A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems
Open accessAbstract
Considerable scientific and technological efforts have been devoted to develop neuroprostheses and hybrid bionic systems that link the human nervous system with electronic or robotic prostheses, with the main aim of restoring motor and sensory functions in disabled patients.
AI snapshot
Written from the paper’s full text.
TL;DR
This review surveys peripheral nervous system interfaces for neuroprostheses and hybrid bionic systems, comparing non-invasive, muscle, extraneural, intraneural, and regenerative electrodes. It concludes that no single interface is universally optimal; selection depends on required selectivity, invasiveness, clinical maturity, and application.
Key findings
- Electrode invasiveness generally tracks selectivity: surface and muscle electrodes are safer and easier but act only on underlying or implanted muscles; cuff and epineurial electrodes interface many axons; intrafascicular, penetrating, and sieve/regenerative electrodes access smaller axon groups but carry greater risk of nerve damage and are mostly research-stage.
- Clinical PNS neuroprostheses are dominated by stimulation: sacral anterior root stimulators for bladder control reached about 2,000 patients worldwide; phrenic pacing, vagus nerve stimulation (postoperative infection under 3%, device removal 1%), cochlear implants, and pain/continence devices are clinical, while recording interfaces for prosthesis control remain largely research.
- Penetrating arrays: Utah MEAs with 25 or 100 needles, including slanted lengths from 0.5 to 1.5 mm, recorded single-unit mechanoreceptor responses in only 10–20% of electrodes and evoked graded muscle recruitment with 1–20 µA currents; responses were stable over 36 hours, and in cat sciatic nerve up to 7 months stimulation stabilized in 80% of electrodes but sensory recordings did not.
- LIFEs implanted in eight amputees for two days allowed voluntary motor signals linked to intended missing-limb movements and elicited graded touch, joint-movement, and position sensations; regenerative sieve electrodes require hole diameters around 40–65 µm, and no human implants have been reported. EMG-based control works for prostheses and exoskeletons but is limited in degrees of freedom, while ENG can provide event and kinematic feedback.
Methods
- Design: critical narrative review, not a primary experimental study; it covers PNS organization, electrode types, materials and fabrication, and control strategies.
- Data or sample: synthesizes prior animal and human studies and clinical experience; no systematic search, inclusion criteria, or primary dataset is described.
- Main analysis: qualitative comparison using tables of advantages/disadvantages and application/status, with discussion of engineering, biological, and clinical tradeoffs.
Limitations
- Stated: no universal best electrode exists; nerve electrodes may damage nerves; cuff, intraneural, and regenerative designs need more study; regenerative electrodes have no human implants; bioelectronic interfaces are early experimental.
- Stated: long-term stability, selectivity, and biocompatibility remain unresolved, and many recording approaches are not clinically mature.
- Evident: as a narrative review from 2005, it lacks a systematic search or meta-analysis and predates later technological and clinical developments.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
19 of 259 cited sources checked · 8 supported · 0 flagged · accuracy not yet measured
- Premisecitation§ Epineurial and helicoidal electrodes · p. 9
Helical electrodes are widely used for functional electrical stimulation of the vagus nerve for intractable epilepsy, sleep apnea, and depressive syndromes.
“Helical electrodes are widely used for FES of the vagus nerve, which has found therapeutical applications for the control of intractable epilepsy, sleep apnea, and treatment of depressive syndromes (McLachlan, 1997; Fisher and Handforth, 1999) .”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites McLachlan, 1997 source
- cites Fisher and Handforth, 1999 source
The Abstract (page unavailable) describes an electrode implanted around the left vagus nerve and concludes that vagus nerve stimulation is effective for medically intractable partial seizures. It does not identify the electrode as helical, establish that helical electrodes are widely used, or discuss treatment of sleep apnea or depressive syndromes. (LLM, conf=0.87)
Experimental — accuracy not yet measured
- Premisecitation§ Book electrodes · p. 9
An epidural implantation procedure was developed to increase the group of patients who may benefit from this neural prosthesis.
“Besides the intradural surgical implantation, an epidural implantation procedure was developed (Sauerwein et al., 1990) to enhance the group of patients who may benefit from this kind of neural prosthesis.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Sauerwein et al., 1990 source
- Premisecitation§ Book electrodes · p. 9
The sacral anterior root stimulator can also be used to restore male sexual and reproductive function.
“The SARS can be also used to restore male sexual and reproductive function (Stief et al., 1992; Brindley, 1995) .”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites Stief et al., 1992 source
The Abstract (p. 5) says the implanted stimulator could induce penile erection, and the Discussion (p. 6) reports full erections in all three men at certain stimulation frequencies. This supports restoration of an aspect of sexual function, but the paper does not establish restoration of reproductive function, such as ejaculation or fertility. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites Brindley, 1995
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Penetrating microelectrodes · p. 14
Electrode arrays provided neural contacts with low-force recruitment properties similar to those of single wires.
“Although involving a more invasive insertion procedure, electrode arrays provided neural contacts with low-force recruitment properties similar to those of single wires.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Smit et al., 1999 source
- Premisecitation§ Penetrating microelectrodes · p. 14
No strict inverse recruitment was observed, especially for intrafascicular electrodes.
“Especially for intrafascicular electrodes, no strict inverse recruitment was observed (Veltink et al., 1989) .”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Veltink et al., 1989 source
The supplied passages contain bibliographic metadata, not findings about recruitment. Searches for inverse recruitment with intrafascicular electrodes and for recruitment-order reversal returned no relevant passage, so our copy does not establish whether the paper observed strict inverse recruitment. (LLM, conf=0.72)
Experimental — accuracy not yet measured
- Premisecitation§ Penetrating microelectrodes · p. 14
The arrays were capable of selectively evoking threshold forces with high efficiency.
“The arrays were capable of evoking threshold forces selectively with high efficiency.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Smit et al., 1999 source
- Premisecitation§ Penetrating microelectrodes · p. 14
Array results revealed partial blocking of neural conduction, similar to that reported with microneurographic insertion with single needles.
“Array results revealed partial blocking of neural conduction, similar to that reported with microneurographic insertion with single needles.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Smit et al., 1999 source
- Premisecitation§ Penetrating microelectrodes · p. 14
Motor recruitment was more stable with stimulation by intrafascicular multielectrodes than by extraneural electrodes.
“Motor recruitment was found more stable with stimulation by intrafascicular multielectrodes than by extraneural electrodes.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Veltink et al., 1989 source
The available copy contains a title and bibliographic records (the passage labeled Abstract, no page given), but no study findings. Searches for stable motor recruitment and for a comparison of intrafascicular with extraneural stimulation found no relevant passage, so the claimed difference cannot be verified from this copy. (LLM, conf=0.73)
Experimental — accuracy not yet measured
- Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Extraction of information from natural sensors in the body by processing afferent nerve signals is very challenging.
“the extraction of information from natural sensors in the body by processing afferent nerve signals is very challenging (Johnson et al., 1995)”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Johnson et al., 1995 source
- Premisecitation§ EMG-based control of hand prostheses · p. 20
Features extracted in time and frequency domains increase the possibility of controlling more than two degrees of freedom when combined with advanced classification algorithms and on-line learning.
“Several features can be extracted in both time and frequency domains (Zardoshti-Kermani et al., 1995; Englehart, 1998; Pattichis and Pattichis, 1999; Sparto et al., 2000) that increase the possibility of controlling more than two degrees of freedom of the prosthesis when used in combination with advanced classification algorithms (Kelly et al., 1990; Christodoulou and Pattichis, 1999; Micera et al”
Quote in the paperNot a substantial claim1 of 3 sources checked - Findingreasoning§ Interface Electrodes · p. 5
The use of many small electrodes does not always grant the expected selectivity within the praxis.
“However, the use of many small electrodes does not always grant the expected selectivity within the praxis.”
Quote in the paperSubstantial claim - Premisecitation§ Electrical · p. 2
Selective stimulation of different fascicles of axons composing a nerve may be achieved.
“Selective stimulation of different fascicles of axons composing the nerve may also be achieved.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2001 source
- Premisecitation§ Electrical · p. 2
FES has clinical applications for reducing pain.
“for reducing pain”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Electrical · p. 2
FES has clinical applications for phrenic nerve pacing for ventilatory assistance.
“for phrenic nerve pacing for ventilatory assistance”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Electrical · p. 2
FES has clinical applications for activation of lower extremity motion.
“for activation of lower extremity motion”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Electrical · p. 2
FES has clinical applications for control of hand movements.
“for control of hand movements”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Surface electrodes · p. 5
The only non-invasive electrodes are surface electrodes applied to the skin for recording ECG, EMG, or EEG.
“The only non-invasive electrodes are surface electrodes applied to the skin of the subject for the recording of ECG, EMG, or EEG”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Birbaumer et al., 2004 source
- Premisecitation§ Surface electrodes · p. 5
Felt-pad metal electrodes and carbon-rubber electrodes are adequate for transcutaneous motor stimulation considering their electrical impedance, ease of application, durability, and lack of skin reactivity.
“Felt-pad metal electrodes and carbon-rubber electrodes are adequate for transcutaneous motor stimulation considering their electrical impedance and their ease of application, durability, and lack of skin reactivity”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Nelson et al., 1980
- Premisecitation§ Non-electrical interfaces · p. 7
Magnetoneurography is a non-invasive method that allows tracing and visualizing in three dimensions the propagation of compound action currents along peripheral nerves.
“Magnetoneurography is a non-invasive method that allows tracing and visualizing in three dimensions the propagation of compound action currents along peripheral nerves.”
Quote in the paperNot a substantial claim - Premisecitation§ Epimysial and intramuscular electrodes · p. 8
Electrical stimulation delivered through intramuscular or epimysial electrodes has been applied for maintenance of denervated muscles after plexus or peripheral nerve injuries, in order to avoid disuse atrophy.
“Electrical stimulation delivered through intramuscular or epimysial electrodes has been applied for maintenance of denervated muscles after plexus or peripheral nerve injuries, in order to avoid disuse atrophy”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Nicolaidis and Williams, 2001 source
- Premisenone given§ Epimysial and intramuscular electrodes · p. 7
Intramuscular electrodes can be used for both stimulation and recording.
“Intramuscular electrodes can be used for both stimulation and recording.”
Quote in the paperNot a substantial claim - Premisenone given§ Epimysial and intramuscular electrodes · p. 8
Although epimysial and intramuscular electrodes do not directly contact peripheral nerves, they achieve muscle activation by exciting the nerve fibers arborized within the muscle in which they are implanted.
“Although epimysial and intramuscular electrodes do not directly contact peripheral nerves, they achieve muscle activation by exciting the nerve fibers arborized within the muscle in which they are implanted.”
Quote in the paperNot a substantial claim - Premisecitation§ Interface Electrodes · p. 5
Time variations in the current required to generate a particular level of neuromuscular activation are attributable to changes in induced fields from tissue encapsulation or inflammation, changes in electrode position relative to the nerve, and changes in physiological properties of the neuromuscular system including degeneration and regeneration of stimulated nerve fibers.
“Time variations in the current required to generate a particular level of neuromuscular activation are attributable to changes in the induced fields resulting from tissue encapsulation or inflammation, changes in the electrode position relative to the nerve, and changes in the physiological properties of the neuromuscular system, including degeneration and regeneration of stimulated nerve fibers”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Grill and Mortimer, 1998 source
- Interpretationreasoning§ Interface Electrodes · p. 5
The desired selectivity of stimulation or recording from individual nerve fibers or motor units increases in parallel with the invasivity of the electrode implantation.
“Considering the application of different types of electrodes, the desired selectivity of stimulation or recording from individual nerve fibers or motor units increases in parallel with the invasivity of the electrode implantation”
Quote in the paperSubstantial claim - Premisecitation§ Surface electrodes · p. 5
Surface electrodes are used for standing and short-distance walking assistance.
“for standing and short-distance walking assistance”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Graupe and Kohn, 1997 source
- Premisecitation§ Non-electrical interfaces · p. 7
Extremely weak magnetic fields generated by the ion flows of evoked compound nerve action potentials can be detected using SQUID sensors with high spatial and temporal resolution.
“Extremely weak magnetic fields generated by the ion flows of evoked compound nerve action potentials can be detected using superconducting quantum interference device (SQUID) sensors with high spatial and temporal resolution”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hoshiyama et al., 1999 source
- Premisecitation§ Interface Electrodes · p. 5
The theoretical basis of electrode surface and usable characteristics and testable models of electrode design have been optimized.
“The theoretical basis of electrode surface and usable characteristics and testable models of electrode design have been optimized”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Surface electrodes · p. 5
Surface electrodes for stimulation are extensively used in rehabilitation to activate skeletal muscles or reduce chronic pain by activating large afferent fibers in peripheral nerves supplying the affected cutaneous region.
“surface electrodes for stimulation are extensively used in rehabilitation as components of simple stimulating devices to activate skeletal muscles or to reduce chronic pain (TENS units) by activating large afferent fibers in peripheral nerves supplying the affected cutaneous region”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stanton-Hicks and Salamon, 1997 source
- Interpretationnone given§ Epimysial and intramuscular electrodes · p. 8
For muscular electrical stimulation, the control of relatively complex movements may be achieved using a large set of electrodes, each one activating a different muscle.
“For muscular electrical stimulation, the control of relatively complex movements may be achieved using a large set of electrodes, each one activating a different muscle.”
Quote in the paperNot a substantial claim - Premisecitation§ Epimysial and intramuscular electrodes · p. 8
Another problem derives from the lack of selective, size-programmed activation of motor units within each muscle, thus limiting the period of use by muscle fatigue.
“Another problem derives from the lack of selective, size-programmed activation of motor units within each muscle, thus limiting the period of use by muscle fatigue”
Quote in the paperNot a substantial claim0 of 2 sources checked - Premisecitation§ Interface Electrodes · p. 5
Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage.
“Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Non-electrical interfaces · p. 7
The main clinical perspective of magnetoneurography lies in the diagnosis of proximal mononeuropathies (plexus and root lesions), where conventional electrophysiological tests sometimes fail.
“The main clinical perspective of magnetoneurography lies in the diagnosis of proximal mononeuropathies (plexus and root lesions), where conventional electrophysiological tests sometimes fail”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Mackert, 2004 source
- Premisecitation§ Interface Electrodes · p. 5
Good candidates for ohmic electrode-tissue contacts are gold, platinum, platinum-iridium, tungsten, and tantalum.
“Good candidates for ohmic electrode-tissue contacts are gold, platinum (Pt), platinum-iridium, tungsten, and tantalum”
Quote in the paperNot a substantial claim0 of 2 sources checked - Interpretationreasoning§ Epimysial and intramuscular electrodes · p. 8
In denervated muscles, voltage pulses should be higher and of longer duration than for activation of innervated muscles so as to directly excite the muscle fiber membrane.
“In this situation, voltage pulses should be higher and of longer duration than for activation of innervated muscles so as to directly excite the muscle fiber membrane.”
Quote in the paperNot a substantial claim - Premisecitation§ Surface electrodes · p. 5
Surface electrodes are used for control of postural hypotension.
“for control of postural hypotension”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Taylor et al., 2002 source
- Premisecitation§ Epimysial and intramuscular electrodes · p. 7
Epimysial electrodes have proven their suitability in current complex motor prostheses that require coordinated stimulation of several muscles, for example, grasping system for tetraplegics (Freehand System) and standing and walking systems for paraplegics.
“Epimysial electrodes have proven their suitability in current complex motor prostheses that require coordinated stimulation of several muscles, for example, grasping system for tetraplegics (Freehand System) (Bhadra et al., 2001) and standing and walking systems for paraplegics”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Non-electrical interfaces · p. 7
Myoacoustical signals obtained via a single microphone have been proved feasible for the control of a prosthetic hand.
“Myoacoustical signals obtained via a single microphone have been proved feasible for the control of a prosthetic hand.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Barry et al., 1986
- Premisecitation§ Surface electrodes · p. 5
Surface electrodes are used in FES systems for correction of foot drop in hemiplegic patients.
“They are also used in more sophisticated FES systems for correction of foot drop in hemiplegic patients”
Quote in the paperNot a substantial claim0 of 2 sources checked - Premisenone given§ Non-electrical interfaces · p. 7
Contracting skeletal muscle produces vibrations and emits sounds that are easily recorded with a standard microphone or an accelerometer.
“Contracting skeletal muscle produces vibrations and emits sounds that are easily recorded with a standard microphone or an accelerometer.”
Quote in the paperSubstantial claim - Premisecitation§ Cuff electrodes · p. 10
The reduced size and thickness of recent polymer cuffs opens the possibility of implanting several small cuffs around different fascicles or branches of nerves and achieving selective functional stimulation of a higher number of muscles.
“The reduced size and thickness of recent polymer cuffs (Stieglitz et al., 2000) also opens the possibility of implantation of several small cuffs around different fascicles or branches of nerves and, consequently, the achievement of selective functional stimulation of a higher number of muscles (Rodrı ´guez et al., 2000; Stieglitz et al., 2003) .”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites Stieglitz et al., 2000 source
- cites Rodrı ´guez et al., 2000
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Stieglitz et al., 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Cuff electrodes · p. 11
Long and snug-fitting cuffs generally provide larger recorded nerve signals.
“In general, long (interelectrode distance close to the wavelength of the action potential) and snug-fitting cuffs provide larger signals (Loeb et al.,”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Loeb et al., 1977 source
- cites Hoffer and Kallesøe, 2001
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Cuff electrodes · p. 11
Chronically implanted tripolar cuff electrodes in animals can be used for long-term recording of both afferent and efferent nerve activity.
“Research in animals with chronically implanted tripolar cuff electrodes showed that they can be used for long-term recording of both afferent and efferent nerve activity (Stein et al., 1977; Popovic et al., 1993; Hoffer and Kallesøe, 2001) .”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Flat-interface nerve electrodes · p. 11
The flat-interface nerve electrode is an extraneural electrode designed to reshape peripheral nerves into a geometry favorable for selective stimulation.
“The FINE is an extraneural electrode designed to reshape peripheral nerves into a favorable geometry for selective stimulation (Fig. 8).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Tyler and Durand, 2002 source
- Premisecitation§ Flat-interface nerve electrodes · p. 11
Modeling studies suggest that greater nerve reshaping yields better selectivity for stimulation and recording.
“Modeling studies have suggested that the more the nerve is reshaped, the better the selectivity for stimulation and recording”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Perez-Orive and Durand, 2000
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Choi et al., 2001 source
- Premisecitation§ Epineurial and helicoidal electrodes · p. 9
The open structural shape of the helicoidal electrode explains its low selectivity.
“However, its open structural shape explains the low selectivity of the electrode (Agnew et al., 1989) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Agnew et al., 1989 source
- Premisecitation§ Cuff electrodes · p. 10
Nerves can be damaged by the presence of a cuff because of the delicacy of nerve tissue and the physical properties of the electrodes, especially in limb peripheral nerves subjected to a wide range of motion.
“In spite of the fact that cuff electrodes have been implanted in patients for decades, nerves can be damaged by the presence of the cuff due to the delicacy of the nerve tissue and the physical properties of the electrodes (Nielson et al., 1976; Waters et al., 1985; Krarup et al., 1989; Naples et al., 1990; Larsen et al., 1998) , especially in peripheral nerves of the limbs, subjected to a wide”
Quote in the paperSubstantial claim0 of 3 sources checked - Premisecitation§ Cuff electrodes · p. 10
Multichannel cuff electrodes enable selective stimulation of separate axonal fascicles within a nerve, each supplying innervation to a different muscle.
“Multichannel cuff electrodes enable selective stimulation of separate axonal fascicles within the nerve, each one supplying innervation to a different muscle (Rozman et al., 1993; Veraart et al., 1993; Grill and Mortimer, 1996a; Walter et al., 1997; Navarro et al., 2001) .”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Interfascicular electrodes · p. 12
Experimental results showed that interfascicular stimulation provided additional recruitment to enhance surface stimulation.
“Experimental results showed that interfascicular stimulation provided additional recruitment to enhance surface stimulation.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Tyler and Durand, 1997 source
- Premisecitation§ Epimysial and intramuscular electrodes · p. 9
BION microstimulators have been used in human patients for electrical muscle stimulation to prevent disuse atrophy and secondary complications.
“These microstimulators have been used in human patients for muscle electrical stimulation to prevent disuse atrophy and secondary complications (Dupont-Salter et al., 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Dupont-Salter et al., 2004 source
- Premisecitation§ Book electrodes · p. 9
Book electrodes are widely used clinically for urinary bladder management in spinal cord-injured persons.
“Book electrodes are of widespread clinical use for urinary bladder management in spinal cord-injured persons.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Brindley, 1972
- Premisecitation§ Epimysial and intramuscular electrodes · p. 9
Compared with surface stimulation of the common peroneal nerve, BION stimulation gave more selective activation of specific muscles such as tibialis anterior and extensor digitorum longus.
“Compared with surface stimulation of the common peroneal nerve, stimulation with BIONs provided more selective activation of specific muscles such as tibialis anterior and extensor digitorum longus (Weber et al., 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Weber et al., 2004 source
- Premisecitation§ Cuff electrodes · p. 10
Cuff electrodes confine the stimulating current to the inner space of the electrode, avoiding stimulation of neighboring nerves and tissues.
“Cuff electrodes placed around the nerves have several advantages compared with surface, intramuscular, and epimysial electrodes: they allow for correct positioning of electrode leads to minimize mechanical distortion and lead failure, and the stimulating current is confined to the inner space of the electrode, thus avoiding the stimulation of other neighboring nerves and tissues (Loeb and Peck, 19”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Loeb and Peck, 1996 source
- Premisecitation§ Epineurial and helicoidal electrodes · p. 9
In functional electrical stimulation systems for phrenic nerve and pain control, cuff electrodes have replaced button-type epineurial electrodes to reduce the risk of traumatic nerve injury.
“Interestingly, in FES systems for phrenic nerve and for pain control, the cuff electrodes used in the past have been substituted for button-type epineurial electrodes to reduce the risk for traumatic injury to the nerve (Chervin and Guilleminault, 1997; Stanton-Hicks and Salamon, 1997) .”
