Reliability and Validity of Commercially Available Wearable Devices for Measuring Steps, Energy Expenditure, and Heart Rate: Systematic Review
Open accessPublished version of a preprint
Abstract
Background: Consumer-wearable activity trackers are small electronic devices that record fitness and health-related measures. Objective: The purpose of this systematic review was to examine the validity and reliability of commercial wearables in measuring step count, heart rate, and energy expenditure. Methods: We identified devices to be included in the review. Database searches were conducted in PubMed, Embase, and SPORTDiscus, and only articles published in the English language up to May 2019 were considered. Studies were excluded if they did not identify the device used and if they did not examine the validity or reliability of the device. Studies involving the general population and all special populations were included. We operationalized validity as criterion validity (as compared with other measures) and construct validity (degree to which the device is measuring what it claims). Reliability measures focused on intradevice and interdevice reliability. Results: We included 158 publications examining nine different commercial wearable device brands. Fitbit was by far the most studied brand. In laboratory-based settings, Fitbit, Apple Watch, and Samsung appeared to measure steps accurately. Heart rate measurement was more variable, with Apple Watch and Garmin being the most accurate and Fitbit tending toward underestimation. For energy expenditure, no brand was accurate. We also examined validity between devices within a specific brand. Conclusions: Commercial wearable devices are accurate for measuring steps and heart rate in laboratory-based settings, but this varies by the manufacturer and device type. Devices are constantly being upgraded and redesigned to new models, suggesting the need for more current reviews and research.
AI snapshot
Written from the paper’s full text.
TL;DR
This systematic review of 158 publications found that commercial wearables are generally accurate for step count and heart rate in controlled settings, but not for energy expenditure. Accuracy varied by brand, device, and setting, while interdevice reliability was generally very strong and intradevice reliability data were limited mainly to step count.
Key findings
- The review included 158 publications, 169 studies/substudies, 5,934 participants, and 1,838 comparisons; Fitbit was most studied (144 studies, 12 models), followed by Garmin (42 studies, 13 models), Apple (28 studies, 2 models), Polar (15 studies, 6 models), Misfit (13 studies, 2 models), Withings (12 studies, 2 models), Samsung (8 studies, 4 models), Mio (6 studies, 2 models), and Xiaomi (2 studies, 2 models).
- For step count in controlled settings, 45.2% (364/805) of comparisons were within ±3% measurement error, with an overall tendency to underestimate (mean −9%, median −2%); in free-living settings, 42% (29/69) were within ±10%, with slight overestimation (mean 5%, median 6%).
- For heart rate in controlled settings, 56.5% (100/177) of comparisons were within ±3%, with a slight underestimation tendency (median −1%); Apple Watch was within ±3% 71% (35/49) of the time, Fitbit 51% (36/71), and Garmin 49% (23/47). For energy expenditure in controlled settings, only 9.2% (28/305) were within ±3%, and no brand was within ±3% more than 13% of the time.
- Reliability analyses included 19 studies and 166 comparisons; 153/166 comparisons reported interdevice reliability. Interdevice correlations were mostly very strong for step count (35 very strong, 3 strong) and energy expenditure (75% of 16 reported correlations very strong), while intradevice step-count reliability was variable (mean correlation coefficient 0.58).
Methods
- Systematic review reported per PRISMA; searched PubMed, Embase, and SPORTDiscus through May 28, 2019, limited to English-language articles published from 2000 onward, including consumer wearable brands Apple, Empatica, Fitbit, Garmin, Jawbone, Mio, Misfit, Polar, Samsung, UnderArmour, Withings, and Xiaomi.
- From 34,890 unique citations, 21,083 remained after duplicate removal, 542 underwent full-text screening, and 158 publications were included; these yielded 169 studies/substudies, with 168 examining validity and 19 examining reliability, and 1,838 comparisons (1,672 validity, 166 reliability).
- Because of heterogeneous testing conditions and outcomes, meta-analysis was not conducted; the authors used narrative synthesis, criterion validity via correlation and group percentage difference, acceptable limits of ±3% in controlled settings and ±10% in free-living settings, and modified COSMIN-based risk-of-bias assessment.
Limitations
- Stated by the authors: the consumer wearable market is volatile (e.g., Google purchased Fitbit for US $2.3 billion on November 1, 2019), many included devices are outdated or discontinued, and restricting to English-language studies may introduce bias.
- Stated/evident: meta-analyses were not possible because of wide ranges in testing conditions and reported outcomes; reliability studies were fewer than validity studies, and between-brand or between-device reliability comparisons were not possible because of small sample sizes; no intradevice reliability data were reported for heart rate or energy expenditure.
- Evident from the text: 82.8% of studies (140/169; 1,640/1,838 comparisons) were rated fair or poor for sample size (<50 participants), with only 7 excellent (≥100 participants) and 22 good (50-99 participants) studies; many tested devices are no longer available, and no brand stood out as a “gold standard” or had complete assessment across all measures.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 39 cited sources checked · 0 supported · 0 flagged
- Findingown data
Fitbit was the most studied brand in the review.
“Fitbit was by far the most studied brand.”
- Also stated in: § Study and Participant Characteristics · plain
“Fitbit consumer-grade wearables were examined most frequently (144 studies examining 12 models), followed by Garmin (42 studies, 13 models), Apple (28 studies, 2 models), Polar (15 studies, 6 models), Misfit (13 studies, 2 models), Withings (12 studies, 2 models), Samsung (8 studies, 4 models), Mio (6 studies, 2 models), and Xiaomi (2 studies, 2 models)”
Quote in the paperNot a substantial claim - Also stated in: § Study and Participant Characteristics · plain
- Findingnone given
In laboratory-based settings, Fitbit, Apple Watch, and Samsung appeared to measure steps accurately.
“In laboratory-based settings, Fitbit, Apple Watch, and Samsung appeared to measure steps accurately.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
- Findingown data
Those three comparisons noted underestimation of heart rate that varied depending on activity intensity.
“and noted underestimation of heart rate that varied depending on activity intensity”
Quote in the paperNot a substantial claim - Findingown data
The correlation coefficient was moderate in one free-living heart rate comparison examining estimation during high-intensity activity.
“and moderate in one comparison examining estimation during high-intensity activity.”
Quote in the paperNot a substantial claim - Findingown data
Despite similar ±3% measurement error rates, Fitbit appeared to underestimate heart rate more than Apple Watch and Garmin.
“Despite similar ±3% measurement error rates, Fitbit appeared to underestimate heart rate more than Apple Watch and Garmin.”
Quote in the paperNot a substantial claim
- Findingown data
A total of 42 studies (84 comparisons) examined wearable device step count measurements compared with accelerometry in free-living settings.
“A total of 42 studies (84 comparisons) examined wearable device step count measurements compared with the reference standard criterion measure of accelerometry [33,45,48,49, 56,59,60,64,74-76,89,96,101,106-112,120,131-136, 149,154-156,159,167,168,172-175] in free-living settings”
Quote in the paperNot a substantial claim- Rests on an assumption: Accelerometry is a valid reference standard criterion measure for step count in free-living settings.
- Findingown data
For heart rate, measurement error varied by device brand.
“For heart rate, measurement error also varied by device brand (Figure 5).”
Quote in the paperSubstantial claim - Findingown data
Of the 69 step count comparisons with group measurement error, 17% (n=12) were below −10% measurement error.
“17% (n=12) were below −10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
Of the 69 step count comparisons with group measurement error, 42% (n=29) were within ±10% measurement error.
“Of these, 42% (n=29) were within ±10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
Error level for step count varied by device brand.
“We observed that the error level varied by device brand (Figure 5).”
Quote in the paperNot a substantial claim - Findingown data
The overall tendency for step count in free-living settings was slight overestimation (mean 5%, median 6%).
“The overall tendency was slight overestimation of step count (mean: 5%, median: 6%).”
Quote in the paperSubstantial claim - Findingown data
There were relatively few studies on wearable device validity in free-living conditions.
“There were relatively few studies on wearable device validity in free-living conditions.”
Quote in the paperSubstantial claim - Findingreasoning
Three of the four remaining free-living heart rate comparisons examined the Fitbit Charge HR in adults.
“Three of the four remaining comparisons examined the Fitbit Charge HR in adults”
Quote in the paperNot a substantial claim - Findingown data
No brand had more than 10 studies for heart rate or energy expenditure validity in free-living conditions.
“no brands had more than 10 studies for heart rate or energy expenditure.”
Quote in the paperSubstantial claim - Findingown data
Of the 11 step count comparisons reporting MAPE, 60% (n=9) were above 10% measurement error.
“60% (n=9) were above 10% measurement error.”
Quote in the paperNot a substantial claim - Findingown data
Correlation coefficients were strong to very strong in four of the five free-living heart rate comparisons.
“Correlation coefficients were strong to very strong in four of the five comparisons”
Quote in the paperNot a substantial claim - Findingown data
No apparent systematic bias for measurement error was seen beyond a small number of comparisons showing extreme overestimation (four in step count and five in energy expenditure).
“In these examinations, we saw no apparent systematic bias for measurement error beyond a small number of comparisons showing extreme overestimation (four comparisons in step count and five comparisons in energy expenditure).”
Quote in the paperSubstantial claim - Findingown data
Of the 11 step count comparisons reporting MAPE, 40% (n=6) were below 10% measurement error.
“of which 40% (n=6) were below 10% measurement error”
Quote in the paperNot a substantial claim - Findingnone given
For energy expenditure, no brand was accurate.
“For energy expenditure, no brand was accurate.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
- Findingown data
There was a slight overall tendency toward underestimation of heart rate (median error −1%).
“There was a slight overall tendency toward underestimation of heart rate (estimated median error: −1%).”
Quote in the paperNot a substantial claim - Findingown data
Group measurement error was reported or calculable for 69 of the 84 step count comparisons in free-living settings.
“Group measurement error was reported or calculable for 69 of the 84 comparisons, regardless of the criterion measure.”
Quote in the paperNot a substantial claim - Findingown data
For energy expenditure, no brand was within ±3% measurement error more than 13% of the time.
“For energy expenditure estimates, no brand of wearable was within ±3% measurement error more than 13% of the time (Figure 5).”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
- Findingown data
Of 169 studies, 140 (82.8%) were rated fair or poor for sample size (<50 participants).
“Of 169 studies, 140 (82.8%; 1640 of 1838 [89.23%] comparisons) were rated fair or poor for sample size (<50 participants), but were not excluded from the analysis owing to the paucity of studies with excellent (≥100 participants, n=7) and good (50-99 participants, n=22) sample sizes.”
Quote in the paperSubstantial claim - Findingown data
Fitbit was the only brand with more than 10 studies published for step count validity in free-living conditions.
“Fitbit was the only brand with more than 10 studies published for step count validity in free-living conditions”
Quote in the paperSubstantial claim - Findingcitation
Three studies (five comparisons) examined wearable device heart rate compared with a Polar brand chest strap in free-living settings.
“Three studies (five comparisons) examined wearable device heart rate compared with the reference standard criterion measure of a Polar brand chest strap in free-living settings [75,77,78].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [75] source
Not checked in the free run.
- cites [77] Brazendale, K, 2019, Int J Exerc Sci
Not checked in the free run.
- cites [78] source
Not checked in the free run.
- Rests on an assumption: A Polar brand chest strap is a valid reference standard criterion measure for heart rate in free-living settings.
- Findingown data
Of the 69 step count comparisons with group measurement error, 41% (n=28) were above 10% measurement error.
“41% (n=28) were above 10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
Nine studies (22 comparisons) examined energy expenditure in free-living settings compared with doubly labeled water and accelerometry.
“Nine studies (22 comparisons) examined energy expenditure in free-living settings compared with the criterion measures of doubly labeled water [104] and accelerometry [29,49,79,101,131,172,174,175].”
Quote in the paperSubstantial claim0 of 9 sources checked- cites [104] source
Not checked in the free run.
- cites [29] source
Not checked in the free run.
- cites [49] source
Not checked in the free run.
- cites [79] source
Not checked in the free run.
- cites [101] source
Not checked in the free run.
- cites [131] source
Not checked in the free run.
- cites [172] source
Not checked in the free run.
- cites [174] source
Not checked in the free run.
- cites [175] source
Not checked in the free run.
- Rests on an assumption: Doubly labeled water and accelerometry are valid criterion measures for energy expenditure in free-living settings.
- Findingown data
The overall tendency was to underestimate step count (mean −9%, median −2%).
“The overall tendency was to underestimate step count (mean: −9%, median: −2%).”
Quote in the paperSubstantial claim - Findingown data
Group measurement error was reported or calculable for 17 of the 22 free-living energy expenditure comparisons.
“Group measurement error was reported or calculable for 17 of the 22 comparisons, regardless of the criterion measure.”
Quote in the paperNot a substantial claim - Findingown data
The MPE values for Fitbit, Samsung, and Withings fell within ±3% on average.
“while the MPE values for Fitbit, Samsung, and Withings fell within ±3% on average.”
Quote in the paperNot a substantial claim - Findingnone given
It appears that Fitbit devices may underestimate heart rate depending on activity intensity.
“but it appears that Fitbit devices may underestimate heart rate depending on activity intensity.”
Quote in the paperNot a substantial claim - Findingnone given
In both controlled and free-living studies, when not correctly estimating steps, devices tended to underestimate values.
“In both controlled and free-living studies, when not correctly estimating steps, devices tended to underestimate values.”
Quote in the paperNot a substantial claim - Findingown data
Twelve studies (51 comparisons) with sample sizes ranging from 13 to 56 reported interdevice reliability for step count.
“Twelve studies (51 comparisons) with sample sizes ranging from 13 [117,138] to 56 [151] reported on interdevice reliability for step count [50,58,72,85,94,110,113,116,117,121,125, 138,151,161,171].”
Quote in the paperNot a substantial claim - Findingown data
Of the 16 interdevice energy expenditure reliability comparisons with correlation coefficients, 75% (n=12) were rated very strong.
“75% (n=12) were rated very strong.”
Quote in the paperNot a substantial claim - Findingown data
No studies reported intradevice reliability for heart rate or energy expenditure.
“No studies reported intradevice reliability for heart rate or energy expenditure.”
Quote in the paperNot a substantial claim - Findingown data
Nineteen studies (166 comparisons) with sample sizes ranging from 11 to 56 reported inter- or intradevice reliability for Apple, Fitbit, Garmin, Polar, and Withings.
