Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrations

Gehring, H., Hornberger, C., Dibbelt, L., Roth‐Isigkeit, A., Gerlach, K., Schumacher, J., Schmucker, P. · 2002 · Acta Anaesthesiologica Scandinavica
Cited by 98

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background: While point‐of‐care testing (POCT) is being used increasingly as a basis for deciding on perioperative erythrocyte transfusion, no valid standards currently exist concerning the accuracy of Hb concentration measurements. For clinical employment, however, the confidence limits (±2 SD) of these measurements should lie close to 5 g/l. The aim of the present study was to evaluate the accuracy and precision of point‐of‐care testing for blood hemoglobin concentration (cHb in g/l) measurements in critically ill patients. Methods: Fifty blood samples from 50 postoperative patients requiring intensive care treatment were withdrawn from a cannula in the radial artery into a 2‐ml heparinized syringe (containing wet sodium heparinate in the conus), in a 2‐ml Monovette with 50 IE lithium heparinate, and into a 2.7‐ml cuvette with 1.6 mg potassium EDTA/ml blood. The POCT battery consisted of two blood gas analyzers (ABL TM 625 and 725, Radiometer, Copenhagen), the HemoCue® system (Mallinckrodt Medical, Germany), and an automated hematology analyzer (M‐2000®, Sysmex, Germany). The cyanmethemoglobin method served as the reference ‘gold standard’ procedure. The blood gas analyzer and HemoCue® systems were tested using dry and wet heparinized blood samples. Results: Hemoglobin concentrations of the reference measurements ranged from 73.9 to 159.4 g/l. The automated hematology analyzer method did reveal a small but systematic deviation for higher cHb values. For the blood gas analyzer and HemoCue® system procedures there was no systematic deviation of bias for either the first measurement or the averaged data. Bland & Altman analysis revealed a larger scattering for the wet heparinized samples. Conclusions: The above‐stated requirement for POCT systems, i.e. that the confidence limits should lie close to 5 g/l cHb, held true for the dry heparinized samples of the blood gas analyzer (1st measurement and mean of 2), the HemoCue® system (mean of 3) and the automated hematology analyzer.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1034/j.1399-6576.2002.460809.x

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 98

Metadata

Authors
Gehring, H., Hornberger, C., Dibbelt, L., Roth‐Isigkeit, A., Gerlach, K., Schumacher, J., & Schmucker, P.
Venue
Acta Anaesthesiologica Scandinavica
Publisher
Wiley
Published
2002-08-21
Volume
46
Issue
8
Pages
980-986
ISSN
  • 0001-5172 (ISSN-L)
  • 1399-6576
OpenAlex
W2051615659
Cited by
98
References
2
FWCI
0.55
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrations. https://app.cochat.ai/papers/10.1034/j.1399-6576.2002.460809.x
The paper
Gehring, H., Hornberger, C., Dibbelt, L., Roth‐Isigkeit, A., Gerlach, K., Schumacher, J., & Schmucker, P. (2002). Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiologica Scandinavica. https://doi.org/10.1034/j.1399-6576.2002.460809.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.