MALDI Mass Spectrometry Imaging of Bioactive Lipids in Mouse Brain with a Synapt G2-S Mass Spectrometer Operated at Elevated Pressure: Improving the Analytical Sensitivity and the Lateral Resolution to Ten Micrometers
The DOI matches its Crossref record.
Abstract
Mass spectrometers from the Synapt-G1/G2 family (Waters) are widely employed for matrix-assisted laser desorption/ionization mass spectrometry imaging (MALDI-MSI). A lateral resolution of about 50 μm is typically achieved with these instruments, that is, however, below the often desired cellular resolution. Here, we show the first MALDI-MSI examples demonstrating a lateral resolution of about ten micrometers obtained with a Synapt G2-S HDMS mass spectrometer without oversampling. This improvement became possible by laser beam shaping using a 4:1 beam expander and a circular aperture for spatial mode filtering and by replacement of the default focusing lens. We used dithranol as an effective matrix for imaging of acidic lipids such as sulfatides, gangliosides, and phosphatidylinositols in the negative ion mode. At the same time, the matrix enables MS imaging of more basic lipids in the positive ion mode. Uniform matrix coatings with crystals having average dimensions between 0.5 and 3 μm were obtained upon spraying a chloroform/methanol matrix solution. Increasing the cooling gas pressure in the MALDI ion source after adding an additional gas line was furthermore found to increase the ion abundances of labile lipids such as gangliosides. The combined characteristics are demonstrated with the MALDI-MSI analysis of fine structures in coronal mouse brain slices.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Kompauer, M., Heiles, S., & Spengler, B. (2017) · Nature Methods · doi:10.1038/nmeth.4071 · Cited by 600
Transmission-mode MALDI-2 mass spectrometry imaging of cells and tissues at subcellular resolution
Niehaus, M., Soltwisch, J., Belov, M. E., & Dreisewerd, K. (2019) · Nature Methods · doi:10.1038/s41592-019-0536-2 · Cited by 429
Self-citationMass spectrometry imaging with laser-induced postionization
Soltwisch, J., Kettling, H., Vens-Cappell, S., Wiegelmann, M., Müthing, J., & Dreisewerd, K. (2015) · Science · doi:10.1126/science.aaa1051 · Cited by 382
Self-citationPaglia, G., Angel, P., Williams, J. P., Richardson, K., Olivos, H. J., Thompson, J. W., Menikarachchi, L., Lai, S., Walsh, C., Moseley, A., et al. (2015) · Analytical Chemistry · doi:10.1021/ac503715v · Cited by 295
Mass Spectrometry Imaging of Biomolecular Information
Spengler, B. (2015) · Analytical Chemistry · doi:10.1021/ac504543v · Cited by 290
Körber, A., Anthony, I. G. M., & Heeren, R. M. A. (2025) · Analytical Chemistry · doi:10.1021/acs.analchem.4c05249 · Cited by 88
Harvey, D. J. (2018) · Mass Spectrometry Reviews · doi:10.1002/mas.21530 · Cited by 61
Metadata
- Authors
- Kettling, H., Vens-Cappell, S., Soltwisch, J., Pirkl, A., Haier, J., Müthing, J., & Dreisewerd, K.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2014-07-22
- Volume
- 86
- Issue
- 15
- Pages
- 7798-7805
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2323819752
- Cited by
- 79
- References
- 0
- FWCI
- 4.90
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.