“Nineteen studies (166 comparisons) with sample sizes ranging from 11 [94] to 56 [151] reported inter- or intradevice reliability for Apple (seven comparisons), Fitbit (92 comparisons), Garmin (22 comparisons), Polar (one comparison), and Withings (44 comparisons).”
Quote in the paperNot a substantial claim - Findingcitation
Two studies (13 comparisons) reported intradevice reliability for step count, with sample sizes of 20 and 24.
“Two studies (13 comparisons) reported on intradevice reliability for step count, with sample sizes of 20 [82] and 24 [150].”
Quote in the paperNot a substantial claim0 of 2 sources checked - Findingown data
Of the 17 energy expenditure comparisons with group measurement error, 29% (n=5) were above 10% measurement error.
“29% (n=5) were above 10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
Of the 17 energy expenditure comparisons with group measurement error, 18% (n=3) were within ±10% measurement error.
“Of these, 18% (n=3) were within ±10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
The mean intradevice reliability correlation coefficient for step count was 0.58.
“The mean correlation coefficient was 0.58.”
Quote in the paperNot a substantial claim - Findingown data
The majority of correlation coefficients for step count interdevice reliability were very strong (n=35).
“The majority of correlation coefficients for step count interdevice reliability were very strong (n=35)”
Quote in the paperNot a substantial claim - Findingreasoning
In controlled settings across all devices, heart rate was accurately measured with only a very small tendency for underestimation.
“In controlled settings across all devices, heart rate was accurately measured with only a very small tendency for underestimation.”
Quote in the paperSubstantial claim - Findingreasoning
All devices were within acceptable measurement error for heart rate.
“All devices were within acceptable measurement error for heart rate.”
Quote in the paperSubstantial claim - Findingown data
There was an overall tendency to underestimate energy expenditure in free-living settings (mean −3%, median −11%).
“There was an overall tendency to underestimate energy expenditure (mean: −3%, median: −11%).”
Quote in the paperSubstantial claim - Findingnone given
Heart rate validity was only sufficiently tested in Apple Watch, Fitbit, and Garmin devices.
“Heart rate validity was only sufficiently tested in Apple Watch, Fitbit, and Garmin devices.”
Quote in the paperNot a substantial claim - Findingnone given
There were few studies examining the validity of heart rate measures in free-living conditions.
“There were few studies examining the validity of heart rate measures in free-living conditions”
Quote in the paperNot a substantial claim - Findingown data
Only a small number (n=3) of step count interdevice reliability correlation coefficients were reported as strong.
“with only a small number (n=3) being reported as strong.”
Quote in the paperNot a substantial claim - Findingnone given
All of the brands measured heart rate to within ±3% on average in controlled settings.
“All of the brands measured heart rate to within ±3% on average in controlled settings.”
Quote in the paperSubstantial claim - Findingown data
Criterion validity of commercial wearables varied by study type (controlled or free-living), brand, and device.
“Criterion validity of commercial wearables varied by study type (controlled or free-living), brand, and device.”
Quote in the paperSubstantial claim - Findingnone given
Validity compared with criteria was the best for Apple Watch and Garmin.
“Validity compared with criteria was the best for Apple Watch and Garmin”
Quote in the paperSubstantial claim - Findingown data
Of the 16 interdevice energy expenditure reliability comparisons with correlation coefficients, 6% (n=1) were rated moderate.
“6% (n=1) were rated moderate”
Quote in the paperNot a substantial claim - Findingown data
Of the 16 interdevice energy expenditure reliability comparisons with correlation coefficients, 6% (n=1) were rated strong.
“6% (n=1) were rated strong”
Quote in the paperNot a substantial claim - Findingreasoning
In many of the studies, there did appear to be a tendency for systematic over or underestimation of energy expenditure.
“In many of the studies, there did appear to be a tendency for systematic over or underestimation.”
Quote in the paperSubstantial claim - Findingown data
The majority of reliability comparisons (153/166) reported interdevice reliability for step count, heart rate, or energy expenditure.
“The majority of comparisons (153/166) reported interdevice reliability for step count, heart rate, or energy expenditure.”
Quote in the paperNot a substantial claim - Findingown data
Of the 17 energy expenditure comparisons with group measurement error, 53% (n=9) were below −10% measurement error.
“53% (n=9) were below −10% measurement error”
Quote in the paperNot a substantial claim - Findingown data
Of the 16 interdevice energy expenditure reliability comparisons with correlation coefficients, 13% (n=2) were rated very weak.
“Of these, 13% (n=2) were rated very weak”
Quote in the paperNot a substantial claim - Findingnone given
Apple, Fitbit, and Garmin were accurate nearly 50% of the time.
“and Apple, Fitbit, and Garmin were accurate nearly 50% of the time.”
Quote in the paperSubstantial claim - Findingown data
The review was not able to conduct comparisons between brands or devices owing to small sample sizes.
“and we were not able to conduct comparisons between brands or devices owing to small sample sizes.”
Quote in the paperSubstantial claim - Findingnone given
No specific device or brand involved a complete assessment across all measures.
“There was no specific device or brand that involved a complete assessment across all measures”
Quote in the paperSubstantial claim - Findingnone given
Sufficient data for intradevice reliability was only available for step count.
“Sufficient data for intradevice reliability was only available for step count.”
- Also stated in: § Conclusion · plain
“Sufficient data for intradevice reliability were only available for step count”
Quote in the paperSubstantial claim - Also stated in: § Conclusion · plain
- Findingnone given
Compared with validity studies, there were fewer reliability studies.
“However, compared with validity studies, there were fewer reliability studies”
Quote in the paperSubstantial claim - Findingnone given
No brand fell within the acceptable accuracy limits for energy expenditure.
“No brand fell within the acceptable accuracy limits for energy expenditure.”
Quote in the paperSubstantial claim - Findingown data
There does not appear to be a relationship among more accurate estimates of energy expenditure in devices that include heart rate.
“There does not appear to be a relationship among more accurate estimates of energy expenditure in devices that include heart rate (Multimedia Appendix 7).”
Quote in the paperNot a substantial claim - Findingreasoning
No specific brand stood out as the "gold standard" in fitness wearables.
“and no specific brand stood out as the “gold standard” in fitness wearables.”
Quote in the paperSubstantial claim - Findingreasoning
Many of the devices utilized in the studies included in this review are so out of date that they are no longer available on the market.
“many of the devices utilized in the studies included in this review are so out of date that they are no longer available on the market.”
Quote in the paperNot a substantial claim - Findingnone given
Apple and Samsung had the highest validity for step count.
“Apple and Samsung had the highest validity for step count”
Quote in the paperSubstantial claim - Interpretationreasoning
The five extreme outliers for energy expenditure were likely true reflections of tendencies to overestimate energy expenditure during sedentary and low-intensity activities.
“Upon closer examination of the five extreme outliers for energy expenditure (four occurred in a study with greater than 60 participants [21] and one occurred in a study with fewer than 40 participants [41]), we determined that these were likely true reflections of tendencies to overestimate energy expenditure during sedentary and low-intensity activities, and therefore, we included these five”
Quote in the paperSubstantial claim
- May combine two claims. 41
- Classified as probably not a substantial claim. 22
- The quote only partly states this claim. 1
May combine two claims.
Findingown dataThe systematic review included 158 publications examining nine different commercial wearable device brands.
“We included 158 publications examining nine different commercial wearable device brands.”
- Also stated in: § Results · plain
“As a result, a total of 158 publications were included in this systematic review (Figure 1) [14].”
Quote in the paperNot a substantial claim- Also stated in: § Results · plain
May combine two claims.
Findingown dataHeart rate measurement was more variable across devices, with Apple Watch and Garmin being the most accurate and Fitbit tending toward underestimation.
“Heart rate measurement was more variable, with Apple Watch and Garmin being the most accurate and Fitbit tending toward underestimation.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). The quote only partly states this claim.
Findingown dataAll of those three comparisons reported MAPE values within 10% measurement error.
“but all reported that MAPE values fell within 10% measurement error.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor energy expenditure in controlled settings, group measurement error was reported or calculable for 305 of 312 comparisons; 9.2% within ±3%, 54.1% below −3%, 36.7% above 3%.
“Group measurement error was reported or calculable for 305 of the 312 comparisons, regardless of the criterion measure. Of these, 9.2% (n=28) were within ±3% measurement error, 54.1% (n=165) were below −3% measurement error, and 36.7% (n=112) were above 3% measurement error.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataFitbit devices tended to provide inaccurate measures compared with the criterion, underestimating 48.4% (76/157) and overestimating 39.5% (62/157) of the time.
“Fitbit devices tended to provide inaccurate measures compared with the criterion, underestimating 48.4% (76/157) of the time and overestimating 39.5% (62/157) of the time, despite the boxplot in Figure 5 showing a reasonable median value for accuracy.”
Quote in the paperNot a substantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataGroup measurement error was reported or calculable for one of the five free-living heart rate comparisons, with the Fitbit Charge HR falling within ±10% measurement error in healthy children.
“Group measurement error was reported or calculable for one of the five comparisons, with the Fitbit Charge HR falling within ±10% measurement error in the study examining healthy children.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataAmong the remaining step count comparisons, 11 of 15 reported MAPE.
“Among the remaining comparisons, 11 of 15 reported MAPE”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor step count in controlled settings, group measurement error was reported or calculable for 805 of 979 comparisons; 45.2% were within ±3%, 42.7% below −3%, and 12.1% above 3%.
“Group measurement error was reported or calculable for 805 of the 979 comparisons, regardless of the criterion measure. Of these, 45.2% (n=364) were within ±3% measurement error, 42.7% (n=344) were below −3% measurement error, and 12.1% (n=97) were above 3% measurement error.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataStudies showed a tendency to underestimate energy expenditure and to provide inaccurate measures of energy expenditure compared with the criterion.
“Studies showed a tendency to underestimate energy expenditure and to provide inaccurate measures of energy expenditure compared with the criterion.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataApple wearables overestimated energy expenditure 58% (18/31) of the time and Polar 69% (9/13) of the time.
“Conversely, Apple wearables overestimated energy expenditure 58% (18/31) of the time and Polar wearables overestimated energy expenditure 69% (9/13) of the time.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataGarmin wearables underestimated energy expenditure 69% (37/51) of the time and Withings 74% (34/46) of the time.
“Underestimation of energy expenditure (less than −3%) was observed in Garmin wearables 69% (37/51) of the time and in Withings wearables 74% (34/46) of the time.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataWithings and Misfit wearables consistently underestimated step count, and Apple and Samsung had less measurement variability than other brands.
“Withings and Misfit wearables consistently underestimated step count, and Apple and Samsung had less measurement variability than other brands.”
Quote in the paperNot a substantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataFor heart rate in controlled settings, group measurement error was reported or calculable for 177 of 266 comparisons; 56.5% within ±3%, 24.9% below −3%, 18.6% above 3%.
“Group measurement error was reported or calculable for 177 of 266 comparisons, regardless of the criterion measure. Of these, 56.5% (n=100) were within ±3% measurement error, 24.9% (n=44) were below −3% measurement error, and 18.6% (n=33) were above 3% measurement error.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataApple Watch was within ±3% 71% (35/49) of the time, Fitbit 51% (36/71), and Garmin 49% (23/47).
“Apple Watch was within ±3% 71% (35/49) of the time, while Fitbit wearables were within ±3% 51% (36/71) of the time and Garmin wearables were within ±3% 49% (23/47) of the time.”
Quote in the paperSubstantial claim- Rests on an assumption: The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
May combine two claims.
Findingown dataWithin brands, devices appeared to vary, with Fitbit Classic tending to overestimate steps, while Fitbit Charge tending to underestimate steps.
“Within brands, devices appeared to vary, with Fitbit Classic tending to overestimate steps, while Fitbit Charge tending to underestimate steps”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFive studies (50 comparisons) reported interdevice reliability for energy expenditure, with analyzed sample sizes ranging from 13 to 29.
“Five studies (50 comparisons) reported on interdevice reliability [85,113,116,117,166], with analyzed sample sizes ranging from 13 [117] to 29 [113].”
Quote in the paperNot a substantial claim0 of 5 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataAll five interdevice energy expenditure reliability studies recruited healthy adults (mean age 26.3 years, SD 3.9 years).
“All five studies recruited healthy adults (mean age 26.3 years, SD 3.9 years).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIntradevice reliability correlations for step count were very weak (n=1), weak (n=2), moderate (n=5), strong (n=2), and very strong (n=3).
“Intradevice reliability correlations were very weak (n=1), weak (n=2), moderate (n=5), strong (n=2), and very strong (n=3).”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationSimilar standard typical error levels were seen in maximum heart rate measured during a single incremental maximal oxygen uptake test and heart rate taken from the highest 30-s mean heart rate, with somewhat lower correlation coefficients [24].
“Similar standard typical error levels were seen in maximum heart rate measured during a single incremental maximal oxygen uptake test performed on a treadmill and heart rate taken from the highest 30-s mean heart rate, with somewhat lower correlation coefficients [24].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] source
Not checked in the free run.
May combine two claims.
FindingreasoningOn average, only Fitbit measured energy expenditure to within acceptable limits, but there was wide variation around the estimate.
“On average, only Fitbit measured energy expenditure to within acceptable limits, but there was wide variation around the estimate.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe review included nine brands and 45 devices, with the number of comparisons ranging from 201 for the Fitbit Zip to one for the Garmin Forerunner 405CX and the Polar M600.
“Our review included nine brands and 45 devices with the number of comparisons ranging from 201 for the Fitbit Zip to one for the Garmin Forerunner 405CX and the Polar M600.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationDuring the same tasks, Garmin Vivosmart HR+ showed good reliability during all tasks and had narrower limits of agreement than Fitbit.
“During the same tasks, Garmin Vivosmart HR+ showed good reliability during all tasks and had narrower limits of agreement than Fitbit.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [58] source
Not checked in the free run.
May combine two claims.
FindingcitationApple Watch showed very good interdevice reliability at 5-s epochs during treadmill bouts at 4, 7, and 10 km/h, with reliability increasing and standard typical error decreasing with increasing pace [26].