Quote in the paperNot a substantial claim0 of 2 sources checked - Premisecitation§ Cuff electrodes · p. 10
Cuff electrodes allow correct positioning of electrode leads to minimize mechanical distortion and lead failure.
“Cuff electrodes placed around the nerves have several advantages compared with surface, intramuscular, and epimysial electrodes: they allow for correct positioning of electrode leads to minimize mechanical distortion and lead failure, and the stimulating current is confined to the inner space of the electrode, thus avoiding the stimulation of other neighboring nerves and tissues (Loeb and Peck, 19”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Loeb and Peck, 1996 source
- Premisecitation§ Epimysial and intramuscular electrodes · p. 9
BIONs have been reported to be useful in a functional electrical stimulation system for correcting foot drop.
“Recently, the BIONs were reported to be useful in a FES system to correct foot drop.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Weber et al., 2004 source
- Premisecitation§ Regenerative electrodes · p. 16
Regenerative electrodes are designed to interface a high number of nerve fibers using an array of holes with electrodes built around them, implanted between the severed stumps of a peripheral nerve.
“Regenerative electrodes are designed to interface a high number of nerve fibers by using an array of holes, with electrodes built around them, implanted between the severed stumps of a peripheral nerve (Llina ´ s et al., 1973; Edell, 1986; Kovacs et al., 1992; Dario et al., 1998) .”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Penetrating microelectrodes · p. 15
Intraneural Utah multielectrode arrays made of silicon with 25 and 100 individual needle electrodes were inserted into a peripheral nerve in experimental animals using a pneumatic insertion device without significantly disturbing nerve function.
“Intraneural Utah MEAs made of silicon with 25 and 100 individual needle electrodes have been inserted into a peripheral nerve in experimental animals using a pneumatic insertion device without significantly disturbing nerve function.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2001 source
- Premisecitation§ Intrafascicular electrodes · p. 12
Comparatively smaller stimulus intensities can be used with intrafascicular electrodes to achieve equivalent levels to those of extraneural electrodes.
“Comparatively smaller stimulus intensities can be used to achieve equivalent levels to those of extraneural electrodes (Yoshida et al., 2000) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Yoshida et al., 2000 source
- Premisecitation§ Intrafascicular electrodes · p. 13
Electrophysiological tests conducted for two consecutive days after surgery indicated that volitional motor nerve activity associated with missing limb movements could be recorded.
“Electrophysiological tests conducted for 2 consecutive days after the surgery indicated that it was possible to record volitional motor nerve activity associated with missing limb movements.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisecitation§ Penetrating microelectrodes · p. 15
The surgical technique highly affected the long-term results of Utah multielectrode array implants.
“The surgical technique highly affected the long-term results.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2004 source
- Premisecitation§ Penetrating microelectrodes · p. 15
A case report on a healthy volunteer with a multielectrode array implanted in the median nerve for three months supports the potential of such devices for bidirectional interfacing with the peripheral nervous system.
“A case report on a healthy volunteer who had a MEA implanted in the median nerve for 3 months gives support to the potential of such devices for bidirectional interfacing the PNS (Warwick et al., 2003) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Warwick et al., 2003 source
- Premisecitation§ Penetrating microelectrodes · p. 13
Despite microlesions caused by electrode insertion into the nerve, the morbidity of microneurography is acceptably low.
“Despite the production of microlesions caused by the electrode insertion into the nerve, the morbidity with the procedure is acceptably low (Gandevia and Hales, 1997) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Gandevia and Hales, 1997 source
- Premisecitation§ Intrafascicular electrodes · p. 13
Longitudinal intrafascicular electrodes have been implanted in severed nerves proximally to the stump in eight subjects with limb amputation.
“LIFEs have been implanted in severed nerves proximally to the stump of eight subjects with limb amputation (Dhillon et al., 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisecitation§ Penetrating microelectrodes · p. 15
Histological analysis indicated that the morphology and fiber density of the nerve around the Utah array electrodes were normal.
“A histological analysis indicated that the morphology and fiber density of the nerve around the electrodes were normal.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2004 source
- Premisecitation§ Intrafascicular electrodes · p. 13
Electrical stimulation through implanted longitudinal intrafascicular electrodes elicited graded sensations of touch, joint movement, and position referring to the missing limb.
“Electrical stimulation through the implanted electrodes elicited graded sensations of touch, joint movement, and position, referring to the missing limb.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisecitation§ Penetrating microelectrodes · p. 15
Electrodes without and with lead wires were implanted for up to seven months in cat sciatic nerves.
“Electrodes without and with lead wires were implanted for up to 7 months in cat sciatic nerves (Branner et al., 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2004 source
- Premisenone given§ Bioelectronic interfaces · p. 19
Transfer of cultured cells from the in vitro environment into the brain remained difficult.
“However, the transfer of the cultured cells from the in vitro environment into the brain remained difficult.”
Quote in the paperNot a substantial claim - Premisenone given§ Regenerative electrodes · p. 16
Multichannel regenerative electrodes can be applied only to transected nerves.
“multichannel regenerative electrodes can be applied only to transected nerves”
Quote in the paperNot a substantial claim - Premisecitation§ EMG-based control of hand prostheses · p. 20
Useful information can be found in the transient burst of myoelectric signal immediately after onset of contraction.
“useful information can be found in the transient burst of myoelectric signal (Hudgins et al., 1993) immediately after the onset of contraction”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hudgins et al., 1993 source
- Premisenone given§ EMG-based control of hand prostheses · p. 20
Parameters from a global steady-state approach (variance, mean absolute value, Fourier spectrum, median frequency) are often insufficient to distinguish more than two classes of movement.
“the parameters that can be extracted using this global steadystate approach (e.g., variance, mean absolute value, Fourier spectrum, and median frequency) are often not sufficient to distinguish between more than two classes of movement.”
Quote in the paperSubstantial claim - Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
If standard electronic processes are directly transferred for implant passivation, degradation during implantation deteriorates the insulation layer and functional integrated structures within 1 year.
“If parts of standard electronic processes were directly transferred for implant passivation, degradation during implantation deteriorated not only the insulation layer but also functional integrated structures within 1 year (Haemmerle et al., 2002)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Haemmerle et al., 2002 source
- Interpretationcitation§ Regenerative electrodes · p. 16
Limits on amplitude and discriminability of single-unit action potentials recordable from nerve fibers inside tubular electrodes may also apply to regenerative electrodes, reducing recorded signal amplitude.
“The limits on the amplitude and discriminability of single-unit action potentials recordable from nerve fibers inside tubular electrodes (Loeb et al., 1977) may also apply to regenerative electrodes, thus decreasing the actual amplitude of recorded signals.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Loeb et al., 1977 source
- Premisenone given§ EMG-based control of neuroprostheses and hybrid bionic systems · p. 20
Residual EMG signals from a paretic muscle can be used to control stimulation of the same or a different muscle in FES systems.
“residual EMG signals recorded from a paretic muscle can be used for the control of stimulation of the same muscle or of a different muscle in FES systems.”
Quote in the paperNot a substantial claim - Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
Silicon-based microdevices have the advantage of monolithic integration of electronic circuits for recording and signal processing.
“One advantage of siliconbased microdevices is the ability of monolithic integration of electronic circuits for recording and signal processing (Najafi and Wise, 1986; Ji et al., 1990)”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Najafi and Wise, 1986 source
- cites Ji et al., 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Regenerative electrodes · p. 16
Efferent neural signals recorded from regenerative electrodes can be used to control the motion of a mechanical prosthesis.
“recording of neural efferent signals can be used for the motion control of a mechanical prosthesis (Edell, 1986)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Edell, 1986 source
- Interpretationreasoning§ Materials for substrates, insulation, and encapsulation · p. 18
Silicon- and polyimide-based interfaces, because they are founded on micromachining, have an advantage over hand-made electrodes in ability to modify designs with high precision and to control active zone size and repeatability.
“Because the silicon- and polyimide-based interfaces are founded on micromachining techniques, they have an advantage over hand-made electrodes in that one has the ability to modify designs with high precision (fast prototyping) and can assert precise control over active zone size and repeatability.”
Quote in the paperNot a substantial claim - Premisecitation§ Regenerative electrodes · p. 18
Amplified motor signals can be obtained from lesioned nerves by allowing them to innervate isolated slips of host muscle, from which EMG signals can be recorded by wire electrodes.
“Amplified motor signals can be obtained from lesioned nerves by allowing them to innervate isolated slips of host muscle, from which EMG signals can be recorded by wire electrodes (Wells et al., 2001)”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Wells et al., 2001
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Bioelectronic interfaces · p. 19
In silicon micromachining wells with gold electrodes and a top grillwork, the grillwork prevented neurons from migrating.
“The grillwork prevented the neurons from migration.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Tatic-Lucic et al., 1993
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Bioelectronic interfaces · p. 19
A regenerative electrode was introduced as a bioelectronic interface between sprouting embryonic neurons and the peripheral nerve to allow neural stimulation for controlled muscle excitation to prevent atrophy and restore some function.
“A regenerative electrode was introduced as a bioelectronic interface (Stieglitz et al., 2002) between the sprouting embryonic neurons and the peripheral nerve to allow neural stimulation for controlled muscle excitation to prevent atrophy and restore some function.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stieglitz et al., 2002
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
Silicon nitride and silicon oxide are not stable as monolayers in a physiologic environment.
“materials such as silicon nitride and silicon oxide are not stable as monolayers in a physiologic environment (Connolly et al., 1992; Edell et al., 1993)”
Quote in the paperSubstantial claim0 of 2 sources checked- cites Connolly et al., 1992 source
- cites Edell et al., 1993
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning§ EMG-based control of neuroprostheses and hybrid bionic systems · p. 20
For EMG-based control, exoskeletons and tele-operated humanlike robots are very similar because their joints can be controlled by homologous muscles.
“for the implementation of an EMG-based control algorithm, the second and third types of application are very similar if the remotely controlled robotic device presents a humanlike structure (typically the upper extremity). In fact, in both these cases, the different joints of the prosthesis can be controlled by the 'homologous' muscles”
Quote in the paperNot a substantial claim - Premisereasoning§ EMG-based control of hand prostheses · p. 20
Little is known about the reliability and accuracy of proposed onset detection methods.
“Several methods and algorithms have been proposed in the literature, but little is known about their reliability and accuracy.”
Quote in the paperNot a substantial claim - Premisecitation§ Bioelectronic interfaces · p. 19
An animal model for a biohybrid approach to cure flaccid paralysis involves transplanting embryonic spinal cord cells into a container adapted to the distal nerve stump to restore skeletal muscle function.
“researchers have developed an animal model for a biohybrid approach to cure flaccid paralysis that involves embryonic spinal cord cell transplantation into a container adapted to the distal nerve stump to restore skeletal muscle function (Katsuki et al., 1997; Thomas et al., 2000; Klinge et al., 2001b)”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ EMG-based control of hand prostheses · p. 20
The generalized likelihood ratio method seems more robust than other methods, especially when muscle activity is weak.
“The generalized likelihood ratio (GLR) method seems to be more robust than other methods especially when muscle activity is weak (Micera et al., 2001b)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Micera et al., 2001b
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning§ EMG-based control of neuroprostheses and hybrid bionic systems · p. 20
Control of prostheses or tele-operation of not-human-like robots is complicated by the need to code the different actions of the robot.
“the control of prostheses (or the tele-operation of not-human-like robots) is complicated by the need for coding the different actions of the robot.”
Quote in the paperNot a substantial claim - Interpretationreasoning§ EMG-based control of hand prostheses · p. 20
Because of the pattern recognition procedure, correct onset of movement must be identified to extract significant features from sEMG signals.
“Because of this pattern recognition procedure, it is important to identify the correct onset of the movement in order to extract significant features from the sEMG signals.”
Quote in the paperNot a substantial claim - Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
Polyimide-based neural implants show good biocompatibility with respect to toxicity and biostability when used as cuff and regenerative-type electrodes.
“Polyimide-based neural implants (Stieglitz et al., 2000) showed good biocompatibility (Haggerty and Lusted, 1989; Richardson et al., 1993) with respect to toxicity as well as biostability (Navarro et al., 1998; Rodrı ´guez et al., 2000) when used as cuff and regenerative-type electrodes.”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ Choice of fabrication technology · p. 19
First chronic results from pilot experiments showed that combination layers of parylene C insulation and silicone elastomer encapsulation were stable over more than 1 year in vivo.
“Even though first chronic results from pilot experiments showed that these combination layers were stable over more than 1 year in vivo (Stieglitz et al., 2003)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stieglitz et al., 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
Multilayer coatings achieve better stability than monolayers.
“Better stability was achieved with multilayers (Cogan et al., 1993; Vogt and Hauptmann, 1995)”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Cogan et al., 1993
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Vogt and Hauptmann, 1995 source
- Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18
For silicon to obtain flexibility, its dimensions must be reduced to a degree that the final product is mechanically fragile.
“For silicon to obtain any degree of flexibility, the dimensions must be reduced to such a degree that the final product is mechanically fragile.”
Quote in the paperNot a substantial claim - Premisecitation§ Regenerative electrodes · p. 18
In a patient with upper arm amputation, residual brachial plexus nerves anastomosed to pectoralis muscles formed few motor units that allowed simultaneous control of two degrees of freedom with a myoelectric prosthesis.
“The few reformed motor units allowed for the simultaneous control of two degrees of freedom with a myoelectric prosthesis (Kuiken et al., 2004)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kuiken et al., 2004 source
- Premisereasoning§ EMG-based control of hand prostheses · p. 20
Due to the stochastic characteristics of sEMG, onset detection is challenging, especially when the response is weak.
“Due to the stochastic characteristics of the sEMG, the onset detection is a challenging task, especially when the response is weak.”
Quote in the paperNot a substantial claim - Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Subjects in that study were able to control a cursor on a PC screen.
“The subjects involved in the study were able to control a cursor on a PC screen”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisecitation§ EMG-based control of exoskeletons · p. 21
A neuro-fuzzy network was used to predict torque information from sEMG of upper extremity muscles and kinematic trajectories to control a shoulder exoskeleton.
“a neuro-fuzzy network was used to predict torque information from the sEMG of upper extremity muscles and from the kinematic trajectories of the different joints to control a shoulder exoskeleton (Kiguchi et al., 2001; 2003)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kiguchi et al., 2001; 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
The non-linear and time-variant characteristics of the neuromuscular system require closed-loop control algorithms for effective and usable systems that need feedback about the executing motor task.
“the non-linear and time-variant characteristics of the neuromuscular system ask for the use of closed-loop control algorithms to develop effective and usable systems that require feedback information about the executing motor task.”
Quote in the paperNot a substantial claim- Rests on an assumption: The neuromuscular system has non-linear and time-variant characteristics.
- Interpretationreasoning§ EMG-based control of hand prostheses · p. 21
Implementing PNS interfaces to record and stimulate peripheral nerves selectively is a possible solution to overcome the limits of the EMG-based approach.
“A possible solution to overcome the limits of the EMG-based approach is to implement interfaces between the PNS and the artificial device to record and stimulate peripheral nerves in a selective manner.”
Quote in the paperSubstantial claim - Premisecitation§ EMG-based control of exoskeletons · p. 21
Biomechanical studies show that under isometric contractions the relationship between force and EMG can be considered linear.
“This approach is supported by biomechanical studies (Hof, 1997; Onishi et al., 2000) showing that under isometric contractions the relationship between force and EMG can be considered linear.”
Quote in the paperSubstantial claim0 of 2 sources checked - Premisecitation§ EMG-based control of hand prostheses · p. 21
Recent research tried to implement a continuous classifier able to quickly switch from one hand task state to another.
“recent research tried to implement a continuous classifier able to quickly switch from one state (i.e., hand task) to another (Englehart and Hudgins, 2003; Carrozza et al., 2004)”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Englehart and Hudgins, 2003 source
- cites Carrozza et al., 2004
- Premisereasoning§ EMG-based control of exoskeletons · p. 21
sEMG signals have been used to control upper and lower extremity exoskeletons by determining the force/torque to be produced by actuators moving the joints.
“The sEMG signals have been also used in the recent past for the control of upper and lower extremity exoskeletons. In this application, sEMG signals are used to determine the force/torque to be produced by the actuators moving the different joints.”
Quote in the paperNot a substantial claim - Premisenone given§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Signals recorded using neural interfaces have been used to extract useful sensory information to improve the performance of FES systems.
“The signals recorded using different neural interfaces have been used in the past to extract useful sensory information to improve the performance of FES systems.”
Quote in the paperNot a substantial claim - Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
LIFE electrodes have been used to control a hand prosthesis using ENG signals recorded from peripheral nerves of amputee subjects.
“Recently, LIFE electrodes have been used to control a hand prosthesis using the ENG signals recorded from peripheral nerves of amputee subjects (Dhillon et al., 2004)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisenone given§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
In most instances, ENG signals are used to detect an event such as slippage during grasping or heel contact.
“In most instances, ENG signals are used to detect an event such as slippage during grasping or heel contact.”
Quote in the paperNot a substantial claim - Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
ENG signals recorded from muscle spindles have been used in animal models to extract kinematic trajectories.
“Similarly, ENG signals recorded from muscle spindles have been used in animal models to extract kinematic trajectories.”
Quote in the paperNot a substantial claim - Premisecitation§ EMG-based control of exoskeletons · p. 21
Rosen et al. developed a myoprocessor based on the Hill model of muscular activation to extract force/torque information from muscle information.
“Rosen et al. (2001) developed a 'myoprocessor' based on the Hill model of muscular activation to extract force/torque information from muscle information.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rosen et al., 2001 source
- Premisecitation§ EMG-based control of hand prostheses · p. 20
Transient EMG signals have greater classification capacity than steady-state signals.
“transient EMG signals have a greater classification capacity than steady-state signals.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hudgins et al., 1993 source
- Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
For hand grasp neural prosthesis, a cuff around the palmar digital nerve recorded mechanoreceptor activity when a grasped object was slipping to regulate grasping force.
“for hand grasp neural prosthesis, a cuff placed around the palmar digital nerve recorded the activity from mechanoreceptors when a grasped object was slipping to regulate the force of grasping (Haugland and Hoffer, 1994; Haugland et al., 1994; Inmann and Haugland, 2004)”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites Haugland and Hoffer, 1994
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Haugland et al., 1994
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Inmann and Haugland, 2004 source
- Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
It was possible to elicit adequate sensory feedback to users by stimulating afferent nerves through LIFEs.
“In addition, it was possible to elicit adequate sensory feedback to the users by stimulating the afferent nerves through the LIFEs.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Dhillon et al., 2004 source
- Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
For heel contact with the floor detected from a cuff electrode on the sural nerve, advanced statistical techniques are used to improve detection performance.
“In other cases, such as heel contact with the floor detected from a cuff electrode on the sural nerve, more advanced statistical techniques are used to improve the detection performance (Upshaw and Sinkjaer, 1998)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Upshaw and Sinkjaer, 1998 source
- Interpretationcitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Neural signals could solve some problems of EMG-based control of hand prostheses for amputees.
“neural signals could solve some of the problems of the EMG-based control of hand prostheses for amputees.”
Quote in the paperSubstantial claim - Premisecitation§ EMG-based control of exoskeletons · p. 21
Use of EMG for exoskeleton control under dynamic conditions presents problems because of muscle dynamics and possible prostheses.
“the use of EMG for this purpose under dynamic conditions (as in the case of exoskeleton control) presents several problems because of muscle dynamics and possible prostheses (Rainoldi et al., 2000)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rainoldi et al., 2000 source
- Interpretationcitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
ENG signals can be used to achieve more than simple on-off control.
“indicating that ENG signals can be used to achieve more than a simple on-off control”
Quote in the paperSubstantial claim - Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Recorded ENG using cuff electrodes allowed prediction of ankle movements in an animal model.
“Recorded ENG signals using cuff electrodes allowed the prediction of ankle movements in an animal model (Micera et al., 2001a; Jensen et al., 2002; Cavallaro et al., 2003)”
Quote in the paperNot a substantial claim0 of 3 sources checked - Premisecitation§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22
Recently developed algorithms using the neuro-fuzzy approach allow extraction of useful kinematic information mimicking actual trajectories.
“Recently developed algorithms using this approach allow the extraction of useful kinematic information mimicking the actual trajectories (Cavallaro et al., 2003)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Cavallaro et al., 2003 source
- May combine two claims. 94
- Classified as probably not a substantial claim. 46
- The quote only partly states this claim. 1
May combine two claims.
Premisecitation§ Electrical · p. 2The most frequent use of PNS interfaces is in FES, with clinical applications for controlling micturition and defecation by stimulating the sacral roots.
“Nowadays, the most frequent use of PNS interfaces resides in FES, which has clinical applications in various systems designed to control micturition and defecation by stimulating the sacral roots”
Quote in the paperNot a substantial claim1 of 3 sources checked- cites Brindley et al., 1986 source
The Discussion reports that most patients used sacral anterior root stimulators mainly for micturition. Methods and Results also reports that a few used S3 or S4 stimulation to aid defaecation. The paper does not establish that FES is the most frequent use of PNS interfaces overall. (LLM, conf=0.87)
Experimental — accuracy not yet measured - cites Creasey, 1993 source
- cites Van Kerrebroeck et al., 1993 source
May combine two claims.
Premisecitation§ Non-electrical interfaces · p. 7Acoustic signals increase in parallel with surface EMG signals during voluntary and evoked contraction, but acoustic amplitude decays with fatigue whereas EMG amplitude does not.
“Acoustic signals increase in parallel with surface EMG signals during voluntary and evoked contraction, but acoustic amplitude decays with fatigue whereas EMG amplitude does not”
Quote in the paperNot a substantial claim2 of 2 sources checked- cites Barry et al., 1985 source
The Abstract reports that acoustic amplitude decayed with fatigue while surface EMG amplitude did not, and the Figure 5 discussion describes both signals increasing together when voluntary effort increased. The copy does not establish that they increase in parallel during evoked contractions; the Discussion describes external nerve stimulation as a proposed future test. It also notes that surface EMG began to decay below 45% of maximal voluntary contraction, so the claim that EMG amplitude does not decay needs qualification. Page numbers are unavailable in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites Barry et al., 1992 source
The Results (page unavailable) report that vibration amplitude declines with fatigue, but also state that the evoked compound muscle action potential (CMAP) declines, though less rapidly. The discussion (page unavailable) says surface EMG amplitude typically does not decline during moderate voluntary contractions. The paper does not support the broader claim that acoustic and EMG signals increase in parallel during both types of contraction, or that electrical-signal amplitude does not decline with evoked fatigue. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Book electrodes · p. 9After more than 10 years of development, Brindley's group reported the first 50 cases of a radio frequency-controlled sacral anterior root stimulator implant in humans.
