Wearable sensors during drawing tasks to measure the severity of essential tremor

Open access

Abstract

Commonly used methods to assess the severity of essential tremor (ET) are based on clinical observation and lack objectivity. This study proposes the use of wearable accelerometer sensors for the quantitative assessment of ET. Acceleration data was recorded by inertial measurement unit (IMU) sensors during sketching of Archimedes spirals in 17 ET participants and 18 healthy controls. IMUs were placed at three points (dorsum of hand, posterior forearm, posterior upper arm) of each participant’s dominant arm. Movement disorder neurologists who were blinded to clinical information scored ET patients on the Fahn–Tolosa–Marin rating scale (FTM) and conducted phenotyping according to the recent Consensus Statement on the Classification of Tremors. The ratio of power spectral density of acceleration data in 4–12 Hz to 0.5–4 Hz bands and the total duration of the action were inputs to a support vector machine that was trained to classify the ET subtype. Regression analysis was performed to determine the relationship of acceleration and temporal data with the FTM scores. The results show that the sensor located on the forearm had the best classification and regression results, with accuracy of 85.71% for binary classification of ET versus control. There was a moderate to good correlation (r2 = 0.561) between FTM and a combination of power spectral density ratio and task time. However, the system could not accurately differentiate ET phenotypes according to the Consensus classification scheme. Potential applications of machine-based assessment of ET using wearable sensors include clinical trials and remote monitoring of patients.

Overview

CoChat Summary

TL;DR

This study used wearable IMU accelerometers during Archimedes spiral drawing to quantify essential tremor severity and classify ET versus controls. Forearm sensor data combined with task time correlated with clinical FTM scores and classified ET vs control with 85.71% accuracy, but ET subtype classification was poor.

Key findings

  • The forearm sensor (Sensor-2) gave the best results: binary ET versus control classification accuracy was 85.71%, and PSD ratio versus FTM regression had r² = 0.50; Sensor-1 and Sensor-3 had r² = 0.3855 and 0.4184, respectively.
  • Combining PSD ratio and task time estimated clinical FTM scores with r² = 0.561 and RMSE = 8.158.
  • ET participants had higher mean PSD ratios at all three sensor locations and longer mean task times; Mann–Whitney U testing showed significant ET-control differences for Sensor-2 PSD ratio (U = 25.00, p = 0.00) and task time (U = 52.00, p = 0.001).
  • The 4-class SVM separating controls, ET − 0, ET + 1, and ET + 2 achieved only 57.14% accuracy using time plus PSD ratio, so the system could not accurately differentiate ET phenotypes.

Methods

  • Seventeen ET participants and 18 healthy controls had analysable recordings; ET participants were recruited from Monash Health, with mean age 67.2 ± 13.0 years and mean tremor duration 21.7 ± 19.0 years. Two ET and two control participants were excluded for noisy or incomplete IMU data.
  • IMU sensors were placed at three points on the dominant arm: dorsum of hand, mid-forearm, and posterior upper arm. Accelerometer data were recorded during drawing of a predefined Archimedes spiral on a digital tablet; vector magnitude was derived, and the PSD ratio of 4–12 Hz to 0.5–4 Hz plus task time T were used as features.
  • Blinded movement disorder neurologists scored FTM and classified ET/ET plus per the Consensus Statement. Analysis used Mann–Whitney U tests, Kruskal–Wallis tests, linear regression, and SVM with leave-one-out cross-validation in Python 3.8.

Limitations

  • Stated in the paper: each participant was investigated only once, so repeatability was not tested; the sample size was not large enough to perform a detailed analysis of differences within ET subgroups.
  • Evident from the results: the system’s ability to separate ET phenotypes was poor, with only 57.14% accuracy for the 4-class classification, and the paper notes this may reflect the Consensus classification’s emphasis on clinical features other than tremor.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1038/s41598-022-08922-6

