On-Line 1D and 2D Porous Layer Open Tubular/LC-ESI-MS Using 10-μm-i.d. Poly(styrene−divinylbenzene) Columns for Ultrasensitive Proteomic Analysis
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Following on our recent work, on-line one-dimensional (1D) and two-dimensional (2D) porous layer open tubular/liquid chromatography-electrospray ionization-mass spectrometry (PLOT/LC-ESI-MS) platforms using 3.2 mx10 microm i.d. poly(styrene-divinylbenzene) (PS-DVB) PLOT columns have been developed to provide robust, high-performance, and ultrasensitive proteomic analysis. With the use of a PicoClear tee, the dead volume connection between a 50 microm i.d. PS-DVB monolithic micro-SPE column and the PLOT column was minimized. The micro-SPE/PLOT column assembly provided a separation performance similar to that obtained with direct injection onto the PLOT column at a mobile phase flow rate of 20 nL/min. The trace analysis potential of the platform was evaluated using an in-gel tryptic digest sample of a gel fraction (15-40 kDa) of a cervical cancer (SiHa) cell line. As an example of the sensitivity of the system, approximately 2.5 ng of protein in 2 microL of solution, an amount corresponding to 20 SiHa cells, was subjected to on-line micro-SPE-PLOT/LC-ESI-MS/MS analysis using a linear ion trap MS. A total of 237 peptides associated with 163 unique proteins were identified from a single analysis when using stringent criteria associated with a false positive rate of less than 1%. The number of identified peptides and proteins increased to 638 and 343, respectively, as the injection amount was raised to approximately 45 ng of protein, an amount corresponding to 350 SiHa cells. In comparison, only 338 peptides and 231 unique proteins were identified (false positive rate again less than 1%) from 750 ng of protein from the identical gel fraction, an amount corresponding to 6000 SiHa cells, using a typical 15 cmx75 microm i.d. packed capillary column. The greater sensitivity, higher recovery, and higher resolving power of the PLOT column resulted in the increased number of identifications from only approximately 5% of the injected sample amount. The resolving power of the micro-SPE/PLOT assembly was further extended by 2D chromatography via combination of the high-efficiency reversed-phase PLOT column with strong cation-exchange chromatography (SCX). As an example, 1071 peptides associated with 536 unique proteins were identified from 75 ng of protein from the same gel fraction, an amount corresponding to 600 cells, using five ion-exchange fractions in on-line 2D SCX-PLOT/LC-MS. The 2D system, implemented in an automated format, led to simple and robust operation for proteomic analysis. These promising results demonstrate the potential of the PLOT column for ultratrace analysis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Elias, J. E., & Gygi, S. P. (2007) · Nature Methods · doi:10.1038/nmeth1019 · Cited by 3,925
Cited by
Effect of eluent on the ionization process in liquid chromatography–mass spectrometry
Kostiainen, R., & Kauppila, T. J. (2009) · Journal of Chromatography A · doi:10.1016/j.chroma.2008.08.095 · Cited by 402
Wu, R., Hu, L., Wang, F., Ye, M., & Zou, H. (2008) · Journal of Chromatography A · doi:10.1016/j.chroma.2007.09.022 · Cited by 257
Ultrasensitive and Fast Bottom‐up Analysis of Femtogram Amounts of Complex Proteome Digests
Sun, L., Zhu, G., Zhao, Y., Yan, X., Mou, S., & Dovichi, N. J. (2013) · Angewandte Chemie International Edition · doi:10.1002/anie.201308139 · Cited by 212
The Revolution and Evolution of Shotgun Proteomics for Large-Scale Proteome Analysis
Yates, J. R. (2013) · Journal of the American Chemical Society · doi:10.1021/ja3094313 · Cited by 131
Wu, Q., Yuan, H., Zhang, L., & Zhang, Y. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2012.04.010 · Cited by 94
Metadata
- Authors
- Luo, Q., Yue, G., Valaskovic, G. A., Gu, Y., Wu, S.-L., & Karger, B. L.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2007-07-11
- Volume
- 79
- Issue
- 16
- Pages
- 6174-6181
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W1999836467
- Cited by
- 123
- References
- 1
- FWCI
- 4.21
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: On-Line 1D and 2D Porous Layer Open Tubular/LC-ESI-MS Using 10-μm-i.d. Poly(styrene−divinylbenzene) Columns for Ultrasensitive Proteomic Analysis. https://app.cochat.ai/papers/10.1021/ac070583w
- The paper
- Luo, Q., Yue, G., Valaskovic, G. A., Gu, Y., Wu, S.-L., & Karger, B. L. (2007). On-Line 1D and 2D Porous Layer Open Tubular/LC-ESI-MS Using 10-μm-i.d. Poly(styrene−divinylbenzene) Columns for Ultrasensitive Proteomic Analysis. Analytical Chemistry. https://doi.org/10.1021/ac070583w
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.