“After more than 10 years of development, Brindley's group reported the first 50 cases of a radio frequency-controlled implant from series production with medical approval called sacral anterior root stimulator (SARS) in humans (Brindley et al., 1986) .”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Brindley et al., 1986 source
The Abstract reports the first 50 human recipients of sacral anterior root stimulators; Methods and Results describes external radio transmitters, and the References include development work dating to 1972. But the printed sentence goes further: Methods and Results says the first 11 implants were made in the Neurological Prostheses Unit, not series-produced, and our copy does not establish medical approval. (LLM, conf=0.87)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ Book electrodes · p. 9The Finetech-Brindley stimulator was commercialized, and less than a decade later 500 patients had benefited from the system.
“The implant was commercialized as FinetechBrindley stimulator, and less than a decade later 500 patients benefited from the system (Brindley, 1994) .”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Brindley, 1994 source
The Introduction says the 500th London-design stimulator was implanted in February 1992, and Sources of information identifies Finetech Medical Ltd as the manufacturer. The paper does not date commercialization, so it cannot establish the claimed interval of less than a decade. Its benefits discussion describes potential benefits but does not establish that all 500 patients benefited. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Penetrating microelectrodes · p. 14In acute experiments, multiple-wire-microelectrode arrays were inserted into rat nerves to investigate selective stimulation of motor units.
“In acute experiments, multiple-wire-microelectrode arrays were inserted into rat nerves to investigate selective stimulation of motor units (Smit et al., 1999) .”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Smit et al., 1999 source
May combine two claims.
Premisecitation§ Intrafascicular electrodes · p. 13Histology of animal nerves with metal or polymer longitudinal intrafascicular electrodes implanted for six months showed the implants were biocompatible and caused no damage, although a slight increase in connective tissue around the implant site was evident long-term.
“The histology of animal nerves that had metal or polymer LIFEs implanted for a period of 6 months revealed that the implants are biocompatible and that no damage was caused by the presence of the electrode, although a slight increase in connective tissue surrounding the implant site was evident in the longterm (Lefurge et al., 1991; Lawrence et al., 2002) .”
Quote in the paperNot a substantial claim1 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Bioelectronic interfaces · p. 19Activity of embryonic mouse spinal cord cells has been recorded simultaneously over several months on a planar glass substrate with 64 iridium tin oxide electrodes.
“The activity of embryonic spinal cord cells of mice has been recorded simultaneously over a period of several months on a planar glass substrate with 64 iridium tin oxide electrodes (Droge et al., 1986)”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Droge et al., 1986 source
The Cell culture section describes recordings from embryonic mouse spinal-cord cells, and the paper describes simultaneous multisite recording. But its recording matrix had 36 gold-conductor electrodes, not 64 iridium tin oxide electrodes; the Results describe cultures examined 3–5 weeks after seeding, not recordings over several months. No page numbers are available in our copy. (LLM, conf=0.90)
Experimental — accuracy not yet measured
May combine two claims.
Premisecitation§ EMG-based control of exoskeletons · p. 21Different complex algorithms have been developed to extract force/torque information from sEMG signals to address these limitations, although in some cases good results have been achieved using a simple proportional control law.
“Different complex algorithms have been developed to extract force/torque information from sEMG signals that address these limitations, although in some cases good results have been achieved using a simple proportional control law (Gordon and Ferris, 2004)”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Gordon and Ferris, 2004 source
The Virtual object interface section describes a simple proportional law: processed surface EMG amplitudes generated virtual forces, and the Discussion reports that subjects improved their control with practice. The Discussion also mentions signal-processing algorithms for myoelectric prostheses, but our copy does not say those algorithms extract force or torque from sEMG; the study's proportional forces were virtual, not measured force or torque estimates. Page numbers are unavailable in our copy. (LLM, conf=0.85)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingcitation§ Non-electrical interfaces · p. 7Patients learned to open and close the hand reliably after a few minutes of practice.
“The patients learned to open and close the hand reliably after a few minutes of practice”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Barry et al., 1986
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone given§ Interface Electrodes · p. 4Attempts to develop material surfaces that may be recognized as biological and form truly biocompatible interfaces with tissue are under investigation, although they have yet to be integrated in functional neural interfaces.
“Attempts to develop material surfaces that may be recognized as biological and form truly biocompatible interfaces with the tissue are under investigation, although they have yet to be integrated in functional neural interfaces.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ Acritical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems · p. 1Extraneural electrodes such as cuff and epineurial electrodes provide simultaneous interface with many axons in the nerve.
“Extraneural electrodes, such as cuff and epineurial electrodes, provide simultaneous interface with many axons in the nerve”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone given§ Electrical · p. 2Nerve electrodes have the potential to damage the nerves on which they are implanted, whereas muscle electrodes are generally considered safer.
“However, nerve electrodes also have the potential to damage the nerves on which they are implanted, whereas muscle electrodes are generally considered as safer”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone given§ Organization of the PNS · p. 3Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.
“Signals are transmitted by the corresponding axons in series of impulses or action potentials, with intensity of the signal mainly coded in impulse frequency along the peripheral axon.”
Quote in the paperNot a substantial claim- Rests on an assumption: Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.
May combine two claims.
Interpretationnone given§ Electrical · p. 2Patients are able to use implantable neuroprostheses for activities of daily living and to enhance their quality of life and independence.
“Patients are able to use the system for activities of daily living and enhancing their quality of life and independency.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Organization of the PNS · p. 3Nerve fibers are grouped in fascicles surrounded by connective tissue in the peripheral nerve.
“Nerve fibers, both afferent and efferent, are grouped in fascicles surrounded by connective tissue in the peripheral nerve”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Peters et al., 1991
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ Electrical · p. 2Most patients with nervous system injury suffer from an incomplete lesion, so modular, controllable systems for supporting functions are being developed.
“New technological developments are taking into account the fact that the most patients with nervous system injury suffer from an incomplete lesion, and thus modular, controllable systems for supporting these functions are being developed”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rupp and Gerner, 2004 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone given§ Organization of the PNS · p. 3Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.
“Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.”
Quote in the paperNot a substantial claim- Rests on an assumption: Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisenone given§ Organization of the PNS · p. 3The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.
“The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal with respect to proximal segments.”
Quote in the paperNot a substantial claim- Rests on an assumption: The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.
May combine two claims.
Premisecitation§ Electrical · p. 2Nerve electrodes implanted adjacent to, around, or within a peripheral nerve trunk or spinal root reduce the stimulus intensity required for activation compared with muscular or surface electrodes, thereby decreasing hazardous electrochemical processes and power consumption.
“Due to the proximity to the nerve, the stimulus intensity required for activation is reduced in comparison with muscular or surface electrodes and, consequently, hazardous electrochemical processes and power consumption of the stimulator system can be decreased”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Loeb and Peck, 1996 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone given§ Organization of the PNS · p. 3The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.
“The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.”
Quote in the paperNot a substantial claim- Rests on an assumption: The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisenone given§ Acritical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems · p. 1Intrafascicular, penetrating, and regenerative electrodes may contact small groups of axons within a nerve fascicle.
“intrafascicular, penetrating, and regenerative electrodes may contact small groups of axons within a nerve fascicle”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Interface Electrodes · p. 4Compatibility between a technical and a biological system can be divided into structural biocompatibility and surface biocompatibility.
“In general, the compatibility between a technical and a biological system (Bronzino, 1995) can be divided into the structural biocompatibility and the surface biocompatibility”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Bronzino, 1995
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisereasoning§ Epimysial and intramuscular electrodes · p. 7Epimysial electrodes offer a good signal-to-noise ratio because of their position and large surface and provide exact selectivity, namely the implanted muscle.
“In general, they offer a good signal-to-noise ratio because of their position and large surface and provide exact selectivity, namely the implanted muscle.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ Surface electrodes · p. 5For recording purposes, electrodes made of silver/silver chloride in the form of bars (10-1 mm) have been found to give adequate signal-to-noise ratio.
“For recording purposes, electrodes made of silver/silver chloride in the form of bars (10 -1 mm) have been found to give adequate signal-to-noise ratio.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Surface electrodes · p. 5Multichannel surface EMG electrodes have been developed that enable measurements of motor unit potentials size and number and provide topographical information beyond the capability of intramuscular needle electrodes.
“Multichannel (linear and high-density arrays) surface EMG electrodes have been developed that enable measurements of motor unit potentials size and number and provide topographical information that is beyond the capability of intramuscular needle electrodes”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Zwarts and Stegeman, 2003 source
May combine two claims.
Premisecitation§ Surface electrodes · p. 5Surface electrodes can also supply a perceivable sensation to the skin or excite nerve tracts or muscles underneath the skin.
“they can also be employed to supply a perceivable sensation to the skin or to excite nerve tracts or muscles underneath the skin”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
Interpretationnone given§ Non-electrical interfaces · p. 7Disadvantages of acoustic myography include susceptibility to interference by environmental noise and vibrations and limited experience with its use.
“Disadvantagesinclude the susceptibility to interference by environmental noise and vibrations and limited experience with its use.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 8A final issue with percutaneous electrodes is that muscle contractions may result in eventual breakage, although these problems have become less frequent with the improvement of electrodes and lead wires.
“A final issue with percutaneous electrodes is that muscle contractions may result in eventual breakage, although these problems have become less frequent with the improvement of electrodes and lead wires”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Onishi et al., 2000 source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ Surface electrodes · p. 5The active recording electrode should be located near but not on the motor point because at this point the stability of signals suffers from minor physical displacement of the electrode.
“The active recording electrode should be located near but not on the motor point because at this point the stability of signals suffers from minor physical displacement of the electrode”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites De Luca, 2002
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationreasoning§ Non-electrical interfaces · p. 7Further technical developments to increase sensitivity of measurements are needed before magnetic interfaces can be used for single-unit activity recordings.
“Further technical developments to increase sensitivity of measurements are needed before magnetic interfaces can be used for single-unit activity recordings.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone given§ Epimysial and intramuscular electrodes · p. 7Although epimysial electrodes can be used for EMG recording, their main application area is for muscle stimulation.
“Although they can be used for EMG recording, the main application area is for muscle stimulation”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone given§ Interface Electrodes · p. 5The number of electrodes to be used depends on the working application: a low number for robust use and limited functionality, or high numbers for good spatial resolution and selectivity.
“The number of electrodes to be used depends on the working application: low number of electrodes for robust use and limited functionality or high numbers for good spatial resolution and selectivity”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone given§ Non-electrical interfaces · p. 7Muscle sounds have been used to measure twitch force, monitor fatigue, and in the diagnosis of muscle diseases.
“Muscle sounds have been used to measure twitch force, monitor fatigue and in the diagnosis of muscle diseases.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone given§ Non-electrical interfaces · p. 7As a control signal for externally powered prostheses, acoustic myography may present advantages over surface EMG because it is unaffected by changes in skin impedance, less sensitive to placement on the muscle, and requires less amplification and electrical shielding.
“As a control signal for externally powered prostheses, acoustic myography may present some advantages over surface EMG because it is unaffected by changes in skin impedance, less sensitive to placement on the muscle, and requires less amplification and electrical shielding.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Surface electrodes · p. 5The impedance at the electrode-skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type.
“The impedance at the electrode-skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hewson et al., 2003 source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisereasoning§ Interface Electrodes · p. 5Cuff and intrafascicular electrodes can be applied to intact nerves in acute or chronic studies, whereas regenerative sieve electrodes are implanted in transected nerves that need to regenerate across the electrode sieve over several months.
“cuff and intrafascicular electrodes can be applied to intact nerves in acute or chronic studies, whereas, by definition, regenerative sieve electrodes are implanted in transected nerves that need to regenerate across the electrode sieve over several months.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 7Intramuscular electrodes may damage many muscle fibers, but in most instances this loss will not represent a noticeable reduction in muscle force.
“These electrodes may damage many muscle fibers, but in most instances this loss will not represent a noticeable reduction in muscle force”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Mortimer et al., 1980 source
May combine two claims.
Interpretationreasoning§ Non-electrical interfaces · p. 7Measurements for magnetoneurography are usually performed inside a magnetically shielded room to minimize external contamination, limiting its application for control of neuroprostheses.
“To minimize external contamination, measurements are usually performed inside a magnetically shielded room, thus limiting its application for control of neuroprostheses.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 7Compared with nerve cuff electrodes, intramuscular electrodes have been shown to produce more reliable, more graded, and less fatigable recruitment of motor units.
“Compared with nerve cuff electrodes, intramuscular electrodes have been shown to produce more reliable, more graded, and less fatigable recruitment of motor units”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Singh et al., 2000
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ Epimysial and intramuscular electrodes · p. 7High energy consumption is required to activate the muscle with epimysial electrodes.
“In addition, high energy consumption is required to activate the muscle.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationreasoning§ Epimysial and intramuscular electrodes · p. 8To avoid problems derived from excess charge delivery, stimulation is given in cycles, and the system is explanted once regenerating axons reinnervate the muscles of interest.
“To avoid problems derived from excess charge delivery, stimulation is given in cycles, and the system is explanted once regenerating axons reinnervate the muscles of interest.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 8The BION implantable microstimulator consists of a small cylindrical capsule (16-2 mm) whose internal components are connected to electrodes sealed hermetically into its ends.
“This device consists of a small cylindrical capsule (16 -2 mm) whose internal components are connected to electrodes sealed hermetically into its ends.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Loeb et al., 2001 source
May combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 7Electrode and lead failures or infections have been found to be rare and the electrode response was stable over time in patients with intramuscular or epimysial electrodes during more than 3 years.
“Electrode and lead failures or infections have been found to be rare and the electrode response was stable over time in patients with intramuscular or epimysial electrodes during more than 3 years”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kilgore et al., 2003 source
May combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 8The currents required to produce contractions are high and may induce discomfort and tissue damage.
“In addition, the currents required to produce contractions are high and may induce discomfort and tissue damage”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Chae and Hart, 1998 source
May combine two claims.
Interpretationnone given§ Epimysial and intramuscular electrodes · p. 8Adequate EMG signals can be obtained with epimysial and intramuscular electrodes, but the fatigue of the muscle and the derived change of the recorded signal have to be taken into account for long-term use.
“Adequate EMG signals can be obtained with epimysial and intramuscular electrodes, but the fatigue of the muscle and the derived change of the recorded signal have to be taken into account for long-term use.”
Quote in the paperSubstantial claimMay combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 9The BION electrode was found to be stable, biocompatible, and effective in experimental animals when implanted passively or used for muscle stimulation over long-term implantation.
“Tests performed indicated that the BION is a stable, biocompatible, and effective electrode when implanted passively or used to stimulate muscles in experimental animals for long-term implantation (Cameron et al., 1998) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Cameron et al., 1998 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone given§ Epimysial and intramuscular electrodes · p. 7Management may be difficult when many epimysial electrodes on different muscles are used because of the interconnection between them, providing additional risks and possible source of failure.
“Management may be difficult when many epimysial electrodes on different muscles are used because of the interconnection between them, providing additional risks and possible source of failure.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone given§ Surface electrodes · p. 5The main advantage of surface electrodes being non-invasive and easily adaptable is counteracted by disadvantages such as the need for daily placement and frequent calibration, and low reproducibility and quality of acquired signals.
“Their main advantage of being non-invasive and easily adaptable is counteracted by several disadvantages, such as the need for daily placement and frequent calibration, and by the low reproducibility and quality of acquired signals.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone given§ Surface electrodes · p. 5Recorded EMG signals are used to determine the activation time of a muscle, estimate the force produced during muscle contraction, and estimate the rate of muscle fatigue.
“Currently, recorded EMG signals are used to determine the activation time of a muscle, to estimate the force produced during muscle contraction and to estimate the rate of muscle fatigue.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Interface Electrodes · p. 5Capacitive electrodes such as titanium nitride might reduce the risk of corrosion under stimulation conditions, but their performance highly varies with the fabrication technology.
“Capacitive electrodes such as titanium nitride might reduce the risk of corrosion under stimulation conditions, but their performance highly varies with the fabrication technology”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Janders et al., 1996
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Weiland et al., 2002 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ Epimysial and intramuscular electrodes · p. 8For large muscles, the EMG activity mainly reflects the action potentials of motor units located more superficial or closer to the electrode.
“For large muscles, the EMG activity mainly reflects the action potentials of motor units located more superficial or closer to the electrode.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationcitation§ Epimysial and intramuscular electrodes · p. 8This requires the implantation and maintenance of many electrodes subjected to considerable mechanical stress caused by strong muscle contractions.
“This requires the implantation and maintenance of many electrodes subjected to considerable mechanical stress caused by strong muscle contractions”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Devasahayam, 1992 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone given§ Interface Electrodes · p. 5The choice of an adequate electrode size and material, in combination with a structurally compatible design, is always a compromise between electrode impedance, signal-to-noise ratio, and selectivity.
“The choice of an adequate electrode size and material, in combination with a 'structural compatible' design, is always a compromise between electrode impedance, signal-to-noise ratio, and selectivity.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Epimysial and intramuscular electrodes · p. 8Direct muscle stimulation may be used to recover chronically denervated muscles, although activation requires long duration pulses and extremely high current that are usually not generated by common FES devices.
“Direct muscle stimulation may be also used to recover chronically denervated muscles, although activation requires long duration pulses and extremely high current that are usually not generated by common FES devices”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kern et al., 2002 source
May combine two claims.
Premisecitation§ Interfascicular electrodes · p. 12There was no gross evidence of axonal damage one day after implantation of the slowly penetrating interfascicular nerve electrode, but chronic implants have not been reported.
“There was no gross evidence of axonal damage after 1 day, but chronic implants have not been reported.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Tyler and Durand, 1997 source
May combine two claims.
Premisecitation§ Cuff electrodes · p. 10Quasi-trapezoidal stimulation pulses induce more gradual contraction than rectangular pulses, an effect attributed to activation of fatigue-resistant motor units before large fast-fatigue units with cuff electrodes.
“For example, quasi-trapezoidal stimulation pulses have been shown to induce more gradual contraction than rectangular pulses, an effect that was attributed to activation of fatigue-resistant motor units before activation of large fast-fatigue units with cuff electrodes (Fang and Mortimer, 1991; Rozman et al., 1993) .”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Cuff electrodes · p. 10Using short pulsewidths and a subthreshold transverse current from a steering anode restricts the region of excitation of the nerve trunk and significantly improves stimulation selectivity.
“The use of short pulsewidths and the introduction of a subthreshold transverse current from a steering anode allows restriction of the region of excitation of the nerve trunk, significantly improving the selectivity of stimulation (Gorman and Mortimer, 1986; Sweeney et al., 1990; Grill and Mortimer, 1996b; Navarro et al., 2001) .”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites Gorman and Mortimer, 1986
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Sweeney et al., 1990 source
- cites Grill and Mortimer, 1996b
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisecitation§ Cuff electrodes · p. 10Chronic implantation of snug cuffs modifies nerve shape and produces loss of large nerve fibers, which are the most sensitive to compression.
“Although snug-fitting nerve cuffs have been advocated to reduce the stimulus charge injection or to obtain a high signal-to-noise ratio for neural recordings (Naples et al., 1990) , different studies have shown that chronic implantation of snug cuffs modifies the nerve shape and produces a loss of large nerve fibers, which are the most sensitive to compression (Krarup et al., 1989; Grill and Mort”
Quote in the paperSubstantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Cuff electrodes · p. 11A tripolar electrode configuration is preferred to reduce noise pickup and provides a biophysically ideal way to use a differential amplifier to reject large bioelectric signals arising outside the cuff.
“Tripolar (central active electrode vs. two reference electrodes near the ends of the cuff) electrode configuration is preferred to reduce noise pickup (Stein et al., 1977) ; it provides a biophysically ideal way to use a differential amplifier to reject almost completely the large bioelectric signals arising from outside the cuff (e.g., EMG).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stein et al., 1977 source
May combine two claims.
Premisecitation§ Interfascicular electrodes · p. 12Histological cross-sections showed that the slowly penetrating interfascicular nerve electrode rearranges the epineurium and penetrates deep within a multifascicular nerve but does not disrupt the perineurium.
“Histological cross-sections showed that the SPINE rearranges the epineurium and penetrates deep within a multifascicular nerve but did not disrupt the perineurium (Tyler and Durand, 1997) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Tyler and Durand, 1997 source
May combine two claims.
Premisecitation§ Flat-interface nerve electrodes · p. 11Flattening a nerve into a more elliptical shape makes fascicles more accessible and moves central fibers closer to the stimulating electrode compared with cylindrical cuffs.
“By flattening the nerve into a more elliptical shape, fascicles become more accessible and central fibers are moved closer to the stimulating electrode in comparison with cylindrical cuffs.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Tyler and Durand, 2002 source
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Book electrodes · p. 9The first micturition caused by electrical nerve root stimulation was reported in 1973.
“The first micturition because of electrical nerve root stimulation was reported in 1973 (Brindley, 1973) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Brindley, 1973
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisecitation§ Epineurial and helicoidal electrodes · p. 9Postoperative infections after helicoidal electrode implants in vagus nerve patients occur in less than 3% of cases, and removal of the device is required in 1%.
“Postoperative infections after helicoidal electrode implants in vagus nerve patients occur in less than 3% of cases, requiring removal of the device in 1%.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites McLachlan, 1997 source
May combine two claims.
Premisecitation§ Epineurial and helicoidal electrodes · p. 9Helicoidal electrodes conform to the shape of the nerve and minimize mechanical trauma.