“Apple Watch showed very good interdevice reliability at 5-s epochs during treadmill bouts at 4, 7, and 10 km/h, with reliability increasing and standard typical error decreasing with increasing pace [26].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationXiaomi data were not analyzed in a single indirect calorimetry study owing to the lack of data [53].
“Xiaomi data were not analyzed in a single indirect calorimetry study owing to the lack of data [53].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] source
Not checked in the free run.
May combine two claims.
Findingown dataEnergy expenditure estimates varied widely with less than 10% of estimates falling within acceptable limits in controlled settings.
“Energy expenditure estimates varied widely with less than 10% of estimates falling within acceptable limits in controlled settings.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor step count, in controlled laboratory settings, a higher proportion of devices showed accuracy, and this was within a tighter limit of acceptable accuracy compared with free-living conditions.
“For step count, our review showed that in controlled laboratory settings, a higher proportion of devices showed accuracy, and this was within a tighter limit of acceptable accuracy compared with free-living conditions.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationIn healthy older adults, Fitbit Charge 2 showed good reliability during treadmill and overground bouts and poor reliability during hand movement tasks such as dusting [58].
“In the examination of interdevice reliability in healthy older adults, Fitbit Charge 2 showed good reliability during treadmill and overground bouts and poor reliability during hand movement tasks such as dusting [58].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [58] source
Not checked in the free run.
May combine two claims.
FindingcitationThree studies (23 comparisons) examined interdevice reliability for heart rate, with analyzed sample sizes ranging from 13 to 21.
“Three studies (23 comparisons) examined interdevice reliability for heart rate [24,26,58], with analyzed sample sizes ranging from 13 [24] to 21 [26].”
Quote in the paperNot a substantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataCorrelation coefficients were reported for 16 of the 50 interdevice energy expenditure reliability comparisons.
“Correlation coefficients were reported for 16 of 50 comparisons.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenFitbit was studied the most and Xiaomi and Mio were studied the least.
“Among the studies included, Fitbit was studied the most and Xiaomi and Mio were studied the least.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenInterdevice reliabilities for steps, heart rate, and energy expenditure were all very strong.
“Interdevice reliabilities for steps, heart rate, and energy expenditure were all very strong.”
- Also stated in: § Conclusion · boosted
“Interdevice reliabilities for steps, heart rate, and calories were all very strong.”
Quote in the paperSubstantial claim- Also stated in: § Conclusion · boosted
May combine two claims.
Findingown dataThis systematic review of 158 publications included assessments of consumer wearable devices from nine brands (Apple Inc, Fitbit, Garmin, Mio, Misfit, Polar, Samsung, Withings, and Xiaomi).
“This systematic review of 158 publications included assessments of consumer wearable devices from nine brands (Apple Inc, Fitbit, Garmin, Mio, Misfit, Polar, Samsung, Withings, and Xiaomi), with a focus on the reliability and validity of the devices in measuring heart rate, energy expenditure, and step count.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenEnergy expenditure estimates varied by model, with the Fitbit Classic underestimating the value considerably and Fitbit Charge HR overestimating the value.
“Energy expenditure estimates also varied by model, with the Fitbit Classic underestimating the value considerably and Fitbit Charge HR overestimating the value.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe results showed considerable variability within the same device for step count for Fitbit Charge HR, Fitbit Surge, Fitbit Zip, and Garmin Vivofit, with five, five, one, and two comparisons, respectively.
“The results showed considerable variability within the same device for step count for Fitbit Charge HR, Fitbit Surge, Fitbit Zip, and Garmin Vivofit, with five, five, one, and two comparisons, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningThe four extreme outliers for step count involved measurement during sedentary and light physical activity in a single study with fewer than 40 participants and were likely inflated by limited steps accumulated.
“The four extreme outliers for step count involved measurement during sedentary and light physical activity in a single study with fewer than 40 participants [20] and were likely inflated by the limited number of steps accumulated during those bouts.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningCommercial wearable devices are accurate for measuring steps and heart rate in laboratory-based settings, but this varies by the manufacturer and device type.
“Commercial wearable devices are accurate for measuring steps and heart rate in laboratory-based settings, but this varies by the manufacturer and device type.”
Quote in the paperSubstantial claim- Rests on an assumption: The commercial wearable devices included in the review (nine brands) are representative of commercial wearable devices generally.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe review was not registered with PROSPERO.
“The review was not registered with PROSPERO.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductown dataFull-text documents were independently assessed for inclusion by at least two authors, with conflicts resolved by discussion and consensus.
“Following the title and abstract screening, full-text documents of the selected studies were searched and retrieved and were independently assessed for inclusion by at least two authors (EC, JL, and DF). Any conflicts were resolved by discussion and consensus.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationThe systematic review was conducted and reported according to PRISMA guidelines.
“This systematic review was conducted and reported according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [14].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductcitationThe MEDLINE search strategy was developed by a librarian and peer reviewed by a second librarian according to PRESS 2015.
“A librarian (KR) developed the MEDLINE search strategy, which was peer reviewed by a second librarian according to the Peer Review of Electronic Search Strategies (PRESS) 2015 Guideline Statement [15].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationFitbit wearables were highly correlated with criterion measures of step count during laboratory-based assessment and had consistently high interdevice reliability for both step count and energy expenditure.
“Fitbit wearables were highly correlated with criterion measures of step count during laboratory-based assessment and had consistently high interdevice reliability for both step count and energy expenditure [13].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationBunn et al noted tendencies of wearables to underestimate energy expenditure, heart rate, and step count.
“In a recent review of 10 articles, Bunn et al [12] noted tendencies of wearables to underestimate energy expenditure, heart rate, and step count.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [12] Bunn, JA, 2018, Int J Exerc Sci
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFitbit wearables provide accurate measures only in limited circumstances.
“However, this review found that these devices tended to underestimate energy expenditure, which is consistent with a separate review of Fitbit accuracy [14] indicating that Fitbit wearables provide accurate measures only in limited circumstances.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningThere are possible interactions between the number and size of studies and device wear location that may influence brand comparisons.
“There are possible interactions between the number and size of studies and device wear location that may influence the brand comparisons.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductcitationRisk of bias and quality of each study were assessed using a modification of the modified COSMIN validation subscale and an a priori modified COSMIN reliability subscale.
“using a modification of the modified Consensus-Based Standards for the Selection of Health Status Measurement Instruments (COSMIN) validation subscale used by Feehan et al [13] and an a priori modified COSMIN reliability subscale (Multimedia Appendix 2) to assess the quality and risk of bias of each study (EC and DF)”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] source
Not checked in the free run.
May combine two claims.
PremisecitationResearch has shown variation in step count among devices, and the applicability of metrics may vary by device brand and device type.
“However, research has shown variation in step count among devices, and the applicability of these metrics may vary by device brand and device type [6].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [6] source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
PremisecitationAcceptable limits of percentage difference were ±3% in controlled settings and ±10% in free-living settings.
“Our interpretation of measurement accuracy was focused on acceptable limits of percentage difference of ±3% in controlled settings and percentage difference of ±10% in free-living settings, as outlined in previous work [13].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenDetails from each reviewer were compared, and inconsistencies were resolved through consensus before compiling results.
“Details from each reviewer were compared, and inconsistencies were resolved through consensus before compiling the results (EC and DF).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataTitles and abstracts were independently assessed by at least two authors, with conflicts resolved by a third author or consensus.
“The titles and abstracts of the studies included from the initial database search were independently assessed by at least two authors from the team. Conflicts arising during any step of the screening for inclusion/exclusion were resolved by a third author or by consensus.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningThe variability observed could be attributed to differences in the number of comparisons for each device and in wear locations of the devices.
“however, the variability observed could be attributed to differences in the number of comparisons for each device and in wear locations of the devices.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe bias assessment showed no apparent bias toward studies of different sample sizes.
“Our bias assessment showed no apparent bias toward studies of different sample sizes.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThere is a strong overrepresentation of studies with 20 participants.
“However, there is a strong overrepresentation of studies with 20 participants.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationThere is some potential for bias when including only English language studies in systematic reviews.
“There is some potential for bias when including only English language studies in systematic reviews.”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationStudies have shown that the effect of English-language restriction may be small in general but may be difficult to measure for an individual systematic review [183,184].
“However, studies have shown that the effect may be small in general but may be difficult to measure for an individual systematic review [183,184].”
Quote in the paperNot a substantial claim0 of 2 sources checked
The paper relies on this without support.
The criterion measures used in the included studies (manual counting, accelerometry, electrocardiography, Polar chest straps, pulse oximetry, direct and indirect calorimetry) provide valid reference standards for step count, heart rate, and energy expenditure.
“Criterion validity: comparing the devices to a criterion measure of steps, heart rate, or energy expenditure.”
“compared with reference standard criterion measures of manual counting [32,34-38,42,46,47,50-53,57,58,72,80-84,88-102,109, 114-125,138-141,144-147,149-153,158-161,165,169-171,173] and accelerometry [20,60,64-66,85,103,109,126-128,148, 154,164]”
“compared with reference standard criterion measures, including electrocardiography [22,23,38-40,43,44,54,61,62,67-70,142,162,176], Polar brand chest straps [20,21,24-28,58,63,71,163], and pulse oximetry [66]”
The paper relies on this without support.
The commercial wearable devices included in the review (nine brands) are representative of commercial wearable devices generally.
“We included 158 publications examining nine different commercial wearable device brands.”
“Commercial wearable devices are accurate for measuring steps and heart rate in laboratory-based settings, but this varies by the manufacturer and device type.”
The paper relies on this without support.
Research is equivocal about whether commercial wearable devices are valid and reliable methods for estimating steps, heart rate, and energy expenditure.
“Despite their popularity, research is equivocal about whether commercial wearable devices are valid and reliable methods for estimating metrics associated with physical activity including steps, heart rate, and energy expenditure.”
The paper relies on this without support.
Accelerometry is a valid reference standard criterion measure for step count in free-living settings.
“compared with the reference standard criterion measure of accelerometry [33,45,48,49, 56,59,60,64,74-76,89,96,101,106-112,120,131-136, 149,154-156,159,167,168,172-175] in free-living settings”
The paper relies on this without support.
A Polar brand chest strap is a valid reference standard criterion measure for heart rate in free-living settings.
“compared with the reference standard criterion measure of a Polar brand chest strap in free-living settings [75,77,78].”
The paper relies on this without support.
Doubly labeled water and accelerometry are valid criterion measures for energy expenditure in free-living settings.
“compared with the criterion measures of doubly labeled water [104] and accelerometry [29,49,79,101,131,172,174,175].”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The study was unable to account for daily step variations between different types of Fitbit models, seasonal differences, and the COVID-19 pandemic because device data were not available at the time of analyses and data were date-shifted to protect participant privacy.
“We were not able to account for daily step variations between different types of Fitbit models14 and seasonal differences15 as well as the occurrence of the COVID-19 pandemic because device data were not available at the time of analyses and data were date-shifted to protect privacy of participants.”
From Association of step counts over time with the risk of chronic disease in the All of Us Research Program (2022) · cites this paper as [14]
In recent years, the quality and accuracy of wearables have improved.
“In recent years, the quality and accuracy of wearables have improved [7,8], resulting in more clinically approved certifications [9].”
From The impact of wearable technologies in health research: scoping review (2022) · cites this paper as [7]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Discussion reverses the direction of free-living step-count error
MajorThe discussion says devices tended to underestimate steps in free-living studies, but the free-living results report more overestimates than underestimates and describe an overall tendency toward overestimation.
“In both controlled and free-living studies, when not correctly estimating steps, devices tended to underestimate values.”
“Of these, 42% (n=29) were within ±10% measurement error, 17% (n=12) were below −10% measurement error, and 41% (n=28) were above 10% measurement error. The overall tendency was slight overestimation of step count (mean: 5%, median: 6%).”
Garmin energy-expenditure percentage does not match its fraction
MinorThe reported 37 of 51 Garmin comparisons is 72.55%, not 69%.
Reported 69 · recomputed 72.54902
“Underestimation of energy expenditure (less than −3%) was observed in Garmin wearables 69% (37/51) of the time”
MAPE categories exceed the number said to report MAPE
MinorThe six comparisons below 10% error and nine above it total 15, although only 11 comparisons are said to have reported MAPE.
Reported 11 · recomputed 15
“Among the remaining comparisons, 11 of 15 reported MAPE, of which 40% (n=6) were below 10% measurement error and 60% (n=9) were above 10% measurement error.”
Mean comparisons per reliability study is understated
MinorThe stated 166 reliability comparisons across 19 reliability studies give a mean of about 8.74, which rounds to 9 rather than the reported 8.
Reported 8 · recomputed 8.736842
“166 examined reliability (mean 8, SD 11 per reliability study; range 1-40)”
Withings model count conflicts with Table 1
MinorThe study-characteristics summary says two Withings models were examined, while Table 1 lists three: Pulse, Pulse O2, and Pulse Ox.
