ECG artifacts and heart period variability: Don't miss a beat!
Abstract
The impact of artifacts on estimates of heart period variability were evaluated by modeling the effects of missed R‐waves and spurious R‐wave detections in actual and simulated heart period series. Results revealed that even a single artifact, occurring within a 128‐s interbeat interval series, can impart substantial spurious variance into all commonly analyzed frequency bands, including that associated with respiratory sinus arrhythmia. In fact, the spurious variance introduced by a single artifact may be greater than that associated with true basal heart period variability and can far exceed typical effect sizes in psychophysiological studies. The effects of artifacts are not related to a specific analytical method and are apparent in both frequency and time domain analyses. Results emphasize the importance of artifact detection and resolution for studies of heart period variability.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Only one heartbeat makes a difference in the analysis.
“This is eloquently explained by Berntson and Stowell (1998) who stated that only one heartbeat makes a difference in the analysis.”
From Heart Rate Variability and Cardiac Vagal Tone in Psychophysiological Research – Recommendations for Experiment Planning, Data Analysis, and Data Reporting (2017) · cites this paper as Berntson and Stowell (1998)
Artifact correction is a very important step in the pre-processing of the HRV signal.
“As we will detail later artifact correction is very important step in the pre-processing of the signal (Berntson and Stowell, 1998; Shaffer and Combatalade, 2013).”
From Heart Rate Variability and Cardiac Vagal Tone in Psychophysiological Research – Recommendations for Experiment Planning, Data Analysis, and Data Reporting (2017) · cites this paper as Berntson and Stowell, 1998
Cited by
Sorted by citations
An overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185 · see citing works
Heart rate variability: origins, methods, and interpretive caveats
BERNTSON, G. G., THOMAS BIGGER, J., ECKBERG, D. L., GROSSMAN, P., KAUFMANN, P. G., MALIK, M., NAGARAJA, H. N., PORGES, S. W., SAUL, J. P., STONE, P. H., et al. (1997) · Psychophysiology · doi:10.1111/j.1469-8986.1997.tb02140.x · Cited by 3,962 · see citing works
Self-citationLaborde, S., Mosley, E., & Thayer, J. F. (2017) · Frontiers in Psychology · doi:10.3389/fpsyg.2017.00213 · Cited by 2,336 · see citing works
“As we will detail later artifact correction is very important step in the pre-processing of the signal (Berntson and Stowell, 1998; Shaffer and Combatalade, 2013).”
BackgroundThe many metrics of cardiac chronotropy: A pragmatic primer and a brief comparison of metrics
Allen, J. J. B., Chambers, A. S., & Towers, D. N. (2007) · Biological Psychology · doi:10.1016/j.biopsycho.2006.08.005 · Cited by 491 · see citing works
Quintana, D. S., Alvares, G. A., & Heathers, J. A. J. (2016) · Translational Psychiatry · doi:10.1038/tp.2016.73 · Cited by 420 · see citing works
Considerations in the assessment of heart rate variability in biobehavioral research
Quintana, D. S., & Heathers, J. A. J. (2014) · Frontiers in Psychology · doi:10.3389/fpsyg.2014.00805 · Cited by 395 · see citing works
Brage, S., Ekelund, U., Brage, N., Hennings, M. A., Froberg, K., Franks, P. W., & Wareham, N. J. (2007) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00092.2006 · Cited by 303 · see citing works
ARTiiFACT: a tool for heart rate artifact processing and heart rate variability analysis.
Kaufmann, T., Sütterlin, S., Schulz, S. M., & Vögele, C. (2011) · Behavior Research Methods · doi:10.3758/s13428-011-0107-7 · Cited by 284 · see citing works
Gaussian process robust regression for noisy heart rate data
Stegle, O., Fallert, S. V., MacKay, D. J. C., & Brage, S. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.923118 · Cited by 213 · see citing works
Blum, J., Rockstroh, C., & Göritz, A. S. (2019) · Frontiers in Psychology · doi:10.3389/fpsyg.2019.02172 · Cited by 183 · see citing works
“…to be reliable in most conditions (Gillinov et al., 2017; Plews et al., 2017), it does not allow for precise artifact correction via visual inspection of the ECG signal to understand whether artifacts are due to technical or physiological issues (Berntson and Stowell, 1998; Laborde et al., 2017).”