“Helicoidal electrodes, placed circumjacent to the nerve, are made of flexible platinum ribbon with an open helical design (Fig. 5), allowing the electrode to conform to the shape of the nerve and minimizing mechanical trauma (Naples et al., 1990) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Naples et al., 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisecitation§ Flat-interface nerve electrodes · p. 12Flat-interface nerve electrodes that moderately flattened the nerve demonstrated the best selectivity for limb motion measurements, and this selectivity was maintained throughout the implant time.
“Electrodes that moderately flattened the nerve demonstrated the best selectivity for limb motion measurements, which were maintained throughout the implant time (Tyler and Durand, 2003; Leventhal and Durand, 2004) .”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Epineurial and helicoidal electrodes · p. 9Using epineurial electrodes for neuropathic pain has limited benefit because it should be applied only when pain involves one nerve territory in the head or limbs, and it has frequent morbidity from electrode approximation, electrode displacement, and lead failure.
“The latter application is of limited benefit because it should be applied only when pain involves one nerve territory in the head or the limbs and has frequent morbidity associated with electrode approximation, electrode displacement, and lead failure (Hassenbusch et al., 1996) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hassenbusch et al., 1996 source
May combine two claims.
Premisecitation§ Flat-interface nerve electrodes · p. 12Acute experiments showed that individual fascicles of the cat sciatic nerve, and groups of fibers within fascicles, can be selectively activated with the flat-interface nerve electrode, and selectivity strongly depends on the relative locations of the fascicle and electrode contacts.
“Results on acute experiments showed that it is possible to selectively activate individual fascicles of the cat sciatic nerve with the FINE, as well as groups of fibers within the fascicles, and revealed the strong dependency of selectivity on the relative locations of the fascicle and the electrode contacts (Tyler and Durand, 2002; Leventhal and Durand, 2003) .”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Premisereasoning§ Book electrodes · p. 9Ventral root stimulation causes simultaneous contraction of the sphincter and detrusor muscles, so voiding occurs in brief spurts at the end of stimulation when the striated sphincter relaxes faster than the smooth detrusor.
“Because ventral root stimulation produces simultaneous contraction of both sphincter and detrusor muscles, voiding is achieved in brief spurts at the end of the stimulation, when the striated sphincter muscle relaxes faster than the smooth detrusor muscle.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Epineurial and helicoidal electrodes · p. 9Occasional fracture of the electrode wire has been reported for helicoidal electrodes.
“Occasional fracture of the electrode wire has been reported (McLachlan, 1997) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites McLachlan, 1997 source
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Book electrodes · p. 9There are now 2,000 patients worldwide with a urinary bladder stimulator.
“Nowadays, there are 2,000 patients worldwide that carry a urinary bladder stimulator (Rijkhoff, 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rijkhoff, 2004 source
May combine two claims.
Premisecitation§ Epineurial and helicoidal electrodes · p. 9Epineurial electrodes have been used in biomedical functional electrical stimulation for breathing control by phrenic nerve stimulation, foot-drop improvement, and relief of neuropathic pain.
“Epineurial electrodes are used in biomedical FES applications for breathing control by phrenic nerve stimulation (Creasey et al., 1996) , foot-drop improvement (Liberson et al., 1961) , and relief of neuropathic pain (Strege et al., 1994) .”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites Creasey et al., 1996
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Liberson et al., 1961
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Strege et al., 1994 source
May combine two claims.
Premisecitation§ Flat-interface nerve electrodes · p. 12Chronic studies in laboratory animals with flat-interface nerve electrodes implanted for 1 to 3 months showed that high reshaping force induced nerve damage, whereas moderate and small forces caused no detectable change in nerve physiology and histology.
“Chronic studies in laboratory animals with FINE implanted over 1-3 months showed that electrodes applying high reshaping force induced nerve damage, whereas those with moderate and small forces did not cause any detectable change in nerve physiology and histology.”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Regenerative electrodes · p. 16The applicability of regenerative electrodes depends on successful axonal regeneration through the perforations or holes, the possibility of nerve damage from mechanical load or constrictive forces within the holes, and the biocompatibility of the components.
“Applicability of regenerative electrodes is dependent on the success of axonal regeneration through the perforations or holes, the possibility of nerve damage from the mechanical load imposed by the electrode or from constrictive forces within the holes, and the biocompatibility of the components (Rosen et al., 1990; Navarro et al., 1996) .”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15Recruitment curves for the Utah electrode array were broader, with twitch thresholds starting at much lower currents than usual for cuff electrodes.
“Recruitment curves for the electrode array were broader with twitch thresholds starting at much lower currents than usual for cuff electrodes.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2001 source
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Penetrating microelectrodes · p. 13A study used a microelectrode to stimulate the sensory nerve of an operator to indicate contact with an object by a remote robotic hand.
“An interesting study used a microelectrode to stimulate the sensory nerve of an operator to indicate contact with an object by a remote robotic hand (Shimojo et al., 2003) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Shimojo et al., 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Penetrating microelectrodes · p. 15A negative aspect in the volunteer case was gradual degradation of the electrode wire bundle.
“One negative aspect was the gradual degradation of the electrode wire bundle.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Warwick et al., 2003 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15A slanted array with electrodes of varying length from 0.5 to 1.5 mm and 0.1 mm difference between rows of neighboring electrodes was developed to decrease redundant electrodes and provide access to more fascicles within the nerve.
“To decrease the number of redundant electrodes and provide access to more fascicles within the nerve, a slanted array with electrodes of varying length, ranging from 0.5 to 1.5 mm with 0.1 mm difference in length between rows of neighboring electrodes, has been developed (Branner et al., 2001) .”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Branner et al., 2001 source
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16Using multiple-hole silicon arrays, researchers demonstrated axonal regeneration and neural activity recording in peripheral nerves of rat, frog, and fish.
“Using multiple-hole silicon arrays, the researchers demonstrated axonal regeneration and even neural activity recording was demonstrated in peripheral nerves of rat, frog, and fish (Kovacs et al., 1994; Navarro et al., 1996; Bradley et al., 1997; Della Santina et al., 1997; Mensinger et al., 2000) .”
Quote in the paperNot a substantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Intrafascicular electrodes · p. 13Activity from cutaneous receptors in response to stimuli has been recorded in several studies, but the activity was small and difficult to distinguish from background noise.
“Activity arising from cutaneous receptors in response to stimuli has been recorded in several studies, although activity was found to be small and difficult to distinguish from the background noise in the recording (Malmstrom et al., 1998; Yoshida et al., 2000) .”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Penetrating microelectrodes · p. 14Rutten and colleagues concluded that an electrode separation of 120 μm was optimal for selective interfacing of peripheral nerves such as the rat peroneal nerve containing around 350 alphamotor fibers.
“In a series of experimental and modeling studies, Rutten and colleagues concluded that an electrode separation of 120 m m was optimal for selective interfacing peripheral nerves, such as the rat peroneal”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rutten, 2002 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 14A trade-off must be made between selectivity of stimulation and the total number of nerve fibers that can be activated, and optimal selectivity requires redundant electrodes in a 3D configuration.
“Thus, a trade-off must be made between selectivity of stimulation and the total number of nerve fibers that can be activated; optimal selectivity requires the use of redundant electrodes in the 3D configuration (Rutten et al., 1999) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rutten et al., 1999 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 13Microstimulation through a microneurography electrode can activate single axons, although in most cases it affects small groups of axons in the fascicle.
“Microstimulation through the electrode can be used to activate single axons, although in most cases stimulation affects small groups of axons in the fascicle (Ochoa and Torebjo ¨ rk, 1983; Torebjo ¨ rk et al., 1987) .”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Ochoa and Torebjo ¨ rk, 1983 source
- cites Torebjo ¨ rk et al., 1987
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16Regenerating axons grow through the holes of regenerative electrodes, making it possible to record action potentials from and stimulate individual axons or small fascicles.
“Regenerating axons eventually grow through the holes (Fig. 12), making it possible to record action potentials from and to stimulate individual axons or small fascicles.”
Quote in the paperSubstantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15The volunteer received feedback from force and slip sensors in a prosthetic hand and used the array to control the hand for grasping an unseen object.
“The subject received feedback information from force and slip sensors in a prosthetic hand and subsequently used the array to control the hand for grasping an unseen object.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Warwick et al., 2003 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15Stimulation properties stabilized in 80% of the Utah array electrodes over the course of the experiment, but the recorded sensory signals were not stable over time.
“The stimulation properties stabilized in 80% of the electrodes over the course of the experiment, but the recorded sensory signals were not stable over time.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2004 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15The Utah multielectrode array can record more effectively from more dorsal root ganglion neurons than conventional recording techniques, providing a more stable option for chronic implantation.
“The Utah MEA is also capable of recording more effectively from more dorsal root ganglion neurons than has been achieved by conventional recording techniques, providing a more stable option for chronic implantation (Aoyagi et al., 2003) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Aoyagi et al., 2003 source
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16A one-to-one design allowing access to each individual regenerated axon through one hole of 2-10 μm diameter has proved impossible, because nerve regeneration fails with holes of such small size.
“Ideally a one-to-one design would allow access to each individual regenerated axon grown through one hole (2-10 m m in diameter). However, this has been proved impossible; nerve regeneration fails with holes of such small size.”
Quote in the paperSubstantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15Electrodes in the Utah array selectively recorded single-unit responses from mechanoreceptors, although only in 10-20% of the electrodes, and evoked graded recruitment of force in muscle groups in a highly selective fashion with current injections in the 1-20 μA range.
“Electrodes in the array were capable of selectively recording single-unit responses from mechanoreceptors, although only in 10-20% of the electrodes, and evoked graded recruitment of force in muscle groups in a highly selective fashion with current injections in the 1-20 m A range (Branner and Norman, 2000; Branner et al., 2001) .”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Intrafascicular electrodes · p. 13Longitudinal intrafascicular electrodes offer good selectivity for stimulation and for multiunit extracellular recording, and their geometry makes them suitable for long-term implantation.
“The LIFEs offer good selectivity for stimulation (Nannini and Horch, 1991; Yoshida and Horch, 1993) and for multiunit extracellular recording (Goodall et al., 1991; McNaughton and Horch, 1994; Yoshida and Stein, 1999) , and their geometry, make them suitable for longterm implantations (Lefurge et al., 1991) .”
Quote in the paperSubstantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Intrafascicular electrodes · p. 13The use of multistrand Kevlar longitudinal intrafascicular electrodes increases tensile strength and signal-to-noise ratio.
“The use of multistrand Kevlar LIFEs (m-polyLIFE) increases the tensile strength and signal-to-noise ratio (Lawrence et al., 2004) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Lawrence et al., 2004 source
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16Silicon regenerative interfaces cause frequent signs of axonopathy and constitute a physical barrier that limits elongation of regenerating axons depending on the size of the holes.
“However, such silicon interfaces cause frequent signs of axonopathy and constitute a physical barrier that limits the elongation of regenerating axons depending on the size of the holes (Edell, 1986; Rosen et al., 1990; Navarro et al., 1996; Zhao et al., 1997) .”
Quote in the paperSubstantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 15Both recording and stimulation with the Utah array were stable over the 36-hour-long period of the experiments.
“Both the recording and stimulation were stable over the 36-h-long period of the experiments.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Branner et al., 2001 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 14Loss of neurons in the spiral ganglion and cochlear nucleus was found after cochlear nerve microelectrode implants, thought to be due to difficulty with electrode insertion and chronic motion of the implants.
“Nevertheless, the loss of neurons in spiral ganglion and cochlear nucleus was found, thought to be due to difficulty with electrode insertion and chronic motion of the implants.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Arts et al., 2003 source
May combine two claims.
Premisecitation§ Penetrating microelectrodes · p. 14A silicon-based ribbon electrode with high flexibility functioned successfully after chronic implantation of up to one year, but its interconnect cable was not designed for the highly mobile somatic peripheral nerve and was too stiff and brittle for intrafascicular application.
“A silicon-based ribbon electrode with high flexibility demonstrated successful function after chronic implantation of up to 1 year (Hetke et al., 1994) , but this interconnect cable was not designed for implants in the highly mobile somatic peripheral nerve and was too stiff and brittle for intrafascicular application (Najafi and Hetke, 1990) .”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone given§ Bioelectronic interfaces · p. 19In the PNS, bioelectronic interfaces are in an early and experimental stage.
“In the PNS, bioelectronic interfaces are in an early and experimental stage.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Bioelectronic interfaces · p. 19Pine et al. placed single rat hippocampus cells into a 3D silicon structure and demonstrated that the transplanted cells retained capacity for axonal outgrowth and that nerve action potentials could be recorded extracellularly from the soma.
“Pine et al . (1994) managed to place single hippocampus cells of the rat into a 3D silicon structure, which was termed 'cultured neuron probe.' They could demonstrate that the transplanted cells inside the microcompartments retained their capacity for axonal outgrowth and that nerve action potentials could be directly recorded extracellularly from the soma of the cells.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Pine et al., 1994
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Bioelectronic interfaces · p. 19Fromherz et al. obtained a current-free measurement of the membrane potential of a single leech neuron by direct coupling to the gate of a field effect transistor.
“they obtained a current-free measurement of the membrane potential of a single neuron of a leech by direct coupling to the gate of a field effect transistor (FET) (Fromherz et al., 1991a)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Fromherz et al., 1991a
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16Chronically implanted regenerative electrodes have allowed stimulation of different nerve bundles and recording of nerve action potentials in response to functional stimulation.
“been possible to stimulate different nerve bundles and to record nerve action potentials in response to functional stimulation (Navarro et al., 1998; Ceballos et al., 2002)”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Navarro et al., 1998
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Ceballos et al., 2002 source
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ EMG-based control of hand prostheses · p. 20The formal scheme for EMG acquisition and analysis for prosthetic control includes signal conditioning, pre-processing, onset detection, feature extraction, dimensionality reduction, pattern recognition, and off-line and on-line learning.
“the formal scheme for the acquisition and analysis of EMG signals for the control of prosthetic devices is composed of several modules (Zecca et al., 2002) : (1) signal conditioning, pre-processing, and detection of onset movement (Micera et al., 1999) ; (2) feature extraction; (3) dimensionality reduction; (4) pattern recognition; and (5) off-line and on-line learning.”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18Fabrication of silicon-based interconnection cables is technologically feasible.
“The fabrication of silicon-based interconnection cables was shown to be technologically feasible (Hetke et al., 1990; 1994)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Hetke et al., 1990; 1994
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Interpretationreasoning§ Regenerative electrodes · p. 16Interfacing regenerated axons with regenerative electrodes requires time, so acute experiments are not possible.
“some time is needed for interfacing the regenerated axons, thus precluding acute experiments”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Regenerative electrodes · p. 16The presence of a polyimide-based regenerative electrode showed no chronic foreign body response, and the immunohistochemically measured pattern of characteristic proteins was comparable with a healing reaction without any implant.
“The presence of a polyimide-based regenerative electrode showed no chronic foreign body response, and the immunohistochemically measured pattern of characteristic proteins was comparable with a healing reaction without any implant (Klinge et al., 2001a) .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Klinge et al., 2001a source
May combine two claims.
Interpretationnone given§ EMG-based control of hand prostheses · p. 20In EMG-based control of hand prostheses, homologous muscles usually cannot be used to control prosthetic movements, so complex pattern recognition algorithms are required.
“In this case, it is usually not possible to use the 'homologous' muscles to control the movements of the prosthetic device, and the development of a complex algorithm exploiting the potentialities of using advanced pattern recognition techniques is required.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ EMG-based control of hand prostheses · p. 20The EMG signal was modeled as amplitude-modulated Gaussian noise whose variance is related to the force developed by the muscle.
“the EMG signal was modeled as the amplitude-modulated Gaussian noise whose variance is related to the force developed by the muscle.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Graupe and Cline, 1975; 1982
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Graupe et al., 1982 source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone given§ EMG-based control of neuroprostheses and hybrid bionic systems · p. 20sEMG signals are easy to record and provide important access to the neuromuscular system of the user.
“sEMG signals are easy to record and provide an important access to the neuromuscular system of the user.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Bioelectronic interfaces · p. 19The structure's topology, such as the width of grooves, has an important impact on the growth of cultured cells.
“The structure's topology (e.g., the width of grooves) has an important impact on the growth of the cultured cells (Wilkinson et al., 1987; Clark et al., 1990; Connolly et al., 1992)”
Quote in the paperNot a substantial claim0 of 3 sources checkedMay combine two claims.
Premisecitation§ Materials for substrates, insulation, and encapsulation · p. 18Silicone elastomer remains the material of choice for encapsulating cables and for additional coating of hermetically sealed electronic circuits.
“Silicone elastomer is still the material of choice for encapsulation of cables and also as additional coating for hermetically sealed electronic circuits (Stieglitz et al., 2004)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Stieglitz et al., 2004 source
May combine two claims.
Premisecitation§ Regenerative electrodes · p. 16Polyimide-based electrodes have been shown to be biocompatible and stable over months of in vivo implantation and allow much better regeneration than silicon dice.
“Polyimide-based electrodes have been shown to be biocompatible and stable over months of in vivo implantation and allow for much better regeneration than silicon dice (Navarro et al., 1998; Ceballos et al., 2002; Lago et al., 2005) .”
Quote in the paperNot a substantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ Bioelectronic interfaces · p. 19Fromherz et al. demonstrated guided growth of leech nerve cells on a laminin-coated substrate.
“Fromherz et al . (1991b) could demonstrate the guided growth of nerve cells from leeches on a laminin-coated substrate.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Fromherz et al., 1991b
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationreasoning§ Regenerative electrodes · p. 16Regenerative electrodes can be implanted in severed nerves of an amputee's limb to create a bidirectional interface for a feedback-controlled neuroprosthesis.
“One of the most logical and challenging applications of regenerative electrodes consists of their implantation in severed nerves of an amputee's limb for bidirectional interface in a feedback-controlled neuroprosthesis.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Bioelectronic interfaces · p. 19In the cone electrode, a piece of rat sciatic nerve was placed into a cone, and after implantation in rat brain, neurite ingrowth was observed and stable recordings were obtained over 6 months.
“Here, a piece of a rat sciatic nerve was placed into a cone of 1.5 mm length and 100-200 m mdiameter. After implantation in the brain of a rat, neurite ingrowth was observed and stable recordings were obtained over 6 months.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kennedy, 1989 source
May combine two claims.
Premisereasoning§ Bioelectronic interfaces · p. 19Neither the cultured neuron probe nor the cone electrode approach has been investigated in the PNS.
“Neither approach has been investigated in the PNS.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Materials for the electrodes · p. 18Platinum, iridium, tungsten, and stainless steel are typically used as conductors, and silicone elastomer, polytetrafluoroethylene, and polyimide as insulating carriers in neural electrodes.
“Materials typically used are platinum, iridium, tungsten, and stainless steel as conductors and silicone elastomer, polytetrafluoroethylene, and polyimide as insulating carriers (Naples et al., 1990; Heiduschka and Thanos, 1998)”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Naples et al., 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Heiduschka and Thanos, 1998 source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ Bioelectronic interfaces · p. 19With surface topologies on the substrate, axonal outgrowth of embryonic rat nerve cells was observed and electrical activity was recorded in a microstructure with 16 electrodes connected with grooves.
“Having surface topologies on the substrate, axonal outgrowth of embryonic nerve cells from rats was observed and electrical activity was recorded in a microstructure with 16 electrodes that were connected with grooves (Jimbo and Kawana, 1992; Jimbo et al., 1993)”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
Premisecitation§ Bioelectronic interfaces · p. 19Reinnervation of the gastrocnemius and tibialis muscles via the distal stump of the sciatic nerve was observed after a few weeks, although there was no communication with the CNS.
“Reinnervation of the gastrocnemius and tibialis muscles via the distal stump of the sciatic nerve was observed after a few weeks, although there was no communication with the CNS.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Katsuki et al., 1997 source
Classified as probably not a substantial claim.
Premisecitation§ Regenerative electrodes · p. 16Stimulation of afferent nerve fibers in the residual limb can provide sensory feedback from tactile and force sensors to an amputee.
“sensory feedback from tactile and force sensors might be provided to the user through stimulation of afferent nerve fibers within the residual limb (Riso, 1999)”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Riso, 1999 source
May combine two claims.
Interpretationreasoning§ Materials for substrates, insulation, and encapsulation · p. 18Cost, technology, and limited cable length restrict widespread use of silicon-based interconnection cables in neural prostheses.
“but cost, technology, and limited cable length limit widespread use in neural prostheses applications”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone given§ Choice of fabrication technology · p. 18Devices made by precision mechanics are robust, stable, and reliable but have low spatial resolution, low electrode selectivity, and quite large size.
“The devices are robust, stable, and reliable but suffer from a low spatial resolution and low selectivity of the electrodes as well as from a quite large size.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ Bioelectronic interfaces · p. 19Functional regeneration was proven, but the number of surviving cells decreased from 10,000 injected to some tens that finally innervated the muscle.
“Functional regeneration was proven but the number of surviving cells tremendously decreased from the 10,000 that had been injected to some tens that finally innervated the muscle (T. Stieglitz, 2004. The neuron micro probe project, personal communication)”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites T. Stieglitz, 2004. The neuron micro probe project, personal communication source
May combine two claims.
Interpretationreasoning§ Materials for the electrodes · p. 18Deterioration of an implanted electrode may cause implant failure and release toxic products.
“deterioration of the device may result in implant failure and the release of toxic products”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ EMG-based control of exoskeletons · p. 21Using this neuro-fuzzy approach, it was possible to control different degrees of freedom of the shoulder while reducing the muscular effort of the subjects.
“Using this approach, it was possible to control the different degrees of freedom of the shoulder reducing the muscular effort of the subjects.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Kiguchi et al., 2001; 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisereasoning§ ENG-based control of neuroprostheses and hybrid bionic systems · p. 22Neuro-fuzzy models seem most adequate for processing ENG because of their model-free structure.
“Neuro-fuzzy models seem the most adequate because of their model-free structure.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ EMG-based control of exoskeletons · p. 21Experimental tests have shown that use of sEMG significantly improved the mechanical gain of the system while maintaining natural control.
“Experimental tests have shown that the use of sEMG significantly improved the mechanical gain of the system (i.e., the possibility of augmenting capabilities and/or reducing human muscular efforts to carry out the different tasks) while maintaining natural control of the system.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rosen et al., 2001 source
Probably not a substantial claim: may not pass S4 (based).
