Measurement of cerebral oxygenation state in anaesthetized children using the INVOS 5100 cerebral oximeter

Dullenkopf, Alexander, Frey, Bernhard, Baenziger, Oskar, Gerber, Andreas, Weiss, Markus · 2003 · Pediatric Anesthesia
Cited by 127

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Summary: Background: Near‐infrared spectroscopy is a developing technique for monitoring cerebral oxygenation during anaesthesia. The aim of this study was to evaluate absolute values of cerebral oxygenation during stable anaesthesia conditions in otherwise healthy children using the recently introduced INVOS 5100 cerebral oximeter with a paediatric and adult sensor and to compare them with values obtained from the NIRO 300 oximeter. Methods: Thirty paediatric surgical patients (aged 0.23–15.97 years) were studied during general anaesthesia with tracheal intubation and controlled ventilation. Comparative measurements of cerebral oxygenation were performed on the forehead with two probes within 10 min under stable cardiorespiratory and anaesthesia conditions. Cerebral oxygenation values (rSO 2) obtained from the paediatric and adult INVOS 5100 sensors were compared with the tissue oxygenation index (TOI) obtained from the NIRO 300 cerebral oximeter using 4‐ and 5‐cm emitter‐detector separation. Results: Cerebral rSO 2 values and the TOI values both showed a large range of cerebral oxygenation in the children studied (rSO 2: 59–95%, TOI: 48–85%). Cerebral rSO 2 values measured by the INVOS 5100, particularly with the paediatric sensor, were significantly higher than the TOI values obtained from the NIRO 300 (P < 0.0001). Agreement between the INVOS and NIRO oximeter was poor. Conclusion: The large range and the poor agreement of cerebral oxygenation values between the two oximeters makes it difficult to define a normal value. Cerebral oxygenation readings by these monitors, based on one single point measurement during anaesthesia, should be viewed with caution. Actually, there may be little indication for routine use of such monitoring during general anaesthesia.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1046/j.1460-9592.2003.01111.x

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 127

Metadata

Authors
Dullenkopf, A., Frey, B., Baenziger, O., Gerber, A., & Weiss, M.
Venue
Pediatric Anesthesia
Publisher
Wiley
Published
2003-06
Volume
13
Issue
5
Pages
384-391
ISSN
  • 1155-5645 (ISSN-L)
  • 1460-9592
OpenAlex
W2086325155
Cited by
127
References
2
FWCI
6.72
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.