Steam sterilisation’s energy and water footprint
The DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
Objective: The aim of the present study was to quantify hospital steam steriliser resource consumption to provide baseline environmental data and identify possible efficiency gains. We sought to find the amount of steriliser electricity and water used for active cycles and for idling (standby), and the relationship between the electricity and water consumption and the mass and type of items sterilised. Methods: We logged a hospital steam steriliser’s electricity and water meters every 5 min for up to 1 year. We obtained details of all active cycles (standard 134°C and accessory or ‘test’ cycles), recording item masses and types. Relationships were investigated for both the weight and type of items sterilised with electricity and water consumption. Results: Over 304 days there were 2173 active cycles, including 1343 standard 134°C cycles that had an average load mass of 21.2 kg, with 32% of cycles <15 kg. Electricity used for active cycles was 32 652 kWh (60% of total), whereas the water used was 1 243 495 L (79%). Standby used 21 457 kWh (40%) electricity and 329 200 L (21%) water. Total electricity and water consumption per mass sterilised was 1.9 kWh kg–1 and 58 L kg–1, respectively. The linear regression model predicting electricity use was: kWh = 15.7+ 0.14 × mass (in kg; R2 = 0.58, P < 0.01). Models for water and item type were poor. Electricity and water use fell from 3 kWh kg–1 and 200 L kg–1, respectively, for 5-kg loads to 0.5 kWh kg–1 and 20 L kg–1, respectively, for 40-kg loads. Conclusions: Considerable electricity and water use occurred during standby, load mass was only moderately predictive of electricity consumption and light loads were common yet inefficient. The findings of the present study are a baseline for steam sterilisation’s environmental footprint and identify areas to improve efficiencies. What is known about the topic?: There is increasing interest in the environmental effects of healthcare. Life cycle assessment (‘cradle to grave’) provides a scientific method of analysing environmental effects. Although data of the effects of steam sterilisation are integral to the life cycles of reusable items and procedures using such items, there are few data available. Further, there is scant information regarding the efficiency of the long-term in-hospital use of sterilisers. What does this paper add?: We quantified, for the first time, long-term electricity and water use of a hospital steam steriliser. We provide useful input data for future life cycle assessments of all reusable, steam-sterilised equipment. Further, we identified opportunities for improved steriliser efficiencies, including rotating off idle sterilisers and reducing the number of light steriliser loads. Finally, others could use our methods to examine steam sterilisers and many other energy-intensive items of hospital equipment. What are the implications for practitioners?: We provide useful input data for all researchers examining the environmental footprint of reusable hospital equipment and procedures using such equipment. As a result of the present study, staff in the hospital sterile supply department have reduced steam steriliser electricity and water use considerably without impeding sterilisation throughput (and reduced time inefficiencies). Many other hospitals could benefit from similar methods to improve steam steriliser and other hospital equipment efficiencies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Managing the health effects of climate change
Costello, A., Abbas, M., Allen, A., Ball, S., Bell, S., Bellamy, R., Friel, S., Groce, N., Johnson, A., Kett, M., et al. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60935-1 · Cited by 3,102
Thiel, C. L., Eckelman, M., Guido, R., Huddleston, M., Landis, A. E., Sherman, J., Shrake, S. O., Copley-Woods, N., & Bilec, M. M. (2015) · Environmental Science & Technology · doi:10.1021/es504719g · Cited by 416
A Comparison of Reusable and Disposable Perioperative Textiles: Sustainability State-of-the-Art 2012
Overcash, M. (2012) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31824d9cc3 · Cited by 240
The carbon footprint of cataract surgery
Morris, D. S., Wright, T., Somner, J. E. A., & Connor, A. (2013) · Eye · doi:10.1038/eye.2013.9 · Cited by 238
Strategies to reduce greenhouse gas emissions from laparoscopic surgery
Thiel, C. L., Woods, N. C., & Bilec, M. M. (2018) · American Journal of Public Health · doi:10.2105/ajph.2018.304397 · Cited by 222
Life cycle assessment perspectives on delivering an infant in the US
Campion, N., Thiel, C. L., DeBlois, J., Woods, N. C., Landis, A. E., & Bilec, M. M. (2012) · Science of The Total Environment · doi:10.1016/j.scitotenv.2012.03.006 · Cited by 144
A life cycle assessment of reusable and single-use central venous catheter insertion kits
McGain, F., McAlister, S., McGavin, A., & Story, D. (2012) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31824e9b69 · Cited by 140
Self-citationCarbon footprint of robotically‐assisted laparoscopy, laparoscopy and laparotomy: a comparison
Woods, D. L., McAndrew, T., Nevadunsky, N., Hou, J. Y., Goldberg, G., Yi‐Shin Kuo, D., & Isani, S. (2015) · The International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.1640 · Cited by 129
Ecological responsibility in anesthesia practice
Sherman, J. D., & Ryan, S. (2010) · International Anesthesiology Clinics · doi:10.1097/aia.0b013e3181ea7587 · Cited by 67
Cited by
MacNeill, A. J., Lillywhite, R., & Brown, C. J. (2017) · The Lancet Planetary Health · doi:10.1016/s2542-5196(17)30162-6 · Cited by 820
Environmental sustainability in anaesthesia and critical care
McGain, F., Muret, J., Lawson, C., & Sherman, J. D. (2020) · British Journal of Anaesthesia · doi:10.1016/j.bja.2020.06.055 · Cited by 455
Self-citationStrategies to reduce greenhouse gas emissions from laparoscopic surgery
Thiel, C. L., Woods, N. C., & Bilec, M. M. (2018) · American Journal of Public Health · doi:10.2105/ajph.2018.304397 · Cited by 222
Financial and environmental costs of reusable and single-use anaesthetic equipment
McGain, F., Story, D., Lim, T., & McAlister, S. (2017) · British Journal of Anaesthesia · doi:10.1093/bja/aex098 · Cited by 206
“15 kWh kg 1 and 40 litres of water kg (1).(24) All reusable anaesthetic items were washed as an ‘anaesthetic load’ for 10 min at 80 C.”
BackgroundSelf-citationCarbon footprint of general, regional, and combined anesthesia for total knee replacements
McGain, F., Sheridan, N., Wickramarachchi, K., Yates, S., Chan, B., & McAlister, S. (2021) · Anesthesiology · doi:10.1097/aln.0000000000003967 · Cited by 144
“This uncertainty routinely cannot be derived directly from the available information, so a standard procedure was developed to derive uncertainty factors from a qualitative assessment of the data, known as the Pedigree Matrix.27 The Pedigree Matrix is a commonly used qualitative scoring system derived from the seconda…”
BackgroundMethodInfluentialSelf-citation
Metadata
- Authors
- McGain, F., Moore, G., & Black, J.
- Venue
- Australian Health Review
- Publisher
- CSIRO Publishing
- Published
- 2016-04-14
- Volume
- 41
- Issue
- 1
- Pages
- 26-32
- ISSN
- 0156-5788 (ISSN-L)
- 1449-8944
- OpenAlex
- W2270837100
- Cited by
- 61
- References
- 9
- FWCI
- 1.98
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Steam sterilisation’s energy and water footprint. https://app.cochat.ai/papers/10.1071/ah15142
- The paper
- McGain, F., Moore, G., & Black, J. (2017). Steam sterilisation’s energy and water footprint. Australian Health Review. https://doi.org/10.1071/ah15142
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.