Reliability and Validity of the Lactate Pro 2 Analyzer
Open accessThe DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L−1 for blood lactate (BLa) ≤4.0 mmol.L−1 to −1.6 to +1.4 mmol.L−1 for BLa >8.0 mmol.L−1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa ≤ 8 mmol.L−1, and using the LP2 to identify training intensities at lactate threshold.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The lactate analyzer (Lactate Pro 2) was verified as reliable in research using a graded incremental rowing test, comparing it with a laboratory-based analyzer (Crotty et al., 2021).”
From The relationship of muscle oxygen saturation analyzer with other monitoring and quantification tools in a maximal incremental treadmill test (2023) · cites this paper as Crotty et al., 2021
In the citing paper
“which has been shown to have strong intra-device reliability with intraclass correlation coefficients (ICC) = 0.99, with a coefficient of variation (CV) = 3.3% and 95% limits of agreement (LoA) between −1.0 and 1.0 mmol L-1, making it ideal for testing in large team scenarios (Crotty et al., 2021).”
From Monitoring skeletal muscle oxygen saturation kinetics during graded exercise testing in NCAA division I female rowers (2025) · cites this paper as Crotty et al., 2021
In the citing paper
“Previous studies showed that Lactate Pro 2 presented a good reliability compared with blood samples (Crotty et al., 2021; Raa et al., 2020), with higher agreement if the measurement is performed in the earlobe than in the fingertip (Raa et al., 2020), and with Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L -1 to +1.4 mmol.L -1 for concentrations lower than 4.0 mmol.L -1 and higher”
From A comparative analysis of mathematical methods for detecting lactate thresholds using muscle oxygenation data during a graded cycling test (2023) · cites this paper as Crotty et al., 2021
Citations
Cites
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 31,709
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532 · cited 1×
“The ICC (3,1) was used in the current study according to the guidelines of Shrout and Fleiss (1979) and as described by Koo and Li (2016).”
MethodMeasuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
Atkinson, G., & Nevill, A. M. (1998) · Sports Medicine · doi:10.2165/00007256-199826040-00002 · Cited by 3,779
JOURNAL OF SPORTS SCIENCE & MEDICINE
Kallinen, M. (2005) · Cited by 1,481
Lactate Threshold Concepts: How Valid are They?
Faude, O., Kindermann, W., & Meyer, T. (2009) · Sports Medicine · doi:10.2165/00007256-200939060-00003 · Cited by 882 · cited 2×
“L−1 (Faude et al., 2009; Goodwin et al., 2007).”
(2022) · Journal of Sports Medicine · doi:10.1155/5363 · Cited by 416
Doglikuu (2021) · doi:10.21315/mjms · Cited by 406
Bentley, D. J., Newell, J., & Bishop, D. (2007) · Sports Medicine · doi:10.2165/00007256-200737070-00002 · Cited by 404
A New Approach for the Determination of Ventilatory and Lactate Thresholds
Cheng, B., Kuipers, H., Snyder, A., Keizer, H., Jeukendrup, A., & Hesselink, M. (1992) · International Journal of Sports Medicine · doi:10.1055/s-2007-1021309 · Cited by 371
Research works printed but not linked: 5
- Buckley J. D., 2003, Journal of Science and Medicine, 6, 408
- Ghosh A. K., 2004, Malaysian Journal of Medical Sciences, 11, 24
- Goodwin M. L., 2007, Journal of Diabetes, 1, 558
- Mc Grath E., 2019, International Journal of Exercise Science, 12, 1334
- Rowe G. S., 2016, Gazzetta Medica Italiana Archivio per le Scienze Mediche, 175, 378
Cited by
Mentzoni, F., Skaugen, M., Eythorsdottir, I., Roterud, S., JOHANSEN, E. S., & Losnegard, T. (2024) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05572-6 · Cited by 33
Maré, L. D., Boshuizen, B., de Vega, C. V. M., meeûs, C. D., Plancke, L., Gansemans, Y., Nieuwerburgh, F. V., Deforce, D. L. D., de Oliveira, J. E., Hosotani, G., et al. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.792052 · Cited by 26
Conte, D., Lukonaitienė, I., Matulaitis, K., Sniečkus, A., Kniubaitė, A., Kreivytė, R., & Kamandulis, S. (2023) · Biology of Sport · doi:10.5114/biolsport.2023.122478 · Cited by 24
Spitz, S., Schobesberger, S., Brandauer, K., & Ertl, P. (2023) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10604 · Cited by 20
Takei, N., Inaba, T., Morita, Y., Kakinoki, K., Hatta, H., & Kitaoka, Y. (2024) · Temperature · doi:10.1080/23328940.2024.2375693 · Cited by 19
“%) compared to a reliable laboratory-based analyzer [24].”
