The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
A Leadless Intracardiac Transcatheter Pacing System
Open accessAbstract
BACKGROUND: A leadless intracardiac transcatheter pacing system has been designed to avoid the need for a pacemaker pocket and transvenous lead. METHODS: In a prospective multicenter study without controls, a transcatheter pacemaker was implanted in patients who had guideline-based indications for ventricular pacing. The analysis of the primary end points began when 300 patients reached 6 months of follow-up. The primary safety end point was freedom from system-related or procedure-related major complications. The primary efficacy end point was the percentage of patients with low and stable pacing capture thresholds at 6 months (≤2.0 V at a pulse width of 0.24 msec and an increase of ≤1.5 V from the time of implantation). The safety and efficacy end points were evaluated against performance goals (based on historical data) of 83% and 80%, respectively. We also performed a post hoc analysis in which the rates of major complications were compared with those in a control cohort of 2667 patients with transvenous pacemakers from six previously published studies. RESULTS: The device was successfully implanted in 719 of 725 patients (99.2%). The Kaplan-Meier estimate of the rate of the primary safety end point was 96.0% (95% confidence interval [CI], 93.9 to 97.3; P<0.001 for the comparison with the safety performance goal of 83%); there were 28 major complications in 25 of 725 patients, and no dislodgements. The rate of the primary efficacy end point was 98.3% (95% CI, 96.1 to 99.5; P<0.001 for the comparison with the efficacy performance goal of 80%) among 292 of 297 patients with paired 6-month data. Although there were 28 major complications in 25 patients, patients with transcatheter pacemakers had significantly fewer major complications than did the control patients (hazard ratio, 0.49; 95% CI, 0.33 to 0.75; P=0.001). CONCLUSIONS: In this historical comparison study, the transcatheter pacemaker met the prespecified safety and efficacy goals; it had a safety profile similar to that of a transvenous system while providing low and stable pacing thresholds. (Funded by Medtronic; Micra Transcatheter Pacing Study ClinicalTrials.gov number, NCT02004873.).
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Proportional Hazards Model for the Subdistribution of a Competing Risk
Fine, J. P., & Gray, R. J. (1999) · Journal of the American Statistical Association · doi:10.1080/01621459.1999.10474144 · Cited by 14,290
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546
Kirkfeldt, R. E., Johansen, J. B., Nohr, E. A., Jorgensen, O. D., & Nielsen, J. C. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht511 · Cited by 871
Cited by
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
Self-citationKusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.044 · Cited by 1,008
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Self-citationKusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Circulation · doi:10.1161/cir.0000000000000628 · Cited by 603
Blomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · EP Europace · doi:10.1093/europace/euz246 · Cited by 423
Self-citation
Metadata
- Authors
- Reynolds, D., Duray, G. Z., Omar, R., Soejima, K., Neuzil, P., Zhang, S., Narasimhan, C., Steinwender, C., Brugada, J., Lloyd, M., Roberts, P. R., Sagi, V., Hummel, J., Bongiorni, M. G., Knops, R. E., Ellis, C. R., Gornick, C. C., Bernabei, M. A., Laager, V., Stromberg, K., Williams, E. R., Hudnall, J. H., & Ritter, P.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2016-02-11
- Volume
- 374
- Issue
- 6
- Pages
- 533-541
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2178696728
- Cited by
- 933
- References
- 3
- FWCI
- 57.55
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Leadless Intracardiac Transcatheter Pacing System. https://app.cochat.ai/papers/10.1056/nejmoa1511643
- The paper
- Reynolds, D., Duray, G. Z., Omar, R., Soejima, K., Neuzil, P., Zhang, S., Narasimhan, C., Steinwender, C., Brugada, J., Lloyd, M., Roberts, P. R., Sagi, V., Hummel, J., Bongiorni, M. G., Knops, R. E., Ellis, C. R., Gornick, C. C., Bernabei, M. A., Laager, V., . . . Ritter, P. (2016). A Leadless Intracardiac Transcatheter Pacing System. New England Journal of Medicine. https://doi.org/10.1056/nejmoa1511643
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.