Cites

63 of 63 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 a Boxplot of the PSD ratio for ET and controls at the three sensing locations. ETS1—ET Sensor—1, ETS2—ET Sensor—2, ETS3—ET Sensor—3; and for Controls, CS1—Control Sensor—1, CS2—Control Sensor—2, CS3—Control Sensor—3 . (b) Shows the placement of the sensor location, S1, S2 and S3. (c) Boxplot of the Task Time of ET and controls at the sensing locations.
Figure 1 a Boxplot of the PSD ratio for ET and controls at the three sensing locations. ETS1—ET Sensor—1, ETS2—ET Sensor—2, ETS3—ET Sensor—3; and for Controls, CS1—Control Sensor—1, CS2—Control Sensor—2, CS3—Control Sensor—3 . (b) Shows the placement of the sensor location, S1, S2 and S3. (c) Boxplot of the Task Time of ET and controls at the sensing locations.CC BY · PubMed Central
Figure 2 Regression analysis of the PSD ratio versus FTM score at three sensing locations.
Figure 2 Regression analysis of the PSD ratio versus FTM score at three sensing locations.CC BY · PubMed Central
Figure 3 (a) Shows the 3D scatter plot of the FTM, PSD Ratio and Task Time representing the groups of ET − 0 (yellow), ET + 1 (blue) and ET + 2 (red), (b) Shows the 3D regression plot of the regression Eq. (1).
Figure 3 (a) Shows the 3D scatter plot of the FTM, PSD Ratio and Task Time representing the groups of ET − 0 (yellow), ET + 1 (blue) and ET + 2 (red), (b) Shows the 3D regression plot of the regression Eq. (1).CC BY · PubMed Central
Figure 4 Shows the ET Phenotype of the clinical and estimated FTM scores.
Figure 4 Shows the ET Phenotype of the clinical and estimated FTM scores.CC BY · PubMed Central
Figure 5 Participant performing the spiral drawing task with the wearable sensors mounted on the upper limb.
Figure 5 Participant performing the spiral drawing task with the wearable sensors mounted on the upper limb.CC BY · PubMed Central
Figure 6 Sketch of the spiral drawing task performed by both control and ET patient.
Figure 6 Sketch of the spiral drawing task performed by both control and ET patient.CC BY · PubMed Central

Verification

Claims the paper makes

9 Valid · 0 Not valid · 16 Not checked · 44 Own results and interpretations

28 more claims · 2 with their cited sources checked

Create a free account to see all 69 claims and to verify the 27 references still open.

See all claims — free account
  • It may state two separate claims as one. 20
  • Probably not a substantial claim. 22

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 63 references

Research works printed but not linked: 1

  1. Sztyler, T. & Stuckenschmidt, H. in 2016 IEEE International Conference on Pervasive Computing and Communications (PerCom). 1–9.

Cited by

See all 35

Metadata

Authors
Ali, S. M., Arjunan, S. P., Peters, J., Perju-Dumbrava, L., Ding, C., Eller, M., Raghav, S., Kempster, P., Motin, M. A., Radcliffe, P. J., & Kumar, D. K.
Affiliations
  • RMIT University (AU)
  • SRM Institute of Science and Technology (IN)
  • Monash Health (AU)
Corresponding author
Kumar, Dinesh Kant
Venue
Scientific Reports
Published
2022-03-28
Volume
12
Issue
1
Pages
5242-5242
ISSN
2045-2322 (ISSN-L)
OpenAlex
W4220988247
Cited by
35
References
63
Citations by year
2026: 5 · 2025: 5 · 2024: 10 · 2023: 13 · 2022: 2
FWCI
3.95
Type
journal article
Language
en
Open access
Licence
Locations
  • Scientific Reports (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • RMIT Research Repository (RMIT University Library) (submittedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
  • Figshare (submittedVersion, cc-by)
Topics
  • Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
Keywords
essential tremor, wearable sensors, inertial measurement unit, accelerometer, tremor severity assessment, tremor classification, power spectral density, drawing tasks, remote patient monitoring
MeSH terms
Acceleration, Wearable Electronic Devices, Hand, Humans, Tremor, Essential Tremor

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-04, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-04 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (qwen3.8-omni-flash and gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Wearable sensors during drawing tasks to measure the severity of essential tremor. https://app.cochat.ai/papers/10.1038/s41598-022-08922-6
The paper
Ali, S. M., Arjunan, S. P., Peters, J., Perju-Dumbrava, L., Ding, C., Eller, M., Raghav, S., Kempster, P., Motin, M. A., Radcliffe, P. J., & Kumar, D. K. (2022). Wearable sensors during drawing tasks to measure the severity of essential tremor. Scientific Reports. https://doi.org/10.1038/s41598-022-08922-6

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.