Acetaminophen interference with Nova StatStrip® Glucose Meter: case report with bench top confirmation
The DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
Context: Point-of-care glucose meters are an integral part in the assessment of patients with altered mental status. For this reason, glucose meters are checked for interference from commonly encountered substances, including acetaminophen. The Nova StatStrip® glucose meter has previously been reported to be resistant to interference. We report a case of a very high acetaminophen concentration causing interference with this point-of-care glucose meter.Case report: A 25-year-old female presented after an overdose of acetaminophen and diphenhydramine combination product. Patient was minimally responsive, so a point-of-care glucose check was attempted using the Nova StatStrip® glucose meter. Five different meters were attempted, and each showed an error message. Laboratory analysis using Beckman Coulter® Unicel DxC 800 revealed a glucose of 180 mg/dL and an acetaminophen concentration of 465 mg/L. Serum spiked with acetaminophen at different concentrations revealed interference with the Nova StatStrip® glucose meter at a concentration of 399 mg/L and above. To our knowledge, this interference with the Nova StatStrip® glucose meter has not been reported in the medical literature.Conclusion: Very high levels of acetaminophen can interfere with point-of-care glucose meters, even those that have previously been reported to be robust such as the Nova StatStrip® glucose meter. Clinicians should be aware of this possible interference when treating patients with acetaminophen overdose.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
null (2016) · Philosophy Study · doi:10.17265/2159-5313/2016.09.003 · Cited by 28,317
Gummin, D. D., Mowry, J. B., Spyker, D. A., Brooks, D. E., Osterthaler, K. M., & Banner, W. (2018) · Clinical Toxicology · doi:10.1080/15563650.2018.1533727 · Cited by 344
Emergency Department Visits for Overdoses of Acetaminophen-Containing Products
Budnitz, D. S., Lovegrove, M. C., & Crosby, A. E. (2011) · American Journal of Preventive Medicine · doi:10.1016/j.amepre.2011.02.026 · Cited by 178
Ceriotti, F., Kaczmarek, E., Guerra, E., Mastrantonio, F., Lucarelli, F., Valgimigli, F., & Mosca, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814563351 · Cited by 45
Performance of a new interference-resistant glucose meter
Vanavanan, S., Santanirand, P., Chaichanajarernkul, U., Chittamma, A., DuBois, J. A., Shirey, T., & Heinz, M. (2010) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.09.010 · Cited by 39
Klonoff, D. C., Umpierrez, G. E., & Rice, M. J. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818801607 · Cited by 26
Factors interfering with the accuracy of five blood glucose meters used in Chinese hospitals
Lv, H., Zhang, G., Kang, X., Yuan, H., Lv, Y., Wang, W., & Randall, R. (2013) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.21611 · Cited by 20
Tran, N. K., Godwin, Z. R., Steele, A. N., Wolf, S. E., & Palmieri, T. L. (2016) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000000877 · Cited by 18
Self-citationEvaluation of two glucose meters and interference corrections for screening neonatal hypoglycemia
Wada, Y., Nakamura, T., Kaneshige, M., Takahashi, S., Fujinaga, H., Tsukamoto, K., Ito, Y., & Sago, H. (2015) · Pediatrics International · doi:10.1111/ped.12543 · Cited by 12
Russell, W. J., Lee, C., & Milne, D. (2003) · Anaesthesia and Intensive Care · doi:10.1177/0310057x0303100317 · Cited by 6
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 208
Karami, T., Hoshyar, H., & Tavana, A. M. (2021) · International Journal of Surgery Open · doi:10.1016/j.ijso.2020.12.013 · Cited by 5
Erroneous Causes of Point-of-Care Glucose Readings
Majewski, J., Risler, Z. W., & Gupta, K. (2023) · Cureus · doi:10.7759/cureus.36356 · Cited by 5
“Laboratory analysis concluded a glucose level of 180 mg/dl [4].”
BackgroundBiosensor-Based Point-of-Care Devices: Metabolites and Pulse Oximetry
Hwang, I. M., Lou, X. A., Toubian, A. A., & Kamei, D. T. (2021) · Springer eBooks · doi:10.1007/978-981-16-2782-8_4 · Cited by 1
Metadata
- Authors
- Chenoweth, J. A., Dang, L. T., Gao, G., & Tran, N. K.
- Affiliations
- University of California Davis Medical Center (US)
- Corresponding author
- Chenoweth, James A.
- Venue
- Clinical Toxicology
- Publisher
- Informa UK Limited
- Published
- 2020-03-06
- Volume
- 58
- Issue
- 11
- Pages
- 1067-1070
- ISSN
- 0009-9309 (ISSN-L)
- 1556-3650
- 1556-9519
- OpenAlex
- W3009315972
- Cited by
- 10
- References
- 11
- Citations by year
- 2025: 3 · 2023: 4 · 2021: 2 · 2020: 1
- FWCI
- 1.05
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Toxicology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Toxicology › Forensic Toxicology and Drug Analysis
- Health Sciences › Medicine › Emergency Medicine › Poisoning and overdose treatments
- Keywords
- paracetamol, point-of-care glucose testing, assay interference, acetaminophen overdose, diphenhydramine, altered mental status
- MeSH terms
- Acetaminophen, Adult, Blood Glucose, Female, Humans, Reagent Strips, Point-of-Care Systems, Drug Overdose
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Acetaminophen interference with Nova StatStrip® Glucose Meter: case report with bench top confirmation. https://app.cochat.ai/papers/10.1080/15563650.2020.1732404
- The paper
- Chenoweth, J. A., Dang, L. T., Gao, G., & Tran, N. K. (2020). Acetaminophen interference with Nova StatStrip® Glucose Meter: case report with bench top confirmation. Clinical Toxicology. https://doi.org/10.1080/15563650.2020.1732404
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.