Interpretationnone given§ Conclusion · p. 23At present, there is no universal best choice for all possible uses of the described PNS interfaces.
“At the moment, there is no universal best choice for all possible uses of the described interfaces”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Premisecitation§ EMG-based control of exoskeletons · p. 21Using this myoprocessor scheme, a one-degree-of-freedom exoskeleton was controlled using dynamic, kinematic, and physiological feedback.
“Using this schemea one-degree-of-freedom exoskeleton has been controlled using a multiple-feedback approach: (1) dynamic feedback - the moments generated at the interfaces between the human arm, the external load, and the exoskeleton structure; (2) kinematic feedback -the joint angular trajectories recorded using encoders, and the angular velocity and acceleration calculated by finite differences;”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Rosen et al., 2001 source
May combine two claims.
Interpretationnone given§ EMG-based control of hand prostheses · p. 21The continuous classifier approach seems promising for EMG-controlled hand prostheses, but EMG signals seem limited to prostheses with few degrees of freedom, probably no more than four.
“This approach seems promising for the development of EMG-based controlled hand prostheses although the use of EMG signals seems to be limited to prosthesis with a few degrees of freedom (probably no more than four).”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationnone given§ EMG-based control of hand prostheses · p. 21With the EMG approach, the user must initiate all contractions to control the prosthesis from rest, making it difficult to switch between classes and increasing response time in unexpected situations.
“by using this approach, the user must initiate all the contractions to control the prosthesis from rest, making it difficult to switch from one class to another and increasing the response time of the user when dealing with unexpected situations.”
Quote in the paperSubstantial claim
The paper relies on this without support.
Nerve fibers are recruited by electrical stimulation according to their thickness, so large motor fibers are activated earlier than thinner fibers.
“nerve fibers are recruited by electrical stimulation according to their thickness, therefore large motor fibers innervating the fast-fatigue and strong motor units are activated earlier than the physiologically first-recruited thinner fibers controlling slow fatigue-resistant motor units”
The paper relies on this without support.
Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.
“Recruitment of motor units follows an order of size, with slow fatigue-resistant motor units activated first and large fast-fatigue motor units activated only at high levels of tension.”
The paper relies on this without support.
Signals are transmitted by axons in series of impulses or action potentials, with intensity mainly coded in impulse frequency along the peripheral axon.
“Signals are transmitted by the corresponding axons in series of impulses or action potentials, with intensity of the signal mainly coded in impulse frequency along the peripheral axon.”
The paper relies on this without support.
The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal compared with proximal segments.
“The fascicular architecture changes throughout the length of the nerve, with an increasing number of fascicles of smaller size in distal with respect to proximal segments.”
The paper relies on this without support.
The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.
“The perineurium is the main contributor to the tensile strength of the nerve, acts as a diffusion barrier, and maintains the endoneurial fluid pressure.”
The paper relies on this without support.
Using multiple wire needles to interface a high number of nerve fibers has practical problems for long-term use.
“However, the use of multiple wire needles for interfacing a high number of nerve fibers has practical problems for long-term use.”
The paper relies on this without support.
No human implants of regenerative electrodes have been reported.
“No human implants of regenerative electrodes have been reported.”
The paper relies on this without support.
Cuff electrodes are the type of peripheral nervous system interface electrode most investigated in basic and applied research.
“Cuff electrodes are the type of PNS interface electrode most investigated in basic and applied research.”
The paper relies on this without support.
The neuromuscular system has non-linear and time-variant characteristics.
“the non-linear and time-variant characteristics of the neuromuscular system”
The paper relies on this without support.
A traumatic lesion of a peripheral nerve leads to Wallerian degeneration of the distal nerve.
“A traumatic lesion of a peripheral nerve leads to Wallerian degeneration of the distal nerve”
The paper relies on this without support.
Healing of a peripheral nerve lesion occurs only under favorable conditions involving sprouting of axons from the proximal stump, outgrowth along the distal stump, and reinnervation of target tissues.
“healing of such a lesion occurs only under favorable conditions involving sprouting of axons from the proximal nerve stump, outgrowth along the distal nerve stump, and reinnervation of target tissues”
The paper relies on this without support.
Recovery from a peripheral nerve lesion is usually delayed and limited if the lesion site is very proximal.
“Recovery is usually delayed and limited if the lesion site is very proximal.”
The paper relies on this without support.
Muscle atrophy prevents formation of new neuromuscular junctions at a later time when regenerating nerve fibers may reach the muscle.
“muscle atrophy prevents the formation of new neuromuscular junctions at a later time point when regenerating nerve fibers may reach the muscle”
The paper relies on this without support.
Regenerated nerve fibers have smaller than normal diameter and internodal length.
“the smaller than normal diameter and internodal length of regenerated fibers”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Motor and sensory nerve signals gathered using cuff electrodes have been used to guide and fine-tune prostheses such as artificial limbs in human patients.
“Although more challenging to carry out compared to nerve stimulation due to the distance between electrodes and the weak electrical signals transmitted by axons, with an ECM-rich epineurium separating the two, motor and sensory nerve signals gathered using cuff electrodes have been used to guide and fine-tune prostheses such as artificial limbs in human patients (Micera et al., 2001; Navarro et al”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005
The ENG-based control section (p. 22) reports that cuff electrodes recorded sensory signals used to regulate grasping in a hand neuroprosthesis, and that cuff-recorded signals predicted ankle movements in an animal model. It separately reports hand-prosthesis control in human amputees using LIFE electrodes, not cuffs. The paper therefore does not support the claim that cuff-recorded motor and sensory signals were used to guide artificial limbs in human patients. (LLM, conf=0.88)
Experimental — accuracy not yet measuredFor nerve stimulation at the clinical level, epineurial cuffs remain the preferred choice.
“For nerve stimulation at the clinical level, epineurial cuffs remain the preferred choice (Navarro et al., 2005).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005
The Cuff electrodes section (p. 10) says cuffs have been implanted in patients for decades and can stimulate nerves. But it does not call “epineurial cuffs” the preferred clinical choice; the Epineurial and helicoidal electrodes section (p. 9) instead describes button-type epineurial electrodes replacing cuffs in some applications to reduce injury risk. (LLM, conf=0.85)
Experimental — accuracy not yet measuredA thickening of the epineurium as part of FBR can dissipate the already weak electrical signals transmitted by axons, limiting the use of these devices for electrical recording applications in long-term implantations.
“A thickening of the epineurium as part of FBR can dissipate the already weak electrical signals transmitted by axons, limiting the use of these devices for electrical recording applications in long-term implantations (Navarro et al., 2005).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005
The Intrafascicular electrodes section (pp. 12–13) links fibrous encapsulation around an electrode to a gradual decrease in recorded axon-potential amplitude, supporting a recording-related effect of tissue growth. It does not identify thickening of the epineurium as the cause or say that this response limits long-term recording generally; the Cuff electrodes section (p. 11) reports successful long-term nerve-activity recording. (LLM, conf=0.84)
Experimental — accuracy not yet measuredBased on the degree of invasiveness, nerve neuroprosthetic designs can be categorized into three groups: epineurial cuff electrodes, intraneural penetrating electrodes, and regenerative nerve electrodes.
“Based on the degree of invasiveness nerve neuroprosthetic designs can be categorized into three different groups (summarized in Figures 4, 5).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005
The Abstract (p. 1) distinguishes extraneural cuff and epineurial electrodes from intrafascicular, penetrating, and regenerative electrodes; the Cuff electrodes section (p. 10) also compares their invasiveness. The paper does not present the asserted three-group classification: it treats cuff and epineurial electrodes as distinct designs and discusses intrafascicular and interfascicular electrodes separately (p. 12). (LLM, conf=0.84)
Experimental — accuracy not yet measuredMost commonly the nerve lesion for regenerative implants is a full nerve transection where the two ends are cleanly separated and a gap forms.
“Most commonly this is in the form of a full nerve transection, where the two ends of the nerve are cleanly separated and a gap forms between the two (Navarro et al., 2005; Thompson et al., 2015).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Navarro et al., 2005
The Regenerative electrodes section (p. 16) says these electrodes are implanted between severed nerve stumps and that multichannel regenerative electrodes can be applied only to transected nerves. It does not say that full transection is the most common lesion for regenerative implants, or specify that the ends are cleanly separated and a gap forms. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Existing peripheral nerve interfaces have limitations that reduce their clinical usefulness.
“However, current peripheral nerve interfaces have limitations, which minimize their clinical utility (1, 2).”
From A regenerative peripheral nerve interface allows real-time control of an artificial hand in upper limb amputees (2020) · cites this paper as [2]
An extensive, quantitative comparison of peripheral nerve electrode use in neuroprostheses in recent literature is absent.
“Many electrodes have been developed for peripheral nerves (Warren et al., 2016) and a comparison of their performance been conducted more than a decade ago (Navarro et al., 2005), but an extensive, quantitative comparison (on e.g., longevity and spatial resolution) of their use in neuroprostheses in recent literature is absent.”
From Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics (2018) · cites this paper as Navarro et al., 2005
Clinical practice of peripheral nerve stimulation for the control of neuroprosthesis is limited.
“However, clinical practice of peripheral nerve stimulation for the control of neuroprosthesis is limited (Navarro et al., 2005).”
From Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics (2018) · cites this paper as Navarro et al., 2005
It is hard to compare electrodes due to differences in impedance, spatial resolution, and inconsistent SNR results across studies.
“But even with this definition, it is hard to compare the electrodes due to the difference in impedance, spatial resolution (Navarro et al., 2005) and inconsistent SNR results across studies published so far.”
From Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics (2018) · cites this paper as Navarro et al., 2005
MNG can be used to record signals from nerves for neuroprosthetic control, but measurements in a magnetically shielded room restrict its use.
“It can also be used to record signals from nerves for the intent of neuroprosthetic control; however the fact that the measurements are done in a magnetically shielded room restricts its use [12].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [12]
Currently in clinical practice, implanted electrodes are generally extraneural.
“Data collected by Navarro et al. point to the fact that currently in clinical practice, implanted electrodes are generally extraneural [36].”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [36]
The electrode's degree of invasiveness is an important factor in the neural response.
“The electrode's degree of invasiveness is an important factor in the neural response [5].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [5]
Information on specifically interfacing the peripheral nervous system to control advanced prostheses is more sparse than information on related topics.
“However, the information on specifically interfacing the PNS to control these advanced prostheses is more sparse.”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [12]
The Introduction (p. 2) and the section on PNS-interface control (p. 20) discuss interfacing the PNS to control prostheses. The Electrical section (p. 2) says PNS interfaces are used most frequently for functional electrical stimulation, but it does not compare how much information is available on prosthesis control versus related topics. Searches for scarce information and limited research found no such comparison in our copy. (LLM, conf=0.73)
Experimental — accuracy not yet measuredState-of-the-art implants that target the PNS are mainly based on flexible materials.
“State-of-the-art implants that target the PNS are, therefore, mainly based on flexible materials (Navarro et al. 2005).”
From Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations (2018) · cites this paper as Navarro et al. 2005
Cites
The Handbook of brain theory and neural networks
(1996) · Choice Reviews Online · doi:10.5860/choice.33-4484 · Cited by 3,913 · see citing works
Muscle and tendon: properties, models, scaling, and application to biomechanics and motor control.
Zajac, F. E. (1989) · PubMed · Cited by 2,980 · see citing works
A new strategy for multifunction myoelectric control
Hudgins, B., Parker, P., & Scott, R. N. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.204774 · Cited by 2,149 · see citing works
A robust, real-time control scheme for multifunction myoelectric control
Englehart, K., & Hudgins, B. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.813539 · Cited by 1,743 · see citing works
The Biomedical Engineering Handbook
Bronzino, J. D., & Peterson, D. R. (2018) · CRC Press eBooks · doi:10.1201/b18423 · Cited by 1,605 · see citing works
(1991) · Medical Entomology and Zoology · Cited by 1,512 · see citing works
Somatosensory, proprioceptive, and sympathetic activity in human peripheral nerves
Vallbo, A. B., Hagbarth, K. E., Torebjork, H. E., & Wallin, B. G. (1979) · Physiological Reviews · doi:10.1152/physrev.1979.59.4.919 · Cited by 1,507 · see citing works
Wt, L., Holmquest, H. J., Scot, D., & Dow, M. (1961) · PubMed · Cited by 898 · see citing works
Analysis of a Model for Excitation of Myelinated Nerve
McNeal, D. R. (1976) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1976.324593 · Cited by 833 · see citing works
Connecting cortex to machines: recent advances in brain interfaces
Donoghue, J. P. (2002) · Nature Neuroscience · doi:10.1038/nn947 · Cited by 720 · see citing works
Nicolelis, M. A. L. (2001) · Nature · doi:10.1038/35053191 · Cited by 696 · see citing works
Topographical control of cell behaviour: II. multiple grooved substrata
Clark, P., Connolly, P., Curtis, A. S. G., Dow, J. A. T., & Wilkinson, C. D. W. (1990) · Development · doi:10.1242/dev.108.4.635 · Cited by 682 · see citing works
Control of Multifunctional Prosthetic Hands by Processing the Electromyographic Signal
Zecca, M., Micera, S., Carrozza, M. C., & Dario, P. (2002) · Critical Reviews? in Biomedical Engineering · doi:10.1615/critrevbiomedeng.v30.i456.80 · Cited by 606 · see citing works
Self-citationKuiken, T. A., Dumanian, G. A., Lipschutz, R. D., Miller, L. A., & Stubblefield, K. A. (2004) · Prosthetics & Orthotics International · doi:10.3109/03093640409167756 · Cited by 583 · see citing works
A miniature microelectrode array to monitor the bioelectric activity of cultured cells
THOMASJR, C., SPRINGER, P., Loeb, G. E., Berwald‐Netter, Y., & Okun, L. M. (1972) · Experimental Cell Research · doi:10.1016/0014-4827(72)90481-8 · Cited by 580 · see citing works
Control of Multifunctional Prosthetic Hands by Processing the Electromyographic Signal
Zecca, M., Micera, S., Carrozza, M. C., & Dario, P. (2017) · Critical Reviews in Biomedical Engineering · doi:10.1615/critrevbiomedeng.v45.i1-6.150 · Cited by 569 · see citing works
Self-citationA Neuron-Silicon Junction: A Retzius Cell of the Leech on an Insulated-Gate Field-Effect Transistor
Fromherz, P., Offenhäusser, A., Vetter, T. R., & Weis, J. (1991) · Science · doi:10.1126/science.1925540 · Cited by 563 · see citing works
Current trends in Graz brain-computer interface (BCI) research
Pfurtscheller, G., Neuper, C., Guger, C., Harkam, W., Ramoser, H., Schlögl, A., Obermaier, B., & Pregenzer, M. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.847821 · Cited by 518 · see citing works
EMG feature evaluation for movement control of upper extremity prostheses
Zardoshti-Kermani, M., Wheeler, B. C., Badie, K., & Hashemi, R. M. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481972 · Cited by 477 · see citing works
Ochoa, J., & Torebjörk, E. (1983) · The Journal of Physiology · doi:10.1113/jphysiol.1983.sp014873 · Cited by 467 · cited 2× · see citing works
“Microstimulation through the electrode can be used to activate single axons, although in most cases stimulation affects small groups of axons in the fascicle (Ochoa and Torebjörk, 1983; Torebjörk et al., 1987).”
BackgroundA myosignal-based powered exoskeleton system
Rosen, J., Brand, M., Fuchs, M. B., & Arcan, M. (2001) · IEEE Transactions on Systems, Man, and Cybernetics - Part A: Systems and Humans · doi:10.1109/3468.925661 · Cited by 437 · see citing works
Wilkinson, D. W. (1987) · doi:10.1242/dev.99.3.439 · Cited by 418 · see citing works
A neural interface for a cortical vision prosthesis
Normann, R. A., Maynard, E. M., Rousche, P. J., & Warren, D. J. (1999) · Vision Research · doi:10.1016/s0042-6989(99)00040-1 · Cited by 411 · see citing works
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · cited 3× · see citing works
“A design variation of the cuff electrode is the flatinterface nerve electrode (FINE) developed by Durand and coworkers (Tyler and Durand, 2002).”
BackgroundMethodIn vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes
Weiland, J. D., Anderson, D. J., & Humayun, M. S. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.805487 · Cited by 382 · see citing works
A three-dimensional microelectrode array for chronic neural recording
Hoogerwerf, A. C., & Wise, K. D. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.335862 · Cited by 372 · see citing works
Kipke, D. R., Vetter, R. J., Williams, J. C., & Hetke, J. F. (2003) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.814443 · Cited by 364 · see citing works
Drake, K. L., Wise, K. D., Farraye, J., Anderson, D. J., & BeMent, S. L. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7273 · Cited by 364 · see citing works
Single unit recording capabilities of a 100 microelectrode array
Nordhausen, C. T., Maynard, E. M., & Normann, R. A. (1996) · Brain Research · doi:10.1016/0006-8993(96)00321-6 · Cited by 358 · see citing works
Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 358 · cited 6× · see citing works
“…of selectively recording single-unit responses from mechanoreceptors, although only in 10–20% of the electrodes, and evoked graded recruitment of force in muscle groups in a highly selective fashion with current injections in the 1–20 mA range (Branner and Norman, 2000; Branner et al., 2001).”
BackgroundInfluentialFunctional Separation of EMG Signals via ARMA Identification Methods for Prosthesis Control Purposes
Graupe, D., & Cline, W. K. (1975) · IEEE Transactions on Systems, Man, and Cybernetics · doi:10.1109/tsmc.1975.5408479 · Cited by 354 · see citing works
Dhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 2× · see citing works
“Recently, LIFE electrodes have been used to control a hand prosthesis using the ENG signals recorded from peripheral nerves of amputee subjects (Dhillon et al., 2004).”
MethodMicromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343 · see citing works
Self-citationSelective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
The first 500 patients with sacral anterior root stimulator implants: general description
Brindley, G. S. (1994) · Spinal Cord · doi:10.1038/sc.1994.126 · Cited by 318 · see citing works
Neural prostheses : fundamental studies
Agnew, W. F., & McCreery, D. B. (1990) · Medical Entomology and Zoology · Cited by 316 · see citing works
Sacral anterior root stimulators for bladder control in paraplegia: the first 50 cases
Brindley, G. S., Polkey, C. E., Rushton, D. N., & Cardozo, L. (1986) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp.49.10.1104 · Cited by 313 · see citing works
The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation
Gorman, P. H., & Mortimer, J. T. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325041 · Cited by 313 · see citing works
An implantable multielectrode array with on-chip signal processing
Najafi, K., & Wise, K. D. (1986) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1986.1052646 · Cited by 310 · see citing works
A high-yield IC-compatible multichannel recording array
Najafi, K., Wise, K. D., & Mochizuki, T. (1985) · IEEE Transactions on Electron Devices · doi:10.1109/t-ed.1985.22102 · Cited by 303 · see citing works
Multichannel surface EMG: Basic aspects and clinical utility
Zwarts, M. J., & Stegeman, D. F. (2003) · Muscle & Nerve · doi:10.1002/mus.10358 · Cited by 302 · see citing works
Polyimides as biomaterials: preliminary biocompatibility testing
Richardson, R. R., Miller, J. A., & Reichert, W. M. (1993) · Biomaterials · doi:10.1016/0142-9612(93)90183-3 · Cited by 302 · see citing works
Long-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve
Branner, A., Stein, R. B., Fernandez, E., Aoyagi, Y., & Normann, R. A. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.820321 · Cited by 301 · see citing works
Wilkinson, D. G., Bailes, J. A., Champion, J. E., & McMahon, A. P. (1987) · Development · doi:10.1242/dev.99.4.493 · Cited by 293 · see citing works
Selective Electrical Interfaces with the Nervous System
Rutten, W. L. C. (2002) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.4.020702.153427 · Cited by 276 · see citing works
Clinical use of the odstock dropped foot stimulator: Its effect on the speed and effort of walking
Taylor, P. N., Burridge, J. H., Dunkerley, A. L., Wood, D. E., Norton, J. A., Singleton, C., & Swain, I. D. (1999) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(99)90333-7 · Cited by 275 · see citing works
Long-term results of peripheral nerve stimulation for reflex sympathetic dystrophy
Hassenbusch, S. J., Stanton-Hicks, M., Schoppa, D., Walsh, J. G., & Covington, E. C. (1996) · Journal of Neurosurgery · doi:10.3171/jns.1996.84.3.0415 · Cited by 254 · cited 1× · see citing works
“The latter application is of limited benefit because it should be applied only when pain involves one nerve territory in the head or the limbs and has frequent morbidity associated with electrode approximation, electrode displacement, and lead failure (Hassenbusch et al., 1996).”