“Withings (12 studies, 2 models)”
“| Withings | Pulse | 2013 | Discontinued | Collar, pant pocket, waist/hip, or wrist | [89,96,131,166] | | Withings | Pulse O2 | 2013 | Discontinued | Collar, waist/hip, or wrist | [104,122,123,169,175] | | Withings | Pulse Ox | 2014 | Current model | Waist/hip or wrist | [53,58] |”
Cites
PRISMA-P Group, Moher, D., Shamseer, L., Clarke, M., Ghersi, D., Liberati, A., Petticrew, M., Shekelle, P., & Stewart, L. A. (2015) · Systematic Reviews · doi:10.1186/2046-4053-4-1 · Cited by 26,223 · see citing works
PRESS Peer Review of Electronic Search Strategies: 2015 Guideline Statement
McGowan, J., Sampson, M., Salzwedel, D. M., Cogo, E., Foerster, V., & Lefebvre, C. (2016) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2016.01.021 · Cited by 5,700 · see citing works
Morrison, A., Polisena, J., Husereau, D., Moulton, K., Clark, M., Fiander, M., Mierzwinski-Urban, M., Clifford, T., Hutton, B., & Rabb, D. (2012) · International Journal of Technology Assessment in Health Care · doi:10.1017/s0266462312000086 · Cited by 1,257 · see citing works
Systematic review of the validity and reliability of consumer-wearable activity trackers
Evenson, K. R., Goto, M. M., & Furberg, R. D. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0314-1 · Cited by 1,232 · see citing works
How many steps/day are enough? for adults
Tudor-Locke, C., Craig, C. L., Brown, W. J., Clemes, S. A., De Cocker, K., Giles-Corti, B., Hatano, Y., Inoue, S., Matsudo, S. M., Mutrie, N., et al. (2011) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/1479-5868-8-79 · Cited by 1,087 · see citing works
Direction and impact of language bias in meta-analyses of controlled trials: empirical study
Jüni, P., Holenstein, F., Sterne, J., Bartlett, C., & Egger, M. (2002) · International Journal of Epidemiology · doi:10.1093/ije/31.1.115 · Cited by 892 · see citing works
Accuracy of Smartphone Applications and Wearable Devices for Tracking Physical Activity Data
Case, M. A., Burwick, H. A., Volpp, K. G., & Patel, M. S. (2015) · JAMA · doi:10.1001/jama.2014.17841 · Cited by 694 · see citing works
Quantifying the body: monitoring and measuring health in the age of mHealth technologies
Lupton, D. A. (2013) · Critical Public Health · doi:10.1080/09581596.2013.794931 · Cited by 646 · see citing works
Accuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data
Feehan, L. M., Geldman, J., Sayre, E. C., Park, C., Ezzat, A. M., Yoo, J. Y., Hamilton, C. B., & Li, L. C. (2018) · JMIR mHealth and uHealth · doi:10.2196/10527 · Cited by 616 · see citing works
Interventions to Increase Physical Activity Among Healthy Adults: Meta-Analysis of Outcomes
Conn, V. S., Hafdahl, A. R., & Mehr, D. R. (2011) · American Journal of Public Health · doi:10.2105/ajph.2010.194381 · Cited by 486 · see citing works
Ferguson, T., Rowlands, A. V., Olds, T., & Maher, C. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0201-9 · Cited by 468 · see citing works
Translating Physical Activity Recommendations into a Pedometer-Based Step Goal
Marshall, S. J., Levy, S. S., Tudor-Locke, C. E., Kolkhorst, F. W., Wooten, K. M., Ji, M., Macera, C. A., & Ainsworth, B. E. (2009) · American Journal of Preventive Medicine · doi:10.1016/j.amepre.2009.01.021 · Cited by 447 · see citing works
Fitbit®: An accurate and reliable device for wireless physical activity tracking
Diaz, K. M., Krupka, D. J., Chang, M. J., Peacock, J., Ma, Y., Goldsmith, J., Schwartz, J. E., & Davidson, K. W. (2015) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.03.038 · Cited by 405 · see citing works
Validity of Consumer-Based Physical Activity Monitors
LEE, J.-M., KIM, Y., & WELK, G. J. (2014) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000287 · Cited by 401 · cited 3× · see citing works
“device energy expenditure measurements compared with reference standard criterion measures, including direct calorimetry [86,104] and indirect calorimetry [20,21,29-31,38,39,41-43,53,55,63,66,73,85,87,93,95,97,103, 105,116,117,129,130,142,143,146,148,159,165,166,177], in controlled settings.”
MethodAccuracy of wrist-worn heart rate monitors
Wang, R., Blackburn, G., Desai, M., Phelan, D., Gillinov, L., Houghtaling, P., & Gillinov, M. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.3340 · Cited by 386 · see citing works
Reliability and validity of ten consumer activity trackers
Kooiman, T. J. M., Dontje, M. L., Sprenger, S. R., Krijnen, W. P., van der Schans, C. P., & de Groot, M. (2015) · BMC Sports Science, Medicine and Rehabilitation · doi:10.1186/s13102-015-0018-5 · Cited by 360 · see citing works
Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise
GILLINOV, S., ETIWY, M., WANG, R., BLACKBURN, G., PHELAN, D., GILLINOV, A. M., HOUGHTALING, P., JAVADIKASGARI, H., & DESAI, M. Y. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001284 · Cited by 345 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Validation of the Fitbit One activity monitor device during treadmill walking
Takacs, J., Pollock, C. L., Guenther, J. R., Bahar, M., Napier, C., & Hunt, M. A. (2013) · Journal of science and medicine in sport · doi:10.1016/j.jsams.2013.10.241 · Cited by 322 · see citing works
Xie, J., Wen, D., Liang, L., Jia, Y., Gao, L., & Lei, J. (2018) · JMIR mHealth and uHealth · doi:10.2196/mhealth.9754 · Cited by 310 · see citing works
Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices
Rosenberger, M. E., Buman, M. P., Haskell, W. L., McConnell, M. V., & Carstensen, L. L. (2015) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000778 · Cited by 298 · see citing works
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · see citing works
Dooley, E. E., Golaszewski, N. M., & Bartholomew, J. B. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7043 · Cited by 242 · see citing works
Reliability and Validity of Ten Consumer Activity Trackers Depend on Walking Speed
Fokkema, T., Kooiman, T. J. M., Krijnen, W. P., van der Schans, C. P., & de Groot, M. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001146 · Cited by 212 · see citing works
Validity of Consumer-Based Physical Activity Monitors for Specific Activity Types
Nelson, M. B., Kaminsky, L. A., Dickin, D. C., & Montoye, A. H. K. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000933 · Cited by 206 · see citing works
Validity of Wearable Activity Monitors during Cycling and Resistance Exercise
BOUDREAUX, B. D., HEBERT, E. P., HOLLANDER, D. B., WILLIAMS, B. M., CORMIER, C. L., NAQUIN, M. R., GILLAN, W. W., GUSEW, E. E., & KRAEMER, R. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001471 · Cited by 205 · see citing works
Murakami, H., Kawakami, R., Nakae, S., Nakata, Y., Ishikawa-Takata, K., Tanaka, S., & Miyachi, M. (2016) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2016.0152 · Cited by 204 · see citing works
Adam Noah, J., Spierer, D. K., Gu, J., & Bronner, S. (2013) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2013.831135 · Cited by 202 · see citing works
Current State of Commercial Wearable Technology in Physical Activity Monitoring 2015-2017
Bunn, J., Navalta, J. W., Fountaine, C. J., & Reece, J. (2018) · International Journal of Exercise Science · doi:10.70252/njqx2719 · Cited by 194 · see citing works
Accuracy of 2 Activity Monitors in Detecting Steps in People With Stroke and Traumatic Brain Injury
Fulk, G. D., Combs, S. A., Danks, K. A., Nirider, C. D., Raja, B., & Reisman, D. S. (2014) · Physical Therapy · doi:10.2522/ptj.20120525 · Cited by 182 · see citing works
Tully, M. A., McBride, C., Heron, L., & Hunter, R. F. (2014) · BMC Research Notes · doi:10.1186/1756-0500-7-952 · Cited by 172 · see citing works
Lee, J. A., Williams, S. M., Brown, D. D., & Laurson, K. R. (2014) · Journal of Sports Sciences · doi:10.1080/02640414.2014.981848 · Cited by 169 · see citing works
Step Detection and Activity Recognition Accuracy of Seven Physical Activity Monitors
Storm, F. A., Heller, B. W., & Mazzà, C. (2015) · PLoS ONE · doi:10.1371/journal.pone.0118723 · Cited by 168 · see citing works
Validity of the Fitbit activity tracker for measuring steps in community-dwelling older adults
Paul, S. S., Tiedemann, A., Hassett, L. M., Ramsay, E., Kirkham, C., Chagpar, S., & Sherrington, C. (2015) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000013 · Cited by 165 · see citing works
Reddy, R. K., Pooni, R., Zaharieva, D. P., Senf, B., El Youssef, J., Dassau, E., Doyle III, F. J., Clements, M. A., Rickels, M. R., Patton, S. R., et al. (2018) · JMIR mHealth and uHealth · doi:10.2196/10338 · Cited by 157 · see citing works
Video-Recorded Validation of Wearable Step Counters under Free-living Conditions
TOTH, L. P., PARK, S., SPRINGER, C. M., FEYERABEND, M. D., STEEVES, J. A., & BASSETT, D. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001569 · Cited by 154 · see citing works
Sushames, A., Edwards, A., Thompson, F., McDermott, R., & Gebel, K. (2016) · PLOS ONE · doi:10.1371/journal.pone.0161224 · Cited by 152 · see citing works
Fortune, E., Lugade, V. A., Morrow, M. M. B., & Kaufman, K. R. (2014) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2014.02.006 · Cited by 150 · see citing works
Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking.
(2016) · PubMed Central · Cited by 150 · see citing works
Comparison of Consumer and Research Monitors under Semistructured Settings
BAI, Y., WELK, G. J., NAM, Y. H., LEE, J. A., LEE, J.-M., KIM, Y., MEIER, N. F., & DIXON, P. M. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000727 · Cited by 149 · see citing works
Assessment of the Fitbit Charge 2 for monitoring heart rate
Benedetto, S., Caldato, C., Bazzan, E., Greenwood, D. C., Pensabene, V., & Actis, P. (2018) · PLOS ONE · doi:10.1371/journal.pone.0192691 · Cited by 149 · see citing works
Continuous daily assessment of multiple sclerosis disability using remote step count monitoring
Block, V. J., Lizée, A., Crabtree‐Hartman, E., Bevan, C. J., Graves, J., Bove, R. M., Green, A. J., Nourbakhsh, B., Tremblay, M. A., Gourraud, P., et al. (2016) · Journal of Neurology · doi:10.1007/s00415-016-8334-6 · Cited by 148 · see citing works
Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR
Bai, Y., Hibbing, P. R., Mantis, K., & Welk, G. J. (2017) · Journal of Sports Sciences · doi:10.1080/02640414.2017.1412235 · Cited by 147 · see citing works
Effects of a 10,000 Steps per Day Goal in Overweight Adults
SCHNEIDER, P. L., Bassett, D. R., Thompson, D. L., Pronk, N. P., & Bielak, K. M. (2006) · American Journal of Health Promotion · doi:10.4278/0890-1171-21.2.85 · Cited by 144 · see citing works
Alharbi, M., Bauman, A. E., Neubeck, L., & Gallagher, R. D. (2016) · European Journal of Preventive Cardiology · doi:10.1177/2047487316634883 · Cited by 142 · see citing works
Falter, M., Budts, W., Goetschalckx, K., Cornelissen, V., & Buys, R. (2019) · JMIR mHealth and uHealth · doi:10.2196/11889 · Cited by 142 · see citing works
Validity of Different Activity Monitors to Count Steps in an Inpatient Rehabilitation Setting
Treacy, D., Hassett, L. M., Schurr, K., Chagpar, S., Paul, S. S., & Sherrington, C. (2017) · Physical Therapy · doi:10.1093/ptj/pzx010 · Cited by 141 · see citing works
Höchsmann, C., Knaier, R., Eymann, J., Hintermann, J., Infanger, D., & Schmidt‐Trucksäss, A. (2018) · Scandinavian Journal of Medicine and Science in Sports · doi:10.1111/sms.13074 · Cited by 138 · see citing works
Reid, R. E. R., Insogna, J. A., Carver, T. E., Comptour, A. M., Bewski, N. A., Sciortino, C. D., & Andersen, R. E. (2016) · Journal of science and medicine in sport · doi:10.1016/j.jsams.2016.10.015 · Cited by 133 · see citing works
Wahl, Y., Düking, P., Droszez, A., Wahl, P., & Mester, J. (2017) · Frontiers in Physiology · doi:10.3389/fphys.2017.00725 · Cited by 133 · see citing works
How valid are wearable physical activity trackers for measuring steps?
An, H., Jones, G. C., Kang, S., Welk, G. J., & Lee, J. M. (2016) · European Journal of Sport Science · doi:10.1080/17461391.2016.1255261 · Cited by 131 · see citing works
An easy approach to calculating estimated energy requirements.