Martínez‐Ferrán, M., Cuadrado‐Peñafiel, V., Sánchez-Andreo, J. M., Villar-Lucas, M., Montealegre, M. C., Rubio-Martín, A., Morales, C. R., Casla-Barrio, S., & Pareja-Galeano, H. (2022) · Nutrients · doi:10.3390/nu14214635 · Cited by 20
“Lactate Lactate concentrations were measured in a capillary blood sample obtained from participants’ fingertips using a portable analyzer (Lactate 2, Arkray, Kyoto, Japan) [31,32].”
MethodCabre, H. E., Ladan, A. N., Moore, S. R., Joniak, K. E., Blue, M. N. M., Pietrosimone, B. G., Hackney, A. C., & Smith‐Ryan, A. E. (2024) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000004764 · Cited by 14
Zhong, F., Chen, Z., Gu, Z., Wang, X., Holmberg, H., & Li, Y. (2024) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05585-1 · Cited by 14
“Since these factors play key roles in planning personal training in a manner that enhances performance optimally, determination of Bla must be as accurate as possible (Crotty et al. 2021; Tanner et al. 2010).”
Douligeris, A., Methenitis, S. K., Lazou, A., Panayiotou, G., Feidantsis, K. G., Voulgaridou, G., Μanios, Y., Jamurtas, A. Z., Giaginis, C., & Papadopoulou, S. Κ. (2023) · Nutrients · doi:10.3390/nu15102304 · Cited by 13
“Blood lactate concentration (BL) (mmol/L) was measured using a portable blood lactate analyzer (Lactate Pro 2, LT-1730; Arkray, Kyoto, Japan) from a finger prick blood sample (0.3 mL) three minutes after the end of the Yo-Yo IRL1 VO 2max test [52].”
MethodInfluentialBurger, D., Vidondo, B., Gerber, V., Deillon, D., Müller, A., Scheidegger, M. D., Käser, R., & Ramseyer, A. A. (2023) · Equine Veterinary Journal · doi:10.1111/evj.13999 · Cited by 13
Daly, L. S., Catháin, C. Ó., & Kelly, D. T. (2023) · Biology of Sport · doi:10.5114/biolsport.2024.129479 · Cited by 12
Boland, M., Crotty, N. M., Mahony, N., Donne, B., & Fleming, N. (2022) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2021-0090 · Cited by 12
Self-citationFerioli, D., Alcaráz, P. E., Freitas, T. T., Trimarchi, F., Conte, D., Formica, L., Chung, L. H., & Scanlan, A. T. (2024) · Frontiers in Psychology · doi:10.3389/fpsyg.2024.1414339 · Cited by 10
Sendra-Pérez, C., Encarnación-Martínez, A., Oficial-Casado, F., Salvador-Palmer, R., & Priego-Quesada, J. I. (2023) · Physiological Measurement · doi:10.1088/1361-6579/ad1457 · Cited by 9
Osmani, F., Lago-Fuentes, C., Alemany-Iturriaga, J., & Barcala-Furelos, M. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1155037 · Cited by 8
“The lactate analyzer (Lactate Pro 2) was verified as reliable in research using a graded incremental rowing test, comparing it with a laboratory-based analyzer (Crotty et al., 2021).”
MethodJiménez‐Martínez, P., Alix‐Fages, C., Janicijevic, D. N., Miras‐Moreno, S., Chacón‐Ventura, S., Martín-Olmedo, J. J., de la Cruz-Márquez, J. C., Osuna‐Prieto, F. J., Jurado‐Fasoli, L., Amaro‐Gahete, F. J., et al. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1190345 · Cited by 8
“The dependent variables considered in this test were: i) final stage completed in the protocol, ii) VO2max, iii) circulating lactate concentration recorded 90 s after test cessation with a portable lactate analyzer (Lactate PRO2, Arkray, Kyoto, Japan) (Crotty et al., 2021), iv) general (RPEg) and quadriceps (RPEquad)…”
Methodda Cruz, C. J. G., Porto, L. G. G., & Molina, G. E. (2021) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2021.1979980 · Cited by 8
“Due to anaerobic threshold determination’s clinical and functional relevance, the validity and reliability of some methods and devices used to identify this metabolic phenomenon were previously investigated in the literature (Bonafiglia et al., 2018; Crotty et al., 2021; Hailstone & Kilding, 2011).”
