An Entirely Subcutaneous Implantable Cardioverter–Defibrillator
Open accessAbstract
BACKGROUND: Implantable cardioverter-defibrillators (ICDs) prevent sudden death from cardiac causes in selected patients but require the use of transvenous lead systems. To eliminate the need for venous access, we designed and tested an entirely subcutaneous ICD system. METHODS: First, we conducted two short-term clinical trials to identify a suitable device configuration and assess energy requirements. We evaluated four subcutaneous ICD configurations in 78 patients who were candidates for ICD implantation and subsequently tested the best configuration in 49 additional patients to determine the subcutaneous defibrillation threshold in comparison with that of the standard transvenous ICD. Then we evaluated the long-term use of subcutaneous ICDs in a pilot study, involving 6 patients, which was followed by a trial involving 55 patients. RESULTS: The best device configuration consisted of a parasternal electrode and a left lateral thoracic pulse generator. This configuration was as effective as a transvenous ICD for terminating induced ventricular fibrillation, albeit with a significantly higher mean (+/-SD) energy requirement (36.6+/-19.8 J vs. 11.1+/-8.5 J). Among patients who received a permanent subcutaneous ICD, ventricular fibrillation was successfully detected in 100% of 137 induced episodes. Induced ventricular fibrillation was converted twice in 58 of 59 patients (98%) with the delivery of 65-J shocks in two consecutive tests. Clinically significant adverse events included two pocket infections and four lead revisions. After a mean of 10+/-1 months, the device had successfully detected and treated all 12 episodes of spontaneous, sustained ventricular tachyarrhythmia. CONCLUSIONS: In small, nonrandomized studies, an entirely subcutaneous ICD consistently detected and converted ventricular fibrillation induced during electrophysiological testing. The device also successfully detected and treated all 12 episodes of spontaneous, sustained ventricular tachyarrhythmia. (ClinicalTrials.gov numbers, NCT00399217 and NCT00853645.)
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart Failure
Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., Domanski, M., Troutman, C., Anderson, J., Johnson, G., et al. (2005) · New England Journal of Medicine · doi:10.1056/nejmoa043399 · Cited by 6,552
Self-citationThe Sprint Fidelis lead fracture story: What do we really know and where do we go from here?
Ellenbogen, K. A., Wood, M. A., & Swerdlow, C. D. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.08.001 · Cited by 19
Cited by
McDonagh, T. A., Metra, M., Adamo, M., Gardner, R. S., Baumbach, A., Böhm, M., Burri, H., Butler, J., Čelutkienė, J., Chioncel, O., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab670 · Cited by 19,442
2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure
Ponikowski, P., Voors, A. A., Anker, S. D., Bueno, H., Cleland, J. G. F., Coats, A. J. S., Falk, V., González-Juanatey, J. R., Harjola, V.-P., Jankowska, E. A., et al. (2016) · European Heart Journal · doi:10.1093/eurheartj/ehw128 · Cited by 11,429
2016 ESC Guidelines for the Diagnosis and Treatment of Acute and Chronic Heart Failure
Ponikowski, P., Voors, A. A., Anker, S. D., Bueno, H., Cleland, J. G. F., Coats, A. J. S., Falk, V., González-Juanatey, J. R., Harjola, V.-P., Jankowska, E. A., et al. (2016) · European Journal of Heart Failure · doi:10.1002/ejhf.592 · Cited by 5,866
2014 ESC Guidelines on diagnosis and management of hypertrophic cardiomyopathy
Authors/Task Force members, Elliott, P. M., Anastasakis, A., Borger, M. A., Borggrefe, M., Cecchi, F., Charron, P., Hagege, A. A., Lafont, A., Limongelli, G., et al. (2014) · European Heart Journal · doi:10.1093/eurheartj/ehu284 · Cited by 4,332
Priori, S. G., Blomström-Lundqvist, C., Mazzanti, A., Blom, N., Borggrefe, M., Camm, J., Elliott, P. M., Fitzsimons, D., Hatala, R., Hindricks, G., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv316 · Cited by 3,893
Metadata
- Authors
- Bardy, G. H., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R. E., Wright, D. J., Connelly, D. T., Fynn, S. P., Murgatroyd, F. D., Sperzel, J., Neuzner, J., Spitzer, S. G., Ardashev, A. V., Oduro, A., Boersma, L., Maass, A. H., Van Gelder, I. C., Wilde, A. A., van Dessel, P. F., Knops, R. E., Barr, C. S., Lupo, P., Cappato, R., & Grace, A. A.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2010-07
- Volume
- 363
- Issue
- 1
- Pages
- 36-44
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2139305230
- Cited by
- 780
- References
- 2
- FWCI
- 30.86
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An Entirely Subcutaneous Implantable Cardioverter–Defibrillator. https://app.cochat.ai/papers/10.1056/nejmoa0909545
- The paper
- Bardy, G. H., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R. E., Wright, D. J., Connelly, D. T., Fynn, S. P., Murgatroyd, F. D., Sperzel, J., Neuzner, J., Spitzer, S. G., Ardashev, A. V., Oduro, A., Boersma, L., Maass, A. H., . . . Grace, A. A. (2010). An Entirely Subcutaneous Implantable Cardioverter–Defibrillator. New England Journal of Medicine. https://doi.org/10.1056/nejmoa0909545
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.