Gerrior, S., Juan, W., & Peter, B. (2006) · PubMed · Cited by 129 · see citing works
Smart watches for heart rate assessment in atrial arrhythmias
Koshy, A. N., Sajeev, J., Nerlekar, N., Brown, A. J., Rajakariar, K., Zureik, M., Wong, M. C. G., Roberts, L., Street, M., Cooke, J. C., et al. (2018) · International Journal of Cardiology · doi:10.1016/j.ijcard.2018.02.073 · Cited by 127 · see citing works
A Comparison of Energy Expenditure Estimation of Several Physical Activity Monitors
Dannecker, K. L., Sazonova, N. A., Melanson, E. L., Sazonov, E. E., & Browning, R. C. (2013) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e318299d2eb · Cited by 124 · see citing works
Kroll, R. R., Boyd, J. G., & Maslove, D. M. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.6025 · Cited by 123 · see citing works
Validity of FitBit, Jawbone UP, Nike+ and other wearable devices for level and stair walking
Huang, Y., Xu, J., Yu, B., & Shull, P. B. (2016) · Gait & Posture · doi:10.1016/j.gaitpost.2016.04.025 · Cited by 122 · see citing works
Chu, A. H. Y., Ng, S. H.-X., Paknezhad, M., Gauterin, A., Koh, D. S. Q., Brown, M. S., & Müller‐Riemenschneider, F. (2017) · PLoS ONE · doi:10.1371/journal.pone.0172535 · Cited by 121 · see citing works
Imboden, M. T., Nelson, M. B., Kaminsky, L. A., & Montoye, A. H. K. (2017) · British Journal of Sports Medicine · doi:10.1136/bjsports-2016-096990 · Cited by 119 · see citing works
Chowdhury, E. A., Western, M. J., Nightingale, T. E., Peacock, O. J., & Thompson, D. (2017) · PLOS ONE · doi:10.1371/journal.pone.0171720 · Cited by 118 · see citing works
Validation of the Fitbit Wireless Activity Tracker for Prediction of Energy Expenditure
Sasaki, J. E., Hickey, A., Mavilia, M. G., Tedesco, J., John, D., Keadle, S. K., & Freedson, P. S. (2014) · Journal of Physical Activity and Health · doi:10.1123/jpah.2012-0495 · Cited by 116 · see citing works
Tedesco, S., Sica, M., Ancillao, A., Timmons, S., Barton, J., & O’Flynn, B. (2019) · PLOS ONE · doi:10.1371/journal.pone.0216891 · Cited by 116 · see citing works
Thomson, E. A., Nuss, K. J., Comstock, A., Reinwald, S., Blake, S., Pimentel, R. E., Tracy, B. L., & Li, K. (2019) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1560644 · Cited by 112 · see citing works
Understanding the error of our ways: Mapping the concepts of validity and reliability
Higgins, P. A., & Straub, A. J. (2006) · Nursing Outlook · doi:10.1016/j.outlook.2004.12.004 · Cited by 109 · see citing works
Floegel, T. A., Flórez-Pregonero, A., Hekler, E. B., & Buman, M. P. (2016) · The Journals of Gerontology Series A · doi:10.1093/gerona/glw098 · Cited by 108 · see citing works
Alley, S., Schöeppe, S., Guertler, D., Jennings, C. A., Duncan, M. J., & Vandelanotte, C. (2016) · BMJ Open · doi:10.1136/bmjopen-2016-011243 · Cited by 105 · see citing works
Dominick, G. M., Winfree, K. N., Pohlig, R. T., & Papas, M. A. (2016) · JMIR mhealth and uhealth · doi:10.2196/mhealth.6281 · Cited by 103 · see citing works
Almalki, M., Gray, K. M., & Martin-Sanchez, F. J. (2015) · Health Information Science and Systems · doi:10.1186/2047-2501-3-s1-s1 · Cited by 98 · see citing works
Gorny, A. W., Liew, S. J., Tan, C. S., & Müller-Riemenschneider, F. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.8233 · Cited by 94 · see citing works
Validity and Reliability of the Apple Watch for Measuring Heart Rate During Exercise
Khushhal, A., Nichols, S., Evans, W., Gleadall-Siddall, D., Page, R., O'Doherty, A., Carroll, S., Ingle, L., & Abt, G. (2017) · Sports Medicine International Open · doi:10.1055/s-0043-120195 · Cited by 92 · see citing works
Chow, J. J., Thom, J. M., Wewege, M. A., Ward, R. E., & Parmenter, B. J. (2017) · Gait & Posture · doi:10.1016/j.gaitpost.2017.06.012 · Cited by 91 · see citing works
Accuracy of Wristband Activity Monitors during Ambulation and Activities
CHEN, M.-D., KUO, C.-C., PELLEGRINI, C. A., & HSU, M.-J. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000984 · Cited by 90 · see citing works
Gomersall, S. R., Ng, N., Burton, N. W., Pavey, T., Gilson, N. D., & Brown, W. J. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.5531 · Cited by 90 · see citing works
Klassen, T. D., Simpson, L. A., Lim, S. B., Louie, D. R., Parappilly, B., Sakakibara, B. M., Zbogar, D., & Eng, J. J. (2015) · Physical Therapy · doi:10.2522/ptj.20140611 · Cited by 89 · see citing works
Brewer, W., Swanson, B. T., & Ortíz, A. (2017) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2017-000254 · Cited by 89 · see citing works
Price, K. J., Bird, S. R., Lythgo, N., Raj, I. S., Wong, J., & Lynch, C. (2016) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2016.1253795 · Cited by 87 · see citing works
The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers
Cadmus‐Bertram, L., Gangnon, R. E., Wirkus, E. J., Thraen-Borowski, K. M., & Gorzelitz, J. (2017) · Annals of Internal Medicine · doi:10.7326/l16-0353 · Cited by 80 · see citing works
Simpson, L. A., Eng, J. J., Klassen, T. D., Lim, S. B., Louie, D. R., Parappilly, B., Sakakibara, B. M., & Zbogar, D. (2015) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-1993 · Cited by 79 · see citing works
Concurrent Validity of Wearable Activity Trackers Under Free-Living Conditions
Brooke, S. M., An, H.-S., Kang, S.-K., Noble, J. M., Berg, K. E., & Lee, J.-M. (2017) · Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000001571 · Cited by 79 · see citing works
Alinia, P., Cain, C. R., Fallahzadeh, R., Shahrokni, A., Cook, D. J., & Ghasemzadeh, H. G. (2017) · JMIR mhealth and uhealth · doi:10.2196/mhealth.6321 · Cited by 77 · see citing works
Accuracy of wearable physical activity trackers in people with Parkinson’s disease
Lamont, R. M., Daniel, H. L., Payne, C. L., & Brauer, S. (2018) · Gait & Posture · doi:10.1016/j.gaitpost.2018.04.034 · Cited by 75 · see citing works
Burton, E., Hill, K., Lautenschlager, N. T., Thøgersen‐Ntoumani, C., Lewin, G. F., Boyle, E., & Howie, E. K. (2018) · BMC Geriatrics · doi:10.1186/s12877-018-0793-4 · Cited by 71 · see citing works
Hwang, J., Kim, J., Choi, K., Cho, M. S., Nam, G., & Kim, Y. (2019) · Korean Circulation Journal · doi:10.4070/kcj.2018.0323 · Cited by 66 · see citing works
Klassen, T. D., Semrau, J. A., Dukelow, S. P., Bayley, M. T., Hill, M. D., & Eng, J. J. (2017) · Stroke · doi:10.1161/strokeaha.117.018175 · Cited by 65 · see citing works
Leth, S., Hansen, J., Nielsen, O., & Dinesen, B. (2017) · Sensors · doi:10.3390/s17010211 · Cited by 64 · see citing works
An Evaluation of Commercial Pedometers for Monitoring Slow Walking Speed Populations
Beevi, F. H. A., Miranda, J., Pedersen, C. F., & Wagner, S. (2015) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2015.0120 · Cited by 62 · see citing works
Haghayegh, S., Khoshnevis, S., Smolensky, M. H., & Diller, K. R. (2019) · Chronobiology International · doi:10.1080/07420528.2019.1596947 · Cited by 62 · see citing works
Validity of Commercial Activity Trackers in Children With Congenital Heart Disease
Voss, C., Gardner, R. F., Dean, P. H., & Harris, K. C. (2016) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2016.11.024 · Cited by 61 · see citing works
Macridis, S., Johnston, N., Johnson, S. T., & Vallance, J. K. (2018) · PLoS ONE · doi:10.1371/journal.pone.0189298 · Cited by 60 · see citing works
Gusmer, R., Bosch, T. A., Watkins, A. N., Ostrem, J. D., & Dengel, D. R. (2014) · The Open Sports Medicine Journal · doi:10.2174/1874387001408010011 · Cited by 59 · see citing works
Mobile Device Accuracy for Step Counting Across Age Groups
Modave, F. P., Guo, Y. M., Bian, J., Gurka, M. J., Parish, A., Smith, M. D., Lee, A. M., & Buford, T. W. (2017) · JMIR mhealth and uhealth · doi:10.2196/mhealth.7870 · Cited by 57 · see citing works
Balto, J. M., Kinnett-Hopkins, D., & Motl, R. W. (2016) · Multiple Sclerosis Journal - Experimental Translational and Clinical · doi:10.1177/2055217316634754 · Cited by 56 · see citing works
Validity of wrist-worn consumer products to measure heart rate and energy expenditure
Thiebaud, R. S., Funk, M. D., Patton, J. C., Massey, B. L., Shay, T. E., Schmidt, M. G., & Giovannitti, N. (2018) · Digital Health · doi:10.1177/2055207618770322 · Cited by 56 · see citing works
Collins, J. E., Yang, H. Y., Trentadue, T. P., Gong, Y., & Losina, E. (2019) · PLOS ONE · doi:10.1371/journal.pone.0211231 · Cited by 55 · see citing works
Validation of the Fitbit Zip for monitoring physical activity among free-living adolescents
Schneider, M. L., & Chau, L. (2016) · BMC Research Notes · doi:10.1186/s13104-016-2253-6 · Cited by 54 · see citing works
Step count accuracy and reliability of two activity tracking devices in people after stroke
Schaffer, S. D., Holzapfel, S. D., Fulk, G., & Bosch, P. R. (2017) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2017.1354412 · Cited by 54 · see citing works
Støve, M. P., Haucke, E., Nymann, M. L., Sigurdsson, T., & Larsen, B. T. (2018) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1535563 · Cited by 51 · see citing works
Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities
Claes, J., Buys, R., Ávila, A., Finlay, D., Kennedy, A., Guldenring, D., Budts, W., & Cornelissen, V. (2017) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2017.1333166 · Cited by 49 · see citing works
Physical activity among children: objective measurements using Fitbit One® and ActiGraph
Hamari, L., Kullberg, T., Ruohonen, J., Heinonen, O. J., Díaz-Rodríguez, N., Lilius, J., Pakarinen, A., Myllymäki, A., Leppänen, V., & Salanterä, S. (2017) · BMC Research Notes · doi:10.1186/s13104-017-2476-1 · Cited by 48 · see citing works
Tracking Steps on Apple Watch at Different Walking Speeds
Veerabhadrappa, P., Moran, M. D., Renninger, M. D., Rhudy, M. B., Dreisbach, S. B., & Gift, K. M. (2018) · Journal of General Internal Medicine · doi:10.1007/s11606-018-4332-y · Cited by 48 · see citing works
Sañudo, B., De Hoyo, M., Muñoz-López, A., Perry, J., & Abt, G. (2019) · Journal of Medical Systems · doi:10.1007/s10916-019-1325-2 · Cited by 48 · cited 2× · see citing works
“A total of 29 studies (266 comparisons) examined wearable device heart rate measurements compared with reference standard criterion measures, including electrocardiography [22,23,38-40,43,44,54,61,62,67-70,142,162,176], Polar brand chest straps [20,21,24-28,58,63,71,163], and pulse oximetry [66], in controlled setting…”
BackgroundThe bit doesn’t fit: Evaluation of a commercial activity-tracker at slower walking speeds
Wong, C. K., Mentis, H. M., & Kuber, R. (2018) · Gait & Posture · doi:10.1016/j.gaitpost.2017.10.010 · Cited by 46 · see citing works
Singh, A. K., Farmer, C., van den Berg, M., Killington, M., & Barr, C. (2015) · Disability and health journal · doi:10.1016/j.dhjo.2015.10.011 · Cited by 45 · see citing works
Gruwez, A., Libert, W., Ameye, L., & Bruyneel, M. (2017) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2017.03.008 · Cited by 45 · see citing works
Thorup, C. B., Andreasen, J. J., Sørensen, E. E., Grønkjær, M., Dinesen, B., & Hansen, J. (2017) · BMJ Open · doi:10.1136/bmjopen-2016-011742 · Cited by 45 · see citing works
The Validity of Consumer-Level Activity Monitors in Healthy Older Adults in Free-Living Conditions
Farina, N., & Lowry, R. (2017) · Journal of Aging and Physical Activity · doi:10.1123/japa.2016-0344 · Cited by 45 · see citing works
Validity of sports watches when estimating energy expenditure during running
Roos, L., Taube, W., Beeler, N., & Wyss, T. (2017) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-017-0089-6 · Cited by 45 · see citing works
Validity and Wearability of Consumer-based Fitness Trackers in Free-living Children
Brazendale, K., Decker, L., Hunt, E. T., Perry, M. W., Brazendale, A. B., Weaver, R. G., & Beets, M. W. (2019) · International journal of exercise science · doi:10.70252/bbtu4889 · Cited by 44 · see citing works
Park, W., Lee, V. J., Ku, B., & Tanaka, H. (2014) · Journal of Exercise Science & Fitness · doi:10.1016/j.jesf.2014.01.003 · Cited by 42 · see citing works
Ummels, D., Beekman, E., Theunissen, K., Braun, S., & Beurskens, A. J. H. M. (2018) · JMIR mhealth and uhealth · doi:10.2196/mhealth.8524 · Cited by 42 · see citing works
National Cancer Institute (2020) · Definitions · doi:10.32388/t0uxhu · Cited by 42 · see citing works
The Fitbit One Physical Activity Tracker in Men With Prostate Cancer: Validation Study
Blarigan, E. L. V., Kenfield, S. A., Tantum, L. K., Cadmus‐Bertram, L., Carroll, P. R., & Chan, J. M. (2017) · JMIR Cancer · doi:10.2196/cancer.6935 · Cited by 41 · see citing works
The validity and inter-device variability of the Apple Watch™ for measuring maximal heart rate
Abt, G., Bray, J., & Benson, A. C. (2018) · Journal of Sports Sciences · doi:10.1080/02640414.2017.1397282 · Cited by 41 · see citing works
Mahendran, N., Kuys, S. S., Downie, E., Ng, P., & Brauer, S. (2016) · Brain Impairment · doi:10.1017/brimp.2016.13 · Cited by 40 · see citing works
When a Step Is Not a Step! Specificity Analysis of Five Physical Activity Monitors
O’Connell, S., OLaighin, G. M., & Quinlan, L. R. (2017) · PLoS ONE · doi:10.1371/journal.pone.0169616 · Cited by 40 · see citing works
Validity of the Fitbit One for Measuring Activity in Community-Dwelling Stroke Survivors
Hui, J., Heyden, R., Bao, T., Accettone, N., McBay, C., Richardson, J. A., & Tang, A. (2017) · Physiotherapy Canada · doi:10.3138/ptc.2016-40.ep · Cited by 40 · see citing works
Accuracy of Activity Trackers in Parkinson Disease: Should We Prescribe Them?