MethodEserhaut, D. A., DeLeo, J. M., Provost, J. A., Ackerman, K. E., & Fry, A. C. (2025) · Frontiers in Physiology · doi:10.3389/fphys.2025.1538465 · Cited by 7
Turner, O., Chessor, R. J., & Mitchell, N. (2025) · Performance Nutrition · doi:10.1186/s44410-025-00001-x · Cited by 7
“…Lactate Meter has been shown to have high reliability (intraclass correlation coefficient [ICC] = 0.99), although displays poor construct validity in comparison to a calibrated laboratory-based analyser at high intensities (intraclass correlation coefficient at 4.1–8.0 mmol ⋅ L − 1 = 0.30) [58].”
InfluentialWhy Train Together When Racing Is Performed Alone? Drafting in Long-Track Speed Skating
van den Brandt, F. A. P., Stoter, I. K., Otter, R., & Elferink-Gemser, M. T. (2021) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2020-0913 · Cited by 7
Mihăilescu, L. N., Chiriac, P. B., Mihailescu, L. E., Manolachi, V., & Potop, V. (2023) · PeerJ · doi:10.7717/peerj.15477 · Cited by 7
“LAC was evaluated immediately after exercise in each individual athlete, according to the test performed (Crotty et al., 2021).”
MethodClague-Baker, N., Tyson, S. F., Leslie, K., Dawes, H., Bull, M., & Hilliard, N. (2023) · Fatigue Biomedicine Health & Behavior · doi:10.1080/21641846.2023.2245584 · Cited by 7
“Compared to a laboratory-based analyser it was found to be valid [28].”
MethodSalem, A., Ammar, A., Kerkeni, M., Boujelbane, M. A., Uyar, A. M., Köbel, L. M., Selvaraj, S. A., Zare, R., Heinrich, K. M., Jahrami, H., et al. (2025) · Biology of Sport · doi:10.5114/biolsport.2025.151658 · Cited by 6
“The Lactate Pro 2 device used for lactate assessment has a reported coefficient of variation (CV) of approximately 3%, indicating good measurement precision [58].”
Hidouri, S., Driss, T., Tagougui, S., Kammoun, N., Chtourou, H., & Hammouda, O. (2024) · Sensors · doi:10.3390/s24030909 · Cited by 6
“%) between 3 and 3.6% [42].”
InfluentialAshcroft, S. K., Basclain, K., Woolnough, C., Hoon, M. W., Walsh, S. J., Starc, L. C., Johnson, L. G., Kuys, S. S., & Thompson-Butel, A. G. (2025) · Journal of Neuroscience Research · doi:10.1002/jnr.70058 · Cited by 5
“The measurement of lactate concentration is also easier than BDNF concentration in clinical practice, requiring a smaller blood sample, and is within the scope of practice of health professionals delivering exercise interventions (Bayles et al. 2023; Crotty et al. 2021).”
Golluceli, B., Kamandulis, S., Pernigoni, M., Lukonaitienė, I., Kreivytė, R., & Conte, D. (2025) · Biology of Sport · doi:10.5114/biolsport.2025.148547 · Cited by 4
“The Lactate Pro 2 CT-1730 analyser (Arkray Co., Kyoto, Japan) was used to exam - ine blood samples [27].”
Faria, V. S., & Egan, B. (2024) · European Journal of Sport Science · doi:10.1002/ejsc.12090 · Cited by 4
“The Lactate Pro2 analyzer has a manufacturer ‐ reported CV of ~3%, which is in broad agreement with published studies that is, 2.7%–4.3% at concentrations between 5.0 and 14.9 mM (Bonaventura et al., 2015) and 3.3% at concentrations > 8.0 mM (Crotty et al., 2021).”
BackgroundInfluentialChua, M. T., Sim, A., & Burns, S. F. (2025) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2024-0423 · Cited by 4
Eserhaut, D. A., DeLeo, J. M., Provost, J. A., & Fry, A. C. (2025) · Physiological Reports · doi:10.14814/phy2.70455 · Cited by 3
“Immediately following sample acquisition, untreated whole blood was analyzed for BLa concentrations using the Lactate Pro 2 handheld analyzer (Arkay, Kyoto, Japan), which possesses strong intra-device reliability (Crotty et al., 2021).”
