Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator
Open accessAbstract
BACKGROUND: The presence of a cardiovascular implantable electronic device has long been a contraindication for the performance of magnetic resonance imaging (MRI). We established a prospective registry to determine the risks associated with MRI at a magnetic field strength of 1.5 tesla for patients who had a pacemaker or implantable cardioverter-defibrillator (ICD) that was "non-MRI-conditional" (i.e., not approved by the Food and Drug Administration for MRI scanning). METHODS: Patients in the registry were referred for clinically indicated nonthoracic MRI at a field strength of 1.5 tesla. Devices were interrogated before and after MRI with the use of a standardized protocol and were appropriately reprogrammed before the scanning. The primary end points were death, generator or lead failure, induced arrhythmia, loss of capture, or electrical reset during the scanning. The secondary end points were changes in device settings. RESULTS: MRI was performed in 1000 cases in which patients had a pacemaker and in 500 cases in which patients had an ICD. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred during MRI. One ICD generator could not be interrogated after MRI and required immediate replacement; the device had not been appropriately programmed per protocol before the MRI. We observed six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset. Changes in lead impedance, pacing threshold, battery voltage, and P-wave and R-wave amplitude exceeded prespecified thresholds in a small number of cases. Repeat MRI was not associated with an increase in adverse events. CONCLUSIONS: In this study, device or lead failure did not occur in any patient with a non-MRI-conditional pacemaker or ICD who underwent clinically indicated nonthoracic MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in accordance with the prespecified protocol. (Funded by St. Jude Medical and others; MagnaSafe ClinicalTrials.gov number, NCT00907361 .).
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Poole, J. E., Gleva, M. J., Mela, T., Chung, M. K., Uslan, D. Z., Borge, R., Gottipaty, V., Shinn, T., Dan, D., Feldman, L. A., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.976076 · Cited by 790
Lead Extraction in the Contemporary Setting: The LExICon Study
Wazni, O., Epstein, L. M., Carrillo, R. G., Love, C., Adler, S. W., Riggio, D. W., Karim, S. S., Bashir, J., Greenspon, A. J., DiMarco, J. P., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.08.070 · Cited by 584
Wilkoff, B. L., Byrd, C. L., Love, C. J., Hayes, D. L., Sellers, T. D., Schaerf, R., Parsonnet, V., Epstein, L. M., Sorrentino, R. A., & Reiser, C. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00074-1 · Cited by 438
Self-citationMagnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla
Martin, E. T., Coman, J. A., Shellock, F. G., Pulling, C. C., Fair, R., & Jenkins, K. (2004) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2003.12.016 · Cited by 422
Self-citationCurrent Clinical Issues for MRI Scanning of Pacemaker and Defibrillator Patients
KALIN, R., & STANTON, M. S. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50024.x · Cited by 381
Nazarian, S., Roguin, A., Zviman, M. M., Lardo, A. C., Dickfeld, T. L., Calkins, H., Weiss, R. G., Berger, R. D., Bluemke, D. A., & Halperin, H. R. (2006) · Circulation · doi:10.1161/circulationaha.105.607655 · Cited by 356
Roguin, A., Zviman, M. M., Meininger, G. R., Rodrigues, E. R., Dickfeld, T. M., Bluemke, D. A., Lardo, A., Berger, R. D., Calkins, H., & Halperin, H. R. (2004) · Circulation · doi:10.1161/01.cir.0000137121.28722.33 · Cited by 355
Pacemaker and ICD Generator Malfunctions: Analysis of Food and Drug Administration Annual Reports
Maisel, W. H., Moynahan, M., Zuckerman, B. D., Gross, T. P., Tovar, O. H., Tillman, D.-B., & Schultz, D. B. (2006) · JAMA · doi:10.1001/jama.295.16.1901 · Cited by 353
Nazarian, S., Hansford, R., Roguin, A., Goldsher, D., Zviman, M. M., Lardo, A. C., Caffo, B. S., Frick, K. D., Kraut, M. A., Kamel, I. R., et al. (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-7-201110040-00004 · Cited by 302
Wilkoff, B. L., Bello, D., Taborsky, M., Vymazal, J., Kanal, E., Heuer, H., Hecking, K., Johnson, W. B., Young, W., Ramza, B., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.10.002 · Cited by 266
Cited by
2023 ESC guidelines for the management of cardiomyopathies
Arbelo, E., Protonotarios, A., Gimeno, J. R., Arbustini, E., Barriales-Villa, R., Basso, C., Bezzina, C. R., Biagini, E., Blom, N. A., de Boer, R. A., et al. (2023) · European Heart Journal · doi:10.1093/eurheartj/ehad194 · Cited by 2,844
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
Self-citationKusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R., Cha, Y.-M., Clancy, J., Deharo, J.-C., Ellenbogen, K. A., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.09.001 · Cited by 1,199
Self-citation2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Self-citationIndik, J. H., Gimbel, J. R., Abe, H., Alkmim-Teixeira, R., Birgersdotter-Green, U., Clarke, G. D., Dickfeld, T.-M. L., Froelich, J. W., Grant, J., Hayes, D. L., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.04.025 · Cited by 438
Self-citation
Metadata
- Authors
- Russo, R. J., Costa, H. S., Silva, P. D., Anderson, J. L., Arshad, A., Biederman, R. W. W., Boyle, N. G., Frabizzio, J. V., Birgersdotter-Green, U., Higgins, S. L., Lampert, R., Machado, C. E., Martin, E. T., Rivard, A. L., Rubenstein, J. C., Schaerf, R. H. M., Schwartz, J. D., Shah, D. J., Tomassoni, G. F., Tominaga, G. T., Tonkin, A. E., Uretsky, S., & Wolff, S. D.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2017-02-23
- Volume
- 376
- Issue
- 8
- Pages
- 755-764
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2588043918
- Cited by
- 373
- References
- 30
- FWCI
- 34.25
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. https://app.cochat.ai/papers/10.1056/nejmoa1603265
- The paper
- Russo, R. J., Costa, H. S., Silva, P. D., Anderson, J. L., Arshad, A., Biederman, R. W. W., Boyle, N. G., Frabizzio, J. V., Birgersdotter-Green, U., Higgins, S. L., Lampert, R., Machado, C. E., Martin, E. T., Rivard, A. L., Rubenstein, J. C., Schaerf, R. H. M., Schwartz, J. D., Shah, D. J., Tomassoni, G. F., . . . Wolff, S. D. (2017). Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. New England Journal of Medicine. https://doi.org/10.1056/nejmoa1603265
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.