Quality assurance for polychromatic flow cytometry using a suite of calibration beads
Abstract
The quality assurance program presented here provides a means to maximize and maintain the performance of individual flow cytometers in a facility. To optimize performance, we recommend performing all three steps (optimization, calibration and standardization) in this program when a new flow cytometer is installed or whenever the flow cytometer's optical path is altered (e.g., lasers, filters or detectors are replaced). The complete process requires 3-4 h. On a more frequent basis, only a subset of these procedures need to be performed as a part of daily maintenance routines. The data generated can be tracked to monitor the instrument and determine whether service is needed. In addition, the data can provide a metric for whether repairs and upgrades have improved or harmed performance, and for future instrument-to-instrument comparisons. In sum, the procedures presented here represent an updated framework for optimizing, calibrating and standardizing a flow cytometer for daily use.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Voltage titration is the most widely accepted approach to optimize detector voltages and determines the minimally acceptable voltage yielding optimal resolution for each detector.
“The most widely accepted approach for this is a voltage titration [3], which will determine the minimally acceptable voltage yielding optimal resolution for each detector.”
From Guidelines for the use of flow cytometry and cell sorting in immunological studies (third edition) (2021) · cites this paper as [3]
Cites
Shapiro, H. M. (2003) · Wiley eBooks · doi:10.1002/0471722731 · Cited by 2,698 · see citing works
A deep profiler's guide to cytometry
Bendall, S. C., Nolan, G. P., Roederer, M., & Chattopadhyay, P. K. (2012) · Trends in Immunology · doi:10.1016/j.it.2012.02.010 · Cited by 653 · see citing works
Self-citationQuantum dot semiconductor nanocrystals for immunophenotyping by polychromatic flow cytometry
Chattopadhyay, P. K., Price, D. A., Harper, T. F., Betts, M. R., Yu, J., Gostick, E., Perfetto, S. P., Goepfert, P., Koup, R. A., De Rosa, S. C., et al. (2006) · Nature Medicine · doi:10.1038/nm1371 · Cited by 372 · see citing works
Self-citationQuality assurance for polychromatic flow cytometry
Perfetto, S. P., Ambrozak, D. R., Nguyen, R., Chattopadhyay, P. K., & Roederer, M. (2006) · Nature Protocols · doi:10.1038/nprot.2006.250 · Cited by 200 · see citing works
Self-citationCytometry: Today’s technology and tomorrow’s horizons
Chattopadhyay, P. K., & Roederer, M. (2012) · Methods · doi:10.1016/j.ymeth.2012.02.009 · Cited by 134 · see citing works
Self-citationFormalization of the MESF unit of fluorescence intensity
Schwartz, A., Gaigalas, A. K., Wang, L., et al. (2003) · Cytometry Part B Clinical Cytometry · doi:10.1002/cyto.b.10066 · Cited by 131 · see citing works
Performance evaluation of the FACSCount system: A dedicated system for clinical cellular analysis
Strauss, K. W., Hannet, I., Engels, S., Shiba, A. S., Ward, D. M., Ullery, S., Jinguji, M. G., Valinsky, J. E., Barnett, D. B., Orfao, A., et al. (1996) · Cytometry · doi:10.1002/(sici)1097-0320(19960315)26:1 · Cited by 84 · see citing works
Resolution of dimly fluorescent particles: A practical measure of fluorescence sensitivity
Chase, E. S., & Hoffman, R. A. (1998) · Cytometry · doi:10.1002/(sici)1097-0320(19981001)33:2 · Cited by 81 · see citing works
Data quality assessment of ungated flow cytometry data in high throughput experiments
Meur, N. L., Rossini, A. J., Gasparetto, M., et al. (2007) · Cytometry Part A · doi:10.1002/cyto.a.20396 · Cited by 46 · see citing works
Quality Control in Clinical Flow Cytometry
Harvath, L. (2008) · Pathology and Immunopathology Research · doi:10.1159/000157127 · Cited by 14 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Guidelines for the use of flow cytometry and cell sorting in immunological studies (third edition)
Cossarizza, A., Chang, H., Radbruch, A., Abrignani, S., Addo, R., Akdis, M., Andrä, I., Andreata, F., Annunziato, F., Arranz, E., et al. (2021) · European Journal of Immunology · doi:10.1002/eji.202170126 · Cited by 531 · see citing works
Human bone marrow assessment by single-cell RNA sequencing, mass cytometry, and flow cytometry
Oetjen, K. A., Lindblad, K. E., Goswami, M., Gui, G., Dagur, P. K., Lai, C., Dillon, L. W., McCoy, J. P., & Hourigan, C. S. (2018) · JCI Insight · doi:10.1172/jci.insight.124928 · Cited by 272 · see citing works
Lucas, F., Gil‐Pulido, J., LaMacchia, J., Preffer, F., Wallace, P. K., & Lopez, P. (2020) · Cytometry Part A · doi:10.1002/cyto.a.23940 · Cited by 10 · see citing works
- Authors
- Perfetto, S. P., Ambrozak, D., Nguyen, R., Chattopadhyay, P. K., & Roederer, M.
- Affiliations
- National Institutes of Health (US)
- Corresponding author
- Perfetto, Stephen P
- Venue
- Nature Protocols
- Published
- 2012-11-08
- Volume
- 7
- Issue
- 12
- Pages
- 2067-2079
- ISSN
- 1750-2799 (ISSN-L)
- 1754-2189
- OpenAlex
- W2034043217
- Cited by
- 148
- References
- 10
- Citations by year
- 2026: 6 · 2025: 10 · 2024: 10 · 2023: 15 · 2022: 14 · 2021: 15 · 2020: 12 · 2019: 16 · 2018: 8 · 2017: 13 · 2016: 10 · 2015: 8 · 2014: 6 · 2013: 4 · 2012: 1
- FWCI
- 4.08
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Nature Protocols (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (Z99 AI999999)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Health Professions › Medical Laboratory Technology › Quality and Safety in Healthcare
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
- Keywords
- polychromatic flow cytometry, instrument optimization, instrument maintenance, performance monitoring
- MeSH terms
- Calibration, Flow Cytometry, Fluorescent Dyes, Quality Control
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review