BackgroundBION <sup>™</sup> system for distributed neural prosthetic interfaces
Loeb, G. E., Peck, R. A., Moore, W. H., & Hood, K. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00011-x · Cited by 243 · see citing works
New Trends in Neuromodulation for the Management of Neuropathic Pain
Aló, K., & Holsheimer, J. (2002) · Neurosurgery · doi:10.1097/00006123-200204000-00003 · Cited by 241 · see citing works
Criteria for the Selection of Materials for Implanted Electrodes
Geddes, L. A., & Roeder, R. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1581292 · Cited by 240 · see citing works
The application of neural networks to myoelectric signal analysis: a preliminary study
Kelly, M. F., Parker, P. A., & Scott, R. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.52324 · Cited by 237 · see citing works
A multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats
Branner, A., & Normann, R. A. (2000) · Brain Research Bulletin · doi:10.1016/s0361-9230(99)00231-2 · Cited by 234 · see citing works
3D flexible multichannel neural probe array
Takeuchi, S., Suzuki, T. J., Mabuchi, K., & Fujita, H. (2003) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/14/1/014 · Cited by 221 · see citing works
Modulation of Muscle Force by Recruitment During Intramuscular Stimulation
Crago, P. E., Peckham, P. H., & Thrope, G. B. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326592 · Cited by 216 · see citing works
The effect of stimulus pulse duration on selectivity of neural stimulation
Grill, W. M., & Mortimer, J. T. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.481985 · Cited by 215 · see citing works
Cuff electrodes for chronic stimulation and recording of peripheral nerve activity
Loeb, G. E., & Peck, R. A. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00123-9 · Cited by 215 · see citing works
Selective stimulation of peripheral nerve fibers using dual intrafascicular electrodes
Yoshida, K., & Horch, K. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.243412 · Cited by 208 · see citing works
Acoustic myography: A noninvasive monitor of motor unit fatigue
Barry, D. T., Geiringer, S. R., & Ball, R. D. (1985) · Muscle & Nerve · doi:10.1002/mus.880080303 · Cited by 208 · see citing works
Polyimide cuff electrodes for peripheral nerve stimulation
Rodri´guez, F. J., Ceballos, D., Schu¨ttler, M., Valero, A., Valderrama, E., Stieglitz, T., & Navarro, X. (2000) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(00)00192-8 · Cited by 207 · see citing works
Self-citationThe application of implant technology for cybernetic systems
Warwick, K., Gasson, M., Hutt, B., Goodhew, I., Kyberd, P., Andrews, B., Teddy, P., & Shad, A. (2003) · Archives of Neurology · doi:10.1001/archneur.60.10.1369 · Cited by 206 · see citing works
Graupe, D., Salahi, J., & Kohn, K. H. (1982) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(82)90021-8 · Cited by 205 · cited 1× · see citing works
“From the first works by Graupe and colleagues (Graupe and Cline, 1975; 1982; Graupe et al., 1982), the EMG signal was modeled as the amplitude-modulated Gaussian noise whose variance is related to the force developed by the muscle.”
BackgroundBrummer, S. B., Robblee, L. S., & Hambrecht, F. T. (1983) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1983.tb31628.x · Cited by 205 · see citing works
A nerve cuff technique for selective excitation of peripheral nerve trunk regions
Sweeney, J. D., Ksienski, D. A., & Mortimer, J. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55681 · Cited by 196 · see citing works
An Implanted Upper-Extremity Neuroprosthesis. Follow-up of Five Patients*
KILGORE, K. L., PECKHAM, P. H., KEITH, M. W., THROPE, G. B., WUOLLE, K. S., BRYDEN, A. M., & HART, R. L. (1997) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199704000-00008 · Cited by 192 · see citing works
Geometrical factors in surface EMG of the vastus medialis and lateralis muscles
Rainoldi, A., Nazzaro, M., Merletti, R., Farina, D., Caruso, I., & Gaudenti, S. (2000) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(00)00024-9 · Cited by 192 · cited 2× · see citing works
“However, the use of EMG for this purpose under dynamic conditions (as in the case of exoskeleton control) presents several problems because of muscle dynamics and possible prostheses (Rainoldi et al., 2000).”
BackgroundJimbo, Y., Robinson, H. P. C., & Kawana, A. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.238465 · Cited by 192 · cited 1× · see citing works
“Having surface topologies on the substrate, axonal outgrowth of embryonic nerve cells from rats was observed and electrical activity was recorded in a microstructure with 16 electrodes that were connected with grooves (Jimbo and Kawana, 1992; Jimbo et al., 1993).”
MethodAn exoskeletal robot for human shoulder joint motion assist
Kiguchi, K., Iwami, K., Yasuda, M., Watanabe, K., & Fukuda, T. (2003) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2003.809168 · Cited by 191 · see citing works
Rattay, F. (1990) · doi:10.1007/978-3-7091-3271-5 · Cited by 190 · see citing works
Unsupervised pattern recognition for the classification of EMG signals
Christodoulou, C. I., & Pattichis, C. S. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.740879 · Cited by 186 · see citing works
Agnew, W. F., McCreery, D. B., Yuen, T. G. H., & Bullara, L. A. (1989) · Annals of Biomedical Engineering · doi:10.1007/bf02364272 · Cited by 185 · see citing works
INTRANEURAL MICROSTIMULATION IN MAN: ITS RELATION TO SPECIFICITY OF TACTILE SENSATIONS
TOREBJÖRK, H. E., VALLBO, Å. B., & OCHOA, J. L. (1987) · Brain · doi:10.1093/brain/110.6.1509 · Cited by 184 · see citing works
Long-term pain control by direct peripheral-nerve stimulation.
Nashold, B. S., Goldner, J. L., Mullen, J. B., & Bright, D. S. (1982) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198264010-00001 · Cited by 184 · cited 2× · see citing works
“, 1997), for reducing pain (Nashold et al., 1982; Strege et al., 1994; StantonHicks and Salamon, 1997), for phrenic nerve pacing for ventilatory assistance (Glenn et al.”
BackgroundLago, N., Ceballos, D., J Rodrı́guez, F., Stieglitz, T., & Navarro, X. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.06.025 · Cited by 178 · see citing works
Self-citationEdell, D. J. (1986) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1986.325892 · Cited by 177 · see citing works
Silicon ribbon cables for chronically implantable microelectrode arrays
Hetke, J. F., Lund, J. L., Najafi, K., Wise, K. D., & Anderson, D. J. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.284959 · Cited by 174 · see citing works
Chronically implanted intrafascicular recording electrodes
Lefurge, T., Goodall, E., Horch, K., Stensaas, L., & Schoenberg, A. (1991) · Annals of Biomedical Engineering · doi:10.1007/bf02368469 · Cited by 172 · see citing works
The Biomedical Engineering Handbook
von Maltzahn, W. W. (1999) · The electrical engineering handbook/Electrical engineering handbook series/Electrical engineering handbook · doi:10.1201/9781420049510.sec8 · Cited by 172 · see citing works
Fisher, R. S., & Handforth, A. (1999) · Neurology · doi:10.1212/wnl.53.4.666 · Cited by 169 · see citing works
Cutaneous whole nerve recordings used for correction of footdrop in hemiplegic man
Haugland, M. K., & Sinkjaer, T. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481970 · Cited by 168 · see citing works
Regeneration microelectrode array for peripheral nerve recording and stimulation
Kovacs, G. T. A., Storment, C. W., & Rosen, J. M. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.256422 · Cited by 164 · cited 2× · see citing works
“Regenerative electrodes are designed to interface a high number of nerve fibers by using an array of holes, with electrodes built around them, implanted between the severed stumps of a peripheral nerve (Llinás et al., 1973; Edell, 1986; Kovacs et al., 1992; Dario et al., 1998).”
BackgroundNeural prostheses: Replacing motor function after disease or disability
Hilton‐Jones, D. (1993) · Neuromuscular Disorders · doi:10.1016/0960-8966(93)90069-v · Cited by 164 · see citing works
Kovacs, G. T. A., Storment, C. W., Halks-Miller, M., Belczynski, C. R., Santina, C. C. D., Lewis, E. R., & Maluf, N. I. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.293244 · Cited by 161 · see citing works
Stable long-term recordings from cat peripheral nerves
Stein, R. B., Nichols, T. R., Jhamandas, J., Davis, L., & Charles, D. (1977) · Brain Research · doi:10.1016/0006-8993(77)90233-5 · Cited by 160 · see citing works
Sensory nerve recording for closed‐loop control to restore motor functions
Popovic, D. B., Stein, R. B., Jovanovic, K. L., Dai, R., Kostov, A., & Armstrong, W. W. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.247801 · Cited by 159 · see citing works
An exoskeletal robot for human elbow motion support‐sensor fusion, adaptation, and control
Kiguchi, K., Kariya, S., Watanabe, K., Izumi, K., & Fukuda, T. (2001) · IEEE Transactions on Systems, Man and Cybernetics, Part B (Cybernetics) · doi:10.1109/3477.931520 · Cited by 159 · see citing works
Waters, R. L., McNeal, D. R., Faloon, W., & Clifford, B. (1985) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198567050-00016 · Cited by 156 · cited 4× · see citing works
“…for ventilatory assistance (Glenn et al., 1986; Creasey et al., 1996; Chervin and Guilleminault, 1997), for activation of lower extremity motion (Waters et al., 1985; Popovic, 1992; Haugland and Sinkjaer, 1995; Triolo et al., 1996; Graupe and Kohn, 1997; Popovic et al., 1998; Taylor et al.,…”
BackgroundInfluentialKobetic, R., Sharma, M., Triolo, R. J., Uhlir, J. P., Bieri, C., Wibowo, M., Polando, G., Marsolais, E. B., Davis, J. A., & Ferguson, K. A. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.808942 · Cited by 155 · see citing works
Kennedy, P. R. (1989) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(89)90142-8 · Cited by 151 · see citing works
Biostability of micro‐photodiode arrays for subretinal implantation
Hämmerle, H., Kobuch, K., Kohler, K., Nisch, W., Sachs, H., & Stelzle, M. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00185-5 · Cited by 146 · see citing works
Strength characterization of silicon microprobes in neurophysiological tissues
Najafi, K., & Hetke, J. F. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55638 · Cited by 146 · see citing works
Chronic Response of the Rat Sciatic Nerve to the Flat Interface Nerve Electrode
Tyler, D. J., & Durand, D. M. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1569263 · Cited by 144 · cited 1× · see citing works
“Electrodes that moderately flattened the nerve demonstrated the best selectivity for limb motion measurements, which were maintained throughout the implant time (Tyler and Durand, 2003; Leventhal and Durand, 2004).”
BackgroundA micromachined silicon sieve electrode for nerve regeneration applications
Akin, T., Najafi, K., Smoke, R. H., & Bradley, R. M. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.284958 · Cited by 144 · see citing works
(1986) · PubMed · Cited by 137 · see citing works
Quantification of recruitment properties of multiple contact cuff electrodes
Grill, W. M., & Mortimer, J. T. (1996) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.506402 · Cited by 136 · see citing works
Recruitment properties of monopolar and bipolar epimysial electrodes
Grandjean, P. A., & Mortimer, J. T. (1986) · Annals of Biomedical Engineering · doi:10.1007/bf02364648 · Cited by 134 · see citing works
Neural Networks: Concepts, Applications, and Implementations
Antognetti, P., & Milutinović, V. M. (1991) · Medical Entomology and Zoology · Cited by 132 · see citing works
Three-dimensional microfabrication using synchrotron radiation
Ehrfeld, W., & Münchmeyer, D. (1991) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/0168-9002(91)90289-3 · Cited by 127 · see citing works
Electrical stimulation and recording from cultured neurons using a planar electrode array
Jimbo, Y., & Kawana, A. (1992) · Bioelectrochemistry and Bioenergetics · doi:10.1016/0302-4598(92)80067-q · Cited by 126 · cited 1× · see citing works
“Having surface topologies on the substrate, axonal outgrowth of embryonic nerve cells from rats was observed and electrical activity was recorded in a microstructure with 16 electrodes that were connected with grooves (Jimbo and Kawana, 1992; Jimbo et al., 1993).”
MethodImplantable bioelectronic interfaces for lost nerve functions
Heiduschka, P., & Thanos, S. (1998) · Progress in Neurobiology · doi:10.1016/s0301-0082(98)00013-6 · Cited by 126 · see citing works
Stieglitz, T., Beutel, H., & Meyer, J.-U. (1997) · Sensors and Actuators A: Physical · doi:10.1016/s0924-4247(97)01494-5 · Cited by 125 · see citing works
Self-citationA method to effect physiological recruitment order in electrically activated muscle
Fang, Z.-P., & Mortimer, J. T. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.76384 · Cited by 125 · see citing works
Twenty Years of Experience in Phrenic Nerve Stimulation to Pace the Diaphragm
GLENN, W. W. L., PHELPS, M. L., ELEFTERLEADES, J. A., DENTZ, B., & HOGAN, J. F. (1986) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1986.tb06627.x · Cited by 125 · see citing works
Intramuscular electrical stimulation: Tissue damage
Mortimer, J. T., Kaufman, D., & Roessmann, U. (1980) · Annals of Biomedical Engineering · doi:10.1007/bf02364479 · Cited by 123 · see citing works
Skin contact force information in sensory nerve signals recorded by implanted cuff electrodes
Haugland, M. K., Hoffer, J. A., & Sinkjær, T. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296346 · Cited by 122 · see citing works
Neural Prostheses for Restoration of Sensory and Motor Function
Chapin, J. K., & Moxon, K. A. (2000) · CRC Press eBooks · doi:10.1201/9781420039054 · Cited by 122 · see citing works
Selectivity of multiple-contact nerve cuff electrodes: a simulation analysis
Choi, A. Q., Cavanaugh, J. K., & Durand, D. M. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.909637 · Cited by 121 · see citing works
Applications of cortical signals to neuroprosthetic control: a critical review
Lauer, R. T., Peckham, P. H., Kilgore, K. L., & Heetderks, W. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.847817 · Cited by 120 · see citing works
Biopotentials as command and feedback signals in functional electrical stimulation systems.
Sinkjær, T., Haugland, M., Inmann, A., Hansen, M., & Nielsen, K. D. (2003) · Medical Engineering and Physics · doi:10.1016/s1350-4533(02)00178-9 · Cited by 118 · see citing works
Stimulation of the Central and Peripheral Nervous System for the Control of Pain
Stanton-Hicks, M., & Salamon, J. (1997) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-199701000-00004 · Cited by 115 · see citing works
A micromachined silicon depth probe for multichannel neural recording
Yoon, T. H., Hwang, E. J., Shin, D. Y., Park, S. I., Oh, S., Jung, S. C., Shin, H. C., & Kim, S. J. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.855936 · Cited by 110 · see citing works
Droge, M., Gross, G., Hightower, M., & Czisny, L. (1986) · The Journal of Neuroscience · doi:10.1523/jneurosci.06-06-01583.1986 · Cited by 109 · cited 1× · see citing works
“The activity of embryonic spinal cord cells of mice has been recorded simultaneously over a period of several months on a planar glass substrate with 64 iridium tin oxide electrodes (Droge et al., 1986).”
MethodA slowly penetrating interfascicular nerve electrode for selective activation of peripheral nerves
Tyler, D. J., & Durand, D. M. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559349 · Cited by 108 · cited 1× · see citing works
“Histological cross-sections showed that the SPINE rearranges the epineurium and penetrates deep within a multifascicular nerve but did not disrupt the perineurium (Tyler and Durand, 1997).”
BackgroundAn intrafascicular electrode for recording of action potentials in peripheral nerves
Malagodi, M. S., Horch, K. W., & Schoenberg, A. A. (1989) · Annals of Biomedical Engineering · doi:10.1007/bf02368058 · Cited by 107 · see citing works
(1998) · PubMed · Cited by 107 · see citing works
Self-citationLawrence, S. M., Dhillon, G. S., Jensen, W., Yoshida, K., & Horch, K. W. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.831491 · Cited by 106 · cited 1× · see citing works
“The use of multistrand Kevlar LIFEs (m-polyLIFE) increases the tensile strength and signal-to-noise ratio (Lawrence et al., 2004).”
MethodOnishi, H., Yagi, R., Akasaka, K., Momose, K., Ihashi, K., & Handa, Y. (2000) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(99)00020-6 · Cited by 105 · see citing works
Time-scale analysis of motor unit action potentials
Pattichis, C. S., & Pattichis, M. S. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.797992 · Cited by 105 · see citing works
A 64-site multishank CMOS low-profile neural stimulating probe
Kim, C., & Wise, K. D. (1996) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.535406 · Cited by 103 · see citing works
Grill, W. M., & Mortimer, J. T. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.736150 · Cited by 103 · see citing works
The relationship between electromyogram and muscle force
Hof, A. L. (1997) · Sportverletzung · Sportschaden · doi:10.1055/s-2007-993372 · Cited by 103 · cited 1× · see citing works
“This approach is supported by biomechanical studies (Hof, 1997; Onishi et al., 2000) showing that under isometric contractions the relationship between force and EMG can be considered linear.”
Jezernik, S., Craggs, M., Grill, W. M., Creasey, G., & Rijkhoff, N. J. M. (2002) · Neurological Research · doi:10.1179/016164102101200294 · Cited by 102 · see citing works
Triolo, R. J., Bieri, C., Uhlir, J., Kobetic, R., Scheiner, A., & Marsolais, E. B. (1996) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(96)90133-1 · Cited by 102 · see citing works
Procházka, A., Mushahwar, V. K., & McCreery, D. B. (2001) · The Journal of Physiology · doi:10.1111/j.1469-7793.2001.0099b.x · Cited by 101 · see citing works
Kern, H., Hofer, C., Mödlin, M., Forstner, C., Raschka‐Högler, D., Mayr, W., & Stöhr, H. (2002) · Artificial Organs · doi:10.1046/j.1525-1594.2002.06933.x · Cited by 100 · see citing works
Hewson, D. J., Hogrel, J.-Y., Langeron, Y., & Duchêne, J. (2003) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(02)00097-4 · Cited by 99 · cited 1× · see citing works
“The impedance at the electrode–skin interface is largely variable between individuals at low frequencies, whereas at high frequencies the decline of impedance values with time depends on the electrode type (Hewson et al., 2003).”
BackgroundRiso, R. R. (1999) · Technology and Health Care · doi:10.3233/thc-1999-7604 · Cited by 98 · see citing works
Development of percutaneous intramuscular electrode for multichannel FES system
Handa, Y., Hoshimiya, N., Iguchi, Y., et al. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32102 · Cited by 98 · see citing works
Larsen, J. O., Thomsen, M., Haugland, M., & Sinkjær, T. (1998) · Acta Neuropathologica · doi:10.1007/s004010050907 · Cited by 96 · see citing works
Improving detection of muscle activation intervals
Micera, S., Vannozzi, G., Sabatini, A. M., & Dario, P. (2001) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.982274 · Cited by 95 · see citing works
Self-citationAn algorithm for detecting the onset of muscle contraction by EMG signal processing
Micera, S., Sabatini, A. M., & Dario, P. (1998) · Medical Engineering & Physics · doi:10.1016/s1350-4533(98)00017-4 · Cited by 95 · see citing works
Self-citationMuscle recruitment with intrafascicular electrodes
Nannini, N., & Horch, K. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83589 · Cited by 94 · see citing works
Fabrication and characteristics of an implantable, polymer-based, intrafascicular electrode
Lawrence, S. M., Dhillon, G. S., & Horch, K. W. (2003) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(03)00231-0 · Cited by 93 · see citing works
Plasma‐deposited passivation layers for moisture and water protection
Vogt, M., & Hauptmann, R. (1995) · Surface and Coatings Technology · doi:10.1016/0257-8972(95)08268-9 · Cited by 92 · cited 1× · see citing works
“Better stability was achieved with multilayers (Cogan et al., 1993; Vogt and Hauptmann, 1995).”
Implanted stimulators for restoration of function in spinal cord injury
Bhadra, N., Kilgore, K. L., & Peckham, P. H. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00012-1 · Cited by 92 · see citing works
Micera, S., Sabatini, A. M., Dario, P., & Rossi, B. (1999) · Medical Engineering & Physics · doi:10.1016/s1350-4533(99)00055-7 · Cited by 90 · see citing works
Self-citationJanders, M., Egert, U., Stelzle, M., & Nisch, W. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1996.656936 · Cited by 90 · see citing works
Vagus nerve stimulation for intractable epilepsy: a review
McLachlan, R. S. (1997) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-199709000-00002 · Cited by 89 · see citing works
Sparto, P. J., Parnianpour, M., Barría, E. A., et al. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.867887 · Cited by 86 · see citing works
Durability of implanted electrodes and leads in an upper-limb neuroprosthesis
Kilgore, K. L., Peckham, P. H., Keith, M. W., Montague, F. W., Hart, R. L., Gazdik, M. M., Bryden, A. M., Snyder, S. A., & Stage, T. G. (2003) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2003.11.0457 · Cited by 85 · see citing works
Mannard, A., Stein, R. B., & Charles, D. (1974) · Science · doi:10.1126/science.183.4124.547 · Cited by 85 · see citing works
Subfascicle stimulation selectivity with the flat interface nerve electrode
Leventhal, D. K., & Durand, D. M. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1569266 · Cited by 81 · see citing works
Electrical stimulation of sacral roots for micturition after spinal cord injury
Creasey, G. H. (1993) · Urologic Clinics of North America · doi:10.1016/s0094-0143(21)00511-5 · Cited by 80 · see citing works
Worldwide experience with the Finetech‐Brindley sacral anterior root stimulator
Van Kerrebroeck, P. E. V., Koldewijn, E. L., & Debruyne, F. M. J. (1993) · Neurourology and Urodynamics · doi:10.1002/nau.1930120511 · Cited by 80 · see citing works
Haugland, M. K., & Hoffer, J. A. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296345 · Cited by 80 · see citing works
(1996) · PubMed · Cited by 79 · see citing works
Stieglitz, T., Ruf, H. H., Gross, M., Schüettler, M., & Meyer, J.-U. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00019-2 · Cited by 77 · see citing works
Self-citationCharacterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes
Yoshida, K., & Stein, R. B. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.740885 · Cited by 76 · see citing works
Chronic recording of regenerating viiith nerve axons with a sieve electrode
Mensinger, A. F., Anderson, D. J., Buchko, C. J., Johnson, M. A., Martin, D. C., Tresco, P. A., Silver, R. B., & Highstein, S. M. (2000) · Journal of Neurophysiology · doi:10.1152/jn.2000.83.1.611 · Cited by 76 · see citing works
Metallized polymer fibers as leadwires and intrafascicular microelectrodes
McNaughton, T. G., & Horch, K. W. (1996) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(96)00111-2 · Cited by 76 · see citing works
Upshaw, B., & Sinkjaer, T. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.681183 · Cited by 74 · see citing works
Chae, J., & Hart, R. (1998) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/00002060-199811000-00013 · Cited by 72 · see citing works
Satisfaction with and usage of a hand neuroprosthesis
Wuolle, K. S., Van Doren, C. L., Bryden, A. M., Hunter Peckham, P., Keith, M. W., Kilgore, K. L., & Grill, J. H. (1999) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(99)90123-5 · Cited by 70 · see citing works
Long term chronic recordings from peripheral sensory fibers using a sieve electrode array
Bradley, R. M., Cao, X., Akin, T., & Najafi, K. (1997) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(97)02225-5 · Cited by 69 · see citing works
Retina implant—a bioMEMS challenge
Meyer, J.-U. (2002) · Sensors and Actuators A Physical · doi:10.1016/s0924-4247(01)00807-x · Cited by 69 · see citing works
(1998) · Library and Archives Canada (Government of Canada) · Cited by 68 · see citing works
Aoyagi, Y., Stein, R. B., Branner, A., Pearson, K. G., & Normann, R. A. (2003) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(03)00143-2 · Cited by 68 · see citing works
Rat sciatic nerve regeneration through a micromachined silicon chip
Zhao, Q., Drott, J., Laurell, T., Wallman, L., Lindström, K., Bjursten, L. M., Lundborg, G., Montelius, L., & Danielsen, N. (1997) · Biomaterials · doi:10.1016/s0142-9612(96)00112-3 · Cited by 67 · see citing works
Neuroprostheses to treat neurogenic bladder dysfunction: current status and future perspectives
Rijkhoff, N. J. M. (2004) · Child's Nervous System · doi:10.1007/s00381-003-0859-1 · Cited by 66 · see citing works
First clinical experience with BION implants for therapeutic electrical stimulation
Salter, A.-C. D., Bagg, S. D., Creasy, J. L., Romano, C., Romano, D., Richmond, F. J. R., & Loeb, G. E. (2004) · Neuromodulation: Technology at the Neural Interface · doi:10.1111/j.1525-1403.2004.04005.x · Cited by 66 · see citing works
Krarup, C., Loeb, G. E., & Pezeshkpour, G. H. (1988) · Muscle & Nerve · doi:10.1002/mus.880110906 · Cited by 66 · see citing works
Stieglitz, T., & Gross, M. (2002) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(01)01021-8 · Cited by 65 · see citing works
Self-citationControl of neural prostheses for grasping and reaching
Popović, M. B. (2002) · Medical Engineering & Physics · doi:10.1016/s1350-4533(02)00187-x · Cited by 64 · see citing works
Klinge, P. M., Vafa, M. A., Brinker, T., Brandis, A., Walter, G. F., Stieglitz, T., Samii, M., & Wewetzer, K. (2001) · Biomaterials · doi:10.1016/s0142-9612(00)00420-8 · Cited by 64 · see citing works
Self-citationThe frequency response of frog muscle spindles under various conditions
Kirkwood, P. A. (1972) · The Journal of Physiology · doi:10.1113/jphysiol.1972.sp009791 · Cited by 63 · see citing works
González, C., & Rodrı́guez, M. (1997) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(96)02202-9 · Cited by 63 · see citing works
Information contained in sensory nerve recordings made with intrafascicular electrodes
Goodall, E. V., Lefurge, T. M., & Horch, K. W. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83604 · Cited by 63 · see citing works
Extradural implantation of sacral anterior root stimulators.
