Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parameters
Abstract
In this study we analysed the accuracy of computed tomography (CT) measurements in assessing cortical bone. We determined the dependency of thickness and density measurements on the true width and density of the cortex and on the spatial resolution in the CT images using two optimized segmentation methods. As a secondary goal, we assessed the ability of CT to reflect small changes in cortical thickness. Two different bone-mimicking phantoms with varying cortical thickness were scanned with single-slice CT on a Somatom Plus 4 scanner. Images were reconstructed with both a standard and a high-resolution convolution kernel. Two special operator-independent segmentation methods were used to automatically detect the edges of the cortical shell. We measured cortical thickness and density and compared the phantom measurements with theoretical computations by simulating a cross-sectional shape of the cortical shell. Based on the simulations, we calculated CT's power to detect small changes in cortical thickness. Simulations and phantom measurements were in very good agreement. Cortical thickness could be measured with an error of less than 10% if the true thickness was larger than 0.9 (0.7) mm for the standard (high-resolution) kernel which is close to the full width at half maximum (FWHM) of the point spread functions for these kernels and our scanner. Density measurements yielded errors of less than 10% for true cortical thickness values above two to three times the FWHM corresponding to 2.5 (2) mm in our case. The simulations showed that a 10% change in cortical width would not be detected with satisfying probability in bones with a cortical shell thinner than 1.2 mm. An accurate determination of the cortical thickness is limited to bones with a thickness higher than the FWHM of the scanner's point spread function. Therefore, the use of a high-resolution reconstruction kernel is crucial. Cortical bone mineral density can only be measured accurately in bones two to three times thicker than this number. In thinner bones, the measured density becomes dependent on the thickness. Changes in cortical thickness can only be assessed if the change is rather large or if the measured bone has sufficient thickness. Therefore, assessing density or thickness of the vertebral shell by CT should be treated with caution.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Kalender, W. A., Felsenberg, D., Genant, H. K., Fischer, M., Dequeker, J., & Reeve, J. (1995) · European Journal of Radiology · doi:10.1016/0720-048x(95)00631-y · Cited by 263
Self-citation
Cited by
Cortical and Trabecular Bone Mineral Loss From the Spine and Hip in Long-Duration Spaceflight
Lang, T., LeBlanc, A., Evans, H., Lu, Y., Genant, H., & Yu, A. (2004) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.040307 · Cited by 828
Engelke, K., Adams, J. E., Armbrecht, G., Augat, P., Bogado, C. E., Bouxsein, M. L., Felsenberg, D., Ito, M., Prevrhal, S., Hans, D. B., et al. (2008) · Journal of Clinical Densitometry · doi:10.1016/j.jocd.2007.12.010 · Cited by 617
Self-citationQuantitative computed tomography
Adams, J. E. (2009) · European Journal of Radiology · doi:10.1016/j.ejrad.2009.04.074 · Cited by 458
Subject-Specific Finite Element Model of the Pelvis: Development, Validation and Sensitivity Studies
Anderson, A. E., Peters, C. L., Tuttle, B. D., & Weiss, J. A. (2005) · Journal of Biomechanical Engineering · doi:10.1115/1.1894148 · Cited by 347
Adaptation of the proximal femur to skeletal reloading after long-duration spaceflight
Lang, T. F., Leblanc, A. D., Evans, H. J., & Lu, Y. (2006) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.060509 · Cited by 302
Metadata
- Authors
- Prevrhal, S., Engelke, K., & Kalender, W. A.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 1999-01-01
- Volume
- 44
- Issue
- 3
- Pages
- 751-764
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2071503181
- Cited by
- 226
- References
- 1
- FWCI
- 3.15
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parameters. https://app.cochat.ai/papers/10.1088/0031-9155/44/3/017
- The paper
- Prevrhal, S., Engelke, K., & Kalender, W. A. (1999). Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parameters. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/44/3/017
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.