Bruno, J., Thapa, R. K., Santos, Â., Montoro-Bombú, R., & Sarmento, H. (2026) · Sports Medicine - Open · doi:10.1186/s40798-026-00979-1 · Cited by 3
Rodríguez-Muñoz, A., Martínez‐Montoro, J. I., Sojo-Rodriguez, B., Benítez‐Porres, J., de Albornoz-Gil, M. C., Carrasco-Fernández, L., Subiri-Verdugo, A., Molina-Ramos, A. I., Cobos-Díaz, A., Tinahones, F. J., et al. (2024) · Journal of the International Society of Sports Nutrition · doi:10.1080/15502783.2024.2433740 · Cited by 3
Myers, N. D., Lee, S., Chun, H., & Silverman, S. (2022) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2022.2071126 · Cited by 3
“…in 2022, we identify an example of a high-quality manuscript recently published in MPEES that is consistent with each section of the journal: ● Crotty et al. (2021) for the exercise science section; ● Eddy et al. (2021) for the physical education pedagogy section; ● Crotti et al. (2021) for…”
Inhibitory effects of NaF on mitochondrial energy generation in human platelets in vitro
Tsujino, T., Kasahara, T., Kawabata, H., Watanabe, T., Nishiyama, K., Kitamura, Y., Watanabe, T., Okudera, H., Mochizuki, T., Ushiki, T., et al. (2024) · Frontiers in Toxicology · doi:10.3389/ftox.2024.1421184 · Cited by 2
“Lactate levels in extracellular plasma were determined using a lactate meter that employs speci fi c test strips based on the lactate oxidase enzyme electrode method (Lactate Pro2; Arkray, Kyoto, Japan) (Crotty et al., 2021).”
Methoddel‐Cuerpo, I., Jerez‐Mayorga, D., Ríos, L. J. C., Navarrete, F. C., & Delgado‐Floody, P. (2025) · PeerJ · doi:10.7717/peerj.18215 · Cited by 2
Acute alterations in blood lactate in the setting of transient stress induced myocardial ischaemia
O’Driscoll, J. M., Smith, E., Bibat, M., Edwards, J. J., Compton, C., Kipourou, K., Coleman, D. A., Wiles, J. D., Cunliffe, E., Marciniak, A., et al. (2025) · Experimental Physiology · doi:10.1113/ep092429 · Cited by 1
“Previous work has reportedhighreliabilityoftheLactatePro2analyserwithacoefficient of variation of 3.3% and Bland–Altman 95% limits of agreement of Crotty et al., 2021).”
InfluentialMeixner, B. J., Matzka, M., Renner, P., Achtzehn, S., & Sperlich, B. (2026) · Biology Open · doi:10.1242/bio.062543 · Cited by 1
Kishalı, N. F., Kıyıcı, F., Yılmaz, H. H., Özbay, S., Gençoğlu, C., Öget, F., Hoos, O., Kaufmann, S., Atasever, G., SEREN, K., et al. (2024) · Research Square · doi:10.21203/rs.3.rs-4666482/v1 · Cited by 1
Matzka, M., Lauber, L., Lenk, M., Engel, F. A., & Sperlich, B. (2025) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000003677
Gehlert, S., Khurshudyan, A., Weber, S., Widdershoven, J., & Schuylenbergh, R. V. (2025) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2025.1546117
Optimizing Player Health, Recovery, and Performance in Basketball, volume II
Ferioli, D. (2025) · Frontiers research topics · doi:10.3389/978-2-8325-6145-4
Etiology and recovery of knee extensor muscle fatigue following simulated basketball match-play
Ferioli, D., Freitas, T. T., Mannucci, C., Chung, L. H., Mombelli, A., Alcaráz, P. E., & Maffiuletti, N. A. (2025) · Biology of Sport · doi:10.5114/biolsport.2025.148546
“At the end of each quarter, blood lactate concentration (Blact) was measured from the earlobe immediately after the completion of the last circuit using a portable amperometric microvolume lactate analyzer with support - ed validity and reliability [21] (Lactate Pro 2, Arkray, Kyoto, Japan).”
Mochizuki, T., Ushiki, T., Koga, H., Suzuki, K., Sato, M., Osawa, M., Kamimura, M., Ishiguro, H., Suwabe, T., & Kawase, T. (2025) · Health Science Reports · doi:10.1002/hsr2.70784
“The lactate levels in extracellular plasma were determined using a lactate meter with specific test strips based on the lactate oxidase enzyme electrode method (Lactate Pro2; Arkray, Kyoto, Japan) [10].”
Mason, A., & Korostynska, O. (2025) · IEEE Microwave Magazine · doi:10.1109/mmm.2024.3476808
“Notably, these devices are not as accurate as lab-oratory-based analyzers [14], [15] and cannot provide continuous lactate measurements because of their invasive nature.”
