Validation of the Fitbit One, Garmin Vivofit and Jawbone UP activity tracker in estimation of energy expenditure during treadmill walking and running
The DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
Objectives: To determine the validity of energy expenditure estimation made by the Fitbit One, Garmin Vivofit and Jawbone UP activity trackers during treadmill walking and running. Determining validity of such trackers will inform the interpretation of the data they generate.Design: Cross-sectional study.Method: Fourteen adults walked at 0.70, 1.25, 1.80 ms−1 and ran at 2.22, 2.78, 3.33 ms−1 on a treadmill wearing a Fitbit One, Garmin Vivofit and Jawbone UP. Estimation of energy expenditure from each tracker was compared to measurement from indirect calorimetry (criterion). Paired t-tests, correlation coefficients and Bland–Altman plots assessed agreement and proportional bias. Mean percentage difference assessed magnitude of difference between estimated and criterion energy expenditure for each speed.Results: Energy expenditure estimates from the Fitbit One and Garmin Vivofit correlated significantly (p< 0.01; r= 0.702; 0.854) with criterion across all gait speeds (0.70–3.33 ms−1). Fitbit One, Garmin Vivofit and Jawbone UP correlated significantly (p < 0.05; r = 0.729; 0.711; 0.591) with criterion across all walking speeds (0.70–1.80 ms−1). However, only the Garmin Vivofit correlated significantly (p< 0.05; r = 0.346) with energy expenditure estimations from criterion across running speeds (2.22–3.33 ms−1). Bland–Altman plots showed proportional bias for the Fitbit One and Garmin Vivofit. Energy expenditure estimations of single speeds were overestimated by the Fitbit One and underestimated by the Garmin Vivofit.Conclusions: Energy expenditure reported by the devices distinguished between walking and running, with a general increase as exercise intensity increased. However, the reported energy expenditure from these devices should be interpreted with caution, given their potential bias and error.Practical implicationsAlthough devices report the same outcome of EE estimation, they are not equivalent to each other and differ from criterion measurements during walking and running.These devices are not suitable as research measurement tools for recording precise and accurate EE estimates but may be suitable for use in interventions of behaviour change as they provide feedback to user on trends in energy expenditure.If intending to use these devices in studies where precise measurements of energy expenditure are required, researchers need to undertake specific validation and reliability studies prior to interventions and the collection of cross-sectional data.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
An analysis of variance test for normality (complete samples)
SHAPIRO, S. S., & WILK, M. B. (1965) · Biometrika · doi:10.1093/biomet/52.3-4.591 · Cited by 23,480
An Analysis of Variance Test for Normality (Complete Samples)
Shapiro, S. S., & Wilk, M. B. (1965) · Biometrika · doi:10.2307/2333709 · Cited by 23,469
Conducting accelerometer-based activity assessments in field-based research
TROST, S. G., MCIVER, K. L., & PATE, R. R. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000185657.86065.98 · Cited by 2,033
Randomized Trial of a Fitbit-Based Physical Activity Intervention for Women
Cadmus-Bertram, L. A., Marcus, B. H., Patterson, R. E., Parker, B. A., & Morey, B. L. (2015) · American Journal of Preventive Medicine · doi:10.1016/j.amepre.2015.01.020 · Cited by 469 · cited 1×
“and sedentary behaviour interventions.[8,9] Pilot and”
Ferguson, T., Rowlands, A. V., Olds, T., & Maher, C. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0201-9 · Cited by 468 · cited 4×
“Exaggerated hip movement at slow walking speed may have contributed to higher values found in the Fitbit One.”