Wendel, N. C., Macpherson, C. E., Webber, K., Hendron, K. L., DeAngelis, T. R., Colón-Semenza, C., & Ellis, T. D. (2018) · Physical Therapy · doi:10.1093/ptj/pzy054 · Cited by 39 · see citing works
Influence of Pedometer Position on Pedometer Accuracy at Various Walking Speeds: A Comparative Study
Ehrler, F., Weber, C., & Lovis, C. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.5916 · Cited by 37 · see citing works
Montoye, A. H. K., Mitrzyk, J. R., & Molesky, M. J. (2017) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2017.1331166 · Cited by 37 · see citing works
Accuracy of the vivofit activity tracker
Alsubheen, S., George, A. M., Baker, A., Rohr, L. E., & Basset, F. A. (2016) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2016.1193238 · Cited by 36 · see citing works
Rozanski, G. M., Aqui, A., Sivakumaran, S., & Mansfield, A. (2018) · JMIR Cardio · doi:10.2196/cardio.8199 · Cited by 36 · see citing works
Mobile Phone-Connected Wearable Motion Sensors to Assess Postoperative Mobilization
Appelboom, G., Taylor, B. E. S., Bruce, E. M., Bassile, C. C., Malakidis, C., Yang, A., Youngerman, B., D’Amico, R. S., Bruce, S., Bruyère, O., et al. (2015) · JMIR mhealth and uhealth · doi:10.2196/mhealth.3785 · Cited by 35 · see citing works
Boeselt, T., Spielmanns, M., Nell, C., Storre, J. H., Windisch, W., Magerhans, L., Beutel, B., Kenn, K., Greulich, T., Alter, P., et al. (2016) · PLOS ONE · doi:10.1371/journal.pone.0157229 · Cited by 35 · see citing works
A Comparison of Accelerometer Accuracy in Older Adults
Phillips, L. J., Petroski, G. F., & Markis, N. E. (2015) · Research in Gerontological Nursing · doi:10.3928/19404921-20150429-03 · Cited by 34 · see citing works
VALIDITY AND RELIABILITY OF THE FITBIT FLEX™ AND ACTIGRAPH GT3X+ AT JOGGING AND RUNNING SPEEDS
Jones, D., Crossley, K. M., Dascombe, B. J., Hart, H. F., & Kemp, J. L. (2018) · International Journal of Sports Physical Therapy · doi:10.26603/ijspt20180860 · Cited by 33 · see citing works
The Accuracy of Pedometers in Measuring Walking Steps on a Treadmill in College Students
Husted, H. M., & Llewellyn, T. L. (2017) · International journal of exercise science · doi:10.70252/ftda3470 · Cited by 31 · see citing works
Accuracy of a Wrist-Worn Heart Rate Sensing Device during Elective Pediatric Surgical Procedures
Pelizzo, G., Guddo, A., Puglisi, A., Silvestri, A. D., Comparato, C., Valenza, M., Bordonaro, E., & Calcaterra, V. (2018) · Children · doi:10.3390/children5030038 · Cited by 31 · see citing works
Duclos, N. C., Aguiar, L. T., Aïssaoui, R., Faria, C. D. C. D. M., Nadeau, S. M., & Duclos, C. (2019) · PM&R · doi:10.1002/pmrj.12080 · Cited by 31 · see citing works
Comparison of wrist-worn and hip-worn activity monitors under free living conditions
Hargens, T. A., Deyarmin, K. N., Snyder, K. M., Mihalik, A. G., & Sharpe, L. E. (2017) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2016.1271046 · Cited by 30 · see citing works
Mooses, K., Oja, M., Reisberg, S., Vilo, J., & Kull, M. (2018) · BMC Public Health · doi:10.1186/s12889-018-5752-7 · Cited by 30 · see citing works
Validation of the Fitbit One® for physical activity measurement at an upper torso attachment site
Diaz, K. M., Krupka, D. J., Chang, M. J., Shaffer, J. A., Ma, Y., Goldsmith, J., Schwartz, J. E., & Davidson, K. W. (2016) · BMC Research Notes · doi:10.1186/s13104-016-2020-8 · Cited by 29 · see citing works
Accuracy of Consumer Monitors for Estimating Energy Expenditure and Activity Type
WOODMAN, J. A., CROUTER, S. E., BASSETT, D. R., FITZHUGH, E. C., & BOYER, W. R. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001090 · Cited by 29 · see citing works
Kendall, B., Bellovary, B. N., & Gothe, N. P. (2018) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1481723 · Cited by 29 · see citing works
Breteler, M. J. M., Janssen, J. H., Spiering, W., Kalkman, C. J., van Solinge, W. W., & Dohmen, D. (2019) · JMIR Formative Research · doi:10.2196/11489 · Cited by 29 · see citing works
Accuracy and Acceptability of Commercial-Grade Physical Activity Monitors in Older Adults
Hergenroeder, A. L., Gibbs, B. B., Kotlarczyk, M. P., Perera, S., Kowalsky, R. J., & Brach, J. S. (2018) · Journal of Aging and Physical Activity · doi:10.1123/japa.2018-0036 · Cited by 28 · see citing works
Methods for Activity Monitor Validation Studies: An Example With the Fitbit Charge
DeShaw, K. J., Ellingson, L. D., Bai, Y., Lansing, J. E., Perez, M., & Welk, G. J. (2018) · Journal for the Measurement of Physical Behaviour · doi:10.1123/jmpb.2018-0017 · Cited by 28 · see citing works
A Validation Study of the Fitbit One in Daily Life Using Different Time Intervals
Middelweerd, A., van der Ploeg, H. P., van Halteren, A., Twisk, J. W. R., Brug, J., & Velde, S. J. T. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001225 · Cited by 27 · see citing works
Gaz, D. V., Rieck, T. M., Peterson, N. W., Ferguson, J. A., Schroeder, D. R., Dunfee, H. A., Henderzahs-Mason, J. M., & Hagen, P. T. (2018) · American Journal of Health Promotion · doi:10.1177/0890117118763273 · Cited by 27 · see citing works
Validity of Garmin Vívofit 1 and Garmin Vívofit 3 for School-Based Physical Activity Monitoring
Šimůnek, A., Dygrýn, J., Jakubec, L., Neuls, F., Frömel, K., & Welk, G. J. (2018) · Pediatric Exercise Science · doi:10.1123/pes.2018-0019 · Cited by 27 · see citing works
Morris, C. E., Wessel, P., Tinius, R. A., Schafer, M. A., & Maples, J. M. (2019) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.2019.1603989 · Cited by 27 · see citing works
Validity and reliability of the Fitbit Zip as a measure of preschool children’s step count
Sharp, C., Mackintosh, K. A., Erjavec, M., Pascoe, D., & Horne, P. J. (2017) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2017-000272 · Cited by 25 · see citing works
O’Connell, S., OLaighin, G. M., Kelly, L. S., Murphy, E., Beirne, S., Burke, N., Kilgannon, O., & Quinlan, L. R. (2016) · PLoS ONE · doi:10.1371/journal.pone.0154956 · Cited by 24 · see citing works
Validation study of Polar V800 accelerometer
Hernández-Vicente, A., Santos-Lozano, A., De Cocker, K., & Garatachea, N. (2016) · Annals of Translational Medicine · doi:10.21037/atm.2016.07.16 · Cited by 23 · see citing works
Šimůnek, A., Dygrýn, J., Gába, A., Jakubec, L., Stelzer, J., & Chmelík, F. (2016) · Acta Gymnica · doi:10.5507/ag.2016.014 · Cited by 23 · see citing works
Assessment of step accuracy using the Consumer Technology Association standard
Bunn, J. A., Jones, C., Oliviera, A., & Webster, M. J. (2018) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1491941 · Cited by 23 · see citing works
Validation of Electronic Activity Monitor Devices During Treadmill Walking
Tam, K. M., & Cheung, S. Y. (2018) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2017.0263 · Cited by 22 · see citing works
Nuss, K. J., Thomson, E. A., Courtney, J. B., Comstock, A., Reinwald, S., Blake, S., Richard, E., Tracy, B. L., & Li, K. (2019) · American Journal of Health Behavior · doi:10.5993/ajhb.43.3.5 · Cited by 22 · see citing works
Comparative accuracy of fitness tracking modalities in quantifying energy expenditure
Dondzila, C. J., & Garner, D. P. (2016) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2016.1197978 · Cited by 21 · see citing works
Schmal, H., Holsgaard‐Larsen, A., Izadpanah, K., Brønd, J. C., Madsen, C. F., & Lauritsen, J. L. (2018) · Rehabilitation Research and Practice · doi:10.1155/2018/3521271 · Cited by 21 · see citing works
Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers
Cadmus‐Bertram, L., Gangnon, R. E., Wirkus, E. J., Thraen-Borowski, K. M., & Gorzelitz, J. (2017) · Annals of Internal Medicine · doi:10.7326/l17-0380 · Cited by 21 · see citing works
Arch, E. S., Sions, J. M., Horne, J. R., & Bodt, B. (2018) · Prosthetics and Orthotics International · doi:10.1177/0309364618767138 · Cited by 20 · see citing works
Validation of Four Smartwatches in Energy Expenditure and Heart Rate Assessment During Exergaming
Pope, Z., Lee, J. E., Zeng, N., & Gao, Z. (2019) · Games for Health Journal · doi:10.1089/g4h.2018.0087 · Cited by 20 · see citing works
The accuracy of personal activity monitoring devices
Battenberg, A. K., Donohoe, S., Robertson, N. B., & Schmalzried, T. P. (2017) · Seminars in Arthroplasty JSES · doi:10.1053/j.sart.2017.07.006 · Cited by 19 · see citing works
Validation of the Fitbit Zip and Fitbit Flex with pregnant women in free-living conditions
St-Laurent, A., Mony, M., Mathieu, M.-E., & Ruchat, S.-M. (2018) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2018.1472822 · Cited by 19 · see citing works
Validity and Reliability Evaluation of Four Commercial Activity Trackers' Step Counting Performance
Tophøj, K. H., Petersen, M. G., Sæbye, C., Baad-Hansen, T., & Wagner, S. (2018) · Telemedicine and e-Health · doi:10.1089/tmj.2017.0264 · Cited by 18 · see citing works
Validation of the Fitbit Flex in an Acute Post–Cardiac Surgery Patient Population
Daligadu, J., Pollock, C. L., Carlaw, K., Chin, M., Haynes, A., Kopal, T. T., Tahsinul, A., Walters, K., & Colella, T. J. F. (2018) · Physiotherapy Canada · doi:10.3138/ptc.2017-34 · Cited by 18 · see citing works
Evaluation of Earbud and Wristwatch Heart Rate Monitors during Aerobic and Resistance Training
Bunn, J. A., Wells, E. K., Manor, J., & Webster, M. J. (2019) · International journal of exercise science · doi:10.70252/loox1045 · Cited by 18 · see citing works
Prieto-Centurion, V., Bracken, N. E., Norwick, L. M., Zaidi, F., Mutso, A. A., Morken, V., Coultas, D., Rand, C. S., Márquez, D. X., & Krishnan, J. A. (2016) · Chronic Obstructive Pulmonary Diseases Journal of the COPD Foundation · doi:10.15326/jcopdf.3.3.2015.0164 · Cited by 17 · see citing works
Congruent Accuracy of Wrist-worn Activity Trackers during Controlled and Free-living Conditions
Dondzila, C. J., Lewis, C., Lopez, J. R., & Parker, T. (2018) · International Journal of Exercise Science · doi:10.70252/ubbt6886 · Cited by 16 · cited 3× · see citing works
“[20,21,29,32,34,36,38,45,53,64-82] Wrist Discontinued 2015 Charge HR Fitbit”
BackgroundChandrasekar, A., Hensor, E. M. A., Mackie, S. L., Backhouse, M. R., & Harris, E. (2018) · Gait & Posture · doi:10.1016/j.gaitpost.2018.01.035 · Cited by 15 · see citing works
Accuracy of swimming wearable watches for estimating energy expenditure
Lee, M., Lee, H., & Park, S. (2018) · IJASS(International Journal of Applied Sports Sciences) · doi:10.24985/ijass.2018.30.1.80 · Cited by 15 · see citing works
Measuring the validity and reliability of the Apple Watch as a physical activity monitor
Zhang, P., Godin, S., & Owens, M. V. (2019) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.18.08339-1 · Cited by 15 · see citing works
The validity of using activity monitors to detect step count after lumbar fusion surgery
Gilmore, S. J., Davidson, M., Hahne, A. J., & McClelland, J. A. (2018) · Disability and Rehabilitation · doi:10.1080/09638288.2018.1509140 · Cited by 14 · see citing works
Measuring Reliability of Movement With Accelerometry: Fitbit ® Versus ActiGraph ®
Rowe, V. T., & Neville, M. (2019) · American Journal of Occupational Therapy · doi:10.5014/ajot.2019.030692 · Cited by 14 · see citing works
Clevenger, K. A., Molesky, M. J., Vusich, J., & Montoye, A. H. K. (2019) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2019.1577737 · Cited by 13 · see citing works
Wearable Device Validity in Determining Step Count During Hiking and Trail Running
Navalta, J. W., Montes, J., Bodell, N. G., Aguilar, C. D., Lujan, A., Guzman, G., Kam, B. K., Manning, J. W., & DeBeliso, M. (2018) · Journal for the Measurement of Physical Behaviour · doi:10.1123/jmpb.2018-0010 · Cited by 12 · see citing works
Sala, D. A., Grissom, H., DelSole, E. M., Chu, M. L., Godfried, D. H., Bhattacharyya, S., Karamitopoulos, M. S., & Chu, A. (2019) · Developmental Medicine & Child Neurology · doi:10.1111/dmcn.14207 · Cited by 12 · see citing works
The validity and accuracy of wrist-worn activity monitors in lower-limb prosthesis users
Smith, J. D., Guerra, G., & Burkholder, B. G. (2019) · Disability and Rehabilitation · doi:10.1080/09638288.2019.1587792 · Cited by 12 · see citing works
Activity trackers are not valid for step count registration when walking with crutches
Ridder, R. D., & Blaiser, C. D. (2019) · Gait & Posture · doi:10.1016/j.gaitpost.2019.02.009 · Cited by 11 · see citing works
Accuracy of the Fitbit Zip for Measuring Steps for Adolescents With Visual Impairments
Haegele, J. A., Brian, A., & Wolf, D. L. (2017) · Adapted Physical Activity Quarterly · doi:10.1123/apaq.2016-0055 · Cited by 8 · see citing works
Sears, T., Alvalos, E., Lawson, S., McAlister, I. P., Eschbach, L. C., & Bunn, J. A. (2017) · International journal of exercise science · Cited by 8 · see citing works
Boolani, A., Towler, C., LeCours, B., Blank, H., LaRue, J. L., & Fulk, G. D. (2019) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/cpt.0000000000000107 · Cited by 6 · see citing works
Impact of EPOC adjustment on estimation of energy expenditure using activity monitors
Přibyslavská, V., Caputo, J. L., Coons, J. M., & Barry, V. (2018) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2018.1472823 · Cited by 5 · see citing works
Fitbit Flex Energy Expenditure and Step Count Evaluation
Montes, J., Young, J. C., Tandy, R. D., & Navalta, J. W. (2017) · Cited by 5 · see citing works
Poojary, J., Arora, E., Britto, A., Polen, Z., Arena, R. A., & Babu, A. S. (2018) · Mayo Clinic Proceedings · doi:10.1016/j.mayocp.2018.09.003 · Cited by 4 · see citing works
(2011) · SpringerReference · doi:10.1007/springerreference_6922 · Cited by 2 · see citing works
Fuller, D. (2019) · Harvard Dataverse · doi:10.7910/dvn/o7gqim · Cited by 2 · see citing works
Self-citationValidation Of A Self-monitoring Tool For Use In Post-concussion Syndrome Therapy
Powierza, C. S., Clark, M. D., Hughes, J. M., Carneiro, K. A., & Mihalik, J. P. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000487390.22076.d1 · Cited by 1 · see citing works
Flagged first, then by citations
1 of the 180 references couldn’t be matched to a known paper.
