Design, Characterization, and One-Point in vivo Calibration of a Subcutaneously Implanted Glucose Electrode

Open access
Csoeregi, Elisabeth., Quinn, Chris P., Schmidtke, David W., Lindquist, Sten-Eric., Pishko, Michael V., Ye, Ling., Katakis, Ioanis., Hubbell, Jeffrey A., Heller, Adam. · 1994 · Analytical Chemistry
Cited by 110

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

A 0.29-mm-diameter flexible electrode designed for subcutaneous in vivo amperometric monitoring of glucose is described. The electrode was designed to allow "one-point" in vivo calibration, i.e., to have zero output current at zero glucose concentration, even in the presence of other electroreactive species of serum or blood. A valid zero point, along with a measurement of the glucose concentration in a withdrawn sample of blood at which the current is known, defined the sensitivity in the linear response range. The electrode was four-layered, with the layers serially deposited within a 0.125-mm recess upon the tip of a polyimide-insulated 0.25-mm gold wire. The recessed structure reduced the sensitivity to movement and allowed, through control of the depth of the recess, control of the transport of glucose and thus the range of linearity. The recess contained the four polymeric layers, with a total mass less than 5 micrograms and no leachable components. The bottom glucose concentration-to-current transducing layer consisted of the enzyme "wiring" redox polymer poly[(vinylimidazole)Os(bipyridine)2Cl]+ , complexed with recombinant glucose oxidase and cross-linked with poly(ethylene glycol) diglycidyl ether, to form an electron-conducting hydrogel. The layer was overcoated with an electrically insulating layer of polyaziridine-cross-linked poly(allylamine), on which an immobilized interference-eliminating horseradish peroxidase based film was deposited. An outer biocompatible layer was formed by photo-cross-linking derivatized poly(ethylene oxide). The current output of a typical electrode at 10 mM glucose and at 37 degrees C was 35 nA, the apparent Km was 20 mM, and the 10-90% response time was approximately 1 min.(ABSTRACT TRUNCATED AT 250 WORDS)

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1021/ac00091a022

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 110

Metadata

Authors
Csoeregi, E., Quinn, C. P., Schmidtke, D. W., Lindquist, S.-E., Pishko, M. V., Ye, L., Katakis, I., Hubbell, J. A., & Heller, A.
Venue
Analytical Chemistry
Published
1994-10-01
Volume
66
Issue
19
Pages
3131-3138
ISSN
  • 0003-2700 (ISSN-L)
  • 1520-6882
OpenAlex
W2069720757
Cited by
110
References
2
FWCI
10.25
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Design, Characterization, and One-Point in vivo Calibration of a Subcutaneously Implanted Glucose Electrode. https://app.cochat.ai/papers/10.1021/ac00091a022
The paper
Csoeregi, E., Quinn, C. P., Schmidtke, D. W., Lindquist, S.-E., Pishko, M. V., Ye, L., Katakis, I., Hubbell, J. A., & Heller, A. (1994). Design, Characterization, and One-Point in vivo Calibration of a Subcutaneously Implanted Glucose Electrode. Analytical Chemistry. https://doi.org/10.1021/ac00091a022

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.