering Handbook.
  3. CoganSF LiuY‐P McCaffreyM TwardochUM(1993). Thin film hermetic coatings. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 5 N44‐NS‐2‐2311.
  4. De Luca CJ, 2002, Surface Electromyography: Detection and Recording.
  5. EdellDJ DevaneyL DurfeeWK GleasonEF GleasonKK HerndonTO ThompsonCV WrightonMS(1993). Bioresistant integrated circuits. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 9 NO1‐NS‐0‐2399.
  6. Hoffer JA, 2001, Neural Prostheses for Restoration of Sensory and Motor Function., 139
  7. Llinás R, 1973, Brain Unit Activity During Behaviour., 105
  8. Naples GG, 1990, Neural Prostheses: Fundamental Studies, Biophysics and Bioengineering Series., 107
  9. Nishikawa D, 1999, IEEE IROS'99, Kyongju, 1592
  10. PineJ TaiY‐C BuzsakiG BraginA CarpiD(1994). The cultured neuron probe. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 4 NO1‐NS‐3‐2393.
  11. Popovic DB, 1992, Neural Prostheses Replacing Motor Function After Disease or Disability., 233
  12. Roblee LS, 1990, Neural Prostheses: Fundamental Studies, Biophysics and Bioengineering Series., 26
  13. Santa‐Cruz MC, 2000, IFESS 2000, 436
  14. SinghJ PeckRA LoebGE(2001). Development of BIONs technology for functional electrical stimulation: hermetic packaging. Proceedings of the 23th Annual International Conference of IEEE Eng Med and Bio Soc.
  15. Stein RB, 1992, Neural Prostheses. Replacing Motor Function After Diseases or Disability
  16. StieglitzT HabererW KammerS SchanzeT(2003). A simple model to investigate the stability of flexible micromachined retina stimulators. Proceedings of the 8th Annual International Conference of the International Functional Electrical Stimulation Society pp69–73.
  17. Tatic‐LucicS TaiY‐C WrightJA PineJ DenisonT(1993). Silicon‐micromachined neurochips for in vitro studies of cultured neural networks. Proceedings of the International Conference on Solid State Sensors and Actuators pp943–946.
  18. Tyler DJ, 1994, Interfascicular electrical stimulation for selectively activating axons, IEEE–EMBS Mag, 13, 575

Cited by

Sorted by citations

Authors
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P.
Affiliations
  • Universitat Autònoma de Barcelona (ES)
  • University of Freiburg (DE)
  • Scuola Superiore Sant'Anna (IT)
Corresponding author
Navarro, Xavier
Venue
Journal of the Peripheral Nervous System
Publisher
Wiley
Published
2005-09
Volume
10
Issue
3
Pages
229-258
ISSN
  • 1085-9489 (ISSN-L)
  • 1529-8027
OpenAlex
W2108068079
Cited by
845
References
259
Citations by year
2026: 18 · 2025: 27 · 2024: 36 · 2023: 32 · 2022: 39 · 2021: 53 · 2020: 49 · 2019: 58 · 2018: 53 · 2017: 49 · 2016: 60 · 2015: 57 · 2014: 58 · 2013: 45 · 2012: 28
FWCI
2.19
Type
journal article
Language
en
Open access
Locations
  • Journal of the Peripheral Nervous System (publishedVersion)
  • PubMed (publishedVersion)
  • Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion)
  • Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
  • CINECA IRIS Institutional Research Information System (Sant'Anna School of Advanced Studies) (submittedVersion)
Funders
European Commission (IST-001917)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
  • Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
Keywords
neural prostheses, peripheral nerve interfaces, peripheral nerve stimulation, neuromuscular stimulation, electromyography, surface electrodes, cuff electrodes, intrafascicular electrodes, electrode design, tissue damage, robotic prostheses
MeSH terms
Animals, Bionics, Electrodes, Electromyography, Humans, User-Computer Interface, Peripheral Nervous System, Prostheses and Implants

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-01.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-01.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review