Research works printed but not linked: 14
- CostelloKGartner Says Worldwide Wearable Device Sales to Grow 26 Percent in 2019Gartner, Inc20182020-08-17https://www.gartner.com/en/newsroom/press-releases/2018-11-29-gartner-says-worldwide-wearable-device-sales-to-grow-
- Bunn, JA, 2018, Int J Exerc Sci, 11, 503
- Swinscow, T, 2009, Statistics at Square One
- Bunn, J, 2019, Int J Exerc Sci, 12, 374
- Sears, T, 2017, International Journal of Exercise Science, 10, 764
- Husted, HM, 2017, Int J Exerc Sci, 10, 146
- Keating, XD, 2012, ICHPER-SD Journal of Research, 9, 28
- Dondzila, C, 2018, International Journal of Exercise Science, 11, 575
- Jo, E, 2016, J Sports Sci Med, 15, 540
- Brazendale, K, 2019, Int J Exerc Sci, 12, 471
- Montes, J, 2017, Journal Exercise Physiology Online, 20, 134
- Stackpool, C, 2014, J Fit Res, 3, 32
- Gerrior, S, 2006, Prev Chronic Dis, 3, A129
- RallsJFitbit to Be Acquired by GoogleFitbit20192019-11-01https://investor.fitbit.com/press/press-releases/press-release-details/2019/Fitbit-to-Be-Acquired-by-Google/default.aspx
Cited by
Sorted by citations
The impact of wearable technologies in health research: scoping review
Huhn, S., Axt, M., Gunga, H.-C., Maggioni, M. A., Munga, S., Obor, D., Sié, A., Boudo, V., Bunker, A., Sauerborn, R., et al. (2022) · JMIR mHealth and uHealth · doi:10.2196/34384 · Cited by 379 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Kim, K. B., & Baek, H. J. (2023) · Electronics · doi:10.3390/electronics12132923 · Cited by 238 · see citing works
Master, H., Annis, J., Huang, S., Beckman, J. A., Ratsimbazafy, F., Marginean, K., Carroll, R., Natarajan, K., Harrell, F. E., Roden, D. M., et al. (2022) · Nature Medicine · doi:10.1038/s41591-022-02012-w · Cited by 221 · see citing works
Physical Activity Over the Lifecourse and Cardiovascular Disease
Perry, A., Dooley, E. E., Master, H., Spartano, N. L., Brittain, E. L., & Gabriel, K. P. (2023) · Circulation Research · doi:10.1161/circresaha.123.322121 · Cited by 194 · see citing works
Remote Monitoring for Heart Failure Management at Home
Stevenson, L. W., Ross, H. J., Rathman, L. D., & BOEHMER, J. P. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.04.010 · Cited by 189 · see citing works
Wearable activity trackers–advanced technology or advanced marketing?
Shei, R., Holder, I. G., Oumsang, A. S., Paris, B. A., & Paris, H. L. R. (2022) · European Journal of Applied Physiology · doi:10.1007/s00421-022-04951-1 · Cited by 169 · see citing works
A Survey on Wearable Sensors for Mental Health Monitoring
Gomes, N., Pato, M. P. M., Lourenço, A., & Datia, N. (2023) · Sensors · doi:10.3390/s23031330 · Cited by 147 · see citing works
Wearable inertial sensors for human movement analysis: a five-year update
Picerno, P., Iosa, M., D’Souza, C., Benedetti, M. G., Paolucci, S., & Morone, G. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.1988849 · Cited by 146 · see citing works
Technologies in Home-Based Digital Rehabilitation: Scoping Review
Arntz, A. R., Weber, F., Handgraaf, M., Lällä, K., Korniloff, K., Murtonen, K.-P., Chichaeva, J., Kidritsch, A., Heller, M., Sakellari, Ε., et al. (2023) · JMIR Rehabilitation and Assistive Technologies · doi:10.2196/43615 · Cited by 138 · see citing works
Doherty, C., Baldwin, M., Keogh, A., Caulfield, B., & Argent, R. (2024) · Sports Medicine · doi:10.1007/s40279-024-02077-2 · Cited by 114 · see citing works
AI-Enabled Wearable Medical Internet of Things in Healthcare System: A Survey
Subhan, F., Mirza, A., Su’ud, M. B. M., Alam, M. M., Nisar, S., Habib, U., & Iqbal, M. Z. (2023) · Applied Sciences · doi:10.3390/app13031394 · Cited by 117 · see citing works
Consumer Wearable Health and Fitness Technology in Cardiovascular Medicine
Petek, B. J., Al‐Alusi, M. A., Moulson, N., Grant, A. J., Besson, C., Guseh, J. S., Wasfy, M. M., Grémeaux, V., Churchill, T. W., & Baggish, A. L. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.04.054 · Cited by 106 · see citing works
Wearable Devices for Respiratory Monitoring
Vicente, B. A., Sebastião, R., & Sencadas, V. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202404348 · Cited by 96 · see citing works
Wearable and Biodegradable Sensors for Clinical and Environmental Applications
Baldo, T. A., de Lima, L. F., Mendes, L. F., de Araujo, W. R., Paixão, T. R. L. C., & Coltro, W. K. T. (2020) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.0c00735 · Cited by 93 · see citing works
Verdicchio, C. V., Freene, N., Hollings, M., Maiorana, A., Briffa, T. G., Gallagher, R. D., Hendriks, J., Abell, B., Brown, A. D. H., Colquhoun, D., et al. (2023) · Heart Lung and Circulation · doi:10.1016/j.hlc.2023.06.854 · Cited by 93 · see citing works
Chandrasekaran, R., Katthula, V., & Moustakas, E. (2021) · Health Informatics Journal · doi:10.1177/14604582211058073 · Cited by 90 · see citing works
Sanal‐Hayes, N. E. M., Mclaughlin, M., Hayes, L. D., Mair, J. L., Ormerod, J. A., Carless, D., Hilliard, N., Meach, R., Ingram, J., & Sculthorpe, N. F. (2023) · Journal of Translational Medicine · doi:10.1186/s12967-023-04587-5 · Cited by 90 · see citing works
A roadmap for the development of human body digital twins
Tang, C., Yi, W., Occhipinti, E., Dai, Y., Gao, S., & Occhipinti, L. G. (2024) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-024-00025-w · Cited by 90 · see citing works
Wang, W., Cheng, J., Song, W., & Shen, Y. (2022) · JMIR mhealth and uhealth · doi:10.2196/32435 · Cited by 89 · see citing works
Engineering the Comfort‐of‐Wear for Next Generation Wearables
Shimura, T., Sato, S., Zalar, P., & Matsuhisa, N. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202200512 · Cited by 86 · see citing works
Lugones-Sanchez, C., Recio-Rodriguez, J. I., Agudo-Conde, C., Repiso-Gento, I., G Adalia, E., Ramirez-Manent, J. I., Sanchez-Calavera, M. A., Rodriguez-Sanchez, E., Gomez-Marcos, M. A., Garcia-Ortiz, L., et al. (2022) · Journal of Medical Internet Research · doi:10.2196/30416 · Cited by 85 · see citing works
Hajj‐Boutros, G., Landry-Duval, M.-A., Comtois, A. S., Gouspillou, G., & Karelis, A. D. (2021) · European Journal of Sport Science · doi:10.1080/17461391.2021.2023656 · Cited by 84 · see citing works
Chevance, G., Golaszewski, N. M., Tipton, E., Hekler, E. B., Buman, M. P., Welk, G. J., Patrick, K., & Godino, J. (2022) · JMIR mhealth and uhealth · doi:10.2196/35626 · Cited by 77 · see citing works
Ngueleu, A.-M., Barthod, C., Best, K. L., Routhier, F., Otis, M. J.-D., & Batcho, C. S. (2022) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-022-01085-5 · Cited by 73 · see citing works
Migliaccio, G. M., Padulo, J., & Russo, L. (2024) · Applied Sciences · doi:10.3390/app14156649 · Cited by 71 · see citing works
He, D., Wang, Z., Khalil, E. B., Donmez, B., Qiao, G., & Kumar, S. (2022) · Transportation Research Record Journal of the Transportation Research Board · doi:10.1177/03611981221090937 · Cited by 69 · see citing works
Köhler, C., Bartschke, A., Fürstenau, D., Schaaf, T., & Salgado, E. (2024) · Journal of Medical Internet Research · doi:10.2196/58936 · Cited by 68 · see citing works
Patel, V., Orchanian‐Cheff, A., & Wu, R. C. (2021) · JMIR mhealth and uhealth · doi:10.2196/17411 · Cited by 65 · see citing works
Wang, Y., Zhang, Y., Bennell, K. L., White, D. K., Wei, J., Wu, Z., He, H., Liu, S., Luo, X., Hu, S., et al. (2020) · Journal of Medical Internet Research · doi:10.2196/21632 · Cited by 64 · see citing works
Actigraphically measured psychomotor slowing in depression: systematic review and meta-analysis
Wüthrich, F., Nabb, C. B., Mittal, V. A., Shankman, S. A., & Walther, S. (2022) · Psychological Medicine · doi:10.1017/s0033291722000903 · Cited by 60 · see citing works
Chandrasekaran, R., Muhammed Sadiq, T., & Moustakas, E. (2025) · Journal of Medical Internet Research · doi:10.2196/63879 · Cited by 59 · see citing works
Golbus, J. R., Pescatore, N. A., Nallamothu, B. K., Shah, N., & Kheterpal, S. (2021) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00138-2 · Cited by 57 · see citing works
Brahmbhatt, D. H., Ross, H. J., & Moayedi, Y. (2021) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2021.11.014 · Cited by 56 · see citing works
Polar Vantage and Oura Physical Activity and Sleep Trackers: Validation and Comparison Study
Henriksen, A., Svartdal, F., Grimsgaard, S., Hartvigsen, G., & Hopstock, L. A. (2022) · JMIR Formative Research · doi:10.2196/27248 · Cited by 54 · see citing works
Willingham, T. B., Stowell, J. F., Collier, G., & Backus, D. A. (2024) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph21010079 · Cited by 54 · see citing works
Christensen, J. C., Blackburn, B. E., Anderson, L. A., Gililland, J. M., Peters, C. L., Archibeck, M. J., & Pelt, C. E. (2023) · The Journal of Arthroplasty · doi:10.1016/j.arth.2023.03.060 · Cited by 53 · see citing works
Portz, J. D., Moore, S. L., & Bull, S. S. (2024) · Journal of Medical Internet Research · doi:10.2196/62790 · Cited by 53 · see citing works
Akbar, F., Mark, G. J., Prausnitz, S. R., Warton, E. M., East, J. A., Moeller, M. F., Reed, M. E., & Lieu, T. A. (2021) · JMIR Medical Informatics · doi:10.2196/24014 · Cited by 52 · see citing works
Employing of machine learning and wearable devices in healthcare system: tasks and challenges
Saad, H. S., Zaki, J. F. W., & Abdelsalam, M. M. (2024) · Neural Computing and Applications · doi:10.1007/s00521-024-10197-z · Cited by 51 · see citing works
Intelligence-Based Spine Care Model: A New Era of Research and Clinical Decision-Making
Mallow, G. M., Siyaji, Z. K., Galbusera, F., Orías, A. A. E., Giers, M. B., Lundberg, H. J., Ames, C. P., Karppinen, J. I., Louie, P. K., Phillips, F. M., et al. (2020) · Global Spine Journal · doi:10.1177/2192568220973984 · Cited by 51 · see citing works
Molina‐García, P., Notbohm, H. L., Schumann, M., Argent, R., Hetherington‐Rauth, M., Stang, J. S., Bloch, W., Cheng, S., Ekelund, U., Sardinha, L. B., et al. (2022) · Sports Medicine · doi:10.1007/s40279-021-01639-y · Cited by 50 · see citing works
Argent, R., Hetherington‐Rauth, M., Stang, J. S., Tarp, J., Ortega, F. B., Molina‐García, P., Schumann, M., Bloch, W., Cheng, S., Grøntved, A., et al. (2022) · Sports Medicine · doi:10.1007/s40279-022-01665-4 · Cited by 49 · see citing works
Merrigan, J. J., Stovall, J. H., Stone, J. D., Stephenson, M., Finomore, V. S., & Hagen, J. A. (2022) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2022.2161820 · Cited by 49 · see citing works
Natalucci, V., Marmondi, F., Biraghi, M., & Bonato, M. (2023) · Nutrients · doi:10.3390/nu15040913 · Cited by 48 · see citing works
Wearable Artificial Intelligence for Detecting Anxiety: Systematic Review and Meta-Analysis
Abd-Alrazaq, A. A., AlSaad, R., Harfouche, M., Aziz, S., Ahmed, A., Damseh, R., & Sheikh, J. I. (2023) · Journal of Medical Internet Research · doi:10.2196/48754 · Cited by 48 · see citing works
Wearable Healthcare and Continuous Vital Sign Monitoring with IoT Integration
Taherdoost, H. (2024) · Computers, Materials & Continua · doi:10.32604/cmc.2024.054378 · Cited by 47 · see citing works
Validity of the Wrist-Worn Polar Vantage V2 to Measure Heart Rate and Heart Rate Variability at Rest
Nuuttila, O., Korhonen, E., Laukkanen, J. A., & Kyröläinen, H. (2021) · Sensors · doi:10.3390/s22010137 · Cited by 47 · see citing works
Application of Exercise Snacks across Youth, Adult and Clinical Populations: A Scoping Review
Weston, K. L., Little, J. P., Weston, M., McCreary, S., Kitchin, V., Gill, A. J., Niven, A. G., McNarry, M. A., & Mackintosh, K. A. (2025) · Sports Medicine - Open · doi:10.1186/s40798-025-00829-6 · Cited by 48 · see citing works
Giurgiu, M., Timm, I., Becker, M. L., Schmidt, S., Wunsch, K., Nissen, R., Davidovski, D., Bussmann, J. B. J., Nigg, C. R., Reichert, M., et al. (2022) · JMIR mhealth and uhealth · doi:10.2196/36377 · Cited by 46 · see citing works