Sauerwein, D., Ingunza, W., Fischer, J., Madersbacher, H., Polkey, C. E., Brindley, G. S., Colombel, P., & Teddy, P. (1990) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp.53.8.681 · Cited by 62 · see citing works
Multielectrode intrafascicular and extraneural stimulation
Veltink, P. H., van Alsté, J. A., & Boom, H. B. K. (1989) · Medical and Biological Engineering and Computing · doi:10.1007/bf02442165 · Cited by 62 · see citing works
Muscle fatigue measured with evoked muscle vibrations
Barry, D. T., Hill, T., & Im, D. (1992) · Muscle & Nerve · doi:10.1002/mus.880150308 · Cited by 62 · cited 1× · see citing works
“Acoustic signals increase in parallel with surface EMG signals during voluntary and evoked contraction, but acoustic amplitude decays with fatigue whereas EMG amplitude does not (Barry et al., 1985; 1992) .”
BackgroundCeballos, D., Valero‐Cabré, A., Valderrama, E., Schüttler, M., Stieglitz, T., & Navarro, X. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10099 · Cited by 61 · see citing works
Self-citationEmptying the bladder by stimulating sacral ventral roots.
Gs, B. (1974) · PubMed · Cited by 61 · see citing works
Selective Detrusor Activation By Electrical Sacral Nerve Root Stimulation in Spinal Cord Injury
Rijkhoff, N., Wijkstra, H., Kerrebroeck, P. V., & Debruyne, F. M. J. (1997) · The Journal of Urology · doi:10.1016/s0022-5347(01)65032-8 · Cited by 60 · see citing works
Chronic peripheral nerve pain treated with direct electrical nerve stimulation
Strege, D. W., Cooney, W. P., Wood, M. B., Johnson, S. J., & Metcalf, B. J. (1994) · The Journal of Hand Surgery · doi:10.1016/0363-5023(94)90092-2 · Cited by 60 · see citing works
Embryonic Cord Transplants in Peripheral Nerve Restore Skeletal Muscle Function
Thomas, C. K., Erb, D. E., Grumbles, R. M., & Bunge, R. P. (2000) · Journal of Neurophysiology · doi:10.1152/jn.2000.84.1.591 · Cited by 60 · see citing works
Neural interfaces for regenerated nerve stimulation and recording
Dario, P., Garzella, P., Toro, M., Micera, S., Alavi, M., Meyer, U., Valderrama, E., Sebatiani, L., Ghelarducci, B., Mazzoni, C., et al. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.736149 · Cited by 58 · see citing works
Self-citationMicera, S., Jensen, W., Sepulveda, F., Riso, R. R., & Sinkjaer, T. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930903 · Cited by 58 · see citing works
Diaphragm Pacing for Respiratory Insufficiency
Chervin, R. D., & Guilleminault, C. (1997) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-199709000-00003 · Cited by 56 · see citing works
Arrays for chronic functional microstimulation of the lumbosacral spinal cord
McCreery, D. B., Pikov, V., Lossinsky, A. S., Bullara, L. A., & Agnew, W. F. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.827223 · Cited by 56 · see citing works
Rupp, R., & Gerner, H. J. (2007) · Springer eBooks · doi:10.1007/978-3-211-33079-1_55 · Cited by 56 · see citing works
Long‐term biocompatibility of implanted polymer‐based intrafascicular electrodes
Lawrence, S. M., Larsen, J. O., Horch, K. W., Riso, R., & Sinkjær, T. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10303 · Cited by 55 · see citing works
Modeling study of peripheral nerve recording selectivity
Perez-Perive, J., & Durund, D. M. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.867874 · Cited by 55 · see citing works
Negredo, P., Castro, J. E., Lago, N., Navarro, X., & Avendaño, C. (2004) · Neuroscience · doi:10.1016/j.neuroscience.2004.07.017 · Cited by 55 · see citing works
Self-citationChronic Measurement of the Stimulation Selectivity of the Flat Interface Nerve Electrode
Leventhal, D. K., & Durand, D. M. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.827535 · Cited by 55 · see citing works
Separation of action potentials in multiunit intrafascicular recordings
Goodall, E. V., & Horch, K. W. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.125014 · Cited by 53 · see citing works
Long-term biocompatibility of a miniature stimulator implanted in feline hind limb muscles
Cameron, T., Liinamaa, T. L., Loeb, G. E., & Richmond, F. J. R. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.704871 · Cited by 53 · see citing works
(2001) · PubMed · Cited by 52 · see citing works
Self-citationMotor prosthesis for restoration of upper extremity function
Peckham, P. H. (1992) · Medical Entomology and Zoology · Cited by 51 · see citing works
EMG prosthetic hand controller discriminating ten motions using real-time learning method
Nishikawa, D., Yu, W., Yokoi, H., & Kakazu, Y. (2003) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.1999.811706 · Cited by 50 · see citing works
Malmstrom, J. A., McNaughton, T. G., & Horch, K. W. (1998) · Annals of Biomedical Engineering · doi:10.1114/1.35 · Cited by 50 · cited 1× · see citing works
“Activity arising from cutaneous receptors in response to stimuli has been recorded in several studies, although activity was found to be small and difficult to distinguish from the background noise in the recording (Malmstrom et al., 1998; Yoshida et al., 2000).”
Background(2000) · PubMed · Cited by 50 · see citing works
Intrafascicular electrodes for stimulation and recording from mudpuppy spinal roots
Yoshida, K., Jovanović, K., & Stein, R. B. (2000) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(99)00176-4 · Cited by 49 · see citing works
Wallman, L. (2001) · Biomaterials · doi:10.1016/s0142-9612(00)00342-2 · Cited by 49 · see citing works
Peripheral nerve injury from implantation of chronic stimulating electrodes for pain control.
Kd, N., Watts, C., & Wk, C. (1976) · PubMed · Cited by 47 · see citing works
Loeb, G. E., Marks, W. B., & Beatty, P. G. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442964 · Cited by 46 · see citing works
Peripheral nerve regeneration through microelectrode arrays based on silicon technology
Navarro, X., Calvet, S., Butí, M., Gómez, N., Cabruja, E., Garrido, P., Villa, R., & Valderrama, E. (1996) · Restorative Neurology and Neuroscience · doi:10.3233/rnn-1996-9303 · Cited by 46 · see citing works
Self-citationCavallaro, E., Micera, S., Dario, P., Jensen, W., & Sinkjaer, T. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.816075 · Cited by 44 · cited 2× · see citing works
“Recorded ENG signals using cuff electrodes allowed the prediction of ankle movements in an animal model (Micera et al., 2001a; Jensen et al., 2002; Cavallaro et al., 2003) with the idea of using this information for the closed-loop control of standing in paraplegic subjects.”
BackgroundMethodSelf-citationFunctional electrical stimulation using microstimulators to correct foot drop: a case study
Weber, D. J., Stein, R. B., Chan, K. M., Loeb, G. E., Richmond, F. J. R., Rolf, R., James, K., Chong, S. L., Thompson, A. K., & Misiaszek, J. (2004) · Canadian Journal of Physiology and Pharmacology · doi:10.1139/y04-078 · Cited by 44 · see citing works
Rutten, W. L. C., Frieswijk, T. A., Smit, J. P. A., Rozijn, T. H., & Meier, J. H. (1995) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(95)96802-6 · Cited by 43 · see citing works
Guided outgrowth of leech neurons in culture
Fromherz, P., Schaden, H., & Vetter, T. (1991) · Neuroscience Letters · doi:10.1016/0304-3940(91)90724-8 · Cited by 43 · see citing works
Flagged first, then by citations
Research works printed but not linked: 18
- Brindley GS, 1995, Neuroprostheses., 15
- Bronzino JD, 1995, Biomedical Engineering Handbook.
- CoganSF LiuY‐P McCaffreyM TwardochUM(1993). Thin film hermetic coatings. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 5 N44‐NS‐2‐2311.
- De Luca CJ, 2002, Surface Electromyography: Detection and Recording.
- EdellDJ DevaneyL DurfeeWK GleasonEF GleasonKK HerndonTO ThompsonCV WrightonMS(1993). Bioresistant integrated circuits. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 9 NO1‐NS‐0‐2399.
- Hoffer JA, 2001, Neural Prostheses for Restoration of Sensory and Motor Function., 139
- Llinás R, 1973, Brain Unit Activity During Behaviour., 105
- Naples GG, 1990, Neural Prostheses: Fundamental Studies, Biophysics and Bioengineering Series., 107
- Nishikawa D, 1999, IEEE IROS'99, Kyongju, 1592
- PineJ TaiY‐C BuzsakiG BraginA CarpiD(1994). The cultured neuron probe. Quarterly Progress Report NIH‐NINDS Neural Prosthesis Program No. 4 NO1‐NS‐3‐2393.
- Popovic DB, 1992, Neural Prostheses Replacing Motor Function After Disease or Disability., 233
- Roblee LS, 1990, Neural Prostheses: Fundamental Studies, Biophysics and Bioengineering Series., 26
- Santa‐Cruz MC, 2000, IFESS 2000, 436
- SinghJ PeckRA LoebGE(2001). Development of BIONs technology for functional electrical stimulation: hermetic packaging. Proceedings of the 23th Annual International Conference of IEEE Eng Med and Bio Soc.
- Stein RB, 1992, Neural Prostheses. Replacing Motor Function After Diseases or Disability
- StieglitzT HabererW KammerS SchanzeT(2003). A simple model to investigate the stability of flexible micromachined retina stimulators. Proceedings of the 8th Annual International Conference of the International Functional Electrical Stimulation Society pp69–73.
- Tatic‐LucicS TaiY‐C WrightJA PineJ DenisonT(1993). Silicon‐micromachined neurochips for in vitro studies of cultured neural networks. Proceedings of the International Conference on Solid State Sensors and Actuators pp943–946.
- Tyler DJ, 1994, Interfascicular electrical stimulation for selectively activating axons, IEEE–EMBS Mag, 13, 575
Cited by
Sorted by citations
Brain–machine interfaces: past, present and future
Lebedev, M. A., & Nicolelis, M. A. L. (2006) · Trends in Neurosciences · doi:10.1016/j.tins.2006.07.004 · Cited by 1,875 · see citing works
“37 Navarro, X. et al. (2005) A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and”
BackgroundRestoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001 · see citing works
“In parallel, the rapid development of neural interfaces for the peripheral nervous system (16) has provided potential for new tools through which bidirectional communication with nerves in the stump could be potentially restored.”
BackgroundSelf-citationSurface Electromyography Signal Processing and Classification Techniques
Chowdhury, R. H., Reaz, M. B. I., Ali, M. H., Bakar, A. A. A., Chellappan, K., & Chang, T. (2013) · Sensors · doi:10.3390/s130912431 · Cited by 1,013 · see citing works
“For this reason, they are widely used as surface electrodes [7].”
MethodFarina, D., Jiang, N., Rehbaum, H., Holobar, A., Graimann, B., Dietl, H., & Aszmann, O. C. (2014) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2305111 · Cited by 999 · see citing works
A comprehensive review on emerging artificial neuromorphic devices
Zhu, J., Zhang, T., Yang, Y., & Huang, R. (2020) · Applied Physics Reviews · doi:10.1063/1.5118217 · Cited by 763 · see citing works
Literature Review on Needs of Upper Limb Prosthesis Users
Cordella, F., Ciancio, A. L., Sacchetti, R., Davalli, A., Cutti, A. G., Guglielmelli, E., & Zollo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00209 · Cited by 766 · see citing works
Syringe-injectable electronics
Liu, J., Fu, T.-M., Cheng, Z., Hong, G., Zhou, T., Jin, L., Duvvuri, M., Jiang, Z., Kruskal, P., Xie, C., et al. (2015) · Nature Nanotechnology · doi:10.1038/nnano.2015.115 · Cited by 687 · see citing works
“(14) After injection of electronics as described in steps 10-11, the I/O region of electronics was unfolded on the surface of the ceramic scaffold.”
MethodFlexible Hybrid Electronics for Digital Healthcare
Ma, Y., Zhang, Y., Cai, S., Han, Z., Liu, X., Wang, F., Cao, Y., Wang, Z., Li, H., Chen, Y., et al. (2020) · Advanced Materials · doi:10.1002/adma.201902062 · Cited by 620 · see citing works
Hassler, C., Boretius, T., & Stieglitz, T. (2011) · Journal of Polymer Science Part B: Polymer Physics · doi:10.1002/polb.22169 · Cited by 501 · see citing works
Self-citationNanomaterials for Neural Interfaces
Kotov, N. A., Winter, J. O., Clements, I. P., Jan, E., Timko, B. P., Campidelli, S., Pathak, S., Mazzatenta, A., Lieber, C. M., Prato, M., et al. (2009) · Advanced Materials · doi:10.1002/adma.200801984 · Cited by 504 · see citing works
Ortiz-Catalan, M., Håkansson, B., & Brånemark, R. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008933 · Cited by 484 · see citing works
Implanted Neural Interfaces: Biochallenges and Engineered Solutions
Grill, W. M., Norman, S. E., & Bellamkonda, R. V. (2009) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-061008-124927 · Cited by 488 · see citing works
“The closer a stimulation or recording electrode is to a nerve fiber that is buried within the 3D nerve cable, the more selective or specific the signal evoked or recorded (14).”
BackgroundA transverse intrafascicular multichannel electrode (TIME) to interface with the peripheral nerve
Boretius, T., Badia, J., Pascual-Font, A., Schuettler, M., Navarro, X., Yoshida, K., & Stieglitz, T. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.05.010 · Cited by 470 · see citing works
Self-citationForeign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454 · see citing works
“Although more challenging to carry out compared to nerve stimulation due to the distance between electrodes and the weak electrical signals transmitted by axons, with an ECM-rich epineurium separating the two, motor and sensory nerve signals gathered using cuff electrodes have been used to guide and fine-tune prosthes…”
BackgroundInfluentialOn the Shared Control of an EMG-Controlled Prosthetic Hand: Analysis of User–Prosthesis Interaction
Cipriani, C., Zaccone, F., Micera, S., & Carrozza, M. C. (2008) · IEEE Transactions on Robotics · doi:10.1109/tro.2007.910708 · Cited by 450 · see citing works
Self-citationLi, N., Zhang, X., Song, Q., Su, R., Zhang, Q., Kong, T., Liu, L., Jin, G. Y., Tang, M., & Cheng, G. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.08.065 · Cited by 428 · see citing works
Neural Stimulation with a Carbon Nanotube Microelectrode Array
Wang, K., Fishman, H. A., Dai, H., & Harris, J. S. (2006) · Nano Letters · doi:10.1021/nl061241t · Cited by 416 · see citing works
Control of Hand Prostheses Using Peripheral Information
Micera, S., Carpaneto, J., & Raspopović, S. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2085429 · Cited by 405 · see citing works
Self-citationDouble nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · see citing works
“Peripheral nerve interfaces aim to detect electrical activity of the nerve fibres and/or to excite them as selectively as possible (Navarro et al., 2005; Stieglitz et al., 2005; Tesfayesus and Durand, 2007; Micera et al., 2008).”
BackgroundSelf-citationConducting Polymers for Neural Prosthetic and Neural Interface Applications
Green, R. A., & Abidian, M. R. (2015) · Advanced Materials · doi:10.1002/adma.201501810 · Cited by 397 · see citing works
Upper limb amputees can be induced to experience a rubber hand as their own
Ehrsson, H. H., Rosén, B., Stockselius, A., Ragnö, C., Kohler, P., & Lundborg, G. (2008) · Brain · doi:10.1093/brain/awn297 · Cited by 343 · see citing works
“In principle, by connecting sensors in the artificial limbs to electrodes in the primary somatosensory cortex (London et al., 2008) or peripheral nerves (Navarro et al., 2005; Kuiken et al., 2007), one could effectively create tactile sensibility in the prosthetic limb.”
BackgroundInfluentialDesign of a cybernetic hand for perception and action
Carrozza, M. C., Cappiello, G., Micera, S., Edin, B. B., Beccai, L., & Cipriani, C. (2006) · Biological Cybernetics · doi:10.1007/s00422-006-0124-2 · Cited by 339 · see citing works
“Finally, PNS invasive interfaces can be used to discriminate different neural signals (Navarro et al. 2005; Dhillon et al. 2004; Citi et al. 2006).”
BackgroundSelf-citationSensory feedback for limb prostheses in amputees
Raspopovic, S., Valle, G., & Petrini, F. M. (2021) · Nature Materials · doi:10.1038/s41563-021-00966-9 · Cited by 308 · see citing works
Zhang, Y., Zheng, N., Cao, Y., Wang, F., Wang, P., Ma, Y., Lu, B., Hou, G., Fang, Z., Liang, Z., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw1066 · Cited by 314 · see citing works
Vu, P. P., Vaskov, A. K., Irwin, Z. T., Henning, P. T., Lueders, D. R., Laidlaw, A. T., Davis, A. J., Nu, C. S., Gates, D. H., Gillespie, R. B., et al. (2020) · Science Translational Medicine · doi:10.1126/scitranslmed.aay2857 · Cited by 291 · see citing works
“However, current peripheral nerve interfaces have limitations, which minimize their clinical utility (1, 2).”
BackgroundToward higher-performance bionic limbs for wider clinical use
Farina, D., Vujaklija, I., Brånemark, R., Bull, A. M. J., Dietl, H., Graimann, B., Hargrove, L. J., Hoffmann, K.-P., Huang, H., Ingvarsson, T., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00732-x · Cited by 263 · see citing works
“User intent in lower and upper limbs can also be decoded from electrical activity recorded from efferent axons in peripheral nerves 102–105 .”
BackgroundSelf-citationEnhancing functional abilities and cognitive integration of the lower limb prosthesis
Petrini, F. M., Valle, G., Bumbasirevic, M., Barberi, F., Bortolotti, D., Cvancara, P., Hiairrassary, A., Mijovic, P., Sverrisson, A. Ö., Pedrocchi, A., et al. (2019) · Science Translational Medicine · doi:10.1126/scitranslmed.aav8939 · Cited by 265 · see citing works
“We engineered wireless communication modules to control the parameters of PNES with a latency of less than 50 ms. Petrini et al., Sci. Transl.”
BackgroundSelf-citationFlexible and stretchable micro-electrodes for in vitro and in vivo neural interfaces
Lacour, S. P., Benmerah, S., Tarte, E. J., FitzGerald, J. J., Serra, J., McMahon, S. B., Fawcett, J. W., Graudejus, O., Yu, Z., & Morrison, B. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0644-8 · Cited by 265 · see citing works
“Dense arrays of up to 256 electrodes are used in vitro to monitor/stimulate individual neurons or acute tissue slices [14, 27] and in vivo to treat or repair damaged brain tissues or peripheral nerves [16, 31].”
BackgroundRestoration of sensory information via bionic hands
Bensmaı̈a, S. J., Tyler, D. J., & Micera, S. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00630-8 · Cited by 256 · see citing works
Self-citationWater-responsive supercontractile polymer films for bioelectronic interfaces
Yi, J., Zou, G., Huang, J., Ren, X., Tian, Q., Yu, Q., Wang, P., Yuan, Y., Tang, W., Wang, C., et al. (2023) · Nature · doi:10.1038/s41586-023-06732-y · Cited by 252 · see citing works
“Besides recording compound action potential with large amplitudes, our WRAP electrode can also record sensory neural signals with small amplitude 43 .”
Proprioception from a neurally controlled lower-extremity prosthesis
Clites, T. R., Carty, M. J., ULLAURI, J. G. B., Carney, M. E., Mooney, L. M., Duval, J.-F., Srinivasan, S. S., & Herr, H. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aap8373 · Cited by 232 · see citing works
“Although sometimes effective in preventing neuropathic pain, this technique creates a hurdle for neural interfaces because of the limited longitudinal viability of direct contact between synthetic interfaces and peripheral nerve tissues, especially for neural recording (16, 17).”