BackgroundBenevides, T. W., & Lane, S. J. (2015) · Journal of Autism and Developmental Disorders · doi:10.1007/s10803-013-1971-z · Cited by 137 · see citing works
On the validity of using the Polar RS800 heart rate monitor for heart rate variability research
Quintana, D. S., Heathers, J. A. J., & Kemp, A. H. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-012-2453-2 · Cited by 118 · see citing works
Heart rate variability is associated with disease severity in psychosis spectrum disorders
Benjamin, B. R., Valstad, M., Elvsåshagen, T., Jönsson, E. G., Moberget, T., Winterton, A., Haram, M., Høegh, M. C., Lagerberg, T. V., Steen, N. E., et al. (2021) · Progress in Neuro-Psychopharmacology and Biological Psychiatry · doi:10.1016/j.pnpbp.2020.110108 · Cited by 52 · see citing works
Rogers, B., Giles, D., Draper, N., Mourot, L., & Gronwald, T. (2021) · Sensors · doi:10.3390/s21030821 · Cited by 54 · see citing works
Beda, A., Simpson, D. M., Carvalho, N. C., & Carvalho, A. R. S. (2014) · Psychophysiology · doi:10.1111/psyp.12163 · Cited by 43 · see citing works
Mathewson, K. J., Drmic, I. E., Jetha, M. K., Bryson, S. E., Goldberg, J. O., Hall, G. B., Santesso, D. L., Segalowitz, S. J., & Schmidt, L. A. (2011) · Autism Research · doi:10.1002/aur.176 · Cited by 40 · see citing works
Escalona-Vargas, D., Bolin, E. H., Lowery, C. L., Siegel, E. R., & Eswaran, H. (2020) · Physiological Measurement · doi:10.1088/1361-6579/abc353 · Cited by 26 · see citing works
Shared hearts and minds: physiological synchrony during empathy
Qaiser, J., Leonhardt, N. D., Le, B. M., Gordon, A. M., Impett, E. A., & Stellar, J. E. (2023) · Affective Science · doi:10.1007/s42761-023-00210-4 · Cited by 21 · see citing works
Czamanski-Cohen, J., Wiley, J., & Weihs, K. (2020) · BMJ Open · doi:10.1136/bmjopen-2020-037521 · Cited by 19 · see citing works
- Authors
- Berntson, G. G., & Stowell, J. R.
- Affiliations
- The Ohio State University (US)
- Corresponding author
- Berntson, Gary G.
- Venue
- Psychophysiology
- Publisher
- Wiley
- Published
- 1998-01
- Volume
- 35
- Issue
- 1
- Pages
- 127-132
- ISSN
- 0048-5772 (ISSN-L)
- 1469-8986
- OpenAlex
- W1969980989
- Cited by
- 223
- References
- 0
- Citations by year
- 2026: 2 · 2025: 8 · 2024: 12 · 2023: 12 · 2022: 15 · 2021: 13 · 2020: 14 · 2019: 12 · 2018: 15 · 2017: 16 · 2016: 13 · 2015: 10 · 2014: 16 · 2013: 8 · 2012: 15
- FWCI
- 0.26
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Psychophysiology (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Heart, Lung, and Blood Institute (HL52321)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- ECG artifacts, heart rate variability, R-wave detection, respiratory sinus arrhythmia, artifact detection, frequency domain analysis, time-domain analysis
- MeSH terms
- Computer Simulation, Data Interpretation, Statistical, Electrocardiography, Heart Rate, Humans
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review