Validation of Oura ring energy expenditure and steps in laboratory and free-living
Kristiansson, E., Fridolfsson, J., Arvidsson, D., Holmäng, A. B., Börjesson, M., & Andersson-Hall, U. K. (2023) · BMC Medical Research Methodology · doi:10.1186/s12874-023-01868-x · Cited by 46 · see citing works
Wearables for Measuring Health Effects of Climate Change–Induced Weather Extremes: Scoping Review
Koch, M., Matzke, I., Huhn, S., Gunga, H., Maggioni, M. A., Munga, S., Obor, D., Sié, A., Boudo, V., Bunker, A., et al. (2022) · JMIR mhealth and uhealth · doi:10.2196/39532 · Cited by 45 · see citing works
Li, S., Qian, J., Liu, J., Xue, Y., Zhang, J., Liu, Y., Hu, X., Jing, X., & Zhang, W. (2023) · Mechanical Systems and Signal Processing · doi:10.1016/j.ymssp.2023.110679 · Cited by 43 · see citing works
Wearable photoplethysmography devices
Marozas, V., & Charlton, P. H. (2021) · Zenodo (CERN European Organization for Nuclear Research) · doi:10.5281/zenodo.4601547 · Cited by 43 · see citing works
Saito, T., Suzuki, H., & Kishi, A. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.861808 · Cited by 42 · see citing works
Predicting lying, sitting, walking and running using Apple Watch and Fitbit data
Fuller, D., Anaraki, J. R., Simango, B., Rayner, M., Dorani, F., Bozorgi, A., Luan, H., & Basset, F. A. (2021) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2020-001004 · Cited by 41 · see citing works
Self-citationDetection and Monitoring of Viral Infections via Wearable Devices and Biometric Data
Goergen, C. J., Tweardy, M. J., Steinhubl, S. R., Wegerich, S., Singh, K., Mieloszyk, R. J., & Dunn, J. (2021) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-103020-040136 · Cited by 41 · see citing works
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 40 · see citing works
Huhn, S., Matzke, I., Koch, M., Gunga, H., Maggioni, M. A., Sié, A., Boudo, V., Ouédraogo, W. A., Compaoré, G., Bunker, A., et al. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.972177 · Cited by 39 · see citing works
The Current and Future Role of Technology in Respiratory Care
Honkoop, P., Usmani, O. S., & Bonini, M. (2022) · Pulmonary Therapy · doi:10.1007/s41030-022-00191-y · Cited by 38 · see citing works
Sun, Y., Chen, J., Ji, M., & Li, X. (2025) · Journal of Medical Internet Research · doi:10.2196/69077 · Cited by 38 · see citing works
Jamieson, A., Chico, T., Jones, S. H., Chaturvedi, N., Hughes, A. D., & Orini, M. (2025) · npj Cardiovascular Health · doi:10.1038/s44325-025-00082-6 · Cited by 38 · see citing works
Assessing Physical Activity After ACL Injury: Moving Beyond Return to Sport
Kuenze, C. M., Collins, K. A., Pfeiffer, K. A., & Lisee, C. M. (2021) · Sports Health A Multidisciplinary Approach · doi:10.1177/19417381211025307 · Cited by 38 · see citing works
Romagnoli, S., Ripanti, F., Morettini, M., Burattini, L., & Sbrollini, A. (2023) · Sensors · doi:10.3390/s23063350 · Cited by 37 · see citing works
Giurgiu, M., Kolb, S., Nigg, C., Burchartz, A., Timm, I., Becker, M. L., Rulf, E., Doster, A.-K., Koch, E. D., Bussmann, J. B. J., et al. (2022) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2021-001267 · Cited by 36 · see citing works
Promoting child and adolescent health through wearable technology: A systematic review
Zhang, W., Xiong, K., Zhu, C., Evans, R. D., Zhou, L. M., & Podrini, C. (2024) · Digital Health · doi:10.1177/20552076241260507 · Cited by 36 · see citing works
Hurwitz, E. G., Butzin-Dozier, Z., Master, H., O’Neil, S. T., Walden, A., Holko, M., Patel, R. C., & Haendel, M. A. (2024) · JMIR mhealth and uhealth · doi:10.2196/54622 · Cited by 35 · see citing works
Ash, G. I., Stults-Kolehmainen, M., Busa, M. A., Gaffey, A. E., Angeloudis, K., Muniz-Pardos, B., Gregory, R., Huggins, R. A., Redeker, N. S., Weinzimer, S. A., et al. (2021) · Sports Medicine · doi:10.1007/s40279-021-01543-5 · Cited by 30 · see citing works
Dudarev, V., Barral, O., Zhang, C., Davis, G., & Enns, J. T. (2023) · Sensors · doi:10.3390/s23135863 · Cited by 35 · see citing works
More Physical Activity after Concussion Is Associated with Faster Return to Play among Adolescents
Seehusen, C. N., Wilson, J. C., Walker, G. A., Reinking, S. E., & Howell, D. R. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18147373 · Cited by 33 · see citing works
Ni, Z., Wu, T., Wang, T., Sun, F., & Li, Y. (2022) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2022.3163429 · Cited by 33 · see citing works
Challenges and Opportunities for Applying Wearable Technology to Sleep
Tobin, S. Y., Williams, P., Baron, K. G., Halliday, T. M., & Depner, C. M. (2021) · Sleep Medicine Clinics · doi:10.1016/j.jsmc.2021.07.002 · Cited by 32 · see citing works
Femiano, R., Werner, C., Wilhelm, M., & Eser, P. (2022) · Gait & Posture · doi:10.1016/j.gaitpost.2021.11.035 · Cited by 31 · see citing works
Siddi, S., Bailon, R., Giné-Vázquez, I., Matcham, F., Lamers, F., Kontaxis, S., Laporta, E., Garcia, E., Lombardini, F., Annas, P., et al. (2023) · Psychological Medicine · doi:10.1017/s0033291723001034 · Cited by 31 · see citing works
Open-source FlexNIRS: A low-cost, wireless and wearable cerebral health tracker
Wu, K.-C., Tamborini, D., Renna, M., Peruch, A., Huang, Y., Martin, A., Kaya, K., Starkweather, Z., Zavriyev, A. I., Carp, S. A., et al. (2022) · NeuroImage · doi:10.1016/j.neuroimage.2022.119216 · Cited by 31 · see citing works
Wearable photoacoustic watch for humans
Zhang, T., Guo, H., Qi, W., & Xi, L. (2024) · Optics Letters · doi:10.1364/ol.514238 · Cited by 31 · see citing works
Roth, A. M., Tran, N. K., Cocchiaro, B. F., Mitchell, A. K., Schwartz, D. G., Hensel, D. J., Ataiants, J., Brenner, J., Yahav, I., & Lankenau, S. E. (2021) · Drug and Alcohol Dependence · doi:10.1016/j.drugalcdep.2021.109138 · Cited by 31 · see citing works
Kane, W. J., Hassinger, T. E., Myers, E. L., Chu, D. L., Charles, A. N., Hoang, S. C., Friel, C. M., Thiele, R. H., & Hedrick, T. L. (2021) · Surgical Endoscopy · doi:10.1007/s00464-021-08449-3 · Cited by 30 · see citing works
Strain, T., Wijndaele, K., Pearce, M., & Brage, S. (2022) · Journal for the Measurement of Physical Behaviour · doi:10.1123/jmpb.2021-0046 · Cited by 30 · see citing works
Viciana, J., Casado-Robles, C., Guijarro-Romero, S., & Mayorga-Vega, D. (2022) · Journal of Sports Science and Medicine · doi:10.52082/jssm.2022.356 · Cited by 30 · see citing works
Hu, X., Sgherza, T. R., Nothrup, J. B., Fresco, D. M., Naragon‐Gainey, K., & Bylsma, L. M. (2024) · Behavior Research Methods · doi:10.3758/s13428-024-02387-3 · Cited by 29 · see citing works
Howell, D. R., Wingerson, M. J., Kirkwood, M. W., Grubenhoff, J. A., & Wilson, J. C. (2022) · Physical Therapy in Sport · doi:10.1016/j.ptsp.2022.04.010 · Cited by 29 · see citing works
Wearable Smart Bandage-Based Bio-Sensors
Levin, A., Gong, S. P., & Cheng, W. (2023) · Biosensors · doi:10.3390/bios13040462 · Cited by 29 · see citing works
Wattanapisit, A., Ng, C. J., Angkurawaranon, C., Wattanapisit, S., Chaovalit, S., & Stoutenberg, M. (2022) · Heliyon · doi:10.1016/j.heliyon.2022.e11259 · Cited by 28 · see citing works
Belabbaci, N. A., Anaadumba, R., & Alam, M. A. U. (2025) · JMIR mhealth and uhealth · doi:10.2196/60569 · Cited by 28 · see citing works
Ha, Y., Lee, S.-H., Lee, D., Kang, Y.-H., Choi, W.-J., & An, J. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph19021014 · Cited by 28 · see citing works
Greenberg, J. K., Javeed, S., Zhang, J. K., Benedict, B. C., Frumkin, M. R., Xu, Z., Zhang, J., Rodebaugh, T. L., Lu, C., Piccirillo, J. F., et al. (2023) · Journal of Neurosurgery Spine · doi:10.3171/2022.12.spine221302 · Cited by 28 · see citing works
Falz, R., Bischoff, C., Thieme, R., Tegtbur, U., Hillemanns, P., Stolzenburg, J., Aktas, B., Bork, U., Weitz, J., Lässing, J., et al. (2023) · BMC Medicine · doi:10.1186/s12916-023-03010-6 · Cited by 28 · see citing works
Pini, N., Ong, J. L., Yılmaz, G., Chee, N. I. Y. N., Siting, Z., Awasthi, A., Biju, S., Kishan, K., Patanaik, A., Fifer, W. P., et al. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.974192 · Cited by 27 · see citing works
Giurgiu, M., Ketelhut, S., Kubica, C., Nissen, R., Doster, A.-K., Thron, M., Timm, I., Giurgiu, V., Nigg, C. R., Wöll, A., et al. (2023) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-023-01473-7 · Cited by 27 · see citing works
Henriksen, A., Johannessen, E., Hartvigsen, G., Grimsgaard, S., & Hopstock, L. A. (2021) · JMIR Public Health and Surveillance · doi:10.2196/23806 · Cited by 27 · see citing works
Early adverse physiological event detection using commercial wearables: challenges and opportunities
Phipps, J. F., Passage, B., Sel, K., Martinez, J., Saadat, M., Koker, T. E., Damaso, N., Davis, S. K., Palmer, J., Claypool, K. T., et al. (2024) · npj Digital Medicine · doi:10.1038/s41746-024-01129-1 · Cited by 26 · see citing works
Ravindran, K. K. G., della Monica, C., Atzori, G., Lambert, D., Hassanin, H., Revell, V. L., & Dijk, D. (2024) · JMIR mhealth and uhealth · doi:10.2196/53643 · Cited by 25 · see citing works
Větrovský, T., Král, N., Pfeiferová, M., Kühnová, J., Novák, J., Wahlich, C., Jaklova, A., Jurková, K., Janek, M., Omcirk, D., et al. (2023) · BMC Public Health · doi:10.1186/s12889-023-15513-1 · Cited by 25 · see citing works
Casado-Robles, C., Mayorga-Vega, D., Guijarro-Romero, S., & Viciana, J. (2023) · Journal of Sports Science and Medicine · doi:10.52082/jssm.2023.196 · Cited by 25 · see citing works
Schiefer, L. M., Treff, G., Treff, F., Schmidt, P., Schäfer, L., Niebauer, J., Swenson, K. E., Swenson, K. E., Berger, M. M., & Sareban, M. (2021) · Sensors · doi:10.3390/s21196363 · Cited by 24 · see citing works
Wang, J., Fang, Y., Frank, E., Walton, M. A., Burmeister, M., Tewari, A., Dempsey, W., NeCamp, T., Sen, S., & Wu, Z. (2023) · npj Digital Medicine · doi:10.1038/s41746-022-00746-y · Cited by 24 · see citing works
White, J. W., Finnegan, O., Tindall, N., Nelakuditi, S., Brown, D. E., Pate, R. R., Welk, G. J., de Zambotti, M., Ghosal, R., Wang, Y., et al. (2024) · PLoS ONE · doi:10.1371/journal.pone.0286898 · Cited by 24 · see citing works
Hermans, F., Arents, E., Blondeel, A., Janssens, W., Cardinaels, N., Calders, P., Troosters, T., Derom, E., & Demeyer, H. (2024) · JMIR mhealth and uhealth · doi:10.2196/56027 · Cited by 23 · see citing works
- Authors
- Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., Slade, L., & Taylor, N. G. A.
- Affiliations
- Memorial University of Newfoundland (CA)
- University of Oregon (US)
- Dalhousie University (CA)
- Corresponding author
- Fuller, Daniel
- ORCID
- Venue
- JMIR mHealth and uHealth
- Publisher
- JMIR Publications Inc.
- Published
- 2020-09-08
- Volume
- 8
- Issue
- 9
- Pages
- e18694
- ISSN
- 2291-5222 (ISSN-L)
- PMC
- PMC7509623
- OpenAlex
- W3037734338
- Cited by
- 715
- References
- 180
- Citations by year
- 2026: 130 · 2025: 155 · 2024: 153 · 2023: 116 · 2022: 101 · 2021: 55 · 2020: 5
- FWCI
- 8.82
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- JMIR mhealth and uhealth (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Memorial University Research Repository (Memorial University) (acceptedVersion)
- HARVEST (University of Saskatchewan) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Memorial University Research Repository (Memorial University) (submittedVersion)
- Topics
- Health Sciences › Medicine › Physiology › Physical Activity and Health
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Keywords
- wearable devices, wearable activity trackers, step counting, heart rate measurement, energy expenditure, device validity, device reliability, Fitbit, Apple Watch, Garmin, consumer health technology, inter-device agreement
- MeSH terms
- Fitness Trackers, Wearable Electronic Devices, Energy Metabolism, Heart Rate, Humans, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review