BackgroundBadia, J., Boretius, T., Andreu, D., Azevedo-Coste, C., Stieglitz, T., & Navarro, X. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/3/036023 · Cited by 227 · see citing works
Self-citationSolderable and electroplatable flexible electronic circuit on a porous stretchable elastomer
Jeong, G. S., Baek, D.-H., Jung, H., Song, J. H., Moon, J. H., Hong, S. W., Kim, I. Y., & Lee, S. (2012) · Nature Communications · doi:10.1038/ncomms1980 · Cited by 227 · see citing works
Flexible inorganic bioelectronics
Chen, Y., Zhang, Y., Liang, Z., Cao, Y. K., Han, Z., & Feng, X. (2020) · npj Flexible Electronics · doi:10.1038/s41528-020-0065-1 · Cited by 231 · see citing works
“In additional to the recording of CNS, the recording of PNS plays an important role in both neuroscience research and clinical practices.(107) Aiming to balance the invasiveness and high selectivity, researchers have kept optimizing the bio-electrodes for PNS from traditional intraneural electrodes such as Michigan an…”
BackgroundLago, N., Yoshida, K., Koch, K. P., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886617 · Cited by 227 · see citing works
“secure high precision and repeatability, and that a higher number of active sites can be positioned within a small surface [ 1 ].”
BackgroundInfluentialSelf-citationPeripheral nerve stimulation for neuropathic pain
Slavin, K. (2011) · Neurotherapeutics · doi:10.1016/j.nurt.2007.11.005 · Cited by 199 · see citing works
Brain-computer interface technologies: from signal to action
Ortiz-Rosario, A., & Adeli, H. (2013) · Reviews in the Neurosciences · doi:10.1515/revneuro-2013-0032 · Cited by 212 · see citing works
“These methods provide the highest spatial resolution yet are the most invasive, typically causing damage to tissue surrounding the recording site and cells in the area (Navarro et al., 2005).”
MethodSafety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
Clemente, F., D'Alonzo, M., Controzzi, M., Edin, B. B., & Cipriani, C. (2016) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2015.2500586 · Cited by 209 · see citing works
Triboelectric Nanogenerators for Therapeutic Electrical Stimulation
Conta, G., Libanori, A., Tat, T., Chen, G., & Chen, J. (2021) · Advanced Materials · doi:10.1002/adma.202007502 · Cited by 205 · see citing works
He, K., Wang, C., He, Y., Su, J., & Chen, X. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00527 · Cited by 202 · see citing works
Neural Data-Driven Musculoskeletal Modeling for Personalized Neurorehabilitation Technologies
Sartori, M., Llyod, D. G., & Farina, D. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2538296 · Cited by 188 · see citing works
Biomimetic approaches to bionic touch through a peripheral nerve interface
Saal, H. P., & Bensmaia, S. J. (2015) · Neuropsychologia · doi:10.1016/j.neuropsychologia.2015.06.010 · Cited by 179 · see citing works
“Here, we describe recent developments in peripheral nerve interfaces, along with their respective advantages and disadvantages (for a more in depth review of peripheral interfaces, see Navarro et al., 2005).”
Schultz, A. E., & Kuiken, T. A. (2011) · PM&R · doi:10.1016/j.pmrj.2010.06.016 · Cited by 186 · see citing works
“One common type of extraneural electrode, the nerve cuff electrode, surrounds the nerve and therefore records and stimulates a few to several combined nerve fascicles, depending on the number and arrangement of contacts (Figure 5A) [39].”
BackgroundMethodAtzori, M., & Müller, H. (2015) · Frontiers in Systems Neuroscience · doi:10.3389/fnsys.2015.00162 · Cited by 184 · see citing works
“Man-machine interfaces have been developed to control hand prostheses via the brain (Lebedev and Nicolelis, 2006), peripheral nerves (Navarro et al., 2005) or the muscles (Cipriani et al., 2011).”
BackgroundKung, T. A., Langhals, N. B., Martin, D. C., Johnson, P. J., Cederna, P. S., & Urbanchek, M. G. (2014) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000000168 · Cited by 181 · see citing works
Incidence of Traumatic Peripheral Nerve Injuries and Amputations in Sweden between 1998 and 2006
Asplund, M., Nilsson, M., Jacobsson, A., & von Holst, H. (2009) · Neuroepidemiology · doi:10.1159/000197900 · Cited by 179 · see citing works
Long-term usability and bio-integration of polyimide-based intra-neural stimulating electrodes
Wurth, S., Capogrosso, M., Raspopovic, S., Gandar, J., Federici, G., Kinany, N., Cutrone, A., Piersigilli, A., Pavlova, N., Guiet, R., et al. (2017) · Biomaterials · doi:10.1016/j.biomaterials.2017.01.014 · Cited by 176 · see citing works
Self-citationReal-time encoding and compression of neuronal spikes by metal-oxide memristors
Gupta, I., Serb, A., Khiat, A., Zeitler, R., Vassanelli, S., & Prodromakis, T. (2016) · Nature Communications · doi:10.1038/ncomms12805 · Cited by 187 · see citing works
Implantable neurotechnologies: a review of micro- and nanoelectrodes for neural recording
Patil, A. C., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1430-4 · Cited by 175 · see citing works
Multi-Modal Sensing Techniques for Interfacing Hand Prostheses: A Review
Fang, Y., Hettiarachchi, N., Zhou, D., & Liu, H. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2450211 · Cited by 173 · see citing works
“Clinical peripheral nerve electrodes can be classified into three groups: intraneural, extraneural and miscellaneous, as discussed in [79] and [80].”
BackgroundKwon, K. Y., Sirowatka, B., Weber, A., & Li, W. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2013.2282318 · Cited by 166 · see citing works
Advances in functional electrical stimulation (FES)
Popović, D. B. (2014) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2014.09.008 · Cited by 171 · see citing works
“Based on the discussion presented above, the following conclusions can be stated: (1) Array multi-pad electrodes can be implanted in or around the peripheral nerves (Navarro et al., 2005).”
BackgroundPolymer integration for packaging of implantable sensors
Qin, Y., Howlader, M. M. R., Deen, M. J., Haddara, Y. M., & Selvaganapathy, P. R. (2014) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2014.05.063 · Cited by 163 · see citing works
Materials for microfabricated implantable devices: a review
Scholten, K., & Meng, E. F. (2015) · Lab on a Chip · doi:10.1039/c5lc00809c · Cited by 161 · see citing works
Noninvasive Neural Prostheses Using Mobile and Wireless EEG
Lin, C., Ko, L., Chiou, J., Duann, J., Huang, R., Liang, S., Chiu, T.-W., & Jung, T. (2008) · Proceedings of the IEEE · doi:10.1109/jproc.2008.922561 · Cited by 160 · see citing works
“During the current decade, hundreds of thousands of patients have already been helped by neuroprosthetic technologies that restore vision [1], hearing [2] or movement [ 3 ] and relieve chronic pain or neurological disorders [4].”
BackgroundSemiconducting Polymers for Neural Applications
Dimov, I. B., Moser, M., Malliaras, G. G., & McCulloch, I. A. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00685 · Cited by 160 · see citing works
Flexible Electrodes for In Vivo and In Vitro Electrophysiological Signal Recording
Zhu, M., Wang, H., Li, S., Liang, X., Zhang, M., Dai, X., & Zhang, Y. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100646 · Cited by 158 · see citing works
Aszmann, O. C., Roche, A. D., Salminger, S., Paternostro-Sluga, T., Herceg, M., Sturma, A., Hofer, C., & Farina, D. (2015) · The Lancet · doi:10.1016/s0140-6736(14)61776-1 · Cited by 152 · see citing works
Flexible, all-polymer microelectrode arrays for the capture of cardiac and neuronal signals
Blau, A. W., Murr, A., Wolff, S., Sernagor, E., Medini, P., Iurilli, G., Ziegler, C., & Benfenati, F. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.11.014 · Cited by 155 · see citing works
Effects of surface charges of graphene oxide on neuronal outgrowth and branching
Tu, Q., Pang, L., Chen, Y., Zhang, Y., Zhang, R., Lu, B., & Wang, J. (2013) · The Analyst · doi:10.1039/c3an01796f · Cited by 151 · see citing works
Sikdar, S., Rangwala, H., Eastlake, E. B., Hunt, I. A., Nelson, A. J., Devanathan, J., Shin, A., & Pancrazio, J. J. (2013) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2013.2274657 · Cited by 146 · see citing works
Ortiz-Catalan, M., Brånemark, R., Håkansson, B., & Delbeke, J. (2012) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-11-33 · Cited by 144 · see citing works
“[7], and in 2008 by Cogan with the emphasis on microelectrodes for stimulation [8].”
BackgroundClinical applications of penetrating neural interfaces and Utah Electrode Array technologies
Normann, R. A., & Fernandez, E. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/6/061003 · Cited by 135 · see citing works
“However there are limitations associated with the use of cuff electrodes (Navarro et al 2005, Micera and Navarro 2009) that could be overcome by using intrafascicular penetrating electrode arrays such as the USEA.”
BackgroundA review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
A Haptic Feedback System for Lower-Limb Prostheses
Fan, R. E., Culjat, M. O., King, C., Franco, M. L., Boryk, R. J., Bisley, J. W., Dutson, E., & Grundfest, W. S. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.920075 · Cited by 138 · see citing works
“Furthermore, implanted electrodes require invasive surgery, and errors determined from postoperative testing would require additional invasive surgical correction [22].”
BackgroundViana, D., Walston, S. T., Masvidal‐Codina, E., Illa, X., Rodríguez‐Meana, B., del Valle, J., Hayward, A., Dodd, A., Loret, T., Prats‐Alfonso, E., et al. (2024) · Nature Nanotechnology · doi:10.1038/s41565-023-01570-5 · Cited by 140 · see citing works
“To broaden the range of clinical uses of neural interfaces, improvements in efficacy are needed so that the treatment benefit outweighs the associated risks 6–9 .”
BackgroundInfluentialSelf-citationCho, Y., Park, S., Lee, J., & Yu, K. J. (2021) · Advanced Materials · doi:10.1002/adma.202005786 · Cited by 136 · see citing works
Deer, T. R., Krames, E. S., Mekhail, N. A., Pope, J. E., Leong, M. S., Stanton-Hicks, M. D., Golovac, S., Kapural, L., Alo, K., Anderson, J., et al. (2014) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.12204 · Cited by 135 · see citing works
“Presently, electrical devices such as DBSs are used in the brain for restoring normal function for movement disorders (14), PD (30), epilepsy (19), and psychiatric disease (207), SCS over the spinal cord for restoration of motor function (208) and amelioration of pain (209), PNS for functional restoration (210) and pa…”
BackgroundControl of prosthetic hands via the peripheral nervous system
Ciancio, A. L., Cordella, F., Barone, R., Romeo, R. A., Bellingegni, A. D., Sacchetti, R., Davalli, A., Di Pino, G., Ranieri, F., Di Lazzaro, V., et al. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00116 · Cited by 131 · see citing works
“This work was supported partly by the National Institute for Insurance against Accidents at Work (INAIL) with the PPR 2 project (CUP: E58C13000990001), partly by the Italian Ministry of Instruction, University and Research with PRIN HANDBOT project (CUP: B81J12002680008) and partly by the Italian…”
BackgroundMethodInfluentialChai, G., Sui, X., Li, S., He, L., & Lan, N. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/6/066002 · Cited by 131 · see citing works
Self‐Assembled On‐Chip‐Integrated Giant Magneto‐Impedance Sensorics
Karnaushenko, D., Karnaushenko, D. D., Makarov, D., Baunack, S., Schäfer, R., & Schmidt, O. G. (2015) · Advanced Materials · doi:10.1002/adma.201503127 · Cited by 125 · see citing works
Lee, J. H., Kim, H., Kim, J. H., & Lee, S. (2016) · Lab on a Chip · doi:10.1039/c5lc00842e · Cited by 124 · see citing works
Adaptive self-healing electronic epineurium for chronic bidirectional neural interfaces
Song, K.-I., Seo, H., Seong, D., Kim, S., Yu, K. J., Kim, Y.-C., Kim, J., Kwon, S. J., Han, H.-S., Youn, I., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-18025-3 · Cited by 120 · see citing works
Micera, S., Navarro, X., Carpaneto, J., Citi, L., Tonet, O., Rossini, P. M., Carrozza, M. C., Hoffmann, K. P., Vivo, M., Yoshida, K., et al. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.2006207 · Cited by 120 · see citing works
Self-citationEMG-driven control in lower limb prostheses: a topic-based systematic review
Cimolato, A., Driessen, J. J. M., Mattos, L. S., Momi, E. D., Laffranchi, M., & Michieli, L. D. (2022) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-022-01019-1 · Cited by 122 · see citing works
“Instead, whereas numerous other reviews focus specifically on myoelectric control, they either do so for generic HRI [1, 3, 93, 94, 97, 103, 109], or for upper limb prosthetics specifically [47, 102, 107].”
BackgroundInfluentialThin films and microelectrode arrays for neuroprosthetics
Ordonez, J., Schuettler, M., Boehler, C., Boretius, T., & Stieglitz, T. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.117 · Cited by 118 · see citing works
Self-citationRaspopović, S., Capogrosso, M., & Micera, S. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2151878 · Cited by 113 · see citing works
“They range from cuff electrodes, having low invasiveness and low selectivity, to sieve electrodes, having higher selectivity but, unfortunately, also higher invasiveness [ 9 ].”
BackgroundSelf-citationInterfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
Valle, G., Petrini, F. M., Strauss, I., Iberite, F., D’Anna, E., Granata, G., Controzzi, M., Cipriani, C., Stieglitz, T., Rossini, P. M., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34910-w · Cited by 117 · see citing works
Self-citationCiti, L., Carpaneto, J., Yoshida, K., Hoffmann, K., Koch, K., Dario, P., & Micera, S. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.04.025 · Cited by 117 · see citing works
“Currently, there are various electrodes under investigation as neural interfaces (see Navarro et al., 2005, for a review on the available electrodes).”
Self-citationDevelopment of a regenerative peripheral nerve interface for control of a neuroprosthetic limb
Urbanchek, M. G., Kung, T. A., Frost, C. M., Martin, D. C., Larkin, L. M., Wollstein, A., & Cederna, P. S. (2016) · BioMed Research International · doi:10.1155/2016/5726730 · Cited by 114 · see citing works
“Other investigators have attempted to provide prosthetic control through direct neural interfacing of the central or peripheral nervous systems [1, 2].”
BackgroundNeuromorphic Artificial Touch for Categorization of Naturalistic Textures
Rongala, U. B., Mazzoni, A., & Oddo, C. M. (2015) · IEEE Transactions on Neural Networks and Learning Systems · doi:10.1109/tnnls.2015.2472477 · Cited by 113 · see citing works
“Provided that proper neural interfaces will be available for upper limb neuroprostheses [5], [6], neuromorphic sensors can emulate human mechanoreceptors to target the partial restoration of natural tactile feedback, which is a major need in manipulation tasks [7].”
BackgroundInvasive Intraneural Interfaces: Foreign Body Reaction Issues
Lotti, F., Ranieri, F., Vadalà, G., Zollo, L., & Pino, G. D. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00497 · Cited by 117 · see citing works
“…epimysial and cuff electrodes as it stabilizes the implant (Grandjean and Mortimer, 1986; Grill and Mortimer, 1998), FBR is an important issue to be solved in intraneural electrodes (Navarro et al., 2005), limiting device function over time, often forcing a premature removal (Badia et al., 2011).”
BackgroundPino, G. D., Guglielmelli, E., & Rossini, P. M. (2009) · Progress in Neurobiology · doi:10.1016/j.pneurobio.2009.03.001 · Cited by 112 · see citing works
“…issues, such as neural inflammatory response, structural and chemical characteristic of the implanted tissue and temporal stability of the implant (Navarro et al., 2005; Schwartz, 2004), but further knowledge is needed primarily on the central nervous system and its reorganization following limb…”
Current Solutions and Future Trends for Robotic Prosthetic Hands
Mendez, V., Iberite, F., Shokur, S., & Micera, S. (2021) · Annual Review of Control Robotics and Autonomous Systems · doi:10.1146/annurev-control-071020-104336 · Cited by 112 · see citing works
Self-citationCho, S., Lu, H., Cauller, L. J., Romero‐Ortega, M. I., Lee, J., & Hughes, G. A. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2008.2006261 · Cited by 112 · see citing works
“microelectrodes that can detect or stimulate neural activity throughout the nervous system [3].”
BackgroundNeuromorphic hardware for somatosensory neuroprostheses
Donati, E., & Valle, G. (2024) · Nature Communications · doi:10.1038/s41467-024-44723-3 · Cited by 115 · see citing works
Bridging the gap in peripheral nerve repair with 3D printed and bioprinted conduits
Dixon, A. R., Jariwala, S. H., Bilis, Z., Loverde, J. R., Pasquina, P. F., & Alvarez, L. M. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.09.010 · Cited by 110 · see citing works
Electrical Stimuli in the Central Nervous System Microenvironment
Thompson, D. M., Koppes, A. N., Hardy, J. G., & Schmidt, C. E. (2014) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-121813-120655 · Cited by 107 · see citing works
A Review: Electrode and Packaging Materials for Neurophysiology Recording Implants
Yang, W., Gong, Y. P., & Li, W. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.622923 · Cited by 106 · see citing works
“The tissue should produce minimal glial encapsulation surrounding the implant and only mild foreign body reaction without evidence of necrosis or implant rejection (Navarro et al., 2005; Márton et al., 2020), as shown in Figure 2B.”
BackgroundNeural interfaces for somatosensory feedback: bringing life to a prosthesis
Tyler, D. J. (2015) · Current Opinion in Neurology · doi:10.1097/wco.0000000000000266 · Cited by 105 · see citing works
“There are several options being explored for peripheral nerve interfaces [84,85].”
BackgroundInterfaces with the Peripheral Nerve for the Control of Neuroprostheses
del Valle, J., & Navarro, X. (2013) · International Review of Neurobiology · doi:10.1016/b978-0-12-420045-6.00002-x · Cited by 105 · see citing works
Self-citationBiomimetic Microelectronics for Regenerative Neuronal Cuff Implants
Karnaushenko, D., Münzenrieder, N., Karnaushenko, D. D., Koch, B., Meyer, A. K., Baunack, S., Petti, L., Tröster, G., Makarov, D., & Schmidt, O. G. (2015) · Advanced Materials · doi:10.1002/adma.201503696 · Cited by 103 · see citing works
Badia, J., Boretius, T., Pascual-Font, A., Udina, E., Stieglitz, T., & Navarro, X. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2153850 · Cited by 103 · see citing works
“Various approaches have been developed to interface the peripheral nervous system and to be able to record from and/or to stimulate its nerve fibers [1], [2].”
BackgroundSelf-citationMicera, S., Citi, L., Rigosa, J., Carpaneto, J., Raspopović, S., Pino, G. D., Rossini, L., Yoshida, K., Denaro, L., Dario, P., et al. (2010) · Proceedings of the IEEE · doi:10.1109/jproc.2009.2038726 · Cited by 100 · see citing works
Self-citationMicera, S., Rossini, P. M., Rigosa, J., Citi, L., Carpaneto, J., Raspopovic, S., Tombini, M., Cipriani, C., Assenza, G., Carrozza, M. C., et al. (2011) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-8-53 · Cited by 100 · see citing works
“Several invasive PNS interfaces have been developed in the past [18].”
BackgroundSelf-citationNeural machine interfaces for controlling multifunctional powered upper-limb prostheses
Ohnishi, K., Weir, R. F. F., & Kuiken, T. (2006) · Expert Review of Medical Devices · doi:10.1586/17434440.4.1.43 · Cited by 103 · see citing works
Mastinu, E., Doguet, P., Botquin, Y., Håkansson, B. E. V., & Ortiz-Catalan, M. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2694710 · Cited by 98 · see citing works
“Human-Machine Gateway (OHMG) combines the benefits of skeletal attachment with the reliability and increased information provided by implanted muscular electrodes [13], [14].”
BackgroundThin film multi-electrode softening cuffs for selective neuromodulation
González-González, M. A., Kanneganti, A., Joshi-Imre, A., Hernandez-Reynoso, A. G., Bendale, G., Modi, R., Ecker, M., Khurram, A., Cogan, S. F., Voit, W. E., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34566-6 · Cited by 97 · see citing works
- Authors
- Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P.
- Affiliations
- Universitat Autònoma de Barcelona (ES)
- University of Freiburg (DE)
- Scuola Superiore Sant'Anna (IT)
- Corresponding author
- Navarro, Xavier
- Venue
- Journal of the Peripheral Nervous System
- Publisher
- Wiley
- Published
- 2005-09
- Volume
- 10
- Issue
- 3
- Pages
- 229-258
- ISSN
- 1085-9489 (ISSN-L)
- 1529-8027
- OpenAlex
- W2108068079
- Cited by
- 845
- References
- 259
- Citations by year
- 2026: 18 · 2025: 27 · 2024: 36 · 2023: 32 · 2022: 39 · 2021: 53 · 2020: 49 · 2019: 58 · 2018: 53 · 2017: 49 · 2016: 60 · 2015: 57 · 2014: 58 · 2013: 45 · 2012: 28
- FWCI
- 2.19
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of the Peripheral Nervous System (publishedVersion)
- PubMed (publishedVersion)
- Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- CINECA IRIS Institutional Research Information System (Sant'Anna School of Advanced Studies) (submittedVersion)
- Funders
- European Commission (IST-001917)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- neural prostheses, peripheral nerve interfaces, peripheral nerve stimulation, neuromuscular stimulation, electromyography, surface electrodes, cuff electrodes, intrafascicular electrodes, electrode design, tissue damage, robotic prostheses
- MeSH terms
- Animals, Bionics, Electrodes, Electromyography, Humans, User-Computer Interface, Peripheral Nervous System, Prostheses and Implants
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-01.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-01.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review