Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer
Abstract
With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode. The skin doses for seven prostate patients were measured on an IRB-approved protocol. TLD capsules were placed on the patient's skin at the central axis of three beams: AP, left lateral (Lt Lat) and right lateral (Rt Lat). To avoid the ring artefacts centred in the prostate, the treatment couch was dropped 3 cm from the patient's tattoo (central axis). The measured AP skin doses ranged 3-6 cGy for 20-33 cm separation. The larger the patient size the less the AP skin dose. Lateral doses did not change much with patient size. The Lt Lat dose was approximately 4.0 cGy, which was approximately 40% higher than the Rt Lat dose of approximately 2.6 cGy. To verify this dose asymmetry, surface doses on an IMRT QA phantom (oval shaped, 30 cm x 20 cm) were measured at the same three sites using TLD capsules with 3 cm table-drop. The dose asymmetry was due to: (1) kV source rotation which always starts from the patient's Lt Lat and ends at Lt Lat. Gantry rotation gets much slower near the end of rotation but dose rate stays constant and (2) 370 degrees scan rotation (10 degrees scan overlap on the Lt Lat side). In vivo doses were measured inside a Rando pelvic heterogeneous phantom using TLDs. The left hip (femoral head and neck) received the highest doses of approximately 10-11 cGy while the right hip received approximately 6-7 cGy. The surface and in vivo doses were also measured for phantoms at the central-axis setup. The difference was less than approximately 12% to the table-drop setup.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Compressed sensing based cone‐beam computed tomography reconstruction with a first‐order methoda)
Choi, K., Wang, J., Zhu, L., Suh, T., Boyd, S., & Xing, L. (2010) · Medical Physics · doi:10.1118/1.3481510 · Cited by 255
Noise correlation in CBCT projection data and its application for noise reduction in low‐dose CBCT
Zhang, H., Ouyang, L., Ma, J., Huang, J., Chen, W., & Wang, J. (2014) · Medical Physics · doi:10.1118/1.4865782 · Cited by 47
Metadata
- Authors
- Wen, N., Guan, H., Hammoud, R., Pradhan, D., Nurushev, T., Li, S., & Movsas, B.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2007-04-02
- Volume
- 52
- Issue
- 8
- Pages
- 2267-2276
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2027638943
- Cited by
- 173
- References
- 0
- FWCI
- 19.48
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer. https://app.cochat.ai/papers/10.1088/0031-9155/52/8/015
- The paper
- Wen, N., Guan, H., Hammoud, R., Pradhan, D., Nurushev, T., Li, S., & Movsas, B. (2007). Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/52/8/015
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.