Willman, H., & Hytönen, L. (2023) · DiVA (Digitala Vetenskapliga Arkivet)
How pre-exercise glucose affects lactate accumulation and perceived exertion in recreational runners
Johansson, T., & Rebecka, S. (2024) · Hogskolan Ihalmstad (Halmstad University)
Study of the local oxygen saturation in cycling and differences between sexes
Sendra‐Pérez, C. (2025) · Repository of Digital Objects for Teaching Research and Culture (University of Valencia)
Newbury, J. W. (2023) · BCU Open Access Repository (Birmingham City University)
Training prescription and monitoring in rowing: Coaches' perspectives and current practices
Watts, S. (2025) · UWA Profiles and Research Repository (University of Western Australia) · doi:10.26182/a9r2-7v21
Suzuki, Y., Takei, N., Tanji, F., Yamanaka, R., & Saeki, T. (2026) · Scandinavian Journal of Medicine and Science in Sports · doi:10.1111/sms.70286
Kafrawi, F. R., Wahyudi, H., Ningsih, Y. F., Boby, F. A., Rusdiawan, A., Irmawati, F., Dimyati, A., & Pramono, M. (2026) · Pedagogy of Physical Culture and Sports · doi:10.15561/26649837.2026.0308
“Finger-prick samples were collected using sterile lancets, and analysis was performed immediately according to manufacturer guidelines [19].”
Assessment of blood lactate measurement accuracy in athletes using three portable analyzers
Dikunets, M. A., Dudko, G. A., Arkhipkin, A. A., Kryuchkov, A. S., & Геньевич, А. А. (2026) · Sports medicine research and practice · doi:10.47529/2223-2524.2026.1.7
Beram, S., & Stručić, D. (2026) · Hrvatski športskomedicinski vjesnik · doi:10.69589/hsv.41.1.12
Shekhawat, D. S., Kumar, S., Thapa, R. K., Robustelli, A., Choudhary, S., Tomar, T., & Weldon, A. (2026) · Journal of Physical Education and Sport · doi:10.7752/jpes.2026.06121
Aldhahli, A., Amara, S., Al-Hadabi, B., & Mkaouer, B. (2026) · Balneo and PRM Research Journal · doi:10.12680/balneo.2026.1048
The FITP: A Controlled Intermittent Running Test for Futsal Athletes
Ribeiro, J. N., Yousefian, F., Bartolomeu, R. F., Martins, F. S. B., Spyrou, K., Travassos, B. F. R., Vila-Chã, C. J. F., & Esteves, P. (2026) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk11030296
“This device has been previously validated and shown to be reliable for these measures [35].”
Kolumbet, A. N., Yeltsov, D., Пивоваров, А. А., & Ногас, А. О. (2026) · Реабілітаційні та фізкультурно-рекреаційні аспекти розвитку людини (Rehabilitation & recreation) · doi:10.67034/2786-8354.2026.20.2.14
“It can be assessed through analysis of the structure of functional responses - for example, the predominance of increases in tidal and stroke volumes, along with a reduction in respiratory rate and heart rate under conditions of physical, hypercapnic load [3, 29].”
Čabarkapa, D., Batra, A., Johnson, Q. R., Cabarkapa, D. V., Wetmore, A. B., & Munce, T. A. (2026) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2026.1923398
“In addition to CMJ testing, finger-stick blood lactate samples were collected using a Lactate Pro device with single-use disposable lancets (Arkray, Kyoto, Japan), which has been shown to demonstrate strong intra-device reliability (22).”
Metadata
- Authors
- Crotty, N. M., Boland, M., Mahony, N., Donne, B., & Fleming, N.
- Affiliations
- Trinity College Dublin (IE)
- Corresponding author
- Crotty, Nora May
- Venue
- Measurement in Physical Education and Exercise Science
- Publisher
- Informa UK Limited
- Published
- 2021-01-25
- Volume
- 25
- Issue
- 3
- Pages
- 202-211
- ISSN
- 1091-367X (ISSN-L)
- 1532-7841
- OpenAlex
- W3122543005
- Semantic Scholar
- 40d45e26c798723ea7b3b0a9fdefeb406bb0ee5a
- Cited by
- 50
- References
- 19
- Citations by year
- 2026: 8 · 2025: 11 · 2024: 10 · 2023: 15 · 2022: 4 · 2021: 2
- FWCI
- 7.65
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- cc-by-nc-sa
- Locations
- Measurement in Physical Education and Exercise Science (publishedVersion)
- Trinity's Access to Research Output (TARA) (Trinity College Dublin) (submittedVersion, cc-by-nc-sa)
- Funders
- Trinity College Dublin
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Sports Performance and Training
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Health Sciences › Health Professions › Physical Therapy, Sports Therapy and Rehabilitation › Physical Education and Training Studies
- Keywords
- blood lactate measurement, measurement reliability, measurement validity, lactate threshold, oxygen consumption, heart rate, exercise intensity
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.