BackgroundResultInfluentialFitbit®: An accurate and reliable device for wireless physical activity tracking
Diaz, K. M., Krupka, D. J., Chang, M. J., Peacock, J., Ma, Y., Goldsmith, J., Schwartz, J. E., & Davidson, K. W. (2015) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.03.038 · Cited by 405
Validity of Consumer-Based Physical Activity Monitors
LEE, J.-M., KIM, Y., & WELK, G. J. (2014) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000287 · Cited by 401
Estimating energy expenditure using accelerometers
Crouter, S. E., Churilla, J. R., & Bassett, D. R. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0307-5 · Cited by 366
Accuracy and reliability of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 metabolic systems
Crouter, S. E., Antczak, A., Hudak, J. R., DellaValle, D. M., & Haas, J. D. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0255-0 · Cited by 240 · cited 1×
“The TrueOne 2400 has demonstrated reliable results for the measurement of VE, VO2, and VCO2 compared to the Douglas Bag method at work rates ranging from rest through 50 W–250 W.[13,14] The TrueOne 2400 flowmeter was calibrated using a 3-L syringe.”
MethodLyden, K., Kozey, S. L., Staudenmeyer, J. W., & Freedson, P. S. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1639-8 · Cited by 222 · cited 2×
“[16] No one equation is able to accurately predict energy expenditure across a wide range of activities [21] and current prediction techniques only perform well within a specified range of intensities and/or specific activity types.[4] However, several new activity tracker models have combined physiological measures,…”
BackgroundAdam Noah, J., Spierer, D. K., Gu, J., & Bronner, S. (2013) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2013.831135 · Cited by 202 · cited 1×
“97) for kilocalories expended.[16] No one equation is able to accurately predict energy expenditure across a wide range of activities [21] and current prediction techniques only perform well within a specified range of intensities and/or specific activity types.”
Cited by
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715
“device energy expenditure measurements compared with reference standard criterion measures, including direct calorimetry [86,104] and indirect calorimetry [20,21,29-31,38,39,41-43,53,55,63,66,73,85,87,93,95,97,103, 105,116,117,129,130,142,143,146,148,159,165,166,177], in controlled settings.”
MethodAccuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data
Feehan, L. M., Geldman, J., Sayre, E. C., Park, C., Ezzat, A. M., Yoo, J. Y., Hamilton, C. B., & Li, L. C. (2018) · JMIR mHealth and uHealth · doi:10.2196/10527 · Cited by 616
“or indirect (16 studies) calorimetry (Multimedia Appendix 3) [29,34,37,39,46,68-80].”
BackgroundHenriksen, A., Haugen Mikalsen, M., Woldaregay, A. Z., Muzny, M., Hartvigsen, G., Hopstock, L. A., & Grimsgaard, S. (2018) · Journal of Medical Internet Research · doi:10.2196/jmir.9157 · Cited by 542
“In 22 of the studies, one or more Garmin devices were used [32,33,46,49,50,62,77-92].”
MethodSember, V., Meh, K., Sorić, M., Starc, G., Rocha, P., & Jurak, G. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17197161 · Cited by 266
Germini, F., Noronha, N., Borg Debono, V., Abraham Philip, B., Pete, D., Navarro, T., Keepanasseril, A., Parpia, S., de Wit, K., & Iorio, A. (2022) · Journal of Medical Internet Research · doi:10.2196/30791 · Cited by 220
“For the Fitbit One, a study reported a mean bias of 2.91 (SD 4.35) kcal per minute [49], whereas for the Fitbit Ultra, the Pearson correlation coefficient ranged from 0.24 to 0.67 for different physical activities [58].”
Influential
Metadata
- Authors
- Price, K., Bird, S. R., Lythgo, N., Raj, I. S., Wong, J. Y. L., & Lynch, C.
- Venue
- Journal of Medical Engineering & Technology
- Publisher
- Informa UK Limited
- Published
- 2016-12-05
- Volume
- 41
- Issue
- 3
- Pages
- 208-215
- ISSN
- 0309-1902 (ISSN-L)
- 1464-522X
- OpenAlex
- W2560086155
- Cited by
- 87
- References
- 11
- FWCI
- 8.21
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.